Alegria Clinical ELISA Analyser: Quick Guide

Uploaded by

Gustavo Rivas
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
  • Contents: The table of contents lists all sections with their respective pages for easy navigation.
  • Introduction and Reference: Provides essential prerequisites and information about proper use of the Alegria® system.
  • Intended Use: Outlines the intended user and purpose of the Quick Guide, specifying its usage in familiar contexts for system operations.
  • Quick Guide: Summarizes the types of protocols and steps to take the user directly to test procedures.
  • Start of Instrument: Instructions for starting the instrument, detailing the initial setup steps and user interaction.
  • LIMS Protocol - Requests Sorted by Patient ID: Details the process for sorting and handling test requests using the patient ID with LIMS.
  • LIMS Protocol - View all Requests: Describes the procedure for listing all test requests currently stored.
  • Manual Paper Protocol: Provides step-by-step instructions for manual handling of tests without automated systems.
  • Touch Screen Protocol - Entering Patient ID before Scanning: Covers the protocol for entering patient information through touch screen before scanning samples.
  • Touch Screen Protocol - Entering Patient ID after Scanning: Instructions for inputting patient IDs after completing scanning procedures.
  • External Barcode Reader Protocol: Guidance for operating the device using external barcode readers for identification purposes.
  • Service Centre: Contact information and references for authorized service support for the Alegria system.

Quick Guide

Alegria® Clinical ELISA Analyser


Quick Guide for use
Preamble

The contents of this Quick Guide are the property of ORGENTEC Diagnostika GmbH.

Any reproduction in whole or in part is strictly prohibited.

At the time of printing, the Quick Guide correctly describes the device and its functions.

However, as modifications may have been carried out since printing, the provided system

package consisting of this Quick Guide, the Instructions for Use (on CD), and an

Installation Guide may contain one or more amendments to the manuals .

The Instructions for Use additionally to this Quick Guide and any amendments must be

thoroughly read before using the device.

ORGENTEC Diagnostika GmbH is only responsible for the reliability and performance

of the device if the following is strictly observed:

u Authorized personnel (see back of manual for authorized service centres) carry out
all service, repairs and modifications.

u The device must be used in accordance with the intended use and the instructions
put forth in the Safety Information section of the Instructions for Use.

u The user has to perform the daily, weekly and the monthly maintenance. The daily
maintenance contains the rinsing of the system at the end of the day.
Once a week, thoroughly clean all Alegria reagent containers using a brush
and a commercially available laboratory detergent. The monthly maintenance
is a cleaning of the system with Flush Routine Solution (order no.: ORG 305).

ORGENTEC Diagnostika GmbH offers a one-year warranty from the day of purchase on
defective material and assembly. If the below points are not strictly observed, the warranty
will be considered invalid.

u The warranty does not cover damage resulting from incorrect use or user mainten-
ance or from unauthorized software modification.

u ORGENTEC Diagnostika GmbH is only responsible for replacement of defective


parts excluding wear parts.

u ORGENTEC Diagnostika GmbH is not responsible for any personal injury or any
damage resulting from incorrect use of the analyser.

2
Contents

Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3

Introduction and Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4

Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5

Quick Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5

Start of Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5

LIMS Protocol - Requests Sorted by Patient ID . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6

LIMS Protocol - View all Requests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7

Manual Paper Protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8

Touch Screen Protocol - Entering Patient ID before Scanning . . . . . . . . . . . . . . . . . .9

Touch Screen Protocol - Entering Patient ID after Scanning . . . . . . . . . . . . . . . . . .10

External Barcode Reader Protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11

Service Centre . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12

3
Introduction and Reference

WARNING

This Quick Guide is for Alegria® users that had read the Alegria® Instructions for Use and

know all safety informations required for safe and proper use of the instrument.

Prerequisites are:

u The user has read the Alegria® Instructions for Use !

u The device was unpacked and installed according to the Instructions for Use !

Note

The topics listed below are described in the Alegria® Instructions for Use:

u Safety information

u Symbols
u Packaging/unpackaging

u Intended use
u Application

u System setup
u Program overview and detail views

u Replacement procedures
u Calibration

u Service
u Maintenance

u Technical specifications

4
Intended Use

This manual is to be used as a Quick Guide for the user that knows the instrument.

If not, please consult the Alegria® Instructions for use.

The Quick Guide is intended to take the user the fastest way through the test protocol(s).

Quick Guide
The instrument can be used with different types of protocols. The quick guide describes

the following protocol types:

(1) LIMS (Laboratory Information Management System) Protocol (page 6)

(2) Manual Paper Protocol (Page 8)

(3) Touch Screen Protocol (Page 9)

(4) External Barcode Reader Protocol (Page 11)

Note
Depending on the device configuration (with or without LIMS) four respectively three

Protocols will be visible after having started the instrument.

Start of Instrument
u Push "Start Analyze" button
u Enter a user name by using touch screen keyboard

u Push "Ok" button


u Push the respectively protocol button and follow instructions below

5
(1) LIMS Protocol - Requests Sorted by Patient ID
u By using external barcode reader, scan first the patient ID and the appropriate test

requests are shown in the run list

u Insert the corresponding test strips into the SysTray

u Highlight the test request on the display by tapping the corresponding line showing

the patient ID and the test name; scan the corresponding test strip barcode with the

external barcode reader; a message appears that prompts if the two scanned IDs

are to be linked together; push the "OK" or "Cancel" button

u A list can be printed by pushing "Print" button, this can be helpful when pipetting the

samples

u Pipette the samples into the strips and ensure to pipette into the correct strip positi-

on in the SysTray

u Push "Open Lid" button

u Insert SysTray into instrument

u To insert more than one SysTray push the "Rotate" button to rotate to the next free

SysTray position

u Push "Close Lid" button

u Push "Scan Strips" button

Note
After scanning of the test strips with the internal barcode reader inside the instrument, the
position of the test request on the display is set according to the position of the test strip
inserted to the SysTray.
u Push "Analyze" button

u Results will be transferred after the test run depending on the chosen LIMS export
mode either automatically to the LIMS, or can be exported manually from the archive

u Printing is done automatically


u Remove SysTray

u To remove more than one SysTray, push the "Rotate" button to rotate to the next
SysTray

u Push "New" button to start a new test or push "Exit" button to end analyse session

6
(1) LIMS Protocol - View all Requests
u Push "View all" button to call up all test requests stored on the Alegria to the display

u Insert the corresponding test strips into the SysTray

u Highlight the test request on the display by tapping the corresponding line showing

the patient ID and the test name; scan the corresponding test strip barcode with the

external barcode reader; a message appears that prompts if the two scanned IDs

are to be linked together, push the "OK" or "Cancel" button

u A list can be printed by pushing "Print" button, this can be helpful when pipetting the

samples

u Pipette the samples into the strips and ensure to pipette into the correct strip positi-

on in the SysTray

u Push "Open Lid" button

u Insert SysTray into instrument

u To insert more than one SysTray push the "Rotate" button to rotate to the next free

SysTray position

u Push "Close Lid" button

u Push "Scan Strips" button

Note
After scanning of the test strips with the internal barcode reader inside the instrument, the
position of the test request on the display is set according to the position of the test strip
inserted to the SysTray.

u Push "Analyze" button

u Results will be transferred after the test run depending on the chosen LIMS export

mode either automatically to the LIMS, or can be exported manually from the archive

u Printing is done automatically

u Remove SysTray

u To remove more than one SysTray, push the "Rotate" button to rotate to the next

SysTray

u Push "New" button to start a new test or push "Exit" button to end analyse session

7
(2) Manual Paper Protocol

u Insert strips into SysTray

2 Note on paper protocol on which position the strips are entered in the SysTray.

u Pipette the samples into the strip and ensure to pipette into the correct strip position

in the SysTray

2 Note on paper protocol in which strip the sample is filled.

u Push "Open Lid" button

u Insert SysTray into instrument

u To insert more than one SysTray push the "Rotate" button to rotate to the next free

SysTray position

u Push "Close Lid" button

u Push "Scan Strips" button

u Push "Analyze" button


u Printing is done automatically

u Remove SysTray
u To remove more than one SysTray, push the "Rotate" button to rotate to the next

SysTray
u Push "New" button to start a new test or push "Exit" button to end analyse session

8
(3) Touch Screen Protocol - Entering Patient ID before Scanning

u Insert strips into SysTray

u Push "Keyboard" button and enter patient ID by using the touch screen, or use the

external barcode reader and note in which position the patient ID is entered

u Pipette sample into strip and ensure to pipette into the correct strip position in the

SysTray according to entered patient ID on the touch screen

u Push "Open Lid" button

u Insert SysTray into instrument

u To insert more than one SysTray push the "Rotate" button to rotate to the next free

SysTray position

u Push "Close Lid" button

u Push "Scan Strips" button

u Push "Analyze" button

u Printing is done automatically

u Remove SysTray

u To remove more than one SysTray, push the "Rotate" button to rotate to the next

SysTray

u Push "New" button to start a new test or push "Exit" button to end analyse session

9
(3) Touch Screen Protocol - Entering Patient ID after Scanning
u Insert strips into SysTray

u Push "Open Lid" button

u Insert SysTray into instrument

u To insert more than one SysTray push the "Rotate" button to rotate to the next free

SysTray position

u Push "Close Lid" button

u Push "Scan Strips" button

u Push "Keyboard" button and enter patient ID by using the touch screen, note in

which position the patient ID is entered

u A list can be printed by pushing "Print" button, this can be helpful when pipetting the

samples

u Push "Open Lid" button

u Remove SysTray

u To remove more than one SysTray, push the "Rotate" button to rotate to the next

SysTray

u Push "Close Lid" button

u Pipette sample into strip and ensure to pipette into the correct strip position in the

SysTray according to entered patient ID on the touch screen

u Push "Open Lid" button

u Insert SysTray into instrument

u Push "Close Lid" button

u Push "Scan Strips" button

u Push "Analyze" button

u Printing is done automatically

u Remove SysTray

u To remove more than one SysTray, push the "Rotate" button to rotate to the next

SysTray

u Push "New" button to start a new test or push "Exit" button to end analyse session
10
(4) External Barcode Reader Protocol

u Insert strips into SysTray

u By using external barcode reader scan first the patient ID and then the strip ID; a

message appears that prompts if the two scanned IDs are to be linked together,

push the OK or Cancel button

u A list can be printed by pushing "Print" button, this can be helpful when pipetting the

samples

u Pipette sample into strip and ensure to pipette into the correct strip position in the

SysTray

u Push "Open Lid" button

u Insert SysTray into instrument

u To insert more than one SysTray push the "Rotate" button to rotate to the correct

SysTray position

u Push "Close Lid" button

u Push "Scan Strips" button

Note
After scanning of the test strips with the internal barcode reader inside the instrument, the
position of the test request on the display is set according to the position of the test strip
inserted to the SysTray.

u Push "Analyze" button

u Printing is done automatically

u Remove SysTray

u To remove more than one SysTray, push the "Rotate" button to rotate to the next

SysTray

u Push "New" button to start a new test or push "Exit" button to end analyse session

11
Service Centre

Authorized Service Representative:

Manufactured by:

ORGENTEC Diagnostika GmbH, Carl-Zeiss-Straße 49-51, 55129 Mainz, Germany

Tel. +49 (0) 6131 / 9258-0

Fax +49 (0) 6131 / 9258-58

[email protected]
2011-04