Labtec PRO User Manual Overview
Labtec PRO User Manual Overview
User Manual
Labtec PRO
No Content Page
1 foreword 2
2 safety instructions 2
3 device description 3
4 technical data 4
5 component description 5
6 operating elements 6
7 commissioning 7
8 care and maintenance 7
Attachments: Perform tests manually
Labtec PRO are electronic test devices to A 230-240 V / 50-60 Hz power supply is re-
be operated via a PC for the testing of brea- quired to operate the test stand.
thing masks according to DIN EN 136, self-
contained breathing apparatus (SCBA) ac- The high pressure supply required for opera-
cording to DIN EN 137 and chemical protec- tion is provided by a 300 or 200 bar brea-
tive suits (CPS) according to DIN EN 943-1. thing air cylinder or by a central air supply
with 300 bar.
The tests are performed in the low (-50 to 50
mbar), medium (0 to 20 bar) and high pres- As medium pressure connection Euro-
sure measuring range (0 to 300 bar). coupling and nipple are available as stan-
dard. If units with other coupling connections
For testing other breathing apparatus or che- are tested, appropriate intermediate adap-
mical protective suits, contact Labtec. Only ters must be used. Please contact the manu-
breathing air according to DIN EN 12021 facturer of the units for this.
may be used.
Labtec PRO are fully automatic, "high-end" The scope of delivery of the device includes
test devices without any restrictions in test a high-power-connection cable incl. screw
operation in the low, medium and high pres- fitting, microphone cable (only PRO), power
sure range. supply cable and a USB cable.
Labtec Pro light are automatic test devices Adapters for the connection of CPS, regula-
incl. artificial variable lung for static and dy- tors and sealing caps for respirators must be
namic testing of respirators during breathing. ordered separately depending on the
In contrast to Labtec PRO they do not have connection used (e.g. ESA, PE45, RD40,
the possibility of automatic sealing of de- manufacturer-specific connections).
mand valves, no microphone for automatic
warning signal detection, no pneumatic re- Labtec offers a wide range of accessories for
dundancy and no possibility of SCBA safety test devices.
valve testing by medium pressure increase.
The range includes test adapters for adap-
Compared to Labtec PRO light, Labtec PRO ting all respiratory protection devices on the
static testers do not have an artificial variab- market, quick release devices for quick de-
le lung and are therefore only suitable for vice adaptation as well as holders for holding
static testing of breathing apparatus. helmet-mask combinations.
The devices have several expansion options Further information and brochures
(modules) which are identified by a letter can be found at www.labtec.de
code:
IP rating IP 54
A
I
C
B
K
D J
Switch button for Artifical Lung (main valve has to be open, Only Labtec PRO + PRO light)
When pressed, the artificial lung is switched on and ventilated through the mouth opening of the
test head. Pressing button again switches it off.
Establish high-pressure connection The demand valve is inserted into the mouth
opening of the test head using a suitable
Connect the high-pressure line to the air supp- adapter.
ly connection of the test device.
Connect the power supply to a 230 V / 50-60 The test instrument should be protected from
Hz power source using the supplied power external harmful influences such as water, mo-
cable. Switch on the unit at the switch. isture in the form of steam and dust.
Establish connection to the PC The test head should be cleaned and maintai-
ned regularly. This can significantly increase
Install the LabNet Profi test software accord- the service life of the search unit.
ing to the installation instructions. Establish the
connection between PC and test bench via The test head should always be covered with
USB cable. the enclosed protective hood after the instru-
ment has been used.
> See separate installation and operating
instructions LabNet Profi > See separate test head care instructions
Device connection (breathing mask) The test device must be serviced at least an-
nually by a certified Labtec service technician.
The test head must be completely vented. The
breathing connection of the mask is closed The test device is recalibrated and deviations
with a suitable sealing cap. Alternatively, in the previous test cycle can be excluded.
connect the mask to a suitable demand valve The maintenance is carried out with a special
and then make a device connection (SCBA). calibration device.
The mask is put on the test head. The inner The maintenance is confirmed by a calibration
mask should rest correctly on the nose of the certificate issued directly by Labtec. The main-
probe. tenance may only be carried out by authorized
and trained service technicians.
Then tighten the fasteners of the head mask
starting from the bottom. The test head can In order to ensure a cost-effective and timely
now be inflated until a tight fit of the mask is maintenance, it is possible to conclude a main-
ensured. If necessary, readjust the mask. tenance contract with Labtec.
Konformitätserklärung
Certificate of Conformity
Attestation de Conformité
We do hereby certify that the above mentioned product meets the requirements set
forth in EEC-Guideline 2016/42/EG and EMC 2014/30/EU including all changes and
addendums to date thereto. The above mentioned product meets the following
standards and guidelines:
Nous Vous Confirmons que l’appareil cité ci-dessus correspond aux exigences des
directives 2016/42/EG ainsi qu’ à la directive CEM 2014/30/EU ci-inclus toues les
modifications ainsi que tous les suppléments publiés jusqu’à ce jour. L’appareil
mentionné correspond aux nomes cités ci-après :
operating
Step Activity Comment
elements
Close breathing connection to mask with sealing If necessary, moisten the exhalation
1.1.1
plug. valve of the mask.
Place the mask on the test head and tighten the
The inflation time varies from mask
1.1.2 mask strap. Fill test head with head inflation
type to mask type.
button
Set test vacuum with negative pressure button Adjust the test pressure with the low
1.1.3
to setpoint value - # mbar. Switch on stop watch. pressure relief button if necessary.
Seal the exhalation valve of the mask with the Test only possible with tightly fitted
1.2.1 adapter. Close breathing connection to mask exhalation valve. Use the manufac-
with sealing plug. turer's specific adapters.
Place the mask on the test head and tighten the
The inflation time varies from mask
1.2.2 mask strap. Fill test head with head inflation
type to mask type.
button
Set test vacuum with negative pressure button Adjust the test pressure with the low
1.2.3
to setpoint value - # mbar. Switch on stop watch. pressure relief button if necessary.
operating
Step Activity Comment
elements
Insert the test adapter into the mouth opening of
Select suitable adapter according to
2.1.1 the test head and connect it to the demand val-
function and connection type
ve.
Seal the medium pressure hose of the demand
2.1.2 valve with a sealing cap or connect it to the me-
dium pressure connection on the device.
Set test negative pressure with button to gener-
Adjust the test pressure with the low
2.1.3 ate negative pressure to setpoint + # mbar.
pressure relief button if necessary.
Switch on stop watch.
Set test overpressure with button to generate
Adjust the test pressure with the low
opt. positive pressure to setpoint + # mbar. Switch on
pressure relief button if necessary.
stop watch.
After a preset time # min, read the pressure
2.1.4 change on the display (LP) and compare with
the setpoint.
All further tests require high pressure! Check correct device connection!
The Labtec test bench + connected test device parts must be completely vented before
connecting or removing test devices.
operating
Step Activity Comment
elements
Connect demand valve in combination with a
2.3.1 Check correct device connection!
SCBA base unit to the test device.
Open high pressure main valve with main valve High pressure flows into the unit and
2.3.2
switch. is present in the unit.
until the regulator opens and the
Switch on vacuum pump with switch 10l/min
2.3.3 maximum negative pressure point is
Inhalation to produce constant flow.
reached
Read the opening / closing pressure of the de-
2.3.4
mand valve on the display (LP).
Switch off the pump by pressing the 10/lmin In-
2.3.5
halation switch again.
Close high pressure main valve with main valve
2.3.6
switch.
operating
Step Activity Comment
elements
Connect demand valve in combination with a
3.1.1 Check correct device connection!
SCBA base unit to the test device.
Open high pressure main valve with main valve
3.1.2 High pressure flows into the unit.
switch.
Close high pressure main valve with main valve Test instrument/test specimen are
3.1.3
switch. under pressure.
Vent high pressure with button for high pressure
relief (fast) or medium pressure relief (slower) to
3.1.4 Read off the SCBA manometer
test pressure until PA pressure gauge shows the
set point.
Read the high pressure on the display (HP) and
3.1.5
compare it with instrument pressure.
operating
Step Activity Comment
elements
Connect demand valve in combination with a
5.1.1 Check correct device connection!
SCBA base unit to the test device.
Open high pressure main valve with main valve
5.1.2 High pressure flows into the unit.
switch.
Close high pressure main valve with main valve Test instrument/test specimen are
5.1.3
switch. under pressure.
Vent high pressure with button for high pressure
5.1.4 relief (fast) or medium pressure relief (slower) to
to setpoint + # bar inlet pressure
Read the medium pressure on the display (MP)
5.1.5
and compare with setpoint.
6 Complete Device
6.1 High pressure tightness test
operating
Step Activity Comment
elements
Connect demand valve in combination with a
5.3.1 Check correct device connection!
SCBA base unit to the test device.
Open high pressure main valve with main valve High pressure flows into the unit and
5.3.2
switch. is present in the unit.
8 Valves
8.1 Tightness of the suit valves after # min. at - # mbar
(repeat the test depending on the number of valves)
operating
Step Activity Comment
elements
Close the high-pressure device connection with Device connection must be closed
7.1.1
a plug. during tightness test.
Connect one tightening valve with test adapter, CPS test adapter CGA 200 and brand
7.1.2
seal all other tightening valves with plugs. -specific adapters
Couple the filling hose of the test adapter with
7.1.3 CPS test adapter CGA 200
plug nipple of CPS test connector.
Insert the measuring hose of the test adapter
7.1.4 CPS test adapter CGA 200
into the mouth of the test head.
Open high pressure main valve with main valve High pressure flows into the unit and
7.1.5
switch. is present in the unit.
7.1.6 Fill CPS by pressing the CPS inflation switch CPS is filled to +# mbar
Durch erneutes Drücken des Schalters CSA when the required tightening internal
7.1.7
füllen das Befüllen des Anzugs stoppen. pressure is reached.
manufacturer-dependent adapter,
Connect tightening valve with valve test adapter
8.1.2 depending on the number of
from inside
existing valves.
Set test vacuum with negative pressure button to Adjust the test pressure with the low
8.1.3
setpoint value - # mbar. Switch on stop watch. pressure relief button if necessary.