Technical Sales
(866) 531-6285
[email protected]
Requirements and Compatibility | Ordering Information | Detailed Specifications | Pinouts/Front Panel Connections
For user manuals and dimensional drawings, visit the product page resources tab on ni.com.
Last Revised: 2014-11-06 07:14:23.0
NI 9219
24-Bit Universal Analog Input
250 Vrms channel-to-channel isolation
Thermocouple, RTD, resistance, voltage, and current measurements
Built-in quarter-, half-, and full-bridge support
CJC per channel for accurate thermocouple measurement
Built-in voltage and current excitation
100 S/s/ch simultaneous inputs (50 S/s/ch for thermocouple)
Overview
The NI 9219 is a four-channel universal C Series module designed for multipurpose testing in any NI CompactDAQ or CompactRIO chassis. With the NI 9219, you can measure
several signals from sensors such as strain gages, RTDs, thermocouples, load cells, and other powered sensors. The channels are individually selectable, so you can perform a
different measurement type on each of the four channels. Measurement ranges differ for each type of measurement and include up to 60 V for voltage and 25 mA for current.
Please see the manual for detailed specifications and ranges.
Back to Top
Requirements and Compatibility
OS Information
Driver Information
Software Compatibility
Real-Time OS
NI-DAQmx
LabVIEW
Windows
NI-RIO
LabWindows/CVI
Measurement Studio
SignalExpress
Visual Basic
Visual C#
Visual C++
Visual Studio
Visual Studio .NET
Back to Top
Comparison Tables
Thermocouple Module
Channels
Sample Rate
Resolution
Feature
NI 9213
16
1,200 S/s
24-bit
Lowest cost/channel
NI 9219
50 S/s/ch
24-bit
Channel-to-channel isolation
NI 9211
14 S/s
24-bit
Low-channel count
Back to Top
1/9
www.ni.com
Application and Technology
Because of the driver design, the NI 9219 does not limit the overall speed of an NI CompactDAQ system when used with faster sampling modules. With 250 Vrms of
channel-to-channel isolation, the NI 9219 protects not only the surrounding modules, chassis, and connected computer system but also the other channels within the same
module. In addition to increased safety, channel-to-channel isolation eliminates problems associated with ground loops.
Connectivity Accessories
The NI 9219 uses six-position spring terminal connectors in each channel for direct signal connectivity. You can purchase additional connectors to reduce signal connection time
for multiple test units. In addition to extra connectors, a strain relief kit is available to secure the signal wires. NI recommends:
NI 9972 - strain relief backshells for signal wire security and high-voltage protection (qty 4)
NI 9973 - extra connectors for six-position connector modules (qty 10)
NI C Series Compatibility
The NI C Series hardware family features more than 50 measurement modules and several chassis and carriers for deployment. With this variety of modules, you can mix and
match measurements such as temperature, acceleration, flow, pressure, strain, acoustic, voltage, current, digital, and more to create a custom system. Install the modules in one
of several carriers to create a single module USB, Ethernet, or Wi-Fi system, or combine them in chassis such as NI CompactDAQ and CompactRIO to create a
mixed-measurement system with synchronized measurements. You can install up to eight modules in a simple, complete NI CompactDAQ USB data acquisition system to
synchronize all of the analog output, analog input, and digital I/O from the modules. For a system without a PC, CompactRIO holds up to eight modules and features a built-in
processor, RAM, and storage for an embedded data logger or control unit. For higher-speed control, CompactRIO chassis incorporate a field-programmable gate array (FPGA)
that you can program with NI LabVIEW software to achieve silicon-speed processing on I/O data from C Series modules.
Advanced Features
When used with CompactRIO, NI C Series analog input modules connect directly to reconfigurable I/O (RIO) FPGA hardware to create high-performance embedded systems. The
reconfigurable FPGA hardware within CompactRIO provides a variety of options for custom timing, triggering, synchronization, filtering, signal processing, and high-speed
decision making for all C Series analog modules. For instance, with CompactRIO, you can implement custom triggering for any analog sensor type on a per-channel basis using
the flexibility and performance of the FPGA and the numerous arithmetic and comparison function blocks built into the LabVIEW FPGA Module.
Key Features
High-accuracy, high-performance analog measurements for any CompactRIO embedded system, R Series expansion chassis, or NI CompactDAQ chassis
Screw terminals, BNC, D-Sub, spring terminals, strain relief, high voltage, cable, solder cup backshell, and other connectivity options
Available channel-to-earth ground double-isolation barrier for safety, noise immunity, and high common-mode voltage range
CompactRIO Extreme Industrial Certifications and Ratings
Built-in signal conditioning for direct connection to sensors and industrial devices
Visit ni.com/compactrio or ni.com/compactdaq for up-to-date information on module availability, example programs, application notes, and other developer tools.
Back to Top
Ordering Information
For a complete list of accessories, visit the product page on ni.com.
Products
Part Number
Recommended Accessories
Part Number
No accessories required.
Back to Top
Software Recommendations
LabVIEW Professional Development System for Windows
Advanced software tools for large project development
Automatic code generation using DAQ Assistant and Instrument I/O Assistant
Tight integration with a wide range of hardware
Advanced measurement analysis and digital signal processing
Open connectivity with DLLs, ActiveX, and .NET objects
Capability to build DLLs, executables, and MSI installers
Back to Top
Support and Services
2/9
www.ni.com
Support and Services
System Assurance Programs
NI system assurance programs are designed to make it even easier for you to own an NI system. These programs include configuration and deployment services for your NI PXI,
CompactRIO, or Compact FieldPoint system. The NI Basic System Assurance Program provides a simple integration test and ensures that your system is delivered completely
assembled in one box. When you configure your system with the NI Standard System Assurance Program, you can select from available NI system driver sets and application
development environments to create customized, reorderable software configurations. Your system arrives fully assembled and tested in one box with your software preinstalled.
When you order your system with the standard program, you also receive system-specific documentation including a bill of materials, an integration test report, a recommended
maintenance plan, and frequently asked question documents. Finally, the standard program reduces the total cost of owning an NI system by providing three years of warranty
coverage and calibration service. Use the online product advisors at ni.com/advisor to find a system assurance program to meet your needs.
Calibration
NI measurement hardware is calibrated to ensure measurement accuracy and verify that the device meets its published specifications. To ensure the ongoing accuracy of your
measurement hardware, NI offers basic or detailed recalibration service that provides ongoing ISO 9001 audit compliance and confidence in your measurements. To learn more
about NI calibration services or to locate a qualified service center near you, contact your local sales office or visit ni.com/calibration.
Technical Support
Get answers to your technical questions using the following National Instruments resources.
Support - Visit ni.com/support to access the NI KnowledgeBase, example programs, and tutorials or to contact our applications engineers who are located in NI sales
offices around the world and speak the local language.
Discussion Forums - Visit forums.ni.com for a diverse set of discussion boards on topics you care about.
Online Community - Visit community.ni.com to find, contribute, or collaborate on customer-contributed technical content with users like you.
Repair
While you may never need your hardware repaired, NI understands that unexpected events may lead to necessary repairs. NI offers repair services performed by highly trained
technicians who quickly return your device with the guarantee that it will perform to factory specifications. For more information, visit ni.com/repair.
Training and Certifications
The NI training and certification program delivers the fastest, most certain route to increased proficiency and productivity using NI software and hardware. Training builds the skills
to more efficiently develop robust, maintainable applications, while certification validates your knowledge and ability.
Classroom training in cities worldwide - the most comprehensive hands-on training taught by engineers.
On-site training at your facility - an excellent option to train multiple employees at the same time.
Online instructor-led training - lower-cost, remote training if classroom or on-site courses are not possible.
Course kits - lowest-cost, self-paced training that you can use as reference guides.
Training memberships and training credits - to buy now and schedule training later.
Visit ni.com/training for more information.
Extended Warranty
NI offers options for extending the standard product warranty to meet the life-cycle requirements of your project. In addition, because NI understands that your requirements may
change, the extended warranty is flexible in length and easily renewed. For more information, visit ni.com/warranty.
OEM
NI offers design-in consulting and product integration assistance if you need NI products for OEM applications. For information about special pricing and services for OEM
customers, visit ni.com/oem.
Alliance
Our Professional Services Team is comprised of NI applications engineers, NI Consulting Services, and a worldwide National Instruments Alliance Partner program of more than
700 independent consultants and integrators. Services range from start-up assistance to turnkey system integration. Visit ni.com/alliance.
Back to Top
Detailed Specifications
The following specifications are typical for the range 40 to 70 C unless otherwise noted.
Input Characteristics
Number of channels
4 analog input channels
ADC resolution
24 bits
Type of ADC
Delta-sigma (with analog prefiltering)
Sampling mode
Simultaneous
Type of TEDS supported
IEEE 1451.4 TEDS Class II (Interface)
3/9
www.ni.com
Mode input ranges
Mode
Nominal Range(s)
Actual Range(s)
Voltage
60 V, 15 V, 4 V, 1 V, 125 mV 60 V, 15 V, 4 V, 1 V, 125 mV
Current
25 mA
25 mA
4-Wire and 2-Wire Resistance 10 k, 1 k
10.5 k, 1.05 k
Thermocouple
125 mV
125 mV
4-Wire and 3-Wire RTD
Pt 1000, Pt 100
5.05 k, 505
Quarter-Bridge
350 , 120
390 , 150
Half-Bridge
500 mV/V
500 mV/V
Full-Bridge
62.5 mV/V, 7.8 mV/V
62.5 mV/V, 7.8125 mV/V
Digital In
060 V
Open Contact
1.05 k
Conversion time, no channels in TC mode
High speed
10 ms for all channels
Best 60 Hz rejection
110 ms for all channels
Best 50 Hz rejection
130 ms for all channels
High resolution
500 ms for all channels
Conversion time, one or more channels in TC mode
High speed
20 ms for all channels
Best 60 Hz rejection
120 ms for all channels
Best 50 Hz rejection
140 ms for all channels
High resolution
510 ms for all channels
Overvoltage protection
Terminals 1 and 2
30 V
Terminals 3 through 6, across any combination
60 V
Input impedance
Voltage and Digital In modes (60 V, 15 V, 4 V)
1 M
Current mode
<40
All other modes
>1 G
Accuracy
Mode, Range
Gain Error (Percent of Reading) Offset Error (ppm of Range)
Typ (25 C, 5 C), Max (40 to 70 C)
Voltage, 60 V
0.3, 0.4
20, 50
Voltage, 15 V
0.3, 0.4
60, 180
Voltage, 4 V
0.3, 0.4
240, 720
Voltage, 1 V
0.1, 0.18
15, 45
Voltage/Thermocouple, 125 mV
0.1, 0.18
120, 360
Current, 25 mA
0.1, 0.6
30, 100
4-Wire and 2-Wire
Resistance, 10 k 0.1, 0.5
120, 320
4-Wire and 2-Wire 1 Resistance, 1 k
0.1, 0.5
1200, 3200
4-Wire and 3-Wire RTD, Pt 1000
0.1, 0.5
240, 640
4-Wire and 3-Wire RTD, Pt 100
0.1, 0.5
2400, 6400
Quarter-Bridge, 350
0.1, 0.5
2400, 6400
Quarter-Bridge, 120
0.1, 0.5
2400, 6400
Half-Bridge, 500 mV/V
0.03, 0.07
300, 450
Full-Bridge, 62.5 mV/V
0.03, 0.08
300, 1000
4/9
www.ni.com
Accuracy
Mode, Range
Gain Error (Percent of Reading) Offset Error (ppm of Range)
Typ (25 C, 5 C), Max (40 to 70 C)
Full-Bridge, 7.8 mV/V
0.03, 0.08
Cold-junction compensation sensor accuracy
2200, 8000
1C typ
Stability
Mode, Range
Gain Drift (ppm of Read-ing/C) Offset Drift (ppm of Range/C)
Voltage, 60 V
20
0.2
Voltage, 15 V
20
0.8
Voltage, 4 V
20
3.2
Voltage, 1 V
10
0.2
Voltage/Thermocouple, 125 mV
10
1.6
Current, 25 mA
15
0.4
4-Wire and 2-Wire Resistance, 10 k 15
4-Wire and 2-Wire Resistance, 1 k
15
30
4-Wire and 3-Wire RTD, Pt 1000
15
4-Wire and 3-Wire RTD, Pt 100
15
60
Quarter-Bridge, 350
15
120
Quarter-Bridge, 120
15
240
Half-Bridge, 500 mV/V
20
Full-Bridge, 62.5 mV/V
20
Full-Bridge, 7.8 mV/V
20
Input noise in ppm of Range rms
Mode, Range
Conversion Time
High speed Best 60 Hz re-jection Best 50 Hz re-jection High resolution
Voltage, 60 V
7.6
1.3
1.3
0.5
Voltage, 15 V
10.8
1.9
1.9
0.7
Voltage, 4 V
10.8
2.7
2.7
1.3
Voltage, 1 V
7.6
1.3
1.3
0.5
Voltage/Thermocouple, 125 mV
10.8
1.9
1.9
1.0
Current, 25 mA
10.8
1.9
1.9
1.0
4-Wire and 2-Wire Resistance, 10 k 4.1
1.3
0.8
0.3
4-Wire and 2-Wire Resistance, 1 k
7.1
1.8
1.2
0.7
4-Wire and 3-Wire RTD, Pt 1000
7.6
1.7
1.1
0.4
4-Wire and 3-Wire RTD, Pt 100
10.8
1.9
1.9
0.9
Quarter-Bridge, 350
5.4
1.0
1.0
0.7
Quarter-Bridge, 120
5.4
1.0
1.0
0.7
Half-Bridge, 500 mV/V
3.8
0.5
0.5
0.2
Full-Bridge, 62.5 mV/V
5.4
1.0
1.0
0.8
Full-Bridge, 7.8 mV/V
30
4.7
4.7
2.3
Input bias current
<1 nA
INL
15 ppm
CMRR (in = 60 Hz)
>100 dB
NMRR
Best 60 Hz rejection
90 dB at 60 Hz
5/9
www.ni.com
Best 50 Hz rejection
80 dB at 50 Hz
High resolution
65 dB at 50 Hz and 60 Hz
Excitation level for Half-Bridge and Full-Bridge modes
Mode
Half-Bridge
Full-Bridge
Load Resistance ()
Excitation (V)
700
2.5
240
2.0
350
2.7
120
2.2
Excitation level for Resistance, RTD, and Quarter-Bridge modes
Load Resistance ()
Excitation (mV)
120
50
350
150
1k
430
10 k
2200
384,716 hours at 25 C; Bellcore Issue 2, Method 1, Case 3, Limited Part
Stress Method
MTBF
Note Contact NI for Bellcore MTBF specifications at other temperatures or for MIL-HDBK-217F specifications.
Power Requirements
Power consumption from chassis
Active mode
750 mW max
Sleep mode
25 W max
Thermal dissipation (at 70 C)
Active mode
625 mW max
Sleep mode
25 W max
Physical Characteristics
Note For two-dimensional drawings and three-dimensional models of the C Series module and connectors, visit ni.com/dimensions and search by module number.
Spring-terminal wiring
18 to 28 AWG copper conductor wire with 7 mm (0.28 in.) of insulation stripped
from the end
Weight
156 g (5.5 oz)
Safety
If you need to clean the module, wipe it with a dry towel.
Safety Voltages
Connect only voltages that are within the following limits.
Channel-to-channel
Continuous
250 VAC, Measurement Category II
Withstand
1,390 VAC, verified by a 5 s dielectric withstand test
Channel-to-earth ground
Continuous
250 VAC, Measurement Category II
Withstand
2,300 VAC, verified by a 5 s dielectric withstand test
Zone 2 hazardous locations applications in Europe channel-to-channel and
channel-to-earth ground
60 VDC, Measurement Category I
Measurement Category I is for measurements performed on circuits not directly connected to the electrical distribution system referred to as MAINS 2 voltage. This category is for
measurements of voltages from specially protected secondary circuits. Such voltage measurements include signal levels, special equipment, limited-energy parts of equipment,
circuits powered by regulated low-voltage sources, and electronics.
Caution Do not connect to signals or use for measurements within Measurement Categories II, III, or IV.
6/9
www.ni.com
Measurement Category II is for measurements performed on circuits directly connected to the electrical distribution system. This category refers to local-level electrical
distribution, such as that provided by a standard wall outlet (e.g., 115 V for U.S. or 230 V for Europe). Examples of Measurement Category II are measurements performed on
household appliances, portable tools, and similar products.
Caution Do not connect to signals or use for measurements within Measurement Categories III or IV.
Hazardous Locations
U.S. (UL)
Class I, Division 2, Groups A, B, C, D, T4; Class I, Zone 2, AEx nA IIC T4
Canada (C-UL)
Class I, Division 2, Groups A, B, C, D, T4; Class I, Zone 2, Ex nA IIC T4
Europe (DEMKO)
Ex nA IIC T4
Safety Standards
This product is designed to meet the requirements of the following standards of safety for electrical equipment for measurement, control, and laboratory use:
IEC 61010-1, EN 61010-1
UL 61010-1, CSA 61010-1
Note For UL and other safety certifications, refer to the product label or the Online Product Certification section.
Electromagnetic Compatibility
This product meets the requirements of the following EMC standards for electrical equipment for measurement, control, and laboratory use:
EN 61326 (IEC 61326): Class A emissions; Industrial Immunity
EN 55011 (CISPR 11): Group 1, Class A emissions
AS/NZS CISPR 11: Group 1, Class A emissions
FCC 47 CFR Part 15B: Class A emissions
ICES-001: Class A emissions
Note For the standards applied to assess the EMC of this product, refer to the Online Product Certification section.
Note For EMC compliance, operate this device with shielded cables.
CE Compliance
This product meets the essential requirements of applicable European Directives, as amended for CE marking, as follows:
2006/95/EC; Low-Voltage Directive (safety)
2004/108/EC; Electromagnetic Compatibility Directive (EMC)
Note For the standards applied to assess the EMC of this product, refer to the Online Product Certification section.
Online Product Certification
Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for this product, visit
ni.com/certification, search by module number or product line, and click the appropriate link in the Certification column.
Shock and Vibration
To meet these specifications, you must panel mount the system.
Operating vibration
Random (IEC 60068-2-64)
5 grms, 10 to 500 Hz
Sinusoidal (IEC 60068-2-6)
5 g, 10 to 500 Hz
Operating shock (IEC 60068-2-27)
30 g, 11 ms half sine, 50 g, 3 ms half sine, 18 shocks at 6 orientations
Environmental
National Instruments C Series modules are intended for indoor use only but may be used outdoors if installed in a suitable enclosure. Refer to the manual for the chassis you are
using for more information about meeting these specifications.
Operating temperature (IEC 60068-2-1, IEC 60068-2-2)
40 to 70 C
Storage temperature (IEC 60068-2-1, IEC 60068-2-2)
40 to 85 C
Ingress protection
IP 40
Operating humidity (IEC 60068-2-56)
10 to 90% RH, noncondensing
Storage humidity (IEC 60068-2-56)
5 to 95% RH, noncondensing
Maximum altitude
2,000 m
Pollution Degree (IEC 60664)
Environmental Management
National Instruments is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous
substances from our products is beneficial not only to the environment but also to NI customers.
7/9
www.ni.com
For additional environmental information, refer to the NI and the Environment Web page at ni.com/environment. This page contains the environmental regulations and directives
with which NI complies, as well as other environmental information not included in this document.
Waste Electrical and Electronic Equipment (WEEE)
EU Customers At the end of their life cycle, all products must be sent to a WEEE recycling center. For more information about WEEE recycling centers and National
Instruments WEEE initiatives, visit ni.com/environment/weee.htm.
Calibration
You can obtain the calibration certificate for this device at ni.com/calibration.
Calibration interval
1 year
2-Wire Resistance mode accuracy depends on the lead wire resistance. This table assumes 0 of lead wire resistance.
MAINS is defined as the (hazardous live) electrical supply system to which equipment is designed to be connected for the purpose of powering the equipment. Suitably rated
measuring circuits may be connected to the MAINS for measuring purposes.
Back to Top
8/9
www.ni.com
Pinouts/Front Panel Connections
Back to Top
2010 National Instruments. All rights reserved. CompactRIO, CVI, FieldPoint, LabVIEW, National Instruments, National Instruments Alliance Partner, NI, ni.com, NI CompactDAQ, and SignalExpress are
trademarks of National Instruments. The mark LabWindows is used under a license from Microsoft Corporation. Windows is a registered trademark of Microsoft Corporation in the United States and other countries.
Other product and company names listed are trademarks or trade names of their respective companies. A National Instruments Alliance Partner is a business entity independent from National Instruments and has
no agency, partnership, or joint-venture relationship with National Instruments.
My Profile | RSS | Privacy | Legal | Contact NI 2014 National Instruments Corporation. All rights reserved.
9/9
www.ni.com