0% found this document useful (0 votes)
42 views28 pages

DHO1000U DataSheet

Uploaded by

Ninad Patange
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
42 views28 pages

DHO1000U DataSheet

Uploaded by

Ninad Patange
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd

Product Features

Product Features

• Built on RIGOL's "Centaurus" technical platform


• Ultra-low noise floor for pure signals, measuring low level signals accurately
• 12-bit vertical resolution[1]
• 200 MHz analog bandwidth, 2/4 analog channels, and 1 EXT channel
• Up to 2 GSa/s real-time sample rate
• Max. memory depth: 50 Mpts (optional)
• Vertical sensitivity range: 500 μV/div to 10 V/div
• Up to 500,000 wfms/s waveform capture rate in UltraAcquire mode
• 10.1'' 1280 x 800 HD touch display
• User-friendly Flex Knobs, bringing smoother interaction
• Standard photoelectric encoder operating knobs, effectively prolonging its service life
• Standard USB Device & Host, LAN, and HDMI interfaces

DHO1000U series digital oscilloscope is designed to meet the requirements for the design, debug,
and test of the mainstream oscilloscope market. Built on RIGOL's "Centaurus" technical platform,
this series delivers a fast waveform capture rate of 500,000 wfms/s in UltraAcquire mode, 50 Mpts
memory depth, 12-bit vertical resolution, all combined with excellent noise floor performance and
vertical accuracy to meet your requirements for more accurate measurements, bringing
extraordinary T&M experience for you.

NOTE:

[1]: 16 bits in High Resolution mode.

Page 1
RIGOL Probes and Accessories
Supported
Model Type Description

Passive High-impedance Probe

• Attenuation: 10:1/1:1
• 1X BW: DC to 35 MHz
Passive High-
impedance Probe • 10X BW: DC to 150 MHz
• Compatibility: All models of RIGOL's digital
PVP2150 oscilloscopes

• Attenuation: 10:1/1:1
• 1X BW: DC to 35 MHz
Passive High-
impedance Probe • 10X BW: DC to 350 MHz
• Compatibility: All models of RIGOL's digital
oscilloscopes
PVP2350

• Attenuation: 10:1/1:1
• 1X BW: DC to 20 MHz
Passive High-
impedance Probe • 10X BW: DC to 150 MHz
• Compatibility: All models of RIGOL's digital
PVP3150 oscilloscopes

• Attenuation: 10:1
• BW: DC to 500 MHz
Passive High- • Compatibility: MSO/DS4000, DS6000, MSO/
impedance Probe DS7000, MSO8000, MSO8000A,
DHO4000/1000, DS70000, and DS80000
RP3500A series

High-voltage Single-end Probe

• Attenuation: 1000:1
• BW: DC to 40 MHz
• DC: 0 to 10 kV DC
High-voltage Probe • AC: pulse ≤20 kVp-p
• AC: sine ≤7 kVrms
RP1010H
• Compatibility: All models of RIGOL's digital
oscilloscopes

Page 2
Model Type Description

• Attenuation: 1000:1
• BW: DC to 150 MHz
• DC+ACpeak: 18 kV CAT II
High-voltage Probe
• ACrms: 12 kV CAT II
RP1018H • Compatibility: All models of RIGOL's digital
oscilloscopes

• Attenuation: 100:1
• BW: DC to 300 MHz
• CAT I 2000 V (DC+AC)
High-voltage Probe
• CAT II 1500 V (DC+AC)
• Compatibility: All models of RIGOL's digital
RP1300H oscilloscopes

High-voltage Differential Probe

• BW: DC to 70 MHz
High-voltage • Max. voltage ≤ 1500 Vpp
Differential Probe • Compatibility: All models of RIGOL's digital
oscilloscopes
PHA0150

• BW: DC to 100 MHz


High-voltage • Max. voltage ≤ 1500 Vpp
Differential Probe • Compatibility: All models of RIGOL's digital
oscilloscopes
PHA1150

• 50X BW: DC to 160 MHz


• 500X BW: DC to 200 MHz
High-voltage
Differential Probe • Max. voltage ≤ 1500 Vpp
• Compatibility: All models of RIGOL's digital
PHA2150 oscilloscopes

• BW: DC to 25 MHz
High-voltage • Max. voltage ≤ 1400 Vpp (DC + AC P-P)
Differential Probe • Compatibility: All models of RIGOL's digital
oscilloscopes
RP1025D

• BW: DC to 50 MHz
High-voltage • Max. voltage ≤ 7000 Vpp (DC + AC P-P)
Differential Probe • Compatibility: All models of RIGOL's digital
oscilloscopes
RP1050D

Page 3
Model Type Description

• BW: DC to 100 MHz


High-voltage • Max. voltage ≤ 7000 Vpp (DC + AC P-P)
Differential Probe • Compatibility: All models of RIGOL's digital
oscilloscopes
RP1100D

Current Probe

• BW: DC to 300 kHz


• Maximum Input

AC: ±100 A
Current Probe AC P-P: 200 A

AC RMS: 70 A
RP1001C • Compatibility: All models of RIGOL's digital
oscilloscopes

• BW: DC to 1 MHz
• Maximum Input

AC: ±70 A
Current Probe AC P-P: 140 A

AC RMS: 50 A
RP1002C • Compatibility: All models of RIGOL's digital
oscilloscopes

• BW: DC to 50 MHz
• Maximum Input

AC P-P: 50 A (non-continuous)
Current Probe
AC RMS: 30 A
• Compatibility: All models of RIGOL's digital
RP1003C
oscilloscopes
• Required to order RP1000P power supply.

• BW: DC to 100 MHz


• Maximum Input

AC P-P: 50 A (non-continuous)
Current Probe
AC RMS: 30 A
• Compatibility: All models of RIGOL's digital
RP1004C
oscilloscopes
• Required to order RP1000P power supply.

Page 4
Model Type Description

• BW: DC to 10 MHz
• Maximum Input

AC P-P: 300 A (non-continuous), 500 A


(@pulse width ≤ 30 us)
Current Probe
AC RMS: 150 A
• Compatibility: All models of RIGOL's digital
RP1005C
oscilloscopes
• Required to order RP1000P power supply.

• BW: DC to 2 MHz
• Maximum Input

AC P-P: 700 A peaks, non-continuous


Current Probe
AC RMS: 500 A
• Compatibility: All models of RIGOL's digital
RP1006C oscilloscopes
• Required to order RP1000P power supply.

Power supply for RP1003C, RP1004C, RP1005C,


4CH Power Supply and RP1006C; supporting 4 channels.

RP1000P

Page 5
Specifications
All the specifications are guaranteed except the parameters marked with "Typical" and the
oscilloscope needs to operate for more than 30 minutes under the specified operation temperature.

Overview of the DHO1000U Series Technical Specifications


Overview of the DHO1000U Series Technical Specifications

Model DHO1202U DHO1204U

Analog Bandwidth (-3 dB) 200 MHz 200 MHz

No. of Input Channels 2 + EXT 4 + EXT

Rise Time
≤1.75 ns ≤1.75 ns
(10% to 90%, typical)

Sampling Mode Real-time Sampling

Two-channel model: 2 GSa/s (single-channel[1]), 1 GSa/s (full-


channel[3])
Max. Sample Rate of
four-channel model:
Analog Channel
2 GSa/s (single-channel[1]), 1 GSa/s (half-channel[2]), 500 MSa/s (full-
channel[3])

25 Mpts (single-channel[1]), 10Mpts (half-channel[2]), 1Mpts (full-


Standard Memory Depth
channel[3])

50 Mpts (single-channel[1]), 25Mpts (half-channel[2]), 10Mpts (full-


Optional Memory Depth
channel[3])

Max. Waveform Capture 30,000 wfms/s (in Vector Mode)


Rate 500,000 wfms/s (in UltraAcquire Mode)

Vertical Resolution 12 bits

Hardware Real-time
Waveform Recording and Max. 500,000 frames
Playing

Peak Detection Capture 2 ns glitches

LCD Size and Type 10.1 capacitive multi-touch screen

Display Resolution 1280×800

Page 6
Vertical System Analog Channel
Vertical System Analog Channel

Input Coupling DC, AC, or GND

Input Impedance 1 MΩ ± 1%

Input Capacitance 19 pF ± 3 pF

0.0001X, 0.0002X, 0.0005X, 0.001X, 0.002X, 0.005X,


Probe Attenuation 0.001X, 0.002X, 0.005X, 0.01X, 0.02X, 0.05X, 0.1X, 0.2X,
Coefficient 0.5X, 1X, 2X, 5X, 10X, 20X, 50X, 100X, 200X, 500X, 1000X,
2000X, and 5000X

CAT I 300 Vrms, 400 Vpk (DC + Vpeak)

No transient overvoltage allowed whether the probe is


Maximum Input Voltage used or not.
Remarks Please use the instrument dedicated for the specified
measurement category (not applicable to CAT II, III, and
IV).

Vertical Resolution 12 bits

Effective Number Of Bits[4]


>8
(ENOB, Typical)

Vertical Sensitivity Range[5] 500 μV/div to 10 V/div

±1 V (≤65 mV/div )

±10 V (>65 mV/div, ≤270 mV/div )


Offset Range
±20 V (>270 mV/div, ≤2.75 V/div )

±100 V (>2.75 V/div, ≤10 V/div )

Dynamic Range ±4 div (12 bits)

Bandwidth Limit (Typical) 20 MHz, FULL; selectable for each channel

DC Gain Accuracy[5] ± 2% of full scale

≤200 mV/div (±0.1 div ± 2 mV ± 1.5% of offset value)


DC Offset Accuracy
>200 mV/div (±0.1 div ± 2 mV ± 1.0% of offset value)

Channel-to-Channel
≥100:1
Isolation

ESD Tolerance ±8 kV (on input BNCs)

Page 7
Noise Floor
Noise floor, typical (sample rate 2 Gsa/s, memory depth 1 Mpts, timebase 20 μs)
500 μV/div 75 μVrms
1 mV/div 77 μVrms
2 mV/div 77 μVrms
5 mV/div 95 μVrms
10 mV/div 101 μVrms
20 mV/div 144 μVrms
50 mV/div 305 μVrms
100 mV/div 1.4mVrms
200 mV/div 1.8mVrms
500 mV/div 6mVrms
1 V/div 7.9mVrms
2 V/div 13mVrms
5 V/div 49mVrms
10 V/div 74mVrms

Horizontal System--Analog Channel


Horizontal System--Analog Channel

2 ns/div~1 ks/div
Range of Time Base
Fine

Time Base Resolution 400 ps

Time Base Accuracy ±5 ppm ± 1 ppm/year

Pre-
-5 div
trigger
Time Base Delay Range
Post- 1 s or 100 div, whichever is greater
trigger

±(Time Base Accuracy x Readout) ± (0.001 x Screen Width) ±


Delta Time Accuracy
50 ps

Channel-to-Channel Skew Channel-to-Channel Skew Correction Range ±100 ns,


Correction Accuracy ± 1 ps

Analog Channel-to-
≤2 ns[6]
Channel Delay (Typical)

Page 8
Horizontal System--Analog Channel

YT Default

XY Channel 1/2/3/4

Horizontal Mode SCAN Time base ≥ 200 ms/div

Time base ≥50 ms/div or ≥ 100 ms/div (selectable), available


ROLL to enter or exit the ROLL mode by rotating the Horizontal
SCALE knob

Acquisition System
Acquisition System

Two-channel model: 2 GSa/s (single-channel[1]), 1 GSa/s (full-channel[3])


Max. Sample Rate of
Analog Channel Four-channel model: 2 GSa/s (single-channel[1]), 1 GSa/s (half-channel[2]),
500 MSa/s (full-channel[3])

Optional: 50 Mpts (single-channel[1]), 25Mpts (half-channel[2]), 10Mpts (full-


channel[3])
Max. Memory Depth
of Analog Channel Standard: 25 Mpts (single-channel[1]), 10Mpts (half-channel[2]), 1Mpts (full-
channel[3])

Normal Default

Peak
capture 2 ns glitches
Detection

Acquisition Mode Average 2, 4, 8, 16…65536 are available for you to choose

High
14 bits, 16 bits
Resolution

UltraAcquire Waveform capture rate up to 500,000 wfms/s

Trigger System
Trigger System

Trigger Source Analog channel (CH1 to CH4), EXT TRIG, AC Line

Trigger Mode Auto, Normal, Single

Page 9
Trigger System

DC DC coupling trigger

AC AC coupling trigger

High
Trigger Coupling High frequency rejection, cut-off frequency to 75 kHz (internal
Frequency
trigger only)
Rejection

Low Frequency Low frequency rejection, cut-off frequency to 75 kHz (internal


Rejection trigger only)

Increases delay for the trigger circuit (internal trigger only),


Noise Rejection
On/Off

Holdoff Range 8 ns to 10 s

Internal
Analog Bandwidth
Trigger
Trigger
Bandwidth
External
200 MHz
Trigger

Internal 0.50 div, ≥50 mV/div;


Trigger 0.7 div (with noise rejection enabled)
Trigger Sensitivity
External 200 mVpp, DC to100 MHz
Trigger 500 mVpp, 100 MHz to 200 MHz

Input
1 MΩ±1%, BNC connector
Impedance

EXT TRIG <1 nsrms


Jitter
Normal acquisition, Edge trigger, trigger level located near 50%
(Typical)
of EXT input signal

Internal
± 5 div from the center of the screen
Trigger

Trigger Level
External
Range ±5 V
Trigger

AC Line fixed 40%-60%

Page 10
Trigger Type
Trigger Type

Edge trigger, Pulse trigger, Slope trigger, Video trigger, Pattern trigger,
Trigger Type Duration trigger, Timeout trigger, Runt trigger, Window trigger, Delay trigger,
Setup/Hold trigger, and Nth Edge trigger, RS232/UART, I2C, SPI, CAN, and LIN

Triggers on the threshold of the specified edge of the input signal. The edge
Edge types can be Rising, Falling, or Either.
Source channel: CH1 to CH4, EXT, or AC Line

Triggers on the positive or negative pulse with a specified width. The pulse
Pulse width is greater or smaller than a certain value or within a certain time range.
Source channel: CH1 to CH4.

Triggers on the positive or negative slope of the specified time. The slew time
Slope is greater or smaller than a certain value or within a certain time range.
Source channel: CH1 to CH4.

Triggers on all lines, specified line, add field, or even field that conforms to the
video standards. The supported video standards include NTSC, PAL/SECAM,
480p/60Hz, 576p/50Hz, 720p/60Hz, 720p/50Hz, 720p/30Hz, 720p/25Hz, 720p/
Video
24Hz, 1080p/60Hz, 1080p/50Hz, 1080p/30Hz, 1080p/25Hz, 1080p/24Hz, 1080i/
60Hz, and 1080i/50Hz.
Source channel: CH1 to CH4.

Identifies a trigger condition by searching for a specified pattern. The pattern is


a combination of multiple selected channel sources. The logic pattern of each
Pattern
channel is H, L, X, Rising, or Falling.
Source channel: CH1 to CH4.

Triggers when the specified pattern meets the specified duration condition.
The pattern is a combination of multiple selected channel sources. The logic
Duration pattern of each channel is H, L, and X. The duration is greater or smaller than a
certain value, or within a certain time range, or outside a certain time range.
Source channel: CH1 to CH4.

Triggers when duration of a certain event exceeds the specified time. The event
Timeout can be specified as Rising, Falling, or Either.
Source channel: CH1 to CH4.

Triggers when the pulses pass through one threshold but fail to pass through
Runt another threshold.
Source channel: CH1 to CH4.

Triggers in a specified window state when the rising edge of the signal crosses
the upper threshold or the falling edge crosses the lower threshold. The
Window
window state can be Enter, Exit, or Time.
Source channel: CH1 to CH4.

Page 11
Trigger Type

Triggers when the time difference between the specified edges of Source A
and Source B meets the preset time. The duration is greater or smaller than a
Delay
certain value, or within a certain time range, or outside a certain time range.
Source channel: CH1~CH4

When the setup time or hold time between the input clock signal and the data
Setup/Hold signal is smaller than the specified time.
Source channel: CH1 to CH4.

Triggers on the Nth edge that appears after the specified idle time. The edge
Nth Edge can be specified as Rising or Falling.
Source channel: CH1 to CH4.

Triggers on the Start, Error, Check Error, or Data frame of the RS232/UART bus
RS232/UART (up to 20 Mb/s).
Source channel: CH1 to CH4.

Triggers on the Start, Stop, Restart, MissedACK, Address (7 bits, 8 bits, or 10


I2C bits), Data, or Address Data of the I2C bus.
Source channel: CH1 to CH4.

Triggers on the specified pattern of the specified data width (4~32) of SPI bus.
SPI CS and Timeout are supported.
Source channel: CH1 to CH4.

Triggers on the start of a frame, end of a frame, Remote ID, Overload, Frame ID,
Frame Data, Data&ID, Frame Error, Answer Error, Check Error, Format Error, Bit
CAN Fill, and Random of the CAN signal (up to 5Mb/s). The supported CAN bus
signal types include CAN_H, CAN_L, TX/RX, and DIFF.
Source channel: CH1 to CH4.

Triggers on the Sync, ID, Data (length settable), Data&ID, Wakeup, Sleep, and
LIN Error of the LIN bus signal (up to 20 Mb/s).
Source channel: CH1 to CH4.

Search&Navigation
Search&Navigation

Type Edge, Pulse

Source Analog channel

Copies the search settings from or to the trigger settings mutually, including
Copy
threshold setting and search condition settings

Displays in event table form; can be exported to the external or internal


Result Display
memory

Page 12
Search&Navigation

Time navigation: navigates to the acquired waveforms in time order.

Event navigation: uses the navigation keys to scroll through the event search
Navigation results and navigates to the specified event.

Frame navigation: navigates to the specified frame segment in UltraAcquire


mode.

Waveform Measurement
Waveform Measurement

Number of
2 pairs of XY cursors
Cursors

Voltage deviation between cursors (∆Y)

Manual Mode Time deviation between cursors (∆X)

Reciprocal of ∆X (Hz) (1/∆X)

Fix Y-axis to track X-axis waveform point's voltage and time


values
Cursor Track Mode
Fix X-axis to track Y-axis waveform point's voltage and time
values

Auto
Allows to display cursors during auto measurement
Measurement

Measures the voltage parameters of the corresponding channel


XY Mode waveforms in XY time base mode.

X = Channel 1, Y = Channel 2

Page 13
Waveform Measurement

Number of 41 auto measurements; and up to 14 measurements can be


Measurements displayed at a time.

Measurement
CH1-CH4, Math1-Math4
Source

Measurement
Main, Zoom
Range (Region)

Displays 33 measurement items (vertical and horizontal) for the


All
current measurement channel; the measurement results are
Measurement
updated continuously.
Auto
Measurement Vmax, Vmin, Vpp, Vtop, Vbase, Vamp, Vupper, Vmid, Vlower, Vavg,
Vertical
VRMS, Per. VRMS, Overshoot, Preshoot, Area, and Period Area.

Period, Frequency, Rise Time, Fall Time, +Width, -Width, +Duty, -


Duty, Positive Pulse Count, Negative Pulse Count, Rising Edge
Horizontal
Count, Falling Edge Count, Tvmax, Tvmin, +Slew Rate, and -Slew
Rate

Delay(A↑-B↑), Delay(A↑-B↓), Delay(A↓-B↑), Delay(A↓-B↓), Phase(A↑-


Others
B↑), Phase(A↑-B↓), Phase(A↓-B↑), and Phase(A↓-B↓)

Items: Current, Average, Max, Min, Standard Deviation, Count


Statistics
Statistical times settable

Waveform Calculation
Waveform Calculation

No. of Math Functions 4; 4 math functions available to be displayed at a time

A+B, A-B, A×B, A/B, FFT, A&&B, A||B, A^B, !A, Intg, Diff, Sqrt, Lg,
Operation
Ln, Exp, Abs, AX+B, LowPass, HighPass, BandPass, and BandStop

Color Grade FFT supported

Record Length Max. 1 Mpts

Rectangular, Blackman-Harris, Hanning (default), Hamming,


Window Type
FFT Flattop, and Triangle.

a maximum of 15 peaks, confirmed by the settable threshold and


Peak Search
offset threshold set by users

Page 14
Waveform Analysis
Waveform Analysis

Stores the signal under test in segments according to the trigger


events, that is, saves all the sampled waveform data as a segment
to the RAM for each trigger event. The maximum number of the
sampled segments reaches 500,000.
Waveform
Recording
Source All enabled analog channels

Support playing frame by frame or continuous playing; capable


Analysis
of calculating, measuring, and decoding the played waveforms

Compares the signal under test with the user-defined mask to


provide the test results: the number of successful tests, failed
tests, and the total number of tests. The pass/fail event can
PassFail
enable immediate stop, beeper, and the screenshot.

Source Any analog channel

Provide a dimensional view for color grade waveforms, color


grade >16, 256-level color scale display

Source Any analog channel


Color Grade

Color Theme Temperature and intensity

Mode Supports all modes

Serial Decoding
Serial Decoding

Number of
4, four protocol types can be decoded and enabled at the same time
Decodings

Decoding Type Standard: Parallel, RS232/UART, I2C, SPI, LIN, and CAN

Up to 4 bits of Parallel decoding, supporting any analog channel Support user-


Parallel defined clock and auto clock settings.

Source channel: CH1~CH4

Decodes the RS232/UART (up to 20 Mb/s) bus's TX/RX data (5-9 bits), parity
RS232/UART (Odd, Even, or None), and stop bits (1-2 bits)

Source channel: CH1~CH4

Page 15
Serial Decoding

Decodes the address (with or without the R/W bit) of the I2C bus, data, and
I2C ACK.

Source channel: CH1~CH4

Decodes the MISO/MOSI data (4-32 bits) of the SPI bus. The available mode
SPI includes "Timeout" and "CS".

Source channel: CH1~CH4

Decodes the remote frame (ID, byte number, CRC), overload frame, and data
frame (standard/extended ID, control domain, data domain, CRC, and ACK) of
CAN the CAN bus (up to 5 Mb/s). The supported CAN bus signal types include
CAN_H, CAN_L, TX/RX, and DIFF.

Source channel: CH1~CH4

Decodes the protocol version (1.X or 2.X) of the LIN bus (up to 20 Mb/s). The
LIN decoding displays sync, ID, data, and check sum.

Source channel: CH1~CH4

Auto
Auto

AutoScale Min voltage > 10 mVpp, duty cycle > 1%, frequency > 35 Hz

Digital Voltmeter
Digital Voltmeter

Source Any analog channel

Function DC, AC+DCrms, ACrms

Resolution ACV/DCV: 4 bits

Limits Beeper Sounds an alarm when the voltage value is within or outside of the limit range

High-precision Frequency Counter


High-precision Frequency Counter

Source Any analog channel and EXT

Measure Frequency, period, totalizer

Page 16
High-precision Frequency Counter

Resolution 3-6 digits, user-defined


Counter
Max. Frequency Max. analog bandwidth

48-bit totalizer
Totalizer
Counts the number of the rising edges

Time Reference Internal reference

Command Set
Command Set

Common Commands Support IEEE488.2 Standard

Error Message Definition Error messages

Support Status Report Mechanism Status Reporting

Support Syn Mechanism Synchronization

Display
Display

LCD 10.1-inch capacitive multi-touch screen, gesture enabled


operation

Resolution 1280×800 (Screen Region) 16:9

Graticule (10 vertical divisions) x(8 horizontal divisions)

Persistence Off, Infinite, variable persistence (100 ms to 10 s)

Brightness 256 intensity levels (LCD, HDMI)

Processor System
Processor System

Processor Cortex-A72 up to 1.8 GHz, 6-core processor

System Memory 4 GB RAM

Operating System Android

Internal Non-volatile Memory 8 GB

Page 17
I/O
I/O

USB3.0 Host 2 on the front panel

USB3.0 Device 1 on the rear panel

LAN 1 on the rear panel, 10/100/1000 Base-T, supporting LXI-C

Supports Web Control interface (input the IP address of


Web Remote
the oscilloscope into the Web browser to display the
Control
operation interface of the oscilloscope)

BNC output on the rear panel.

Vo (H) ≥ 2.5 V open circuit, ≥1.0 V 50 Ω to GND

Vo (L) ≤ 0.7 V to load ≤4 mA, ≤0.25 V 50 Ω to GND

Trig Out Outputs a pulse signal when the oscilloscope is triggered


AUX Out

Outputs a pulse signal when a pass/fail event occurs.


Pass/Fail Supports user-defined pulse polarity and pulse time (100
ns to 10 ms)

Rise Time ≤1.5 ns

Input Interface 1, BNC connector on the rear panel

Output Interface 1, BNC connector on the rear panel


10 MHz REF

IN/OUT 50 Ω, with the amplitude 130 mVpp to 4.1 Vpp (-10 dBm,
Input Interface
20 dBm), frequency 10 MHz ± 10 ppm

Output Interface 50 Ω, 1.5 Vpp sine waveform

HDMI 1 on the rear panel, HDMI 1.4, A plug. Used to connect to


Video Output an external monitor or projector

Probe
Compensation 1 kHz frequency, 0.3 V amplitude, Square
Output

Power Supply
Power Supply

Power Voltage AC 100 V to 240 V, 50 Hz to 60 Hz

Power 400 VA max (connect to various interfaces, USB, active probes)

Page 18
Power Supply

Fuse 3.15 A, T degree, 250 V

Environment
Environment

Operating 0℃ to +50℃
Temperature
Range
Non-operating -30℃ to +60℃

below +30℃: ≤90% RH (without condensation)

Operating +30℃ to +40℃, ≤75% RH (without condensation)


Humidity Range
+40℃ to +50℃, ≤45% RH (without condensation)

Non-operating below 60℃: ≤90% RH (without condensation)

Operating below 3,000 m


Altitude
Non-operating Below 15,000 m

Warranty and Calibration Interval


Warranty and Calibration Interval

Warranty Three years for the mainframe, excluding the probes and accessories.

Recommended
18 months
Calibration Interval

Page 19
Regulations
Regulations

Compliant with EMC DIRECTIVE 2014/30/EU, compliant with or higher than


the standards specified in IEC 61326-1:2013/EN 61326-1:2013 Group 1 Class
A

CISPR 11/EN 55011

IEC 61000-4-2:2008/EN ±4.0 kV (contact discharge), ±8.0 kV (air


61000-4-2 discharge)

IEC 61000-4-3:2002/EN 3 V/m (80 MHz to 1 GHz); 3 V/m (1.4 GHz to


61000-4-3 2 GHz); 1 V/m (2.0 GHz to 2.7 GHz)

Electromagnetic IEC 61000-4-4:2004/EN


Compatibility 1 kV power line
61000-4-4

0.5 kV (phase-to-neutral voltage); 1 kV


IEC 61000-4-5:2001/EN
(phase-to-earth voltage); 1 kV (neutral-to-
61000-4-5
earth voltage)

IEC 61000-4-6:2003/EN
3 V, 0.15-80 MHz
61000-4-6

Voltage dip: 0% UT during half cycle; 0% UT


IEC 61000-4-11:2004/EN during 1 cycle ; 70% UT during 25 cycles
61000-4-11
short interruption: 0% UT during 250 cycles

EN 61010-1:2019

EN 61010-031:2015

IEC 61010-1:2016

IEC 61010-2-030:2017

Safety UL 61010-1:2012 R7

UL 61010-2-31:2017 R2

CAN/CSA-22.2 No. 61010-1-12:2017

CAN/CSA-22.2 No. 61010-2-30:2018

CAN/CSA-22.2 No. 61010-031-07:201

Meets GB/T 6587; class 2 random


Vibration
Meets MIL-PRF-28800F and IEC60068-2-6; class 3 random

Page 20
Regulations

Meets GB/T 6587-2012; class 2 random

Meets MIL-PRF-28800F and IEC 60068-2-27; class 3 random


Shock
In non-operating conditions: 30 g, half-sine wave, 11 ms duration, 3 shocks
along the main axis, total of 18 shocks

Mechanical Characteristics
Mechanical Characteristics

Dimensions 358.14 mm (W) x 214.72 mm (H) x 120.62 mm (D)

Rack Mount Kit 5U

Package excluded: 3.8 kg


Weight[7]
Package included: 5.37 kg

Non-volatile Memory
Non-volatile Memory

Setup/Image setup (*.stp), image (*.png, *.bmp, *.jpg)

Data/File Storage CSV waveform data (*.csv), binary waveform data


Waveform Data (*.bin,), list data (*.csv), and reference waveform data
(*.ref, *.csv, *.bin)

Internal Capacity 8 GB

Reference Waveform Displays 10 internal waveforms

Setting Storage is limited by the capacity

Supports the USB storage device that conforms to the


USB Capacity
industry standard

NOTE:

[1]: Single-channel mode: If any one of the channels is enabled, it is called single-channel mode.

[2]: Half-channel mode: For four-channel models, if any two of the channels are enabled, it is called
half-channel mode.

[3]: Full-channel mode: For two-channel models, if all of the two channels are enabled, it is called
full-channel mode; for four-channel models, if any three channels or all of the four channels are
enabled, it is called full-channel mode.

[4]: 11.23046875 MHz, Full Scale, 20 mV/div.

Page 21
[5]: 500 μV/div is a magnification of 1 mV/div setting. For vertical accuracy calculations, use full scale
of 8 mV.

[6]: For any channel, under the same input impedance with DC-coupled, the Volts/div setting is the
same for 100mV/div and 200mV/div.

[7]: Standard configuration.

Page 22
Order Information and Warranty
Period

Order Information

Order Information Order No.

Model
200 MHz, 2 GSa/s, 25 Mpts, 2CH DHO DHO1202U
200 MHz, 2 GSa/s, 25 Mpts, 4CH DHO DHO1204U
Standard Accessories
Power Cord Conforming to the Standard of the Destination ——
Country
USB Cable ——
4 Passive HighZ Probes (350 MHz) Standard for DHO1204U, 2 PVP2350
Passive HighZ Probes (350 MHz) Standard for DHO1202U
Memory Depth Upgrade Option
50 Mpts Memory Depth Upgrade Option DHO1000U-RLU

NOTE:

For all the mainframes, accessories, and options, please contact the local office of RIGOL.

Warranty Period

Three years for the mainframe, excluding the probes and accessories.

Page 23
Option Ordering and Installation
Process

1. According to the usage requirements, please purchase the specified function options from RIGOL

Sales Personnel, and provide the serial number of the instrument that needs to install the option.

2. After receiving the option order, the RIGOL factory will mail the paper software product

entitlement certificate to the address provided in the order.

3. Log in to RIGOL official website for registration. Use the software key and instruments serial

number provided in the entitlement certificate to obtain the option license code and the option

license file.

4. Download the option license file to the root directory of the USB storage device, and connect the

USB storage device to the instrument properly. After the USB storage device is successfully

recognized, the Option install menu is activated. Press this menu key to start installing the

option.

Page 24

You might also like