Calibration Certificate for 34972A Unit

This certificate documents the calibration of a data acquisition/switch unit and its temperature sensors. The unit was calibrated by comparison to standard platinum resistance thermometers in a liquid bath. The results show the temperature readings from the unit compared to the standard at various temperature points, along with any measurement corrections. The calibration establishes traceability to national measurement standards.

Uploaded by

soeminhtoo7792
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
39 views6 pages

Calibration Certificate for 34972A Unit

This certificate documents the calibration of a data acquisition/switch unit and its temperature sensors. The unit was calibrated by comparison to standard platinum resistance thermometers in a liquid bath. The results show the temperature readings from the unit compared to the standard at various temperature points, along with any measurement corrections. The calibration establishes traceability to national measurement standards.

Uploaded by

soeminhtoo7792
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd

CERTIFICATE OF CALIBRATION

Date of Issue : 07 January 2023 Certificate Number : QR20-2284


Order Item No.: 2012068 Page : 1 of 6

www.qreborn.com

Quality Reborn Co.,Ltd.


42/267 Leab Klong pasricharoen fangnue 8/1
Nongkham, Bangkok 10160 NSC-TISI-TIS17025
Tel : +662-4447-382, Fax: +662-4447-383 CALIBRATION 0292

Customer: DKSH (Myanmar) Ltd


No.13, Thitsar Road, Yankin Township, Yangon, Myanmar.
Date Received: 27 December 2022
Date of Calibration: 28 December 2022
Due of Calibration: 28 December 2024
Instrument: Description: Data Acquisition/Switch Unit with Sensor
Model: 34972A
Serial Number: MY59002090
ID Number: -
Manufacturer: Keysight
Site: Quality Reborn Office
Location: Calibration Room 1
Environmental Conditions
Temperature: 25 ̊C ± 3 ̊C
Relative Humidity: 55% ± 25 %
Calibration Method Used
This instrument was calibrated by comparison of indication with the Standard Platinum
Resistance Thermometer in liquid calibration bath according to calibration procedure no. CP-T01-01.
Traceability of Measurement
This certificate of calibration documents the traceability to national standards, which
realize the units of measurement according to the International System of Units (SI).
Calibrated By: Mr.Wutthichai Rujiwut Approved By: .

[ ]Mr. Thanat Sutthinate

[ ]Mr. Jatuporn Juijai-ngam

This certificate may not be reproduced other than in full except with the prior written approval of Quality Reborn Co.,Ltd
Certificate No. : QR20-2284
www.qreborn.com
Order Item No.: 2012058
Page : 2 of 6

Details of Calibration
1. Reference Standard Equipment Used
Description Certificate No. Due Date
Platinum Resistance Thermometer “Accu Mac” S/N 1621951 91037 03-April 2023
Reference Thermometer “Isotech” milliK S/N39622-1 C20/00002 08-January 2023
2. The results reported in this certificate refer to the condition of the instrument on the date of calibration
and carry no implication regarding the long-term stability of the instrument.
3. Condition of calibration item: normal condition, no indication find for any damage or malfunction
4. Immersion Depth of Calibration = 150 mm.
Result of Calibration: (Without Adjustment)
Function: Temperature Measurement of Slot 100 S/N MY60000817
Description of Sensor: RTD Model:Pt100 Diameter 1.6mm Length: 200mm
Ch: 101 SensorS/N. 5002915-001 Parameter Setting: Gain=1.0 Offset=0.0000
Standard Temperature UUC* Temperature Uncertainty of
(C̊ ) Reading ( C
̊ ) Correction ( C
̊ ) Measurement (± ̊C)
-39.9953 -40.10 0.1047 0.040
-20.0098 -20.05 0.0402 0.040
-0.0088 0.02 -0.0288 0.040
500084 50.20 -0.0181 0.040
99.9917 100.33 -0.3383 0.040

Ch: 102 SensorS/N.: 5002915-002 Parameter Setting: Gain=1.0 Offset=0.0000


Standard Temperature UUC* Temperature Uncertainty of
(C̊ ) Reading ( C
̊ ) Correction ( C
̊ ) Measurement (± ̊C)
-39.9953 -40.10 0.1147 0.040
-20.0098 -20.05 0.0402 0.040
-0.0088 0.03 -0.0388 0.040
50.0084 50.21 -0.2016 0.040
99.9917 100.34 -0.3483 0.040

Ch: 103 SensorS/N.: 5002915-003 Parameter Setting: Gain=1.0 Offset=0.0000


Standard Temperature UUC* Temperature Uncertainty of
(C̊ ) Reading ( C
̊ ) Correction ( C
̊ ) Measurement (± ̊C)
-39.9953 -40.09 0.0947 0.040
-20.0098 -20.03 0.0202 0.040
-0.0088 0.04 -0.0488 0.040
50.0084 50.18 -0.1716 0.040
99.9917 100.26 -0.2683 0.040

UUC*: Unit Under Calibration


Certificate No. : QR20-2284
www.qreborn.com
Order Item No.: 2012058
Page : 3 of 6

Ch: 104 SensorS/N. 5002915-004 Parameter Setting: Gain=1.0 Offset=0.0000

Standard Temperature UUC* Temperature Uncertainty of


(C̊ ) Reading ( C
̊ ) Correction ( C
̊ ) Measurement (± ̊C)
-39.9953 -40.10 0.1047 0.040
-20.0098 -20.03 0.0202 0.040
-0.0088 0.04 -0.0488 0.040
500084 50.18 -0.1716 0.040
99.9917 100.28 -0.2883 0.040

Ch: 105 SensorS/N.: 5002915-005 Parameter Setting: Gain=1.0 Offset=0.0000


Standard Temperature UUC* Temperature Uncertainty of
(C̊ ) Reading ( C
̊ ) Correction ( C
̊ ) Measurement (± ̊C)
-39.9953 -40.10 0.1147 0.040
-20.0098 -20.05 0.0402 0.040
-0.0088 0.03 -0.0388 0.040
50.0084 50.21 -0.2016 0.040
99.9917 100.34 -0.3483 0.040
Ch: 103 SensorS/N.: 5002915-002 Parameter Setting: Gain=1.0 Offset=0.0000
Standard Temperature UUC* Temperature Uncertainty of
(C̊ ) Reading ( C
̊ ) Correction ( C
̊ ) Measurement (± ̊C)
-39.9953 -40.10 0.1047 0.040
-20.0098 -20.04 0.0302 0.040
-0.0088 0.03 -0.0388 0.040
50.0084 50.19 -0.1816 0.040
99.9917 100.30 -0.3083 0.040

UUC*: Unit Under Calibration


Certificate No. : QR20-2284
www.qreborn.com
Order Item No.: 2012058
Page : 4 of 6

Result of Calibration: ( Without Adjustment)


Function: Temperature Measurement of Slot 200 S/N MY60001201
Description of Sensor: RTD Model: Pt 100 Diameter 3.2 mm Length: 50 mm
Ch: 201 SensorS/N. 1 Parameter Setting: Gain=1.0 Offset=0.0000

Standard Temperature UUC* Temperature Uncertainty of


(C̊ ) Reading ( C
̊ ) Correction ( C
̊ ) Measurement (± ̊C)
-39.9954 -40.25 0.2553 0.040
-20.0017 -20.22 0.2183 0.040
-0.0105 -0.19 0.1795 0.040
99.9949 99.89 0.1049 0.040
200.0061 199.65 0.3561 0.040

Ch:202 SensorS/N.: 2 Parameter Setting: Gain=1.0 Offset=0.0000

Standard Temperature UUC* Temperature Uncertainty of


(C̊ ) Reading ( C
̊ ) Correction ( C
̊ ) Measurement (± ̊C)
-39.9947 -40.23 0.2353 0.040
-20.0017 -20.17 0.1683 0.040
-0.0105 -0.12 0.1095 0.040
99.9949 99.98 0.0149 0.040
200.0061 199.78 0.2261 0.040

Ch: 203 SensorS/N.:3 Parameter Setting: Gain=1.0 Offset=0.0000


Standard Temperature UUC* Temperature Uncertainty of
(C̊ ) Reading ( C
̊ ) Correction ( C
̊ ) Measurement (± ̊C)
-39.9947 -40.24 0.2453 0.040
-20.0017 -20.20 0.1983 0.040
-0.0105 -0.17 0.1595 0.040
99.9949 99.87 0.1249 0.040
200.0061 199.62 0.3861 0.040

Ch: 204 SensorS/N.:4 Parameter Setting: Gain=1.0 Offset=0.0000

Standard Temperature UUC* Temperature Uncertainty of


(C̊ ) Reading ( C
̊ ) Correction ( C
̊ ) Measurement (± ̊C)
-39.9947 -40.22 0.2253 0.040
-20.0017 -20.20 0.1983 0.040
-0.0105 -0.18 0.1695 0.040
99.9949 99.88 0.1149 0.040
200.0061 199.63 0.3761 0.040

UUC*: Unit Under Calibration


Certificate No. : QR20-2284
www.qreborn.com
Order Item No.: 2012058
Page : 5 of 6

Ch: 205 SensorS/N. 5 Parameter Setting: Gain=1.0 Offset=0.0000

Standard Temperature UUC* Temperature Uncertainty of


(C̊ ) Reading ( C
̊ ) Correction ( C
̊ ) Measurement (± ̊C)
-39.9954 -40.26 0.2653 0.040
-20.0017 -20.23 0.2283 0.040
-0.0105 -0.20 0.1895 0.040
99.9949 99.86 0.1349 0.040
200.0061 199.64 0.3661 0.040

Ch:206 SensorS/N.: 6 Parameter Setting: Gain=1.0 Offset=0.0000


Standard Temperature UUC* Temperature Uncertainty of
(C̊ ) Reading ( C
̊ ) Correction ( C
̊ ) Measurement (± ̊C)
-39.9947 -40.20 0.2053 0.040
-20.0017 -20.18 0.1783 0.040
-0.0105 -0.15 0.1395 0.040
99.9949 99.94 0.0549 0.040
200.0061 199.73 0.2761 0.040

Ch: 207 SensorS/N.:7 Parameter Setting: Gain=1.0 Offset=0.0000


Standard Temperature UUC* Temperature Uncertainty of
(C̊ ) Reading ( C
̊ ) Correction ( C
̊ ) Measurement (± ̊C)
-39.9947 -40.24 0.2453 0.040
-20.0017 -20.18 0.1783 0.040
-0.0105 -0.12 0.1095 0.040
99.9949 99.99 0.0049 0.040
200.0061 199.81 0.1961 0.040

Ch: 208 SensorS/N.:8 Parameter Setting: Gain=1.0 Offset=0.0000


Standard Temperature UUC* Temperature Uncertainty of
(C̊ ) Reading ( C
̊ ) Correction ( C
̊ ) Measurement (± ̊C)
-39.9947 -40.28 0.2853 0.040
-20.0017 -20.24 0.2383 0.040
-0.0105 -0.18 0.1695 0.040
99.9949 99.89 0.1049 0.040
200.0061 199.67 0.3361 0.040

UUC*: Unit Under Calibration


Certificate No. : QR20-2284
www.qreborn.com
Order Item No.: 2012058
Page : 6 of 6

Ch: 209 SensorS/No.9 Parameter Setting: Gain=1.0 Offset=0.0000

Standard Temperature UUC* Temperature Uncertainty of


(C̊ ) Reading ( C
̊ ) Correction ( C
̊ ) Measurement (± ̊C)
-39.9947 -40.24 0.2453 0.040
-20.0017 -20.21 0.2083 0.040
-0.0105 -0.17 0.1595 0.040
99.9949 99.90 0.0949 0.040
200.0061 199.66 0.3461 0.040

Ch:210 SensorS/N.: 10 Parameter Setting: Gain=1.0 Offset=0.0000


Standard Temperature UUC* Temperature Uncertainty of
(C̊ ) Reading ( C
̊ ) Correction ( C
̊ ) Measurement (± ̊C)
-39.9947 -40.21 0.2153 0.040
-20.0017 -20.16 0.1583 0.040
-0.0105 -0.12 0.1095 0.040
99.9949 99.99 0.0049 0.040
200.0061 199.80 0.2061 0.040

UUC*: Unit Under Calibration


This reported uncertainty of measurement was based on a standard uncertainty multiplied by a
coverage factor k=2, providing a level of confidence of approximately 95%.