Quality Control in Pathology Labs
Quality Control in Pathology Labs
INTERNAL QUALITY CONTROoL 1 Test ordering and utilization: Sources of error may
Through internal quality control, the laboratory can be introduced due to inappropriate tests, bad
ensure that the results being issued by laboratory are handwriting (which may not be legible), wrong
reliable Internal quality control is the study of errors patient identification and due to delayed order of
and introduction_of procedures to récognize and testsand inappropriate control of 'stat' tests.With
minimize them The errors include all those arise within proper measures, sources of error arising due to
theTaboratorybetween the teceipt of the specimen and above-mentioned aspects can be controlled.
dispatch of the report./A laboratory that meets quality 2 Patient preparation: Laboratory tests are affected
requirements has fewer reruns and complaints, and this by many factors such as recent intake of food, al-
saves money Meeting quality requirements satisfies cohol, drug etc., food intake (3 days prior to test).
clinicians (users of the laboratory), giving the laboratory Smoking, exercise, stress, sleep and posture dur-
a greater competitive edge and providing job security ing specimen collection also can affect laboratory
for the senior staff. It also leads to better and more tests. Laboratory management must define the in-
efficient patient care. structions and procedures for patient preparation.
These procedures should be included in hospital
For internal quality control, it is necessary to understand
the following
procedure manuals and should be transmitted to
requirements of clinicians patients in both oral and written instructions and
1 An adequate test menu to meet clinical decision- Compliance with these instructions can be moni-
making needs. tored by the phlebotomist.
2 Information and education about the tests that are 3 Specimen collection, acquisition and patient iden-
appropriate for a particular clinical problem, tification: Various problems can arise due to in-
3 Correct tube (or container), incorrect order of
Instructions on preparing the patient prior to
specimen collection and on the appropriate vacutainers, incorrect anticoagulant, incorrect
specimen for an assay. patient identification, inadequate volume, hemo-
4 Phlebotomy performed safely and efficiently without yzed, lipemia or icteric sera, specimens collected
at wrong time and due to improper transport
166 TEXTBOOK OF MEDICAL LABORATORY
TECHNOLOGY
containers
conditions. Prolonged tourniquet in making plastic
application the plasticizers used materials
causes local anoxia to cells and
excessive venous interfere with drug
analysis. Some plastic such as
back pressure. The anoxia causes small solutes amounts of analyses
some
adsorb trace are not
(such as potassium) to leak from cells and the
parathyroid
hormone. Glass Containers
and trace
venous pressure concentrates
cells, proteins, and suitable for the determination
ofcalcium
substances bound to proteins (such as
Blood collected from an arm into which calcium). elementS.
an
venous infusion is running can be diluted or
intra
Variables
taminated. con Control of Analytical
to
variables, it is necessary
The greatest For the control of analytical
promise for correct identification of terms:
understand the following
patient specimens is the introduction of the laser between the
beam wand, which can read bar code or optical Trueness is closeness of
agreement
series of test
characters from labels. Various measures are obtained from a large
average value
suggested in chapter No. 7 to prevent hemolysis of reference value.
results and an accepted
sera and use of between a
specimen blanks or dichromatic of agreement
Accuracy is the degree Statistically it is
readings to tackte problems arising due to measured valyeand its true value^
hemolyzed, icteric or lipemia sera. which is defined as
represented by analytical bias,
between measured
the percentage of the difference
Note method gives
value and true value. An accurate
1 As much as 84 percent of laboratory errors can be correct result.
to the
attributed to the preanalytical phase of specimen Precision refers to reproducibility. It refers
measurements: It is
testing. agreement between replicate
Standard Deviation
2 A quality test result always starts with a quality quantitatively expressed as
or more precisely as Coefficient of
Variation
sample. (S.D.)
(C.V.) of the results in a set of replicate
3 High sensitivity assays that measure-smallamounts measurements. Hence good precision means least
such as TSH, troponin,hCG, troponin are more C.V. A precise method gives same results, if
Susceptible to errors
repeated.
4 Citical care staff such as phlebotomists, nurses A sensitive method measures low concentration
and collection centre technicians should be well of analytes. Sensitivity is a measure of the incidence
trained for coFrect specimen collection, separation of positive results in patients known to have a
of serum/plasma, preservation and transportation
condition, thatis True Positive (TP). A sensitivity of
ofspecimen. 90% implies that only90% of people known to have
5 inadequate clotting time leads to formation of the disease would be diagnosed as having it on the
micro-clots in the specimen and instrument probe basis of that test alone, 10% would be False
obstruction. Negative (FN).
Inadequate mixing of anticoagulated sample in the TP
tube also leads to formation' of microclots, latent Sensitivity = x 100
All with disease + FN
fibrin formation and platelet clumping. This may
lead to decrease in hematocrit, hemoglobin and A specific method is not subject to interference
also sample probe obstruction by other substances. Specificity is a measure of
theincidente of negative results in persons known
Inadequate spinning of sample tube leads to to be free of disease, that is
leakage of red blood cetts, which leads to *True Negative' (TN).
A specificity of 90% implies
contamination of serum with potassium, inorganic that
free people would be classified as
10% of disease-,
phosphorus, LDH and SGOT. on the basis of
having the diseasee
arrive in the
test result; 10% would have a 'False
8 Laboratory logs: When the specimens Positive' result.
laboratory, various logging and monitoring systems
TN
are necessary. Specificity =
All with 100
9 Specimen transport: The stability of specimen disease + TN
during transport from the patient to the laboratory
should be monitored carefully. Particularly for Note
specimen collected at collection centers and those Factors which increase the specificity of a test tend
sent to specialized laboratory should be transported to
decrease the sensitivity and vice versa, as there is
at 2-8°C. almost always an overlap between test results seen in
Collection tubes,
10 Careful specimen separation: health and in disease.
tubes are sources of
pipettes and aliquot determinations. Some of
contamination for many Whether to maximize specificity or
sensitivity depends
CHAPIER6 TOTAL QUALITY MANAGEMENT 167
A Number
of B Number
results Method 1
of
Method 3 Method 4
results
Method 2
Mean Test result True Mean Test result
value value value
Method 1 is
precise compared to method 2 Accuracy
Methods 3 and 4 are equally precise but method
4 does not offertrue value
Fig. 6.1A & B: Precision and
accuracy
on the nature of the condition.
For example,
is very important in a sensitivity and issued by an organization are accepted as
screening test for
condition, since it is possible to a harmful
The aim of the validation of test and calibration influence each should be and may vary
methods than it
be higher or lower
is to demonstrate that the method is fit for the little in magnitude.
intended Considerably or
purpose and that the results have an acceptable
reliability. could occur due to wrong
errors
Systematic analytical
and incorrect
procedure, incorrect standard
Sources of Variation in Analytical Phase of Laboratory
Tests standardization procedure.
Error is a difference between a single result estimate Note
of a quantity and its true value. This difference or de- 1 Determinate (systematic) error (Fig. 6.2A & B)
viation (positive or negative) may be expressed either
in the units in which the quantity is measured or as a can be traced to a specific cause. Common causes
of inaccuracy include constant or proportional
percentage of the true value. Many kinds of error may
errors. Constant systematic error is an error that
skew the findings of laboratory tests. These include
random error, random variation, constant and propor- is always in the same direction and of the same
tional determinate (systematic) error, and methodologi- magnitude even as the concentration of the analyte
cal or physiologic interference. changes. Proportional systematic error is an error
that always occurs in one direction.
Analytical phase can be affected by various types of A classic example of systematic error is a method
errors, imprecision and interference of certain related error caused by the
Components in the sample such as drugs. presence of interfering
substances in the specimen These interfering
substances may be endogenous metabolites, such
Analytical Errors: The errors are classified into random asbilirubin, chylomicrons, drugs-received by the
analytical errors and systematic analytical errors.
Random analytical errors indicate poor precision and patient or excess of anticoagulant
systematic errors indicate poor accuracy. (Refer to Drug interference may be methodologic (in vitro)
figures 6.10-6.20). or
physiologic (in vivo) in
interference occurs when origin. Methodology
the drug is actually
Random analytical errors include pipetting error, detected during measurement as if it were the
transcription error, wrong sample numbering and substance of interest. Physiologic interference,
A
40
40+
Constatemor-
301
30
Ne
o ro
Proportionalerror
20
20
10
10-
(0,0) 0 20 0 40
(0,0) 10 20
Reference values 30 40
Reference values
Fig. 6.2A & B: (A) Systematic errors and (B) Random errors.
MANAGEMENT 169
CHAPTER 6: TOTAL OUALITY
actually represents biological variation when a drug9 average value. The formula
for finding the S.D. is,
received by the patient induces an increase in the
substance of interest being measured. (K-x)
3 Random and systematic errors can be S.D.
detected by n-1
using quality control charts such as "Levey-Jenning
and Cusum charts (Figs. = Sum of
6.10-6.22). x Averagevalue
X = Any single observed value
Postanalytic Errors
n = Total number of observed values
These are mainly transcriptional errors. To avoid these
errors in the results of the test, the reporting
or signatory Note
analyst should verify the results entered manually or control chart, the first
In order to prepare a quality
through on line instrument interfaces before the results the S.D. for the procedure
are reported or dispatched. thing to do so is to determine
it is necessary to
to be controlled. In a practical basis,
At the end of the month,
run a control serum each day.
FORMULATING QUALITY CONTROL CHARTS at least 20 values of various analytes
can be obtained
which can be used for the preparation
of Levey-Jenning
Following are the additional "Definitions" needed in
formulating Quality Control charts such as "Levey- chart for each analyte.
Jenning and Cusum charts:
EXPERIMENT NO. 6.1
creatinine, SGPT, SGOT, etc.), that is tested in the same 2 Glassware to perform tests
tested in order to ensure that
way patient samples are 3 Equipments (e. g. push button pipettes,
water bath,
This
methodology of a laboratory is working accurately. centrifuge, photometer or spectrophometer, etc.).
is a sample (control serum) that is chemically and 4 Diagnostic kits (for the determination of glucose,
physically similar to the unknown specimen. urea nitrogen, creatinine, etc.).
in the name of 5 Pencils, rulers and graph papers
Analyte: It is a component represented
a measurable quantity (i. e.
creatinine in serum, glucose
in plasma, or, urea in control, etc.). Procedure
1 Use a control sample for various tests such
serum
recorded values
Median: A middle value of a set of as glucose, urea, creatinine, uric acid, SGPT,
etc.
above as below (Refer
with an equal number of values every day.
to L-J chart, Fig. 6.4). 2 Note control values for various tests every day.
from the 3 At the end of the 3 weeks, at least 20 values of
Drift: The condition of values moving away various tests will be obtained.
mean line over a period of
time (Refer to L-J chart,
4 Prepare "Levey-Jenning chart for every test
Fig.6.6). to following "calculation procedure"
according
values suddenly move (Table 6.1).
Shift: The situation where control
from clustering around the mean
line to a line above or 5 Example- Test: Glucose. In the column "A" of
Table
below the median (Refer to L-J chart, Fig. 6.7).
listed.
6.1, 20 values of control glucose are
value
i. e. a b Calculate average control serum glucose
Bias: It is a measure of degree of trueness, (Equation 1: It is 100 mg/dl).
measure of systematic error (Fig. 6.14 to the "calculation
7 Prepare column "B" according
from procedure" in Table 6.1.
outliers: Results which appreciably different
are
the "calculation
the general cluster of values (Fig. 6.12). 8 Prepare column "C" according to
procedure" in Table 6.1.
of the dispersion of
randem 8 in Table
Imprecision: A measure 9 Prepare "Equation 2" according to step
errors. 6.1
10 Calculate "Standard Deviation" (1SD,
2SD, 3SD)
Standard deviation: This a statistical expression values using following formula:
around a central
of the scatter or dispersion of values
172 TEXTBOOK OF MEDICAL LABORATORY TECHNOLOGY
S.D. 3R 240
CV 100 2P 160
mean (x) 1R
Example
1 Glucose determination: Method: Folin-Wu.
If = 98 mg/dl
S.D. = t 2.5
2.5 x 1000
C.
98
2.
2 Glucose determination: method: Glucose oxidase
If 78 mg %
=
S. D. = 0.9 mg %
0.9 x100
C.V.=
78
1.2%
Fig. 6.10: Display of quality control charts (L-J) on
autoanalyzer screen
Interpretation
Thus the variation inherent in the reducing sugar method IMPORTANCE OF REFERENCE RANGE
(Folin-Wu) is more than twice as great as that in the (NORMAL
glucose oxidase method. VALUES)
The college of American
standards for their Pathologists, in setting
IMPORTANCE OF DETERMINATION OF % BIAS
inspection and Accreditation Program
for Clinical Laboratories, indicates
that reference values
Determination of percent bias of a control test result (i.e., normal values, the term used in the
test should be provided with past) for each
gives idea of any systematic error present in a particular mind two modern
the report by keeping in
batch analysis. An estimate of bias is the average value laboratory realities:
of a series of measurement minus a reference value. 1 A reference range is required for the
of
Percent Bias gives idea of percent degree systematic
formula:
of most laboratory tests, and
interpretation
calculated using following
error. It is 2 It may not be
Reported value X 100/ True value
possible to provide an
appropriate
% Bias True value
=
- reference range in all cases.
170 TEXTBOOK OF MEDICAL LABORATORY
TECHNOLOGY
t 6.0
Sum 2nd S.D.
=
100
90 1 5 6 7&9 10 11 12 13 14
80 Days
1 2 3 45 6 7 8 9 10 11 12 13 14 15 This type of shift is observed in the case
Fig. 6.8: Interpretation:
Days of reagent or standard deterioration of concentration at one point of
Mg/dl time and later on there was no rurtner deterioration in reagent or
Note: Average value of QC serum
glucose =
100 mg/d standard.
Standard deviation (S. D.) =+10
mg/d
Fig. 6.5: Interpretation: Random error is seen on
mistake committed by the laboratory technician. Afterday
4, due to
day 9 higher
detect systematic errors. When a systematic error
readings of glucose are observed. Jhis may be due to use of is present, the
decreased concentration (all low values or all high values) cusum
of
primary standard (may be due to values will steadily increase. At a certain state the error
deterioration) will be too large to be accepted and method will be
considered out of controlCusum technique generally
Excessive scatter provides better determination of systematic errors than
can be obtained by Levey-Jenning charts. A Cusum chart
can be prepared as follows (Refer to the glucose
120
determination values used for the Levey-Jenning chart
Table 6.1).
100
Fig. 6.6: Interpretation: Excessive scatter may be observed due to Calculation of Cusum
inconsistent ways of working of laboratory technician.
Refer to Table 6.2. The Cusum is calculated by adding
the values in column 4 (Difference), i.e.Cusum for first
Drift or trend in the control values day is (average value:100 individual test value, 98 =
2), Cusum for day 13.2.89 = +2 -1 = +1. Cusum for
day 15.2.89 = sum of +2, -1, -4, +3 = 0. The Cusum
120 for the next day can be calculated by adding the new
difference to the previous Cusum. The Cusum chart is
100 drawn by plotting days on X-axis and Cusum values on
Y-axis (refer to the graphic presentation).
1 2 3 4 5 67 89 10 11 12 13 14
+10
Fig. 6.7: Interpretation: In type of drift or trend in the control
values may be seen either due to deterioration of reagent or due to
increased concentration of primary
standard.
0.0
5 7 8 9 10
Cusum Charts (Fig. 6.9) Days
In order to check the accuracy of the analysis run and Drift or trend after the day 5
also the quality control sera, a Cusum chart is plotted
from the quality control sera results. Such chart can Fig. 6.9: Cusum chart.
MANAGEMENT 173
CHAPTER 6: TOTAL QUALITY
Control of Analytical Quality using Patient Data To be totally "successful" in a given category, a
Intra-laboratory duplicates: Samples can be divided laboratory must produce correct results on four out of
into two aliquots and analyzed and the duplicate five specimens for each of the
analytes in that category
can be used for control purposes. and have an over all score of at least 80% for three
consecutive challenges.
Delta Checks with previous test: Certain errors,
particularly errors in specimen identification, can
FIVE Q' FRAMEWORK
be detected by comparing many laboratory test
results with values obtained on previous specimens The "Five Q" framework also defines how
from the same patient (except some tests like blood be managed objectively using the quality can
"Scientific method or
and urine glucose and cardiac enzyme profile tests). the PDCA cycle, i.e., Plan, Do,
Check, ACT. By means of
Limit Checks: A patient's test result should be "Quality planning" program, Quality laboratory
reviewed to check that they are within the processes" can be
implemented and maintained at
physiological ranges compatible with life. The limit excellent level using "Quality control"
checks are helpful for detecting clerical errors such "Quality assessment' measures, defectsmeasures. By
in "Quality
as transposed digits or misplaced decimal points. processes" can be rectified and improved by "Quality
Mean of normal: Determination of mean of normal improvement" program and accordingly, the
values of various tests can be useful in the detection planning" is revised. "Quality
of systematic errors.
Quality planning
External Quality Control (QP)
Internal QC and External QC are complementary Quality improvement
activities. If internal QC is necessary for the daily Quality laboratory
(Q)
monitoring of the precision and accuracy of the analytical processes (QLP)
method, external QC assessment is important for
maintaining the long term accuracy of the analytical Quality assessment
(QA)
Quality control
methods. (QC)
182 TEXTBOOK OF MEDICAL LABORATORY
TECHNOLOG3Y member position in eac
ach
between
staff
in order to work with other relationship everyone involved
people to get things done. section. This
helps
It requires laboratory or command (authority)
people-oriented and task-oriented leaders the chain of
understand
who have well-developed
management skills. These to
skills include, prompt decision
making, good verbal and (Fig. 6.24).
written
Department chairman
communication, negotiation, sensitivity to the
need of
(Laboratory Directory)
others, and being visionary. Continuing
education provides ample opportunities to
acquire,
improve, reaffirm and maintain current knowledge and
skills in leadership and
management. Following are the Department manager
basic managerial responsibilities:
Offering quality leadership
Laboratory medicine Support
Personnel management Anatomical services
directors
pathology
Directing preparation of policy manual
Directing preparation of standard operation Central processing
procedures (SOPs) of management and laboratory Histology supervisor Chemistry Phlebotomy
Autopsy service Coagulation
procedures Hematology Clerical services
Laboratory information
Directing criteria-based job descriptions Immunology services
Microbiology
Managing recruitment and staffing Toxicology
Managing in-service and continuing education chart.
Fig. 6.24: Generalized organizational
Conducting staff meetings
Controlling personal records Good Management Skills and Personal Characteris-
Maintaining performance evaluation/appraisals tics
Maintaining discipline The personal characteristics and management skills of
Caring legislative and regulatory issues the laboratory manager determine the daily work
Financial management environment of the clinical laboratory. The summary
chart of CLIA88 personnel qualifications for a laboratory
Effective marketing
director for moderate complexity testing (refer to
Managing medico-legal concerns chapter 1) are as follows: