Experiment
Displacement Measurement using Potentiometer type
Displacement sensor
THEORY:
Displacement measurement can be of two types: contact and
noncontact types. Besides the measurement principles can be
classified into two categories: electrical sensing and optical
sensing. In electrical sensing, passive electrical sensors are used
variation of either inductance or capacitance with displacement is
measured. On the other hand the optical method mainly works on
the principle of intensity variation of light with distance.
Potentiometers are simplest type of displacement sensors. They
can be used for linear as well as angular displacement
measurement. They are the resistive type of transducers and the
output voltage is proportional to the displacement. They are the
resistive type of transducers and the output voltage is proportional
to the displacement and is given by:
where xi is the input displacement, xt is the total displacement and
E is the supply voltage.
APPARATUS:
1. Potentiometer Type Displacement Sensor
2. Cantilevered Aluminum Beam
3. Loads
4. Height Gauge
PROCEDURE:
CALIBRATION:
1. Position the height gauge and Potentiometer type
displacement sensor at one point.
2. Apply tip loads and measure the displacement from the
height gauge.
3. Measure the corresponding voltage from the displacement
sensor.
4. Calibrate the voltage against displacement by fitting it with
the straight line.
LOAD VS DISPLACEMENT:
1. Place the sensor at a position other than the previous one.
2. Apply different loads at the tip.
3. Measure the displacement using the sensor at the given point.
4. Do the same for 2 other points along the beam.
5. Plot the load vs. displacement curve.
OBSERVATIONS:
1. CALIBRATION
VOLTAGE (IN mV)
DISPLACEMENT (IN mm)
-97.0
0
855
5
1900
10
2940
15
3960
20
4990
25
6040
30
7050
35
8050
40
2. LOAD VS. VOLTAGE
i) Distance from fixed support = 30cm
LOAD (in gm) VOLTAGE (in mV)
0.00 -1040
119.15 610
239.85 22
360.80 620
482.65 1300
ii) Distance from fixed support = 25cm
LOAD (in gm) VOLTAGE (in mV)
0 -1050
119.15 -640
239.85 -180
360.8 357
482.65 920
iii) Distance from fixed support = 20cm
LOAD (in gm) VOLTAGE (in mV)
0 -925
119.15 -672
239.85 391
360.8 9
482.65 421
RESULT:
1. CALIBRATION
Voltage vs. Displacement
9000
8000 8050
7000 7050
6000 6040
5000 4990 Voltage vs. Displacement
4000 3960
Linear (Voltage vs.
3000 2940 Displacement)
2000 1900
1000 855
0 -97
0 5 10 15 20 25 30 35 40
-1000
Average slope of the graph = 202.675 mV/mm
!
Therefore, = 4.934 /
!
2. LOAD VS. DISPLACEMENT
1500
1000
500
Distance 30
0 Distance 25
0 100 200 300 400 500 600
Distance 20
-500
-1000
-1500
SUBMITTED BY:
SHUBHAM (150696)
RAJAT PANDEY (150555)
SHREY AGARWAL (150692)
YAZUR GUPTA (150842)