GE Power & Water
Water & Process
Technologies
Color
Range: as PtCo (unit)
Method: DR 3900 (Spectrophotometric)
Testing Time: 5 min
1. Select the Program Color using the DR3900.
Blank Preparation
2. Fill 50mm cell with distilled water sample. This will be the Blank.
Reading
3. Wipe then Insert the blank into the cell holder and press Zero.
4. Fill 50mm cell with “Water sample”. Wipe it and insert it into the cell holder and
press Read.
5. Results are in PtCo units.
Notes:
Always use the distilled water and Press Zero, before taking sample readings,
especially in case of multiple samples to insure the accurate results.
Reagents Required
No Reagents needed to perform this test.