CHROMATOGRAPHY
Preparative vs Resolution vs Speed
Gas Chromatography
Achieve separation by using suitable;
Columns with proper stationary phase, diameter of
column, stationary phase loading, column length.
Injection modes/ oven temperatures to optimize the of
loading and separation of the sample mixture on the
column.
Temperature (or pressure) programs for the column
Detector that is suitable for the analyte(s)of interest
2
Instrumentation
Instruments;
Carrier Gas
Flow regulators and meters
Sample injection system
Columns & ovens
Detectors
3
Schematic Diagram of Gas
Chromatograph
4
H2 inlet
(detector) N2 inlet
Air inlet
(make-up gas)
(detector)
He inlet
(carrier gas)
Process Flow Schematic
Detector (flame
Sample injection ionization
detector or FID)
Carrier gas
Air
(nitrogen or
helium) Hydrogen
Long Column (30 m)
Carrier Gas (mobile phase)
Requirements:
It should be inert and available at low cost
High purity
Easily available
Less risk of explosion or fire hazards
Pressure:
-Inlet 10 to 50 psi
-packed column 25 to 150 mL/min.
- capillary column 1 to 25 mL/min.
Sample injection port
Calibrated micro syringes are used to inject liquid
sample
Purge: volatile components are removed from
sample by gentle heating
Rubber or silicone diaphragm(septum)
Sample port T: over 50°C than column
Packed C: sample sizes,1 to 20 μL
Capillary C: 10 to 30 μL
splitter is used to deliver a fraction of
injection (1:50 to 1:500)
Avoid over loading
Slow injection & oversized samples cause band
spreading & poor resolution
Split Injection:
Injection is split, with only a fraction of
the sample (usually 1% - 20%) actually
making it to the column
The most common method of injecting
samples onto small diameter, open-tubular
columns.
Even for injections 20 μL, only a
fraction (adjustable) makes it on
to the column
Not for analytes with a wide range of
boiling points
some may be swept out the split
vent before they are volatilized
Splitless Injection:
Sample - vaporized in the injector and ALL of
the sample is swept onto the column by the
carrier gas
Relatively small samples (≤10 μL)
Sample spends a large amount of time in the
injector
Best for trace (1 -100 ppm range)
concentrations of high boiling point analytes in
low boiling point solvents
extra time in the injector helps
volatilize the analytes.
On-Column Injection:
used widely in packed-column GC, less
in capillary GC
sample - deposited directly on the
column
Good for thermally unstable
compounds
Good for quantitative analysis at low
concentrations
entire sample reaches the detector
Smaller injections (Capillary GC)
Column Configurations
Two types of columns are used in gas
chromatography, packed and open tubular
or capillary.
Packed column length from less than 2 m to
5m
Capillary columns from few m to 100 m
They are constructed of stainless steel, glass,
fused silica, or Teflon.
Column
Types of columns
1.packed columns
2. Open tubular or capillary.
Packed column-3m Capillary column- 30m
PLOT
Column ovens
Column temperature is very important in GC
The column is ordinarily housed in a thermostatic oven.
they are usually formed as coils having diameters of 10 to 30
cm.
The optimum column temperature depends upon the boiling
point of the sample and the degree of separation required.
Roughly, a temperature equal to or slightly above the average
boiling point of a sample results in a reasonable elution time
(2 to 30 min).
Length vs. resolution
Diameter vs. resolution
Stationary phase thickness vs.
resolution
Sample capacity vs. resolution
Sample capacity: the amount of sample that can be
injected onto a column without overloading. Often
expressed as grams of sample per gram of packing.
Overloading is defined as the point at which the sample
mass injected makes the column efficiency decrease by
10% from its normal value; sometimes called sample
loading.
packed (preparative, larger capacity, low resolution)
capillary (analytical, smaller capacity, high resolution)
Capillary column
Pack column
Column Temperature
Temperature too high, causes co-elution: poor resolution but faster separation
Temperature too low, longer elution times: compromise temperature or program
The Stationary Phase
requirements are:
Low vapor pressure
Thermal stability
Low viscosity (for fast mass transfer)
High selectivity for compounds of interest
Detectors
Detect the difference between a pure carrier gas & eluted compound
Ideal detector:
High sensitivity to even small concentration
linearity, ie, less response to low concentration &proportional
response to high concentration
Large linear dynamic range
Compatible with the column, carrier gas, solvent, etc.
Useful at a range of temperatures
Good stability and reproducibility
Rapid response time
Easy to use
Stable, Predictable response
Inexpensive
operation from RT to 400 oC
Types of detectors
1. Thermal Conductivity Detector(TCD)
2. Flame Ionization Detector(FID)
3. Atomic Emission Detector(AED)
4. Electron Capture Detector(ECD)
5. Nitrogen Phosphoroes Detector(NPD)
6. Photo Ionization Detector(PID)
7. Flame Photometric detector(FPD)
8. Electrolytic conductivity detector (Hall detector)
9. Absolute Mass Detector(AMD)
10.Thermionic Detector(TD)
Flame Ionization Detector (FID)
Most widely used, Air-H2 flame
Number of ions depends on
number of reduced (methylene)
carbons in molecule
The positive ions will be attracted to the
cylindrical cathode.
Negative ions and electrons will be attracted to
the jet anode.
Organic compounds Produces ions and
electrons pyrolyzed(temp of flame) burner
tip and electrode
Ions &electrons move to ward the collector
less sensitive for non-hydrocarbon groups
Insensitive to noncombustible gases(CO2, SO2,
NO2 and H2O)
Insensitive to functional group Adv: High sensitivity, low noise ,wide
(carbonyl, alcohol, halogen and amine) linear range, easy to use, fast response
D.Adv: Destroy the sample
Thermal Conductivity Detector (TCD)
Element (platinum, gold or tungsten wire) is
electrically heated at constant power
– Temperature depends on thermal
conductivity (He & H)of surrounding gas.
Hydrogen and helium have higher thermal
conductivity and carrier gas provide best Thermal conductivity detector
sensitivity cell
Six times greater than the Organic compounds
Poorer sensitivity than FID, but more universal
Advantages: simplicity, large range,
inexpensive,linearity is excelent.
organic & inorganic species
DA: low sensitivity ng/mL
Arrangement of the twin detectors
Electron Capture Detectors (ECD)
The sample elute from a column is passed over a
radioactive β-emitter(nickel-63)
Selectively to halogen-containing organic sample
like pesticides and, polychlorinated biphenyls
Ni-63: radioactive β-emitter-- electron --ionization
of carrier gas (N2)
High electronegative group (halogen, peroxide,
quinones and nitro group) in the sample capture
the electron
Highly selective and sensitive, nondestructive
Insensitive to amines, alcohols and Hydrocarbons
AD: High sensitivity, analyse the polyhalogenated
organic compounds
Small linear range
Thermionic detector (nitrogen phosphorus
detector)
N or P containing organic compounds
phosphorus atom is approximately ten times
greater than to a nitrogen atom and 104 to 106 larger
than a carbon atom.
Compared with the FID , the thermionic detector
is approximately 500 times more sensitive to
phosphorus-containing compounds and 50 times
more sensitive to nitrogen bearing species.
Column effluant + H2 +air(hot gas) electrically
hearted Rb2SiO4 (rubidium silicate) bead at 180 V
plasma (600 – 800°C ) ions to determine
compounds
useful for detecting and determining the many
phosphorus-containing pesticides.
Photo Ionization Detector (PID)
UV light
(10.2 eV H2 or 11.7 eV Ar lamp)
photoionization of molecular
current to flow between based
electrodes
Most sensitive for Aromatic and S, P
easily photoionized molecules
Linear range is high
Qualitative analysis:
Retention time data should be useful for
identification of mixtures
Comparing the retention time of the sample
as well as the standard
Checking the purity of a compound: compare
the standard and sample
Additional peaks are obtained…..impurities
are present….compound is not pure
Quantitative analysis:
Direct comparison method:
comparing the area of the peak, peak height, width of peak.
Calibration curves:
standards of varying concentration are used determine peak areas
Internal standard method:
A known concentration of the internal standard is added separately
to the standard solution
The peak area ratio of sample and internal standard….
unknown concentration is easily determined
Elemental analysis
Determination of C,H ,O ,S and N .
Determination of mixture of drugs
Isolation and identification of drugs
Isolation and identification of mixture of
components(amino acids ,plant extracts ,volatile
oils)
Column Bleed
At high temperature stationary phase may vaporize into the carrier
gas, resulting in column “bleed:
GC - Derivatization
Why is chemical derivatization needed?
GC is best for separation of volatile compounds
which are thermally stable.
Not always applicable for compounds of high
molecular weight or containing polar functional
groups. These groups are difficult to analyze by GC
either because they are not sufficiently volatile, tail
badly, are too strongly attracted to the stationary
phase, thermally unstable or even decomposed.
GC - Derivatization
Chemical derivatization prior to analysis is generally done to:
increase the volatility and decrease the polarity of
compounds;
reduce thermal degradation of samples by increasing their
thermal stability;
increase detector response
improve separation and reduce tailing
Derivatizing Reagents: Common derivatization methods can be
classified into 4 groups depending on the type of reaction
applied:
Silylation
Acylation
Alkylation
Esterification
Multidimensional GC
The use of two or more columns to resolve a sample.
• Normal column chromatography frequently fails to
resolve all components in a complex mixture.
Chromatographic analysis
Best - single column single analysis
Second best - two separate assays using
different columns or conditions
Last resort - multi dimensional GC - when
Sample is limited
Time is limited
Equipment is limited
Approaches
Enrichment
• Used to increase amounts of trace components
Approaches
Heart cutting
Grabs an unresolved portion of a sample for improved separation
Approaches
Back flushing
Reverse column flow to drive off highly retained components
Chiral Separations
Chromatography with a chiral (optically active) stationary phase
is one of the few ways to separate enantiomers.
Eg: -Cyclodextrin a cyclic sugar made of seven glucose molecules.
Programmed temperature (120 oC – 200 oC) chiral separation on a 0.25 mm x 25 m open tubular column with a
stationary phase containing 10 wt% fully methylated -Cyclodextrin chemically bonded to dimethyl polysiloxane.
Helium is the most common carrier gas and is compatible with most detectors.
For a flame ionization detector, N2 gives a lower detection limit than He.
H2, He, and N2 give essentially the same optimal plate height (0.3 mm) at
significantly different flow rates.
Fastest separations can be achieved with H2 as carrier gas, and H2 can be run
much faster than its optimal velocity with little penalty in resolution.
There are drawbacks to using H2. It can catalytically react with unsaturated compounds
on metal surfaces, and it cannot be used with a mass spectrometric detector, because H2
breaks down vacuum pump oil in the detector.
Separation of two polyaromatic hydrocarbons on a wall-coated open tubular column with different carrier gases.
Resolution, R, increases and analysis time decreases as we change from N2 to He to H2 carrier gas.