Tata Power SOP
Tata Power SOP
GUIDELINES
FOR
ANALYTICAL TEST PROCEDURES
FOR
WATER AND STEAM SYSTEM
Contents
1 INTRODUCTION ........................................................................................................... 4
2 COAGULATION/ FLOCCULATION JAR TEST ........................................................ 5
3 COLOR.......................................................................................................................... 11
4 CONDUCTIVITY ......................................................................................................... 14
5 CORROSIVITY OF COOLING WATER (Coupon Test)............................................ 19
6 CRUD (Colour Comparison Method) ........................................................................... 25
7 pH .................................................................................................................................. 26
8 OXIDATION – REDUCTION POTENTIAL (ORP) ................................................... 32
9 TURBIDITY (Nephelometric) ...................................................................................... 38
10 SUSPENDED AND TOTAL DISSOLVED SOLIDS .................................................. 41
11 SUSPENDED AND TOTAL DISSOLVED SOLIDS .................................................. 44
12 ALKALINITY (Titration Method, 10 to 500mg/litre) .................................................. 47
13 ALKALINITY DUE TO HYDROXIDE ...................................................................... 52
14 ALUMINIUM ............................................................................................................... 55
15 ALUMINIUM (ATOMIC ABSORPTION METHOD) ................................................ 59
16 AMMONIA (Indophenol method) ................................................................................ 62
17 AMMONIA by Nesseler’s Reagent .............................................................................. 65
18 AMMONIA BY ION SELECTIVE ELECTRODE ...................................................... 69
19 CARBON DIOXIDE (Bicarbonate Titration Method) ................................................. 73
20 CARBON DIOXIDE (Direct Titration of Free Carbon Dioxide) ............................... 75
21 CHLORIDE (Mercuric Thiocyanate Method) ............................................................. 78
22 CHLORIDE (Mercuric Thiocyanate Method, Modified, 2 to 100 micrograms/litre) . 81
23 CHLORIDE (Silver Nitrate Method, 5mg/litre or more) ............................................ 84
24 CHLORINE DEMAND ................................................................................................ 87
25 CHLORINE RESIDUAL (DPD Method, 0.02 to 4.0mg/litre) ..................................... 92
26 CHLORINE, RESIDUAL (Starch Iodide Thiosulphate Method) ................................ 97
27 COPPER (Neocuprine method, 2 to 1000 micrograms/ litre Cu) ............................... 101
28 COPPER (Atomic Absorption Spectrophotometry Method) ...................................... 107
29 EQUIVALENT AND FREE MINERAL ACIDITY .................................................. 109
30 FLUORIDE (SPANDS Dye Method) ......................................................................... 112
31 HARDNESS (TOTAL, CALCIUM AND MAGNESIUM) ....................................... 118
32 HYDRAZINE (p-Dimethylamino Benzaldehyde Method)......................................... 124
33 IRON (Bathophenanthroline Method) ......................................................................... 128
34 IRON (Ferrozine Method) ........................................................................................... 134
35 IRON ATOMIC ABSORPRION METHOD (Graphite Furnace)............................ 137
36 NICKEL (Glyoxime Method) ..................................................................................... 139
37 NICKEL by ATOMIC ABSORPTION METHOD (Direct) ....................................... 143
1. INTRODUCTION
Chemical analysis of water and steam is one of the most important tool for the
Chemists in the power plant. The analytical data obtained in power plant by the
combined system of online and offline monitoring system is a very effective
tool to monitor the operating status of the plant and to understand the trend of
operating parameters. Hence, the frequency of offline tests and collection of on-
line data are adequately timed, so as to take timely corrective actions. Each of
our stations do have robust testing facility and system in place, to measure the
chemistry parameters, however inorder to consolidate and bring in uniformity in
approaches, this guideline has been developed to consolidate all the analytical
test procedures adopted in our thermal power plants.
This Guideline has been designed and finalized in consultation with the
Knowledge Team- Chemistry to ensure that all the essential analytical test
procedures are incorporated in this document. These Guidelines shall be
adopted in all Tata Power thermal stations. The Chemistry Head of each
Division shall use this document to decide on the test methods to be used, based
on their plant configuration and comply with the same. The reference is taken
from ASTM/BIS/ISO for making this document. Consequent to issuance of this
Guideline, it is expected that each of the stations shall endeavour to adopt these
and maintain their parametric chemistry regime optimum values.
The coagulation - flocculation jar test is carried out to determine chemicals and
their dosages, and conditions required in order to reduce suspended, colloidal, and
non-settleable matter from water by chemical coagulation - flocculation, followed
by gravity settling to achieve optimum results. The variables to be investigated are
(i) chemical additives, (ii) pH, (iii) temperature and (iv) order of addition and
mixing condition.
2.2 APPARATUS
2.2.1 Multiple Stirrers - with continuous speed variation from about 20 to 150 rpm. The
stirring paddles should be of light gage, corrosion - resistant material, all of the
same configuration and size. An illuminated base is useful to observe the floc
formation.
2.2.3 Reagent Racks - for introducing each test solution to all beakers simultaneously.
There should be at least one rack for each test solution or suspension. A typical
configuration is given in Fig-I.
2.3 REAGENTS
2.4 PROCEDURE
2.4.1 Place equal volumes (1000m1) of sample into each beaker (1500m1 capacity) and
record the temperature of the sample. Load the test chemicals in the reagent racks.
Final volume of each tube should be same after required make up.
2.4.2 Start the multiple stirrers at flash mix speed (approximately 120 rpm). Add the test
solutions or suspensions at predetermined dosage levels and sequence. Flash mix
for approximately 1 minute after the addition of chemicals. Record the flash mix
time and mixer speed (rpm).
2.4.3 Reduce the speed to the minimum required, to keep floc particles uniformly
suspended throughout the "slow mix" period. Slow mix for 20 minutes. Record the
mixer speed (rpm).
2.4.4 After the slow mix period, withdraw the paddles and observe settling of floc
particles. Record the time required for the bulk of the particles to settle.
2.4.5 After 15 minutes of settling, record the sample temperature and by means of a
pipette, withdraw supernatant liquor for conducting colour, turbidity, pH, non-
reactive and/or colloidal silica and other required analysis.
2.5 INTERFERENCES
2.5.1 Temperature changes during test - Thermal or convection currents may occur,
interfering with the settling of coagulated particles.
2.5.2 Gas release during test - Floatation of coagulated floc may occur due to gas bubble
formation, caused by mechanical agitator, temperature increase or chemical
reaction.
2.5.3 Testing period - Biological activity or other factors may alter the coagulation
characteristics of water upon prolonged standing. Therefore, the period between
sampling and testing should be kept to a minimum.
2.6 NOTES
2.6.1 All polyelectrolytes are classified as anionic, cationic, or neutral, depending upon
their composition. A small dosage, under 1 ppm, may permit a reduction in the
dosage or complete elimination of the coagulant. In case the coolant aid is in
powdered form, prepare 0.1 % solution to provide proper dose. Always add
powdered aids to dissolving water rather than reverse.
LOCATION –
SAMPLE –
pH –
TURBIDITY –
COLOUR –
TEMPERATURE –
1. Chemicals,* mg/litre
6. Temperature, °C.
9. Settling rate
10 Supernatent Tests:
a. Turbidity
b. pH
c. Colour
d. Non-reactive/colloidal Silica
The preferred sequence of addition of chemical is (a) alum & or polymer -----
à (b) chlorine and ------à (c) lime (if needed)
3.0 COLOR
3.1 SUMMARY OF METHOD
3.2 APPARATUS
3.2.2 pH meter.
3.3 REAGENTS
Prepare standards having colours of 5, 10, 15, 20, 25, 30, 35, 40, 45, 50, 60 and
70 units by diluting suitable volumes of standard stock solution (3.2) to 50 ml
with water in Nessler tubes.
3.4 PROCEDURE
3.4.1 Place 50 ml of sample in a Nessler tube.
3.4.2 Compare the colour of the sample with standard colour by looking vertically
downwards against a white surface.
3.4.3 If the colour exceeds 70 units, dilute with water until the colour is within the
range of the standards.
3.5 CALCULATIONS
Colour unit = A x 50
Where:
B = milliliters of sample.
1 -50 1
51 -100 5
101 -250 10
251 -500 20
3.6 INTERFERENCES
3.6.1 Even a slight turbidity causes the apparent colour to be noticeably higher than
the true colour; therefore turbidity should be removed before measurement of
true colour.
3.6.2 The colour value of water is extremely pH dependent and invariably increases
as the pH of the water is raised. When reporting a colour value, specify the pH
at which colour is determined.
4 CONDUCTIVITY
4.1 SUMMARY OF METHOD
The conductivity cell is dipped in the sample contained in a beaker and the
conductivity is read directly from the conductivity meter.
4.2 APPARATUS
4.2.3 Thermometer, accurate to 0.5°C, when the instrument is not provided with
manual or automatic temperature compensation.
4.3 REAGENTS
4.4 PROCEDURE
4.4.1.1 Rinse the conductivity cell with at least three portions of standard potassium
chloride solution.
Kx
Where:
4.4.2.1.1 Use a flow type conductivity cell. Adjust the sample stream to a proper flow
rate and bring the temperature to a steady value as near 25°C as possible Read
the temperature to the nearest 0.5oC.
4.4.2.2.1 Either a flow-type, dip-type, or pipette-type cell may be used. If a flow-type cell
is used, proceed in accordance with 4.2.1.
4.4.2.2.2 If another type of cell is used, rinse the cell thoroughly several times with water
and then two or more times with the sample. Measure the conductivity and the
temperature (to the nearest 0.5°C) on successive portions of the sample until a
constant value is obtained.
4.5 CALCULATIONS
Where:
A = cell constant.
4.6 PRECISION
4.6.1 Results obtained should not differ by more than 1% of the conductivity.
4.7 INTERFERENCES
4.7.1 Exposure of a sample to the atmosphere may cause changes in conductivity due
to loss or gain of dissolved gases.
4.7.2 The carbon dioxide, normally present in the air, can drastically change the
conductivity of pure water. Contact with air should be avoided by using flow
through or inline cells.
4.8 NOTES
4.8.1 The unit of conductivity is siemens per centimeter. The conductance is directly
proportional to the cross-sectional area, A (cm2) and inversely proportional to
the length of the path, L (cm)
A
Conductance = K X -------
L
4.8.2 Recommended cell constants for various conductivity ranges are given below:
(Microsiemens/cm) (Cm-1 )
10 to 200 0.1 to 1
200 to 5000 1 to 10
4.8.3 The conductivity of water and aqueous solutions depends strongly upon the
temperature. To avoid making a correction, it is necessary to hold the
Carefully prepared and weighed metal coupons are installed in contact with
flowing cooling water for a specified length of time. After removal from the
system, these coupons are examined, cleaned and reweighed. The corrosivity and
fouling characteristics of the water are determined from the difference in weight,
the depth and distribution of pits and the weight and characteristics of the foreign
matter on the coupons.
5.2 APPARATUS
5.2.2 Insulating washer, screw, and nut -for attaching the coupon to the phenolic rod.
The insulating washer has a sleeve that fits into the coupon hole and around the*
screw.
5.2.3 Phenolic rod -use .a 152mm length of canvas based 13mm outside diameter
phenolic rod attached at one end to a drilled pipe plug and having a flat surface
and a hole at the other end, suitable for attachment of the coupon.
5.3 REAGENTS
5.3.2 Benzene.
5.3.4 Hydrochloric Acid (1+4) - Mix 1 volume of concentrated hydrochloric acid (sp gr
1.19) with 4 volumes of water.
5.3.5 Hydrochloric Acid (1+1.8, inhibited) - mix 357m1. of hydrochloric acid (sp gr
1.19) and 5.0 g of inhibitor. Dilute to 1 litre with water.
5.3.7 Methyl orange Indicator (0.5g/litre) - Dissolve 0.05g of methyl orange in water
and dilute to 100m1.
5.3.8 Trichloroethylene.
5.3.9 Tripoli – finely granulated, porous, of siliceous rock;' amorphous silica, soft,
porous, and free of sharp edges.
5.4.2.1 Shear 20 gage sheet metal of the type desired to the dimensions of 3 by 102mm.
5.4.2.2 Drill a 7mm hole with its center about 8mm from one end of the coupon.
5.4.2.3 Deburr all sharp edges on the coupon specimen using a file or emery belt and
deburr' the hole with an oversize drill.
5.4.2.4 Stamp identifying numbers on the small coupon area between the edge and the
mounting hole.
5.4.3 Cleaning of coupons (ferrous, copper, brass and cupronickel coupons) - Remove
oil by immersion in benzene. Dry. Immerse in hydrochloric acid (1+4) for 30
minutes at room temperature. Remove acid from the coupon by three rapid
successive rinses in separate water baths; the last rinse water bath shall contain
methyl orange solution and must be kept neutral (yellow). Rinse successively
in isopropyl alcohol and benzene, and dry with a clean cloth. Store in a
desiccator.
5.5 PROCEDURE
5.5.2 Attach the coupon to the phenolic rod, using an insulating washer, the screw and
nut assembly.
5.5.4.1 Use short time intervals (4 to 7 days) for the first time series to establish the rate
at which passivity occurs.
5.5.4.2 Use long time intervals (1 to 3 months) for the second time series to establish
the mean steady state corrosion rate.
5.5.5 Clean the ferrous coupons, prior to reweighing, with a plastic knife. Remove
oily and greasy deposits by soaking in trichloroethylene. Remove remaining
loose corrosion product by brushing with a soft bristle brush. Immerse in
inhibited hydrochloric acid (1+1.8) for 30 seconds at room temperature.
Remove the specimen from acid bath and rinse with water. Rub with granular
trisodium phosphate and then with tripoli. Rinse with water and then with
isopropyl alcohol. Dry between paper towels followed by warm air drying.
5.5.6 Clean the copper and copper alloy coupons, prior to reweighing, with a plastic
knife. Remove oily or greasy deposits by soaking in trichloroethylene. Immerse
the coupons in inhibited hydrochloric acid (1+1.8) for 30 seconds. Rinse with
water, then with isopropyl alcohol and finally with benzene. : Dry between
paper towels and place in a desiccator for 1 hour.
5.5.8 Subject a weighed blank coupon of the same material to the identical cleaning
procedure (5.5 and 5.6) used for the test specimens and reweigh to determine
the blank correction factor.
5.5.9 Record the appearance of the cleaned, weighed coupon a “protected", 'moderate
localized', moderate pitting" or 'severe pitting'.
5.6 CALCULATIONS
Where :
5.6.2 In systems, where all variables are closely controlled, apply the following
equation which gives more precise corrosion rate. Calculate the corrosion rate
for a rectangular coupon as follows:
1
P = --------------------------------------------- x [(W1- W2)/W1.D] x 1.825 x 102
[(1/H) + (1/X) + (1/Y)]
Where:
5.6.3 Express the relationship between corrosion rate in grams per square meter per
day (g/m2d) and penetration in millimeters per year (mmpy) as follows:
Where:
5.7 INTERFERENCES
5.7.2 The presence of different metals in close proximity to the coupon (within
76mm) even if they are insulated from the coupon, constitutes a source of error
in the results.
5.7.3 Deviations in the velocity and direction of flow past the coupons may influence
the precision of the results.
5.7.4 Results are directly comparable only for the water temperature to which the
coupon is exposed.
5.8 NOTES
5.8.1 The ferrous metal coupons, until ready for use, should be stored in desiccator in
separate envelopes made from vapour phase inhibitor - impregnated paper.
5.8.1.1 Non ferrous metal coupons should be stored in sealed plastic envelopes or
wrapped in plastic film.
5.8.2 Since the corrosion rate will be very high initially and then fall to a lower,
nearly constant rate, two time series are chosen. Short time interval is for 4 to 7
days and long time interval is for 1 to 3 months.
5.8.3 The interim period between removal of specimens and reweighing should be
kept to a minimum and in no case should it exceed 1 week. Protect the specimen
if it cannot be examined, cleaned and reweighed immediately after removal, by
storing as described in 8.1.
6.2 APPARATUS
6.3 PROCEDURE
6.3.1 Filter 1-litre of the sample through the membrane filter by applying vacuum.
6.3.2 Dry the membrane filter at room temperature and compare the coloration on the
membrane filter with the standard Babcock & Wilcox Comparison Chart.
6.4 CALCULATIONS
6.4.1 Determine the crud level in the sample from the matched colour of the enclosed
Babcock & Wilcox Comparison Chart.
6.4.2 The method is approximation only. For determining the exact crud level
weighment method to be adopted after passing adequate amount of water
through membrane using in-line filter holder.
7 pH
7.1 SUMMARY OF METHOD
The pH meter and associated electrodes are standardized against two reference
buffer solutions, which are close to the anticipated sample pH. The Sample
measurement is made under specified conditions and prescribed techniques.
7.2 APPARATUS
7.2.1 Laboratory pH meter together with its associated glass and reference electrodes.
7.3 REAGENTS
7.3.2 Borax Reference Buffer Solution (pH = 9.18 at 25°C) - Dissolve 3.80g of
sodium tetraborate decahydrate (Na2 B407.1OH20) in water and dilute to 1-litre.
7.3.3 Phosphate Reference Buffer Solution (pH = 6.86 at 25°C) - Dissolve 3.39g of
potassium dihydrogen phosphate (dried at 130°C for 2 hours) (KH2PO4) and
3.53g of anhydrous disodium hydrogen phosphate (dried at 130°C for 2 hours)
(Na2HPO4) in water and dilute to 1-litre.
7.3.4 Phthalate Reference Buffer Solution (pH = 4.00 at 25°C) - Dissolve 10.12g of
potassium hydrogen phthalate (dried at 110°C for 2 hours) (KHC8H4O4) in water
and dilute to 1-litre.
7.3.5 Tetraoxalate Reference Buffer solution (pH = 1.68 at 25°C) - Dissolve 12.61g of
potassium tetroxalate dihydrate (KHC2H4H2C2O4.2H2O) in water and dilute to 1
litre.
7.3.6 Sodium Bicarbonate - Sodium Carbonate Reference Suffer Solution (pH a 10.01
at 25°C) - Dissolve 2.092g of sodium bicarbonate (NaHCO3) and 2.640g of
sodium carbonate (dried at 275°C for 2 hours) (Na2CO3) in water and dilute to
1-litre.
7.4 PROCEDURE
7.4.2 Select two reference buffer solutions, the pH values of which are close to the
anticipated sample pH and if possible bracket the sample pH.
7.4.3 Standardize the pH meter with the above two (4.2) reference buffer solutions in
accordance with the manufacturer's instructions.
7.4.4 Wash the electrodes with water and fill the beaker (provided with a
thermometer) with water sample. Insert the electrodes into the beaker and record
the pH of the water sample when the drift is less than 0.02 units in 1-minute.
7.5 CALCULATIONS
7.5.1 Most pH meters are calibrated in pH units and the pH of the sample is obtained
directly by reading the meter scale.
7.5.3 Report the pH of the test solution to the nearest 0.01 pH units when the pH
measurement lies between 1.0 and 12.0.
7.5.4 Report the pH of the test solution to the nearest 0.1 pH units when the pH
measurement is less than 1.0 and greater than 12.0.
7.6 PRECISION
7.6.1 The precision of this method is 0.05 pH units for pH measurements between 1.0
and 12.0.
7.6.2 When the pH is less than 1.0 and greater than 12.0, the precision is 0.1 pH units.
7.6.3 In order to attain this precision the condition of the instrumentation and the
technique for standardization and operation is extremely important.
7.7 INTERFERENCES
7.7.1 The true pH of an aqueous solution is affected by the temperature, which can be
compensated automatically in many instruments or can be manually compensated
in most other instruments. The temperature compensation corrects for the effect
of the water temperature on the instrument, including the electrodes, but does not
correct for temperature effects on the chemical system being monitored. It does
not adjust the measured pH to a common temperature; therefore, the temperature
should be reported for each pH measurement.
7.7.2 The glass electrode reliably measures pH in nearly all aqueous solutions and in
general, is not subject to solution interference from colour, turbidity, colloidal
matter, oxidants or reductants.
7.7.3 The pH response of most glass electrodes is imperfect at both ends of the pH
scale. The indicated pH value of highly alkaline solutions will be too low. This is
minimized by the selection of proper glass electrode.
7.7.4 The-indicated pH value of strong aqueous solutions of salts and strong acids
having a pH less than 1, will often be higher than the true pH value. This is
termed the negative error and the pH indicated is somewhat greater than the true
pH.
7.7.5 The pH response of the glass electrode may be impaired by a few coating
substances such as oily materials and particulates. The electrodes can be restored
to normal by an appropriate cleaning procedure recommended by the
manufacturer.
7.8 NOTES
7.8.1 The pH is the negative logarithm to the base ten of the conventional hydrogen
ion activity. It is derived from the electromotive force (emf) of the cell,
(E – Er ) F
pH = pHr = ---------------
2.3026 RT
Where :
F = Faraday constant.
R = Gas constant.
T = absolute temperature.
7.8.2 New glass electrodes and those which have been stored dry, shall be
conditioned and maintained as recommended by the manufacturer. It is
necessary to keep the immersible ends of the electrodes in water between
measurements. For prolonged storage, glass electrodes may be allowed to
become dry, but the junction and filling openings of reference electrodes should
be capped to decrease evaporation. Glass electrodes should be stored as
recommended by the manufacturer and reference electrodes in saturated
potassium chloride solution.
7.8.3 Both the saturated Calomel electrode and silver-silver chloride electrode are
satisfactory for measurement at room temperature. The silver-silver chloride
7.8.4 Where emulsions of free oil and water are to be measured for pH, it is necessary
to clean the glass electrodes thoroughly after each measurement. The cleaning is
done by washing with soap or detergent and water, followed by several rinse
with water, after which, the lower third of the electrode is immersed in
hydrochloric acid (1+9) and finally washed thoroughly with water.
7.8.5 If the sample contains sticky soaps or suspended particles, the cleaning is done
with a suitable solvent or by chemical treatment, to dissolve the deposited
coating. After cleaning with solvent the lower third is immersed in hydrochloric
acid (1+9) followed by thorough washing with water.
7.8.6 If glass electrode has failed to respond the treatment as described in 8.4, it is
immersed in chromic acid cleaning solution for several minutes. This drastic
treatment limits the life of electrode and is used only as an alternative to
discarding it. After chromic acid treatment, the electrode is allowed to stand in
water overnight.
ORP is defined as the electromotive force between a noble metal electrode and a
reference electrode when immersed in a solution. The ORP electrodes are inert
and measure the ratio of the activities of the oxidized to the reduced species
present.
8.2 APPARATUS
what in excess of that outside the junction. In non pressurized applications this
requirement can be met by maintaining the inside solution level higher than that
the outside solution level. If the reference electrode is non flowing junction
type, these outward flow and pressurization considerations shall not apply.
Aqua regia - Mix 1 volume of con. Nitric acid (HNO3, sp. gravity 1.18) with
three volumes of con. Hydrochloric acid (HCl, sp.gr 1.18)
used in each solution so that solid crystals are always present. These reference
solutions are stable only for 8 hrs.
ORP , (mv)
Buffer solution, pH 4 7
-------------------- ---------------------
Temperature 0 C 20 25 30 20 25 30
Reference electrode
Silver/silver chloride 268 263 258 92 86 79
Reference electrode 25 0 C 25 0 C 30 0C
Preparation
8.9 STANDARDIZATION
Checking the response of the electrode to standard redox solutions – Wash the
electrodes with three changes of water or by means of a flowing steam from a
wash bottle. Use one or more of the above mentioned reference solutions to
check the response of the electrode. Fill the sample container with fresh redox
standard solution and immerse the electrodes. The reading should be within 30
mV of the value expected for the standard solution. The second reading should
not differ from the first by more than 10 mV.
8.10 PROCEDURE
After the electrode / meter assembly has been standardized, wash the electrodes
with changes of water or by means of flowing water. Place the sample in a clean
glass beaker or sample cup and insert the electrodes. Provide adequate agitation
throughout the measurement period. Read the millivolt potential of the solution
allowing sufficient time for the system to stabilize. Measure successive portions
of the sample until readings on two successive portions differ by no more than
10 mV. A system that is very slow to stabilize probably will not yield a
meaningful ORP.
Process ORP analyzers with their rugged electrodes and electrode chambers
provide continuous measurements which are the basis for fully automatic
control. Make selection of the electrodes and electrode chamber to suit the
physical and chemical characteristics of the process material. Locate the
submersion-style electrode chamber so that fresh solution representative of the
process stream or batch continuously passes across the electrodes. Agitation
may be employed to make the stream/ batch more homogeneous. The ORP
value is usually displayed continuously and can be noted at any specific time.
Frequently the pH value is also recorded yielding a permanent record.
8.11 REPORT
Report the ORP value to the nearest 1 mV, interpolating the meter scale as
required. When considered appropriate, the temperature at which the ORP and /
or pH were measured may also be reported.
Precision and bias of the measurement depends largely on the condition of the
electrode system and on the chemical system being measured. In the absence of
substances that coat or poison the electrode, the precision is +/- 10 mV.
9 TURBIDITY (Nephelometric)
9.1 SUMMARY OF METHOD
9.2 APPARATUS
9.2.2 Sample tubes. – It must be clear, colourless glass and clean from both inside and
outside. Index marked so that repeated exact placements into the instrument
cuvette well for measurement can be made.
9.3 REAGENTS
9.3.5 In a 100m1 volumetric flask, mix 5.0ml solution A with 5.0 m1 Solution B.
Allow to stand for 24 hours at 25 +/ - 3 0C. Make up to the mark with turbidity
free water and mix well. The turbidity of this suspension is 400 NTU and this
stock solution to be prepared on monthly basis.
9.3.6.1 Dilute l0.0m1 of stock turbidity suspension (3.2.3) to 100ml with turbidity free
water. The turbidity of this suspension is 40 NTU. Prepare weekly this
suspension.
9.3.7.1 Dilute portions of the standard turbidity suspension (3.3.1) with turbidity free
water, as required. Prepare weekly. But pl. note that dilution below 4.0 NTU
should be prepared daily. (Standards below 1.0 NTU are difficult to prepare
accurately. Refer to specific manufacture instruction to determine if such a
standard is needed.)
9.4 PROCEDURE
9.4.1 Calibrate the Nephelometer with standard turbidity suspension for each range,
in accordance with the manufacturer's instructions.
9.4.2 Replace the standard by the sample in the same tube after thoroughly washing
the tube with turbidity -free water or in an optically identical tube and record the
reading.
Shake the sample thoroughly to disperse the solids. Allow air bubble to
disappear, then proceed.
Dilute the sample with one or more equal volumes of reagent water until
turbidity is below 40 NTU after mixing and degassing. Then treat this less than
40 NTU solution as above. The turbidity of the original sample can be
calculated based on dilution data and original sample volume.
9.4.5 After calibration check and adjustment, empty the cell of turbidity standard and
rinse the cell with the water to be tested. The measured NTU value of the water
sample to be determined using the instrument value and the appropriate
calibration curve.
9.5 CALCULATIONS
40 – 100 5.0
10.2 APPARATUS
10.2.2 Automatic Evaporation Assembly - A dust shield, constant level device, heater
and evaporation dish.
10.2.3 Sampling Device - A cooling coil with overflow pipe and solenoid valve
suitable for sampling from a water source to a continuous sample evaporator.
(The cooling coil is necessary, only when, sample is above room temperature).
10.2.4 Membrane Filter Assembly - A borosilicate glass or stainless _steel funnel with
a flat, fritted base of the same material, and membrane filters (0.45 micron pore
size) to fit.
10.3 REAGENTS
10.4 PROCEDURE
10.4.2.1 Place the membrane filter in a petri dish and dry in an oven at 103°C for 15
minutes or in a vacuum desiccator for 30 minutes. Weigh the filter to the nearest
0.1 mg.
10.4.2.2 Filter the sample through membrane filter (4.2.1) using the filtration assembly
and the vacuum pump or water aspirator. Wash the residue with chloroform or
benzene. Place the filter in the petri dish.
10.4.2.3 Place the petri dish in the oven at 103°C for 30 minutes. Reweigh the filter and
record the weight of the residue on the filter.
10.4.3.1 Weigh a platinum dish that has been dried at 103°C for 1 hour and cooled in a
desiccator. Using evaporation assembly start the evaporation of the selected
volume of the sample for total solids (4.1) or the filtrate from the suspended
solids determination (4.2).
10.4.3.2 When the evaporation is almost complete remove the dish from the assembly
and dry at 103°C for 1 hour in an oven: Cool in a desiccator and weigh. Record
the weight of the residue.
10.5 CALCULATIONS
W1
W2
Suspended solids, mg/litre = ---------- x 1000
V
W3
Total dissolved solids, mg/litre = ------------- x 1000
V
Where:
10.6 NOTES
10.6.1 Some evaporation residues readily absorb moisture, therefore rapid weighing
should be done.
Samples containing 25mg/litre or less of total solids on which only the total solids
content is, to be determined shall be immediately acidified with 0.2ml of hydrochloric
acid (sp gr 1.19) per litre of water. If suspended solids is to be separately determined,
the sample, regardless of total solids content, shall be filtered, as soon as possible and
then acidified 11)
11.2 APPARATUS
11.2.2 Membrane Filter Assembly - A borosilicate glass or stainless steel funnel with a
flat, fritted base of the same material, and membrane filters (0.45 micron pore
size) to fit.
11.2.5. Heater - Hot plate or steam bath for maintaining the temperature of the
evaporating sample near the boiling point.
11.3 REAGENTS
11.4 PROCEDURE
11.4.2.1 Place the membrane filter in a petri dish and dry in an oven at 103°C for 15
minutes or in a vacuum desiccator for 30 minutes. Weigh the filter to the nearest
0.1 mg.
11.4.2.2 Filter the sample through the membrane filter (4.2.1) using the filtration
assembly and the vacuum pump or water aspirator. Place the filter in the petri
dish.
11.4.2.3 Place the petri dish in the oven at 103°C for 30 minutes. Reweigh the filter and
record the weight of the residue on the filter,
11.4.3.1 Transfer the sample for total solids determination (4.1) or the filtrate from
suspended solid determination (4.2) to a sample reservoir.
11.4.3.2. Fill an evaporating dish (previously dried at 103°C for 1 hour and weighed) to
within 6.3mm of the top, with sample.
11.4.3.3 Evaporate the sample on a hot plate or steam bath. Periodically, add sample
from the reservoir to the dish until the reservoir is empty.
11.4.3.4 Dry the dish at 103°C for 1 hour. Cool in a desiccator and weigh. Record the
weight of the residue in the dish.
11.5 CALCULATIONS
W1
Total Solids, mg/litre = ---------- x 1000
W2
Suspended solids, mg/litre = -------- x 1000
V
W3
Total dissolved solids, mg/litre = --------- x 1000
V
Where :
11.6 NOTES
11.6.1 Some evaporation residues readily absorb moisture, therefore rapid weighing
should be done.
11.6.2 Suspended solids are defined as those solids, exclusive of gases and in non-
liquid state, which are dispersed in water to give a heterogeneous mixture.
Dissolved solids (exclusive of gases) are dispersed in water to give a
homogeneous liquid and total solids is the sum of suspended and dissolved
solids.
The sample is titrated with a standard acid solution to a designated pH and the
end point is determined using internal indicator.
12.2 RANGE
12.3 REAGENTS
12.3.3 Standard Hydrochloric Acid (0.02N) - Dilute 1.66m1 of hydrochloric acid (sp gr
1.19) to 1 litre with water. For standardization - see notes.
12.3.4 Mixed Bromocresol Green - Methyl Red Indicator Solution -Dissolve 20mg of
methyl red and 100mg of bromocresol green (sodium salt) in either 100m1 of
water or 100m1 of ethyl alcohol (95%).
12.3.5 Methyl Orange Indicator Solution (0.5g/litre) - Dissolve 0.05g of methyl orange
in water and dilute to 100 m1.
12.4 PROCEDURE
12.4.1.1 Place 50m1 sample in a titration flask and add 2 drops of phenolphthalein
indicator.
12.4.1.2 Titrate over a white surface with 0.02N standard hydrochloric acid from a pink
colour to a colourless end point.
12.4.2.1 Add 3 drops of the mixed indicator to the solution in which the phenolphthalein
alkalinity has been determined.
12.4.2.2 Titrate over a white surface with 0.02N standard hydrochloric acid to the
required end point.
12.4.3.1 Add 2 drops of methyl orange indicator to the solution in which the
phenolphthalein alkalinity has been determined.
12.4.3.2 Titrate over a white surface with 0.02 N standard hydrochloric acid to the
required end point. (At pH 4.6 the colour changes to orange and at pH 4.0 to
pink).
12.5 CALCULATIONS
V
Where:
The following table gives the stoichiometric classification of the three principal
forms of alkalinity present in water.
Where :
P = Phenolphthalein Alkalinity.
12.6 PRECISION
12.7 INTERFERENCES
12.7.1 Free residual chlorine markedly affects the indicator colour response in some
water samples through its bleaching action. It can be removed by the addition of
sodium thiosulphate.
12.7.2 Natural colour or the formation of a precipitate during titration may mask the
colour change.
12.7.3 Salts of weak organic and inorganic acids also affect the titration.
12.8 NOTES
12.8.1 Reagent grade chemicals should be used for preparing the reagents.
12.8.3 The following pH values are suggested as the equivalence points for the
corresponding alkalinity concentration as calcium carbonate:
12.8.3.1 Indicators effective in these ranges which give the most reliable results are
mixed indicator for higher pH values and methyl orange for pH values below
4.6.
carbonate and add 2 drops of 0.1% solution of methyl red in alcohol. Titrate
with hydrochloric acid to the first appearance of a red colour, and boil the
solution carefully until the colour is discharged. Cool to room temperature and
continue the titration. Repeat the process of boiling and titration until a faint red
colour is obtained that is not discharged on further heating.
12.8.5 Sulphuric acid can also be used in place of hydrochloric acid for titration.
13.2 REAGENTS
13.2.2 Hydrochloric acid (0.02N) - Dilute 1.66m1 of hydrochloric acid (sp gr 1.19) to
1 litre with water. For standardization - see notes.
13.3 PROCEDURE
13.3.2 Add quickly, while swirling the flask, 1 ml of strontium chloride solution for
each milligram of carbonate or phosphate ion in the sample aliquot, plus a 4ml
excess.
13.3.3 Stopper the flask loosely, boil the contents for a few seconds, and then cool to
room temperature.
13.3.4 Add 4 drops of phenolphthalein indicator solution and quickly titrate with
standard hydrochloric acid to a colourless end-point.
13.4 CALCULATIONS
N x V1 x 17000
Hydroxide ion, mg/litre as OH = ---------------------------------------
V
Where:
V = milliliters of sample.
13.5 PRECISION
13.5.1 The single operator precision of the method can be expressed as follows:
S = 0.05 mg/litre.
13.6 INTEREFENCES
13.6.2 The effects of carbonates and phosphates are eliminated by the addition of
strontium chloride in excess.
13.7 NOTES
13.7.1 Reagent grade chemicals should be used for preparing the reagents.
14 ALUMINIUM
14.1 SUMMARY OF METHOD
14.2 RANGE
14.3 APPARATUS
14.4 REAGENTS
14.5 CALIBRATION
14.5.1.1 To eleven 100m1 volumetric flasks, add respectively 0.0, 2.0, 4.0, 6.0, 8.0, 10.0,
12.0, 14.0, 16.0, 18.0 and 20.Om1 of standard aluminium solution (4.2, l ml =
0.010 mg A1). Dilute to 100m1 with water.
14.5.1.2 Using the procedure described in section 6.0, develop the colour and measure
the absorbance of each standard at 370nm.
14.5.1.3 Prepare a calibration curve by plotting the absorbance against the concentration.
14.5.2 Iron Correction Curve
14.5.2.1 To nine 100m1 volumetric flasks, add respectively 0.0, 5.0, 10.0, 15.0, 20.0,
25.0, 30.0, 35.0 and 40.0m1 of standard iron solution (4.3, l ml = 0.010 mg Fe)
Dilute to 100m1 with water.
14.5.2.2 Using the procedure described in section 6.0 develop the colour and measure the
absorbance of each standard at 370 and 520nm.
14.6 PROCEDURE
14.6.1 Place 25 ml sample into a beaker or flask, and add 2.0 ml of hydroxylamine
hydrochloride solution. Allow to stand for 30 minutes.
14.6.3 Add 2m1 of sodium acetate solution and allow to stand for 10 minutes.
14.6.4 Measure the absorbance at 370 and 520 nm against the zero standard prepared
by adding the reagents (6.2, 6.3) to 25m1 of water.
14.7 CALCULATIONS
14.7.2 Using the iron correction curve locate the absorbance value measured at 520
nm. At the same iron concentration read the corresponding absorbance at 370
nm. Designate this value as B.
14.7.3 A minus B is the corrected absorbance. Using the aluminium calibration curve,
find the aluminium concentration corresponding to the corrected absorbance.
14.8 PRECISION
So = 0.035
14.9 INTERFERENCES
14.9.1 Each mg/litre of iron increases the aluminium reading by about 0.01 mg/litre.
14.9.2 Several other metals and anions cause interference, but only manganese, lead,
cobalt and fluoride show sufficiently pronounced effects to require correction.
Orthophosphate upto 5mg/litre and residual chlorine up to 5mg/litre do not
interfere. Natural colour and turbidity interfere and a correction is usually
required.
14.10 NOTES
14.10.1 Reagent grade chemicals should be used for preparing the reagents.
14.10.2 Filter the sample at the time of collection through a 0.45 micron filter and
perform the analysis as soon as possible after sample collection.
15.4 PROCEDURE
15.4.1 Calibration
Prepare a reagent blank and sufficient standards containing 0 to 60 mg/l of
aluminium by diluting suitable volumes of standard aluminium solution with
nitric acid (1:499) to 100 ml in volumetric flasks. Add 2 ml of KCl solution to
each of the volumetric flasks. Aspirate the reagent blank and carry out
zero63adjustment. Aspirate sequentially the standard solutions and measure the
absorbance at 309.3 nm.
Add 0.5 ml of concentrated nitric acid to 100 ml of the sample taken in a 250-ml
beaker. (If total recoverable aluminium is to be determined, use 100 ml of the
sample obtained after a pretreatment as given below. Add 5 ml of concentrated
hydrochloric acid to 100ml of the sample taken in a 250-ml beaker. Heat on a
hot plate to reduce the volume to about 50 ml and filter the sample through
0.45mm membrane. Transfer quantitatively the contents of the beaker to a 100ml
volumetric flask. Make up to the mark in volumetric flask.) Add 2ml of KCl
solution to this. Prepare a reagent blank with 100ml of water. Rinse the atomizer
by aspirating water containing 1.5 ml conc HNO3/l. Aspirate the reagent blank
and carry out zero adjustment. Aspirate the sample solution and measure the
absorbance at 309.3 nm. From the absorbance data, determine the micrograms
of aluminium present in 100 ml of the final solution.
15.5 Calculations
Aluminium, mg/l = M
V
Where
The relative standard deviation reported in the literature for aluminium in the
4.5mg/l concentration range is 4.2 percent.
16.2 RANGE
10 to 500 micrograms/litre as N.
16.3 APPARATUS
16.4 REAGENTS
16.5 CALIBRATION
16.5.1 Transfer 0.0, 1.0, 5.0, 10.0, 15.0 and 20.0ml of the standard ammonia nitrogen
solution. (1ml =0.5 microgram N) to 25ml volumetric flasks,
16.5.3 Add 0.5ml of hypochlorous acid. solution and add immediately but slowly
0.6m1 of Phenolate solution. Dilute to 25m1 with water.
16.5.4 Measure the absorbance of each standard at 630nm against the zero standard
(blank).
16.6 PROCEDURE
16.6.1 Place 10ml (or other suitable volume containing not more than 10 micrograms
ammonia nitrogen) of the sample in a 25m1 volumetric flask.
16.6.2 Proceed in accordance with section 5.0 .(5.2 to 5.4).
16.7 CALCULATIONS
A x 1000
Ammonia, micrograms/litre as. NH3 = ------------------------------
V
Where:
V = milliliters of sample.
16.8 INTERFERENCES
16.8.1 More than 500 mg/litre of alkalinity, more than 100 mg / litre of acidity, colour
and turbidity interfere.
17.2 RANGE
0.1 to 2mg/litre as N.
17.3 APPARATUS
17.4 REAGENTS
17.4.3.3 Sodium Potassium Tartarate Solution (300 g/litre) - Dissolve 30g of sodium
potassium tartarate tetrahydrate in 100m1 of water.
17.5 CALIBRATION
17.5.1 Prepare a series of standards by diluting 0.0, 1.0, 3.0, 5.0, 8.0 and l0.0m1 of
standard ammonia nitrogen solution (l ml = 0.01 mg N) to 50m1 in Nessler
cylinders.
17.5.2 Add l ml of Nessler reagent, mix and allow to stand for 20 minutes.
17.5.3 Measure the absorbance of each standard (5.1) at 425nm against the zero
standard (blank) using l0mm matched pair of cells.
17.6 PROCEDURE
17.6.1 Place 50ml sample (or other suitable volume diluted to 50ml) containing not
more than 0.lmg of ammonia nitrogen in a Nessler cylinder.
17.6.2 Add l ml of Nessler reagent, mix, and allow to stand for 20 minutes.
17.6.3 Measure the absorbance of the sample at 425nm against the reagent blank
prepared by taking 50 ml of water in a Nessler cylinder and following the step
6.2.
17.7 CALCULATIONS
A x 1000
Ammonia nitrogen, mg/litre as N = ------------------------
V
Where:
V = milliliters of sample.
17.8 PRECISION
17.9 INTERFERENCES
17.9.1 Glycerine, hydrazine, and some amines will react with Nessler's reagent to give
the characteristic yellow colour in the time required for the test.
Add l ml of Zinc sulphate solution to 100 m1 sample and mix. Add sodium
hydroxide solution to raise the pH to about 10.5 (check with a pH paper). Allow
to settle and filter through whatman No.40 filter paper. To prevent cloudiness
add 2 drops of sodium potassium tartarate solution or disodium dihydrogen
ethylenediamine tetraacetate.
17.10 NOTES
17.10.1 Reagent grade chemicals should be used for preparing the reagents.
17.10.2 The Nesseler reagent should give the characteristic colour with ammonia within
10 minutes after addition, and should not produce a precipitate with small
amounts of ammonia (0.04mg in 50 m1 volume). The solution may be used
without 5 day storage if it is filtered 'through a 0.45 - micron membrane filter
shortly before use.
18.5 REAGENTS
18.6.2 Treat the standards as directed in 1 and measure the potential of each standard
and record in millivolts. The standards and the sample must be at the same
temperature, preferably about 250C.
18.6.3 Using semilogarithmic graph paper, plot the concentration of ammonia nitrogen
in milligrams per litre on the log axis against the corresponding electrode
potential, in millivolts, on the linear axis.
18.6.4 Check the Calibration curve every 3 h when analyzing a series of samples.
18.6.5 ISE Meter – Refer to the manufacturer’s instruction manual for proper operation
of the meter. Prepare calibration curves with three standard solutions (see 15.3),
bracketing the expected concentrations of the samples.
18.6.6 Check the calibration curve every 3 h when analyzing a series of samples;
otherwise, calibrate daily.
18.7 PROCEDURE
18.7.6 Check the pH of the sample with pH paper. The pH must be greater than 11.0. If
less than 11.0, add additional NaOH solution (see 15.5) in 0.1-ml increments
until the pH of the solution exceeds 11.0.
18.7.7 When the electrode comes to equilibrium, measure the electrode potential of the
ammonia nitrogen concentration as directed below.
Note 3 – The time required for the electrode to come to equilibrium is
dependent on the ammonia content of the sample. For concentrations above 0.5
mg/L, the response time is about 30 s.
18.7.8 Sample Measurement – Determine the ammonia nitrogen concentration by
means of a pH meter of a specific-ion meter.
18.7.9 pH Meter – Record the observed potential in millivolts and convert to
milligrams per litre of ammonia nitrogen by means of the calibration curve (see
17.2).
18.7.10 ISE Meter – Record the concentration reading directly from the logarithmic
scale as milligrams of ammonia nitrogen per litre.
18.8 CALCULATIONS
19.2 APPARATUS
pH meter.
19.3 REAGENTS
19.3.2 Hydrochloric Acid, Standard (0.04N) - Dilute 3.42 m1 of hydrochloric acid (sp
gr 1.19) to 1 litre with water and standardize with sodium carbonate (dried at
250°C for 4 hours) using methyl red indicator.
19.3.3 Methyl Red Indicator Solution (5g/litre) - Dissolve 0.5g of methyl red in 100
m1 of 95 % ethanol.
19.4 PROCEDURE
19.4.2 Place 50m1 or less of sample water containing no more than 80mg of
bicarbonate ion into a 200m1 beaker.
19.4.3 If the pH of the sample is above 8.3, titrate with 0.04N hydrochloric acid to this
pH value using the pH meter for end-point detection.
19.5 CALCULATIONS
19.5.1 Calculate the bicarbonate ion concentration using the following equation:
2440 X (V2-Vl)
Bicarbonate (HC03), mg/litre = -------------------------
V
Where:
V = volume of sample in milliliters.
19.5.2 Calculate the free carbon dioxide concentration in mg/ litre by using the
following equation for waters with pH values from 6 to 9:
Where:
20.2 APPARATUS
pH meter.
20.3 REAGENTS
20.3.6 Methyl Red Indicator solution (5g/litre) - Dissolve 0.5g of methyl red in 100 m1
of 95% ethanol.
20.4 PROCEDURE
20.4.1 Place 100m1 of sample in a 250m1 beaker and add 5 drops of phenolphthalein
indicator solution.
20.4.2 If the sample remains colourless, titrate rapidly with standard sodium hydroxide
solution until the first faint pink colour is detectable in the solution.
20.4.2.1 Alternatively, titrate the sample to pH 8.3 using a pH meter to detect the end
point.
20.5 CALCULATIONS
20.5.1 Calculate the free carbondioxide content of the water in mg/litre using the
following equation:
20.6 PRECISION
20.6.1 Under the most favourable conditions, precision is approximately 10% of the
indicated carbondioxide content.
20.7 INTERFERENCES
chromium and copper are examples of metal ions that may yield erroneous
results.
20.7.2 Abnormal results also may be obtained in the presence of ammonia, amines,
phosphate, borate, sulphide and nitrate.
The sample is treated with ferric ammonium sulphate and mercuric thiocyanate
solutions. The chloride ion reacts with mercuric thiocyanate to release the
thiocyanate ion which combines with ferric ion to form red ferric thiocyanate.
The intensity of the colour is measured at a wavelength of 463nm.
21.2 RANGE
21.3 APPARATUS
21.4 REAGENTS
[Fe (NH4)2 (SO4)2.6H20] in 20m1 of water. Add 38m1 of nitric acid (sp gr 1.42)
and boil to oxidise the iron and remove the oxides of nitrogen. Dilute to 100 ml
with water.
21.5 CALIBRATION
21.5.1 Prepare a series of standards by diluting 0, 0.5, 2.5, 5.0, 7.5, 10 and 14m1 of the
standard sodium chloride solution (1ml = 0.01mg chloride) to 100 ml with water
in volumetric flasks.
21.6 PROCEDURE
21.6.2 Add 5.0 ml of ferric alum solution and 2.5ml of mercuric thiocyanate solution,
mix thoroughly and allow to stand for 10 minutes.
21.6.3 Measure the intensity of colour at 463m, against the reagent blank, prepared by
using 25 m1 of water and following step 6.2, using 50 mm matched absorption
cells.
21.7 CALCULATIONS
21.7.1 Read the concentration of chloride ion in mg/litre directly from the calibration
curve prepared in accordance with section 5.0:
21.8 PRECISION
Sr = 0.054 X
So = 0.013 X
Where:
21.9 INTERFERENCES
21.9.3 Colour may also interfere depending upon its spectral absorbance.
21.10 NOTES
21.10.1 Reagent grade chemicals should be used for preparing all the reagents.
21.10.2 Mercuric salts are very poisonous. Due precautions should be observed when
using these salts. -
21.10.4 Soak all new glassware in hot nitric acid (1+19) for several hours and in water
(halide free) between tests. Discard all glassware that appear etched or
scratched.
21.10.5 For best results, the temperatures of the standard solutions should be within 1.0
o
C of the reagent blank, and the samples.
22.2 RANGE
22.3 APPARATUS
22.4 REAGENTS
22.4.3 Lead Nitrate Solution - Dissolve 20g of lead nitrate [Pb(N03)2] in water and
dilute to 1 litre.
22.5 CALIBRATION
22.5.1 Prepare a series of standards by diluting 0, 1.0, 5.0, 10.0, 15.0, and 25.0 m1 of
Standard Sodium Chloride Solution (l ml = 1 microgram of chloride) to 250 m1
in 250 ml glass stoppered bottles.
22.6 PROCEDURE
22.6.2 Add 5.0 ml of the lead nitrate solution to the bottle. Cap the bottle and mix.
Allow to stand for 2 minutes.
22.6.3 Add 5.0m1 of sodium phosphate solution and mix. Allow to stand for 5
minutes.
22.6.4 Centrifuge the capped bottle solution at 1500 rpm for 6 minutes. Decant the
supernatent liquid immediately after centrifuging.
22.6.5 Add 15.0 m1 of ferric nitrate solution and mix to dissolve the precipitate.
22.6.6 Add 1.0 m1 of mercuric thiocyanate solution and mix. Dilute to 25 m1 with
water. Allow to stand for 10 minutes.
22.6.7 Measure the absorbance, against reagent blank prepared by taking 250 m1
instead of sample and repeating the steps 6.2 to 6.6, at 463nm using 50mm
matched cells.
22.7 CALCULATIONS
22.8 INTERFERENCES
22.8.1 See mercuric thiocyanate method (Section 9.0) for the determination of
Chloride.
22.9 NOTES
22.9.1 See mercuric thiocyanate method (Section 10.0) for the determination of
chloride.
22.9.2 Lead nitrate is very toxic. Due precautions should be observed when using this
chemical.
22.9.4 For standard method of further low level determination of Cl, PO4 and SO4 in
high purity water, ASTM D-5542 -94 may be referred.
The sample is adjusted to a pH of 8.3 and titrated with silver nitrate solution
using potassium chromate indicator to a brick red colour.
23.2 RANGE
23.3 REAGENTS
23.3.2 Standard Silver Nitrate Solution (0.025N) - Dissolve 4.247g of silver nitrate
(dried to constant weight at 40°C) in water and dilute to 1 litre. Standardize
against standard sodium chloride solution.
23.3.3 Standard Sodium Chloride Solution (0.025N) - Dissolve 1.461g of sodium
chloride (dried at 600°C for 1 hour) in water and dilute to 1 litre.
23.3.8 Sulphuric Acid Solution (1+19) - Mix 1 volume of sulphuric acid (sp gr 1.84)
with 19 volumes of water.
23.4 PROCEDURE
23.4.2 If Sulphite is present, add 0.5 ml of hydrogen peroxide solution and mix for 1
minute,
23.4.3 Adjust the pH to the phenolphthalein endpoint (pH 8.3), using sulphuric acid
solution, or sodium hydroxide solution.
23.4.5 Titrate with standard silver nitrate solution to a brick red (or pink) colour which
persisted through out the sample when illuminated with a yellow light.
23.4.6 Repeat 4.1 to 4.5 using 25m1 of sample diluted to 50ml with water.
23.5 CALCULATIONS
23.5.1 Calculate the chloride ion concentration in the sample, in milligrams per litre, as
follows:
(V1-V2) x N x 70906
Chloride, mg/litre as Cl = --------------------------------
V
Where:
Where:
23.6 PRECISION
ST = 0.013X + 0.70
So = 0.007X + 0.53
Where:
23.7 INTERFERENCES
23.7.1 Bromide, iodide, and sulphide are titrated along-with the chloride.
23.8 NOTES
23.8.1 Reagent grade chemicals should be used for preparing all the reagents.
23.8.2 If the titration requires more than 25m1 of silver nitrate in 4.5, use a smaller
sample size.
24 CHLORINE DEMAND
24.2 APPARATUS
24.2.1 pH meter.
24.3 REAGENTS
24.3.2 Calcium Hydroxide Solution (10.7g/litre) - Weigh 10.7g of 100% hydrated lime
[Ca(OH)2] and suspend in water. Dilute to 1 litre.
24.3.4 Chlorine Water (l ml = 0.5 to 3mg available chlorine) - Pass gaseous chlorine
through water until the solution contains 0.5 to 3.0 mg available chlorine per ml.
For standardization add l0ml of chlorine water to a flask containing l0ml of
acetic acid (1+1) and l0ml of potassium iodide solution (5%). Titrate with 0.l0N
sodium thiosulphate solution using starch indicator.
V1 x 3.546
24.3.5 Hydrochloric Acid (1+1) - Mix equal volumes of hydrochloric acid (sp gr
1.19) and water.
24.4 PROCEDURE
24.4.1.1 Ascertain the range of pH, time of chlorine contact, and the chlorine application
concentration to achieve the objective of Chlorination from past experience,
from literature survey, by experimentation, or from plant conditions.
24.4.1.2 Determine the pH of each test and additions of chlorinating solutions such that
there is not less than five equal increments of the chlorine application
concentrations.
24.4.1.3 In each of a series of clean 1 litre glass containers, place a 500 ml portion of the
sample.
24.4.2.1 To the first of the series of 500 ml portions of the samples, add the maximum
anticipated amount of chlorinating solution. Determine the pH of the solution.
24.4.2.3 Allow the chlorinated sample to stand for a minimum predetermined time.
Determine total and free available residual chlorine. Withdraw successive
samples at selected time intervals to cover the estimated range of minimum to
maximum contact times.
24.4.3 Chlorination
24.4.3.1 On the basis of information obtained by the trial chlorination, add desired
increments of chlorinating solution to separate 500 m1 portions of the sample.
Determine the pH of each portion.
24.4.3.3 Allow each portion of chlorinated sample to stand for a predetermined time.
Withdraw a portion of the sample and determine the total residual chlorine.
24.5 CALCULATIONS
24.5.1 Calculate the chlorine dosage, in mg/litre, for each increment of chlorinating
solution as follows:
Where:
24.5.2 Determine the chlorine consumed, mg/litre, for each increment of chlorine
application as follows:
On log-log graph paper, plot, for a given chlorine application, temperature, and
pH, the chlorine consumed versus the contact time in hours. Determine the
value of the chlorine consumed at the intercept of the line with the co-ordinate
corresponding to a contact time of 1 hour. Designate the value of this intercept
as K. Determine the slop of the line and designate as n. The straight lines for
each chlorine application at each temperature and pH are of the general form:
DT = KT t n
Where :
n = Slope of curve.
24.6 NOTES
24.6.2 When the anticipated chlorine requirement is less than 600mg/litre, use the
chlorinating solution which is to be used in ultimate plant treatment. When the
anticipated chlorine requirement is 600 mg/litre or more, use the appropriate
hypochlorite solution.
24.6.3 pH Adjustment
24.6.3.1 If the pH of the chlorinated sample is higher than the desired range, add
hydrochloric acid (1+1) to the chlorinated sample until the pH of sample
reaches the upper limit of the desired range.
24.6.3.2 If the pH of the chlorinated sample is lower than the desired range, discard the
sample and proceed with another series of sample portions, as follows: add
24.6.3.3 If the chlorinating solution is chlorine water, add an additional 0.l ml of calcium
hydroxide solution for each milligram of available chlorine to be applied to the
sample.
In the absence of iodide ion, free chlorine reacts with para-amino diethylaniline
(NN-Diethyl-p-Phenylene Diamine abbreviated as DPD) to produce a red
colour. Stepwise colour change is carried out to identify monochloramine,
dichloramine, and nitrogen trichloride. The individual fractions are determined
by titration with ferrous ammonium sulphate.
25.2 RANGE
25.3 REAGENTS
25.3.4 Sodium Arsenite Solution- Dissolve 0.5 g sodium. arsenite (NaAsO2) in 100 m1
of water.
25.3.6 Potassium Iodide solution (5g/litre)- Dissolve 0.5g of potassium iodide in water
and dilute to 100m1. Store in a brown coloured glass bottle. Discard when
yellow colour develops.
25.4 PROCEDURE
25.4.1.1 Place 5.0 m1 of DPD solution and 5.0m1 of phosphate buffer in a 250m1
titration flask.
25.4.1.3 Titrate with ferrous ammonium sulphate solution until the red colour is
discharged.
25.4.1.4 Record the volume of ferrous ammonium sulphate in ml used in the titration as
A.
25.4.2.1 To the solution after titration for free chlorine (4.1) add 2 drops of potassium
iodide solution (5g/litre) and continue the titration.
25.4.2.2 Record the total volume of ferrous ammonium sulphate in ml as B. For free
chlorine + monochloramine)
25.4.3.1 To the solution after titration for monochloramine (4.2) add about l g of
potassium iodide and mix rapidly to dissolve, and allow to stand for 2 minutes.
25.4.3.3 Record the total volume of ferrous ammonium sulphate in ml as C (for free
chlorine + monochloramine + dichloramine).
25.4.4.1 Place a small crystal of potassium iodide in a 250m1 titration flask, add 100m1
of sample and mix.
25.4.4.2 Transfer the contents (4.4.1) to another flask containing 5m1 each of buffer
solution and DPD solution.
25.4.5.1 Place 1g of potassium iodide in a 250m1 titration flask, add 100m1 of sample
and mix.
25.4.5.2 Transfer the contents (4.5.1) to another flask containing 5 m1 each of buffer
solution and DPD solution and allow to stand for 2 minutes.
25.4.5.3 Titrate with ferrous ammonium sulphate solution and record the volume in ml
as V1.
ST = 0.008 X + 0.9
Where:
ST = overall precision.
25.5 INTERFERENCES
25.5.1 Cr interferes when its conc. exceeds 5 times of Cu. The interference from
organic matter, sulphide, cyanide and chromium can be eliminated by as
follows:
25.6 NOTES
25.6.1 Reagent grade chemicals should be used for preparing the reagents.
25.6.2 A polythene bottle must be used for sample collection. Hydrochloric acid (sp gr
1.19) should be added to the filtered sample for total recoverable copper
immediately at the time of collection. The volume of acid should be sufficient to
neutralize the sample to pH 4 (using narrow range pH paper) and then add 2.0
m1 for each litre of sample.
25.6.3 Soak all new glassware in hot nitric acid (1+9) for several hours. To ensure the
conditioning of glassware, rinse it with water and run a copper determination
(blank) on copper free water. Repeat until the copper value is less than 4
micrograms per litre. After carrying out the test, always rinse the glassware with
organic solvent, followed by water. Always keep the glassware soaked in nitric
acid (1+9) until used again. Discard any glassware that appears etched or
scratched.
25.6.4 If the sample contains more than maximum concentration of copper specified in
the range, a smaller size sample should be diluted to 200m1 with. copper free
water containing 0.4m1 of hydrochloric acid (sp gr 1.19) per 200 m1 of
solution.
25.6.6 The blank determination made for calibration in section 5.0 compensates for
copper in both the reagents and 200 m1 of water. When the test water contains
less than 10 micrograms/litre of copper, it is important (in 6.7) to compensate
only for the copper in the reagents and not to include the few micrograms per
litre of copper found in copper free water.
The reagent blank is found, by extracting the copper, from two 200 m1 aliquots
of copper free water. In one aliquot the normal values of reagents i.e.
hydrochloric acid, hydroxylamine hydrochloride, sodium acetate and
neocuprine solution are used and in the other aliquot twice the normal values of
reagents are used. The organic extract from the normal blank is used as
reference solution for initial spectrophotometer setting and the blank obtained
from double reagents is measured against the normal blank. The correct value
for copper is found in the unknown sample (6.7) by subtracting from it the value
for the reagent blank.
The sample is treated with potassium iodide at a pH between 3.0 and 4.0. The
librated iodine which is stoichiometrically proportional to the total chlorine
present is titrated with standard solution of sodium thiosulphate.
26.2 RANGE
0.04mg/litre or more, as Cl
26.3 REAGENTS
26.3.5 Sodium Thiosulphate Standard solution (0.01N)- Dilute 25m1 of 0.1N sodium
thiosulphate solution (3.4) to 250 m1 in a volumetric flask.
26.3.6 Starch Indicator Solution - Make a paste of l g of soluble starch and mix into 1
litre of boiling water. Add 20 g of potassium hydroxide, mix, and allow to stand
for 2 hours. Add 6 m1 glacial acetic acid, mix, and add sufficient hydrochloric
acid (sp gr 1.19) to adjust the pH to 4.0 (check. with a narrow range pH paper).
This has a shelf life of 1 year.
26.4 PROCEDURE
26.4.1 For residual chlorine concentrations of l0 mg/litre or less, use a 200 m1 sample,
and for concentrations above l0 mg/litre, use proportionately less sample.
26.4.2 Place 5 m1 or other required volume of glacial acetic acid in a 500 m1 titration
flask in order to reduce the pH of the sample between 3.0 to 4.0.(To be checked
before hand with narrow range pH paper)
26.5 CALCULATIONS
the sample.
titration.
V= milliliters of sample.
26.6 INTERFERENCES
26.6.1 Free available chlorine cannot be determined quantitatively when organic matter
is present that rapidly and continuously consume free available chlorine.
26.7 NOTES
26.7.1 Reagent grade chemicals should be used for preparing the reagents.
26.7.2 A suggested method for preparation of chlorine demand free water is to add
approximately 20mg/litre of available chlorine to Type III reagent grade water.
Allow the chlorinated water to stand about 1 week in the absence of sunlight
until no residual chlorine remains.
26.7.3 Test should be made as soon as possible after collection of the sample (not more
than 5 minutes) because the residual chlorine may diminish with time, due to
chlorine demand of the-sample.
W
N= ---------------------
0.04904 x V
Where:
27.2 RANGE
27.3 APARATUS
27.4 REAGENTS
27.4.2 Copper Stock Solution (200mg/litre) - Place 0.200g electrolytic grade copper in
a 250m1 beaker, add 3m1 of water and 3ml of nitric acid (sp gr 1.42).After the
metal has completely dissolved, add l ml sulphuric acid (sp gr 1.84) and
evaporate on a hot plate to nearly dryness. Dissolve the residue in water and
dilute to 1 litre.
27.4.3 Copper Standard Solution (2mg/litre) - Dilute 100m1 of copper stock solution
(4.2) to 1 - litre. Again dilute 100 m1 of this diluted solution to 1 litre.
27.4.8 Chloroform Solvent - Mix 9 volumes of chloroform (CHCl3) with one volume
of isopropyl alcohol.
27.5 CALIBRATION
27.5.1 Prepare a series of standards (at least five concentrations) to cover the expected
range of copper concentrations by diluting appropriate volumes of copper
standard solution (4.3, l ml = 2 micrograms Cu) as follows:
27.5.1.1 Place the required volumes of copper standard solution (4.3) in 250m1
separatory funnels.
27.5.1.2 Add 0.4 ml hydrochloric acid (sp gr 1.19) to each funnel and add water to make
200m1.
27.5.1.3 Prepare a blank (zero standard) by diluting 0.4 m1 hydrochloric acid (sp gr
1.19) to 200 m1 with water.
27.5.1.4 Proceed in accordance with section 6.0 (6.2 to 6.7) and measure the absorbance
of each individual standard.
27.5.1.5 Use the organic liquid from the blank as a reference solution for the initial
spectrophotometer setting.
27.5.1.6 Prepare a calibration curve by plotting the absorbance of the standards against
the copper' content in micrograms.
27.6 PROCEDURE
27.6.1 Transfer 200 ml of acidified (with 0.4m1 hydrochloric acid, sp gr 1.19) and
unfiltered sample (for total. copper) or 200m1 of filtered (through 0.45 micron
filter) and acidified sample (for dissolved copper) into a 250m1 separatory
funnel.
27.6.4 Add 10 m1 of neocuprine solution and shake the funnel and contents for 1
minute.
27.6.5 Add 20 m1 of chloroform solvent shake vigorously for at least 1 minute and
allow to stand for 5 minutes.
27.6.6 Transfer the organic layer into a dry 50 m1 Erlenmeyer flask and add 20 ml of
isopropyl alcohol to clear the solution. Make upto 50 m1 with chloroform
solvent or isolamyl alcohol depending on the extractant used.
27.6.7 Measure the absorbance of the organic solution (6.6) at 457 nm (when
chloroform solvent is the extractant) using a mixture of 25 m1 of chloroform
solvent and l0 ml of isopropyl alcohol as a reference solution for initial
spectrophotometer setting or at 454 nm (when isoamyl alcohol is the extractant)
using a mixture of 25 m1 of isoamyl alcohol and l0 ml of isopropyl alcohol as a
reference solution.
27.6.8 Carry out a blank determination on 200m1 of water, with all reagents and
extracting in the same manner as for the sample.
27.7 CALCULATIONS
27.7.1 Calculate the concentration of copper in micrograms per litre in the sample, as
follows:
27.8 PRECISION
ST = 0.008 X + 0.9
Where:
ST = overall precision.
27.9 INTERFERENCES
27.9.1 Cr interferes when its conc. exceeds 5 times of Cu. The interference from
organic matter, sulphide, cyanide and chromium can be eliminated by as
follows:
27.10 NOTES
27.10.1 Reagent grade chemicals should be used for preparing the reagents.
27.10.2 A polythene bottle must be used for sample collection. Hydrochloric acid (sp gr
1.19) should be added to the filtered sample for total recoverable copper
immediately at the time of collection. The volume of acid should be sufficient to
neutralize the sample to pH 4 (using narrow range pH paper) and then add 2.0
m1 for each litre of sample.
27.10.3 Soak all new glassware in hot nitric acid (1+9) for several hours. To ensure the
conditioning of glassware, rinse it with water and run a copper determination
(blank) on copper free water. Repeat until the copper value is less than 4
micrograms per litre. After carrying out the test, always rinse the glassware with
organic solvent, followed by water. Always keep the glassware soaked in nitric
acid (1+9) until used again. Discard any glassware that appears etched or
scratched.
27.10.4 If the sample contains more than maximum concentration of copper specified in
the range, a smaller size sample should be diluted to 200m1 with. copper free
27.10.6 The blank determination made for calibration in section 5.0 compensates for
copper in both the reagents and 200 m1 of water. When the test water contains
less than 10 micrograms/litre of copper, it is important (in 6.7) to compensate
only for the copper in the reagents and not to include the few micrograms per
litre of copper found in copper free water.
The reagent blank is found, by extracting the copper, from two 200 m1 aliquots
of copper free water. In one aliquot the normal values of reagents i.e.
hydrochloric acid, hydroxylamine hydrochloride, sodium acetate and
neocuprine solution are used and in the other aliquot twice the normal values of
reagents are used. The organic extract from the normal blank is used as
reference solution for initial spectrophotometer setting and the blank obtained
from double reagents is measured against the normal blank. The correct value
for copper is found in the unknown sample (6.7) by subtracting from it the value
for the reagent blank.
28.2 APPARATUS
Data storage and reduction device: This shall be utilized for collection, storage,
reduction and problem recognition. Automatic sampling is recommended.
Inject a measured aliquot of sample into the furnace device following the
direction as provided by the particular instrument manufacturer.
29.1.1 Equivalent Mineral Acidity - The sample is passed through a column containing
strong acid cation resin in hydrogen form and effluent produced is titrated with
a standard solution of sodium hydroxide using screened methyl orange as
indicator.
29.1.2 Free Mineral Acidity - The sample is titrated with a standard solution of sodium
hydroxide using screened methyl orange as indicator.
29.2 REAGENTS
29.2.5 Screened Methyl Orange Indicator Solution - Dissolve 0.02g of methyl orange
in a mixture of 28 m1 of methylated spirit and 25 m1 of water. (Solution A).
29.3 PROCEDURE
29.3.1.1 Transfer 50m1 of strong acid cation resin into a cation column. Backwash with
water maintaining 50 percent expansion of the bed. Continue backwashing until
the effluent is clear.
29.3.1.3 Pass sample through the column. Reject first 50m1 and collect 100m1 sample in
a 500 m1 conical flask.
29.3.1.4 Add 2-3 drops of screened methyl orange indicator and titrate with standard
sodium hydroxide solution till the sample turns smoky grey.
29.3.2.2 Add 2-3 drops of screened methyl orange indicator and titrate with standard
sodium hydroxide solution till the sample turns smoky grey.
29.4 CALCULATIONS
V1 X N X 50,000
EMA, mg/litre as CaCO3 = -----------------------
V
Where:
V = milliliters of sample.
V2 X N X 50,000
V = milliliters of sample.
29.5 NOTES
29.5.1 Reagent grade chemicals should be used for preparing the reagents.
29.5.2 For standardization of 0.02N sodium hydroxide solution place 0.19 + 0.005g of
potassium hydrogen phthalate (dried at 120°C for 2 hours) in a 500 m1 conical
flask. Add 100 m1 of water and 3 drops of phenolphthalein indicator (1%
solution in ethanol) and titrate with sodium hydroxide solution.
W
N = -----------------
(0.20423xV)
Where:
The fluoride is distilled from the sample as hydrofluorosilicic acid which is then
treated with SPANDS dye. Decrease in the color-intensity of SPANDS dye
which is proportional to the concentration of fluoride is measured at a
wavelength of 570nm.
30.2 RANGE
30.3 APPARATUS
30.4 REAGENTS
30.4.4 Acid Zirconyl - SPANDS Reagent - Mix equal volumes of Zirconyl acid
reagent and SPANDS solution.
30.5 CALIBRATION
30.5.1 Prepare a series of standards in the range from 0 to 1.40 mg/litre by diluting
standard sodium fluoride solution (l ml = 0.01 mg F) to 50 m1 with water in a
volumetric flask.
30.5.2 If the sample to be analyzed contains free chlorine, add 2 drops of sodium
arsenite solution to each standard.
30.5.3 Proceed in accordance with section 6.0 (6.2.1, 6.2.3 and 6.2.4).
30.6 PROCEDURE
30.6.1 Distillation
30.6.1.1 Place 400 m1 of water in the distillation flask and add 200 m1 of sulphuric acid
(sp gr 1.84). Assemble the apparatus and heat the solution to 180°C. Discard the
distillate.
30.6.1.2 Cool the solution to below 100 oC, add slowly 300 m1 of sample, mix
thoroughly and distill at 180°C.
30.6.1.3 Stop the distillation when the volume of distillate reaches 300m1.
30.6.2 Analysis
30.6.2.1 Place 50 m1 of the distillate (or sample, if distillation not required) in a 100 m1
polythene bottle.
30.6.2.2 If the sample contains free chlorine, add 1 drop of sodium arsenite solution for
each 0.l mg of chlorine present. Add 2 drops of sodium arsenite solution in
excess.
30.6.2.4 Measure the absorbance against reference solution (4.5) at 510 nm using l0 mm
matched absorption cells.
30.7 CALCULATIONS
30.7.1 Calculate the concentration of fluoride in the sample, in mg/litre, directly from
the calibration curve prepared in accordance with section 5.0.
Where:
30.8 PRECISION
30.9 INTFRFERENCES
30.9.1.1 Aluminium in excess of 300 mg/litre, colloidal silica in excess of 400 mg/litre,
and silica as silicate in excess of 300 mg/litre will retard fluoride recovery.
30.9.1.2 Fluoride in excess of 3.0 mg/litre will hold up in the condenser to a certain
extent causing low results and act as a positive interference for subsequent
samples of lower fluoride content. In such cases, the condenser should be
flushed with 300 to 400 m1 of water and added to the distillate.
30.9.1.3 Chloride concentrations greater than 2500 mg/litre may distil over and act as an
interference in the subsequent colour reaction. To eliminate chloride
interference, add silver sulphate (Ag2S04) to the distillation mixture (6.1.2) at a
rate of 5 mg/litre of chloride.
30.9.1.4 Samples containing oily matter should be extracted initially with a suitable
solvent to remove the oily material and then warmed on a steam bath to remove
traces of the solvent.
30.9.2 The Table 1 lists the interferences of various constituents on the SPANDS
procedure.
TABLE 1
Sulphate (SO4-2)
30.10 NOTES
30.10.1 Reagent grade chemicals should be used for preparing all the reagents.
30.10.2 Distillation (6.1) may be omitted without loss of precision if the concentration
of interfering substances are below the values given in Table 1 (9.2).
30.10.3 The SPANDS solution (4.3) is stable indefinitely if stored in a well sealed bottle
protected from sunlight.
30.10.4 A new standard curve (5.0) should be prepared whenever a new batch of
SPANDS reagent is prepared.
30.10.6 The acid water distillation solution (6.1.1) may be used repeatedly until the
build up of interfering materials equal the concentration given in section 9.0.
30.10.7 The 50m1 distillate or sample (6.2.1) should contain less than 0.028 mg of
fluoride. If fluoride is more than 0.028 mg, a smaller aliquot diluted to 50 m1
should be used.
30.10.9 The use of reference solution (4.5) permits the analyst to adjust the instrument
so that the meter reading of the blank may be used on the optimum portion of
the meter scale, 0.0 to 0.8 absorbance units.
30.10.9.1 Instrument can also be adjusted to some convenient point using the 0.00 mg F
standard. The fixed reading of 0.80 absorbance is suggested.
31.2 RANGE
31.3 REAGENTS
31.3.4 Ammonium Purpurate - Mix thoroughly 1.0g of ammonium purpurate with 200
g of sucrose.
31.4 PROCEDURE
31.4.1.1 Pipet 50.0 m1 of sample into a titration flask and adjust the pH to 7-10 by the
dropwise addition of ammonium hydroxide (sp gr 0.90).
31.4.1.4 Titrate with EDTA standard solution. The end point will be indicated by colour
change from pink to clear blue.
31.4.2.1 Pipet 50.0 m1 of sample into a titration flask and add l ml of sodium hydroxide
solution.
31.4.2.3 Titrate with EDTA standard solution. The endpoint will be indicated by colour
change from pink to purple.
31.4.2.4 Record the volume of EDTA solution required to titrate the calcium.
31.4.2.5 Determine a reagent blank correction by similarly 50.0m1 of water including all
added reagents.
31.5 CALCULATIONS
V1 x M
31.5.1 Total hardness, mg/litre as CaCO3 = ------------------------- x 10,000
V2 x M
31.5.2 Calcium hardness, mg/litre as CaCO3 = ------------------------- x 10,000
Where:
31.6 PRECISION
31.6.1 The precision of this method for calcium (13 to 88 mg/ litre as Ca) may be
expressed as follows:
Sr = 0.006 X +0.62
So = 0.006 X +0.51
Where:
Sr = overall precision.
So = single operator precision.
31.6.2 The precision of this method for magnesium (2.5 to 36 mg/ litre, as Mg) may be
expressed as follows:
ST = 0.017 X + 0.85
So = 0.002 X + 0.70
Where:
ST = overall precision.
31.7 INTERFERENCES
31.7.1 EDTA reacts with several metallic ions. The interference due to these ions can
be minimized by addition of hydroxylamine and cyanide. Metal concentrations
as high as 5mg/litre Fe, l0mg/ litre Mn, l0 mg/litre Cu, l0 mg/litre Zn and l0
mg/litre Pb can be tolerated when hydroxylamine and cyanide are added.
31.7.2 In the titration of total hardness the higher oxidation states of manganese above
2 reacts rapidly with the indicator to form discoloured oxidation products.
Hydroxylamine hydrochloride reagent is used to reduce manganese to divalent
state. The divalent manganese interference can be eliminated by addition of one
or two small crystals of potassium ferrocyanide.
31.7.4 In the titration of calcium, ammonium purpurate reacts with strontium but not
with magnesium or barium. In the presence of strontium, the endpoint is slow
and the titration is not strictly stoichiometric. Barium does not titrate as calcium,
but affects the indicator in some unknown way so that no endpoint or a poor
endpoint is obtained. Barium can be removed by precipitation with sulphuric
acid.
31.8 NOTES
31.8.1 Reagent grade chemicals should be used for preparing the reagents.
31.8.2 If total recoverable calcium and magnesium concentration are being determined,
acidify the sample with nitric acid (sp gr 1.42) to a pH of 2 or less (check with
the help of narrow range pH paper) immediately at the time of collection;
normally about 2 m1/litre is required.
31.8.3 If dissolved calcium and magnesium concentrations are being determined, filter
the samples through a 0.45 micron membrane filter and acidify the filtrate with
nitric acid (sp gr 1.42), 2 m1/litre.
31.8.4 The upper and lower limits of concentration given in range (3.0) may be
extended either by dilution or use of micro apparatus.
31.8.5 The titration of the sample with EDTA should be completed within 5 minutes of
the buffer addition. If more than 15 m1 titrant is required, take a smaller sample
aliquot and repeat the test.
32.2 RANGE
32.3 APPARATUS
32.4 REAGENTS
32.4.4 Hydrochloric Acid (1+9) - Mix 1 volume of hydrochloric acid (sp gr 1.19) with
9 volumes of water.
32.4.5 Hydrochloric Acid (1+99) – Mix 1volume of hydrochloric acid (sp gr 1.19) with
99 volumes of water.
32.5 CALIBRATION
32.5.2 Pipet 50m1 portions of the hydrazine standard solutions as prepared in section
5.1 into 100 m1 cylinders and proceed in accordance with section 6.0 (6.3-6.4).
32.6 PROCEDURE
32.6.2 Transfer a portion of the sample, to the above cylinder (6.1) that will contain
approximately 0.20 to 5.0 micrograms of hydrazine and make the final volume
to 50m1 with water.
32.7 CALCULATIONS
W x 1000
Hydrazine, micrograms/litre = ----------------------------------
Where:
V = milliliters of sample.
32.8 PRECISION
So = (0.99 X + 0.041)/ V
St = (1.08 X + 0.081)/V
Where:
32.9 INTERFERENCES
32.9.1 The hydrazine content may be diminished by oxidising agents collected with the
sample or absorbed by it prior to testing.
32.10 NOTES
32.10.1 Reagent grade chemicals should be used for preparing the reagents.
32.10.4 The sample should be analyzed as quickly as possible after collection since
hydrazine undergoes auto-oxidation, as well as, oxidation by oxidising agents.
Such agents may be in the sample or may enter the sample from the atmosphere.
If it is suspected that oxidation of the hydrazine in the sample is occurring in the
interval between collection and analysis or if the sample is not to be analyzed
immediately then the sample is to be collected under acid by placing 5.0 m1 of
hydrochloric acid (1+9) in a 50m1 volumetric flask, and collecting sufficient
sample to make total volume to 50 m1.
When the sample is collected under acid, the step 6.1 of section 6.0 should be
deleted and in step 6.2 hydrochloric acid (1+99) to be used, instead of water for
dilution.
33.2 RANGE
33.3 APPARATUS
33.4 REAGENTS
33.4.5 Hydrochloric Acid (1+1) - Mix equal volumes of hydrochloric acid (sp gr 1.19)
and water.
33.4.7 Iron Standard Solution (l ml = 1 microgram Fe) - Dissolve 0.1000g of pure iron
in l0ml of hydrochloric acid (1+1) and l ml of bromine water. Boil to remove
excess bromine. Add 200m1 of hydrochloric acid (1+1), cool, and dilute to 1
litre with water (solution A). To l0ml of solution A add 12m1 of hydrochloric
acid (1+1) and dilute to 1 litre with water.
33.4.8 Hydrochloric Acid (1+9) - Mix 1 volume of hydrochloric acid (sp gr 1.19) with
9 volumes of water.
33.5 CALIBRATION
33.5.1 Prepare a series of standards (at least five concentrations) to cover the expected
range of iron concentrations by diluting appropriate volumes of Iron standard
solution (4.7, l ml = 1 microgram Fe) as follows:
33.5.1.1 Place the required volumes of Iron standard solution (4.7) in 125m1 separatory
funnels.
33.5.1.5. Add ammonium hydroxide (1+1) dropwise with mixing until a distinct turbidity
forms. Add hydrochloric acid (1+9) dropwise with mixing until 1 drop clears
the solution. Allow to stand for 1 minute.
33.5.1.7 Simultaneously carry out a blank determination containing no added iron using
50m1 of water and all reagents.
33.6 PROCEDURE
33.6.1 Transfer a volume of sample (filtered through 0.45 micron membrane filter)
containing not more than 8 micrograms of iron, to a 125m1 separatory funnel.
33.6.4 Add ammonium hydroxide (1+1) dropwise with mixing until a distinct turbidity
forms. Add hydrochloric acid (1+9) dropwise with mixing until 1 drop clears
the solution. Allow to stand for 1 minute.
33.6.5 Add 15.0m1 of n-hexyl or isoamyl alcohol and shake vigorously for 1 minute.
Allow to stand for 15 minutes.
33.6.6 Discard the aqueous layer and transfer the alcohol layer into a 25m1 volumetric
flask.
33.6.7 Add l0ml of methyl, ethyl or isopropyl/alcohol to the funnel and wash the
internal surfaces by rolling and tumbling the funnel. Transfer this alcohol into
the previous alcohol extract (6.6). Dilute to the 25m1 mark with the alcohol
used for extraction (6.6).
33.6.8 Measure the colour of the alcohol solution at 533nm, adjusting the
spectrophotometer to zero absorbance reading with a reference solution of 15m1
of alcohol used in step 6.5 and l0ml of alcohol used in step 6.7.
33.6.9 Carry out a blank determination on 50m1 of water, with all reagents and
extracting in the same manner as for the sample.
33.7 CALCULATIONS
33.7.1 Calculate the concentration of iron in micrograms per litre in the sample as
follows:
W x 1000
Iron, micrograms/litre = ----------------------------
V
Where:
W = micrograms of iron, read from the calibration curve:
33.8 PRECISION
33.8.1 The single operator and overall precision varies with the determined
concentration and may be expressed as follows:
So = 0.008 X + 0.92
St = 0.039 X + 1.47
Where:
33.9 INTERFERENCES
33.9.1 If pH is between 3.3 and 3.7 a l mg/litre concentration of the following ions
does not interfere with the test: copper, manganese, aluminium, zinc,
magnesium, sodium, silica, nitrate and orthophosphate.
33.10 NOTES
33.10.1 Reagent grade chemicals should be used for preparing the reagents.
33.10.3 Soak all new glassware in hot hydrochloric acid (1+1) for 2 hours. Drain and
rinse at least 3 times with iron free water. Before and after use, clean all
glassware by making an iron extraction of each piece (without separating the
alcohol - water layers). Drain and flush with iron free methyl alcohol, ethyl
alcohol, or isopropyl alcohol. ,
33.10.4 If iron content is high in hydrochloric acid (4.5) causing a high blank, distil in
an all glass apparatus, rejecting the first 50m1 and the last 100m1 of distillate.
make an extraction with alcohol alone and allow to stand for a long time to
remove all of the alcohol layer. Discard the alcohol layer.
33.10.6 For total iron determination, heat the sample for 1 hour at 60 0°C with 4m1 of
hydrochloric acid (1+1) and 2m1 of hydroxylamine hydrochloride solution.
Thioglycolic acid can also be used for solubilising unreactive iron.
Soluble iron [both Fe (II) and Fe (III)) is complexed with ferrozine [3-(2-
pyridyl) - 5,6 bis (4 phenylsulphonic acid) - 1,2,4 triazine, disodium salt) at pH
4.4 and its concentration, is determined colorimetrically at a wavelength of
562nm.
34.2 RANGE
34.3 APPARATUS
34.4 REAGENTS
34.4.2 Combined Reagent and Ammonia/Acetic Acid Buffer Solution - Dissolve 2.00g
of ferrozine in approximately 300 m1 of water and add 253m1 of acetic acid.
Add slowly and with stirring 200 m1 of ammonium hydroxide (sp gr 0.90), cool
and dilute to 1 litre. Store in a borosilicate glass bottle: The pH of this solution
should be approximately 5.1. This solution is stable for at least 2 months.
34.4.3 Iron Standard Solution (l ml = 1 microgram Fe) - Dissolve 0.1000g of pure iron
in l0ml of hydrochloric acid (1+1) and l ml of bromine water. Boil to remove
excess bromine. Add 200m1 of hydrochloric acid (1+1), cool, and dilute to 1
litre with water (Solution A). To l0 ml of solution A add 12 m1 of hydrochloric
acid (1+1) and dilute to 1 litre with water.
34.5 CALIBRATION
34.5.1 Prepare a series of standards (at least five concentrations) to cover the expected
range of iron concentrations by diluting appropriate volumes of Iron Standard
solution (4.5, l ml =1 microgram Fe) as follows:
34.5.1.1 Place the required volumes of iron standard solution (4.5) in 50ml volumetric
flasks.
34.5.1.3 Add 4.0 ml of the combined reagent/ buffer solution and mix
34.5.1.5 Measure the absorbance of each standard at 562 nm against the blank prepared
by mixing 46 ml of water and 4 ml of combined reagent/ buffer solution.
34.5.1.6 Prepare a calibration curve by plotting the absorbance of the standards against
the iron content in micrograms /litre.
34.6 PROCEDURE
34.6.2 Add 4.0 ml of the combined reagent / buffer solution and mix and make up to
50ml mark with water.
34.6.3 Measure the absorbance of the solution at 562nm against water as reference. Let
the absorbance be A1.
34.6.4 For reagent blank determination, place 40ml of the combined reagent/buffer
solution in a 50ml volumetric flask. Repeat stages 6.3 and 6.4. Let the
absorbance be A2.
34.7 CALCULATIONS
A3 = A1 – 0.1 A2
34.7.2 From A3 and the calibration curve, calculate the concentration of iron in the
sample.
34.8 INTERFERENCES
34.8.1 100 micrograms of copper per litre gives rise to a value of iron 1.4 micrograms
per litre. None of the other substances, likely to be present, cause significant
interference.
34.9 NOTES
34.9.1 Reagent grade chemicals should be used fore preparing the reagents.
34.9.2 If ammonium hydroxide solution (4.2) gives a high blank, purify it as follows:
34.9.3 Place 600ml of ammonia (sp gr 0.90) in the base of a 25 cm desiccator and 400
ml of water in a polythene beaker supported above the ammonia. (The
polythene beaker should be filled with ammonia and kept for 1 day and then
washed thoroughly with water before use). Replace the lid of the desiccator and
store in a cool place. After several days, the solution in the polythene beaker
will be approximately 9N with respect to ammonia, actual concentration can be
checked by titration with a standard solution of hydrochloric acid.
34.9.5 Distil 800ml of acetic acid in an all glass apparatus. Reject the first 100ml
distillate and the last 100ml remaining in the distillation flask.
35.1 PRINCIPLE
This test method covers the determination of dissolved and total recoverable
iron in most waters and waste waters in the range from 5 to 100 µg/L of iron
using a 20 ml µL injection. The iron content of the sample is determined by
atomic absorption spectrophotometer used in conjunction with a graphite
furnace. A sample is place in a graphite tube , evaporated to dryness, charred
and atomized. The absorption signal generated during atomization is recorded
and compared to standards.
35.2 APPARATUS
Atomic absorption spectrophotometer for use at 248.3 nm with background
correction. The manufacturer’s recommendation is to be followed for all
instrumental parameters.
Iron hallow cathode lamp: A single element lamp is preferred but multielement
lamps may be used.
Data storage and reduction device: It shall be utilized for collection, storage,
reduction and problem recognition. Automatic sampling should be used if
possible.
35.3 REAGENTS
(a) Iron solution: Stock (1.0 ml = 1000 µg of Fe) – Dissolve 1.0 g of pure iron in
100 ml of HCL (1+1) with the aid of heat. Cool and dilute to 1 L with water.
(b) Iron solution intermediate (1.0 ml = 10 µg of Fe): Dilute 10.0 ml of above stock
solution (a) and 1 ml of HNO3 (Sp.Gr. 1.42) to 1 L with water
(c) Iron solution intermediate (1.0 ml = 0.2 µg of Fe): Dilute 20.0 ml of above stock
solution (b) and 1 ml of HNO3 (Sp.GR 1.42) to 1 L with water. This standard is
used to prepare the working standards at the time of analysis.
(e) Argon standard, welder grade, commercially available. Nitrogen may also
be used if recommended by the manufacturer.
35.4 Inject the measured aliquots of sample into the furnace device following the
directions as provided by the particular instrument manufacturer.
36.2 RANGE
36.3 APPARATUS
36.4 REAGENTS
36.4.6 Nickel standard Solution (1ml = 0.02mg Ni) – place 1.00g of Nickel (minimum
purity 99.9%) in a 250 ml beaker. Add 10ml each of water and nitric acid (sp gr
1.42) and allow to stand until dissolved. Add 1ml of sulphuric acid (sp gr 1.84)
and evaporate to nearly dryness. Cool, add 5ml more of sulphuric acid and
evaporate to dense white fumes. Dilute to 500ml with water (solution A). Dilute
50ml of this diluted solution to 250ml.
36.5 CALIBRATION
36.5.1 Prepare a series of standards (at least five concentrations) to cover the expected
range of Nickel concentrations by diluting kjappropriate volumes of Nickel
standard solution (4.6, 1 ml + 20 microgram of Nickel)
36.5.1.1 Place the required volumes of nickel standard solution (4.6) in 100ml
volumetric flasks.
36.5.1.3 Proceed in accordance with section 6.0 (6.2, 6.3 and 6.5) using 50 ml water as
the reference.
36.5.1.4 Prepare a calibration curve by plotting the absorbance versus milligrams per
litre of nickel.
36.6 PROCEDURE
36.6.1 Measure 50 ml portions of the sample into each of two 100ml volumetric flasks.
36.6.2 Add 10 ml of ammonium citrate solution and 5ml of iodine solution to each
flask.
36.6.4 To the other flask add 20ml of ammonium hydroxide (1+1). Dilute to 100ml
with water and allow to stand for minutes. Use this solution as the reference
solution.
36.6.5 Balance the spectrophotometer at zero absorbance with the reference solution
and read the absorbance at 530nm.
36.7 CALCULATIONS
36.7.1 Read the nickel concentration, in mg/litre directly from the calibration curve.
36.8 PRECISION
ST = 0.06 X
SO = 0.04 X
Where:
ST = overall precision.
36.9 INTERFERENCES
36.10 NOTES
36.10.1 Reagent grade chemicals should be used for preparing the reagents.
36.10.2 Samples shall be preserved with nitric acid (sp gr 1.42) to a pH of 2 or less
immediately at the time of collection, normally about 2 ml/ litre. If only
dissolved nickel is to be determined, the sample shall be filtered through a 0.45
micron membrane filter before acidification.
37.1 PRINCIPLE
37.2 INTERFERENCE
The presence of high concentration of chromium may increase the nickel signal.
Most interference can be eliminated by the use of nitrous oxide acetylene flame
but nickel sensitivity will be lowered.
37.3 APPARATUS
37.4 REAGENTS
Pipette out 10.0 ml of stock nickel solution into a 1000 ml volumetric flask and
make upto mark with water ( 1.0 ml = 10µg of Ni )
37.5 PROCEDURE
Calibration curve
Determination of Nickel
For the determination of dissolved nickel content, filtration through 0.45 micron
membrane filter, at the time of sampling is required. For total recoverable
nickel, HNO3 –H2SO4 digestion is to be carried out.
Add 0.5 ml of nitric acid to a suitable volume of the sample taken in a 100 ml
volumetric flask. Make upto the mark. Rinse the nebulizer by aspirating water
containing 1.5 ml of con.HNO3/l. Aspirate the sample solution and measure the
absorbance at 232 nm. Determine microgram of nickel in the solution from the
absorbance reading, by referring to the calibration curve.
37.6 CALCULATION
V1 x V2
The relative standard deviation reported in the literature for the nickel at 4.0
mg/l concentration level is 9.8%.
38 MANGANESE
38.1 PRINCIPLE
Manganese is oxidized to higher valence state (permanganic state) giving
permanganic colour by boiling with potassium periodate in acid condition and
in the presence of silver nitrate catalyst. The colour is proportional to the
manganese concentration.
Periodate method is more sensitive than persulphate in lower concentrations and
colour produced is stable for longer time.
38.2 INTERFERENCE:
Chlorides, organic matter, iron, colour and turbidity interfere in the estimation
of Mn.
38.3 APPARATUS
Colorimeter to operate at 525 nm.
Nessler’s tube cap. 100 ml.
Water bath.
38.4 REAGENTS
H2SO4 conc.
HNO3 conc.
H3PO4 85%.
KIO4 A.R. grade
AgNO3 salt
H2O2 30%
38.5 STANDARD MN SOLUTION: Prepare 0.1 N KmnO4 solution by dissolving
3.2 gm KmnO4 in distilled water and making up to litre. Age for several weeks,
filter and standardize against sodium oxalate.
38.6 PROCEDURE
(a) In case, pretreatment is required take suitable aliquot of sample and add 5 ml
conc. H2SO4 followed by 5 ml conc. HNO3 evaporate till SO3 fumes
disappear. This will remove interfering radicals such as organization qatter,
Chlorides and other oxidizable substances. This result in dehydration of
silica and production of turbidity, cool it and filter.
(b) Take filtrate add 85 ml distilled water, 5 ml HNO3 and 5 H3 PO4 85%. Mix
well. Addition of H3PO4 decolorizes Fe +++ion and prevent precipitation of
iodates or periodates of Mn.
(c) If pretreatment is not required add 5 ml conc. H2SO4 5 ml conc. HNO3 and
5 ml H3PO4 85%. Mix well.
(f) Heat till boiling for 1 hr. Cool immediately and dilute to 100 ml.
Calculate the conc. Of Manganese in the sample from calibration curve and
express in mg/1 as Mn.
39.2 RANGE
39.3 APPARATUS
39.3.1.1 Pipet volumetric, 100 ml capacity, with 200 mm delivery stem. The top of the
bulb shall be cut off and the delivery tip curved to form a shallow hook.
39.3.1.2 Tubing glass or vinyl 1mm inside diameter and about 350mm in length.
39.3.6 Pipettes, serological 1 and 10ml capacity, calibrated to 0.1ml to deliver with
ejection.
39.4 REAGENTS
39.4.2 Ammonium chloride solution – dissolve 85g of ammonium chloride (NH4 Cl)
in water and dilute to 1litre.
39.4.7 Hydrochloric Acid (1+1) - Mix equal volumes of hydrochloric acid (sp gr 1.19)
and water.
39.5.1.1 Clean about 10g of cadmium granules with hydrochloric acid (1+1) and then
rinse with water.
39.5.1.2 Swirl the clean cadmium granules in 100ml of copper sulphate solution for
5minutess or until the blue colour partially fades.
39.5.1.3 Decant and repeat with a fresh 100m1 copper sulphate solution solution until
the first trace of a brown colloidal precipitate appears.
39.5.1.4 Wash the granules with water at least ten times to remove all the precipitated
copper.
39.5.2.1 Insert a small plug of glass wool in the tip of the pipette.
39.5.2.3 Pour sufficient cadmium granules into the apparatus to produce a column
300mm in length.
39.6 CALIBRATION
39.7 PROCEDURE
39.7.1 Mix 20m1 of sample (pH between 6 to 8) with 80m1 of ammonium chloride
solution.
39.7.2 Pour about 60m1 of mixed sample (7.1) into the reservoir of the reduction
column.
39.7.4 Place a clean 50m1 graduated cylinder under the siphon outlet and collect 25m1
of effluent.
39.7.5 Place l0 ml of the collected effluent (7.3) into a 15m1 test tube.
39.7.6 Add 3.0 m1 of colour reagent, mix, and allow to stand for 15 minutes.
39.7.7 Measure the absorbance of solution against water as reference at 543nm using l0
mm matched absorption cells.
39.8 CALCULATIONS
Where:
Where:
39.9 INTERFERENCES
39.9.1 Turbidity of sample results in a buildup on the reduction column that restricts
sample flow. For turbidity removal the sample is filtered through 0.45 micron
membrane filter.
39.9.2 Sample colour that absorbs at wavelengths between 520 and 540nm interferes
with the absorbance measurement. When colour is suspected, analyze a sample
39.9.3 Oil and grease in the sample coat the surface of the cadmium and prevent
complete reduction of nitrate to nitrite.
For oil and grease removal, adjust the pH of the sample to 2 with hydrochloric
acid. Extract with two 25 m1 portion of trichlorotrifluoroethane and discard the
organic layer.
39.9.4 In acidic samples (pH less than 4.5) nitrate is not reduced in the cadmium
column. The pH is adjusted between 6 to 8 with hydrochloric acid or
ammonium hydroxide.
39.9.5 Certain metal ions such as mercury, copper, iron and manganese, in
concentrations above 35mg/litre, may cause interference.
39.10 NOTES
39.10.1 Reagent grade chemicals should be used in preparing all the reagents.
39.10.2 When not in use, the cadmium column should be covered with ammonium
chloride solution. If air enters the packing or it is dried out, prepare another
column.
39.10.3 If nitrite is present in the sample, it can be determined separately by carrying out
the procedure and omitting the cadmium reduction step.
39.10.4 Prepare at least one calibration standard from potassium nitrite at the same
concentration as one of the nitrate standards to verify the efficiency of the
reduction column. The nitrite standard will fall on the nitrate standard curve if
the reduction of nitrate was complete. If the nitrite standard is 5% higher,
prepare a fresh reduction column and repeat the analysis of the nitrate standards.
40.1 PRINCIPLE
Nitrate reacts with phenol disulphonic acid and produces a nitro-derivative
which in alkaline solution develops yellow colour due to rearrangement of its
structure. The colour produced follows Beer’s law and is proportional to the
concentration of NO3 present in the sample.
40.2 INTERFERENCE
Chlorides and nitrate, are the two main sources of interference. Pretreatment of
sample is necessary when the interfering radicals are present.
40.3 APPARATUS
a. Colorimeter or spectrophotometer having a range of 300-700 nm.
b. Nessler’s tubes capacity 100 ml.
c. Beakers capacity 100 ml.
d. Water bath capacity 100 ml.
40.4 REAGENTS
a) Standard silver sulfate: Dissolve 4.40 g Ag2SO4 in distilled water and
dilute to 1000 ml, 1 ml = 1mg Cl.
1 ml = 100m g N.
40.5 PROCEDURE
a) Pretreatment of sample
1. Colour removal: If the sample has a colour in excess of 10 units; add 3
ml aluminium hydroxide to 150 ml sample. Stir well and allow to settle for
a few min. Filter and use the filtrate discarding the first portion of the
filtrate.
2. Nitrite removal:
(1) Generally NO2 occurs along with NH3 and gets eliminated in the
routine test due to decomposition of NO2 and NH3 to N2,
(ii) Oxidize NO2 to NO3 under acid condition using KMnO4, (iii) Add
sulfamic acid to the sample to suppress NO2 interference.
3. Chloride removal: determine the Cl contents of the sample and
precipitate out as AgCl. One should be very careful while adding Ag2SO4
because excess Ag will precipitate out as silver oxide when alkali is added
to sample.
b) Colour Development
41.2 INTERFERENCE
Organic matter can cause a positive but variable interference, the degree
depending on the nature and concentration of the organic material.
Note: Clean all glassware thoroughly and rinse to reduce the error that might
result from streaks or particles on the outside of the cuvettes, as well as
traces of surfactants or dichromate cleaning solution that might adhere
on the interior glass surfaces.
41.3 APPARATUS
41.4 REAGENTS
41.5 PROCEDURE
41.6 CALCULATION
For correction for dissolved organic matter, subtract 2 times the reading at 275
at 275 nm from the reading at 220 nm to obtain the absorbance due to nitrate.
Convert this absorbance value into equivalent nitrate by reading the nitrate.
Value from a standard calibration curve obtained at 220 nm.
Calculate as follows:
42.3.1 Principle
Dissolved or emulsified oil and grease is extracted from water by hexane and
estimation is made gravimetrically.
42.3.2 Interferences
The solvent extracts are not only oil and grease but other organic substances
also.
42.3.3 Apparatus
42.3.4 Reagents
42.3.4.2 Hexane
42.3.5 Procedure
42.3.5.1 Transfer the acidified sample to a separating funnel. Carefully rinse the sample
bottle with 30ml of hexane and add the solvent washings to the separating
funnel.
42.3.6 CALCULATION
Oil and grease, mg/l = M/V x1000
Where
42.4.1 PRINCIPLE
42.4.2 APPARATUS
Soxhlet Apparatus
Vacuum Pump
42.4.3 REAGENTS
42.4.4 PROCEDURE
Prepare a filter consisting of a muslin cloth disc overlaid with filter paper. Wet
the cloth and paper. Pass 100 ml of filter aid suspension through the prepared
filter using vacuum and wash with 1 litre of distilled water. Filter the acidified
sample (pH<52). Apply vacuum until no more liquid sample passes through
filter paper. Using forceps transfer the filter paper to a watch glass. Add
material adhering to edges of muslin cloth disc. Wipe sides and bottom of
collecting vessel and Buchner funnel with filter paper soaked in solvent, taking
care to remove all films caused by grease and collect all the solid material. Add
pieces of filter paper to filter paper on watch glass. Roll all filter papers
containing sample and put into a paper extraction thimble. Add any pieces of
material remaining on watch glass. Wipe the watch glass with a filter paper
soaked in solvent and place it in the thimble. Dry the filled thimble at 103°C for
30 minutes in an oven. Fill the thimble with glass wool or small glass beads.
Weigh the extraction flask and extract oil and grease in a Soxhlet apparatus,
using hexane at a rate of 20 cycles per hour for four hours counting from first
cycle. Distil solvent from extraction flask over a water bath maintained at 70°C.
Place the flask on a water bath at 70°C for lb minutes and draw air through it by
vacuum for the final 1 minute. Cool in a desiccator for 30 minutes and weigh.
42.4.5 CALCULATION
Oil and grease, mg/l = Mx1000
V
Where
42.5.1 PRINCIPLE
To estimate the petroleum hydrocarbon content of oil and grease this modified
method is used. Silica gel has the ability to adsorb all constituents of oil and
grease except hydrocarbons. The material not adsorbed by silica gel is
designated as hydrocarbons by this method.
42.5.2 INTERFERENCE
The method is not accurate for the reason that any compound other than
hydrocarbon and fatty matter recovered by the solvent interfere by coming
either in hydrocarbon part or in fatty matter
42.5.3 REAGENTS
All reagents required for partition gravimetric method shall be applicable for
this method also. In addition, silica gel of 75 to 150 micron size, dried at 110°C
for 24 hours and stored in tightly sealed container is required.
42.5.4 PROCEDURE
Follow the extraction procedure as given in 5.5. To the extracted solution, add 3
to 4 g of silica gel. Stopper the container and stir gently 5or 10 minutes. Filter
the solution through filter paper and wash silica gel and filter paper with 10 ml
of solvent. Collect the filtrate. Distil off the solvent from the filtrate and weigh
the residue as hydrocarbons.
42.5.5 CALCULATION
Hydrocarbons, mg / l = Mx1000
V
Where
M = mass, in mg, of residue; and
Consumption Method)
43.2 REAGENTS
43.2.4 Sulphuric Acid (1+3) - Mix 1 volume of sulphuric acid (sp gr 1.84) with 3
volumes of water.Add standard permanganate solution until a very faint pink
colour persists after 4 hours.
43.2.8 Indicator Solution - Make a paste of l g of soluble starch and mix into 1 litre of
boiling water. Add 20g of potassium hydroxide, mix, and allow to stand for 2
hours. Add 6m1 of glacial acetic acid, mix and, add sufficient hydrochloric acid
to adjust the pH to 4.0 (Check with a narrow range PH paper). This has a shelf
life of 'l year.
43.3 PROCEDURE
43.3.1 Place 100m1 of the sample into a clean, glass stoppered bottle of 250 m1
capacity and place in a thermostat at 27°C. .
43.3.2 When the temperature of the sample becomes 27°C, add 4ml of sulphuric acid
(1+3) and l0 ml of potassium permanganate solution (N/80). Mix well and allow
to stand for 4 hours at 27°C protected from sunlight.
43.3.3 Add few crystals of potassium iodide (0.2-0.3 g) and titrate the librated iodine
with standard sodium thiosulphate solution (N/80) using starch indicator.
43.4 CALCULATIONS
43.4.1 Calculate the milligrams of oxygen consumed per litre of sample as follows:
Where:
Dissolved oxygen reacts, under alkaline conditions, with the indigo carmine
solution to produce a progressive colour change from yellow-green through red
to blue and blue-green. The colour developed in standards dissolved the sample
is compared with colour representing different concentrations of oxygen.
44.2 RANGE
44.3 APPARATUS
44.3.2 Sampling Bucket, with an overflow at least 20mm above the top of the sampling
vessel.
44.3.3 Sampling Vessels - Nessler type 60 m1 tubes or 300 m1 BOD bottles having a
raised lip around the neck and glass stoppers ground to a conical lower tip.
44.4 REAGENTS
44.4.2.5 Hydrochloric Acid (1+99)-Mix 1 volume of hydrochloric acid (sp gr 1.19) with
99 volumes of water.
44.4.2.6 Indigo Carmine Solution- Dissolve 0.18 g of indigo carmine and 2.0 g of
dextrose (or glucose) in 50 m1 of water. Add 750 m1 of glycerin and mix
thoroughly.
44.5 CALIBRATION
0 0.75 05.0 -
5 5.0 20.0 -
10 6.25 12.5 -
15 9.4 10.0 -
20 13.0 5.4 -
25 14.4 5.5 -
44.6 PROCEDURE
44.6.1 Place a clean sampling vessel in the sampling bucket and collect the sample
under water. Allow the sample to overflow for several minutes.
44.6.2 Fix a burette such that its tip dips into the overflowing sample to a depth of 10
to 15mm.
44.6.3 Fill the burette with indigo carmine-potassium hydroxide reagent. Drain
about l ml of reagent into the overflowing sample, and allow the sample to flush
for 1 minute.
44.6.5 Quickly introduce 0.8m1 of the reagent if 60m1 tube is used or 4m1 of reagent
if a BOD bottle is used, stopper the vessel and mix by inversion.
44.6.6 Place the vessel on a white surface and match its colour with the standard by
viewing at a 45° angle using a 'Cool" white fluorescent lamp for illumination,
44.7 PRECISION
44.7.1 The single operator precision of this method may be expressed as follows:
So = 0.052 X + 0.7
Where:
44.8 INTFRFERENCES
44.8.3 Ferrous iron will produce low results and copper will cause high results.
44.8.4 In samples, where ferrous iron and copper are present, their combined effect is
frequently zero.
44.9 NOTES
44.9.1 Reagent grade chemicals should be used for preparing the reagents.
44.9.2 All colour stock solutions should be stored in coloured bottles to prevent fading.
44.9.3 Indigo carmine solution (4.2.6) is stable for 30 days if stored in a refrigerator.
44.9.4 In the procedure (6.1), the sample flow should be between 500 to
1000m1/minute when using 300m1 bottle, or 100 to 200m1/minute when using
60m1 sample tubes.
44.9.5 In the procedure (6.6), the colours should be matched as soon as possible after
mixing the reagents and sample, since the colours are not stable for more than
30 minutes and air leakage may cause a change in colour.
44.9.6 The sample should be analysed as soon as possible after the collection.
45.2 RANGE
45.3 APPARATUS
45.3.1 Sample bottles, 300m1 capacity, with tapered ground glass stoppers.
45.4 REAGENTS
45.4.6 Starch Solution - Make a paste of l g of soluble starch and mix into 1-litre of
boiling water. Add 20 g potassium hydroxide, mix, and allow to stand for 2
hours. Add 6m1 of glacial acetic acid, mix, and add sufficient hydrochloric acid
(sp gr. 1.19) to adjust the pH to 4.0 (check with a narrow range pH paper). This
has a shelf life of 1-year.
45.5 PROCEDURE
45.5.2 Add 2.0 m1 of manganous sulphate solution and 2.0 m1 of alkaline iodide-
sodium azide solution well below the surface of the sample. Replace the stopper
and mix by inversion.
45.5.3 When the floc has settled, add 2.0 m1 of sulphuric acid (sp gr 1.84). Restopper,
and mix by inversion.
45.5.4 Titrate 203 m1 of solution (equivalent to 200m1 sample) with standard solution
of sodium thiosulphate using starch indicator.
45.6 CALCULATIONS
V1 X 0.2
Dissolved oxygen, mg/litre = -------------------------- x 100
200
Where:
45.7 PRECISION
45.7.1 The precision of this method was determined by six operators in three
laboratories using a saturated sample of reagent water. The mean concentration
was 9.0 m1/litre and the pooled single operator precision in these samples was
0.052 mg/litre.
45.8 INTERFERENCES
45.8.2 Ferric iron interferes unless l ml of potassium fluoride solution is used, in which
case 100 to 200mg/litre can be tolerated.
45.9 NOTES
45.9.1 Reagent grade chemicals should be used for preparing the reagents.
45.9.2 In the procedure (5.4), 203m1 of the solution is titrated to take care of the
displaced sample by 2m1 each of manganous sulphate solution and alkaline
iodide-solution.
300
45.9.3.1 Add 0.7m1 of sulphuric acid (sp gr 1.84) and 1.0m1 of potassium permanganate
to the sample. When high iron is present, also add 1.0ml of potassium fluoride
solution. Stopper and mix by inversion. Add sufficient potassium permanganate
to maintain a violet tinge for 5 - minutes.
45.9.3.2 After 5 minutes, destroy the violet colour by adding 0.5 to 1.0m1 of potassium
oxalate solution. Mix and allow to stand in dark. Complete decolorization
should be obtained in 2 to 10 minutes.
45.9.4 Do not delay the determination of dissolved oxygen. Samples may be preserved
4 to 8 hours by adding 0.7m1 of sulphuric acid (sp gr 1.84) and l ml of sodium
azide solution (20g/litre) to the sample.
W
N = ------------------------------------------
0.04904 X V
Where:
Loss Method)
46.1 SUMMARY OF METHOD
46.2 APPARATUS
46.2.1 Incubation Bottles - 250 to 300m1 capacity with ground glass stoppers.
46.3 REAGENTS
46.3.1 Dilution Water - Add 0.3g of sodium bicarbonate per litre of Type II water.
46.4 PROCEDURE
46.4.1 Adjust the temperature of a suitable portion of the well mixed sample to 20°C.
Remove the oxygen or excess air by maintaining the sample under vacuum for
10 minutes using laboratory vacuum pump.
46.4.2 Fill completely two incubation bottles (250 or 300m1 capacity) with the sample
as treated above (4.1). Allow to stand for 15 minutes.
46.4.3 Determine the dissolved oxygen in one bottle by the Iodometric method and in
the other after 5 days" incubation in darkness in the stoppered bottle at 20oC.
46.5 CALCULATIONS
46.5.1 Calculate the BOD of the sample as follows: Biochemical oxygen demand
(BOD), mg/litre = Dl - D2 (5 days at 20°C).
Where:
46.6 INTERFERENCES
46.6.1 Samples for BOD analysis may degrade significantly during storage, resulting
in low BOD values. This can be minimized by analyzing the sample promptly
or cooling it to 4°C or below.
46.7 NOTES
46.7.1 The dissolved oxygen content of the sample before incubation shall be
approximately 9mg/litre or preferably less.
46.7.2 For samples of doubtful purity, the sample should be mixed with dilution water
in the ratio 1:1 at 20°C. Further dilutions shall be used if necessary to ensure
that not more than half the oxygen is consumed during the incubation.
Determine the dissolved oxygen before and after incubation and calculate the
result using the appropriate dilution factor.
Method)
47.1 SUMMARY OF METHOD
The sample is treated with a standard potassium dichromate solution in 50%
sulphuric acid solution. The excess dichromate is titrated with a standard ferrous
ammonium sulphate solution, using orthoprenan-throline-ferrous complex as an
internal indicator.
47.2 RANGE
Upto 800mg/litre.
47.3 APPARATUS
47.4 REAGENTS
47.4.9 Sulphuric Acid - Silver Sulphate Solution - Dissolve 15g of powdered silver
sulphate (Ag2SO4) in 300m1 of Sulphuric acid (sp gr 1.84) and dilute to 1 litre
with sulphuric acid (sp gr 1.84).
47.5 PROCEDURE
47.5.2 Place 50m1 of water in a separate reflux flask for blank determination,
47.5.3 Place the reflux flasks in an ice bath and add l g of powdered mercuric sulphate,
5.0 m1 of sulphuric acid (sp gr 1.84) and several glass beads or boiling stones to
each flask and mix well to complete dissolution.
47.5.4 Add 70m1 of sulphuric acid - silver sulphate solution slowly such that the
solution temperature is maintained as low as possible (below 400C).
47.5.6 Allow the flasks to cool and wash down the condensers with about 25m1 of
water. Dilute the solution to about 300 m1 with water and allow to cool to room
temperature.
47.5.8 For waters of low COD (10 to 50 mg/litre) use 0.025N potassium dichromate
and 0.025N ferrous ammonium sulphate solutions.
47.6 CALCULATIONS
Where:
47.7 INTERFERENCES
47.7.1 The method does not uniformly oxidise all organic materials. Volatile
organics which are difficult to oxidise may be partially lost before oxidation is
achieved. Care .in maintaining a low solution temperature (about 40°C) and
permitting oxidation to proceed at the lower temperature for a period of time
before reflux is initiated will result in higher recoveries of theoretical COD.
47.7.2 Chloride ion is quantitatively oxidised by dichromate in acid solution, with each
mg/litre chloride exerting the equivalent of 0.226mg/litre COD. Reaction of
chloride ion (up to 1000mg/litre) is inhibited by addition of mercuric sulphate.
47.7.3 Oxidizable inorganic ions, such as ferrous, nitrite, sulphites, and sulphides are
oxidised and measured also as organic constituents.
47.8 NOTES
47.8.2 For standardizing ferrous ammonium sulphate (0.25N), place 25m1 of 0.25N
potassium dichromate in a 500m1 titration flask and dilute to about 250m1. Add
20m1 of sulphuric acid (sp gr 1.84) and allow the solution to cool. Titrate with
ferrous ammonium sulphate, using phenanthroline-ferrous sulphate indicator.
47.8.3 Calculate the normality (N) of ferrous ammonium sulphate solution as follows:
V1 X N1
Normality, N = ------------------------------
Where :
N = normality of ferrous sulphate solution.
V1 = milliliters of potassium solution.
If less than 50m1 of the sample is used (procedure, 5.1) add necessary amount
of dilution water to the reflux flask, then add the sample aliquot and mix.
Samples containing more than 800mg/litre COD are diluted and mixed precisely
with water and 50m1 of the diluted sample are placed in the reflux flask.
47.8.4 To check the validity of the test (procedure, 5.0) results, make a standard
determination using potassium acid phthalate solution (4.6). A COD of 500
mg/litre should be obtained on a 25m1 aliquot of the standard solution diluted to
50m1.
47.8.5 In the procedure (5.8) if the COD determined is higher than 50mg/litre after
using 0.025N potassium dichromate and 0.025N ferrous ammonium sulphate
solutions, reanalyze the sample using more concentrated reagents.
47.8.6 Preserve samples by cooling to 4°C if analyzed within 24 hours after sampling
or preserve for up to 7 days at pH < 2 at 4°C.
47.8.7 Silver sulphate acts as a catalyst in the reaction to facilitate the oxidation of
organics which are difficult to oxidise.
Since 10 mol of K2Cr2O7 have the same oxidative power as 15 mol of O2, the
equivalent reaction is:
48.2 RANGE
48.3 APPARATUS
48.4 REAGENTS
48.4.2 ANSA Solution - Dissolve the following chemicals (in the given order) in
100m1 of water: 3.7g of sodium sulphite (Na2SO3), 0.100g of 1-amino-2-
naphthol-4 sulphonic acid, and 6.2g of sodium metabisulphite (Na2S2O5). Store
in an amber coloured bottle and make fresh every two weeks.
48.4.5 Phosphate Standard Solution (l0mg/litre PO4) - Dilute 10.0m1 of the stock
solution (4.4) to 100m1 with water.
48.4.6 Sulphuric Acid Solution (A) - Add 370m1 of sulphuric acid (sp gr 1.84), slowly
and with stirring, to 600m1 of water. Cool and dilute to 1 litre with water.
48.5 CALIBRATION
48.5.1.1 Prepare a series of phosphate standards by diluting 0.0, 0.5, 1.0, 2.0, 5.0 and
l0.0m1 of phosphate stock solution (4.4, l m1 = 0.lmg PO4) to 100m1 with
water in a volumetric flask.
485.1.3 Measure the absorbance of the each standard against reagent blank (zero
standard) at 650nm using 25mm matched absorption cells.
48.5.1.4 Prepare a calibration curve by plotting absorbance versus milligrams per litre of
phosphate.
48.6 PROCEDURE
48.6.1.4 Add 5.0m1 of ANSA solution, mix, and allow to stand for 10 minutes.
48.6.1.5 Measure the absorbance against reagent blank (prepared by adding all the
reagents to 100 m1 water) at 650nm using 25mm matched absorption cells.
48.6.2.1 Proceed in accordance with 6.1, substituting sulphuric acid solution (B) in place
of sulphuric acid solution (A) in step 6.1.2.
48.7 CALCULATIONS
48.7.1 Read the concentration of phosphate in mg/litre directly from the calibration
curve prepared in accordance with section 5.0.
Where:
48.8 INTERFERENCES
48.8.1 A silica concentration fifty times that of the phosphate will cause an error of
less than 2 percent.
48.8.2 Up to 40mg/litre iron, 75 mg/litre chromate, and 50,000 mg/litre chloride will
not interfere with the test. Not more than 1500 mg/litre chloride should be
present when the bismuth salt modification is used.
48.9 NOTES
48.9.1 Reagent grade chemicals should be used in preparing all the reagents.
48.9.2 Only orthophosphate forms a blue colour in the test. Other forms of phosphorus
such as polyphosphate and organic phosphorus can be converted to
orthophosphate, if present in the sample, before applying above mentioned
methods.
48.9.3 All glassware and sampling bottles should be soaked in hydrochloric acid (1+3)
and rinsed with water before use.
49.2 RANGE
49.3 APPARATUS
49.4 REAGENTS
49.4.1 Water - conforming to specifications Type II and filtered through 0.2/0.1 micron
membrane fitter.
49.4.2 Hydrofluoric Acid (1+4) - Dilute 50m1 of 40 % hydrofluoric acid with 200m1
of water. If the hydrofluoric acid is not exactly 40 % adjust the dilution
accordingly.
49.4.3 Hydrochloric Acid (1+2) - Dilute 50m1 of hydrochloric acid (sp gr 1.19) with
100 m1 of water.
49.4.6 Oxalic Acid Solution (10%) - Dissolve l0g of oxalic acid (H2C2O4.H2O) in
water and dilute to 100m1.
49.4.7 Amino - Naphthol - Sulphonic Acid Solution - Dissolve 0.5g of 1-amino -2-
naphthol -4- sulphonic acid in 50 m1 of a solution containing 1g of sodium
sulphite (Na2SO3). Add the solution to 100m1 of a solution containing 30g of
sodium hydrogen sulphite (NaHSO3). Make up to 200m1 and store in a dark,
plastic bottle. Prepare fresh solution every 2 weeks.
49.4.8 Silica Stock Solution (lml = 1mg SiO2) - Dissolve 4.732g of sodium
metasilicate (Na2SiO3.9H20) in water and dilute to 1 litre. Standardize the
solution by acid dehydration procedure (see method for the determination of
soluble silica, Note 10.7).
49.5 CALIBRATION
49.5.1.1 Place 0, 0.5, 1.0, 2.0, 3.0, 4.0 and 5.0m1 of standard silica solution A (l ml =10
micrograms SiO2) in 100m1 volumetric flasks and make up to 100m1 with
water.
49.5.2.1 Place 0.0, 0.5, 1.0, 2.0, 3.0, 4.0 and 5.0m1 of standard silica solution B (l ml = 1
microgram Si02) in 100m1 volumetric flasks and make up to 100 m1 with
water.
49.6 PROCEDURE
49.6.1.2 Add 2m1 of hydrochloric acid solution, 4.5m1 of water and 2m1 of ammonium
molybdate solution. Mix thoroughly and allow to stand for 10 minutes.
49.6.1.3 Add 3 m1 of oxalic acid solution and mix. Allow to stand for 10 minutes.
49.6.1.4 Add 2 m1 of amino naphthol sulphonic acid solution and allow to stand for 10
minutes.
49.6.1.5 Measure the absorbance of the sample against the zero standard (blank) (E1) at
815nm.
49.6.1.6 For blank, place 100m1 of water in a 200m1 polythene conical flask and
proceed in accordance with 6.1.2 to 6:2.4
49.6.2.2 Add l ml of hydrochloric acid and 2m1 of hydrofluoric acid and mix
thoroughly. Allow to stand for 10 minutes.
49.6.2.3 Add 3.5m1 of zirconium solution and, drop by drop with stirring, 2m1 of
ammonium - molybdate solution. Mix thoroughly. Allow to stand for 10
minutes.
49.6.2.4 Add 3m1 of oxalic acid solution and mix thoroughly. Allow to stand for 10
minutes.
49.6.2.5 Add 2m1 of amino-naphthol-sulphonic acid solution and allow to stand for 10
minutes.
49.6.2.6 Measure the absorbance of sample against the blank solution (E2) at 815nm.
49.6.2.7 For blank, place 100m1 of water in a 200m1 polythene conical flask and
proceed in accordance with 6.2.2 to 6.2.5.
49.7 CALCULATIONS
49.7.1 From the value E2 using the calibration curve, calculate the, concentration of
reactive plus non reactive silica in the sample.
49.7.2 From the value (E2-El), using the calibration curve, calculate the concentration
of non reactive silica in the sample.
Where:
49.8 INTERFERENCES
49.8.1 See method for the determination of soluble silica (section 9.0). ,
49.9 NOTES
49.9.1 See method for the determination of soluble silica (Section 10.0).
49.9.2 In the analytical procedure the HF/Zr ratio must be kept constant and should be
equal to 6/1. Values greater than this ratio correspond to an excess of
hydrofluric acid which then again reacts with silicic acid to form fluorosilicic
acid and thus introducing error. Values lower than this ratio corresponds to an
excess of Zr+4 ions that give rise to a white precipitate in the presence of silico-
molybdic complex. Hence, it is very important to know precisely, the zirconium
content in the ZrOCl2.8H20 reagent and the HF titre in 40% hydrofluoric acid to
adjust the amount of zirconium and HF solutions.
W1 x 7403
Zr content (% w/w in zirconium oxychloride) = ----------------
W x 28.317
Where:
The sample is filtered through a 0.2 micron membrane filter and the residue left
on the membrane filter is treated with 48% hydrofluoric acid. The soluble silica
thus produced is reacted with ammonium molybdate and the concentration of
resulting complex is determined calorimetrically.
50.2 RANGE
50.3 APPARATUS
50.3.1 Membrane Filter, 0.2 micron pore size and 47mm diameter.
50.3.6 Water Bath to accommodate at least 6 beakers and with a temperature control
between 35 - 50°C.
50.4 REAGENTS
50.4.2 Boric Acid Solution (5%) - Dissolve 100g of boric acid (H3BO3) in 2 litres of
warm water.
50.4.4 Sulphuric Acid (l0 N) - Cautiously add, with constant stirring, 555m1 of
sulphuric acid (sp gr 1.84) to 1.4 litres of water. Cool to room temperature and
dilute to 2 litres with water. .
50.5 CALIBRATION
50.5.1.1 Place 0.0, 0.5, 1.0, 2:0, 3.0, 4.0 and 5.0 m1 of standard silica solution A (l ml =
10 micrograms Si02) in 200m1 polythene flasks. Add water to make 75m1.
50.5.1.2 Add 4.0m1 of ammonium molybdate reagent and mix. Allow to stand for 20
minutes at 35-40°C in a water bath.
50.5.1.3 Remove the flasks from the water bath and add 20m1 of sulphuric acid (10 N)
rapidly while stirring.
50.5.1.4 Measure the absorbance of each standard at 420nm against zero standard
(blank) using l0mm absorption cells,
50.5.2.1 Place 0.0, 1.0, 2.0, 4.0, 6.0, 8.0 and l0.0 m1 of standard silica solution B (l ml =
1 microgram Si02) in 200m1 polythene flasks. Add water to make 75m1.
50.5.2.2 Add 4.Om1 of ammonium molybdate reagent and mix. Allow to stand for 20
minutes at 35-40°C in a water bath.
50.5.2.3 Remove the flasks from the water bath and add 20m1 of sulphuric acid (10 N)
rapidly while stirring. Allow to stand for 5 minutes.
50.5.2.5 Measure the absorbance of each standard at 815nm against zero standard
(blank) using 50mm absorption cells.
50.6 PROCEDURE
50.6.1 Filter 500 m1 of the water sample through a 0.2 micron membrane filter by
applying vacuum.
50.6.2 Place the membrane filter in the bottom of a 150m1 polythene beaker and add
0.5 ml of 48% hydrofluoric acid with the help of a long and thin, calibrated
polythene dropper. Place a polythene disc on the membrane filter and cover the
beaker with a polythene cover. Allow to stand for 30 minutes.
50.6.3 Add 25m1 of water and 50m1 of boric acid solution. Mix well and place the
beaker in a water bath controlled at 40-50°C. Allow to stand for 15 minutes.
50.6.4 Add 40m1 of ammonium molybdate reagent while stirring. Allow to stand for
20 minutes at 35-40°C in a water bath.
50.6.5 Remove the beaker from the water bath and add 20m1 of sulphuric acid
50.6.8 Run a blank alongwith the sample, using a 0.2 micron membrane filter and
proceeding in accordance with 6.2 to 6.6.
50.7 CALCULATIONS
50.8 NOTES
50.8.1 See method for the determination of soluble silica (section 10.0).
50.8.2 Reagent grade chemicals should be used for preparing all the reagents.
50.8.3 Hydrofluoric acid used should contain least possible amount of fluorosilicic
acid.
50.8.4 0.1 micron membrane filters should be preferred.
50.8.5 Presence of boric acid up to a concentration of 16,200 mg/litre does not interfere
significantly.
The sample containing reactive silica is reacted with molybdate ion in an acidic
medium (pH 1.2 to 1.5) and the resulting green-yellow coloured complex is
reduced to a blue complex with 1-amino-2- naphthol-4-sulphonic acid the
intensity of which is measured at a wavelength of 815nm.
51.2 RANGE
51.3 APPARATUS
51.4 REAGENTS
51.4.2 Amino - Naphthol - Sulphonic Acid Solution - Dissolve 0.5g of 1-amino -2-
naphthol-4-sulphonic acid in 50m1 of a solution containing l g of sodium
sulphite (Na2SO3). Add the solution to 100 ml of a solution containing 30g of
sodium hydrogen sulphite (NaHSO3). Make up, to 200m1 and store in a dark
plastic bottle. Prepare fresh solution every 2 weeks.
51.4.4 Hydrochloric Acid (1+1) - Mix equal volumes hydrochloric acid (sp gr
1.19) and water.
51.4.5 Oxalic Acid Solution (10%) - Dissolve l0g of oxalic acid (H2C2O4.2H2O) in
water and dilute to 100m1.
51.5 CALIBRATION
51.5.1 Prepare a series of standards (at least five concentrations) to cover the range 5 to
140 micrograms Si02/litre by diluting appropriate volumes of silica standard
solution (4.7, l ml = 1 microgram SiO2) to 50 m1 in 100 m1 polythene bottles.
51.6 PROCEDURE
51.6.2 Add 1.0 m1 of hydrochloric acid (1+1) and mix, then add 2.0ml of ammonium
molybdate solution and mix well.
51.6.3 After 5 minutes, add 1.5m1 of oxalic acid solution and mix well.
51.6.5 Prepare a reagent blank by treating 50.0m1 of water as directed in 6.2 through
6.4.
Measure the absorbance at 815nm against the reagent blank using 50mm
matched absorption cells.
51.7 CALCULATIONS
Read the concentration of silica in micrograms/ litre as SiO2 directly from the
calibration curve prepared in section 5.0.
Where:
51.8 PRECISION
So = 0.005 X + 0.7
St = 0.03 X + 1.3
Where:
SiO2/litre.
51.9 INTEREFERENCES
51.9.1 Samples containing less than 1.Omg/litre of Si02 do not normally contain a
significant concentration of interfering substances.
51.9.2 Colour and turbidity will interfere if not removed by filtration or dilution.
51.9.5 Strong oxidising and reducing agents may interfere in the reduction step.
51.10 NOTES
51.10.1 Reagent grade chemicals should be used for preparing all the reagents.
51.10.2 Samples should be collected in plastic or stainless steel sample bottles provided
with rubber or plastic stoppers.
51.10.3 All reagents to be used in this method should be stored in polythene or other
suitable plastic bottles.
51.10.7 The strength of silica stock solution (4.6) may be checked as follows:
Place 100m1 of Silica stock solution (4.6) in a 400m1 scratch-free glass beaker.
Neutralize with hydro-chloric acid (sp gr 1.19) to methyl orange end point and
add 5m1 in excess. Evaporate the solution to dry-ness on a water bath, with
periodic additions of three more 5m1 increments of hydrochloric acid (sp gr
1.19). Dry the evaporated residue in an oven at 110°C for 1 hour. Add 5ml of
hydrochloric acid (sp gr: 1.19) and then 50m1 of water to the beaker, warm the
beaker and its contents. Stir the mixture. Filter the warm solution through an
ashless, medium texture filter paper. Wash the residue on the paper 15 times
with hydrochloric acid (1+19) and then with several small increments of water.
Reserve the paper and its residue for later ignition,
Return the filtrate and washings obtained above to the original evaporating
beaker, and evaporate to dryness on a water bath with periodic addition of two
5m1 increments of hydrochloric acid (sp gr 1.19). Dry the evaporated residue in
an oven at 110°C for 1 hour. Add 5m1 of hydrochloric acid (sp gr 1.19) and
then 50m1 of water to the beaker, warm the beaker and its contents and stir the
mixture. Filter the warm solution through an ashless, medium texture filter
paper. Wash the residue. on the paper 15 times with hydrochloric acid (1+19)
and then with several small increments of water.
Place both filter papers with their residues in a weighed platinum crucible, dry
and char the paper without flaming it, and then ignite the charred residue for 30
minutes at 1000 to 1200°C. Cool in a desiccator and weigh. Repeat the ignition,
cooling, and weighing until a constant weight is obtained.
Add several drops of sulphuric acid (sp yr 1.84) and about 5m1 of hydrofluoric
acid (48%) to the weighed residue in, the crucible and evaporate to dryness on a
low temperature hot plate under a fume hood. Reignite the residue at 1000 to
1200°C and weigh. Repeat the ignition, cooling, and weighing until a constant
weight is obtained.
(Wl-W3) - (W2-W4)
Si02, mg/litre = -----------------------------------------
V
Where:
W2 = weight of crucible and residue, mg, after treatment with hydrofluoric acid
and reignite it.
52.2 RANGE
52.3 APPARATUS
Steady operating conditions of load, drum water level, steaming rate and boiler
water chemistry
The operating parameters of load, steam flow, drum pressure and drum level are
to be maintained for a minimum 4 hours.
Determine the sodium content of both steam and boiler water samples.
5 samples to be collected for analysis at the end of the steady state of boiler
operation.
52.4.2 Determine total solids in the boiler water sample by evaporation method.
52.5 CALCULATIONS
WT
St = Ss x ---------------------
WS
Where:
St = Concentration of total dissolved solid (TDS) in steam, mg/litre
Ss = Concentration of Sodium in steam, mg/ litre
WT = Concentration of total dissolved solids (TDS) in boiler water,
mg/litre. {Coductivity (µs/cm) / 1.8}
WS = Concentration of sodium in boiler water, mg/litre
52.6 NOTES
52.6.2 A continuously flowing sample must be employed for the sodium ion electrode
method. (By on-line analyzer)
53.2 RANGE
20 to 100mg/litre as SO4.
53.3 REAGENTS
53.3.3 Hydrochloric Acid (1+9) - Mix 1 volume of hydrochloric acid (sp gr 1.19) with
9 volumes of water.
53.3.5. Methyl Orange Indicator Solution (0.5g/litre) - Dissolve 0.05g of methyl orange
in water and dilute to 100m1.
53.4 PROCEDURE
53.4.1 Place clear sample, containing sulphate ion equivalent to 10 to 50mg of barium
sulphate into a 500m1 beaker.
53.4.3 Adjust the acidity of the sample to methyl orange end point and add l0 ml
excess of hydrochloric acid (1+9).
53.4.4 Boil the solution and add slowly 5m1 of hot barium chloride solution while
stirring vigorously. Keep for 2 hours on a .steam bath.
53.4.5 Filter through Whatman No.42 or equivalent filter paper and wash the
precipitate with hot water until the washings are substantially free of chloride
(check with silver nitrate solution).
53.4.6 Place the filter paper and its contents in a weighed platinum crucible and ignite
at 800°C for 1 hour.
53.4.7 Add a drop of sulphuric acid and a few drops of hydrofluoric acid, and
evaporate under a hood to expel silica.
53.5 CALCULATIONS
W X 411,500
Sulphate, mg/litre as SO4 = --------------------------
V
Where:
Where:
53.6 PRECISION
53.6.1 Results of this method are precise to 1.0% of the amount of sulphate ion
present.
53.7 INTERFERENCES
53.7.1 Sulphites and sulphides may oxidise and precipitate with sulphate.
53.7.2 Turbidity caused by silica or other insoluble material .would interfere if allowed
to be present.
53.8 NOTES
53.8.1 Reagent grade chemicals should be used for preparing all the reagents.
53.8.2 Turbidity may be removed by filtration 'through a fine filter paper (eg.,
Whatman No.42).
53.8.3 Silica may be removed, prior to analysis, by acid dehydration procedure (see
method for the determination of soluble silica, Note 10.7). In this case, the
ignition described in 4.6 need not be done in a platinum crucible.
53.8.4 Faster precipitation and a coarser precipitate can be obtained by adding l0ml of
saturated picric acid solution and boiling the sample for.5 minutes before adding
barium chloride.
53.8.5 Do not attempt to obtain a complete removal for chloride. Discontinue washing,
when not more than a faint opalescence is produced in the test for chloride.
54.2 RANGE
5 to 1000mg/litre as S04.
54.3 APPARATUS
54.3.2 Ion Exchange Column - Glass column, 500 ± 5mm long and 9 to l0 mm inside
diameter. The top of column widens to a reservoir of 50 to 55mm inside
diameter and 100 ± 5mm in length. 30cm of wet strong acid cation resin is
poured into the column which is held by a filter plug. The flow is controlled by
a length- of 2mm bore capillary tubing joined to the bottom of the column. This
tubing is bent into a U-shape and rises to about 13mm above the resin bed and
then makes a U-bend downward for about 50 to 70mm.
54.4 REAGENTS
3.8 to 4.0 with dilute hydrochloric acid. Standardize in accordance with section
5.0.
54.4.5 Hydrochloric Acid (1+99) - Mix 1 volume of hydrochloric acid (sp gr 1.19)
with 99 volumes of water.
54.4.8 Ion Exchange Resin - Strong acid cation exchange resin, 710 to 850 micron. (IR
120 or equivalent). Regeneration is to be done with hydrochloric acid (I+4).
54.4.11 Starch Indicator - Make a paste of l g of starch with cold water. Pour the paste
into 100m1 of boiling water and boil for several minutes. This is stable for 2 to
3 days.
54.5.1 Prepare a series of standard sulphate solution: by diluting with water 0.0, 2.0,
5.0, 10.0, 15.0,25.0, 35.0 and 50.0m1 of the standard sulphate solution (l ml=
0.l00mg SO4) to 50m1 in volumetric flasks.
54.5.2 Determine the blank and sulphate equivalent of barium chloride solution (note
12:2) in accordance with 6.3, 6.4 and 6.5.
54.6.1 Place 25m1 of filtered sample into a 100m1 beaker. Add 0.5 m1 of starch
indicator and titrate the sulphite with iodine solution.
54.6.2 Add 2 to 3 drops of phenolphthalein indicator and adjust the pH to about 10.3
with hydrochloric acid (1+99) or ammonium hydroxide (1+99) using narrow
range pH paper. Add 0.3 to 0.5g magnesium carbonate and boil for 5 minutes.
Cool to 10°C and filter through Whatman No.41 or equivalent filter paper into a
50 m1 volumetric flask. Wash the precipitate with three 5 m1 portions of water
at 10°C. Add few drops of hydrogen peroxide, shake, and adjust the volume to
50 m1 with water.
54.6.3 Pass the solution through the ion exchange column and discard the first 25 to
30m1 of effluent. Place l0.0m1 of the next effluent into a small white porcelain
dish (100 to 125m1 capacity).
54.6.4 Add 40m1 of alcohol and 2 drops of thorin indicator. Adjust the pH to 3.8 to 4.0
by dropwise addition of ammonium hydroxide (1+99) until the- solution just
turns pink. Add hydrochloric acid (1+99) dropwise until the pink colour
disappears.
54.6.5 Prepare a blank using water and reagents described in 6.1 to 6.4 and record the
iodine solution used for the sulphite correction of the blank. Titrate the sample
with barium chloride solution, using the untitrated yellow blank as a colour
54.7.1 Pass 50m1 of the filtered sample directly through the ion exchange column.
Collect l0.0ml of effluent and proceed in accordance with 6.3, 6.4 and 6.5.
54.8.1 Directly titrate l0.0m1 of the filtered sample in accordance with 6.4 and 6.5.
54.9 CALCULATIONS
Where:
54.10 PRECISION
54.11 INTERFERENCES
54.11.1 Both cations and anions may cause co-precipitation errors with barium sulphate
precipitate. Most metallic ions also interfere by forming coloured complexes
with the thorin indicator, especially in alcohol water mixtures.
54.11.6 Chloride masks the pink endpoint if present in concentration more than
1000mg/litre when the sulphate present is low (about 5mg/litre).
54.12 NOTES
54.12.1 Reagent grade chemicals should be used for preparing all the reagents.
54.12.2 A solution of known sulphate concentration should be run with each series of
tests or new reagents to check the standardization curve. The blank used to
determine sulphate content is preferably that determined from the
standardization curve extrapolated to zero.
54.12.3 Phosphate ion is almost completely precipitated (step 6.2) at or below 10°C, but
solubility increases with increasing temperature.
54.12.4 If the ammonium hydroxide--(step 6.4) is added too fast, it is possible to over-
run the colour change from yellow to pink and the sample continues to be
yellow. It is then impossible to develop the pink colour by addition of
ammonium-hydroxide.
54.12.5 For very low sulphate concentrations a less concentrated barium chloride
solution (l ml = 0.200mg sulphate) should be used. A standard sulphate solution
may be added to the sample to raise the total sulphate concentration to 10 to
15mg/litre. This added sulphate must be subtracted from the final result.
The sample is mixed with hydrochloric acid and potassium iodide and titrated
against potassium iodate solution in presence of starch indicator. The iodine
liberated reacts with sulphite and the endpoint is indicated by the formation of
blue colour.
55.2 RANGE
55.3 REAGENTS
55.3.2 Hydrochloric Acid Solution (1+3) - Mix one volume of hydrochloric acid (sp gr
1.19) with three volumes of water.
55.3.5 Starch Indicator - Make a paste of 6g of starch with cold water. Pour the paste
into 1 litre boiling water add 20g of potassium hydroxide, mix, and allow to
stand for 2 hours. Add 6m1 of glacial acetic acid (sp gr 1.05) and mix. Add
hydrochloric acid (sp gr 1.19) to adjust the pH value of the solution to 4.0
(check with the help of pH-paper). This has a shelf life of 1 year.
55.4 PROCEDURE
55.4.1.1 Add 5m1 of hydrochloric acid solution and 5 m1 of potassium iodide solution to
a 250 m1 conical flask.
55.4.1.3 Add 2 to 3m1 of starch indicator solution and titrate with potassium iodate
solution to the first persistent blue colour.
55.4.1.4 Record the volume (ml) of potassium iodate solution used as ‘ V1’ .
55.4.1.5 Carry out a blank determination using 100 m1 of water instead of sample and
repeating the steps 4.1.1 to 4.1.3. Record the volume of potassium iodate
solution used as 'V2’
55.4.2.2 Add about 0.lg of sulphamic acid and stir until crystals are completely dissolved
and no further gas evolution is noted.
55.4.2.3 Add 5m1 of hydrochloric acid solution and stir again until no further gas
evolution is noted.
55.5 CALCULATIONS
(V1-V2) X N X 1000
Sulphite, mg/litre as Na2S03 = -----------------------------------------
V
Where:
Where:
Where:
55.6 PRECISION
SO = 1
Where:
55.7 INTERFERENCES
55.7.1 Catalysts, such as copper, cause rapid oxidation of the sulphite when the sample
is exposed to air. This reaction is accelerated by an increase in temperature.
EDTA may be added to the sample to complex such metallic ions.
55.7.2 Other reducing agents, such as sulphides and ferrous-iron, will react like
sulphite.
55.7.3 Nitrite, if present, will oxidise sulphite when the sample is acidified. Nitrite,
interference can be eliminated by the addition of sulphamic acid.
55.8 NOTES
55.8.1 Reagent grade chemicals should be used for preparing all the reagents.
55.8.2 To avoid premature oxidation of sulphite, the titration should be carried out
immediately after collection.
The SDI of water can serve as a useful indication of the quantity of particulate
matter in water.The index can be used to determine the effectiveness of
various processes such as filteration or clarification to remove the particulate
matter.The index is also used to correlate the fouling tendency of reverse
osmosis plants.
56.3 APPARATUS
56.5.11 SDI assembly is schematically shown in fig.1. All the wetted part shall be made
up of high quality SS or plastic to prevent the contamination by
corrosion.Suitable filter holders (SS or plastic) to withstand pressure of 30psi
shall be used.
56.3.2 Membrane filter of 47 mm diameter with mean pore size of 0.45 μm ±0.02 μm.
56.4 PROCEDURE
56.4.2. Flush the apparatus with the water to be tested to remove entrained
contaminants.
56.4.3. Measure the temperature of water.
56.4.4. Open the membrane holder & place a47mm dia membrane filter on support
plate.Handle filter membrane with dull tweezers to avoid damage.touching
with finger must be avoided. (Record the manufacturer indentification for the
membrane filter)
56.4.5. Replace the top half of the holder assembly and close loosely.
56.4.6. Crack open the valve to remove the air from filter. Close valve and tighten filter
holder.
56.4.7. Open ball valve, and simultaneously start stopwatch to record time required for
the flow of 500ml. Record time (tᵢ). Leave the valve open for continued flow.
(The time to collect the 500 ml should be within ±10% of the time required to
collect 500ml of reference nonplugging water.Nonplugging reference water can
be obtained from filtering distilled or DM water through 0.2 μm membrane
filter. If time (tᵢ) is less than 90% of nonplugging time membrane might be
cracked; and if it is more than 110% then smaller sample size (250 or 100 ml)
can be used.
56.4.8. Measure and record the time to collect additional 500 ml volume of sample,
starting the collection time at5, 10 and 15 min of total elapsed flow
time.Measure the water temperature and check the pressure as each sample is
collected.(pressure must remain 30 ± 1 psi throughout the test and temperature
should reamin constant ± 1°C.)
56.4.9. After completion of the test membrane can be retain for future reference.
5. Calculation
56.4.10. Calculate silt density index (SDIT)
(For this test method, % P30 should not exceed 75%. If exceed use shorter time
for T (i.e.5 or 10 min); If % P30 exceed 75% after 5 min other test method to be
used)
56.5 REPORT
56.5.1. Report SDI with subscript indicating the total elapsed flow time (T) in minutes.
56.5.2. Water temperature before and after test.
56.5.3. Manufacturer of membrane and manufacturer identification for membrane filter.
The determination of low level of sodium in high pressure boiler feed and steam
is important.Generally no other method is available to cross check the on-line
analyser (sodium sensitive electrode) results.
57.3 INTERFERENCE
57.4 APPARATUS
57.4.5. Pressure reducing valves: The supplies pressure of fuel and oxidant shall be
maintained somewhat higher than operating pressure of instrument.
57.5 APPARATUS
57.5.1. Purity of Reagents: All reagent grade chemicals should be used in all tests.
57.5.2. Purity of water: The reference water must conforming to specification
D1193,Type I (ASTM)
57.5.3. Sodium stock solution: Dry sodium chloride (AR grade) to constant weight at
105 °C.Dissolved 2.5418 grams in water (Type I) and dilute to1L.(1.00 mL= 1.0
mg Na)
(Certified stock sodium solution available in market is better option.)
Standard Sodium solution (1.00mL= 0.1mg Na) : Dilute 100.00mL of stock
solution to 1L with water.
57.5 STANDARDIZATION
57.5.1. Preapre a blank and at least four standard solutions to bracket the expected
sodium range of unknown sample by diluting the Standard sodium solution with
water. Preapre standard each time of the test is to be performed. Select standards
to give zero,middle and maximum points for an analytical curve.
57.5.2. Aspirate the balnk and the standards and record the instrument readings.
(Computerised recording is also available). Aspirate water between standards.
57.6.1. Prepare analytical curve by plotting absorbance versus concentration for each
standard on linear graph. Alternatively store the readings in data system of
instrument if this capability is available.
57.6 PROCEDURE
57.6.2. Aspirate each sample and determine the absorbance or concentration. Aspirate
water between samples. Calculate the concentration of sodium in samle from
curve or direct reading from the instrument.
(Only reagent water was used to obtain the precision statement since this test
method is designated for the determination of low amounts of sodium in low-
solids water.)
Recoveries of known amounts of sodium (from sodium chloride) in the series of
prepared standards for the same laboratories and operators were as given in
Table 1.
58.2 INTERFERENCE
58.3 APPARATUS
58.3.1 Atomic absorption spectrophometer for use at 766.5 nm. (The manufacturer’s
instructions should be followed for all instrumental parameters)
58.3.2 Potassium hollow cathode lamp: Multi elements hollow cathode lamps can be
utilized for analysis.
58.3.3 Pressure reducing valves: The supply of oxidant and fuel shall be maintained
somewhat higher than operating pressure of the instrument by suitable valve.
58.4 REAGENTS
58.4.2 Purity of water: Water conforming to ASTM D1193 Type I utilized for all test
and standard preparation.
58.4.3 Potassium Stock solution : (1.00 mL= 1.0 mg K):Dry potassium chloride
(reagent grade) to constant weight at 105 °C. Dissolve 1.907 gm of dry KCl in
58.4.5 Potassium standard solution : (1.00mL=0.1 mg K)- Dilute 100mL stock solution
to 1 L with water.
58.4.6 Oxidant: Air, which has been passed through a suitable filter to remove oil,
water, and other foreign substances, is the usual oxidant.
58.5 STANDARDIZATION
58.5.1 Prepare 100 mL each of a blank and at least four standard solutions to bracket
the expected potassium concentrationrange of the samples to be analyzed by
diluting the potassium standard solution with water.Prepare the standard each
time the test is performed. Select the standards to give zero, middle, and
maximum points for an analytical curve.
58.5.2 Aspirate the blank and the standards and record the instrument readings.Aspirate
water between standards.
58.5.3 Prepare an analytical curve by plotting the absorbance versus concentration for
each standard on linear graph paper.Alternatively, read directly in
concentration if this capability is provided with the instrument.
58.6 PROCEDURE
The overall and single-operator precision of this test method for eight
laboratories, which include a total of twelve operators analyzing each sample on
three consecutive days, within its range for reagent water varies with the
quantity being measured according to Table 1.
(Only reagent water was used to obtain the precision statement since this test
method is designated for the determination of trace amounts of potassium in
low-solids water.)
These test methods cover the determination of dissolved and total recoverable
lead in water and waste water by atomic-absorption spectrophotometry:
These methods are used to determine lead content if water and waste water.
59.2.2 INTERFERENCE
59.2.2.1 Other metals usually do not interfere in the determination of lead by increasing
or decreasing the amount of absorbed radiation. The most common interference
is caused by a chemical reaction in the flame that prevents conversion of the
lead to the atomic state.
High concentrations of calcium, such as those connected with the coal industry,
will give lead concentrations higher than which actually exist.
59.2.3. APPARATUS
59.2.4. REAGENTS
59.2.5. STANDARDIZATION
59.2.5.1. Prepare 100 mL each of a blank and at least four standard solutions to bracket
the expected lead concentration range to be analyzed by diluting the lead
standard solution with HNO3 (1 + 499). Prepare the standards each time the test
is to be performed.
59.2.5.2. When determining total recoverable lead, add 0.5 mL of HNO3(sp gr 1.42) to
each blank and standard solution and proceed as directed in 6.2 through 6.4.
After the digestion of the blank and standard solutions has been completed in
6.4, return to 5.3 to complete the standardization for total recoverable
determinations. When determining dissolved lead, proceed with 5.3.
59.2.5.3. Aspirate the blank and standards and record the instrument readings. Aspirate
HNO3 (1 + 499) between standards.
59.2.5.4. Prepare an analytical curve by plotting the absorbance versus the concentration
for each standard on linear graph paper. Alternatively, read directly in
concentration if this capability is provided with an instrument.
59.2.6. PROCEDURE
59.2.6.1. Measure 100.0 mL of a well-mixed acidified sample into a 125-mL beaker or
flask. (If only dissolved lead is to be determined, start with 6.5.)
59.2.6.2. Add 5 mL of HCl (sp gr 1.19) to each sample.
59.2.6.3. Heat the samples on a steam bath or hot plate in a well-ventilated hood until the
volume has been reduced to 15 to 20 mL, making certain that the samples do
not boil.
(For samples having appreciable amounts of suspended matter or dissolved
matter, the amount of reduction in volume is left to the discretion of the
analyst.)
59.2.6.4. Cool and filter the samples through a suitable filter such as fine-textured, acid
washed, ashless paper, into 100-mL volumetric flasks. Wash the filter paper two
or three times with water and adjust to volume.
59.2.6.5. Aspirate each filtered and acidified sample and determine its absorbance or
concentration at 283.3 nm. Aspirate HNO3(1 + 499) between samples.
59.2.7. CALCULATION
Calculate the concentration of lead in each sample, in milligrams per litre, using
the calibration curve established in 5.4.
59.3.3 APPARATUS
59.3.3.1. Atomic Absorption Spectrophotometer, for use at 283.3 nm with background
correction.(The manufacturer’s instructions should be followed for all
instrumental parameters.)
59.3.3.2. Lead Electrodeless Discharge Lamps are satisfactory.
59.3.3.3. Graphite Furnace, capable of reaching temperatures sufficient to atomize the
element of interest.
59.3.3.4. Graphite Tubes, compatible with furnace device.
59.3.3.5. Pipets, microlitre with disposable tips. Sizes may range from 5 μL to 100 μL, as
required.
59.3.3.6. Argon, standard, welder’s grade, commercially available. Nitrogen may also be
used if recommended by the instrument manufacturer.
59.3.3.7. Data Storage and Reduction Devices—Computer- and microprocessor-
controlled devices, or strip chart recorder, should be utilized for data collection,
storage, reduction, and problem recognition (drift, incomplete atomization,
changes in sensitivity, etc.). Strip chart recorders shall have a full-scale
deflection time of 0.2 s or less to ensure accuracy.
59.3.4 REAGENTS
59.3.4.1. Lanthanum Nitrate Solution, (1 mL = 50 mg La)— Dissolve 58.64 g of ACS
reagent grade La2O3 in 100 mL of concentrated HNO3 and dilute to 1000 mL
with water. This solution is added to the lead calibration standard and to the
sample solution as well at a rate of 10 mL per 100 mL of standard or sample
solution.
59.3.4.2. Lead Solution, Stock (1.0 mL = 200 μg Pb)
59.3.4.3. Lead Solution, Standard (1.0 mL = 1.0 μg Pb)—Dilute 5.0 mL of lead solution,
stock and 1 mL of HNO3 (sp gr 1.42) to 1 L with water.
59.3.4.4. Lead Solution, Working—Prepare the working standards at the time of analysis
by adding various volumes of standard lead solution to 50 mL of water
containing 0.5 mL of HNO3(sp gr 1.42) adding 10 mL of lanthanum nitrate
solution (40.1) and diluting to 100 mL with water in a volumetric flask.
59.3.4.5. Nitric Acid (sp gr 1.42)—Concentrated nitric acid (HNO3).
59.3.4.6. Nitric Acid (1 + 1)—cautiously dilute 50 mL of nitric acid (sp gr 1.42) to 100
mL with water.
59.3.5 STANDARDIZATION
Initially, set the instrument according to the manufacturer’sspecifications.
59.3.6 PROCEDURE
59.3.6.1. Clean all glassware to be used for preparation of standard solutions or in the
solubilization step, or both, by rinsing first with HNO3(1 + 1) and then with
water.
59.3.6.2. Measure 100.0 mL of a well-mixed sample into a 125-mL beaker or flask. For
total recoverable lead add HNO3 (sp gr 1.42) to each sample at a rate of 5 mL/L
and proceed as directed in 4 through 6 (below).
59.3.6.3. If only dissolved lead is to be determined, filter the sample through a 0.45 μm
membrane filter prior to acidification, add 10 mL of lanthanum nitrate solution
per 100 mL of sample and proceed to 6.
59.3.6.4. Heat the samples at 95°C on a steam bath or hotplatein a well ventilated fume
hood until the volume has beenreduced to 15 to 20 mL, making certain that the
samples do not boil.
59.3.6.5. Cool and filter the sample through a suitable filter (such as fine texture, acid
washed, ashless paper) into a 100-mL volumetric flask. Wash the filter paper 2
or 3 times with water, add 10 mL of lanthanum nitrate solution and bring to
volume.
59.3.6.6. Inject a measured aliquot of sample into the furnace device following the
directions as provided by the particular instrument manufacturer.
59.3.7 CALCULATION
The dissolved lead results obtained from a graph or instrument readout must be
multiplied by 1.1 to compensate for dilution made by the addition of lanthanum
nitrate solution in 3 (at 3.5)
60 Arsenic in Water
60.1 SCOPE
These test methods cover the photometric and atomic absorption determination
of arsenic in most waters and wastewaters.Three test methods are given as
follows:
Concentration
Range
60.5.2 INTERFERENCES
60.5.2.1. Although many samples are relatively free of interferences, several metals,
notably cobalt, nickel, mercury, silver, platinum, copper, chromium, and
molybdenum, may interfere with the evolution of arsine and with the recovery
of arsenic. The presence of any or all of these metals in a sample being analyzed
must be considered as a potential source of interference, and the analyst must
fully determine the extent of actua linterference, if any. This could be
accomplished by spiking.
60.5.2.2. Hydrogen sulfide and other sulfides interfere, but commonly encountered
quantities are effectively removed by the lead acetate scrubber and the
digestion.
60.5.2.3. Antimony interferes by forming stibine, which distills along with the arsine.
Stibine reacts with the color-forming reagent to form a somewhat similar red sol
having maximum absorbance near 510 nm. The sensitivity for antimony at 540
nm is only about 8 % that of arsenic (1 mg/L of antimony will show an apparent
presence of 0.08 mg/L of arsenic).
60.5.2.4. Nitric acid interferes with the test and must be completely eliminated during the
digestion
60.5.3 APPARATUS
60.5.3.1 Arsine Generator, Scrubber, and Absorber, assembled as shown in Fig. 1.
60.5.6.6. Place in each scrubber a plug of borosilicate wool that has been impregnated
with lead acetate solution. Assemble the generator, scrubber, and absorber,
making certain that all parts fit and are correctly adjusted. Add 3.00 mL of
silver diethyldithiocarbamate-pyridine solution to each absorber. Add glass
beads to the absorbers until the liquid just covers them.
(Four millilitres of silver diethyldithiocarbamate-pyridine solution may be used
with some loss of sensitivity.)
60.5.6.7. Disconnect each generator, add 6 g of zinc, and reconnect immediately.
60.5.6.8. Allow 30 min for complete evolution of arsine. Warm the generator flasks for a
few minutes to make sure that all arsine is released.
60.5.6.9. Pour the solutions from the absorbers directly into clean spectrophotometer cells
and within 30 min measure the absorbance of each at 540 nm.
60.5.7 CALCULATIONS
Determine the weight of arsenic in each sample by referring to the analytical
curve. Calculate the concentration of arsenic in the sample in micrograms per
litre:-
Arsenic μg/L = 1000 W/ V
Where : W- volume of sample in mL
V- weight of arsenic in sample, μg.
graphite furnace uses the sample much more efficiently than flame atomization,
the detection of low concentrations of elements in small sample volumes is
possible. Finally, the absorption signal generated during atomization is recorded
and compared to standards.
60.6.2.2. Dissolved arsenic is determined on a filtered and acidified sample with no
pretreatment.
60.6.2.3. Total recoverable arsenic is determined following acid digestion and
centrifugation. Because chlorides interfere with furnace procedures for some
metals, the use of hydrochloric acid in the digestion or solubilization step is to
be avoided.
60.6.3 APPARATUS
60.6.3.1. Atomic-Absorption Spectrophotometer, for use at 193.7 nm with background
correction.
(The manufacturer’s instructions should be followed for all instrumental
parameters.)
60.6.3.2. Centrifuge, capable of holding centrifuge tubes of 15-mL volume.
60.6.3.3. Centrifuge Tubes, graduated centrifuge tubes of 15-mL capacity with stoppers.
60.6.3.4. Graphite Furnace, capable of reaching temperatures sufficient to atomize
arsenic.
60.6.3.5. Graphite Tubes, compatible with the furnace device. Standard graphite tubes are
recommended for the determination of arsenic.
60.6.3.6. Light Source— Arsenic electrodeless discharge lamps are recommended, but
hollow-cathode lamps may be used.
60.6.3.7. Pipets—Microlitre with disposable tips. Sizes may range from 1 to 100 μL, as
required.
60.6.3.8. Strip-Chart Recorder
60.6.3.9. Automatic Sampling, is recommended if available.
60.6.4 REAGENTS
a hot-plate or steam bath, in a well-ventilated fume hood, until the volume has
been reduced to approximately 10 mL.
60.6.6.4. Cool and quantitatively transfer the sample to a 15-mL centrifuge tube. Dilute
to volume with water, cap the tube, and mix the solution thoroughly. If
undissolved material is present, centrifuge the sample for a few minutes to
obtain a clear solution.
60.6.6.5. Pipet 5.0 mL of the supernatant liquid into a 10-mL volumetric flask. Add 1 mL
of nickel solution and dilute to volume.
60.6.6.6. Inject a measured aliquot of sample into the furnace device following the
directions as provided by the particular instrument manufacturer.
60.6.7 CALCULATION
Use curves to determine arsenic in each sample.
vapor passes through a cell positioned in the light path of an atomic absorption
spectrophotometer. Absorbance is measured as a function of mercury
concentration.
61.3 SIGNIFICANCE & USE
The presence of mercury in industrial discharge, domestic discharge, and
potable water is of concern to the public because of its toxicity. Regulations and
standards have been established that require the monitoring of mercury in water.
This test method provides an analytical procedure to measure total mercury in
water.
61.4 INTERFERENCES
61.4.1 Possible interference from sulfide is eliminated by the addition of potassium
permanganate. Concentrations as high as 20 mg/L of sulfide as sodium sulfide
do not interfere with the recovery of added inorganic mercury from distilled
water.
61.4.2 Copper has also been reported to interfere; however, copper concentrations as
high as 10 mg/L have no effect on the recovery of mercury from spiked
samples.
61.4.3 Seawaters, brines, and industrial effluents high in chlorides require additional
permanganate (as much as 25 mL).
61.4.4 During the oxidation step chlorides are converted to free chlorine which will
also absorb radiation at 253.7 nm. Care must be taken to assure that free
chlorine is absent before mercury is reduced and swept into the cell. This may
be accomplished by using an excess of hydroxylamine sulfate reagent (25 mL).
The dead air space in the reaction flask must also be purged before the addition
of stannous sulfate. Both inorganic and organic mercury spikes have been
quantitatively recovered from sea water using this technique.
61.4.5 Volatile organic materials that could interfere will be removed with sample
digestion as described in 9.2 through 9.4.
61.5 APPARARTUS
61.5.1 The schematic arrangement of the closed recirculating system is shown in Fig. 1
and the schematic arrangement of the open one-pass system is shown in Fig. 2.
(Take care to avoid contamination of the apparatus with mercury. Soak all glass
apparatus, pipets, beakers, aeration tubes, and reaction flasks in nitric acid
(HNO3) (1 + 1), and rinse with mercury-free water before use.)
61.5.2 Atomic Absorption Spectrophotometer—A commercial atomic absorption
instrument is suitable if it has an openburner head area in which to mount an
absorption cell, and if it provides the sensitivity and stability for the analyses.
Also instruments designed specifically for the measurement of mercury using
the cold vapor technique in the working range specified may be used.
61.5.3 Mercury Hollow Cathode Lamp.
61.5.4 Simultaneous Background Correction—Background correction may be
necessary to compensate for molecular absorption that can occur at this mercury
wavelength.
Recorder—Any multirange variable speed recorder that is compatible with the
ultraviolet (UV) detection system is suitable.
61.5.5 Absorption Cell—The cell (Fig. 3) is constructed from glass 25.4-mm outside
diameter by 114 mm (Note below). The ends are ground perpendicular to the
longitudinal axis and quartz windows (25.4-mm diameter by 1.6 mm thickness)
are cemented in place. Gas inlet and outlet ports (6.4-mm diameter) are attached
approximately 12 mm from each end. The cell is strapped to a support and
aligned in the light beam to give maximum transmittance.
(An all-glass absorption cell, 18 mm in outside diameter by 200 mm, with inlet
12 mm from the end, 18-mm outside diameter outlet in the center, and with
quartz windows has been found suitable. Methyl methacrylate tubing may also
be used.)
61.5.6 Air Pump—A peristaltic pump, with electronic speed control, capable of
delivering 1 L of air per minute may be used. Regulated compressed air can be
used in the open one-pass system.
61.5.7 Flowmeter, capable of measuring an air flow of 1 L/min.
61.5.8 Aeration Tubing—A straight glass frit having a coarse porosity is used in the
reaction flask. Clear flexible vinyl plastic tubing is used for passage of the
mercury vapor from the reaction flask to the absorption cell.
61.5.9 Lamp—Asmall reading lamp containing a 60-W bulb is used to prevent
condensation of moisture inside the cell. The lamp shall be positioned to shine
on the absorption cell maintaining the air temperature in the cell about 10°C
above ambient. Alternatively, a drying tube, 150 by 18 mm in diameter,
containing 20 g of magnesium perchlorate, may be placed in the line to prevent
moisture in the absorption cell.
(Warning—If the presence of organic vapors is expected, the purity of the
drying agent should be determined to establish the absence of traces of free
perchloric acid in the salt. This is to prevent the formation of perchloric esters,
some of which are known to be violently explosive compounds.)
61.5.10 Reaction Flask—A250 to 300-mL glass container fitted with a rubber stopper
may be used.
61.6 REAGENTS
61.6.1 Purity of Reagents—Reagent grade chemicals shall be used in all tests. Other
grades may be used, provided it is first ascertained that the reagent is of
sufficiently high purity to permit its use without lessening the accuracy of the
determination.
61.6.2 Purity of Water—Unless otherwise indicated, references to water shall be
understood to mean reagent water conforming to Specification D 1193 Type I.
Other reagent water types may be used, provided it is first ascertained that the
water is of sufficiently high purity to permit its use without adversely affecting
the bias and precision of the test method. Type II water was specified at the time
of round-robin testing of this test method.
61.6.3 Mercury Solution, Stock (1 mL = 1 mg Hg)—Dissolve 0.1354 g of mercuric
chloride (HgCl2) in a mixture of 75 mL of water and 10 mL of HNO3(sp gr
1.42) and dilute to 100 mL with water.
Mercury Solution, Intermediate (1 mL = 10 μg Hg)— Pipet 10.0 mL of the
stock mercury solution into a mixture of 500 mL of water and 2 mL of
HNO3(sp gr 1.42) and dilute to 1 L with water. Prepare fresh daily.
61.8.9 Open One-Pass System—Remove the stopper and frit from the reaction flask,
open the valve, and evacuate the system with vacuum until the absorbance
returns to its minimum value. Close the valve and continue aeration.
(Warning—Because of the toxic nature of mercury vapor, precaution must be
taken to avoid its inhalation. Therefore, a bypass has been included in the
system to either vent the mercury vapor into an exhaust hood or pass the vapor
through some absorbing media such as: (a) equal volumes of 0.1 N KMnO4
solution and 10 % H2SO4, or (b) 0.25 % iodine in 3 % KI solution.)
61.8.10 Proceed with the standards and construct a standard curve by plotting peak
height versus micrograms of mercury.
61.9 PROCEDURE
61.9.1 Transfer 100 mL or an aliquot diluted to 100 mL containing not more than 1.0
μg of mercury to a reaction flask.
61.9.2 Add cautiously 5 mL of H2SO4 (sp gr 1.84) and 2.5 mL of HNO3 (sp sr 1.42)
mixing after each addition.(Refer open one-pass system above)
61.9.3 Add 15 mL of KMnO4 solution to each sample bottle. Shake and add additional
portions of KMnO4 solution until the purple color persists for at least 15 min.
61.9.4 Add 8 mL of K2S2O8 solution to each flask and heat for 2 h in a water bath at
95°C and cool to room temperature.
61.9.5 Turn on the circulating system and adjust its rate to 1 L/min. The pump may be
allowed to run continuously throughout the entire series of samples.
61.9.6 Add 6 mL of sodium chloride-hydroxylamine sulfate solution to reduce the
excess permanganate as evident by loss of solution color.
61.9.7 Wait 30 s and add 5 mL of SnSO4 solution to each flask individually and
immediately attach it to the aeration apparatus.
61.9.8 Continued as describe in * above.
61.10 CALCULATION
61.10.1 Determine the peak height of the unknown from the recorder chart and read the
micrograms of mercury from the standard curve.
61.10.2 Calculate the mercury concentration in the sample as follows:
61.11.4 The precision of this test method within the range from 0.2 to 10 μg/L mercury
may be expressed as follows:
In reagent water, Type II:
St = 0.307X 1 0.183
S0 = 0.076X 1 0.293
In natural waters:
St = 0.386X 1 0.107
S0 = 0.145X 1 0.023
where:
St = overall precision,
S0 = single-operator precision, and
X = determined concentration of mercury, μg/L.
61.11.5 Bias—Recoveries of known amounts of mercury are shown in Table 1.
61.11.6 These data apply to fresh water, saline waters, and some industrial and sewage
effluents. It is the analyst’s responsibility to ensure the validity of the test
method to waters of other matrices or to a particular matrix.
62.3.5 Contamination of the sample with both CO2 and organic carbon is a severe
problem as lower levels of analyte are attempted. Throughout this method the
analyst must be vigilant for all potential sources of contamination and must
monitor blanks and adjust operations to prevent contamination.
62.4 APPARATUS
62.4.1 Apparatus for Carbon Determination—A typical instrument consists of reagent
and sample introduction mechanism, reaction vessel, detector, control system,
and a display. Fig. 1 shows a diagram of such an arrangement.
62.4.2 Sampling Needle— A double chambered needle capable of piercing the sample
bottle septum and pulling sample from the bottom of the bottle is used. The
second chamber vents the top of the bottle to prevent vacuum build up as the
sample is withdrawn. Typically this needle is mounted on an autosampler to
provide unattended analysis of several samples.
62.4.3 I.C. Removal—Vacuum degassing requires the manufacturer’s module5 which
includes a vacuum pump and a hollow fiber membrane assembly. Use of this
vacuum degasser will remove essentially all IC as part of the analysis. The
membrane module consists of a tube and shell arrangement of microporous
polypropylene hollow fibers. Sample flows along the inside of the fibers, while
air is passed on the shell side-counterflow to the sample flow. The shell side
pressure is reduced by means of a vacuum pump on the air outlet. The sample is
acidified before introduction into the degasser to facilitate CO2 transport
through the hollow fibers. Sparging requires an inert vessel with provision for
sparging the acidified sample with 50 to 100 mL/min of carbon free gas. This
procedure will remove essentially all IC in 2 to 10 min, depending on design.
62.4.4 Reactor—The sample flow is split after the addition of reagents. Half of the
flow passes to the delay coil while the other half passes into the oxidation
reactor. The effluent from both streams passes over individual membranes that
allow CO2 to pas through the membrane into prepurified water for detection.
62.4.5 Membrane—The membrane is a CO2 selective fluoropolymer which is
hydrophobic and non-porous. Refer to the bibliography for additional details.
62.4.6 Detector—The CO2 that has passed through the membrane into the purified
water is measured by conductivity sensors. The temperature of the conductivity
cell is also automatically monitored so the readings can be corrected for changes
in temperature.
62.4.7 Data Display— The conductivity detector output is related to stored calibration
data and then displayed as parts per million, (ppm = mg of carbon per litre) or
parts per billion, (ppb = μg of carbon per L). Values are given for TC, IC, and
TOC by difference.
62.5 REAGENTS
62.5.1 Purity of Reagents—Reagent grade chemicals shall be used in all tests. Unless
otherwise indicated, it is intended that all reagents conform to the specifications
of the Committee on Analytical Reagents of the American Chemical Society,
where such specifications are available. Other grades may be used, provided it is
first ascertained that the reagent is of sufficient purity to permit its use without
lessening the accuracy of the determination.
62.5.2 Purity of Water— Unless otherwise indicated, references to water shall be
understood to mean reagent water conforming to Type I or Type II in
Specification D 1193. The indicated specification does not actually specify
inorganic carbon or organic carbon levels. These levels can affect the results of
this test method, especially at progressively lower levels of the carbon content
in the samples to be measured.
Where inorganic carbon in reagent water is significant, CO2- free water may be
prepared from reagent water by acidifying to pH 2, then sparging with fritted-
glass sparger using CO2-free gas (time will depend on volume and gas flow
rate, and should be determined by test). The carbon contribution of the reagent
water should be determined and its effect allowed for in preparation of standards
and other solutions. CO2-free water should be protected from atmospheric
contamination. Glass containers are required for storage of water and standard
solutions. Continuous U.V. treatment of water with recycling through
appropriate mixed bed ion exchange resins may be necessary to maintain
adequately low TOC reagent water.
62.5.3 Persulfate Reagent (15 % w/v)—Prepare ammonium persulfate solution to a
concentration of 15 % w/v by dissolving 15 g of ammonium peroxydisulfate in
water and diluting to 100 mL. Verify that it contains less than 2000 μg/L
62.9.1 The sample is introduced into the instrument by piercing the septum with a
double chambered needle. The analyzer in either the grab sample or autosampler
mode will pull sample out of the sample bottle, into the analyzer.
62.9.2 If inorganic carbon is to be removed by vacuum degassing, or if I. C. removal is
unnecessary, no additional sample preparation is required. If inorganic carbon is
to be removed by sparging prior to sample introduction, acidify to
approximately pH 2 with concentrated acid (if not already done) inject through
the needle and sparge through the needle with an appropriate flow of gas. Use of
the double chambered needle prevents contamination of the sample during
sparging.
Other configurations that prevent sample contamination are acceptable. Samples
with high alkali content or buffer capacity may require larger amounts of acid.
In such cases, incorporate this dilution into the calculation of results. If
incomplete sparging of CO2 from IC is suspected, sparge and analyze the
sample and then repeat the procedure until appropriate conditions are
established.
62.9.3 Follow manufacturers’ instructions for introducing the sample into the analyzer.
The sample may be directly aspirated, sampled from an auto sampler, or
connected directly into a source for continuous on-line monitoring.
62.10 CALCULATION
62.10.1 Read carbon values directly from the digital display, printer, or computer
connected to a suitable data interface on the instrument.
62.10.2 Blank correction should be applied to samples that have been prepared or
diluted with reagent water. In the case of materials prepared in reagent water,
the entire blank value obtained in the reagent water should be subtracted. In the
case of dilution, the blank should be corrected for the dilution factor and
subtracted from the obtained value.
62.11 PRECISION & BIAS
Linearity of the response over the entire measurement range allows calibration
at a single higher level concentration. This facilitates preparation of accurate
standards minimizing the effect of contamination. Fig. 2 shows instrument
response for carbon versus carbon concentration over five orders of magnitude
from 0.25 μg/L C to 25 000 μg/L C for two instruments calibrated at 25 000
μg/L C. As stated in Section Calibration, the user should confirm proper
operation of the instrument by running check samples in the range of test
samples.
Annexure – I
ATOMIC WEIGHT/ATOMIC NUMBERS OF VARIOUS ELEMENTS
ANNEXURE - II
DETERMINATION OF SATURATION INDEX (LANGELIER)
AND STABILITY INDEX (RYZNER) OF COOLING WATER
1.0 SUMMARY OF METHOD
The saturation index and the stability index are calculated from the total
alkalinity, calcium hardness, pH, approximate concentration of total dissolved
solids and the temperature of the water.
2.0 PROCEDURE
2.1 Determine the total alkalinity, calcium hardness, pH, total dissolved solids and
the temperature of the water.
2.2 To determine the saturation Index (Langelier Index), obtain the values of A,B,C
and D from table 1 (Section 3.0) and calculate the saturation pH as follows:
pHs = (9.3 + A+B) - (C+D)
Then, using the actual pH of the water, calculate the index as follows:
Saturation Index (Langelier Index) = pH - pHs
2.3 Calculate the stability Index (Ryzner Index) as follows:
Stability Index (Ryzner Index) = 2 pHs - pH
TABLE 1
4.0 NOTES
4.1 The Langelier index or calcium carbonate saturation index of cooling water is
used in prediction the scaling of corrosive tendency of the water.
4.2 If the saturation index (Langelier index) is zero, water is in chemical balance. If
the index is positive, scaling tendencies are indicated. If the index is negative
corrosive tendencies are indicated.
4.3 The Ryzner stability index is a modification of the Langelier saturation index.
All values of this index will be positive and values above 6.0 indicate a
corrosive tendency while values below 6.0 indicate a tendency to form scale.
4.4 The control limits of cooling water are given below:
Min. Max.
Langelier saturation index 0.5 1.5
Ryzner stability index 6.0 6.0
pH 6.0 8.0
Calcium hardness, mg/litre as CaCO3. 20 300
Total dissolved solids mg /litre - 2,500
ANNEXURE - III
prepared fresh
b. One week i
stored at 40 C
22. Barium Chloride 1 year Glass
23. Barium Nitrate 1 year Glass
24. Baryta water 1 month Glass
25. Benzene 1 year Glass
26. Bromine water 1 week Glass
27. Cadmium Acetate 1 year Glass
28. Cadmium Sulphate 1 year Glass
29. Calcium Chloride 1 year Glass
30. Ceric Sulphate 1 month Glass
31 Calcium Nitrate 1 year Glass
32. Calcium Sulphate 1 year Glass
33. Carbon Tetrachloride 1 year Glass
34. Chlorine water 1 week Glass
35. Chromic Acid 1 year Glass
36. Cobalt Chloride 1 year Glass
37. Cobalt Nitrate 1 year Glass
38. Copper Sulphate 1 year Glass
39. Dimethyl Glyoxime Solution 1 month Glass
40. Diphaenyl Amine 1month Glass
41. diphenyl Carbazide 1 week: Glass
4 month:
(3.32g/1)
1 month:
(2.89g/1)
42. Diphenyl Thiocarbzone 1 month Glass
( Satndardize weekly )
90. Pyrogallol 1 month Glass
91. Silver Nitrate (standardize monthly) 1 year Amber Bottle
92. Silver sulphate 1 year Glass
93. Sodium Acetate 1 year Glass
94. Sodium Carbonate 1 year Glass
95. Sodium Arsenite 1 year Dark coloured Glass
Bottle.
96. Sodium chloride 1 year Glass
97. Sodium Hydrogen Carbonate 1 year Glass
98. Sodium Hydroxide (standardize 1 year polythene Bottle
monthly)
ANNEXURE - IV
Maximum total
bacteria count 0/ml 10/ml 100/ml
polishing with a mixed bed of ion exchange materials and a 0.2 micron
membrane filter.
3.2 Type II grade of reagent water shall be prepared by distillation of water having a
conductivity of less than 1.0 microsiemens/cm at 250C . Ion exchange,
distillation, or reverse osmosis may be required as an initial treatment prior to
distillation if the purity cannot be attained by single distillation.
3.3 Type III grade of reagent water shall be prepared by distillation, ion exchange,
reverse osmosis, or a combination thereof, followed by polishing with a 0.45
micron membrane filter.
4.0 USE OF REAGENT WATER
4.1 Type I grade of reagent water shall be used where maximum accuracy and
precision is indicated, provided dissolved organic matter is not a possible
interference.
4.2 Type II grade of reagent water shall be used for most analytical procedures and
all procedures requiring water low in organics.
4.3 Type III grade of reagent water shall be recommended for general laboratory
testing.
5.0 TEST METHODS
5.1 Total Solids – see method B; the determination of suspended and total dissolved
solids.
5.2 Electrical Conductivity – see method for the determination of conductivity.
5.3 pH – see method for the determination of pH.
5.4 Consumption of potassium per magnate – add 0.20ml of potassium
permanganate (0.316g/litre) to a mixture of 500ml of the reagent water and 1
ml of sulphuric acid (sp gr 1.84) in a stoppered bottle of chemically resistant
glass. Consider the reagent water as having passed the test if the permanganate
colour does not disappear completely after standing for the indicated period of
time at room temperature. This test should be run against a blank using water
known to be free of organic substances.
5.5 Silica – see the method for the determination of soluble silica.
ANNEXURE - V