Spectrum Analyzers
U3741/3751
Compact Design with High Performance
Pioneering 3 GHz/8 GHz Spectrum Analyzers are Now Available!
The U3741/3751 portable spectrum analyzer
supports a great range of applications, from
use on production lines to system installation
and maintenance. Its digital IF enables dramatic
improvements in power measurement accuracy
for digitally modulated signals. Moreover, the
U3741/3751 provides twice the throughput of its
predecessor. A light and compact 3 GHz/8 GHz
spectrum analyzer, the U3741/3751 provides basic performance reliably and at a low cost.
Better
measuring speed due to high-speed processing
(twice as fast as its predecessor)
Dramatically improved power measurement accuracy
for digitally modulated signals
Built-in 3 GHz/8 GHz pre-amp standard
Average display noise level:
-155 dBm/Hz@1 GHz, pre-amp ON
Tracking generator covering a frequency range of
100 kHz to 3 GHz/6 GHz
Option available for measurement of phase noise
characteristics
Lightweight and compact design, with a maximum
weight of only 5.6 kg
Continuous operation of up to 2.5 hours with the
battery pack
U3741/ 3751 Web Demonstration
Please access to the http://www.advantest.co.jp/en-index.shtml and click on the following links.
PRODUCTS & SUPPORT
2
U3741/3751-6E Mar. 13
Electronic Measuring Instruments
Products
U3751
Extension-key block
6.5-inch color liquid
crystal display (LCD)
Soft-key block
Exclusive function keys
Front Panel
Data Entry
Main function keys
RF INPUT
POWER
CAL OUT
PHONE
Earphone jack for
demodulated
AM/FM audio
USB
Control keys
TG OUT
Output file data to USB
memory or a USB printer
Image files: PNG, BMP
Setting files: BIN, XML
Tracking generator
(option)
DC INPUT
Battery mount
Mount for the battery pack
(battery sold separately)
Operation with an external
DC power supply
+11 VDC to +17 VDC
GPIB
Rear Panel
IF OUT
21.4-MHz IF
signal output
VIDEO
EXT TRIG
Option Guide
Output to an
external VGA
monitor
USB
EXT REF
LAN
Standard signal input
for external reference
Control and remote operation,
utilizing an external PC
Main unit support
Commons
75 series 1)
50 series 1)
Product name
Model number
Overview
2 Channel input (50 )
OPT.10
Addition of RF INPUT2 (9 kHz to 3 GHz)
Individual RF measurement with RF INPUT 1 and RF INPUT 2
EMC filter
OPT.28
Addition of CISPR bandwidth for EMI measurement, and QP detector
RBW (6 dB Down): 200 Hz, 9 kHz,120 kHz, 1 MHz
High-purity spectrum analysis
(1 ch/2 ch)
OPT.70/71
Spectrum analysis with -102 dBc/Hz @ 10 kHz offset (Typical)
Addition of RBW 30 Hz
U3741
U3751
1ch 2ch 1ch 2ch
OPT.70 OPT.71 OPT.70 OPT.71
Tracking generator (3 GHz)
OPT.76
Frequency range: 100 kHz to 3 GHz
Output level range: 0 to -60 dBm
2)
Tracking generator (6 GHz)
OPT.77
Frequency range: 100 kHz to 6 GHz
Output level range: 0 to -30 dBm
2)
2 Channel input (75 )
OPT.11
RF INPUT 2 (9 kHz to 2.2 GHz) in addition to OPT.15
Individual RF measurement with RF INPUT 1 and RF INPUT 2
1 Channel input (75 )
OPT.15
RF INPUT: 75 (100 kHz to 2.2 GHz) For CATV and TV
picture signal measurement. Channel table data installed.
Tracking generator (2.2 GHz)
OPT.75
Frequency range: 100 kHz to 2.2 GHz.
Output level range: 107 to 47 dBV
High-stability frequency
reference source
OPT.20
Reference oscillator with an aging rate of
2 x 10-8/day, 1 x 10-7/year
Time-domain analysis (1 ch/2 ch)
OPT.53/54
Analyze the basic parameter of RF signal on a time domain (CBW: 3MHz)
(amplitude/phase/frequency/FFT/IQ/IQ output)
OPT.53 OPT.54 OPT.53 OPT.54
Wide-band time-domain
analysis (1 ch/2 ch)
OPT.55/56
Analyze the basic parameter of RF signal on a time domain (CBW: 40MHz)
(amplitude/phase/frequency/FFT/IQ/IQ output)
OPT.55 OPT.56 OPT.55 OPT.56
1) The options of 50 series and 75 series cannot be installed simultaneously. 2): One must be selected from OPT.76/77.
Available
Not available
U3741/3751-6E Mar. 13
Compact Design with High Performance
High throughput
With the U3741/3751, warm-up time has been reduced to a scant
5 minutes (at an ambient temperature of 20 to 30C). This shortened period virtually eliminates pre-warming time as a consideration, and permits quick and accurate measurement.
This spectrum analyzer delivers data transfer speed superior to that
of its predecessor. While the previous model delivered 875 ms, the
U3741/3751 boasts a speed of 350 ms: double the system throughput *2 (using the GPIB interface)*3. This faster speed contributes to
a significant reductions to cost of test on production lines and in
similar applications.
Stable measurement
5-minute warm-up time
min.
U3741/3751
U3741/3751
Our conventional model
Our conventional model
Time
Measuring time
Improvements in overall accuracy
Standard USB (1.1) interface
Digitized IF sections and innovative circuit technology dramatically
improve absolute power measurement accuracy.
Screenshots in BMP or PNG format can easily be sent via USB
external memory. Users can easily store data, and easily paste measurement data into reports.
0.8 dB (10 MHz to 3 GHz: U3741/3751)
1.0 dB (3 to 8 GHz: U3751)
RF Input
ADC
Digital RBW Digital Det
Digital IF Block
Up to 2.5 hours *1 of nonstop battery-driven operation
Compact design
The spectrum analyzer uses one of three power systems: AC (100
V/200 V), DC (+11 V to +17 V), or the battery pack. This flexibility enables measurement in a variety of applications, whether in
the factory or in the field.
At about half the size of its predecessor, this spectrum analyzer
offers a compact design while maintaining the same level of functionality. Its form factor gives it portability, enabling it to be used
anywhere.
*1: Typical value at room temperature, without options
*2: Twice that of its predecessor
*3: Sample case where the frequency and span are specified, and the channel power measurement
result is transferred
Extensive array of measurement functions
U3741/3751-6E Mar. 13
Measurement functions include Channel Power, Total Power,
Avg Power, OBW, ACP, Spurious measurement, Harmonics measurement, IM measurement, Noise/Hz calculation functions,
multi-marker (10 markers), delta marker, peak marker functions,
a channel setting function, and a 3-trace simultaneous sampling
function.
Measurement Functions
RMS Average, essential for power measurement
Pre-Amp covering the 3 GHz/8 GHz bandwidth
Power tends to be spread over a wide frequency range, and the peak
factor tends to be higher in digital modulation, with its expanded
communication capacity. The U3741/3751 allows precise power
measurements by determining the effective values (RMS values)
from instantaneous power values obtained in high-speed sampling
and translating them into a power spectrum. This method also
enables measurement reproducibility of 0.01 dB in power measurement of digitally modulated signals.
The U3741/3751 contains as standard a pre-amp that covers all
frequency bands. In the analysis of faint signals, its input sensitivity can be equivalent to that of high-end models. Also, it effectively compensates for the loss from the antenna when measuring
radio signals in an outdoor environment.
Example of highsensitivity measurement
in high-sensitivity mode
Example of ISDB-T
Channel Power
measurement
Built-in frequency counter with 1-Hz resolution
USER keys
Frequency can be accurately measured by simply positioning the
cursor on the target spectrum selected from multiple spectral lines.
The U3741/3751 is indispensable for measuring the carrier wave
frequency in a general multi-carrier system.
An arbitrary key can be selected from the hierarchical function
keys and assigned to a USER function. Users can thus configure
their own, original setup for operations by assigning frequently
used functions to specific software keys.
Example of multicarrier signal frequency
measurement
Example of
user function
assignment
Zoom function
Spectrum emission mask function
The measuring window and F-F mode can facilitate analysis of a
specific signal in broadband measurement. Also, RBW can be
changed independently, enabling high-speed measurement of the
target signal in both broadband and narrowband. A variety of
other signal analysis functions are also available, including those in
F-T mode or T-T mode.
Using tools such as a spectrum mask and limit line to judge
PASS/FAIL is effective at improving production line throughput
for digital appliances. Using the spectrum emission mask (SEM)
function can facilitate measurement for standards such as wireless
LAN.
Example of two-screen
sample from
measurement in
broadband and
narrowband
Example of S.E.M.
measurement for
wireless LAN
U3741/3751-6E Mar. 13
User-friendly and Convenient Functions
Gated Sweep function
Gated Sweep OFF
A radar or TDMA communication system controls its output
transmission by turning the power on/off intermittently. To monitor the power spectrum during transmission, the Gated Sweep
function is effective at analyzing the spectrum only when the signal
is present and over only the area chosen. This function also
includes an IF trigger that does not require synchronized signals.
Gated Sweep ON
EXT TRIG
Synchronized
signals
RF INPUT
Burst Signal
Ideal for remote operation/monitoring via a LAN
This spectrum analyzer is equipped with a 10/100BASE-T LAN
port as standard, so it can be operated remotely from an external
PC. It can be installed in an unattended radio transmission station,
and remotely operated and monitored from another station.
LAN
PC
Screen of remote operation/monitoring from an external PC via LAN
PC
Server
Searching for the location of a fault in a coaxial cable
When used with its tracking generator option and the sample software for an external PC, the U3741/3751 can measure the distance to the failure point (open/short) in a coaxial cable. This
application permits this distance to be measured from one end of
the coaxial cable.
GPIB
TG OUT
RF INPUT
Power splitter
U3741/3751-6E Mar. 13
PC with sample software
Cable
Screen for measuring the distance to a cable failure point
Extensive Array of Options
2 Channel Input OPT.10 (50 )/11 (75 )
Two-channel input option (OPT.10/OPT.11) offers two independent lines of RF input. Various measurement conditions
including measuring frequency and spans can be set independently for each RF input.
High-speed process by the parallel processing
Simultaneous measurement of standard items (Channel power
and OBW, etc.)
Reduction in time by two-piece simultaneous measurement
Simultaneous measurement of the different system, etc.
Simultaneous measurement of different frequency (1 GHz or
less and micro-wave) etc. at EMC measurement
Simultaneous measurement of Channnel Power and OBW
Applications only possible for a two-channel spectrum
analyzer
Timing measurement between two channels by the synchronized
sweep and synchronized trigger
Simultaneous spectrum observation of the different frequency
by the synchronized sweep when sweeping time is the same
Simultaneous observation of the whole/part by the synchronized
trigger
Simultaneous monitoring of input/output devices
Allocating Connectors on Front Panel (for U3741)
TG OUTPUT (OPT.76)
RF INPUT 1
(standard equipment)
RF INPUT 2 (OPT.10)
FSK signal measurement (required with OPT.54)
Simultaneous measurement of the broadband/narrowband by the synchronized sweep
Simultaneous measurement of input/output for feed-forward amp
Timing measurement of TPMS by the synchronized trigger
U3741/3751-6E Mar. 13
Extensive Array of Options
Time-Domain Analysis OPT.53 (1 ch)/54 (2 ch)
Wide-Band Time-Domain Analysis OPT.55 (1 ch)/56 (2 ch)
By installing this option in addition to the function of the conventional sweeping-type spectrum analyzer, a the time-domain analysis
basic functions is added at low-cost.
Signal observation based on a domain different from
sweeping-type spectrum analyzer
Change in frequency over time by Freq. vs. Time analysis
(ex. analysis of FSK signals, such as keyless entry and TPMS)
Change in phase over time by Phase vs. Time analysis
Change in power over time by Power vs. Time analysis
High resolution (equivalent of 1 Hz RBW) high sensitivity
measurement by FFT
High sensitivity measurement by FFT (RBW 1Hz, -160dBm/Hz (typ))
Time-domain analysis for two singnals (OPT.54/56)
The time-domain basic analysis function in the range of 9 kHz to
8 GHz (on main body) can be installed simultaneously for 2 channels. Unique analysis functions, such as Freq. vs Time during
input and output are realized.
Wide-band time-domain analysis (OPT.55/56)
In the frequency ranges of 9 kHz to 8 GHz (on main body), timedomain analysis for up to the maximum measurement bandwidth
40 MHz is possible.
Measurement using the time-domain basic analysis function
FFT Spectrum
Freq. vs. Time
Power vs. Time
Phase vs. Time
FREQ. vs. Time measurement of the 4 value FSK
Radar wave measurement (OPT.55 for Wide-band time-domain analysis)
Spectrum Analyzer U3700 Series
MIX.
BPF
ADC
RBW
Power DET.
RF Input
LO
I/Q
Detector
FFT
Spectrum
Sample Rate
Control
OPT.53
Time-Domain
Analysis Option
Waveform
Memory
Power Vs.
Time
Freq. Vs.
Time
Display
Switching
Display
Download I/Q data
to external PC
Phase Vs.
Time
IQ Waveform
U3741/3751-6E Mar. 13
GPIB/LAN
Extensive Array of Options
High-Purity Spectrum Analysis OPT.70 (1 ch)/71 (2 ch)
-70
Phase noise measurement is indispensable to evaluation of the
characteristics of high-frequency oscillation circuits or modules.
The high-purity spectrum analysis option offered with the
U3741/3751 can improve the phase noise measurement performance of the spectrum analyzer. Because the performance can be
selected, selecting the most suitable spectrum analyzer for the
device under test (DUT) is simple. At the same time, the added
resolution bandwidth of 30 Hz enables reduction of the display
average noise level and analysis in a high dynamic range.
Phase Noise [dBc/Hz]
-80
@ 2 GHz
@ 5 GHz
-90
-100
-110
-120
-130
-140
2 channel-inputs option (OPT.10/11) is required for OPT.71 installation.
10
100
1,000
10,000
Carrier Offset Frequency [kHz]
Phase noise characteristic graph (representative values)
Tracking Generator OPT.75/76/77
Generates synchronized signals for frequency sweeps by the spectrum analyzer.
OPT.76
Output impedance: 50
Output frequency range: 100 kHz to 3 GHz
Output impedance: 50
Output frequency range: 100 kHz to 6 GHz
OPT.75
Output impedance: 75
OPT.77
Output frequency range: 100 kHz to 2.2 GHz
Functions for evaluating frequency characteristics
The normalize function enables direct measurement of cable loss
and filter characteristics. The frequency offset function of the
tracking generator enables measurement of frequency characteristics and conversion loss characteristics of mixers and other frequency conversion devices.
Characteristics of
a mixer + filter
Characteristics of a filter
RF INPUT
TG OUT
Fin
Frequency
conversion device
Fout
Measurement of mixer frequency conversion loss characteristics
Lo
Function for return loss measurement
The SWR bridge can be used to measure reflection characteristics
of an antenna or filter. It can determine the return loss and evaluate
the VSWR.
RF INPUT
TG OUT
SWR bridge
Filter return loss measurement
Antenna, filter, etc.
U3741/3751-6E Mar. 13
Extensive Array of Options and Accessories
High-Stability Frequency Reference Source OPT.20
Frequency of the high frequency signal was conventionally
counted with a frequency counter. However, multi-carrier method
is often employed for the recent communication system which
uses high frequency signals which contains multiple frequency
components, a frequency counter cannot count the frequency correctly. Therefore, the frequency counter of the spectrum analyzer
attracts attention as an essential function.
In a spectrum analyzer, just by pointing the marker at the spectrum separated as a sine wave of CW, not only the frequency
counting but also faint signal level counting is possible. OPT.20
improves the aging stability of the standard oscillator which determines the frequency counter accuracy of a spectrum analyzer.
Standard
OPT.20
Aging rate
2 x 10-6/year
2 x 10-8/day, 1 x 10-7/year
EMC Filter OPT.28
Option 28 adds 6 dB RBW CISPR bandwidths for EMI measurement of 200 Hz, 9 kHz, 120 kHz, and 1 MHz. A broadband
sweep by the spectrum analyzer is very effective at measuring noise
emitted from electrical devices. Installing OPT.28 allows measurement in CISPR-specified bandwidths. It enables simple, fast measurement using the Positive peak detector and Max Hold, which
makes it effective at compensating for emitted noise. It guarantees
an impulse bandwidth accuracy of 1 MHz. This capability conforms to the standard for noise measurement of 1 GHz or above.
Measurement using EMI sample software
Accessories
Transit case
Many accessories are available, including an easy-to-carry transit
case and a battery pack, useful for field work.
Carrying bag
DC power cable
Charger
Battery pack
10
U3741/3751-6E Mar. 13
50-75
impedance converter
Specifications
Amplitude range
Frequency
Measurement range:
Frequency range
U3741:
Pre-Amp:
Synchronizable
frequency range:
U3751:
Frequency band:
Pre-Amp:
Frequency reading
accuracy:
9 kHz to 3 GHz,
9 kHz to 2.2 GHz (with the OPT.15 installed)
10 MHz to 3 GHz,
10 MHz to 2.2 GHz (with the OPT.15 installed)
9 kHz to 3 GHz
9 kHz to 8 GHz
9 kHz to 3.1 GHz (band 0),
3 GHz to 8 GHz (band 1)
10 MHz to 8 GHz
(marker read value x frequency reference
accuracy + span x span accuracy + residual FM)
Frequency reference stability
Aging rate:
2 x 10-6/year
Temperature stability:
2.5 x 10-6 (0 to 50C)
Frequency counter:
Resolution:
Accuracy:
Resolution bandwidth 100 kHz,
span 100 MHz, signal level: S/N >50 dB
1 Hz to 1 kHz
(counter read value x frequency reference
accuracy + residual FM + 1 LSB)
Frequency stability
Residual FM (zero/span): < 60 Hzp-p/100 ms (internal frequency reference)
Frequency span
Range:
Accuracy:
5 kHz to Full, zero span
1 kHz to Full, zero span
(with the OPT.70 installed)
< 1%
Spectrum purity:
-85 dBc/Hz (offset 10 kHz, span < 200 kHz)
Resolution bandwidth
Range:
U3741:
U3751:
Accuracy:
100 Hz to 1 MHz (1 to 3 steps)
30 Hz to 1 MHz (with the OPT.70/71 installed)
100 Hz to 3 MHz (1 to 3 steps)
30 Hz to 3 MHz (with the OPT.70/71 installed)
< 12%
Video bandwidth range:
10 Hz to 3 MHz (1 to 3 steps)
Sweep
Sweep time
Setting range:
Accuracy:
20 ms to 1000 s (spectrum mode)
50 s to 1000 s (zero span)
< 2% (zero span)
Sweep mode:
Continuous, single, gated
Trigger function
Trigger source:
Free run, video, external, IF
Displayed average noise level to +30 dBm
Displayed average noise level to 134 dBV
(with the OPT.15 installed)
Maximum safe input level: Attenuator 10 dB
Pre-Amp OFF:
+30 dBm, 134 dBV (with the OPT.15 installed)
Pre-Amp ON:
+13 dBm, 120 dBV (with the OPT.15 installed)
U3741:
50 VDC max.
U3751:
15 VDC max.
Input attenuator range:
0 to 50 dB (10 dB steps)
Display range:
100/50/20/10/5 dB, linear
Scale unit:
dBm, dBmV, dBV, dBVemf, dBpW, W, V
Reference level
setting range:
-140 to +40 dBm
-31.2 to 148.8 dBV (with the OPT.15 installed)
Detection mode:
Normal, Positive peak, Negative peak,
Sample, RMS, and Average
Amplitude accuracy
Calibration signal
Frequency:
Level:
Accuracy:
20 MHz
-20 dBm (75 , with the OPT.15 installed)
0.3 dB, 0.4 dB (with the OPT.15 installed)
Scale display accuracy
Log:
0.5 dB/10 dB, 0.5 dB/80 dB, 0.2 dB/1 dB
Overall amplitude
accuracy:
After calibration, with the pre-amp OFF,
and at a temperature ranging from 20 to 30C
Input attenuator 10 dB
U3741:
Reference level 0 dBm,
input signal level -10 to -50 dBm
1.0 dB (9 kHz to 3 GHz)
0.8 dB (10 MHz to 3 GHz)
With the OPT.15 installed: Reference level 108.8 dBV
Input signal level 98.8 to 58.8 dBV
2.1 dB (9 kHz to 2.2 GHz)
0.9 dB (10 MHz to 2.2 GHz)
U3751:
Reference level 0 dBm,
input signal level -10 to -50 dBm
Image suppression OFF
1.5 dB (9 kHz to 10 MHz)
0.8 dB (10 MHz to 3.1 GHz)
1.0 dB (3.1 GHz to 8 GHz)
U3741/3751-6E Mar. 13
11
Dynamic range
Inputs/outputs
Displayed average
noise level:
U3741:
Pre-Amp OFF:
Pre-Amp ON:
U3751:
Pre-Amp OFF:
Pre-Amp ON:
Reference level < -45 dBm (63.8 dBV,
with the OPT.15 installed)
Resolution bandwidth 100 Hz
Frequency 10 MHz to 3 GHz
-123 dBm + 2f (GHz) dB (f < 2.5 GHz)
-123 dBm + 2.5f (GHz) dB (f 2.5 GHz)
-12 dBV + 2f (GHz) dB (f 2.2 GHz,
with the OPT.15 installed)
-138 dBm + 3f (GHz) dB
-27 dBV + 3f (GHz) dB
(with the OPT.15 installed)
Frequency 10 MHz to 8 GHz
-123 dBm + 2f (GHz) dB (f 3.1 GHz, band 0)
-122 dBm + 1f (GHz) dB (f 3 GHz, band 1)
-138 dBm + 3f (GHz) dB (f 3.1 GHz, band 0)
-139 dBm + 1.3f (GHz) dB (f 3 GHz, band 1)
1 dB gain compression
U3741:
Pre-Amp OFF:
Pre-Amp ON:
U3751:
Pre-Amp OFF:
Pre-Amp ON:
Frequency > 20 MHz
> -5 dBm
> 102 dBV (with the OPT.15 installed)
> -25 dBm
> 82 dBV (with the OPT.15 installed)
Frequency > 20 MHz
> -8 dBm
> -25 dBm
Second harmonic distortion
U3741:
U3751:
<-70 dBc (Pre-Amp OFF, Frequency > 20 MHz,
Mixer input level -30 dBm (77 dBV, with
the OPT.15 installed))
<-70 dBc (Pre-Amp OFF, Frequency > 200 MHz,
Mixer input level -40 dBm)
<-75 dBc (typ., Pre-Amp OFF, Frequency
> 300 MHz, Mixer input level -30 dBm)
Third order intermodulation distortion
U3741:
< -60dBc (Pre-Amp OFF, Mixer input level
-20 dBm (88.8 dBV, with the OPT.15
installed), Frequency > 10 MHz,
2 signal separation > 200 kHz)
U3751:
< -50 dBc (Pre-Amp OFF, Mixer input level
-20 dBm, Frequency 10 MHz to 8 GHz,
2 signal separation > 200 kHz)
Image/multiple/out of band response
U3741:
< -60 dBc
(Mixer input level -20 dBm (88.8 dBV, with
the OPT.15 installed))
U3751:
< -60 dBc
(Mixer input level -30 dBm,
Image suppression ON)
Residual response
U3741:
U3751:
12
U3741/3751-6E Mar. 13
< -90 dBm (Frequency > 1 MHz , Pre-Amp OFF)
< 21 dBV (with the OPT.15 installed)
< -80 dBm
(Frequency 10 MHz to 8 GHz, Pre-Amp OFF)
RF input
Connector:
Impedance:
VSWR:
U3741:
U3751:
N-type female
50 (nominal)
75 (nominal, with the OPT.15 installed)
Input attenuator 10 dB
< 1.5 : 1
< 1.6 : 1 (with the OPT.15 installed)
< 1.7 : 1 (10 MHz Frequency 3.0 GHz)
< 2.0 : 1 (Frequency > 3.0 GHz)
Calibration signal output
Connector:
Impedance:
Frequency:
Level:
BNC female
50 (nominal)
75 (nominal, with the OPT.15 installed)
20 MHz
-20 dBm
Frequency reference input
Connector:
Impedance:
Frequency (MHz):
Level:
BNC female
50 (nominal)
1, 1.544, 2.048, 5, 10, 12.8, 13, 13.824, 14.4,
15.36, 15.4, 16.8, 19.2, 19.44, 19.6608,
19.68, 19.8, 20, 26
0 to +16 dBm
External trigger input
Connector:
Impedance:
Level:
BNC female
10 k (nominal), DC coupling
0 to +5 V
21.4-MHz IF output
Connector:
Impedance:
Level:
BNC female
50 (nominal)
Approx. mixer input level + 10 dB
(at a frequency of 20 MHz)
Battery mount
Connector:
AntonBauer QR mount
External DC power input
Connector:
Voltage range:
XLR-4
+11 to +17 V
GPIB:
USB:
Video output connector:
LAN connector:
Audio output:
IEEE-488 bus connector
USB 1.1
D-sub15 pin female
RJ45 type, 10/100 base-T
Small monophonic jack
General specifications
Operating environment range:
Storage environment range:
AC power input:
DC power input:
Power consumption:
Mass
U3741:
U3751:
External dimensions
(W x H x D):
Ambient temperature: 0 to + 50C
Humidity: RH 85% or less (no condensation)
-20 to +60C, RH 85% or less
Automatic switching to 100 VAC or 200 VAC
100 V: 100 to 120 V, 50/60 Hz
200 V: 220 to 240 V, 50/60 Hz
DC + 11 V to +17 V
100 VA or less (AC operation)
70 W or less (DC operation)
5 kg or less (without option)
5.6 kg or less (without option)
Approx. 308 x 175 x 209 mm
(not including protruding parts)
Approx. 337 x 190 x 307 mm
(including the handle and feet)
OPT.10 2 Channel input (50 , 3 GHz)
OPT.55/56 Wide-band time-domain analysis (1 ch/2 ch)
Cross talk between input
channels (between RF input
1 and RF input 2 ):
RF input 2
Connector:
Impedance:
VSWR:
External trigger input:
21.4 MHz IF output:
RF range:
Follows the U3741/3751.
RF amplitude range:
Noise level to +30 dBm *1)
Wave recording method:
I/Q vector time waveform
Measuring bandwidth (CBW): 100 Hz to 30 MHz (1 to 3 steps), 40 MHz
IQ sampling rate:
500 Hz (BW 100 Hz) to 65 MHz (BW 40 MHz)
IQ waveform recording time: 120 msec (BW 40 MHz) to 1000 sec (BW 100 Hz)
Number of IQ waveform
recording samples:
8 M samples (I/Q)
<-90 dBc (Input level -10 dBm, Input
attenuator 0 dB, Preamplifier off)
N type female
50 (nominal)
<1.5 : 1 (Input attenuator > 10 dB)
An external trigger input can be selected as
a trigger input of RF input 2 when installing
the OPT.10. The input connector is only 1
system.
Only IF output which supports RF input 1,
when installing the OPT.10.
Except for all items mentioned above, the frequency, sweep, amplitude range,
amplitude accuracy, dynamic range, input/output, and performance of specifications
follow the standard specifications of the RF input 1 option of the U3741 spectrum
analyzer.
OPT.11 2 Channel input (75 , 2.2 GHz)
Cross talk between input
channels (between RF input
1 and RF input 2 ):
RF input 2
Connector:
Impedance:
VSWR:
External trigger input:
21.4 MHz IF output:
<-90 dBc (Input level 98.8 dBV, Input
attenuator 0 dB, Preamplifier off)
N type female
75 (nominal)
<1.5 : 1 (Input attenuator > 10 dB)
An external trigger input can be selected as
a trigger input of RF input 2 when installing
the OPT.11. The input connector is only 1
system.
Only IF output which supports RF input 1,
when installing the OPT.11.
Except for all items mentioned above, the frequency, sweep, amplitude range,
amplitude accuracy, dynamic range, input/output, and performance of specifications
follow the standard specifications of the RF input 1 option of the U3741 + OPT.15
spectrum analyzer.
OPT.20 High-stability frequency reference source
Frequency reference stability
Aging rate:
Warm-up drift:
Temperature stability:
2 x 10-8/day
1 x 10-7/year
5 x 10-8 (+25C, 10 minutes after power-on)
5 x 10-8 ( 0 to +40C, with reference to 25C)
OPT.28 EMC filter
6 dB bandwidth:
Bandwidth accuracy:
200 Hz, 9 kHz, 120 kHz, 1 MHz
< 10%
OPT.53/54 Time-domain analysis (1 ch/2 ch)
RF range:
Follows the U3741/3751.
RF amplitude range:
Noise level to +30 dBm *1)
Wave recording method:
I/Q vector time waveform
Measuring bandwidth (CBW): 100 Hz to 3 MHz (1 to 3 steps)
IQ sampling rate:
713 Hz (BW 100 Hz) to 21.4 MHz (BW 3 MHz)
IQ waveform recording time: 49 msec (BW 3 MHz) to 1000 sec (BW 100 Hz)
Number of IQ waveform
recording samples:
1 M samples (I/Q)
*1) The noise level follows the dynamic range of the U3741/3751.
*1) The noise level follows the dynamic range of the U3741/3751.
OPT.70/71 High-purity spectrum analysis (1 ch/2 ch)
Frequency span
Range:
Accuracy:
1 kHz to Full, zero span
< 1%
Resolution bandwidth
Range:
Accuracy:
U3741: 30 Hz to 1 MHz (1 to 3 steps)
U3751: 30 Hz to 3 MHz (1 to 3 steps)
< 12%
Spectrum purity:
-98 dBc/Hz (offset 10 kHz, span 1 MHz)
-102 dBc/Hz (Typical)
Displayed average
noise level:
U3741:
Pre-Amp OFF:
Pre-Amp ON:
U3751:
Pre-Amp OFF:
Pre-Amp ON:
Reference level < -45 dBm,
Resolution bandwidth 30 Hz
Frequency 10 MHz to 3 GHz
-126 dBm + 2f (GHz) dB (f < 2.5 GHz)
-126 dBm + 2.5f (GHz) dB (f 2.5 GHz)
-141 dBm + 3f (GHz) dB
Frequency 10 MHz to 8 GHz
-126 dBm + 2f (GHz) dB (f 3.1 GHz, band 0)
-125 dBm + 1f (GHz) dB (f 3 GHz, band 1)
-141 dBm + 3f (GHz) dB (f 3.1 GHz, band 0)
-142 dBm + 1.3f (GHz) dB (f 3 GHz, band 1)
OPT.75 Tracking generator (75 , 2.2 GHz)
Frequency range:
100 kHz to 2.2 GHz
Frequency offset
Range:
Accuracy:
Resolution:
0 Hz to 1 GHz
300 Hz
1 kHz
Output level range:
107 to 47 dBV (0.5 dB steps)
Output level accuracy:
0.5 dB (20 MHz, 97 dBV, +20 to +30C)
Output level flatness:
Using 20 MHz and 97 dBV as a reference
1.0 dB (1 MHz to 1 GHz)
1.5 dB (100 kHz to 2.2 GHz)
Output level switch error:
Frequency offset OFF:
Frequency offset ON:
Using 20 MHz and 97 dBV as a reference
1.0 dB (1 MHz to 1 GHz, 107 to 47 dBV)
2.0 dB (1 MHz to 2.2 GHz, 107 to 47 dBV)
3.0 dB (100 kHz to 2.2 GHz, 107 to 77 dBV)
4.0 dB (100 kHz to 2.2 GHz, 76.5 to 47 dBV)
5.0 dB (100 kHz to 2.2 GHz)
Output spurious:
Harmonic:
Non-harmonic:
Output level 97 dBV
< -15 dBc (100 kHz to 1 MHz)
< -20 dBc (1 MHz to 2.2 GHz)
< -20 dBc (Frequency offset OFF)
TG leakage:
< 31 dBV (Input attenuator 0 dB)
Output impedance:
VSWR:
75 (nominal)
2.0 : 1 (Output level 97 dBV)
Maximum allowable level: 117 dBV, 10 VDC
U3741/3751-6E Mar. 13
13
OPT.76 Tracking generator (50 , 3 GHz)
Ordering information
Frequency range:
Main unit
Spectrum analyzer:
U3741
U3751
Accessories
Operating manual (CD):
Power cable:
Input cable:
With the OPT.15 installed:
N-BNC adapter:
With the OPT.15 installed:
NC-F adapter (with the OPT.15 installed):
Ferrite core:
BU3700S
A01412
A01037-0300
A01045
JUG-201A/U
BA-A165
NCP-NFJ
ESD-SR-120,
E04SR150718
Options
2 Channel input (50 , 3 GHz):
2 Channel input (75 , 2.2 GHz):
1 Channel input (75 ):
High-stability frequency reference source:
EMC filter:
Time-domain analysis (1 ch):
Time-domain analysis (2 ch):
Wide-band time-domain analysis (1 ch):
Wide-band time-domain analysis (2 ch):
High-purity spectrum analyzsis (1 ch):
High-purity spectrum analyzsis (2 ch):
Tracking generator (75 , 2.2 GHz):
Tracking generator (50 , 3 GHz):
Tracking generator (50 , 6 GHz):
OPT.10
OPT.11
OPT.15
OPT.20
OPT.28
OPT.53
OPT.54
OPT.55
OPT.56
OPT.70
OPT.71
OPT.75
OPT.76
OPT.77
Accessories
Japanese operating manual (printed manual):
English operating manual (printed manual):
Battery pack:
Charger:
75 input impedance converter:
DC power cable:
Transit case:
Rack mount kit (JIS):
Rack mount kit (EIA):
JU3700S
EU3700S
A870008
A870009
ZT-130NC
A114020
A129002
A122003
A124004
100 kHz to 3 GHz
Frequency offset
Range:
Accuracy:
Resolution:
0 Hz to 1 GHz
300 Hz
1 kHz
Output level range:
0 to -60 dBm (0.5 dB steps)
Output level accuracy:
0.5 dB (20 MHz, -10 dBm, +20 to +30C)
Output level flatness:
Using 20 MHz and -10 dBm as a reference
1.0 dB (1 MHz to 1 GHz)
1.5 dB (100 kHz to 3 GHz)
Output level switch error:
Frequency offset OFF:
Frequency offset ON:
Using 20 MHz and -10 dBm as a reference
1.0 dB (1 MHz to 1 GHz, 0 to -60 dBm)
2.0 dB (1 MHz to 2.6 GHz, 0 to -60 dBm)
3.0 dB (100 kHz to 3 GHz, 0 to -30 dBm)
4.0 dB (100 kHz to 3 GHz, -30.5 to -60 dBm)
5.0 dB (100 kHz to 3 GHz)
Output spurious:
Harmonic:
Non-harmonic:
Output level -10 dBm
< -15 dBc (100 kHz to 1 MHz)
< -20 dBc (1 MHz to 3 GHz)
< -20 dBc (Frequency offset OFF)
TG leakage:
< -80 dBm (Input attenuator 0 dB)
Output impedance:
VSWR:
50 (nominal)
2.0 : 1 (Output level -10 dBm)
Maximum allowable level:
+10 dBm, 10 VDC
OPT.77 Tracking generator (50 , 6 GHz)
Frequency range:
Output level range:
Output level accuracy:
Output level flatness:
TG leakage:
Output impedance:
VSWR:
Maximum allowable level:
*2)
100 kHz to 6 GHz
0 to -30 dBm (0.5 dB step)
0.5 dB (20 MHz, -10 dBm, +20 to +30C)
20 MHz on -10 dBm criterion, at +20 to +30C
1 dB (1 MHz to 1 GHz)
1.5 dB (100 kHz to 3.1 GHz)
2.0 dB (100 kHz to 6 GHz)
-80 dBm (input attenuator: 0 dB)
50 (nominal)
2.0 : 1 (Output level -10 dBm)
+10 dBm, 10 VDC
*2) The OPT.77 is not allowed to be installed on the U3741.
14
U3741/3751-6E Mar. 13
Note on accessories:
The operating manual on the CD is supplied as standard.
The printed version of the operating manual is offered as an accessory.
Sample software
to be downloaded free from homepage
ADVANTEST provides various kinds of sample software
shown below :
Useful sample software for EMI measurement and Radio
waves monitor, etc.
Module software with source code to control a Spectrum
analyzer for developers.
http://www.advantest.co.jp/en-index.shtml
PRODUCTS & SUPPORT
Electronic Measuring Instruments
Products
U3741 or U3751
Sample Software
EMI measurement software (2 ch)
Radio waves monitor (1 ch/2ch)
Please refer to product manual for complete system specifications.
Specifications may change without notification.
U3741/3751-6E Mar. 13
15
http://www.advantest.co.jp
ADVANTEST CORPORATION
Shin-Marunouchi Center Building, 1-6-2 Marunouchi, Chiyoda-ku, Tokyo 100-0005, Japan Phone: +81-3-3214-7500
2007 ADVANTEST CORPORATION Printed in Japan Bulletin No.U3741/3751-SI6E-137 Mar. 13 S