L&T Star 55 - Manual Servicio
L&T Star 55 - Manual Servicio
Service Manual
Star 55
I
LARSEN & TOUBRO LIMITED
II
LARSEN & TOUBRO LIMITED
Star 55
Service Manual
Star 55
III
Document Information:
CAT No. 3-86-020-0014-68
DRAFT COPY
Copy right:
© 2008 Larsen & Toubro Limited
This service manual is a sole property of Larsen & Toubro Limited and published for limited
circulation only.
This manual or any part of it should not be reproduced without prior and written permission
from Larsen & Toubro Limited.
Star 55
V
Table Of Contents
1. Introduction . . . . . . . . . . . . . . . . . . . . . . . . .
1.1. Audience . . . . . . . . . . . . . . . . . . . . . . . .
1.2. About The System . . . . . . . . . . . . . . . . . . . .
1.3. System Software . . . . . . . . . . . . . . . . . . . . .
2. Safety . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.1. Electrical Safety . . . . . . . . . . . . . . . . . . . . .
2.1.1. Warning . . . . . . . . . . . . . . . . . . . . .
2.1.2. Caution . . . . . . . . . . . . . . . . . . . . . .
2.2. Equipment Safety . . . . . . . . . . . . . . . . . . . . .
2.3. ESD Protection . . . . . . . . . . . . . . . . . . . . . .
2.3.1. ESD Protection Materials . . . . . . . . . . . . . . .
2.3.2. Methods of Protection . . . . . . . . . . . . . . . .
2.3.3. Methods of Elimination . . . . . . . . . . . . . . . .
3. System Architecture . . . . . . . . . . . . . . . . . . . . . .
3.1. System Description . . . . . . . . . . . . . . . . . . . .
3.1.1. Main Board . . . . . . . . . . . . . . . . . . . .
3.1.2. Amplifier Module . . . . . . . . . . . . . . . . . .
3.1.3. NIBP Module . . . . . . . . . . . . . . . . . . .
3.1.4. SpO2 Module . . . . . . . . . . . . . . . . . . . .
3.1.5. CO2 Module . . . . . . . . . . . . . . . . . . . .
3.1.6. Power Supply . . . . . . . . . . . . . . . . . . .
3.1.7. Display . . . . . . . . . . . . . . . . . . . . .
3.1.8. Accessories . . . . . . . . . . . . . . . . . . . .
3.2. System Controls . . . . . . . . . . . . . . . . . . . . .
3.3. Block Diagram . . . . . . . . . . . . . . . . . . . . .
3.4. System Specifications . . . . . . . . . . . . . . . . . . .
3.5. Battery Requirement . . . . . . . . . . . . . . . . . . . .
3.5.1. Battery Specification . . . . . . . . . . . . . . . . .
3.5.2. Characteristics . . . . . . . . . . . . . . . . . . .
3.6. Safety Requirement . . . . . . . . . . . . . . . . . . . .
3.6.1. Product Safety Standard. . . . . . . . . . . . . . . .
3.6.2. EMC . . . . . . . . . . . . . . . . . . . . . . .
3.6.3. Classification. . . . . . . . . . . . . . . . . . . .
4. Getting Started . . . . . . . . . . . . . . . . . . . . . . . .
4.1. Unpacking and Installation . . . . . . . . . . . . . . . . . .
4.2. Turning the System ON and OFF . . . . . . . . . . . . . . . .
4.3. Battery Charging and Checking . . . . . . . . . . . . . . . .
4.4. Connecting / Disconnecting cables for Monitoring . . . . . . . . .
6. Assembly / Disassembly . . . . . . . . . . . . . . . . . . . .
6.1. General. . . . . . . . . . . . . . . . . . . . . . . . .
6.1.1. Removal . . . . . . . . . . . . . . . . . . . . .
6.1.2. Refitting . . . . . . . . . . . . . . . . . . . . .
6.2. Location of Modules . . . . . . . . . . . . . . . . . . . .
6.3. Removal of Rear Cover . . . . . . . . . . . . . . . . . . .
6.4. Removal of Main PWA . . . . . . . . . . . . . . . . . . .
6.5. Removal of SMPS . . . . . . . . . . . . . . . . . . . . .
6.6. Removal of Amplifier Module . . . . . . . . . . . . . . . . .
Star 55
VII
6.7. Removal of SpO2, NIBP and CO2 Module . . . . . . . . . . . . .
6.7.1. NIBP Module . . . . . . . . . . . . . . . . . . . .
6.7.2. CO2 Module . . . . . . . . . . . . . . . . . . . .
6.7.3. SpO2 Module . . . . . . . . . . . . . . . . . . .
6.8. Removal of the Battery . . . . . . . . . . . . . . . . . . .
6.9. Removal of Front Panel . . . . . . . . . . . . . . . . . . .
6.10. Removal of the Keyboard Assembly . . . . . . . . . . . . . .
6.11. Removal of Optical Encoder . . . . . . . . . . . . . . . .
6.12. Removal of Connector plate connectors . . . . . . . . . . . . .
6.13. Removal of Display . . . . . . . . . . . . . . . . . . . .
6.14. Removal of Inverter PWA . . . . . . . . . . . . . . . . . .
6.15. Removal of Alarm PWA . . . . . . . . . . . . . . . . . . .
6.16. Removal of the Recorder . . . . . . . . . . . . . . . . . .
7. Performance Tests . . . . . . . . . . . . . . . . . . . . . . .
7.1. Overview . . . . . . . . . . . . . . . . . . . . . . . .
7.2. Test Equipments . . . . . . . . . . . . . . . . . . . . .
7.3. Power Supply Tests . . . . . . . . . . . . . . . . . . . .
7.3.1. Power Transformer Test . . . . . . . . . . . . . . . .
7.3.2. DC-DC Voltage Test . . . . . . . . . . . . . . . .
7.3.3. Low Battery Checks . . . . . . . . . . . . . . . . .
7.3.4. Battery Charging Test . . . . . . . . . . . . . . . .
7.3.5. Internal / External Battery . . . . . . . . . . . . . . .
7.3.6. Variac Test . . . . . . . . . . . . . . . . . . . .
7.4. Recorder Power Supply Test . . . . . . . . . . . . . . . . .
7.5. Keyboard Test . . . . . . . . . . . . . . . . . . . . . .
7.6. Display Test . . . . . . . . . . . . . . . . . . . . . . .
7.7. Alarm Test . . . . . . . . . . . . . . . . . . . . . . . .
7.8. Inverter Test . . . . . . . . . . . . . . . . . . . . . . .
7.9. ECG Test . . . . . . . . . . . . . . . . . . . . . . . .
7.9.1. HR/PR Switching . . . . . . . . . . . . . . . . . .
7.9.2. LEAD Fault Checks . . . . . . . . . . . . . . . . .
7.9.3. Rate Alarm . . . . . . . . . . . . . . . . . . . . .
7.9.4. SYNC-O/P Check . . . . . . . . . . . . . . . . . .
7.10. Temperature Checks . . . . . . . . . . . . . . . . . . . .
7.11. SpO2 Test . . . . . . . . . . . . . . . . . . . . . . . .
7.12. Recorder Test . . . . . . . . . . . . . . . . . . . . . .
7.13. Analog Output Test . . . . . . . . . . . . . . . . . . . . .
7.14. Skyline Communication Check . . . . . . . . . . . . . . . .
8. Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . .
8.1. System Diagnosis . . . . . . . . . . . . . . . . . . . . .
8.2. Error Codes . . . . . . . . . . . . . . . . . . . . . . .
8.3. Troubleshooting Chart . . . . . . . . . . . . . . . . . . .
Glassary . . . . . . . . . . . . . . . . . . . . . . . . . . .
1.1. Audience
The intended audience of this manual is trained service personnel.
Star 55 offers continuous monitoring for ECG, Respiration, Pulse Oximetry, channel
(optional)Temperature (Along with Delta T), Dual channel Invasive Blood Pressure and Non
Invasive Blood Pressure and optional Capnography with two-channel thermal array recorder
which can record online data and ECG respiration, plethysmograph and CO2 waveforms
along with numerical values of other parameters and also can record tabular trends.
Star 55 has graded and colour coded alarms. Additionally, it also has a red message, yellow
message and white message areas (displayed on the Real Time Screen) which displays
messages during various alarm conditions for better user interpretation. It has 72 hours tabu-
lar and graphical trends with 5 sec resolution for ECG, RR, SP02, Temperature, ETCO2 and
special tabular trend for NIBP to store last 240 readings. Tabular and Graphical Trends can
be viewed at user selectable resolutions. Alarm recall feature offers last 24 alarm conditions.
The basic hardware is comprised of Main board, Power Supply Interface Card(PIC), Bat-
tery Card, Power supply unit (PSU), Module Interface Card(MIC), Communication Interface
Card(CIC), Display module inverter and Parameter modules.
Main board communicates with all the modules and updates the display, resulting in real-time
monitoring of the parameters. The CPU is a 32 Bit Microprocessor with various peripheral
options and controls all the activities of Star 55. VX Works Operating System is used here.
The monitor software resides in a Disk On Chip. It can be upgraded externally through the
USB port using the Software Upload Facility.
Software for other different parameters are embedded in the corresponding modules.
Refer ALL INFO screen for Software versions encompassed in the unit.
Star 55
1
Left blank intentionally
2.1.1. WARNING
• Use insulated tools to adjust controls.
• Do not open the system enclosures in power ON condition.
• All internal adjustments and replacements must be done by a qualified service engineer.
• Do not operate the system in the presence of flammable gases or anaesthetic agents.
This can lead to an explosion.
• Disconnect the AC power adapter from the system before cleaning the system.
• Inspect the AC power adapter cord and plug on a regular basis for any damage.
• Use properly grounded equipment. The grounding wire must not be removed or
disconnected.
• Use accessories and peripherals as recommended for Star 55.
2.1.2. Caution
The following precautions can help reduce ESD (Electro Static Discharge)
• Anti-static spray on carpets
• Anti-static spray on linoleum
• Anti-static mats
• Antistatic wrist bands
• Proper grounding of the work area
Do not block the airflow through the ventilation holes on the back of the system to prevent
rise in temperature.
Caution
• Excessive bending or twisting of cables can cause a failure or intermittent operation.
• Take care when connecting or disconnecting cables. A damaged cable can cause a short
in the electrical circuit.
• Do not plug in or remove PWA (Printed Wire Assembly) cards while the unit is in power
ON condition.
• Improper cleaning or disinfecting or any part of the system with non recommended cleaning
solutions can cause permanent damage.
• Do not use solvents such as thinner, benzene, or abrasive cleaners on any part of the
system.
Star 55
3
• Do not spill liquid on the system.
• Do not handle PWAs without proper static protection. Improper handling may damage components.
• Do not assemble or configure incorrectly. Use proper power source.
• Prevent damage to the connectors by aligning connector pins before you connect the
cable. Misaligned connector pins can cause damage to the system components when
power is swicthed ON.
The damage done by ESD can cause the instant failure of the unit or result in performance
degradation and eventually fail.
The unit gets damaged by the electric field radiating from an electrostatic body. MOS
(Metal Oxide Semicondector) technology devices are most susceptible to voltage or
electrostatic field damage. Bipolar and TTL devices are more susceptible to the range of
electrical current.
Star 55
5
Left blank intentionally
Star 55
7
The Main board also consists of:
• Flash Memory containing the boot loader and image
• Disk On Chip containing the (OS kernel + application software).This also contains
data such as institutional default settings, trend data, etc. that must be retained even
after power OFF
• 2 no 32 MB SDRAM’s from which the code executes
• CPLD chip for address decoding and for generating timing signals like chip
select, signals for RD and write Enable.
• Real Time Clock which maintains system date and time
• Display controller that provides interface between the microprocessor, EDORAM,
and 12.1" TFT display
• EDORAM which stores data for display controller
• Programmable Audio Generator IC for Beep tone and Alarm tones generation.
• 2 no external UART’s with asynchronous communication elements which
communicates with the modules (NIBP, CO2, Recorder) and RS232 ports for
PC Dump and Central Nursing Station (Skyline).
• Keyboard interface which consists of level shifter and buffers which provides the
interface between Main board and keyboard / optical encoder
• Power Supply filters, for power conditioning in 5V,3.3V and 1.8V voltage outputs.
3.1.8. Remote
21 hotkeys along with standby key and an LED is provided on the wireless hand-held
remote. The remote can be used in parallel with the internal keyboard, over a maximum
distance of 20 feet.
Use the Remote in straight line with the remote sensor to avoid changes in other monitors
(55 series) kept near by.
3.1.9. Accessories
ECG
Star 55 offers 3 lead ECG monitoring facility.
• 3 lead ECG cable (I, II, III)
• 5 lead ECG cable (I, II, III, aVR, aVL, aVF, V)
Respiration
Figure 3.1 : 3 Lead ECG Cable Figure 3.2 : 5 Lead ECG Cable*
* Optional
Star 55
9
Respiration can be monitored through same ECG cables 3 / 5 lead or through
Capnography (Refer Page No 12 and 13). Priority for respiration is given to
Capnography, if the option is installed.
Temperature
Figure 3.3 : 3 Lead ECG Cable Figure 3.4 : 5 Lead ECG Cable*
Pulse Oximetry
Figure 3.6 : Temperature Probe (Adult)* Figure 3.7 : Temperature Probe (Neonatal)*
* Optional
* Optional
Star 55
11
NIBP basic configuration comprises of following accessories:
• Adult cuff (14 cm x 37 cm) (Reusable)
• Child cuff (9 cm x 27 cm) (Reusable)
• Neonate cuff (3 cm x 9 cm) (Disposable)
• Hose tube (3 meters) (Reusable)
Capnography (Optional)
Grounding Cable
Star 55
13
3.2. Front Panel
15 17 20
16
18
19 21
1 234 5 6 7 8 9 10 11 12 13 14
Figure 3.24 : Front View
Star 55
15
3.3. Rear Panel
1 3
CIC Connectors
CIC-4 PC Dump + CNS
Connector Description
DB9 PC Dump
USB* Central Nursing Station
Star 55
17
3.4. Left Side Panel
4
1 5
6
2 7
3
10
Star 55
19
3.5. Right Side Panel
Star 55
21
3.6. Interconnection Diagram
SMPS takes input supply from mains or battery and converts it to +/-12V +5V and
8.5V. It charges battery voltage and gives supply to inverter PCB for display.
PIC receives +/-12V +5V ,& 8.5V and distributes to the various modules.
Motherboard loads the system defaults and performs a self test by sensing the
connected module and poping up the respective module.
Motherboard sends the parameter settings to all the boards connected. The data
(digital values and waveforms) from all the boards is collected and displayed
according to the display format selected by the user. This data is also stored for
72hrs. trend. The same data is available for output in RS232 protocol for serial
communication.
Motherboard monitors for fault conditions from all the modules and displays
accordingly. It compares the data to the user set limits and gives alarm whenever
limits are violated.
Power supply monitors the battery charge, if no input supply present. It sends low
battery message for display when battery charge level goes below 20% charge.
Star 55
23
3.8. System Specification
Equipment Classification
Mode of operation : Continuous
Type of protection against electric : Class1
shocks
Degree of protection against electric : Type CF for ECG, IBP, Temperature and Respiration
shocks Type BF- Capnography, NIBP and SpO2
(All with defib protection)
Degree of protection against hazards
of explosion : Not protected
Degree of protection against ingress
of liquids : Drip Proof - IPX1
Power Supply
Voltage : 95 - 265V AC, 50Hz / 60Hz at +/-5% tolerance.
Fuse : 3.15A Fast blow (2 no.s)
Indicator : Amber LED ON indicates Mains ON.
Green LED ON indicates Monitor ON.
Amber LED OFF and green LED ON indicates
unit is in battery mode
Current : 0.7 A max
Wattage : 100 watts
Battery
Type : 14.8 V (4AH / 8AH) Lithium ion
No. of battery : One pack / Two packs
Charging time (min) : 12 Hrs ( one pack )
Discharge time : 2 / 4 hrs
For 2 hrs backup :
One pack fully charged new batteries at 25 0C.
Discharge condition:
ECG, RESP, SpO2, Temp, NIBP running at 15 min interval, without CO2 and recorder printing.
For 4 hrs backup :
Two packs fully charged new batteries at 25 0C.
Discharge condition:
ECG, RESP, SpO2, Temp, NIBP running at 15 min interval without CO2 and recorder printing.
Low battery indication : 20% charge
Battery cut-off voltage : 12.2 +/- 0.2V
Indicator :
Green LED ON and Amber LED OFF indicates battery operation.
Green LED ON and Amber LED ON indicates mains operation and battery charging.
Controls
Front Panel : Switch for unit ON/OFF control.
9 specific function switches.
Remote sensor.
Optical encoder with switch.
Display
Display size : 12.1 " TFT display (800X600).
Trace Speed : 12.5, 25, 50 mm/sec for ECG, SpO2, IBP.
6.25,12.5,25 mm/sec. for Respiration and CO2.
Waveform : 8 waveform display. one standard and six user
selectable format. Waveform and Digital fields are
individually selectable and can be stored.
Parameter colour : Colour of the parameter display ( w/f and digital
fields together ) is user selectable.
SpO2 Strength Indicator : Bar graph showing the signal strength.
Waveform sampling rate : 400 samples/sec for ECG ( max )
100 samples/sec for SpO2 ( fixed )
100 samples/sec for Respiration ( fixed )
10 samples/sec for Capnography.
Patient Modes : Adult/Neonate/Pediatric
Recording indication : RECORDING… message in the white message
area indicate recording in progress.
Arrhythmia indicator : Flashing red alarm and arrhythmia type in message
area on screen provided
Alarm detection status : Selected limits shown if enabled. Crossed Bell symbol
shown if disabled. Depends on mode, default alarm
Trends
Data storage : 5 sec for 72 hours
HR / PR, SpO2%, IBP, Temp, RR, EtCO2 : 72 hours graphical and tabular trend
FiCO2, ST.
NIBP : Last 240 readings will be displayed in Tabular format.
ST : 4 Hrs data with 1 min. resolution ( ST storage ) No
option to record strips.
OXYCRG display : 8 min. online monitoring of HR, SpO2 and RR with
the resolution of 2sec.
Graphical trend time scales : 2, 4, 12, 18, 24, 48 and 72 hrs.
Graphical trend view resolution : 30 sec, 1, 3, 4.5, 6, 12 and 18 min.
Tabular trend view resolution : 5, 10, 15, 30 sec, 1, 2, 4 and 8 min.
Star 55
25
Alarm Trend (Recall) : Tabular trend for display of last 24 patient alarms
Patient information : Max storage of 72 Hrs Data of 200 patients
Storage of ECG strips : 8 ECG strips (strip duration depends on the sweep
selected for ECG) what ever is appearing on the
display at that moment will be stored) ECG strip
recording is possible.
Auto Setting Of Alarms : Provided for HR, SpO2, IBP, RR, EtCO2, FiCO2, ST
and TEMP.
Formulae Used : HR high limit = Present HR value X 8/10 +76
HR low limit = Present HR value X 8/10 +2
SpO2 high limit = 100
SpO2 low limit = (Present SpO2 value - 8) or 80
(whichever is greater)
IBP systolic high limit = Present systolic value +8
IBP systolic low limit = Present systolic value -8
IBP diastolic high limit = Present diastolic value +8
IBP diastolic low limit = Present diastolic value -8
EtCO2 high limit = present value +5
EtCO2 low limit = present value -5
FiCO2 high limit = present value +2
FiCO2 low limit = 0
Temperature high limit = present temperature +2
Temperature low limit = present temperature -2
Respiration high limit = present RR +5
Respiration low limit = present RR -5
ST high limit = present value +2
ST low limit = present value -2
Inputs
Side Panel : Nicholay connector for ECG / Respiration
D-type connector for SpO2
Red colour round connector for IBP
Female coupling for NIBP
YSI connector for Temperature
CO2 filter line
Rear panel : Standard 3-pin IEC power connector.
Line and neutral lines protected by fuse.
ECG
Input : Isolated and floating 3/5 leads (depending on ECG
cables). Protected against surges produced by
Electro-Surgical and defibrillator potentials. Cascade
option available.
Input impedance : >2.5 M Ohms at 10Hz
CMRR : <15mm (max.) (Allowable noise for 3V p-p applied
sine wave of 20Hz)
Patient applied leakage current : < 10uA (Isolated Normal condition)
< 50uA (Isolated single fault condition)
Heart Rate
Range : 20-350 BPM
Accuracy : 2BPM or 2% which ever is greater
HR alarms : Adjustable alarm limits
Min - 20 to 320 BPM.
Max - 50 to 350 BPM
4 sec delay for HR alarms
3 sec R-R interval triggers systole alarms if selected
in Arrhythmia menu.
Time to alarm for Tachy : 5 sec
Heart rate meter response time : Change from 80 to 120 BPM (6 to 7 sec).
Change from 80 to 40 BPM (6 to 8 sec).
Arrhythmias Detected : Irregular, Skip beat, Tachycardia, Bradycardia,
Asystole
Pacemaker pulse rejection capability : When pacer is ON the monitor will display the HR
for all single and double pacemaker pulses either
150 or 250 m sec apart with a pulse width of 0.1 to
2 m sec and amplitudes ±2 mV to ±500 mV without
overshoot.
Note &
Do not rely on the displayed HR value when pacer is OFF and pacer pulses are detected as
QRS.
Star 55
27
ST Analysis
Input : From ECG cable
Measurement range : -10mm to + 10 mm
Resolution : +/- 0.2 mm.
Leads : ST analysis can be performed on any one of I, II, III
incase of 3 lead configuration and any three of I, II,
III, avR, avL, avF, chest in case of 5 lead
configurations.
ST value updation : After every 8 valid QRS samples.
ISO point : -80 ms from QRS peak (Default) Any point in the
range of -150ms to 0 from QRS peak ( user
selectable ).
ST point : 120 ms from QRS peak (Default) any point in the
range of 0 to 200ms from QRS peak ( user
selectable ).
Marker resolution : 2.5ms.
Alarms : Alarm limits adjustable in the range
High : -8 to 12
Low : -12 to -8
Method of measurement : Cyclic
Temperature
Inputs : Isolated and floating
Leakage Current : <10 mA at 240V AC, 50 Hz
Measurement Range : 0 0C - 50 0C
Scale value : Displays direct readings of Temperature input. ‘DT’
value displayed in place of T2. DT is the difference
between T1 and T2.
Warm up : <10 min (Excluding probe)
Accuracy : ±0.2 0C or ±0.4 0F
Alarms : Temperature limits adjustable
Minimum : 12.00 C to 47.0 0C
Maximum : 15.00 C to 50.0 0C
Unit : 0
C or 0F user selectable
SpO2
Tone variation with change in SpO2 : Provided
Measurement range : 0 -100%
Accuracy : Adults : (+1 Std. Dev.)
70 - 100% +2 digits
0 - 69% unspecified
Neonates : (+1 Std. Dev.)
70 - 100% +3 digits
0 - 69% unspecified
Alarms : Adjustable alarm limits
Minimum : 50 to 95%
Maximum : 55 to 100%
Pulse rate range : 20 to 230 BPM
Accuracy (+ 1 Std. Dev.) : +3 BPM
RR : 0-150 BPM
Units : mmHg or kPa or Vol %.
Scale : 20, 40, 60, 80 or 100 mmHg or 2.6, 5.3, 8, 11 kPa /
Vol%
Flow rate : 50 ml/min
Warm up time : 40 sec (Typical)
Star 55
29
Alarms : EtCO2
Upper : 5-80 mmHg
Lower : 0 - 75 mmHg
FiCO2
Upper : 2 - 20 mmHg
RR
Upper : 10 - 150 BPM
Lower : 5 - 145 BPM
Accuracy : 0-70 (±1) BPM
71-120 (±2) BPM
121-150 (±3) BPM
Calibration :
Zero calibration : Automatically performed by the module. Indicated
to the user through a message on the screen.
Note &
No Breath alarm limit is 10-90 with step of 5 sec.
NIBP
Method : Oscillometric
Display : Systolic, Diastolic and Mean
Modes of measurement : Manual, Auto and Stat mode
• In Auto mode, intervals of 2, 3, 4, 5, 10, 15, 30, 60 and 90 minutes are user selectable
• In Stat mode, unit will take as many readings as possible in 5 minutes
• Duration between measurements is 10-12 sec
Unit (User selectable) : mmHg or kPa
Range : 20 - 250 mmHg
Accuracy : +5 mmHg with a standard deviation no greater
than 8 mmHg
Cuffs : Single quick connect hose
Auto zero : Zero pressure reference is automatically established
after every reading
Cuff inflation
Initial inflation
Adult : 160 mmHg
Neonates : 90 mmHg
Pediatric : 120 mmHg
Subsequent inflation approximately 30 mmHg greater than previous systolic pressure
Cuff deflation : Automatic
Alarms : Adjustable alarm limits (for both Sys and Dia)
Upper : 30 to 250 mmHg
Lower : 20 to 240 mmHg
Drug Calculation
25 Standard default drug details are as follows:
AMINOPHYLLINE, DILTIAZEM, DOBUTAMINE, DOPAMINE, EPINEPHRINE, ESMOLOL
HCL, FENTANYL CITRATE, HEPARIN SODIUM, INAMRIONE LACTATE INSULIN REGULAR,
ISO PROTERENOL, HCL LABETOL, HCL LIDOCAINE, HCL LORAZEPAM, MIDAZOLAM,
MILRIONE, NICARDIPINE, NITROGLYCERIN, NITROPRUSSIDE, NOREPINEPHRINE,
PHENLYEPHRINE, PITOCIN, PROCAINAMIDE, HCL PROPOFOL, VASOPRESSIN. Apart
from above drugs there is a provision to add 10 New drugs.
Star 55
31
Alarms
Accessory alarms : Audio ( Alarm beep ), Visual ( Flashing Yellow LED )
and message indication
Patient alarms : Audio ( Alarm beep ), Visual ( Flashing Red LED
alarms ) and message indication
Re-alarm : 3 min after an alarm is acknowledged
Alarms suspend : Continuous Yellow LED to indicate “Alarms Suspend”
condition with display of alarm cross mark.
Alarm volume control : 12 steps (Min to Max)
Outputs
ECG Sync : Provides 5V P-P pulse for synchronizing with
defibrillator
Analog : 15 pin D-type sub connector ( Female ) ( Future
Upgrade )
External monitor : 15 pin D-type sub connector ( Female ) ( Future
Upgrade )
RS485 communication : USB type
Ethernet : RJ45 type (Future Upgrade)
Nurse call : Stereo phono jack type
PC Dump : 9 pin D-type, dumping of all real time data.
Software upgrade : USB type
Recorder
Recorder : Two channel
• Make : Woosim
• Printing method : Thermal recording
• Dot density : 203 DPI
• Paper Width : 58 mm
• Printing Width : 48 mm
• Speed : 25 mm/sec
Recording modes : Direct, Delayed / Continuous modes on red alarm
Recording duration : Direct : 8 sec
Delayed : -6 to +2 sec
Continuous : 30 sec
On alarm : -6 to +2 sec
Communication
CNS Interface : Communicating with Skyline ( L&T’s CNS - Future
Upgrade )
USB : USB connector
Ethernet port : RJ45 connector (Future Upgrade)
PCDUMP : Dumping of all trend data (9-pin D-type)
General
Dimensions ( H * W * D ) : 280 mm * 320 mm * 158 mm
Weight : 5.5 Kg (approx) with one battery pack and recorder
without accessories
LARSEN & TOUBRO LIMITED
32
Mounting option : GCX option
Operating temperature : 0 to 40 0C
Operating humidity : 10 to 90% RH (Non condensing)
Storage and Transportation : -10 to 50 0C
temperature
Storage and Transportation : 0 to 90% RH (Non condensing)
humidity
Operating pressure : 500 to 760 mmHg
Storage pressure : 500 to 760 mmHg
Accessories Supported
ECG / Respiration : 3 lead cable with electrodes
SpO2 : DS ( Dura sensor ) 100A ( Adult )
Oxy-A / N ( Adult / Neonate ) ( Optional )
Dura Y ( Universal ) - ear clip ( Optional )
NIBP : Tuff-Cuff reusable blood pressure cuffs ( Infant to
large adult sizes )
IBP ( Optional ) : Disposable Transducer kit and IBP cable
Temperature ( Optional ) : YSI 400 series temperature probes
CO2 : Microstream filter lines
Remote : Infrared
Standards
Designed to confirm to the following international standards
Class 1 equipment requirement of IEC 60601-1.
EMI/EMC requirements as per IEC 60601-1-2.
IEC 60601-2-27 requirements for ECG.
IEC 60601-2-30 requirements for NIBP.
IEC 60601-2-34 requirements for IBP.
IEC 60601-1-8 requirements for alarm systems.
IEC 60601-2-49 requirements for multifunction patient monitor systems.
ISO 9919 requirements for SpO2.
ISO 21647 requirements for CO2.
AAMI EC 13 for ECG.
AAMI SP 10 for NIBP.
Star 55
33
3.9. Battery Requirement
3.9.2. Characterstics
Capacity 20 hour rate (60mA) 4.0AH
3.10.2. EMC
Conducted and radiated emission as per CISPR11/EN55011 Group1 ClassA,
IEC601-1-2,
Electrostatic discharge as per IEC 1000-4-2,
Radiated susceptibilty as per IEC 1000-4-3,
Electrical fast transient as perIEC 1000-4-4,
Surge immunity as per IEC 1000-4-5.
3.10.3. Classification
The equipment shall be of Class II and type CF.
Ingress of liquid suitability for use in presence of flammable anaesthetic mixture.
4.2.2. Placement
Star 55 is a portable bedside patient monitor. It can be placed on a table, stand, or trolley
or mounted on a pole. Do not place the equipment in an enclosed area that could restrict
heat dissipation from the front or rear of the unit.
Star 55
35
Note &
In case the monitor is not responding or not getting switched OFF you can do a hard shut
down by pressing the ON/Off key for more than 5 sec
Figure 4.1
Notch
ECG IBP
Figure 4.2
NIBP Connection:
• Insert the NIBP hose tube in the socket provided for NIBP connection.
• Press the NIBP socket to release the NIBP hose tube (Refer Figure 4.3).
Figure 4.3
Caution
Do not push the connector forcefully while connecting.
Note:
Read the operating guide carefully to start monitoring the patients condition.
Star 55
37
Left blank intentionally
5.2. Mainboard
The Main board controls the operation of the monitor (Refer to Figure 5.1). Power is supplied
to the Main board from the Power Supply Interface Card(PIC ) using a c/f. The voltage supplied
will be 5V. This 5V is inputted to a Regulator to derive 3.3V which is required for functioning
of the IC’s in the Motherboard. 1.8V is derived from 3.3V forsupplying to Processor core.
The MPC880 microprocessor controls the operation of the monitor. The Reset Control logic
controls the behavior of the board at power up. An on-board 25MHz crystal provides the
basic clock required to operate all devices. A programmable clock generator is used to
generate other frequencies that are used to clock components such as the Microprocessor,
Communication peripherals like USB, Ethernet, the display controller and for the SD Card
Controller.
The Boot Loader is programmed into the Flash IC which loads the initial image (L&T Logo)
and is responsible for loading the OS.
A CPLD is used to generate the chip selects, Read, Write Signals for various ICs. The
processor runs the program out on the SDRAM’s. The code memory resides in the Disk On
Chip. When the processor boots up, it loads the application into the SDRAM and starts
execution. The SDRAM’s also stores temporary data that is used by the monitor during code
execution.
Data that the monitor retains after shutdown is stored in the Disk On Chip. This includes
Date, Time, Trend Data and Institutional settings. The Display controller acts as the interface
between the processor, display memory (EDORAM) and the display.
The Main board has 2 external UARTs(Option is provided for the 3rd one) that control
communication with modules such as NIBP, CO2,Recorder and the RS232 port. In addition
to these external UART’s the 2 processor internal UART’s are used for communication with
the SpO2 and ECG modules. The communication interface between the Main Board and all
the modules in Star 55 are routed through a Module Interface Card(MIC). All Transmit, Receive
signals are transmitted between the MIC card and Main PWA using a Signal C/F. All the
modules are connected directly to the connectors mounted on the MIC Card.
A programmable Audio Generator IC is being used for audio output for Alarm and Beep
tones. This IC contains audio files which is is programmed before assembling into the
Motherboard.
Star 55
39
The SD USB card is connected to the motherboard using a high speed interface connector.
The SD and USB signals and power to SD USB card is transmitted between Motherboard
and SD USB Card using this connector.
The communication Interface Card directly connects to the motherboard using a 100 Pin
High Speed Connector. All the signals and Voltages for various communication modules like
PC Dump, CNS Slave Monitor, Ethernet and Analog Output are transmitted between the
Motherboard and connected Communication Interface Card using this high speed connector.
The motherboard also communicates directly with the Power Supply over the I2C protocol.
Various information’s like Battery Charge %,Temperature, Voltage Level Info and Frequency
info is being sent to the motherboard from the Power Supply using the power c/f between PIC
and Motherboard. This all information can be viewed in the service screen.
In addition to these there is a Alarm driving and keyboard interface circuit consisting of Buffers
and Level Shifters. The keyboard and Alarm LED pulse o/p directly comes out of from the
processor and is interfaced using Buffers and Level Shifters.
ECG section:
Monitor measures the patients ECG signals by measuring the bio-potentials at different skin
surface of the body known as leads like RA,LA,LL for 3 lead configuration and RA,LA,LL,RL
and CL for 5 lead configuration. These leads will sense the muscular activity of the Heart in
form of bio-potential. These signals are of very low voltage signals. Module receives these
bio signals from patient body and signal amplification as well as filtering will takes place in
ECG circuitry of the module.
ECG section comprises Lead selection section, lead signal amplification and filtering section
as well as return path selection section, lead fault detection circuitry, DC level bias adjustment
section , ECG filter mode selection section for MON/DIAG/OT modes, PACER detection
section, SYNC pulse generation section. These Analog signals are fed to the ADC for digital
conversion.
Respiration Section:
Monitor measures the patient respiration signal by ECG leads with the minimal 2 lead combination
of RA-LA or RA-LL. This section comprises of respiration transformer, oscillator, amplifiers
and filters. We will measure the respiration signal by current injection to the body known as
impedence plethysmography. Resulting of this technique module receives the modulated
signal from the body and it will be detected using Demodulation method. The demodulated
signal is passed through low pass filter and high pass filter and Amplifier. Finally this amplified
respiration signal is given to ADC for digital conversion.
Star 55
LEAD-I
RA LA RL RA CHANNEL CARD
LEAD-II
LA
LEAD-III
ESU FILTER, LL
ANA 40 Hz Bend ECG
Defibrillator Instrumentation
LOG Pass and 50Hz A
Pretection and Amplifier
MUX Notch filter N
buffer GAIN STATE &
A 10-BIT PROCESSOR
OFFSET
L ADC
RA ADJUSTMENT
O
SPDT DIODE Respiration Output G
TRANSFORMER
RELAY DETECTOR AND
LA AMPLIFIER M
U OPTO ISOLATOR
CLOCK
TRANSFORMER RECTIFIER FILTER +5V REGULATOR
GENERATOR
NON-ISOLATED ISOLATED
-5V
41
Temperature Section:
Monitor measures patient temperature by temperature probes which will be placed on the
skin surface of the body. These probes are nothing but transducers which will convert temperature
in to equivalent resistance. This resistance will be sensed by the module and hence equivalent
temperature is measured. This equivalent signal will be fed to the ADC for digital conversion.
Pump
J3 – Data power
10
BP Module Connectors
The Model 2 ADVANTAGE OEM BPTM module is an oscillometric OEM blood pressure
system. It operates on +12V supply which is supplied from SMPS through main board. The
Module is controlled via software commands issued from a host system through an
asynchronous serial data port . All Module operations are initiated by the host system. The
Module is designed to take blood pressure samples on demand. After each blood pressure
sample, the Module will discard the previous blood pressure results. Serial communications
baud rate is 9600, with 8-bit data, no parity, and 1 stop bit. Inputs to the Module consist of:
• Cuff Pressure
• Power Outputs from the Module
• Serial Data Port
All Module Operating parameters will reset to default values at power-up. All Module blood
pressure values are discarded when the Module is reset.
+5 µP
Control
Circuit
SP02 Sensor
Block Diagram of SpO2
This module is from Nellcor Inc., USA. It communicates with the main board on serial protocol
providing PR, SP02 % which are refreshed on the display. It oper ates on +5V supplied
from Main board. The finger probe houses an infrared source, red LEDs and a photo detector.
It picks up the signal from the patient and this signal is conditioned by the microcontroller to
derive the oxygen saturation level and pulse rate.
Analog Section
5 Vdc Microstream
‘FRS’ Circuits
Sensor Flow Connector
Assembly System
Gas Outlet
5.6.1. Description
The monitor uses Oridion’s MiniMediCO2 Capnography module. The Main board provides
5V to power the module. The module communicates with Main board using standard
non-isolated RS 232-level serial interface. The module provides inspired/expired carbon
dioxide and respiration rate data, which are refreshed on the display every 1 sec and CO2
waveform data on real time. The method for measuring CO2 is based on non-dispersive
IR absorption of the CO2 in the breath sample using the Oridion IR source. The major components
of the module are as follows:
• CO2 sensor assembly, including the Oridion long-life source
• Flow System, including pump
• Microcontroller + peripherals
• Serial interface
Star 55
43
• Power supply
• FRS to identify the correct usage of a Microstream FilterLine
• Input gas connector with the FRS interface
Before the module begins to measure CO2, it completes the self-test and initialization
procedures required for proper operation. The module enters the power-up mode after
powering the module or after external or software reset. In the Power-up mode, the module
performs a set of self-tests. If the module passes the self-test, it transfers to measurement/
initialization mode. If the Main board fails to detect the module during the self-test, it shuts
down. The power-up takes up to 10 seconds, during which time there is no communication
available.
During the self test, if the either of the modules fail to communicate properly or do not pass
their self test; the respective error is displayed as”CO2 ERROR” or”SPO2 ERROR” on
the Capnostream.
Star 55
45
5.7. Power Supply Unit
AC - DC Converter
Frequency
sensing ckt F sense
Batt +Ve
Battery Battery Batt –Ve
Charger ckt Selector SCL
SDA
DC/DC +5V Output
ON/OFF 5V
Logic Converter
Ground
12 V +12V Output
Converters
Control
F sense
Logic SCL& SDA
ON/OFF to MB
Logic
DC - DC Converter
Description
EMI Filter
This will help to reject the line noise and filter out the power supply generated noise. The
filter winding resistance and inductance can provide the necessary series impedance to
limit the transient current for efficient transient suppression.
Control Logic
The functions of this card are control the ON/OFF logic, protect the DC-DC from short ckt
and over voltages, count the line frequency, collect the battery pack information through SMBUS
and send battery pack information, line frequency and output voltages to the main board
through I2C communication.
1 2
+5V 1 2 Gnd
3 4
+5V 3 4 Gnd
5 6
+5V 5 6 Gnd
7 8
+12V 7 8 Gnd
9 10
+12V 9 10 Gnd
11 12
Bat SCL 11 12 Bat SDA
13 14
-12V 13 14 Key Power ON
15 16
Mains ON 15 16 SDA Host
17 18
+8.5V 17 18 Gnd
19 20
Soft Shutdown 19 20 SCLK Host
21 22
VBat + 21 22 Gnd
23 24
VBat + 23 24 Gnd
Star 55
47
Left blank intentionally
Figure 7.1
7.3.1. Power Transformer Test
With a digital multimeter (DMM) measure the resistance between Line and Neutral pins
of rear panel power socket. See that
a) DMM shows 700KW to 900KW when rear panel switch is On.
b) With the rear panel switch in ‘On’ position Line to Earth resistance should be
>10MW.
c) Resistance between EARTH and CHASSIS should be < 0.4W.
Star 55
61
PIN 13 : -12 volts
7.3.3. Low battery checks
1. Put rear panel switch to OFF position.
2. Give 12 Volt DC to external battery connector from an adjustable power supply. Switch
ON the unit. Wait till the unit comes to the main menu. Slowly reduce the voltage (using
fine adjustment knob of adjustable power supply).
3. Ensure the following points:
“LOW BATT” message and green LED flash at 11.3V to 11.7V.
System shuts OFF at 10.4V to 10.8V.
Star 55
63
2. Check for gain 1 and 2 in SP02 menu. The w/f amplitude should change in screen and
analog o/p.
Note C
The system is repairable through subassembly replacements
Star 55
65
8.2. Troubleshooting Chart
NO Display C
Yes
Yes
Yes
Is unit working? Unit working satisfactorily
No
Yes
Is unit working?
No
Star 55
67
$
Yes
Is unit working? Unit working satisfactorily
No
Yes
Is unit working?
No
Yes
Yes
Is unit working? Replace the Battery
No
Yes
Is unit working?
No
Yes
Is unit working?
No
Star 55
69
+
Yes
Is unit working? Replace Particular Module
No
Yes
Is unit working? Replace the MIC PWA
No
Yes
Is unit working?
No
No
Is Display Back Light #
coming?
Yes
Yes
Yes
Switch ON the Unit Is unit working?
No
No
Boot Bell is coming? Replace Motherboard
Yes
Star 55
71
#
No
Is PIC to Inverter Cable Form Reconnect the Cable Form
connected?
Yes
No
Replace the PIC to Inverter Is unit working?
Cable Form and Switch On the
unit
Yes
Yes
is unit working? Unit working satisfactorily
No
Yes
Is Display working?
No
No
Is Display Back light Replace PIC PWA
coming?
Yes
No
Is Module to MIC Cable Form Reconnect the Cable Form
connected?
Yes
Yes
Is Module detecting? Unit working satisfactorily
No
Yes
Is Module detecting?
No
Yes
Is Module detecting?
No
Replace Motherboard
Star 55
73
E
No
Is Module Sensing? D
Yes
Is Communication Error No
Replace ER2T Module
displayed on the screen?
Yes
Unit working satisfactorily
Yes
No Yes
Is unit working?
No
Is 5V at Pin No.5 in the Replace MIC PWA
Cable Form coming?
Yes
Yes
No
Is unit working? Replace Motherboard
Yes
LARSEN & TOUBRO LIMITED
74
F
No
Is Module Sensing? D
Yes
Is Communication Error No
Replace SpO2 Module
displayed on the screen?
Yes
Unit working satisfactorily
Yes
No Yes
Is unit working?
No
Is 5V at Pin No.5 in the Replace MIC PWA
Cable Form coming?
Yes
Yes
No
Is unit working? Replace Motherboard
Yes
Star 55
75
G
No
Is Module Sensing? D
Yes
Is Communication Error No
Replace CO2 Module
displayed on the screen?
Yes
Unit working satisfactorily
Yes
No Yes
Is unit working?
No
Is 5V at Pin No.5 in the Replace MIC PWA
Cable Form coming?
Yes
Yes
No
Is unit working? Replace Motherboard
Yes
LARSEN & TOUBRO LIMITED
76
H
No
Is Module Sensing? D
Yes
Is Communication Error No
Replace NIBP Module
displayed on the screen?
Yes
Unit working satisfactorily
Yes
No Yes
Is unit working?
No
Is 5V at Pin No.5 in the Replace MIC PWA
Cable Form coming?
Yes
Yes
No
Is unit working? Replace Motherboard
Yes
Star 55
77
I
No
Is Module Sensing? D
Yes
No
Yes
Is unit working?
No
Is Sensor working?
No
Yes
Replace IBP Module
Replace Sensor
No
Is IBP Ext. Cable working? Unit working satisfactorily
Yes
Replace IBP Ext. Cable
No
Is Module detecting?
Yes
No
No
Yes
Replace Motherboard Is unit working?
No
Star 55
79
K
No
Is Module detecting?
Yes
No
No
Yes
Replace Motherboard Is unit working?
No