ORDERING INFORMATION:
100 01
50NB 01 01 01 01 01
SERVICE
AIR
BIOGAS
LPG
ANY OTHER FLUID
STEAM
CODE
25
40
50
80
100
150
AO
LINE SIZE
25 NB
40 NB
50 NB
80 NB
100 NB
150 NB
ANY OTHER PL. SPECIFY
CODE
01
02
DENSITY COMPENSATION
REQUIRED
NOT REQUIRED
CODE
01
02
END CONNECTION
WNRF
SORF
CODE
01
02
DPT
BLIND
WITH DISPLAY
CODE
01
02
FIT
PANEL
WALL
CODE
01
02
PT
BLIND
WITH DISPLAY
Crownsp @ vsnl.com nov 2005-2000
GFM
CODE
100 01
100 02
100 03
100 04
200 05
SAMPLE ORDERING INFORMATION
Due to continuous development specifications are subject to change without prior notice.
manas microsystems pvt. ltd.
TM
REGD. OFFICE : ROW HOUSE NO. 5, PARVATI PARK, DATTAWADI, PUNE 411 030
WORKS : EL 54 ELECTRONIC ZONE, M.I.D.C. BHOSARI, PUNE 411 026. (INDIA)
TEL. : 020-27127044,27127047,27127858 FAX : 020-27127045 E-MAIL : [email protected]
WEBSITE : www.manasmicro.com
...a name that spells trust
AN ISO 9001:2000 COMPANY
Catalog No.: Fc-MFC-01
GFM 100 & GFM - 200
GAS, AIR & STEAM FLOW METERING SYSTEM.
GFM 100 & GFM - 200
GAS, AIR & STEAM FLOW METERING SYSTEM.
INTRODUCTION:
Use our GFM series flow meter for measuring flow rates of
bio-gas, LPG, compressed air, steam (saturated &
superheated) in closed pipes. The flow meter is suited for wide
range of applications where affordability, reliability and
ruggedness are of prime importance.
In conventional system of measurement, the differential
pressure generated by orifice plate is measured by D.P.
transmitter. The output from D.P. transmitter after square
rooting is accepted as proportional to flow - rate. This
assumption is true only when density is constant.
Unfortunately, density of compressible fluid is never
constant. The density of compressible fluid changes with line
pressure & line temperature. Thus, introducing errors in flow
rate measurement.
MANAS MAKE GFM FLOW METERING SYSTEM:
PRINCIPLE OF OPERATION:
GFM series provides on - line density correction to
conventional system.
As per BS 1042 / ISO : 5167 standard, the equation for
mass flow when measured with orifice states:
Qm
r . DP
Where,
Qm = mass flow rate.
r = instantaneous density.
DP = differential pressure.
Thus, by measuring line pressure & temperature & using
the relevant algorithms, instantaneous density can be found
out & also the state of steam (i.e. saturated / superheated) .By
knowing the correct density one can compute the accurate
mass-flow rate. The further operation of integration, square
rooting is similar to ordinary flow Totaliser.
We, at MANAS understand our users' needs & hence offer
two models for orifice flow meter: GFM100 for Biogas, LPG &
air application & GFM 200 specifically for steam.
In GFM 200 the hardware like valve, temperature-sensing
elements etc can be provided with IBR approval.
The complete GFM system is provided in chemically
compatible material with the fluid and includes all the
necessary hardware for the installation. Since the density
calculations for steam and gases differ, the computing units
used for GFM100 & GFM200 are different. GFM series can
also be used for flow measurement of non-compressible
fluids like water, furnace oil etc. which increases its versatility.
PRINCIPAL ADVANTAGES OF GFM SERIES:
1. On - line density compensation possible because of online temperature & pressure measurement.
2. Various sizes of orifice assemblies available with
accurate calculations
3. Various types of needle valves, gate valves & manifolds
with or without IBR approval as per application.
4. Sturdy, rugged field mounting type pressure transmitter
is supplied with standard end connections.
5. On - line display of compensated mass flow rates,
density, temperature; output of DP transmitter is offered.
LED indication for status of steam (saturated or
superheated) is provided.
6. Various models of mass flow integrators (computing
units) CS-3001,CS-3002, CS-3003 available to suit
your application.
7. User friendly. No need to feed all the complicated orifice
constants since the system is intelligent enough to
calculate.
8. Isolated 4-20mA dc output proportional to compensated
flow rate.
9. Disconnection of DPT, PT, Temperature sensor or TT is
indicated by error message.
10. Partial system available to suit your existing
uncompensated flow rate using existing orifice & DP
transmitter or vortex flow meter.
11. Complete system engineered to suit your requirement.
12. Typical Turndown 10:3.
13. Accuracy: 3% of actual reading.
14. It is a standard system & highly reliable.
15. It is easy to maintain. Calibration of orifice, DP
transmitter, pressure transmitter is easy & inexpensive.
16. No moving parts.
ORIFICE
FLANGE
ASSEMBLY
Schematic
Arrangement for
GFM Series
TEMP MEAS.
MANAS MAKE INTEGRAL ORIFICE ASSEMBLY :
The INTEGRAL ORIFICE ASSEMBLY is a flow element
capable of adapting to DP transmitter to make a compact,
complete flow metering transmitting device without the need
of separate impulse piping.
ADVANTAGES OF INTEGRAL ORIFICE :
1. It provides simple & cost-effective solution.
2. Increase in reliability.
3. Installation cost reduces.
4. Time is saved because the unit is supplied as one
complete solution.
5. Available in sizes from to 1 .
APPLICATION :
GFM series is most suitable for compressible fluids,
which are difficult to measure because of variations in their
pressure and temperature.
IDEAL FOR :
l Sectors generating & consuming steam.
l Measuring compressor efficiency.
l Measuring biogas, LPG consumption & many other
similar applications.
SECTORS :
l Chemical / Pharmaceutical.
l Food & Drugs.
l Petrochemicals
l Fertilizers.
l Steel / Aluminium.
l Sugar Factories / Distilleries.
SPECIFICATIONS:
1. Service
Composition
2. Size
: to 14
3. Type of Flow meter
: Differential Pressure Type.
4. Flow Element
: Orifice / Integral Orifice / Nozzle
5. MOC Of Flow element : SS 316
6. Density Compensation : Can be provided.
7. Type of Flanges
: WNRF (Weld Neck Raised
Face) / SORF (Slip On Raised
Face)
8. MOC of Flange
: C.S / S.S / P.P.
9. Flange Rating
: ANSI 150 / ANSI 300 / ANSI 600
10. Type Of Taps
: Flanged Tapping / Corner
Tapping / D & D/2
11. No. Of DP Tappings
: 1 Pair
12. No. Of Drain Tappings : 1 Pair
13. Design Standard
: BS - 1042 / ISO : 5167
14. Accuracy
: 3% of actual reading
* Specifications are subject to change without prior notice.
WHY ORIFICE BASED FLOW METER PREFERRED OVER VORTEX METER?
Sr. No. Parameter
Orifice
Vortex
1.
Well established standards
Available
Not Available
2.
Suitability for high pressure &
temperature application (steam)
Most suitable
Seal fails in majority of cases after
certain duration.
3.
Installation
Easy to install
Critical & expensive because of
requirement of special machined pipe
lengths.
4.
Existing pipe line modifications
for installation
No modifications required.
Design is based on velocity & not on line
size. As a result customer line size &
selected flow meter size may differ.
5.
Recalibration of transmitter
Easy & can be done in house
Has to be done on a flow-rig and hence
is expensive
6.
Changes in Pressure, Temperature
Taken care during density
compensation.
May or may not be taken
care of.
7.
Suitability for line size 1 & below
Integral orifice assembly .
available
Not practical.
8.
Suitability for higher line sizes 8 & above
Suitable.
Higher cost as compared to orifice &
limited output pulse resolution.
9.
Suitability for low velocity measurement
Suitable
Not suitable
10.
Durability
No moving parts and hence
Diaphragm based sensor and hence is
less wear and tear and virtually prone to wear and tear.
maintenance free
GAS / STEAM FLOW
P.T.
3WAY
MANIFOLD
VALVE
ISOLATION
VALVES
CONDENSING
POT
CONDENSING
POT
COMPUTING
UNIT
D.P.T.
DRAIN
VALVE
DRAIN
VALVE
: Biogas, Compressed Air,
Compressed Gas, LPG, Steam.
: User to specify.