Resistance thermometer Pt 100
with screw-in thermowell or with flange connection, Type Series GA251 .
Features
■ Pt 100 connection in 3- or 4-wire technology
■ Measuring insert 1 x Pt 100 or 2 x Pt 100
■ Measuring insert interchangeable
■ Process connection
· for screw-in
· with flange connection
■ Various thermowell designs available
Options
■ Explosion protection
■ Transmitter can be integrated
■ Classification per SIL 2
■ Measuring insert for In-process calibration
Application area Application
· Chemical and petrochemical industry The resistance thermometer is suited for operation on tanks
· Machinery construction and pipes. All standard types of process connections are
available. The change in resistance, dependent on the
measurement temperature, can be detected by a trans-
mitter and converted into a standardized signal. A variety
of transmitters for head mounting is available for different
applications. For In-process calibration the integration of a
special measuriung insert with additional test pipe is possible
(data sheet T4-025-45,Type GA3100, reference sensor:
data sheet T4-025-46, Type GA3110).
Technical Data
Mechanical design Measuring insert For In-process calibration:
Measuring insert interchangeable with Material stainless steel, interchangeable, IBExU 13 ATEX 1017 X
connection head and neck-tube DIN 43735 II 2 G Ex ia IIC T6-T1 Gb
length of measuring insert l5= thermowell
Connection head Ui ≤ 30 V
length L + 10 mm
selective Ø of meas. insert depending on thermowell, Pi ≤ 750 mW
· model B, cap with 2 slotted screws, mat. standard 6 mm Li max. 10 µH/m
aluminium, IP 54 resistor Pt 100 according to DIN EN 60751 Ci max. 500 pF/m
· model BUZH, high spring cover Optional: Measuring insert with connection More technical information see XA_003.
with slotted screw, mat. aluminium, IP 65 socket per DIN 43735 and with additional
· field housing Ø 60 mm, screw cap, test pipe for In-process calibration. Functional safety
stainless steel mat.-no. 1.4305 (303), IP 67 Material: stainless steel, mat..-no. 1.4571 per EN 61508, classification per SIL 2;
further connection heads upon request (316 Ti) (see data sheet T4-025-45) without transmitter, only
Thermowell
Type of sensor/class/circuit Accuracy of the measuring resistor
material stainless steel mat.-no. 1.4571
see order details class A according to EN 60751
(316Ti). Special materials upon request.
Thermowell 9 x 1 bzw. 11 x 2 mm
length see order details Ex-approval
option: certification of material testing per For standard measuring insert:
DIN EN 10204 BVS 04 ATEX E 144 X
Upon request a calculation for thermowells II 2 G Ex ia IIC T4/T6
can be made (for static or dynamic Ui ≤ 30 V
application) with certificate. Pi ≤ 200 mW
More technical information see XA_002
Data Sheet: T4-025-2 Rev. 1A4
Data Sheet: T4-025-2 Rev. 1A4
more Technical Data
Process connections Further process connections upon request b) mounting in the spring cover of the
for screw-in/insertion/weld-in material: stainless steel mat. no. 1.4571 (316Ti) connection head BUZH
· G 1/2 B, G 3/4 B see product group T4 for analog or digital
· G1B Integration of transmitter transmitters
· M 20 x 1.5 suitable Pt 100 transmitters can be integrated
· 1/2" NPT, 3/4" NPT into the connection head. LED-on-site indication
with flange connection Options: programmable LED-on-site indication for
· DN50 PN10/40 model B1 (DIN EN 1092-1) a) instead of terminal block stainless steel field housing (Ø 60 mm),
· DN25 PN10/40 model B1 (DIN EN 1092-1) see data sheet M6-031.
Dimensions
Remark: neck tube M > 60 mm
Connection diagram
Data Sheet: T4-025-2 Rev. 1A4
Order Details - please give additional specifications for models not listed -
Resistance Thermometer Pt 100 with screw-in thermowell or with flange connection
design · with thermowell GA251 .
· without 0
ex-protection
· explosion protection, type of ex-protection s. below 1
· without, for insertion or for welding A01
· G 1/2 B A10 .
· G 3/4 B A11 .
·G1B A12 .
process
· M 20 x 1.5 A13 .
connection
· 1/2" NPT A15 .
· 3/4" NPT A16 .
· flange DN25 PN 10/40 model B1 (DIN EN 1092-1) A21 .
· flange DN50 PN 10/40 model B1 (DIN EN 1092-1) A22 .
material · stainless steel mat.-no. 1.4571 (316Ti) 1
process conn. · varying 9
length L meas. insert L 5
95 mm 105 mm B10
115 mm 125 mm B13
130 mm 140 mm B16
180 mm 190 mm B19
195 mm 205 mm B22
245 mm 255 mm B28
thermowell
265 mm 275 mm B31
length total L
305 mm 315 mm B37
365 mm 375 mm B40
395 mm 405 mm B43
425 mm 435 mm B46
515 mm 525 mm B49
545 mm 555 mm B52
varying B99
· F1 = 9 mm, thermowell 9/7, model 2 per DIN, standard C12 . . . .
thermowell Ø · F1 = 11 mm, thermowell 11/7, model 2 per DIN C13 . . . .
and model · 9/7, reduced tip F3 = Ø 5x20 mm, ID 3.5 mm C16 . . . .
· 12, reduced tip F3 = Ø 9x40 mm, ID 6.5 mm C17 . . . .
as in writing C99 . . . .
thermowell stainless steel mat.-no. 1.4571 (316Ti) 1
material varying 9
immersion length length in mm (e.g. 160 for 160 mm), Umax = L - 60 mm ...
U1 ¹ varying, as in writing 999
diameter, design, material meas. element operating range test pipe
measuring insert · 6 mm, rigid, st. steel, standard thin film -50...+400 °C - D2-M22
as per DIN 43735
(class A) · 6 mm, sheathed element, [Link] ceramic -200...+600 °C - D6-M21
· 6 mm, rigid, st. steel (In-process) -50...+400 °C 28 mm 2 D22-M22
· 1 x Pt 100 in 3-wire technology, standard N2
sensor type · 1 x Pt 100 in 4-wire technology N3
· 2 x Pt 100 in 3-wire technology N5
· model B electrical connection cable gland M20x1.5 T11
· model BUZH nickel plated brass, cable Ø 9-14 T15
cable Ø 3-6.5 T47
polyamide black
cable gland cable Ø 4.5-10 T47.40
connection · field housing
st. steel cable Ø 3-6.5 T47.21
head
with circular connector M12x1 T47.51
· field housing with cable Ø 3-6.5 T49
polyamide black
additional opening cable gland cable Ø 4.5-10 T49.40
for reference sensor st. steel cable Ø 3-6.5 T49.21
additional features (to be indicated in case of need, only)
· II 2G Ex ia IIC T4/T6 3, BVS 04 ATEX E 144 X (standard measuring insert) S68
type of ex-protection
· II 2G Ex ia IIC T6-T1 Gb, IBExU 13 ATEX 1017 X (In-process calibration) S75
incl. transmitter · mounting on the measuring insert (instead of terminal block) Z1
(pls specify separately) · mounting in the spring cover of the connection head BUZH Z2
material certificate per DIN EN 10204-3.1 W1020
functional safety per EN 61508, classification per SIL 2 W2604
transmitter with resistance thermometer calibrated, incl. calibration certificate with 3 meas. points W4057
order code (example): GA2510 A101 B37 C121160 D2-M22 N2 T47
¹ not possible with process connection A01 (insertion/welding)
2
for In-process calibration only (no classification per SIL 2)
3
only with sheathed element
LABOM Mess- und Regeltechnik GmbH · P.O. Box 12 62 ∙ 27795 Hude ∙ Germany · Im Gewerbepark 13 ∙ 27798 Hude ∙ Germany