Digital Multimeters
4 1/2 Digit DMM Series for Diverse Applications
R6441 Series
R6441A: DMM with low-price basic model
R6441B: Multi-functional DMM with Frequency Measurements
R6441C: DMM with Terminals Dedicated for
Floating Current Measurement
R6441D: DMM with Enhanced Current
Measurement
(Photo is R6441C)
R6441 Series
Digital Multimeters
New R6441 series digital multimeters were designed for diverse applications. The series is provided with a variety of
interfaces for use in R&D sections and production lines; it
ensures battery operation for field applications. With dualchannel input and dual display, the R6441 series provides a
new measurement environment.
The series includes three models: R6441A low-price basic
model, R6441B with enhanced AC measurement functions,
R6441C with enhanced very small current and floating method
current measurement functions, and R6441D low-price model
with enhanced current measurement functions.
Memory Card (SRAM Card Conforming to JEIDA
Ver.4) Ensures Data Compatibility with Personal
Computers
Maximum Display of 199999 (with a Sampling Rate
of 2.5 Times/Second) and Maximum Sampling Rate of
80 Times/Second (with Maximum Display of 1999)
Large Easy-to-Read Electron-Ray Indicator Tube
AC Voltage and Current Measurement with True RMS
(R6441B/6441C/6441D), AC + DC Measurement
(R6441B) and Frequency Measurement (R6441B)
Standard RS-232C Interface and Optional GPIB
Interface and BCD Data Output Units
252
Various Interfaces Can be Implemented for Automated
Measurement
Optional Battery Unit Allows the Use as a HighPerformance DMM for Field Measurement
Diverse and Combination Calculation Functions
Memory Function for Panel Settings (Recalls Previous
Condition Settings at Power On)
High-Speed Analog Bar Graph with a Sampling Rate
of 80 Times/Second is Available for Instantaneous
Trendy Check (R6441A)
Wide Power Range (90 to 250 V)
Input Terminal Dedicated for Floating DC/AC Current
(in 2- and 5-A Ranges) (R6441C)
Digital Multimeters
4 1/2 Digit DMM Series for Diverse Applications
R6441 Series
Specifications
Measurement accuracy: 23 5C, 85% RH or less (75% or less is guaranteed
for 1 year at 20-M and 200-M ranges.) The display value is % of reading
digits.
Temperature coefficient: 0.1 (measurement accuracy)/C at 0 to 50C.
The display value is (% of reading digits)/C.
DC voltage measurement
d:digit
Range
20 mV
200 mV
2000 mV
1 V
10 V
100 V
Maximum display
20 V
200 V
1000 V
1 mV
10 mV
100 mV
19999
Resolution
Measurement accuracy
0.04%5d
Input impedance
10999
20 k
Resolution
Measured applied current
20 M
Measurement accuracy
100 m
10
100
1 k
1 mA
100 A
10 A
1 A
100 nA
10 nA
0.07%100d
0.07%20d
Open circuit voltage
Maximum allowable
applied voltage
0.1%20d 0.3%50d
7.5 V or less
500 V
R6441A/6441B
Range
20 mA
200 mA
Maximum display
2000 mA
10 A
19999
1 A
Resolution
Normal mode noise rejection ratio
ratio (unbalanced impedance of 1 k)
2000 k
10 m
DC current measurement
Effective common mode noise rejection
200 k
19999
11.1M1% 10.1M1% 10.0M1%
DC voltage noise rejection ratio
10999
10 A
100 A
0.2%5d
Measurement accuracy
1 mA
0.6%5d
50/60 Hz 0.1%, DC
50/60 Hz 0.1%
Input terminal resistance
1.5 or less *1
Approx. 60 dB
0 dB
Overcurrent protection
0.5 A/250 V IEC 127 sheet 1
15 A/250 V with 10000-A interrupting capacity
Protected by a quick-blowing fuse
Protected by a quick-blowing fuse
Approx. 120 dB
Approx. 60 dB
FAST
MID
SLOW
0.04 or less *1
*1 The resistance of the protection fuse is excluded.
R6441C/6441D
AC voltage measurement
R6441A (with average measurement and rms value display)
Range
200 mV
2000 mV
10 V
200 V
700 V
Resolution
7099
Measurement accuracy
100 V
1 mV
10 mV
100 mV
0.6%40d
0.6%35d
0.6%45d
0.6%45d
0.6%35d
19999
Resolution
20 to 45 Hz
45 to 20 kHz
0.25%35d
0.25%30d
0.25%40d
0.25%40d 0.25%30d
20 to 30 kHz
0.8%40d
0.8%35d
0.8%45d
0.8%45d
0.8%35d
30 to 100 kHz
5%50d
5%50d
5%50d
5%50d
5%50d
1.1 M 10%, 100 pF or less
Input impedance
Maximum allowable
applied voltage
Response time
(0.1% or less of the final value in the same range)
R6441B (True RMS, AC, AC+DC) / R6441C/6441D (True RMS, AC)
With an input of 5% or more of the full scale
2000 mV
20 V
200 V
19999
10 V
Resolution
100 V
1 mV
10 mV
20 Hz to 45 Hz
0.6%35d
45 Hz to 20 kHz
0.2%30d
20 kHz to 30 kHz
0.5%30d
30 kHz to 100 kHz
4%50d
Overcurrent protection
1.1 M10%, 100 pF or less
3:1 at the full scale
Resolution
10 m
100 m
10
Measured applied current
3 mA
1 mA
100 A
10 A
0.07%2d
Resolution
1 to 5 kHz
Input terminal resistance
100 mV
Open circuit voltage
7.5 V or less
Maximum allowable
applied voltage
500 V
2 or less *2
0.1 or less *2
6 A/250 V
with 10000-A
interrupting capacity
Protected by a
quick-blowing fuse
0.5 A/250 V IEC 127 sheet 1
Protected by a quick-blowing fuse
Overcurrent
19999
10999
0.8%40d
0.8%40d
5.0%40d
5.0%40d
1.5 or less *1
0.04 or less *1
0.5 A/250 V IEC 127 sheet 1
15 A/250 V with 10000-A interrupting
Protected by a quick-blowing fuse capacity Protected by a quick-blowing fuse
Approx. 4 seconds for AC current and approx. 2 seconds for AC current filter
(0.1% or less of the final value in the same range)
*1 The resistance of the protection fuse is excluded.
R6441B (True RMS, AC, AC+DC)
With an input of 5% or more of the full scale
20 M
200 M
100
1 k
10 k
100 nA
10 nA
1 A
2%50d 2%5d
102 or less *2
* The AC current filter is 300 Hz to 5 kHz. (Display with input switching is not possible when
an AC current filter is used.)
2000 k
0.1%2d
4999
1 mA
10 A
accuracy
19999
Measurement accuracy 0.07%10d
0.2%5d
1999
100 A
1 mA
7099
Resistance measurement
Maximum display
10 A
10 A
Range
200 mA
10 A
Maximum display
10 A
1 mA
(0.1% or less of the final value in the same range)
200 k
1 A
100 nA
200 mA
700 V
Approx. 1 second
20 k
10 nA
Range
20 Hz to1 kHz
800 Vrms, 1100 Vpeak, 10 VHz
2000
1 nA
2000 mA*1 5 A *1
Maximum display
Measurement
200
200 mA
AC current measurement
R6441A (with average measurement and rms value display)
Response time
Crest factor
Maximum allowable
applied voltage
Response time
20 mA
* When the floating method for 2000-mA and 5-A ranges and the null function are used.
protection
Input impedance
Range
100 pA
Input terminal resistance Approx. 10 k or less*2
* The frequency range of the VAC filter is 300 Hz to 100 kHz.
200 mV
2000 A
19999
*2 The resistance of the protection fuse is excluded.
Approx. 4 seconds for VAC voltage and approx. 2 seconds for VAC voltage filter
Maximum display
200 A
*1 Mounted only on the R6441C.
800 Vrms, 1100 Vpeak, 107 VHz
Range
2 A *1 20 A *1
Range
Maximum display
20 V
Maximum display
Measurement
accuracy
2000
0.04% 2d
1100 V (all ranges, continuous)
Sampling rate
200
Range
Maximum display
* When the null function is used
1 G or more
Maximum allowable
applied voltage
In-circuit resistance measurement
0.3%5d 3.0%10d
Resolution
19999
10999
Measurement 20 Hz to 1 kHz
0.8%40d
0.8%40d
accuracy
5.0%40d
1 kHz to 5 kHz
Crest factor
5.0%40d
3:1 at the full scale
Input terminal resistance
1.5 or less *1
0.04 or less *1
Overcurrent
0.5 A/250 V IEC 127 sheet 1
15 A/250 V with 10000-A interrupting capacity
protection
Protected by a quick-blowing fuse
Protected by a quick-blowing fuse
Response time
Approx. 1 second (0.1% or less of the final value in the same range)
*1 The resistance of the protection fuse is excluded.
* When the null function is used
253
Digital Multimeters
Data Sharing with Personal Computers via Memory Cards
R6441 Series (Continued From Previous Page)
R6441C/6441D (True RMS, AC)
With an input of 5% or more of the full scale
200 A
Range
2000 A
Maximum display
Resolution
10 nA
200 mA
2000 mA *1
100 nA
1 A
10 A
5 A *1
19999
4999
100 A
1 mA
0.8% 40d
Measurement 20Hz to 500Hz
accuracy
20 mA
19999
2% 40d
5.0% 40d
500Hz to 5kHz
Crest factor
3:1 at the full scale
Approx. 102 or less *2
0.1 or less *2
6 A/250 V with 10000-A
0.5 A/250 V IEC 127 sheet 1
interrupting capacity
Protected by a
Protected by a quick-blowing fuse
quick-blowing fuse
Approx. 1 second (0.1% or less of the final value in the same range)
Input terminal resistance
Overcurrent
protection
Response time
2 or less *2
* Floating method is used for 200-mA and 5-A ranges.
*1 Mounted only on the R6441C.
*2 The resistance of the protection fuse is excluded.
Frequency measurement
R6441B
Range
20 Hz
200 Hz
Maximum display
2 kHz
20 kHz
200 kHz
1 Hz
10 Hz
19999
Measurement accuracy
1 mHz
10 mHz
100 mHz
0.02% 2d
Measurement time
General specifications
Measurement method: Integrating type
Input method: Floating type
Range switching: Auto and manual
Data display: 5-digit decimal, 7-segment electron ray indicator tube
Overinput indication: "OL" is displayed for inputs out of the rated
measurement range.
Low-battery indication: If the battery power voltage drops to below
the rated voltage, a low-battery mark is indicated in the display
section.
Dielectric strength: Withstands 450 V continuously applied between
the COM terminal and chassis and between the Com terminal and
AC power line.
Operating environment:
Operating temperature: 0 to 50C
(0 to 40C when the battery is mounted)
Operating humidity: 85% RH or less
Storage temperature: -25 to 70C
(-20 to 50C when the battery is mounted)
Power consumption: 15 VA or less
AC power: Specified at time of ordering.
Option No.
Standard
32
42
44
Power voltage (V)
90 to 110
103 to 132
198 to 242
207 to 250
* Waveform : Sine, square
Duty ratio : 3 or less
Sampling mode: Free-run
Measurement time
Function
SLOW (4 1/2)
DC power supply: 6-hour continuous operation is possible by means
of the R15807 battery unit.
Dimensions: Approx. 212 (W) 88 (H) 310 (D) mm
Mass: 2.2 kg maximum (main unit), 3.5 kg maximum (with options)
Accessories:
FAST (3 1/2)
MID (4 1/2)
DC voltage measurement
12.5 (80)
100 (10)
400 (2.5)
AC voltage measurement (AC coupling)
12.5 (80)
100 (10)
400 (2.5)
Product name
A01402
A01034
Resistance measurement
12.5 (80)
100 (10)
400 (2.5)
Model
Power cable
Input cable x1
DC current measurement
12.5 (80)
100 (10)
400 (2.5)
AC current measurement
12.5 (80)
100 (10)
400 (2.5)
Frequency measurement (R6441B)
210 (4.7)
300 (3.3)
600 (1.5)
Conductive measurement
12.5 (80)
100 (10)
400 (2.5)
Diode measurement
12.5 (80)
100 (10)
400 (2.5)
Unit [ms] (times/second)
Conductive measurement: Measurement range of 200 and
continuity judgment value of 20
Other specifications are the same as those for the 200 range for
resistance measurement.
Diode measurement: Measurement range of 2000 mV
Other specifications are the same as those for the 2000 range for
resistance measurement.
Sampling rate
Number of measurements (times/second)
FAST
MID
SLOW
80
10
2.5
Calculation function: Null, smoothing, dB/dBm, scaling, MAX/MIN,
comparator
254
Standard accessories: RS-232C, baud rate of 9600, 4800, 2400, 1200,
600, and 300
Optional accessories
A08316
Alligator clip adapter
A08317
Miniature clip adapter
A01001
Input cable
A01265
RS-232C cable (For 1 m, 250- and 9-pin (DMM))
A09507
SRAM card (64 kbytes)
TR1116
DC high-voltage probe
TR1111
Terminal adapter
A02464
EIA rack mount kit (twin)
A02463
EIA rack mount kit
A02264
JIS rack mount kit (twin)
A02263
JIS rack mount kit
R16215
Carrying bag