Pet QC PPT2
Pet QC PPT2
Principles of PET/CT
Quality Control and Calibration
skappadath@[Link]
Educational Objectives
Review principles of PET and PET/CT imaging
Overview of PET performance testing
– NEMA NU-2 2001/2007 and ACR
PET/CT Image Quality and Artifacts
Recent advances in PET/CT
1
7/22/2014
Positron Decay
Nuclei with low a neutron-to-proton ratio converts a
proton to a neutron via emission of positron ( +)
p = n + + + ; AXZ = AYZ-1 + + +
Cyclotron (generator) for production of + emitters
13
14
N(p,)11C 16
O(p,)13N C(p,n)13N 14
N(d,n)15O
15
N(p,n)15O 16
O(p,pn)15O 18 18
O(p,n) F 20
Ne(d,)18F
Electron capture competes with positron decay
AX = AY
Z Z-1 + X-rays
– Branching Ratios
– 18F = 0.967
Annihilation Photons
Energy spectrum of + emission is continuous
+ range depends on energy
– 18F: Emax = 0.64 MeV, Range ~1 mm
– 82Ru: Emax = 3.15 MeV, Range ~2 mm
+ annihilation results in simultaneous emission of
– Two 511 keV photons
– Emitted (nearly) 180 degrees apart g: 511 keV
-
+
g: 511 keV
S. Cheenu Kappadath, PhD AAPM 2014 5
electron
positron
Nucleus
Annihilation
photon
2
7/22/2014
PET detectors
3
7/22/2014
Compact, MR compatible
Poor timing resolution
4
7/22/2014
LOR to Sinograms
5
7/22/2014
Correct Random
Normalize
Correct Geometry
Scanner Calibration
Correct Attenuation
Randoms Correction
1) Randoms Rates from Singles Rates, R = 2t x S1 x S2
– Randoms are proportional to S2
– Statistically more accurate since S>>R
2) Randoms Rates from Delayed Prompts (Dt >> 10 ns)
– Real-time subtraction
– Identical deadtime characteristics to Prompts channel
– Requires more memory and statistically less accurate
Detector Detector
10 ns 10 ns
Constant Constant
Delay Fraction
Fraction
Discriminator >> 10ns Discriminator
AND
Normalization
Sinogram
Pre- Post-
Normalization Normalization
6
7/22/2014
Clinical
Imaging
Conditions
Geometry Correction
Ring detector introduces non-uniform sampling of
LOR away from isocenter
Geometry correction maintains uniform pixel size in
transverse plane
Position
7
7/22/2014
P = P1 x P2
= e-ua x e-ub
= e-u(a+b)
= e-uD
of source location
Energy (keV)
Photon energies different between
CT and SPECT
0.200
K≈1 for Compton Scatter
Attenuation at 511keV
8
7/22/2014
Role of CT in PET/CT
Two functions for CT
as part of PET/CT
CT Dose
Requirement
Anatomic
Higher
Localization
AC of PET (Diagnostic )
Ultra-low
Loss of PET accuracy (CT-AC only)
from incorrect CT-AC
9
7/22/2014
2D versus 3D PET
10
7/22/2014
2D versus 3D PET
2D PET: Collimation septa present between detector
planes in axial direction
– Reduces scatter; Uniform AX sensitivity (1 cm bed overlap)
3D PET: No collimation present except at end of ring
– Sensitivity 3D > 2D lower activity needed
– Randoms & Scatter 3D > 2D; Improvements in modeling of
the random and scatter events Standard Acq. mode
– Triangular AX sensitivity profile (~50% detector overlap)
Correct Random
Normalize
Correct Geometry
Scanner Calibration
Correct Attenuation
SAM Question 1
The attenuation of PET coincident events
emitted from the patient depends on the:
11
7/22/2014
SAM Question 2
The well counter calibration for a PET scanner
is used to:
12
7/22/2014
NEMA Tests
Spatial Resolution
Sensitivity
Scatter Fraction/Count
Rate Performance
Image Quality
Accuracy of correction for
count losses and randoms
Daube-Witherspoon M. et al JNM,
43(10) 1398-1409, 2002
13
7/22/2014
Placed at
isocenter
70 cm
14
7/22/2014
15
7/22/2014
16
7/22/2014
Cylinders Background
S. Cheenu Kappadath, PhD AAPM 2014 49
1 2 3
0.95
Relative AC (Measured/True)
0.90
0.85
0.80 Subsets 1
SUV
0.75
0.70
Subsets 3
Subsets 15 mean
Subsets 21
Subsets 45
0.65
Subsets 63
0.60
0 5 10 15 20 25 30 35 40 45
Max AC, SBR: 5-to-1, Sphere ID: 37 mm
1.40 Iterations
1.35
Relative AC (Measured/True)
1.30
1.20
SUV
1.15
max
1.10
1.05
1.00
0 5 10 15 20 25 30 35 40 45
Iterations
SBR: 10-to-1
2000
1600
Background Std-Dev
1200
Subsets 1
Image
Subsets 3
800
Subsets 15
Subsets 21
Noise
400 Subsets 45
Subsets 63
0
0 5 10 15 20 25 30 35 40 45
Iterations
17
7/22/2014
Bailey, Townsend, Valk, and Maisey, “Positron Emission Tomography,” Springer-Verlag, 2005
18
7/22/2014
Truncation Artifacts
SAM Question 3
All of the following affect PET image quality
except:
7% A. Reconstruction parameter
14% B. Scan duration
14% C. CT scan technique
7% D. Patient size
19
7/22/2014
SAM Question 4
The minimum CT dose appropriate for PET/CT
examinations are constrained by:
20
7/22/2014
Time-of-Flight PET
21
7/22/2014
w/o w/
Sensitivity
FOV
Fewer bed positions for same axial coverage
Increased sensitivity time/bed or counts/time
Net reduction in imaging time (or administered
activity) for comparable image quality
Image courtesy: D Townsend
S. Cheenu Kappadath, PhD AAPM 2014 66
22
7/22/2014
SAM Question 5
The main advantage of a TOF PET scanner
over a non-TOF PET scanner is:
23
7/22/2014
Trigger Trigger
1 8 1 8
2
2 7
7
3 3 6
6
4 4 5
5
time
Bin 1 Bin 8
• Prospective fixed forward time binning
• Single FOV Gated PET and Gated CT
• User defined number of bins and bin duration
• Images will be noisy unless acquired for longer durations
Image Courtesy: Tinsu Pan
0.8 mm/s
2’
HD•Chest
2’
0.5 mm/s
2’
2’ 0.8 mm/s
Speed
Higher
2’ 2.0 mm/s
24
7/22/2014
77 years male with follicular lymphoma, 80 kg, 25 BMI, 9.4 mCi, 60 min post injection
25
7/22/2014
References
SR Cherry, JA Sorenson, ME Phelps, “Physics in Nuclear
Medicine, 3rd Edition,” Saunders Elsevier, 2003
DL Bailey, DW Townsend, PE Valk, and MN Maisey, “Positron
Emission Tomography,” Springer-Verlag (London), 2005
M Conti, “Focus on time-of-flight PET: the benefits of
improved time resolution,” EJNMMI 38, 1147-1157, 2011
O Mawlawi, SC Kappadath, T Pan, E Rohren, HA Macapinlac,
“Factors affecting quantification in PET/CT imaging,” Current
Medical Imaging Reviews 4, 34-45, 2008
FH Fahey, MR Palmer, KJ Strauss, RE Zimmerman, RD Badawi,
ST Treves, “Dosimetry and adequacy of CT-based attenuation
correction for pediatric PET: Phantom study,” Radiology 243,
96–104, 2007
26