Electro-Tek Multimeter Calibration Data
Electro-Tek Multimeter Calibration Data
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
Permanent Facility
ELECTRO-
TECHNICAL-
Using 8½ Digit
Alternating AC Current (10 Hz to
1 Precision Multimeter 1 mA to 10 mA 0.06 % to 0.04 %
Current (< 1 10 kHz)
By Direct Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Using 8½ Digit
Alternating AC Current (10 Hz to
2 Precision Multimeter 10 mA to 1 A 0.06 % to 0.07 %
Current (< 1 10 kHz)
By Direct Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Using 8½ Digit
Alternating AC Current (10 Hz to
3 Precision Multimeter 1 A to 20 A 0.01 % to 0.02 %
Current (< 1 5 kHz)
By Direct Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Using 8½ Digit
Alternating AC Current (300 Hz
4 Precision Multimeter 100 µA to 1 mA 0.05 % to 0.06 %
Current (< 1 to 1 kHz)
By Direct Method
GHz)
(Measure)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 2 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
TECHNICAL- Using AC
Alternating AC Voltage (10 Hz to measurement
5 1 V to 100 V 0.03 % to 0.01 %
Current (< 1 40 Hz) Standard By Direct
GHz) Method
(Measure)
ELECTRO-
TECHNICAL- Using AC
Alternating AC Voltage (10 Hz to measurement
6 10 mV to 1 V 0.04 %
Current (< 1 40 Hz) Standard By Direct
GHz) Method
(Measure)
ELECTRO-
TECHNICAL- Using AC
Alternating AC Voltage (10 kHz measurement
7 1 V to 10 V 0.06 %
Current (< 1 to 100 kHz) Standard By Direct
GHz) Method
(Measure)
ELECTRO-
TECHNICAL- Using AC
Alternating AC Voltage (10 kHz measurement
8 10 mV to 1 V 0.4 % to 0.06 %
Current (< 1 to 100 kHz) Standard By Direct
GHz) Method
(Measure)
ELECTRO-
TECHNICAL- Using AC
Alternating AC Voltage (10 kHz measurement
9 10 V to 500 V 0.06 % to 0.07 %
Current (< 1 to 100 kHz) Standard By Direct
GHz) Method
(Measure)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 3 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
TECHNICAL- Using AC
Alternating AC Voltage (100 kHz measurement
10 10 mV to 20 V 0.4 %
Current (< 1 to 1 MHz) Standard By Direct
GHz) Method
(Measure)
ELECTRO-
TECHNICAL- Using AC
Alternating AC Voltage (40 Hz to measurement
11 1 V to 1000 V 0.013 % to 0.27 %
Current (< 1 20 kHz) Standard By Direct
GHz) Method
(Measure)
ELECTRO-
TECHNICAL- Using AC
Alternating AC Voltage (40 Hz to measurement
12 10 mV to 1 V 0.04 % to 0.013 %
Current (< 1 20 kHz) Standard By Direct
GHz) Method
(Measure)
ELECTRO-
TECHNICAL-
Using RLC Digi
Alternating
13 Capacitance @1 kHz bridge by Direct 10 nF to 100 nF 0.07 % to 0.04 %
Current (< 1
Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Using RLC Digi
Alternating
14 Capacitance @1 kHz bridge by Direct 10 pF to 100 pF 0.22 %
Current (< 1
Method:
GHz)
(Measure)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 4 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
TECHNICAL-
Using RLC Digi
Alternating
15 Capacitance @1 kHz bridge by Direct 100 nF to 1000 nF 0.04 % to 0.03 %
Current (< 1
Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Using RLC Digi
Alternating
16 Capacitance @1 kHz bridge by Direct 100 pF to 10 nF 0.2 % to 0.07 %
Current (< 1
Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Using RLC Digi
Alternating
17 Capacitance @1 kHz bridge by Direct 1000 nF to 100 µF 0.03 % to 0.3 %
Current (< 1
Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Using RLC Digi
Alternating
18 Inductance @1 kHz bridge by Direct 1 H to 10 H 0.06 % to 0.04 %
Current (< 1
Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Using RLC Digi
Alternating
19 Inductance @1 kHz bridge by Direct 10 mH to 1 H 0.2 % to 0.06 %
Current (< 1
Method
GHz)
(Measure)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 5 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
TECHNICAL-
Using RLC Digi
Alternating
20 Inductance @1 kHz bridge by Direct 100 µH to 10 mH 0.2 % to 0.2 %
Current (< 1
Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Alternating Using Phase meter
21 Phase Angle -180 ° to +180 ° 0.03 °
Current (< 1 By Direct Method
GHz)
(Measure)
ELECTRO-
TECHNICAL- Using Multiproduct
AC Current (1 kHz to
22 Alternating Calibrator by Direct 20 mA to 10 A 0.05 % to 0.1 %
10 kHz)
Current (< 1 Method
GHz) (Source)
ELECTRO-
TECHNICAL- Using Multiproduct
AC Current (10 Hz to
23 Alternating Calibrator by Direct 200 µA to 100 mA 0.11 % to 0.05 %
40 Hz)
Current (< 1 Method
GHz) (Source)
ELECTRO-
TECHNICAL- Using Multiproduct
AC Current (40 Hz to
24 Alternating Calibrator by Direct 1 A to 10 A 0.05 % to 0.06 %
1 kHz)
Current (< 1 Method
GHz) (Source)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 6 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
TECHNICAL- Using Multiproduct
AC Current (40 Hz to
25 Alternating Calibrator by Direct 1 mA to 100 mA 0.027 % to 0.031 %
1 kHz)
Current (< 1 Method
GHz) (Source)
ELECTRO-
TECHNICAL- Using Multiproduct
AC Current (40 Hz to
26 Alternating Calibrator by Direct 10 A to 20 A 0.06 % to 0.2 %
1 kHz)
Current (< 1 Method
GHz) (Source)
ELECTRO-
TECHNICAL- Using Multiproduct
AC Current (40 Hz to
27 Alternating Calibrator by Direct 100 µA to 1 mA 0.1 % to 0.036 %
1 kHz)
Current (< 1 Method
GHz) (Source)
ELECTRO-
TECHNICAL- Using Multiproduct
AC Current (40 Hz to
28 Alternating Calibrator by Direct 100 mA to 1 A 0.031 % to 0.041 %
1 kHz)
Current (< 1 Method
GHz) (Source)
ELECTRO-
TECHNICAL- Using Multiproduct
AC Current (5 kHz to
29 Alternating Calibrator by Direct 20 mA to 10 A 0.03 % to 0.4 %
10 kHz)
Current (< 1 Method
GHz) (Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
Calibrator and
30 Alternating AC Current @50 Hz 20 A to 1000 A 0.73 % to 0.8 %
current coil by Direct
Current (< 1
Method
GHz) (Source)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 7 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
TECHNICAL- Using multiproduct
AC Voltage (1 kHz to
31 Alternating calibrator by Direct 100 V to 500 V 0.03 % to 0.27 %
100 kHz)
Current (< 1 method
GHz) (Source)
ELECTRO-
TECHNICAL- Using multiproduct
AC Voltage (1 kHz to
32 Alternating calibrator by Direct 2 mV to 2 V 0.45 % to 0.015 %
100 kHz)
Current (< 1 method
GHz) (Source)
ELECTRO-
TECHNICAL- Using multiproduct
AC Voltage (1 kHz to
33 Alternating calibrator by Direct 2 V to 100 V 0.015 % to 0.02 %
100 kHz)
Current (< 1 method
GHz) (Source)
ELECTRO-
TECHNICAL- Using multiproduct
AC Voltage (10 Hz to
34 Alternating calibrator by Direct 1 mV to 1 V 0.79 % to 0.052 %
40 Hz)
Current (< 1 method
GHz) (Source)
ELECTRO-
TECHNICAL- Using multiproduct
AC Voltage (10 Hz to
35 Alternating calibrator by Direct 2 mV to 20 mV 0.26 % to 0.05 %
40 Hz)
Current (< 1 method
GHz) (Source)
ELECTRO-
TECHNICAL- Using multiproduct
AC Voltage (10 Hz to
36 Alternating calibrator by Direct 20 mV to 20 V 0.05 % to 0.03 %
40 Hz)
Current (< 1 method
GHz) (Source)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 8 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
TECHNICAL- Using multiproduct
AC Voltage (10 Hz to
37 Alternating calibrator by Direct 20 V to 200 V 0.03 % to 0.025 %
40 Hz)
Current (< 1 method
GHz) (Source)
ELECTRO-
TECHNICAL- Using Multiproduct
AC Voltage (100 kHz
38 Alternating Calibrator by Direct 2 mV to 20 mV 1.75 % to 0.52 %
to 1 MHz)
Current (< 1 Method
GHz) (Source)
ELECTRO-
TECHNICAL- Using Multiproduct
AC Voltage (100 kHz
39 Alternating Calibrator by Direct 20 mV to 200 mV 0.46 % to 0.4 %
to 1 MHz)
Current (< 1 Method
GHz) (Source)
ELECTRO-
TECHNICAL- Using Multiproduct
AC Voltage (100 kHz
40 Alternating Calibrator by Direct 200 mV to 20 V 0.4 % to 0.25 %
to 1 MHz)
Current (< 1 Method
GHz) (Source)
ELECTRO-
TECHNICAL- Using multiproduct
AC Voltage (40 Hz to
41 Alternating calibrator by Direct 1 mV to 2 mV 0.7 % to 0.3 %
1 kHz)
Current (< 1 method
GHz) (Source)
ELECTRO-
TECHNICAL- Using multiproduct
AC Voltage (40 Hz to
42 Alternating calibrator by Direct 2 mV to 20 mV 0.3 % to 0.03 %
1 kHz)
Current (< 1 method
GHz) (Source)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 9 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
TECHNICAL- Using multiproduct
AC Voltage (40 Hz to
43 Alternating calibrator by Direct 2 V to 20 V 0.006 % to 0.013 %
1 kHz)
Current (< 1 method
GHz) (Source)
ELECTRO-
TECHNICAL- Using multiproduct
AC Voltage (40 Hz to
44 Alternating calibrator by Direct 20 mV to 200 mV 0.03 % to 0.015 %
1 kHz)
Current (< 1 method
GHz) (Source)
ELECTRO-
TECHNICAL- Using multiproduct
AC Voltage (40 Hz to
45 Alternating calibrator by Direct 20 V to 1000 V 0.013 % to 0.038 %
1 kHz)
Current (< 1 method
GHz) (Source)
ELECTRO-
TECHNICAL- Using multiproduct
AC Voltage (40 Hz to
46 Alternating calibrator by Direct 200 mV to 2 V 0.015 % to 0.013 %
1 kHz)
Current (< 1 method
GHz) (Source)
ELECTRO-
TECHNICAL- Using Capacitor
47 Alternating Capacitance @1 kHz Standard by Direct 1 µF to 100 mF 0.58 % to 0.3 %
Current (< 1 Method
GHz) (Source)
ELECTRO-
TECHNICAL- Using Capacitor
48 Alternating Capacitance @1 kHz Standard by Direct 10 nF to 1 µF 0.58 %
Current (< 1 Method
GHz) (Source)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 10 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
TECHNICAL- Using Capacitor
49 Alternating Capacitance @1 kHz Standard by Direct 100 mF to 1 F 0.3 % to 0.6 %
Current (< 1 Method
GHz) (Source)
ELECTRO-
TECHNICAL- Using Capacitor
50 Alternating Capacitance @1 kHz Standard by Direct 220 pF to 10 nF 5.8 % to 0.58 %
Current (< 1 Method
GHz) (Source)
ELECTRO-
TECHNICAL- Using Standard
51 Alternating Inductance @1 kHz Inductors series by 100 µH to 10 H 0.3 % to 0.2 %
Current (< 1 direct method
GHz) (Source)
ELECTRO- Single Phase AC
TECHNICAL- Power (50 Hz to 1 Using Multiproduct
52 Alternating kHz, 120 V to 240 V Calibrator by Direct 9.6 W to 3.84 kW 0.3 %
Current (< 1 , 0.1 A to 20 A @ 0.8 Method
GHz) (Source) PF)
ELECTRO- Single Phase AC
TECHNICAL- Power (50 Hz to 1 Using Multiproduct
53 Alternating kHz, 120 V to 240 V, Calibrator by Direct 2.4 W to 960 W 0.9 %
Current (< 1 0.1 A to 20 A @0.2 Method
GHz) (Source) PF)
ELECTRO- Single Phase AC
TECHNICAL- Power (50 Hz to 1 Using Multiproduct
54 Alternating kHz, 120 V to 240 V, Calibrator by Direct 6 W to 2.4 kW 0.5 %
Current (< 1 0.1 A to 20 A @0.5 Method
GHz) (Source) PF)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 11 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
Single Phase AC
TECHNICAL- Using Multiproduct
Power (50 Hz to 1
55 Alternating Calibrator by Direct 12 W to 4.8 kW 0.2 %
kHz, 120 V to 240 V,
Current (< 1 Method
0.1 A to 20 A @UPF)
GHz) (Source)
ELECTRO-
TECHNICAL- Using 8½ Digit
56 DIRECT DC Current Precision Multimeter 1 µA to 100 µA 0.02 % to 0.003 %
CURRENT By Direct Method
(Measure)
ELECTRO-
TECHNICAL- Using 8½ Digit
57 DIRECT DC Current Precision Multimeter 1 A to 20 A 0.004 % to 0.074 %
CURRENT By Direct Method
(Measure)
ELECTRO-
TECHNICAL- Using 8½ Digit
58 DIRECT DC Current Precision Multimeter 100 µA to 1 A 0.003 % to 0.004 %
CURRENT By Direct Method
(Measure)
ELECTRO-
TECHNICAL- Using 8½ Digit
59 DIRECT DC Current Precision Multimeter 100 nA to 1 µA 0.05 % to 0.02 %
CURRENT By Direct Method
(Measure)
ELECTRO-
Using 8½ Digit
TECHNICAL-
Precision Multimeter
60 DIRECT DC Current 20 A to 100 A 0.35 %
and current shunt by
CURRENT
direct method
(Measure)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 12 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
TECHNICAL- Using 8½ Digit
61 DIRECT DC Voltage Precision Multimeter 1 mV to 10 mV 0.02 % to 0.008 %
CURRENT By Direct Method
(Measure)
ELECTRO-
TECHNICAL- Using 8½ Digit
62 DIRECT DC Voltage Precision Multimeter 10 mV to 100 mV 0.008 % to 0.0008 %
CURRENT By Direct Method
(Measure)
ELECTRO-
TECHNICAL- Using 8½ Digit
63 DIRECT DC Voltage Precision Multimeter 100 µV to 1 mV 0.35 % to 0.02 %
CURRENT By Direct Method
(Measure)
ELECTRO-
TECHNICAL- Using 8½ Digit
64 DIRECT DC Voltage Precision Multimeter 100 mV to 1000 V 0.0008 %
CURRENT By Direct Method
(Measure)
ELECTRO-
TECHNICAL- Using 8½ Digit
10 Mohm to 100
65 DIRECT Resistance (2 wire) Precision Multimeter 0.004 % to 0.015 %
Mohm
CURRENT by direct method
(Measure)
ELECTRO-
TECHNICAL- Using 8½ Digit
100 kohm to 10
66 DIRECT Resistance (2 wire) Precision Multimeter 0.0034 % to 0.004 %
Mohm
CURRENT by direct method
(Measure)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 13 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
TECHNICAL- Using 8½ Digit
100 Mohm to 1
67 DIRECT Resistance (2 wire) Precision Multimeter 0.015 % to 0.28 %
Gohm
CURRENT by direct method
(Measure)
ELECTRO-
TECHNICAL- Using 8½ Digit
68 DIRECT Resistance (4 wire) Precision Multimeter 1 ohm to 10 ohm 0.004 % to 0.002 %
CURRENT by direct method
(Measure)
ELECTRO-
TECHNICAL- Using 8½ Digit
69 DIRECT Resistance (4 wire) Precision Multimeter 10 ohm to 100 ohm 0.002 % to 0.001 %
CURRENT by direct method
(Measure)
ELECTRO-
TECHNICAL- Using 8½ Digit
100 ohm to 100
70 DIRECT Resistance (4 wire) Precision Multimeter 0.001 % to 0.0034 %
kohm
CURRENT by direct method
(Measure)
ELECTRO-
TECHNICAL- Using Multiproduct
71 DIRECT DC Current Calibrator by Direct 1 A to 10 A 0.011 % to 0.06 %
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Using Multiproduct
72 DIRECT DC Current Calibrator by Direct 1 mA to 1 A 0.0051 % to 0.011 %
CURRENT Method
(Source)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 14 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
TECHNICAL- Using Multiproduct
73 DIRECT DC Current Calibrator by Direct 10 µA to 100 µA 0.74 % to 0.012 %
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Using Multiproduct
74 DIRECT DC Current Calibrator by Direct 10 A to 20 A 0.06 % to 0.2 %
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Using multiproduct
75 DIRECT DC Current calibrator by direct 100 µA to 1 mA 0.012 % to 0.0051 %
CURRENT method
(Source)
ELECTRO-
Using multiproduct
TECHNICAL-
calibrator and
76 DIRECT DC Current 20 A to 1000 A 0.35 % to 0.77 %
current coil by direct
CURRENT
method
(Source)
ELECTRO-
TECHNICAL- Using Multifunction
77 DIRECT DC Power calibrator with Direct 1 W to 100 W 0.03 %
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Using Multifunction
78 DIRECT DC Power calibrator with Direct 10 mW to 100 mW 0.03 %
CURRENT Method
(Source)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 15 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
TECHNICAL- Using Multifunction
79 DIRECT DC Power calibrator with Direct 100 mW to 1 W 0.03 %
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Using Multifunction
80 DIRECT DC Power calibrator with Direct 100 W to 1000 W 0.03 % to 0.08 %
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Using Multifunction
81 DIRECT DC Power calibrator with Direct 1000 W to 10000 W 0.08 %
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Using Multifunction
10000 W to 19950
82 DIRECT DC Power calibrator with Direct 0.08 %
W
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Using Multiproduct
83 DIRECT DC Voltage calibrator by direct 1 mV to 100 mV 0.05 % to 0.001 %
CURRENT method
(Source)
ELECTRO-
TECHNICAL- Using Multiproduct
84 DIRECT DC Voltage Calibrator by Direct 100 mV to 100 V 0.001 % to 0.0007 %
CURRENT Method
(Source)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 16 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
TECHNICAL- Using multiproduct
0.0007 % to 0.0009
85 DIRECT DC Voltage calibrator by Direct 100 V to 1000 V
%
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Using Multiproduct
86 DIRECT Resistance Calibrator by Direct 1 ohm to 10 Mohm 0.013 % to 0.0099 %
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Using Standard
87 DIRECT Resistance (2 wire) resistor by Direct 1 Gohm 0.03 %
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Using Standard
88 DIRECT Resistance (2 wire) resistor by Direct 10 Mohm 0.0013 %
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Using Multiproduct
10 Mohm to 100
89 DIRECT Resistance (2 wire) Calibrator by Direct 0.0099 % to 0.015 %
Mohm
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Using Multiproduct
100 Mohm to 1
90 DIRECT Resistance (2 wire) Calibrator by Direct 0.15 % to 1.79 %
Gohm
CURRENT Method
(Source)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 17 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
TECHNICAL- Using Standard
91 DIRECT Resistance (4 wire) resistor by Direct 1 kohm 0.0009 %
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Using Standard
92 DIRECT Resistance (4 wire) resistor by Direct 1 Mohm 0.0008 %
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Using Standard
93 DIRECT Resistance (4 wire) resistor by Direct 1 ohm 0.0005 %
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Using Standard
94 DIRECT Resistance (4 wire) resistor by Direct 10 kohm 0.0008 %
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Using Standard
95 DIRECT Resistance (4 wire) resistor by Direct 10 ohm 0.00044 %
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Using Standard
96 DIRECT Resistance (4 wire) resistor by Direct 100 kohm 0.0009 %
CURRENT Method
(Source)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 18 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
TECHNICAL- Using Standard
97 DIRECT Resistance (4 wire) resistor by Direct 100 ohm 0.0006 %
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Using Oscilloscope
Oscilloscope (Time 0.0001 % to 0.00035
98 ELECTRICAL Calibrator with Head 1 ns to 5 s
base) %
EQUIPMENT by Direct Method
(Source)
ELECTRO- Using Multifunction
TECHNICAL- Oscilloscope - Calibrator and
99 ELECTRICAL Amplitude (DC Oscilloscope 1 mV to 100 mV 0.1 % to 0.06 %
EQUIPMENT Signal) Calibrator with Head
(Source) by Direct Method
ELECTRO-
TECHNICAL- Oscilloscope - Using Oscilloscope
100 ELECTRICAL Amplitude (DC Calibrator with Head 100 mV to 190 V 0.06 % to 0.04 %
EQUIPMENT Signal) by Direct Method
(Source)
ELECTRO-
Amplitude
TECHNICAL-
Modulation With Using Measuring
RF/MICROWAV
101 Frequency 10 MHz Receiver by Direct 5 % to 95 % 2.4 %
E (1 GHZ AND
to 1 GHz (rate: 50 Method
ABOVE)
Hz to 10 kHz)
(Measure)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 19 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
TECHNICAL-
Frequency Using Measuring
RF/MICROWAV
102 Modulation (10 MHz Receiver by Direct 2 kHz to 10 kHz 3%
E (1 GHZ AND
to 1 GHz) Method
ABOVE)
(Measure)
Using Signal
Generator, Low
ELECTRO-
phase noise
TECHNICAL-
reference source, RF
RF/MICROWAV RF Attenuation (100
103 Reference source 1 dB to 50 dB 0.27 dB to 0.4 dB
E (1 GHZ AND kHz to 18 GHz)
and Power meters
ABOVE)
with Sensors, power
(Measure)
sensor by Direct
Method
Using Frequency
ELECTRO-
Counter, Microwave
TECHNICAL-
counter/analyzer,
RF/MICROWAV 0.0000086 % to
104 RF Frequency Spectrum Analyzer 1 Hz to 20 GHz
E (1 GHZ AND 0.00000029 %
& Frequency
ABOVE)
standard by Direct
(Measure)
Method
Using Frequency
ELECTRO-
Counter, Microwave
TECHNICAL-
counter/analyzer,
RF/MICROWAV 0.00000029 % to
105 RF Frequency Spectrum Analyzer 20 GHz to 43.5 GHz
E (1 GHZ AND 0.00000058 %
& Frequency
ABOVE)
standard by Direct
(Measure)
Method
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 20 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
Using Power meter
TECHNICAL-
with Power sensor
RF/MICROWAV RF Power (100 kHz +18 dBm to -70
106 and Power meter 4.71 % to 6 %
E (1 GHZ AND to 40 GHz) dBm
with Power sensor
ABOVE)
by Direct Method
(Measure)
ELECTRO-
Using Power meter
TECHNICAL-
with Power sensor
RF/MICROWAV RF Power (40 GHz to +15 dBm to -60
107 and Power meter 5.0 % to 6 %
E (1 GHZ AND 50 GHz) dBm
with Power sensor
ABOVE)
by Direct Method
(Measure)
ELECTRO-
TECHNICAL- Using Power Sensor,
RF/MICROWAV RF Power (9 kHz to Power Sensor and +13 dBm to -60
108 4.67 % to 4.82 %
E (1 GHZ AND 100 kHz) Spectrum Analyzer dBm
ABOVE) by Direct Method
(Measure)
ELECTRO-
TECHNICAL- Using Network
RF/MICROWAV VSWR (10 MHz to 50 Analyzer and
109 1.05 LU to 2 LU 0.2 LU
E (1 GHZ AND GHz) Calibration kit by
ABOVE) Direct Method
(Measure)
Using low phase
ELECTRO-
noise reference
TECHNICAL-
source, Signal
RF/MICROWAV Bandwidth with
110 Generator with 26 GHz to 33 GHz 0.43 dB to 0.47 dB
E (1 GHZ AND Reference @ 10 MHz
Power meter, Power
ABOVE)
Sensor by Direct
(Source)
Method
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 21 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
Using Oscilloscope
ELECTRO-
Calibrator with Head
TECHNICAL-
Bandwidth with & Low phase Noise
RF/MICROWAV
111 Reference @50 kHz reference RF 1 GHz to 18 GHz 0.26 dB to 0.38 dB
E (1 GHZ AND
and 10 MHz reference source
ABOVE)
with Power sensor
(Source)
by Direct Method
ELECTRO-
Using Oscilloscope
TECHNICAL-
Bandwidth with Calibrator with Head
RF/MICROWAV
112 Reference @50 kHz and Multifunction 10 MHz to 1 GHz 0.23 dB to 0.26 dB
E (1 GHZ AND
and 10 MHz calibrator by Direct
ABOVE)
Method
(Source)
Using low phase
ELECTRO-
noise reference
TECHNICAL-
Bandwidth with source, RF
RF/MICROWAV
113 Reference @50 kHz Reference Source & 18 GHz to 26 GHz 0.38 dB to 0.43 dB
E (1 GHZ AND
and 10 MHz Signal Generator
ABOVE)
with Power meter by
(Source)
Direct Method
ELECTRO-
TECHNICAL-
Using Step
RF/MICROWAV RF Attenuation (100
114 Attenuators by 1 dB to 50 dB 0.23 dB to 0.48 dB
E (1 GHZ AND kHz to 18 GHz)
Direct Method
ABOVE)
(Source)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 22 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
TECHNICAL- Using Analog Signal
RF/MICROWAV Generator & 0.0000086 % to
115 RF Frequency 1 Hz to 20 GHz
E (1 GHZ AND Frequency standard 0.00000029 %
ABOVE) by Direct Method
(Source)
ELECTRO-
TECHNICAL- Using Signal
RF/MICROWAV Generator, 0.00000029 % to
116 RF Frequency 20 GHz to 43.5 GHz
E (1 GHZ AND Frequency Standard 0.00000058 %
ABOVE) by Direct Method
(Source)
ELECTRO- Using RF Signal
TECHNICAL- Generator, Low
RF/MICROWAV RF Power (100 kHz phase Noise +18 dBm to -70
117 7.12 % to 7.25 %
E (1 GHZ AND to 40 GHz) reference source RF dBm
ABOVE) reference Source by
(Source) Direct Method
ELECTRO-
TECHNICAL-
Using RF Signal
RF/MICROWAV RF Power (40 GHz to +15 dBm to -60
118 Generator By Direct 7.12 % to 7.89 %
E (1 GHZ AND 50 GHz) dBm
Method
ABOVE)
(Source)
ELECTRO-
Using Low phase
TECHNICAL-
Noise reference
RF/MICROWAV RF Power (9 kHz to +13 dBm to -60
119 source, Reference 4.66 % to 6 %
E (1 GHZ AND 100 kHz) dBm
Source by Direct
ABOVE)
Method
(Source)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 23 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
TECHNICAL-
Using VSWR
RF/MICROWAV VSWR (10 MHz to 18 1.05, 1.10 LU to 1.5,
120 Standards by Direct 0.09 LU to 0.4 LU
E (1 GHZ AND GHz) 2.0 LU
method
ABOVE)
(Source)
ELECTRO-
TECHNICAL- Using 8½ Digit
E Type
121 TEMPERATURE Precision Multimeter -250 °C to 1000 °C 0.03 °C
Thermocouple
SIMULATION By Direct Method
(Measure)
ELECTRO-
TECHNICAL- Using 8½ Digit
122 TEMPERATURE J Type Thermocouple Precision Multimeter -200 °C to 1200 °C 0.011 °C
SIMULATION By Direct Method
(Measure)
ELECTRO-
TECHNICAL- Using 8½ Digit
K Type
123 TEMPERATURE Precision Multimeter -200 °C to 1372 °C 0.011 °C
Thermocouple
SIMULATION By Direct Method
(Measure)
ELECTRO-
TECHNICAL- Using 8½ Digit
N Type
124 TEMPERATURE Precision Multimeter -200 °C to 1300 °C 0.02 °C
Thermocouple
SIMULATION By Direct Method
(Measure)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 24 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
TECHNICAL- Using 8½ Digit
R Type
125 TEMPERATURE Precision Multimeter 100 °C to 1750 °C 0.2 °C
Thermocouple
SIMULATION By Direct Method
(Measure)
ELECTRO-
TECHNICAL- Using 8½ Digit
126 TEMPERATURE RTD Precision Multimeter -200 °C to 800 °C 0.011 °C
SIMULATION By Direct Method
(Measure)
ELECTRO-
TECHNICAL- Using 8½ Digit
S Type
127 TEMPERATURE Precision Multimeter 100 °C to 1750 °C 0.1 °C
Thermocouple
SIMULATION By Direct Method
(Measure)
ELECTRO-
TECHNICAL- Using 8½ Digit
T Type
128 TEMPERATURE Precision Multimeter -250 °C to 400 °C 0.011 °C
Thermocouple
SIMULATION By Direct Method
(Measure)
ELECTRO-
TECHNICAL- Using Multiproduct
E Type
129 TEMPERATURE Calibrator by Direct -250 °C to 1000 °C 0.49 °C
Thermocouple
SIMULATION Method
(Source)
ELECTRO-
TECHNICAL- Using Multiproduct
130 TEMPERATURE J Type Thermocouple Calibrator by Direct -200 °C to 1200 °C 0.35 °C
SIMULATION Method
(Source)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 25 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
TECHNICAL- Using Multiproduct
K Type
131 TEMPERATURE Calibrator by Direct -200 °C to 1350 °C 0.5 °C
Thermocouple
SIMULATION Method
(Source)
ELECTRO-
TECHNICAL- Using Multiproduct
N Type
132 TEMPERATURE Calibrator by Direct -200 °C to 1300 °C 0.45 °C
Thermocouple
SIMULATION Method
(Source)
ELECTRO-
TECHNICAL- Using Multiproduct
R Type
133 TEMPERATURE Calibrator by Direct 100 °C to 1750 °C 0.67 °C
Thermocouple
SIMULATION Method
(Source)
ELECTRO-
TECHNICAL- Using Multiproduct
134 TEMPERATURE RTD type Calibrator by Direct -200 °C to 800 °C 0.3 °C
SIMULATION Method
(Source)
ELECTRO-
TECHNICAL- Using Multiproduct
S Type
135 TEMPERATURE Calibrator by Direct 100 °C to 1750 °C 0.5 °C
Thermocouple
SIMULATION Method
(Source)
ELECTRO-
TECHNICAL- Using Multiproduct
T Type
136 TEMPERATURE Calibrator by Direct -250 °C to 400 °C 0.7 °C
Thermocouple
SIMULATION Method
(Source)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 26 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
TECHNICAL-
Using Counter by 0.0000089 ns to
137 TIME & Time Period 1 ns to 1 s
Direct Method 0.0000001 s
FREQUENCY
(Measure)
ELECTRO-
TECHNICAL- Using Multifunction
0.000045 Hz to
138 TIME & Frequency calibrator with Direct 10 Hz to 1000 Hz
0.000003 kHz
FREQUENCY Method
(Source)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 27 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
Site Facility
ELECTRO-
TECHNICAL-
Using 8½ Digit
Alternating AC Current (10 Hz to
1 Precision Multimeter 10 mA to 1 A 0.06 % to 0.07 %
Current (< 1 10 kHz)
By Direct Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Using 8½ Digit
Alternating AC Current (300 Hz
2 Precision Multimeter 100 µA to 1 mA 0.05 % to 0.06 %
Current (< 1 to 1 kHz)
By Direct Method
GHz)
(Measure)
ELECTRO-
TECHNICAL- Using AC
Alternating AC Voltage (10 Hz to measurement
3 1 V to 100 V 0.03 % to 0.01 %
Current (< 1 40 Hz) Standard By Direct
GHz) Method
(Measure)
ELECTRO-
TECHNICAL- Using AC
Alternating AC Voltage (10 Hz to measurement
4 10 mV to 1 V 0.04 %
Current (< 1 40 Hz) Standard By Direct
GHz) Method
(Measure)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 28 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
TECHNICAL- Using AC
Alternating AC Voltage (10 kHz measurement
5 1 V to 10 V 0.06 %
Current (< 1 to 100 kHz) Standard By Direct
GHz) Method
(Measure)
ELECTRO-
TECHNICAL- Using AC
Alternating AC Voltage (10 kHz measurement
6 10 mV to 1 V 0.4 % to 0.06 %
Current (< 1 to 100 kHz) Standard By Direct
GHz) Method
(Measure)
ELECTRO-
TECHNICAL- Using AC
Alternating AC Voltage (10 kHz measurement
7 10 V to 500 V 0.06 % to 0.07 %
Current (< 1 to 100 kHz) Standard By Direct
GHz) Method
(Measure)
ELECTRO-
TECHNICAL- Using AC
Alternating AC Voltage (100 kHz measurement
8 10 mV to 20 V 0.4 %
Current (< 1 to 1 MHz) Standard By Direct
GHz) Method
(Measure)
ELECTRO-
TECHNICAL- Using AC
Alternating AC Voltage (40 Hz to measurement
9 1 V to 1000 V 0.013 % to 0.27 %
Current (< 1 20 kHz) Standard By Direct
GHz) Method
(Measure)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 29 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
TECHNICAL- Using AC
Alternating AC Voltage (40 Hz to measurement
10 10 mV to 1 V 0.04 % to 0.013 %
Current (< 1 20 kHz) Standard By Direct
GHz) Method
(Measure)
ELECTRO-
TECHNICAL-
Using RLC Digi
Alternating
11 Capacitance @1 kHz bridge by Direct 10 nF to 100 nF 0.07 % to 0.04 %
Current (< 1
Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Using RLC Digi
Alternating
12 Capacitance @1 kHz bridge by Direct 10 pF to 100 pF 0.22 %
Current (< 1
Method:
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Using RLC Digi
Alternating
13 Capacitance @1 kHz bridge by Direct 100 nF to 1000 nF 0.04 % to 0.03 %
Current (< 1
Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Using RLC Digi
Alternating
14 Capacitance @1 kHz bridge by Direct 100 pF to 10 nF 0.2 % to 0.07 %
Current (< 1
Method
GHz)
(Measure)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 30 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
TECHNICAL-
Using RLC Digi
Alternating
15 Capacitance @1 kHz bridge by Direct 1000 nF to 100 µF 0.03 % to 0.3 %
Current (< 1
Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Using RLC Digi
Alternating
16 Inductance @1 kHz bridge by Direct 1 H to 10 H 0.06 % to 0.04 %
Current (< 1
Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Using RLC Digi
Alternating
17 Inductance @1 kHz bridge by Direct 10 mH to 1 H 0.2 % to 0.06 %
Current (< 1
Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Using RLC Digi
Alternating
18 Inductance @1 kHz bridge by Direct 100 µH to 10 mH 0.2 % to 0.2 %
Current (< 1
Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Alternating Using Phase meter
19 Phase Angle -180 ° to +180 ° 0.03 °
Current (< 1 By Direct Method
GHz)
(Measure)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 31 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
TECHNICAL- Using Multiproduct
AC Current (1 kHz to
20 Alternating Calibrator by Direct 20 mA to 10 A 0.05 % to 0.1 %
10 kHz)
Current (< 1 Method
GHz) (Source)
ELECTRO-
TECHNICAL- Using Multiproduct
AC Current (10 Hz to
21 Alternating Calibrator by Direct 200 µA to 100 mA 0.11 % to 0.05 %
40 Hz)
Current (< 1 Method
GHz) (Source)
ELECTRO-
TECHNICAL- Using Multiproduct
AC Current (40 Hz to
22 Alternating Calibrator by Direct 1 A to 10 A 0.05 % to 0.06 %
1 kHz)
Current (< 1 Method
GHz) (Source)
ELECTRO-
TECHNICAL- Using Multiproduct
AC Current (40 Hz to
23 Alternating Calibrator by Direct 1 mA to 100 mA 0.027 % to 0.031 %
1 kHz)
Current (< 1 Method
GHz) (Source)
ELECTRO-
TECHNICAL- Using Multiproduct
AC Current (40 Hz to
24 Alternating Calibrator by Direct 10 A to 20 A 0.06 % to 0.2 %
1 kHz)
Current (< 1 Method
GHz) (Source)
ELECTRO-
TECHNICAL- Using Multiproduct
AC Current (40 Hz to
25 Alternating Calibrator by Direct 100 µA to 1 mA 0.1 % to 0.036 %
1 kHz)
Current (< 1 Method
GHz) (Source)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 32 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
TECHNICAL- Using Multiproduct
AC Current (40 Hz to
26 Alternating Calibrator by Direct 100 mA to 1 A 0.031 % to 0.041 %
1 kHz)
Current (< 1 Method
GHz) (Source)
ELECTRO-
TECHNICAL- Using Multiproduct
AC Current (5 kHz to
27 Alternating Calibrator by Direct 20 mA to 10 A 0.03 % to 0.4 %
10 kHz)
Current (< 1 Method
GHz) (Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
Calibrator and
28 Alternating AC Current @50 Hz 20 A to 1000 A 0.73 % to 0.8 %
current coil by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO-
TECHNICAL- Using multiproduct
AC Voltage (1 kHz to
29 Alternating calibrator by Direct 100 V to 500 V 0.03 % to 0.27 %
100 kHz)
Current (< 1 method
GHz) (Source)
ELECTRO-
TECHNICAL- Using multiproduct
AC Voltage (1 kHz to
30 Alternating calibrator by Direct 2 mV to 2 V 0.45 % to 0.015 %
100 kHz)
Current (< 1 method
GHz) (Source)
ELECTRO-
TECHNICAL- Using multiproduct
AC Voltage (1 kHz to
31 Alternating calibrator by Direct 2 V to 100 V 0.015 % to 0.02 %
100 kHz)
Current (< 1 method
GHz) (Source)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 33 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
TECHNICAL- Using multiproduct
AC Voltage (10 Hz to
32 Alternating calibrator by Direct 1 mV to 1 V 0.79 % to 0.052 %
40 Hz)
Current (< 1 method
GHz) (Source)
ELECTRO-
TECHNICAL- Using multiproduct
AC Voltage (10 Hz to
33 Alternating calibrator by Direct 2 mV to 20 mV 0.26 % to 0.05 %
40 Hz)
Current (< 1 method
GHz) (Source)
ELECTRO-
TECHNICAL- Using multiproduct
AC Voltage (10 Hz to
34 Alternating calibrator by Direct 20 mV to 20 V 0.05 % to 0.03 %
40 Hz)
Current (< 1 method
GHz) (Source)
ELECTRO-
TECHNICAL- Using multiproduct
AC Voltage (10 Hz to
35 Alternating calibrator by Direct 20 V to 200 V 0.03 % to 0.025 %
40 Hz)
Current (< 1 method
GHz) (Source)
ELECTRO-
TECHNICAL- Using Multiproduct
AC Voltage (100 kHz
36 Alternating Calibrator by Direct 2 mV to 20 mV 1.75 % to 0.52 %
to 1 MHz)
Current (< 1 Method
GHz) (Source)
ELECTRO-
TECHNICAL- Using Multiproduct
AC Voltage (100 kHz
37 Alternating Calibrator by Direct 20 mV to 200 mV 0.46 % to 0.4 %
to 1 MHz)
Current (< 1 Method
GHz) (Source)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 34 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
TECHNICAL- Using Multiproduct
AC Voltage (100 kHz
38 Alternating Calibrator by Direct 200 mV to 20 V 0.4 % to 0.25 %
to 1 MHz)
Current (< 1 Method
GHz) (Source)
ELECTRO-
TECHNICAL- Using multiproduct
AC Voltage (40 Hz to
39 Alternating calibrator by Direct 1 mV to 2 mV 0.7 % to 0.3 %
1 kHz)
Current (< 1 method
GHz) (Source)
ELECTRO-
TECHNICAL- Using multiproduct
AC Voltage (40 Hz to
40 Alternating calibrator by Direct 2 mV to 20 mV 0.3 % to 0.03 %
1 kHz)
Current (< 1 method
GHz) (Source)
ELECTRO-
TECHNICAL- Using multiproduct
AC Voltage (40 Hz to
41 Alternating calibrator by Direct 2 V to 20 V 0.006 % to 0.013 %
1 kHz)
Current (< 1 method
GHz) (Source)
ELECTRO-
TECHNICAL- Using multiproduct
AC Voltage (40 Hz to
42 Alternating calibrator by Direct 20 mV to 200 mV 0.03 % to 0.015 %
1 kHz)
Current (< 1 method
GHz) (Source)
ELECTRO-
TECHNICAL- Using multiproduct
AC Voltage (40 Hz to
43 Alternating calibrator by Direct 20 V to 1000 V 0.013 % to 0.038 %
1 kHz)
Current (< 1 method
GHz) (Source)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 35 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
TECHNICAL- Using multiproduct
AC Voltage (40 Hz to
44 Alternating calibrator by Direct 200 mV to 2 V 0.015 % to 0.013 %
1 kHz)
Current (< 1 method
GHz) (Source)
ELECTRO-
TECHNICAL- Using Capacitor
45 Alternating Capacitance @1 kHz Standard by Direct 1 µF to 100 mF 0.58 % to 0.3 %
Current (< 1 Method
GHz) (Source)
ELECTRO-
TECHNICAL- Using Capacitor
46 Alternating Capacitance @1 kHz Standard by Direct 10 nF to 1 µF 0.58 %
Current (< 1 Method
GHz) (Source)
ELECTRO-
TECHNICAL- Using Capacitor
47 Alternating Capacitance @1 kHz Standard by Direct 100 mF to 1 F 0.3 % to 0.6 %
Current (< 1 Method
GHz) (Source)
ELECTRO-
TECHNICAL- Using Capacitor
48 Alternating Capacitance @1 kHz Standard by Direct 220 pF to 10 nF 5.8 % to 0.58 %
Current (< 1 Method
GHz) (Source)
ELECTRO-
TECHNICAL- Using Standard
49 Alternating Inductance @1 kHz Inductors series by 100 µH to 10 H 0.3 % to 0.2 %
Current (< 1 direct method
GHz) (Source)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 36 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 37 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
TECHNICAL- Using 8½ Digit
56 DIRECT DC Current Precision Multimeter 100 µA to 1 A 0.003 % to 0.004 %
CURRENT By Direct Method
(Measure)
ELECTRO-
TECHNICAL- Using 8½ Digit
57 DIRECT DC Current Precision Multimeter 100 nA to 1 µA 0.05 % to 0.02 %
CURRENT By Direct Method
(Measure)
ELECTRO-
Using 8½ Digit
TECHNICAL-
Precision Multimeter
58 DIRECT DC Current 20 A to 100 A 0.35 %
and current shunt by
CURRENT
direct method
(Measure)
ELECTRO-
TECHNICAL- Using 8½ Digit
59 DIRECT DC Voltage Precision Multimeter 1 mV to 10 mV 0.02 % to 0.008 %
CURRENT By Direct Method
(Measure)
ELECTRO-
TECHNICAL- Using 8½ Digit
60 DIRECT DC Voltage Precision Multimeter 10 mV to 100 mV 0.008 % to 0.0008 %
CURRENT By Direct Method
(Measure)
ELECTRO-
TECHNICAL- Using 8½ Digit
61 DIRECT DC Voltage Precision Multimeter 100 µV to 1 mV 0.35 % to 0.02 %
CURRENT By Direct Method
(Measure)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 38 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
TECHNICAL- Using 8½ Digit
62 DIRECT DC Voltage Precision Multimeter 100 mV to 1000 V 0.0008 %
CURRENT By Direct Method
(Measure)
ELECTRO-
TECHNICAL- Using 8½ Digit
10 Mohm to 100
63 DIRECT Resistance (2 wire) Precision Multimeter 0.004 % to 0.015 %
Mohm
CURRENT by direct method
(Measure)
ELECTRO-
TECHNICAL- Using 8½ Digit
100 kohm to 10
64 DIRECT Resistance (2 wire) Precision Multimeter 0.0034 % to 0.004 %
Mohm
CURRENT by direct method
(Measure)
ELECTRO-
TECHNICAL- Using 8½ Digit
100 Mohm to 1
65 DIRECT Resistance (2 wire) Precision Multimeter 0.015 % to 0.28 %
Gohm
CURRENT by direct method
(Measure)
ELECTRO-
TECHNICAL- Using 8½ Digit
66 DIRECT Resistance (4 wire) Precision Multimeter 1 ohm to 10 ohm 0.004 % to 0.002 %
CURRENT by direct method
(Measure)
ELECTRO-
TECHNICAL- Using 8½ Digit
67 DIRECT Resistance (4 wire) Precision Multimeter 10 ohm to 100 ohm 0.002 % to 0.001 %
CURRENT by direct method
(Measure)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 39 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
TECHNICAL- Using 8½ Digit
100 ohm to 100
68 DIRECT Resistance (4 wire) Precision Multimeter 0.001 % to 0.0034 %
kohm
CURRENT by direct method
(Measure)
ELECTRO-
TECHNICAL- Using Multiproduct
69 DIRECT DC Current Calibrator by Direct 1 A to 10 A 0.011 % to 0.06 %
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Using Multiproduct
70 DIRECT DC Current Calibrator by Direct 1 mA to 1 A 0.0051 % to 0.011 %
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Using Multiproduct
71 DIRECT DC Current Calibrator by Direct 10 µA to 100 µA 0.74 % to 0.012 %
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Using Multiproduct
72 DIRECT DC Current Calibrator by Direct 10 A to 20 A 0.06 % to 0.2 %
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Using multiproduct
73 DIRECT DC Current calibrator by direct 100 µA to 1 mA 0.012 % to 0.0051 %
CURRENT method
(Source)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 40 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
Using multiproduct
TECHNICAL-
calibrator and
74 DIRECT DC Current 20 A to 1000 A 0.35 % to 0.77 %
current coil by direct
CURRENT
method
(Source)
ELECTRO-
TECHNICAL- Using Multiproduct
75 DIRECT DC Voltage calibrator by direct 1 mV to 100 mV 0.05 % to 0.001 %
CURRENT method
(Source)
ELECTRO-
TECHNICAL- Using Multiproduct
76 DIRECT DC Voltage Calibrator by Direct 100 mV to 100 V 0.001 % to 0.0007 %
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Using multiproduct
0.0007 % to 0.0009
77 DIRECT DC Voltage calibrator by Direct 100 V to 1000 V
%
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Using Multiproduct
78 DIRECT Resistance Calibrator by Direct 1 ohm to 10 Mohm 0.013 % to 0.0099 %
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Using Standard
79 DIRECT Resistance (2 wire) resistor by Direct 1 Gohm 0.03 %
CURRENT Method
(Source)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 41 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
TECHNICAL- Using Standard
80 DIRECT Resistance (2 wire) resistor by Direct 10 Mohm 0.0013 %
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Using Multiproduct
10 Mohm to 100
81 DIRECT Resistance (2 wire) Calibrator by Direct 0.0099 % to 0.015 %
Mohm
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Using Multiproduct
100 Mohm to 1
82 DIRECT Resistance (2 wire) Calibrator by Direct 0.15 % to 1.79 %
Gohm
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Using Standard
83 DIRECT Resistance (4 wire) resistor by Direct 1 kohm 0.0009 %
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Using Standard
84 DIRECT Resistance (4 wire) resistor by Direct 1 Mohm 0.0008 %
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Using Standard
85 DIRECT Resistance (4 wire) resistor by Direct 1 ohm 0.0005 %
CURRENT Method
(Source)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 42 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
TECHNICAL- Using Standard
86 DIRECT Resistance (4 wire) resistor by Direct 10 kohm 0.0008 %
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Using Standard
87 DIRECT Resistance (4 wire) resistor by Direct 10 ohm 0.00044 %
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Using Standard
88 DIRECT Resistance (4 wire) resistor by Direct 100 ohm 0.0006 %
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Using Oscilloscope
Oscilloscope (Time 0.0001 % to 0.00035
89 ELECTRICAL Calibrator with Head 1 ns to 5 s
base) %
EQUIPMENT by Direct Method
(Source)
ELECTRO- Using Multifunction
TECHNICAL- Oscilloscope - Calibrator and
90 ELECTRICAL Amplitude (DC Oscilloscope 1 mV to 100 mV 0.1 % to 0.06 %
EQUIPMENT Signal) Calibrator with Head
(Source) by Direct Method
ELECTRO-
TECHNICAL- Oscilloscope - Using Oscilloscope
91 ELECTRICAL Amplitude (DC Calibrator with Head 100 mV to 190 V 0.06 % to 0.04 %
EQUIPMENT Signal) by Direct Method
(Source)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 43 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
Amplitude
TECHNICAL-
Modulation With Using Measuring
RF/MICROWAV
92 Frequency 10 MHz Receiver by Direct 5 % to 95 % 2.4 %
E (1 GHZ AND
to 1 GHz (rate: 50 Method
ABOVE)
Hz to 10 kHz)
(Measure)
ELECTRO-
TECHNICAL-
Frequency Using Measuring
RF/MICROWAV
93 Modulation (10 MHz Receiver by Direct 2 kHz to 10 kHz 3%
E (1 GHZ AND
to 1 GHz) Method
ABOVE)
(Measure)
Using Signal
Generator, Low
ELECTRO-
phase noise
TECHNICAL-
reference source, RF
RF/MICROWAV RF Attenuation (100
94 Reference source 1 dB to 50 dB 0.27 dB to 0.4 dB
E (1 GHZ AND kHz to 18 GHz)
and Power meters
ABOVE)
with Sensors, power
(Measure)
sensor by Direct
Method
Using Frequency
ELECTRO-
Counter, Microwave
TECHNICAL-
counter/analyzer,
RF/MICROWAV 0.0000086 % to
95 RF Frequency Spectrum Analyzer 1 Hz to 20 GHz
E (1 GHZ AND 0.00000029 %
& Frequency
ABOVE)
standard by Direct
(Measure)
Method
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 44 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
Using Frequency
ELECTRO-
Counter, Microwave
TECHNICAL-
counter/analyzer,
RF/MICROWAV 0.00000029 % to
96 RF Frequency Spectrum Analyzer 20 GHz to 43.5 GHz
E (1 GHZ AND 0.00000058 %
& Frequency
ABOVE)
standard by Direct
(Measure)
Method
ELECTRO-
Using Power meter
TECHNICAL-
with Power sensor
RF/MICROWAV RF Power (100 kHz +18 dBm to -70
97 and Power meter 4.71 % to 6 %
E (1 GHZ AND to 40 GHz) dBm
with Power sensor
ABOVE)
by Direct Method
(Measure)
ELECTRO-
Using Power meter
TECHNICAL-
with Power sensor
RF/MICROWAV RF Power (40 GHz to +15 dBm to -60
98 and Power meter 5.0 % to 6 %
E (1 GHZ AND 50 GHz) dBm
with Power sensor
ABOVE)
by Direct Method
(Measure)
ELECTRO-
TECHNICAL- Using Power Sensor,
RF/MICROWAV RF Power (9 kHz to Power Sensor and +13 dBm to -60
99 4.67 % to 4.82 %
E (1 GHZ AND 100 kHz) Spectrum Analyzer dBm
ABOVE) by Direct Method
(Measure)
ELECTRO-
TECHNICAL- Using Network
RF/MICROWAV VSWR (10 MHz to 50 Analyzer and
100 1.05 LU to 2 LU 0.2 LU
E (1 GHZ AND GHz) Calibration kit by
ABOVE) Direct Method
(Measure)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 45 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 46 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
TECHNICAL-
Using Step
RF/MICROWAV RF Attenuation (100
105 Attenuators by 1 dB to 50 dB 0.23 dB to 0.48 dB
E (1 GHZ AND kHz to 18 GHz)
Direct Method
ABOVE)
(Source)
ELECTRO-
TECHNICAL- Using Analog Signal
RF/MICROWAV Generator & 0.0000086 % to
106 RF Frequency 1 Hz to 20 GHz
E (1 GHZ AND Frequency standard 0.00000029 %
ABOVE) by Direct Method
(Source)
ELECTRO-
TECHNICAL- Using Signal
RF/MICROWAV Generator, 0.00000029 % to
107 RF Frequency 20 GHz to 43.5 GHz
E (1 GHZ AND Frequency Standard 0.00000058 %
ABOVE) by Direct Method
(Source)
ELECTRO- Using RF Signal
TECHNICAL- Generator, Low
RF/MICROWAV RF Power (100 kHz phase Noise +18 dBm to -70
108 7.12 % to 7.25 %
E (1 GHZ AND to 40 GHz) reference source RF dBm
ABOVE) reference Source by
(Source) Direct Method
ELECTRO-
TECHNICAL-
Using RF Signal
RF/MICROWAV RF Power (40 GHz to +15 dBm to -60
109 Generator By Direct 7.12 % to 7.89 %
E (1 GHZ AND 50 GHz) dBm
Method
ABOVE)
(Source)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 47 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
Using Low phase
TECHNICAL-
Noise reference
RF/MICROWAV RF Power (9 kHz to +13 dBm to -60
110 source, Reference 4.66 % to 6 %
E (1 GHZ AND 100 kHz) dBm
Source by Direct
ABOVE)
Method
(Source)
ELECTRO-
TECHNICAL-
Using VSWR
RF/MICROWAV VSWR (10 MHz to 18 1.05, 1.10 LU to 1.5,
111 Standards by Direct 0.09 LU to 0.4 LU
E (1 GHZ AND GHz) 2.0 LU
method
ABOVE)
(Source)
ELECTRO-
TECHNICAL- Using 8½ Digit
E Type
112 TEMPERATURE Precision Multimeter -250 °C to 1000 °C 0.03 °C
Thermocouple
SIMULATION By Direct Method
(Measure)
ELECTRO-
TECHNICAL- Using 8½ Digit
113 TEMPERATURE J Type Thermocouple Precision Multimeter -200 °C to 1200 °C 0.011 °C
SIMULATION By Direct Method
(Measure)
ELECTRO-
TECHNICAL- Using 8½ Digit
K Type
114 TEMPERATURE Precision Multimeter -200 °C to 1372 °C 0.011 °C
Thermocouple
SIMULATION By Direct Method
(Measure)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 48 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
TECHNICAL- Using 8½ Digit
N Type
115 TEMPERATURE Precision Multimeter -200 °C to 1300 °C 0.02 °C
Thermocouple
SIMULATION By Direct Method
(Measure)
ELECTRO-
TECHNICAL- Using 8½ Digit
R Type
116 TEMPERATURE Precision Multimeter 100 °C to 1750 °C 0.2 °C
Thermocouple
SIMULATION By Direct Method
(Measure)
ELECTRO-
TECHNICAL- Using 8½ Digit
117 TEMPERATURE RTD Precision Multimeter -200 °C to 800 °C 0.011 °C
SIMULATION By Direct Method
(Measure)
ELECTRO-
TECHNICAL- Using 8½ Digit
S Type
118 TEMPERATURE Precision Multimeter 100 °C to 1750 °C 0.1 °C
Thermocouple
SIMULATION By Direct Method
(Measure)
ELECTRO-
TECHNICAL- Using 8½ Digit
T Type
119 TEMPERATURE Precision Multimeter -250 °C to 400 °C 0.011 °C
Thermocouple
SIMULATION By Direct Method
(Measure)
ELECTRO-
TECHNICAL- Using Multiproduct
E Type
120 TEMPERATURE Calibrator by Direct -250 °C to 1000 °C 0.49 °C
Thermocouple
SIMULATION Method
(Source)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 49 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
TECHNICAL- Using Multiproduct
121 TEMPERATURE J Type Thermocouple Calibrator by Direct -200 °C to 1200 °C 0.35 °C
SIMULATION Method
(Source)
ELECTRO-
TECHNICAL- Using Multiproduct
K Type
122 TEMPERATURE Calibrator by Direct -200 °C to 1350 °C 0.5 °C
Thermocouple
SIMULATION Method
(Source)
ELECTRO-
TECHNICAL- Using Multiproduct
N Type
123 TEMPERATURE Calibrator by Direct -200 °C to 1300 °C 0.45 °C
Thermocouple
SIMULATION Method
(Source)
ELECTRO-
TECHNICAL- Using Multiproduct
R Type
124 TEMPERATURE Calibrator by Direct 100 °C to 1750 °C 0.67 °C
Thermocouple
SIMULATION Method
(Source)
ELECTRO-
TECHNICAL- Using Multiproduct
125 TEMPERATURE RTD type Calibrator by Direct -200 °C to 800 °C 0.3 °C
SIMULATION Method
(Source)
ELECTRO-
TECHNICAL- Using Multiproduct
S Type
126 TEMPERATURE Calibrator by Direct 100 °C to 1750 °C 0.5 °C
Thermocouple
SIMULATION Method
(Source)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
TEKTRONIX INDIA PVT. LTD., SALARPURA PREMIA, SURVEY NO 16,
Laboratory Name :
KADUBEESANAHALLI, VARTHUR HOBLI, BENGALURU, KARNATAKA, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2130 Page No 50 of 50
Validity 22/05/2024 to 21/05/2026 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
TECHNICAL- Using Multiproduct
T Type
127 TEMPERATURE Calibrator by Direct -250 °C to 400 °C 0.7 °C
Thermocouple
SIMULATION Method
(Source)
ELECTRO-
TECHNICAL-
Using Counter by 0.0000089 ns to
128 TIME & Time Period 1 ns to 1 s
Direct Method 0.0000001 s
FREQUENCY
(Measure)
Indicator with sensor
of Environmental Using Humidity
THERMAL-
chambers, Indicator with sensor 30 %rh to 90 %rh @
129 SPECIFIC HEAT 1.6 %rh
Temperature by Comparison 25 °C
& HUMIDITY
Chamber (Single Method
Position)
Using RTD sensor
Temperature with Logger (multi-
THERMAL-
130 Chamber, Oven, position minimum 9 -70 °C to 200 °C 0.98 °C
TEMPERATURE
Furnace Sensors) by
Comparison Method
* CMCs represent expanded uncertainties expressed at approximately the 95% level of confidence, using a coverage factor of k = 2.
This is annexure to 'Certificate of Accreditation' and does not require any signature.