2/17/2018
PCB 4213: ADVANCED WELL TEST ANALYSIS
Interference & Pulse Testing
Credit hour 3
Duration 14 weeks Lecture + Lab/Tutorials
Instructor
Dr. Mohammed Mahbubur Rahman
Outline
Introduction
Interference Test
• Type-Curve Matching
• Semilog Analysis
Pulse Test
• Pulse Test Analysis
• Pulse Test Design
Ref:
Pressure Transient Testing by 3J (SPE Text book series 09)
Well Testing by John Lee (SPE Text book series 01)
1
2/17/2018
Introduction
Single Well Multi Well
PBU Interference
PDD Pulse
Injection
Fall Off
DST
RFT/WFT etc
Introduction
The multiple-well test requires at least one active
(producing or injecting) well and at least one pressure-
observation well.
2
2/17/2018
Introduction
In multiple-well testing, the flow rate at the active
well is varied while bottom-hole pressure response is
measured at the observation wells.
Introduction • The influence region
(the radius of influence)
for pulse or interference
testing is approximately
as indicated in the
figure.
2rinf
• The major difference is
r that the testing time, t, is
much larger in
interference testing
than in pulse testing.
• Thus, rinf, and the total
2rinf +r influence region are
substantially larger in
kt interference testing
rinf 0.029 than in pulse testing.
ct
3
2/17/2018
Interference Test: Type-Curve Matching
To analyze an interference test by type-curve matching:
Plot, on a Move the tracing paper
tracing paper, horizontally and
the observation- vertically until the data
well pressure points match the
data as exponential-integral curve
Δp vs t
Estimate the Choose a
permeability & convenient
the porosity- match point then
compressibility read the match
product point values
4
2/17/2018
Permeability is estimated from
141.2 q ( p D ) M
k
h p M
The porosity-compressibility product is estimated from
0.0002637 k tM
ct
r2 (t D / rD2 ) M
5
2/17/2018
Example 9.1 (JL): Interference-Test Analysis
by Type-Curve Matching
During an interference test, water was injected into
well A for 48 hours. The pressure response in well B,
119 ft away, was observed for 148 hours. Known
reservoir properties are
depth = 2000 ft βw = 1 RB/STB
q = - 170 B/D μw = 1 cp
h = 45 ft r = 119 ft
pi = 0 psig ct = 9*10-6 psi-1
t1 = 48 hours
Example 9.1: Interference-Test Analysis by
Type-Curve Matching
Observed Pressure Data:
Pressure
build up
Pressure
Fall off
Both PBU and PFO segments can be analyzed
6
2/17/2018
Example 9.1: Interference-Test Analysis by
Type-Curve Matching (analyzing the PBU part)
ΔpM = - 100
(pD)M = 0.96
tM = 10 hours
(tD/rD2)M = 0.94
Example 9.1: Interference-Test Analysis by
Type-Curve Matching
At the match point marked:
ΔpM = - 100 psig @ (pD)M = 0.96
tM = 10 hours @ (tD/rD2)M = 0.94
Results: 141.2 17011 0.96
k 5.1 md
45 100
ct
0.0002637 5.1 10 1.01106 psi 1
1192 1 0.94
1.0110 6
0.11
9 10 6
7
2/17/2018
Example 9.1: Interference-Test Analysis by Type-Curve Matching (analyzing the PFO part)
We can estimate the accuracy of the above analysis by using the data from the
declining pressure part of the test (t > 48 hours).
We extrapolate the solid line in Fig 9.6 by
tracing the curve, and estimate the difference
between Δpext and the observed Δp, ΔpΔt. Table
9.2 shows the computations.
The points (plus symbols) fall on just about the same line as the circles, so we
can be confident of the analysis results.
8
2/17/2018
Pulse Test
Pulse testing is a special form of multiple-well testing.
The technique uses a series of short-rate pulses at the active
well.
Pulses generally are alternating periods of production (or
injection) and shut-in, with the same rate during each production
(injection) period.
The pressure response to the pulses is measured at the
observation well.
Although the flow
time and shut-in
time are equal in
that figure, pulse
testing can be
done with
unequal flow and
shut-in times.
However, all flow
times must be the
same and all shut-
in times must be
the same.
9
2/17/2018
Two characteristics of the pressure response at
observation well are used for pulse-test analysis:
1. The time lag
2. The pressure amplitude
Pulse Test Analysis
The ratio of pulse length to the total cycle length is
defined as t p
F'
tc
The dimensionless time lag is defined analogy to any
dimensionless time,
t L D 0.00026372 k t L
ct rw
10
2/17/2018
Pulse Test Analysis
Pulse Test Analysis
Using the wellbore radius of the active well, the
dimensionless distance between the active and
observation wells is
r
rD
rw
The dimensionless-pressure response amplitude is
k h p
pD
141.2 q
where q is the rate at the active well while it is active
11
2/17/2018
Pulse Test Analysis
2
t L
141 . 2 q p D
t
c Fig .
k 2
t
h p L
tc
0.0002637 k t L
ct
(t )
r2 L2 D
rD Fig .
12
2/17/2018
13
2/17/2018
Pulse Test
Wellbore Storage Effects
Wellbore storage effects at the observation well increase the
time lag (tL) and reduce the response pressure amplitude (Δp)
of the first pulse.
We can approximate that the effect of wellbore storage at the
responding well will result in less than a 5-precent increase in
time lag and a virtually unaffected response amplitude when
the distance between pulsed and responding wells satisfies
0.54
c
r 32
ct h
14
2/17/2018
Pulse Test
Example 6.2 (John Lee): Pulse-Test Analysis
Given:
Active well was shut in for 2 hours, then
produced for 2 hours, then shut in for 2 r = 933 ft
hours, etc. μ = 0.8 cp
Amplitude of the 4th pulse, Δp = 0.63 psi h = 26 ft
φ = 0.08
Time lag, tL = 0.4 hours q = 425 STB/D
Bo = 1.26 RB/STB
Analyze the 4th pulse to estimate k, and φct
Pulse Test
Example 6.2 (John Lee): Pulse-Test Analysis
15
2/17/2018
Pulse Test
Pulse Test Design
For best results, pulse tests should be run with a good
combination of pulse length, pulse amplitude, and
pressure-measuring equipment:
Choose value of F` Choose the
For producing maximum Get
wells, try to ΔpD[tL/ΔtC]2 points tL/ΔtC
minimize it from the figures
Calculate
tL Get (tL)D /r2
ΔtC From the
ΔtP figures
Δp
Pulse Test
Example 9.5: Pulse-Test Design
We wish to design a pulse test for reservoir with the
following approximate properties:
q = 100 B/D r = 660 ft
k = 200 md β = 1.1 RB/STB
μw = 3 cp ct = 10*10-6 psi-1
h = 25 ft = 0.12
16
2/17/2018
Pulse Test
Example 9.5: Pulse-Test Design
1. The pulsing well is a producer, so to minimize shut-
in time we choose a short shut-in pulse, such F` = 0.3
2. For initial design calculation we choose the
maximum ΔpD[tL/ΔtC]2 point from Fig. 9.16 for the
first even pulses:
ΔpD[tL/ΔtC]2 = 0.0042 @ F` = 0.3
tL/ΔtC = 0.33
3. From Fig. 9.20:
For that value of tL/ΔtC {(tL)D /r2} = 0.122
Pulse Test
Example 9.5: Pulse-Test Design
17
2/17/2018
Pulse Test
Example 9.5: Pulse-Test Design
2
t L
141.2 q pD
tc Fig .
(t L ) D p
ct r 2 2
2
rD Fig . t
tL kh L
0.0002637 k tc
0.1810 10 6 3660 2 0.122 141.2 1001.130.0042
0.0002637 200
5.4 hours
20025 0.332
Cycle time is given by 0.36 psi
tC t L / tL / tC 5.4 / .33 16.4 hours
The pulse length is
Thus 0.36 psi is the expected
t p F ' tC 4.9 hours
response amplitude for even
pulse analysis.
We would shut in the well for
5 hours, produce for 11 hours,
and so forth
18