Topclean S CPC 30 Ti235cen
Topclean S CPC 30 Ti235cen
TI 235C/07/en/04.03
Automation of
No. 51504335
pH/Redox Measurements
TopClean S CPC 30
086
• Short amortisation times:
– Low procurement price
– Low maintenance costs through
automatic cleaning of electrode
– Low installation effort through
modular design
• Approved for Ex applications
• Communication via PROFIBUS-PA and
HART
TopClean S CPC 30
In the Ex version, the CPC 30 is powered via the power supply/control cable from the CPM 153;
in the non-Ex version, the CPC 30 has its own power supply connection.
2 Endress+Hauser
TopClean S CPC 30
MEAS
DIAG
CAL
PARAM
B
2 4
1
Endress+Hauser
CPG 300
5
3
6
C07-CPC30xxx-14-03-xx-xx-001.EPS
11
12 7
13 8 10
9
Customer supplies:
9: Additional valve
10: Cleaning solution
11: Electrical wiring
12: Compressed air
13: Liquids / cleaning mixture
Endress+Hauser 3
TopClean S CPC 30
- .
Zone 0/1 Zone 1
MEAS
DIAG
CAL
PARAM
B
2 4
1
Endress+Hauser
CPG 300
5
3
6
C07-CPC30xxZ-14-03-xx-xx-001.EPS
11
12 7
13 8
9 10
Customer supplies:
9: Additional valve
10: Cleaning solution
11: Electrical wiring
12: Compressed air
13: Liquids / cleaning mixture
Operating modes For configuration, select the appropriate cleaning programme for your measuring point from the
stored programmes in the instrument. You can freely adapt the cleaning programmes to your
requirements or deactivate them as necessary.
4 Endress+Hauser
TopClean S CPC 30
Select the control for one external additional valve for your device under Order Code item
"External valve control". The "sterilisation" or "sealing water"* functions are only enabled on
devices fitted with a control function for additional external valves. You can freely use additional
external valves as required in the freely definable user programs. For example, for pressurised
media such as superheated steam, a second cleaner, cooling air, organic cleaner etc.
We recommend to use the rinse block CPR 40 for all the media, which shall be controlled by the
additional valves.In any case, you have to convey hot and aggressive media via the additional
valves and the rinse block (see "Materials").
Clean S
Control for 1 valve
(= cleaning + ✓ –
required
sterilisation)
Clean Int
Control for 1 valve
(= cleaning ✓ –
required
interval)
User 1 ✓
1 extra external valve can be used as required, e.g. for
User 2 ✓ superheated steam or organic cleaner.
Device control for 1 valve is required.
User 3 ✓
*Sealing water
In processes with fibrous or adhesive media, assemblies with ball valves are fitted to block the
medium, e.g. CleanFit CPA 473 or CleanFit CPA 475. To keep the rinse chamber free of medium,
the sealing water valve opens automatically before the assembly emerges from the process. The
counterpressure in the rinse chamber caused by the sealing water prevents the ingress of
medium into the chamber. The sealing water pressure must then be greater than the pressure of
the medium.
Endress+Hauser 5
TopClean S CPC 30
Logbooks
There are several logbooks available: The last 30 entries are saved to an error logbook, an
operation logbook and a calibration logbook. You can retrieve the entries by specifying a date and
time.
Data logger
You can record two freely selectable parameters using the integrated data loggers and then view
the results graphically in real time. You can retrieve the last 500 measured values using date and
time. In this way, you can graphically display the process flow. This is a quick way of checking the
process and provides a good possibility of optimising pH control.
➀ ➃
C07-CPM153xx-20-06-00-en-001
Measure
2.00 12.00
7.54 12 15 35 ➄
➁ 01 04 2001
➅
➂ Select
➆
6 Endress+Hauser
TopClean S CPC 30
Simple to control
The follow control functions are used in the CPC 30 :
• Limit value contact
– Two-point controller with hysteresis for simple temperature control, for example
• PID controller
– for one and two-sided processes
– with freely adjustable P, I and D components
– Includes configurable range-dependent gain (kinked curve)
– Distinction between batch and flow (in-line) processes.
• Manipulated variable output
The manipulated variable can be output as a binary signal via relays or as a continuous signal
via the current output:
– binary signal via relay as PWM (pulse length proportional), PFM (pulse frequency propor-
tional) or dynamic PWM
– Current output (0/4 ... 20 mA): Analogue signal to control the actuator (for one or two actuator
drives / "single" or "split range")
DAT module
The DAT module is a memory device (EEPROM) which is inserted in the terminal compartment of
the transmitter. Using the DAT module, you can
• save complete settings, logbooks and the data loggers of the CPC 30 system and
• copy your settings to other CPM 153 transmitters which have identical hardware functionality.
This considerably reduces the effort to install or service several measuring points.
Refer to the graphic on page 24 for information on the Parawin user interface.
Endress+Hauser 7
TopClean S CPC 30
Accurate measurement
• Medium temperature compensation (alpha value compensation)
This allows high-accuracy measurement over wide temperature ranges. In this type of
compensation, the temperature influence on the medium is self-compensated.
• Isothermic intersection compensation
This allows high-accuracy measurement even at temperature fluctuations. Compensates the
deviation between electrode zero point and isothermic intersection point.
8 Endress+Hauser
TopClean S CPC 30
!
Input
Note!
The limit values for Ex version are specified separately, marked with 0 and displayed in italics.
0
Input current under nominal operating conditions < 1.6 · 10–12 A
Sensor output circuit with type of protection EEx ia IIC. This circuit can be connected
to sensors of category 1G (zone 0) as well.
Temperature Offset ± 5K
Endress+Hauser 9
TopClean S CPC 30
0
Input voltage range 6 ... 30 V
Intrinsically safe current inputs for connection with intrinscally safe electric circuits with
type of protection EEx ia IIC or EEx ib IIC.
0
Internal resistance Ri = 5 kΩ
Intrinsically safe optoelectronic coupler for connection with intrinsically safe electric
circuits with type of protection EEx ia IIC or EEx ib IIC.
1
: acc. to IEC 746-1, under nominal operating conditions
CPG 30:
Internal resistance Ri = 5 kΩ
10 Endress+Hauser
TopClean S CPC 30
Output
Output signal pH, redox, temperature
0
Passive current output: Input voltage range 6 ... 30 V
Intrinsically safe current circuits for connection with intrinscally safe electric circuits with
type of protection EEx ib IIC.
0
Output current max. 9 mA
Intrinsically safe current output circuit with type of protection EEx ib IIC.
0
Communication (only internal) RS 485
Intrinsically safe current output circuit with type of protection EEx ib IIC.
Endress+Hauser 11
TopClean S CPC 30
0
With PWM minimum switch-on period 0.4 s
Intrinsically safe relay contact circuits for connection with intrinscally safe electric circuits
with type of protection EEx ia IIC or EEx ib IIC.
1
: acc. to IEC 746-1, under nominal operating conditions
47
48
Zusatz-Ventil
(optional)
N/L–
L/L+
12 Endress+Hauser
TopClean S CPC 30
Electrical connection
Connection diagram Non-Ex area
Installation and connection
by operator
Mycom CPM 153 CPG 30
+ S
81
Digital (Hold) L+ +BN 12V DC L+
input E1* 82 L/L+
+ L– -WH L– N/L– Power supply
93
Digital GN PE
A A )
input E2* 94
B YE CPG comm. B Compressed air
supply 4...6 bar,
+ 85 94 filtered 0.5 µm, free
15 V
Power supply Assembly of oil and condensate
– 86 93 service*
output* AD 6 / ID 4
41 92
Assembly
Alarm relay 42 91 measure*
(NAMUR: Failure
47 90
Relay 1 48 n.c. 41
89
(NAMUR: Mainten. requ.) Valve (optional)
57 88 42
Automatic
Relay 2 58 87 stop*
(NAMUR: Function control)
86
66
Assembly
+ 31
mA 65 service
Current - 32 62
output 1 Assembly
61 measure
+ 33
mA
Current 72
- 34 Program
output 2 71 running
L/L+ 1 *
N/L– 2
Power supply PE Multihose coupling
) Injector
(operator supply) CYR 10
Pipe disconnector
(customer supplied)
C07-CPC30xxx-04-12-00-en-001.EPS
Motive water
2...12 bar
Endress+Hauser 13
TopClean S CPC 30
Ex area
Installation and connection
by operator
Mycom CPM 153 (Ex) CPG 30 (Ex)
+ S
81
Digital L+ +BN 12V DC L+
82
ib
input E1 L– -WH L–
+ 93
Digital A GN A
94 ib
input E2 YE CPG comm. B
Compressed
B air supply 4...6 bar,
filtered 0.5 µm,
+ 94 free of oil and
15 V Assembly
Power supply ib ib service*
condensate
– 86 93
output AD 6 / ID 4
41 92
ib Assembly
Alarm relay 42 91 measure*
47 90
Relay 1 48
ib n.c.
89
88 Valve (optional)
57
ib Automatic
Relay 2 58 87 stop*
86
76
Assembly
+ 31 ib Compressed
mA 75 service
air supply 4...6 bar,
Current - 32 ib 72 filtered 0.5 µm,
output 1 ib Assembly free of oil and
71 measure
+ 33
mA
condensate
Current ib 66
- 34
ib
Program
output 2 65 running
2 1
L/L+
Power supply N/L–
(ex e) PE Multihose
) Injector CYR
(operator supply) 10Z-E12/
-F12
Pipe disconnector
(customer
C07-CPC30xxZ-04-12-00-en-001.EPS
supplied)
Motive water
2...12 bar
Cleaner
14 Endress+Hauser
TopClean S CPC 30
Contacts CPM 153 In the transmitter Mycom S CPM 153, there are six relays available which you can configure using
the software.
The ChemoClean® spray cleaning system with the injector CYR 10 automatically cleans the
electrode. It is controlled by two contacts.
! Note!
• If you use NAMUR contacts (acc. to recommendations of the community of interest process
control engineering of the chemical and pharmaceutical industry), the contacts are set to the
relays as follows:
– Failure to "ALARM"
– Maintenance required to "RELAY 1" and
– Function control to "RELAY 2".
• You can assign up to three relays to the controller, dependent on the instrument version.
Electrical
connection data
Mycom S CPM 153:
Frequency 47 ... 64 Hz
0
Terminals, max. cable cross-sectional area 2.5 mm²
Endress+Hauser 15
TopClean S CPC 30
CPG 30:
Frequency 47 ... 64 Hz
0
Terminals, max. cable cross-sectional area 2.5 mm²
CPG 30 (Ex) : The devices in Ex version are powered from the transmitter CPM 153 (refer above for data).
16 Endress+Hauser
TopClean S CPC 30
Connection examples
90
n.c.
47 89
48 Maintenance 88
Automatic
57 stop 87
58 Function control
Assembly 76
51 service 75 (max. 3 W)
Power 52 pH limit value
72
supply Assembly max. 30 V
measure 71 (max. 3 W)
54
Program 66
55 Relay 4 (max. 3 W)
running 65
44
Air
45 Relay 5 Pneumatic connection
L/L+ Maintenance
N/L– Power supply unit
PE 51
52
pH Multihose Control for 47
electrode Temperature coupling additional valve 48
42 41
pH Ref.PA 11 12 13
1 *
2
1 *
2
Injector
CYR 10
Pipe disconnector
(customer
supplied)
Motive water
C07-CPC30xxx-04-12-00-en-002.EPS
Cleaner
Endress+Hauser 17
TopClean S CPC 30
90
47 n.c.
89 ib
48 Maintenance
88
57 Automatic
stop 87 ib
58 Function control
51 Assembly 76 L+ L–
ib
Power 52 pH limit value service 75
supply 72
Assembly
L+ L– measure 71 ib
54 e.g. 3 x
66
Relay 4 Program
ib 55 running 65 ib STAHL 925O
L+ L–
44
Air
ib 45 Relay 5 Pneumatic connection
L/L+
51 Maintenance
N/L– Power supply unit
Multihose
coupling
52
PE
47
pH
electrode Temperature Control for 48
additional valve
pH Ref.PA 11 12 13
1 2
Compressed
2 1
ia air supply
Injector
CYR 10Z-E12/
-F12
Pipe disconnector
(customer
supplied)
Motive water
e.g. second 2...12 bar
cleaner
C07-CPC30xxZ-04-12-00-en-002.EPS
Alkali Acid
18 Endress+Hauser
TopClean S CPC 30
Performance characteristics
Measured value resolution pH: 0.01
Redox: 1 mV / 1%
Temperature: 0.1 K
1
: acc. to IEC 746-1, under nominal operating conditions
Installation conditions
Installation instructions • Always install the transmitter and the control unit so that the cable entries always point down-
wards.
The components can be installed using the following methods:
Control unit CPG 30 Mounting kit contained not applicable not applicable
in scope of supply.
Refer to figure below.
Endress+Hauser 19
TopClean S CPC 30
If you want to mount the device on the front of an air-tight panel, you must also use a flat seal (see
"Accessories").
Required installation cutout for panel mounting: 161+0.5 x 241+0.5 mm.
The installation depth is approx. 134 mm.
The maximum pipe diameter is 60 mm.
For outdoor use, the weather protection cover CYY 101 is required. This is available as an acces-
sory.
concentrated cleaner
max. 5 m
Motive water volume min. 2 l/min.
max. 10 l/min.
max. 3 m
20 Endress+Hauser
TopClean S CPC 30
Ambient conditions
Ambient temperature –10 ... +55°C (Ex: –10 ... +50°C)
Ambient temperature limit –20 ... +60°C (Ex: –10 ... +50°C)
Electromagnetic compati- Interference emission and interference immunity to EN 61326: 1997 / A1:1998
bility
Process conditions
Temperature range max. 60°C
media conveyed
Pressure of media Aggressive or hot media which are conveyed via additional valves (optional) must not be
conveyed conveyed by the injector CYR 10 of the CPC 30 system. Use the rinse block CPR 40 for that.
Mechanical construction
Design / dimensions
320
300 200 227
Endress+Hauser
CPG 300
24
420
327
350
400
C07-CPC300xx-06-12-01-xx-001.EPS
250
Endress+Hauser 21
TopClean S CPC 30
ENDRESS+HAUSER
MYCOM S CPM 153
Messen
pH 7.54
Wahl
154
MEAS CAL
DIAG PARAM
247
C07-CPM153xx-06-06-00-xx-001.EPS
82
142
100 167
22 Endress+Hauser
TopClean S CPC 30
7.54
7.54
pH
pH ATC
Temperature 25.0 °C
Select Select
Co07-CPM153xx-19-06-00-en-005
Measure
pH 7.54 0 mV
Out 1 10.00 mA
Out 2 0.00 mA
Rel. A 1 2 3 4 5
Select
ENDRESS+HAUSER
1 MYCOM S CPM 153 4
Measure HOLD 5
2
3 pH
Temperature
7.54 25.0 °C
6
Select Off
11
C07-CPM153xx-19-06-00-en-002.EPS
7
12
MEAS CAL
8
9
DIAG PARAM 13
10 ?
Endress+Hauser 23
TopClean S CPC 30
Operating elements There are 4 main menus available for instrument operation:
Mycom S CPM 153 • Measurement ("MEAS")
• Configuration ("PARAM")
• Calibration ("CAL") and
• Diagnostics ("DIAG").
Press the "MEAS", "PARAM", "CAL" and "DIAG" keys to switch to the appropriate selection menu.
Here, the possible options are displayed in plain text and selected elements are displayed in
reverse video. Selections are made using the arrow keys, which are also used to edit the numeric
values.
The PC tool Parawin provides you with a tool for configuring your measuring point on a PC using
a simple and self-explanatory menu structure (an example window can be seen above). Write the
configuration to the DAT module using the RS232 interface on the PC. The data can then be trans-
ferred to the transmitter.
4
EC directives. Endress+Hauser confirms the successful testing of the system by affixing the
symbol.
24 Endress+Hauser
TopClean S CPC 30
Ordering information
Product structure for the Scope of delivery of basic equipment:
complete system Control unit CPG 30, transmitter Mycom S CPM 153 with 6 relays and DAT module, injector CYR
TopClean S CPC 30 10, multihose (5 m), hose clip, power supply cable CPM 153 – CPC (5 m)
Approvals
A Basic equipment: non-Ex
G With ATEX approval II (1) 2G EEx em ib[ia] IIC T4
O With FM approval CI. I, Div. 2, with NI input and output circuits, sensor IS Cl. I Div. 1
P With FM approval CI. I, Div. 2, with NI input and output circuits
S With CSA approval Cl. I, Div. 2, with NI input and output circuits, sensor IS Cl. I Div. 1
T With TIIS approval
Power supply
0 230 V AC
1 100 ... 115 V AC
8 24 V AC / DC
Language versions
A E/D
B E/F
C E/I
D E / ES
E E / NL
F E/J
Cable connection
0 Cable glands M 20 x 1.5
1 Cable glands NPT ½"
3 Cable glands M 20 x 1,5, PROFIBUS-PA M12 plug
4 Cable glands NPT 1/2", PROFIBUS-PA M12 plug
Length of multihose
0 5m
1 with electrical heating, 5 m
2 with electrical heating, 10 m
8 10 m
Additional features
0 Standard version
1 Preparation for housing CYC 300
9 Special version acc. to customer specifications
Configuration
A Factory settings
Endress+Hauser 25
TopClean S CPC 30
Accessories
Offline configuration The Parawin tool provides you with a graphic PC operating program for configuring your measur-
with Parawin ing point at the PC using a simple and self-explanatory menu structure. Write the configuration to
the DAT module using the RS232 interface on the PC. The module can then be plugged into the
transmitter. You can switch the language via software.
The offline configuration system consists of a DAT module, the software and a DAT
interface (RS 232). Required operating sytem: Windows NT/95/98/2000.
Order No.: 51507133 (only Mycom S CPM 153)
Order No.: 51507563 (TopCal S, TopClean S / Mycom S)
DAT module Additional memory device for saving/copying complete settings, logbooks and the data loggers.
Order No.: 51507175
Flat seal Flat seal for air-tight panel mounting of the transmitter CPM 153.
Order No.: 50064975
CeraLiquid Electrodes with ceramic diaphragms and KCl liquid • Process industry
CPS 41/42/43 electrolyte, use with counterpressure, explosion-proof up • Ultrapure water
to 8 bar • Boiler feed water
Technical Information TI 079C/07/en, 50059346 • Detoxification (cyanide)
OrbiPore Gel electrode with open aperture diaphragm for heavily • Process industry
CPS 91 soiled media. Resistant to pressure and concentration • Industrial wastewater
fluctuations. Pressures up to 13 bar.
Technical Information TI 375C/07/en, 51513127
26 Endress+Hauser
TopClean S CPC 30
Connection accessories • CPK 1: Version with pilot wire and additional outer screen, sheathed in PVC, ø 7.2 mm. Exten-
sion with cable CYK 71 possible. Order No.: of CYK 71: 50085333
• CPK 9: For pH/redox electrodes with integrated temperature sensor and TOP68 plug-in head
(version ESA, ESS). Extension with cable CYK 71 possible.
• CPK 12: For IsFET pH sensors and pH/redox electrodes with integrated temperature sensor and
TOP68 plug-in head. Extension with cable CYK 12 possible.
• Junction box VBE: For Ex area Zone 0. Order No.: 50003993
• Junction box VBM: Junction box for extending measuring cable connection between electrode
and transmitter. Two screw unions for e.g. combination electrode.
Material: aluminium casting, ingress protection IP 65. Order No. 50003987
• Junction box VBA: Junction box for extending measuring cable connection between electrode
and transmitter. Four screw unions for e.g. separate reference electrode. Material: aluminium
casting, ingress protection IP 65. Order No. 50003987
Rinse connection adapter Rinse connection adapter CPR 40 for the transport of cleaning agents for use with retractable
assemblies.
Spray cleaning system CYR 10 / CYR 20 ChemoClean Spray Cleaning System for the transport of cleaning agents or
acids for use with retractable assemblies.
Weather protection cover For installing the transmitter outdoors, the weather protection cover CYY 101 is required.
CYY 101 Order No.: CYY101-A
Round post fixture for To fix the weather protection cover to vertical or horizontal posts with
weather protection cover diameters of up to 60 mm. Order No.: 50062121
Weather protection cover CYY 101 Round post fixture for CYY 101
Endress+Hauser 27
TopClean S CPC 30
Housing CYC 300 Housing for TopClean S CPC 30, with rack for buffer and cleaner solutions. Operating panel with
alarm LED and lock for programme start and assembly drive. Window for Mycom S or Memo-
Graph S. For Ex and Non-Ex applications. Material: plastic or stainless steel.
C07-CYC300xx-09-00-06-xx-001.EPS
➃ 9 Operating panel
(Order operating panel separately:
Order No.: 51512891)
➄
➅
840
936
950
40
28 Endress+Hauser
TopClean S CPC 30
Product structure
Housing CYC 300
Certificates
A no approval
G with ATEX approval II (1) 2G EEx em ib[ia] IIC
O with FM approval Cl. I, Div. 2, with NI input and output circuits, sensor IS Cl. I Div. 1
P with FM approval CI. I, Div. 2, with NI input and output circuits
S with CSA approval Cl. I, Div. 2, with NI input and output circuits, sensor IS Cl. I Div. 1
T with TIIS approval
Power supply
1 230 V AC
2 110-115 V AC
3 24 V AC / DC
Materials
A Plastic
G Stainless steel 1.4301 (AISI 304)
Heating
1 no electrical heating
2 with electrical heating
Data recording
A no MemoGraph S
B with MemoGraph S
Allocation
1 empty housing, CPC 30 not mounted
2 order item of associated CPC
Options
1 Basic version
Endress+Hauser 29
TopClean S CPC 30
Supplementary documentation
Supplementary ❑ Operating Instructions TopClean S CPC 30, BA 235C/07/en, Order No. 51504339
documentation ❑ Operating Instructions PROFIBUS-PA, BA 298C/07/en, Order No. 51507116
❑ Operating Instructions HART, BA 301C/07/en, Order No. 51507114
❑ Ex Safety Instructions, XA 236C/07/a3, Order No. 51506729
❑ Technical Information Mycom S CPM 153, TI 233C/07/en, Order No. 51503788
❑ Technical Information CleanFit CPA 471, TI 217C/07/en, Order No. 51502595
❑ Technical Information CleanFit CPA 472, TI 223C/07/en, Order No. 51502644
❑ Technical Information CleanFit CPA 473, TI 344C/07/en, Order No. 51510923
❑ Technical Information CleanFit CPA 474, TI 345C/07/en, Order No. 51510925
❑ Technical Information CleanFit CPA 475, TI 240C/07/en, Order No. 51505598
❑ Technical Information OrbiSint CPS 11/12/13, TI 028C/07/en, Order No. 50052557
❑ Technical Information CeraLiquid CPS 41/42/43, TI 079C/07/en, Order No. 50058726
❑ Technical Information CeraGel CPS 71, TI 245C/07/en, Order No. 51505837
❑ Technical Information OrbiPore CPS 91, TI 375C/07/en, Order No. 51513127
❑ Technical Information TopHit CPS 471, TI 283C/07/en, Order No. 51506687
❑ Technical Information CPK 1-12, TI 118C/07/en, Order No. 50068525
❑ Technical Information CPR 40, TI 342C/07/en, Order No. 51510059
❑ Technical Information CYR 10 / 20, TI 046C/07/en, Order No. 50014223
30 Endress+Hauser
TopClean S CPC 30
Endress+Hauser 31
Subject to modification
Endress+Hauser
Gmbh+Co.
Instruments International
P.O. Box 2222
D-79574 Weil am Rhein
Germany
Internet:
www.endress.com
TI 235C/07/en/04.03
51504335
FM+SGML 6.0 / DT