i.Profiler® plus by ZEISS.
The 4-in-1compact system – ocular wavefront aberrometer,
autorefractometer, ATLAS corneal topographer and keratometer.
ȴȴ Your link to innovative i.Scription® technology
ȴȴ All-in-one system with easy-to-use touch screen
ȴȴ Ocular wavefront measurement up to 7th order Zernike aberration
ȴȴ Measures both eyes automatically in approximately 30 seconds
See more. Live more.
ZEISS precision lenses.
A leader in precision optics since 1846.
i.Profiler® plus by ZEISS.
Ocular wavefront and corneal topography. Easy accurate eye measurement.
i.Profiler combines a high-resolution Hartmann-Shack wave-
® plus
The fully automated measurement procedure, with touch screen
front sensor with the proven ATLAS corneal topographer in a control, enables all measurements of both eyes in
single compact system. approximatly 30 seconds.
Access to i.Scription® technology by ZEISS. Advanced features for faster workflow.
i.Scription technology combines the subjective refraction with
®
Clearly structured functions enable capture, evaluation, pre-
ocular wavefront aberrometry data, creating an individualized sentation and analysis of four measurement modalities: ocular
prescription to 1/100th of a diopter. Integrated with a ZEISS wavefront aberrometer, autorefractometer, ATLAS corneal
high precision lens, i.Scription® technology offers better night topographer and keratometer.
vision, as well as improved color and contrast perception.
Technical Data Wavefront
Measuring range, sphere: –20 D to +20 D
Measuring range, cylinder: 0 D to +8 D
Axis: 0° – 180°
Measuring surface: 2.0 mm to 7.0 mm (three zones)
No. of measuring points: up to 1500
Method: Hartmann-Shack
Reference wavelength:1 555 nm according to ISO 24157
2D ocular wavefront map.
Technical Data Corneal Topography
No. of rings: 22 (18 complete rings)
No. of measuring points: 3,425
Detected corneal surface
dia. 0.75 mm to 9.4 mm
at 42.125 D:
Diopters: measurement range 25 to 65 D
Accuracy: ± 0.05 D (± 0.01 mm)
Reproducibility: ± 0.10 D (± 0.02 mm)
EN_20_010_7138I
Type A: according to ISO 19980
1 Reference wavelength for the interpretation of refractive errors (referring to
2D corneal axial curvature map. maximum luminosity function V(λ) of the human eye in daylight).
Carl Zeiss Vision GmbH
www.vision.zeiss.com
[email protected] The scope of services may deviate according to the local circumstances. We retain the right to change the product or the technical specification.