0% found this document useful (0 votes)
114 views28 pages

EO Gas Sterilizer (General) Email

EO Gas Sterilizer (General) Email

Uploaded by

leyen leyen
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
114 views28 pages

EO Gas Sterilizer (General) Email

EO Gas Sterilizer (General) Email

Uploaded by

leyen leyen
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd

HEAD OFFICE

Add : 166 Pyeongcheonro, Bupyeong-gu,


Incheon, 403-853, Korea
(3-7, Cheongcheon 2-dong, Bupyeong-gu, Incheon)
T e l : ++82-32-500-8800 (Key Number)
T e l : ++82-32-508-8118 (Trading Dept.)
Fax : ++82-32-508-8866
E-mail : mail @hanshinmed.co.kr (former: [email protected])
www.hanshinmed.co.kr
EURO BRANCH
..
Add : Arena-Str. 1, 40474 Dusseldorf, Germany
T e l : ++49-211-4957770
Fax : ++49-211-4957772
E-mail : [email protected]
www.eurohanshin.com

C-GS-C5-E April. 2015 Printed in Korea


As HANSHIN s products are ceaselessly improved, the actual product may differ from the
descriptions, specifications, and pictures in this publication.
Research and
Research and Development
Development to to cherish
cherish Human
Human Life,
Life,
Quality Innovations
Quality Innovations to
to practice
practice Social Contribution !!
Social Contribution
This is the“Quality Management Policy”of HANSHIN MEDICAL. To achieve this
policy, we are consistently carrying forward the intensive quality management
activity and striving for the technical development and management innovation
tirelessly.
We have been conforming to the relevant international directive and standards
strictly including Medical Device Directive (MDD 93/42/EEC) and Quality System
Standards (EN ISO 9001/ EN ISO 13485) designing and manufacturing the
medical sterilizers for several decades, and guaranteeing their excellence of
quality to our customers by obtaining the CE .. mark certificate from the reliable
certificate authorities in the world such as TUV NORD and SGS for every new
developed products.

H A N S H I N M E D I C A L has been developing, manufacturing and supplying the medical


sterilizers, refrigerators and freezers as main products, and devoting ourself to develope the
new medical products consistently required for the health care industry since founded.
HANSHIN MEDICAL is producing the products considering the health and safety for the
health care providers and patients firstly all the time.
HANSHIN MEDICAL sincerely pursues the best quality recognized in the global market
and a honest company remained in customer’s memory as the ultimate company goal.

The EO Gas sterilizers of HANSHIN MEDICAL are designed and


manufactured to comply with the following international directives and
standards:
Medical Devices Directive (MDD ..93/42/EEC)
CE Medical Device Mark.. 0044, TUV NORD CERT GmbH, Germany
EN ISO 9001: 2008, TUV.. NORD CERT GmbH, Germany
EN ISO 13485 : 2012, TUV NORD CERT GmbH, Germany
EN ISO 14971 : 2007, Risk Management to Medical Devices
European Standard, Ethylene Oxide Sterilizers (EN 1422)
European Standard, Safety (EN 61010-1/ EN 61010-2-040)
European Standard, EMC (EN 61326-1)
GOST, Russia Standard Improvement Measurement Committee
TGA, Therapeutic Goods Administration, Australia
KS, Korean Industrial Standards
KGMP, Korea Good Manufacturing Practice

2 HANSHIN
www.hanshinmed.co.kr / www.eurohanshin.com

Product Mix

EO Gas Mixture type Sterilizer

▶ Touch Screen LCD Monitor type


HS-700EO HG-5025EOT
(Automatic Sliding Door type)

▶ Graphic LCD Monitor type


HS-700EO HS-5025EO HS-4085EO HS-3241EO

100% EO Gas type Sterilizer


▶ Touch Screen LCD Monitor type
: HS-4313EO

HANSHIN 3
Ethylene Oxide Gas Mixture type Sterilizer
Technical Advantages of HANSHIN’s EO Gas Sterilizers

Sterilization System

User can select a proper cycle program for an optimum performance of sterilization among 4 kinds of basic cycle programs
comprised of high(56℃) and low(38℃) temperature cycle and change the parameters of basic cycle programs at user’s
pleasure according to the material to be sterilized or sterilization condition. Therefore the sterilization is effective and convenient.
These sterilizers use 3 kinds of EO gas mixture as sterilant that the inert gases, the freon and the carbon dioxide(CO2) gases,
are mixed with Ethylene Oxide respectively, thus user can select the sterilant widely and easily.


The aeration cycle is a practical cycle for the items required the additional aeration even though aerated in the basic or user’s
cycle, and the items processed in the sterilizer without aeration function.


The sterilization temperature in the chamber is sustained using the jacket water heated by the built-in heater and the jacket water is
circulated by the pump for maintaining the temperature distribution uniformly, hence the effectiveness of sterilization is excellent.

Control System

The controller with a microprocessor carries on a sterilization cycle automatically and controls LCD displaying, self diagnosis
function, printing the record. Especially the devices with these functions are integrated as a compact component and the
controller can be changed with a normal one rapidly and easily when a trouble has occurred in the controller. This allows user
to operate and maintain the sterilizer efficiently and conveniently.


The self-diagnosis program detects the operation errors during progressing a cycle and generates an warning alarm
automatically. The cycle in progress is aborted automatically when before discharging gas into the chamber and the sterilizer
returns to the initial state if you press the <START/STOP> key (press the <STOP> button for the touch screen type). But when
after discharging gas into the chamber, the cycle in progress is aborted and the gas in the chamber is exhausted automatically
through the vent and aerate phase and the cycle is completed safely. The aeration phase can be stopped during execution
using the <START/STOP> key (using the <STOP> button for the touch screen type).


The precise humidity control is the important requirement because the sterilizing effect is reduced if the humidity is insufficient in
the chamber. The humidity is formed (40~80%RH) by the steam introduced in the chamber, that the steam is generated in the
vaporizer using water injected from the outside.


If the electric power has suddenly cut off during progressing a cycle, the cycle is retained at the stopped process and is
progressed from that process when the electric power has restored. This is possible that the controller stores the cycle
informations at the stopped process during power failure and is an outstanding function for maintaining the continuity of a cycle
in progress.

4 HANSHIN
Vacuum System


The vacuum pump installed in the sterilizer is different in the sterilizer model respectively as the waterring type (HS-
700EO/5025EO, HG-5025EOT) and the diaphragm type (HS-4085EO/3241EO), and all the pumps have excellent operating
performance such as the powerful exhaust capability, durability, airtightness, full oilless operation and low noise etc.


The vacuum leak test is to check whether the chamber is leaking or not and user should carry on this test once everyday
before sterilizing. If the test result is failed, user must stop operation of the sterilizer and repair after investigating the cause
thoroughly because the gas in the chamber is leaked out to the outside and it may cause human injury and an
environmental contamination.

Ventilation System


The gas in the chamber is exhausted during the vent and aeration process to the outside of building through the ventilation
duct generally, but the EO gas sterilizers of HANSHIN are designed to drain the residual gas mixed with water to the prepared
drainage for the safety of user and for protecting environment. The ethylene oxide is dissolved completely in the 10℃ water.


These sterilizers are equipped with a bacteria retentive high efficiency air filter that is able to filter the particles greater than
0.3㎛ at least 99.999% and the highly filtered air can be introduced into the chamber during the vent and aeration process,
this prevents the sterilized items from recontamination by the air-borne bacteria in the chamber.


When changing an exhausted gas cylinder, the residual gases in the piping or pressure regulator are removed automatically by
operation of the control program for protecting the user’s health and preventing environment of workplace from contamination.

Sterilization Process Checking System



Turning on the power, the exposure temperature and time, the aeration time, the actual chamber pressure and the door state are
displayed on the large graphic or touch screen type LCD monitor. During progressing a cycle, the cycle phase (HEAT,
CONDITION, CHARGE, STERILIZE, VENT, AERATE, COMPLETION) in progress is displayed in order, the executing process
(INITIAL, WAITING, VACUUM, GAS IN, AIR IN, WATER IN, LEAK TEST, GAS VENT) and the current time, the number of cycle
times, the kind of gas, the error code are displayed in real time. This allows user to monitor the progressing of a cycle conveniently.
Especially in case of the touch screen type LCD monitor, user can operate and control the sterilizer conveniently by one touching
the function buttons on the screen.


The built-in panel printer prints out the cycle performance record that is able to check and analyze the results of a executed cycle.
The cycle performance record includes the serial number of sterilizer, the year/month/date of cycle start, the set parameters
(exposure temperature and time, adjusting temperature, aeration time, pressure), the execution result data of each cycle phase
(start and completion time, temperature and pressure), the maximum/minimum value of pressure and temperature during
progressing a cycle, the required time of each cycle phase, the error code occurred during progressing a cycle and other useable
informations (total required time of a executed cycle, total number of the executed cycle times and gas supply times etc.)

HANSHIN 5
EO Gas Mixture type Sterilizer_Touch Screen LCD Monitor type

HS- 7 0 0EO
Ubiquitous EO Gas Sterilizer
- Network Monitoring System

Touch Screen LCD Monitor

Front Side
Network System

CYCLE P1 EO/CO2 CYCLE P2 EO/CO2 CYCLE P3 EO/FREON CYCLE P4 EO/FREON


(High Temp.) (Low Temp.) (High Temp.) (Low Temp.)

CYCLE P5 Aeration CYCLE P6 Helix Test CYCLE P7 VACUUM LEAK TEST

6 HANSHIN
▶ This sterilizer uses the gas mixture of Ethylene Oxide(EO) and Freon or Carbon Dioxide (CO2) as sterilant, and is used for sterilizing the materials
which are unstable under high temperature or items to which the steam sterilization is inapplicable for the structural reasons. Therefore it is
essential for the medical institutions where using the precise and delicate medical devices to use this kind of low temperature sterilizer.
▶ The large capacity of chamber allows user to sterilize conveniently many medical devices and instruments or volumed loads at once.
▶ All the sterilizing cycle is progressing automatically by the controller with a microprocessor, and the controller is designed to separate from the
sterilizer easily for prompt servicing by the supplier or manufacturer. User is able to operate and control the sterilizer conveniently by one
touching on the touch screen, and monitor the state of operating sterilizer in real time on it.
▶ The embedded self diagnostic program monitors the cycle progress in real time and displays the error code on the touch screen monitor
immediately accompanying an audible alarm when any trouble has occurred during operation, and then aborts the cycle in progress
automatically for safety of workplace.
▶ Two gas cylinders are available in succession automatically by the control program and the problems like a gas leak or cycle stop caused by
changing the gas cylinder can be prevented.
▶ The communication program for multi-user that is supporting the ubiquitous function is provided as an optional feature. Each operation
informations from the sterilizers connected to the network system which is consisted of LAN, DB server and Client are saved and preserved
in the DB server safely by this program, and user can check the required informations in real time conveniently with the client computer any
time, any where.
▶ The built-in panel printer prints out the cycle performance for analyzing the results of sterilization
and keeping as a quality evidence.

Gas Same as Same as


N/A N/A
Mixture basic cycle basic cycle
Select one among
Cycle No.
basic cycle
Expos. Temp.
Pressure
Expos. Time
Aeration Time

Rear Side (Double Door type)

HS-700EO

HANSHIN 7
EO Gas Mixture type Sterilizer_Touch Screen LCD Monitor type

HG- 5 0 2 5EOT
Automatic Sliding Door type EO Gas Sterilizer
- Realized the Ergonomic Design for Maximizing Operator’
s Safety and Convenience

Front Side

CYCLE P1 EO/CO2 CYCLE P2 EO/CO2 CYCLE P3 EO/FREON CYCLE P4 EO/FREON VACUUM LEAK TEST
(High Temp.) (Low Temp.) (High Temp.) (Low Temp.)

8 HANSHIN
▶ This sterilizer uses the gas mixture of Ethylene Oxide(EO) and Freon or Carbon Dioxide (CO2) as sterilant, and is used for
sterilizing the materials which are unstable under high temperature or items to which the steam sterilization is inapplicable for
the structural reasons. Therefore it is essential for the medical institutions where using the precise and delicate medical
devices to use this kind of low temperature sterilizer.
▶ The practical medium sized chamber allows user to sterilize conveniently many medical devices and instruments or bulky loads at once.
▶ The door moves up and down smoothly without any physical force by adopting the automatic vertical sliding system, therefore the
large volume sterilizing is much more safe and convenient than the hinged door type. Moreover it has the advantage of saving the
required space to install the sterilizer.
▶ Two gas cylinders are available in succession automatically by the control program and the problems like a gas leak or cycle stop
caused by changing the gas cylinder can be prevented.
▶ The embedded self diagnostic program monitors the cycle progress in real time and displays the error code on the touch screen
monitor immediately accompanying an audible alarm when any trouble has occurred during operation, and then aborts the cycle in
progress automatically for safety of workplace.
▶ The built-in panel printer prints out the cycle performance for analyzing the results of sterilization
and keeping as a quality evidence.

Same as
Gas Mixture N/A
basic cycle
Select one among
Cycle No.
basic cycle
Expos. Temp.
Pressure
Expos. Time
Aeration Time

Rear Side (Double Door type)

HG-5025EOT

HANSHIN 9
EO Gas Mixture type Sterilizer_Graphic LCD Monitor type

HS- 7 0 0EO
Large Capacity Standard EO Gas Sterilizer
- Effective Sterilization for Bulky
Bulhy Loads At Once

Front Side

CYCLE No.1 EO/CO2 CYCLE No.2 EO/CO2 CYCLE No.3 EO/FREON CYCLE No.4 EO/FREON CYCLE No.5
(High Temp.) (Low Temp.) (High Temp.) (Low Temp.) (Vacuum Leak Test)

10 HANSHIN
▶ This sterilizer uses the gas mixture of Ethylene Oxide(EO) and Freon or Carbon Dioxide (CO2) as sterilant, and sterilizes the materials
which are unstable under high temperature and moisture or items to which the steam sterilization is inapplicable for the structural
reasons. Therefore it is essential for the medical institutions where using the precise and delicate medical devices to use this kind of
low temperature sterilizer.
▶ The large capacity of chamber allows user to sterilize conveniently many medical devices and instruments or volumed loads at once.
▶ User can operate and control the sterilizer conveniently by pressing the operation keys arranged on the control panel and can check
the state of a cycle progressing on the message center (LCD monitor).
▶ Two gas cylinders are available in succession automatically by the control program and the problems like a gas leak or cycle stop
caused by changing the gas cylinder can be prevented.
▶ The embedded self diagnostic program monitors the cycle progress in real time and displays the error code on the monitor
immediately accompanying an audible alarm when any trouble has occurred during operation, and then aborts the cycle in progress
automatically for safety of workplace.
▶ The built-in panel printer prints out the cycle performance for analyzing the results of sterilization
and keeping as a quality evidence.

Same as
Gas Mixture N/A
basic cycle
Select one among
Cycle No.
basic cycle
Expos. Temp.
Pressure
Expos. Time
Aeration Time

Rear Side (Double Door type)

HS-700EO

HANSHIN 11
EO Gas Mixture type Sterilizer_Graphic LCD Monitor type

HS- 5 0 2 5EO
Medium Capacity Standard EO Gas Sterilizer
- Efficient Sterilization for Various Medical Devices At Once

Front Side

CYCLE No.1 EO/CO2 CYCLE No.2 EO/CO2 CYCLE No.3 EO/FREON CYCLE No.4 EO/FREON CYCLE No.5
(High Temp.) (Low Temp.) (High Temp.) (Low Temp.) (Vacuum Leak Test)

12 HANSHIN
▶ This sterilizer uses the gas mixture of Ethylene Oxide(EO) and Freon or Carbon Dioxide (CO2) as sterilant, and sterilizes the
materials which are unstable under high temperature and moisture or items to which the steam sterilization is inapplicable for the
structural reasons. Therefore it is essential for the medical institutions where using the precise and delicate medical devices to
use this kind of low temperature sterilizer.
▶ The practical medium sized chamber allows user to sterilize conveniently many medical devices and instruments or bulky loads at once.
▶ User can operate and control the sterilizer conveniently by pressing the operation keys arranged on the control panel and can check
the state of a cycle progressing on the message center (LCD monitor).
▶ Two gas cylinders are available in succession automatically by the control program and the problems like a gas leak or cycle stop
caused by changing the gas cylinder can be prevented.
▶ The embedded self diagnostic program monitors the cycle progress in real time and displays the error code on the monitor
immediately accompanying an audible alarm when any trouble has occurred during operation, and then aborts the cycle in progress
automatically for safety of workplace.
▶ The built-in panel printer prints out the cycle performance for analyzing the results of sterilization
and keeping as a quality evidence.

Same as
Gas Mixture N/A N/A
basic cycle
Select one among
Cycle No. Aerate
basic cycle
Expos. Temp.
Pressure
Expos. Time
Aeration time

Rear Side (Double Door type)

HS-5025EO

HANSHIN 13
EO Gas Mixture type Sterilizer_Graphic LCD Monitor type

HS- 4 0 8 5EO
Practical Capacity Standard EO Gas Sterilizer
- Practical Sterilization Between Medium
and Small Sized Sterilizer

CYCLE No.1 EO/CO2 CYCLE No.2 EO/CO2 CYCLE No.3 EO/FREON CYCLE No.4 EO/FREON VACUUM LEAK TEST
(High Temp.) (Low Temp.) (High Temp.) (Low Temp.)

14 HANSHIN
▶ This sterilizer uses the gas mixture of Ethylene Oxide(EO) and Freon or Carbon Dioxide (CO2) as sterilant, and sterilizes the materials
which are unstable under high temperature and moisture or items to which the steam sterilization is inapplicable for the structural
reasons. Therefore it is essential for the medical institutions where using the precise and delicate medical devices to use this kind of
low temperature sterilizer.
▶ This sterilizer is very effective to load a lot of medical devices because it has a practical sized chamber between large and small
sized sterilizer.
▶ This sterilizer adopts the cylindrical chamber with jacket and the variation of chamber temperature is reduced, the sterilizing
temperature can be maintained evenly during progressing a cycle.
▶ Air pollution caused by the leakage of gas after sterilization cycle can be prevented because the sterilization and aeration process
have done in the same chamber successively.
▶ The embedded self diagnostic program monitors the cycle progress in real time and displays the error code on the monitor
immediately accompanying an audible alarm when any trouble has occurred during operation, and then aborts the cycle in progress
automatically for safety of workplace.
▶ The built-in panel printer prints out the cycle performance for analyzing the results of sterilization
and keeping as a quality evidence.

Same as
Gas Mixture N/A
basic cycle
Select one among
Cycle No.
basic cycle
Expos. Temp.
Pressure
Sterilization Time
Aeration Time

HS-4085EO

HANSHIN 15
EO Gas Mixture type Sterilizer_Graphic LCD Monitor type

HS- 3 2 4 1EO
Small Capacity Standard EO Gas Sterilizer
- Effective Sterilization For Small
Medical Devices and Instruments

CYCLE No.1 EO/CO2 CYCLE No.2 EO/CO2 CYCLE No.3 EO/FREON CYCLE No.4 EO/FREON VACUUM LEAK TEST
(High Temp.) (Low Temp.) (High Temp.) (Low Temp.)

16 HANSHIN
▶ This sterilizer uses the gas mixture of Ethylene Oxide(EO) and Freon or Carbon Dioxide (CO2) as sterilant, and sterilizes the materials
which are unstable under high temperature and moisture or items to which the steam sterilization is inapplicable for the structural
reasons. Therefore it is essential for the medical institutions where using the precise and delicate medical devices to use this kind of
low temperature sterilizer.
▶ This is an economic and substantial small size EO gas sterilizer for sterilizing the small medical devices, instruments and supplies
effectively.
▶ This sterilizer adopts the cylindrical chamber with jacket and the variation of chamber temperature is reduced, the sterilizing
temperature can be maintained evenly during progressing a cycle.
▶ Air pollution caused by the leakage of gas after sterilization cycle can be prevented because the sterilization and aeration process
have done in the same chamber successively.
▶ The embedded self diagnostic program monitors the cycle progress in real time and displays the error code on the monitor
immediately accompanying an audible alarm when any trouble has occurred during operation, and then aborts the cycle in progress
automatically for safety of workplace.
▶ The built-in panel printer prints out the cycle performance for analyzing the results of sterilization
and keeping as a quality evidence.

Same as
Gas Mixture N/A
basic cycle
Select one among
Cycle No.
basic cycle
Expos. Temp.
Pressure
Sterilization Time
Aeration Time

HS-3241EO

HANSHIN 17
● Gas is supplied into the chamber from one gas ● Two gas cylinders are connected to the chamber. If one
cylinder on the bottom side of appliance table. cylinder is exhausted, the other is connected
The available gas cylinders are as follows, automatically by the control program. The available gas
① Cylinder of EO / HCFC (30kg) cylinders are as follows,
② Cylinder of EO / CO2 (10kg) ① Cylinder of EO / HCFC (60kg)
② Cylinder of EO / CO2 (25kg)

NOTE :“Preheat Pressure Regulator for EO Gas Only”and“Hose for EO Gas Only”should be used to the EO gas sterilizer,
otherwise very dangerous situation may be occurred.

Weight Percentage (Wt%) EO 10 / HCFC (124)63 / HCFC (22)27 EO 20 / CO2 80


Volume Percentage (Vol.%) EO 22.7 / HCFC 77.3 EO 20 / CO2 80
Specific Gravity of Gas @30℃ 3.20 1.52
Boiling Point (℃) - 22 - 76
2.1 35
Vapor Pressure (bar) 3.4 45
4.9 57
Weight (Kg) 30 Kg, 60 Kg 10 Kg, 25 Kg
HS-38
-
(Gas Regulator & Pre-Heater)
Gas Regulator
& Pre-Heater EO-57A(Gas Pre-Heater) &
EO-57A(Gas Pre-Heater) & Adapter(OA)
HS-57(Gas Regulator & Pre-Heater)
Remarks Figure (A) Figure (B), Figure (C)

EO-57A

Figure (A) Oxyfume 2002 CAPOX-20

18 HANSHIN
Ethylene Oxide Gas Mixture

If EO +CO2 gas is injected into the sterilizing chamber directly, a trouble may occur in the piping
system due to the high pressure of the gas (about 45bar at 20℃). Decreasing pressure during delivery
to the chamber may cause an abrupt drop in temperature. Therefore when using EO + CO2 gas as a
sterilant, the exclusive heating-type EO gas pressure regulator (refer to the specification table below)
should be used to depressurize to an adequate level.

※ Caution : Manufacturer and supplier do not responsible for any problem arising from the use of a gas pressure
regulator with a different specification, e.g., for CO2 only.
※ Safety : ① Prevention of temperature increase : The temperature is controlled by the thermostat. In case of
malfunction in the thermostat, the temperature fuse is activated to cut off power.
② Prevention of pressure increase : In case of rising secondary pressure due to the leakage of gas in
the valve seat, the safety valve is activated at the pressure of 9 bar to protect the equipment.

Applicable Gas
Primary Pressure
Secondary pressure
Safety valve
Flow Rate
Heater Capacity
Gas pressure Regulator

Applicable Sterilizer

Thermostat
Power Source

HS-38
EO-57A

HS-57

Figure (B) Figure (C)

HANSHIN 19
100% EO Gas type Sterilizer_Touch Screen LCD Monitor type

HS- 4 3 1 3EO
100% EO Gas Cartridge
- Convenient and Safe EO Gas Sterilization

Touch Screen LCD Monitor

CYCLE P1 CYCLE P2 CYCLE P3 CYCLE P4

CYCLE P5 VACUUM LEAK TEST CYCLE

20 HANSHIN
▶ This sterilizer uses 100% ethylene oxide gas as sterilant and sterilizes the medical devices, instruments and supplies effectively by
convenient way of only inserting a small disposable gas cartridge on the holder in the chamber. Cycle progressing is fully automated
by the control system with microprocessor.
▶ The sterilization and aeration process are all completed in the chamber under negative pressure, therefore leakage of EO gas in the
working area can be fully prevented. EO gas being sucked by powerful vacuum pump is exhausted to the prepared drainage with
water safely after sterilization process.
▶ 5 kinds of typical cycle programs and the vacuum leak test cycle are installed in the sterilizer and the operator can select a proper
cycle program according to the type of materials or items to be sterilized and sterilize them conveniently.
▶ The embedded self diagnostic program monitors the cycle progress in real time and displays the error code on the touch screen
monitor immediately accompanying an audible alarm when any trouble has occurred during operation, and then aborts the cycle in
progress automatically for safety of workplace.
▶ The built-in panel printer prints out the cycle performance record for analyzing
the results of sterilization and keeping as a quality evidence.

N/A

Cycle No.

Expos. Temp.

Expos. Time (hh:mm)

Aeration Tme (hh:mm)

Use 100% EO Gas Cartridge Inserted View

HS-4313EO

100% EO Gas Cartridge

HANSHIN 21
EO Gas Monitor

HEGM- 1 0 0 0
Keep All the Time Watch On EO Gas Leakage
For Hospital Safety and Environment !

The monitoring system detects the gas in the air correctly by the gas
sensor, therefore the gas sensor must maintain to the normal state
constantly. The gas sensor should be replaced every two years due to
its life. The zero value and span have to be calibrated after replacing a
new sensor to detect and display the gas concentration correctly.
The replacement of senor and it’ s calibration must be performed by an
authorized technician of the manufacturer.

22 HANSHIN
▶ It is essential to install the gas monitor system in the workplace where operating the EO gas sterilizer for preventing a serious danger
caused by EO gas leakage.
▶ This equipment immediately sounds an audible alarm and takes the emergency measures in case of occurring EO gas leakage for
protecting the health of operators in the workplace.
▶ Rapid and accurate response, stable operating characteristic are technical advantages of this system and it’s convenient to apply
to any special condition because user can change the set values of built-in program.
▶ In case of power failure, the built-in rechargeable battery keeps the monitoring system running continuously about for 3 hours.
Furthermore this system is also available for an efficient surveillance monitor of the other gases by connecting with all kinds of
detectors(sensors) having the output signal of 4~20mA.
▶ The built-in microprocessor for controlling the system detects the monitored data in real time and cuts off the gas supply to the
sterilizer automatically if necessary, and starts the compulsory ventilation fan automatically to eliminate the EO gas in the workplace.


Considering the health hazards caused by EO, the exposure level for operators is restricted strictly within OSHA’s Permissible Exposure
Limit(PEL) of 1 ppm measured as air in an 8 hour time-weighted average(TWA), or Short Term Exposure Limit(STEL) of 5 ppm averaged
over a sampling period of 15 minutes.


EO is flammable when its concentration in the air is over 3%, therefore keep all sources of ignition such as matches, lighted cigarettes,
sparks and static discharge away from the sterilizer. Never put electrical items containing batteries in the loads to be sterilized.

HEGM-1000

HANSHIN 23
Cutting & Sealing Machine for Sterilization Bag

HCS- 4 0 0
Cut & Seal the Roll Pouch for Sterilization
with One Machine Safely and Economically
Using an exclusive machine for cutting and sealing the pouch roll for packing items to
be sterilized is strongly recommended for a safe packing because the items must be
damaged or contaminated if the pouches containing them have damaged while
sterilizing or conserving by reason of wrong packing.
This specially designed machine can cut and seal the pouch roll in one unit, therefore
user is able to do all the work for packing the medical instruments to be sterilized
conveniently and economically in a small space.

24 HANSHIN
▶ The entire process is controlled automatically and displayed digitally on the LED display by a microprocessor.
▶ If the cutting length and quantity are already set in the machine, the roll pouch is attracted to the pouch inlet automatically and the
set quantities of pouch that are sealed bottom side are produced after cutting in the set length.
▶ The pouch contained the item to be sterilized is sealed automatically if the upper side of pouch is inserted in the sealing inlet, and
sealing can be done manually by connecting the foot switch.
▶ The set sealing temperature is maintained constantly by precise control.
▶ The appliance table and the roll stand are provided as an optional items.

The error code is displayed on the LED display panel accompanying


an audible alarm in case of occurring an error during operating.

Err 1 Delay of cutting time


Err 2 Overheating of the heater
Err 3 Delay of discharging time
Err 4 Excess of deviation range for sealing temperature
Err 5 Malfunction of the temperature sensor
Err 6 Delay of preheating time
Err 7 Drop of sealing temperature
Err 8 Delay of the automatic discharging time for the roll bag
Err 9 Failure of Inserting the roll bag
Err10 Delay of Inserting time for roll bag

HCS-400

HANSHIN 25
Self-Contained BI Incubator

HI- 5 6 1 5
Convenient and Effective Incubation of
Self-Contained Biological Indicator

HI-5615

▶ Built-in thermometer for reading the incubation temperature.


▶ Temperature selection switch for simple operation.
▶ 15 incubation cavities and adapters for incubating the SCBIs.
▶ An integrated activation (crushing) cavity to simplify activation.
▶ Neat and transparent plastic top cover for preserving a stable and safe incubation environment.