Caudalimetro MW - Hovedcat
Caudalimetro MW - Hovedcat
Content
Key Facts, Working Principle Seite 2 Pressure drop Seite 7
“MASS-STREAM®” Seite 3 Model number identification Seite 8
Mass Flow Meters Seite 4 Technical specifications Seite 9
Mass Flow Controllers Seite 5 Ultra fast sensor, Digital version Seite 10
Conversion factors, Readout systems Seite 11
Flow profile sensitivity Seite 6 Contact addresses Seite 12
Principle of Operation
Basically the instruments consits of a metal block with a straight Seals
bore. Two stainless steel probes protrude inside the bore; a heater Body
probe and a temperature probe. A constant difference in temperatu- Heater
re (∆T) is created between the two and the energy required to main-
tain this ∆T is dependent of the mass flow rate.
Sensor
Generally speaking we can say that the higher the flow the more
Flow
energy is required to maintain the chosen ∆T, which is usually
approx. 38° C. Overall we can state that King’s law applies to the
relationship between heater energy and mass flow, and the follo-
wing formula can be derived. Basic structure of the MASS-STREAM®-flow sensor
2
“MASS-STREAM®”
Features and Applications
Worth knowing
MASS-STREAM® is the synonym for a metering principle having the
following advantages:
Features
◆ Low pressure drop ◆ Installable in virtually any position
◆ Rugged design ◆ No moving parts
◆ Lower sensivity concerning dirt and humidity ◆ Bodies in stainless steel or as a more economical
◆ Measuring independent of pressure and temperature changes aluminium version
Applications
◆ Gas consumption metering
◆ Exhaust gas metering
◆ Semiconductor industry
◆ Analytical intruments
◆ N2/O2-generators
◆ Fuel cells
◆ Mechanical engineering
◆ And much more
Options
◆ “Low ∆P” version
◆ Integrated totalisation
◆ Integrated actual display
◆ Integrated setpoint potmeter
◆ Readout systems
3
Model D-6210 MFM
The 62xx series working principle is shown For lower flow values the by-pass measu-
on page 2. rement principle is applied.
Model A B C D F G H
D-5110 95 G1/4” G1/4” 30 90 92 35
D-6210 95 G1/4” G1/4” 30 90 92 35
Standard Flow Capacities D-6230 95 G1/4” G1/4” 30 90 92 35
D-6250 95 G1/2” G1/2” 30 90 92 35
Mass Flow Meters Flow capacitiy (Air)
Model (intermediate ranges available) Model D-6270
The modular construction solenoid valve The following kvs-values are available:
is integrated to the base when flows are 6,6E-2, 0,35; 1,0;
up to approx. 500l/min. N2 equivalent. (for higher values please contact factory)
Features
◆ Suitable for almost all gases ◆ Optional: integrated actual flow
and mixtures display
◆ No moving parts in the sensor ◆ Optional: integrated totalisation
◆ Good response times display
◆ PI-control loop ◆ No maintainance required Model A B C D F G H
◆ No inlet pipe necessary ◆ Almost independent to attitude D-5111 95 G1/4” G1/4” 30 90 92 35
(D-62xx series) D-5121 95 G1/2” G1/4” 30 94 96 35
D-6211 95 G1/4” G1/4” 30 90 92 35
D-6231 95 G1/4” G1/4” 30 90 92 35
Best accuracy is reached by calibrating the instruments und actual When using our D-62xx serie with balloon gases like helium and
process conditions. The conversation factor introduces an additional hydrogene it is always recommended to make use of the optional gas
error in abolute accuracy in order of: calibration.
CF ≥ 1 : 2xCF in % FS
CF ≤ 1 : 2/CF in % FS
6
Pressure drop
The pressure drop of the instruments (serie D-62xx) is almost com- the flow profile. These meshes create a certain pressure drop as can
parable to a straight run of pipe of the diameter ans is thus negible. be seen from the table set out below.
However, to make the instruments insensitive to upstream piping For further informations please contact factory.
configurations, a number of mesh screens are required to condition
legend
D 6210
D 6230
D 6250
D 6270
D 6280
D 6290
∆P (mbar)
In some applications a very low pressure drop is required. To fulfil For further information please refer to the second table set
these inquires we are able to build a special version of our instru- out below.
ments where the pressure drop is extremely low.
legend
D 6210
D 6230
D 6250
D 6270
D 6280
D 6290
∆P (mbar)
7
Model Number Identification
Options and Model Numbers MASS-STREAM®
D NNNN AAA AA AA NN A A A
Bus Connection /
Option 7
A = analog
DP = digital / Profibus
DN = digital / device net
DR = digital / RS 232
DF = digital / Flow Bus
8
Technical Specifications
Measurement System
Accuracy (based on air calibration) ± 3% FS incl. non-linearity (better one’s on request)
Repeatability ± 0.5% FS (better one’s on request)
Time constant (63.2%) τ = 0.7 sec (standard, better one’s s. p. 10)
Pressure sensitivity 0.2% / bar typical (air)
Attitude sensitivity ± 0.1% °C
-9
Leak integrity < 2 x 10 mbar l/s He
RFI According to CE
Operating Limits
Range 3…100%
Type of gases All gases compatible with materials chosen
Temperature 0…70°C
Pressure 10 bar; higher on application
Warming up time 30 min for optimum accuracy;
within 30 sec for accuracy ±4% FS
Installation
Series D-5100… 10 D straight pipe upstream
Series D-6200... Unrestricted
Mechanical Part
Sensor AISI 316L
Body AISI 316L or Aluminium (anodised), please specify
Sieves / rings Stainless steel / teflon
Protection IP40 (IP 65 on application)
Electrical Properties
Supply voltage 24 Vdc ±10% or 15 Vdc ±10%
Current peak values
Series D-5100... 75 mA max.
Series D-6200... Inrush current 250 mA max.
No flow 75 mA max.
100% flow 175 mA max.
Control valve, if applicable Add 250 mA max.
Output signal 0…5 Vdc or 4…20 mA
Cable 6-wire DIN or 15-wire SUB-D
9
Ultra-Fast Sensor
"Fast response” version
M+W Instruments Mass Flow Meters and Controllers are used for a
Seals
wide variety of applications and for this reason the sensor has been Body
set up with "smooth-response” characteristics thereby avoiding any
possible overshoot of the setpoint. Heating System
There are, however, applications where the response time of the sen-
sor or the control valve is the decisive factor and for these applica-
tions M+W Instruments has developed a sensor with the following
Heater
features:
Flow
Response time (5τ): up to <100ms
Digital version
All our standard products are equipped with an analog pc-board, a
feature that ensures that they are very economically priced.
However, our well thought-out modular system allows us to offer a
digital pc-board as well thereby giving the options of analog volta-
ge or current output together with Profibus DP, Device Net, RS 232
or Flowbus protocols.
Flow Controller response times down to 0.5 sec and less are possi-
ble when using the "fast response” sensor.
10
Readout Systems Model D-15
Functions
◆ Power Supply for MFM/MFC ◆ Totalization
◆ Indication of flow rate ◆ Setpoint-potentiometer
Electrical data
◆ Power supply 110 or 230 Vac, ◆ Sub-D socket for instrument
50/60 Hz. connection
◆ Suitable for connection of instru- ◆ Max. power +24 Vac,
ments with output signal 0…5 Vac 0.5 A per channel