0% found this document useful (0 votes)
95 views1 page

Water Color Testing Guide

This document provides instructions for using a DR3900 spectrophotometer to test the color of water samples in platinum-cobalt (PtCo) units. It outlines the 3 step process: 1) select the color program, 2) prepare a distilled water blank and zero the instrument, 3) insert the water sample and take a reading, with results displayed in PtCo units. Notes emphasize zeroing the instrument with distilled water for accurate results when testing multiple samples. No reagents are required for this color test.

Uploaded by

teena6506763
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOC, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
95 views1 page

Water Color Testing Guide

This document provides instructions for using a DR3900 spectrophotometer to test the color of water samples in platinum-cobalt (PtCo) units. It outlines the 3 step process: 1) select the color program, 2) prepare a distilled water blank and zero the instrument, 3) insert the water sample and take a reading, with results displayed in PtCo units. Notes emphasize zeroing the instrument with distilled water for accurate results when testing multiple samples. No reagents are required for this color test.

Uploaded by

teena6506763
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOC, PDF, TXT or read online on Scribd

GE Power & Water

Water & Process


Technologies

Color

Range: as PtCo (unit)


Method: DR 3900 (Spectrophotometric)
Testing Time: 5 min

1. Select the Program Color using the DR3900.

Blank Preparation

2. Fill 50mm cell with distilled water sample. This will be the Blank.

Reading

3. Wipe then Insert the blank into the cell holder and press Zero.

4. Fill 50mm cell with “Water sample”. Wipe it and insert it into the cell holder and
press Read.

5. Results are in PtCo units.

Notes:

 Always use the distilled water and Press Zero, before taking sample readings,
especially in case of multiple samples to insure the accurate results.

Reagents Required

No Reagents needed to perform this test.