ECN No.
2015-E-0326
ENGINEERING CHANGE NOTICE (FIELD CHANGE NOTICE)
SUBJECT: Releasing Application Software V2.2 for DRYPIX Lite
Design Change Reason Action Time frame
Specification/Function Change Mandatory Immediately
Pro-active Improvement Recommended (by )
Serviceability Not required (Information only) Next maintenance
Part Replacement Other ( ) Upon failure
Corrective Action At your convenience
Preventive Action Not Applicable
Regulatory Compliance Other( )
Other ( )
DATE: Jul. 30, 2015 Dept.: Medical Systems Business Div.
ORIGINATOR: M.Watanabe EDITED BY: N. Shimokawa APPROVED BY: H.Chikugo
AFFECTED PRODUCTS: DRYPIX Lite
TOTAL PAGE (S): 5 ATTACHMENT: YES NO
RELATED INFORMATION: N/A
***********************************************************************************************************************************
DESCRIPTION OF CHANGE:
We would like to release Application Software V2.2 for DRYPIX Lite.
For the details, see the attachment.
- APPLICABLE TO (Serial Nos. etc.): All
- ESTIMATED WORK TIME: 1 HOUR
- SPECIAL TOOLS / INSTRUMENTS: YES NO
- SUPPLY OF THE ARTICLES: YES NO
IF YES, WHEN: Please download from Libinity One..
NAME / CODE: DRYPIX 2000(DRYPIX Lite) Application Software V2.2
/114Y5001003A11
QUANTITY: N/A
- RETURN OF THE REPLACED PARTS TO FUJI: YES NO
DOCUMENTS AFFECTED: YES NO
AFFECTED CURRENT NEW
DOCUMENTS REV ISSUED DATE REV ISSUED DATE
DETAILS:
The contents described or shown in our ECN documents and/or any information provided from FUJIFILM in verbal or written communication is intended for the
addressee and/or its organization. Transfer and/or copy distribution of the information without written consent of FUJIFILM is not permitted.
PAGE 1 of 2 K12N0028-6-J
ECN No. 2015-E-0326
- NOTE
1) Important information about SUPPLY OF THE ARTICLES:
Charge:
FREE OF CHARGE UNTIL (mmddyyyy): N/A
WITH YOUR CHARGE
The contents described or shown in our ECN documents and/or any information provided from FUJIFILM in verbal or written communication is intended for the
addressee and/or its organization. Transfer and/or copy distribution of the information without written consent of FUJIFILM is not permitted.
PAGE 2 of 2 K12N0028-6-J
ECN No. 2015-E-0326a (1/3)
1. Changes
Corrections
Item Overview
Problem of It is identified that in the case both N-CREATE and N-SET are sent to the
overwritten block of Basic Film Box during DICOM Print communication, Min Density and
DICOM Max Density specified by the modality cannot be reflected normally in some
parameters cases. This failure has been corrected with this version of the software.
There are very few modalities that send both N-CREATE-RQ and N-SET-RQ
in the Basic Film Box. (In overseas countries, Examion has reported only one
case occurring in the DR connection.) Therefore, it appears that the failure
was not recognized.
[Description of the failure]
For instance, Max Density is specified as 360 in the N-CREATE-RQ of Basic
Film Box and if there is no specified value on Max Density in the N-SET-RQ
of Basic Film Box, the film output should be performed with Max Density of
360 normally. However, the parameter of Max Density was overwritten by the
default setting* of the printer when receiving N-SET-RQ and film was
outputted with the default value.
Modality Printer
Parameter of Max Density is
Basic Film Box
overwritten when receiving
N-SET-RQ and default setting
of the machine is used for film
N-CREATE-RQ
output.
(Max Density: 360) If the machine setting is other
than 360, there will be no
problem.
N-SET-RQ
(Max Density: not specified)
* Value of [Max. Density] of the LUT No. for the DI Tool client setting
2. Factory shipment in the future
Serial numbers of the devices to be shipped with the new software will be informed when they are
determined.
3. Upgrading procedure
Upgrade the application software according to the following procedure.
3-1. Turn ON the power of the PC for servicing.
ECN No. 2015-E-0326a (2/3)
3-2. Load the application software CD-R into the CD-ROM drive.
"DPXLiteCopyTool.exe" starts automatically and [DRYPIX Lite Copy Tool] window is displayed.
* If it does not start automatically, double-click "DPXLiteCopyTool.exe" in the CD-R and start it.
3-3. Enter the drive name into [Input drive of FTP-Home].
3-4. Click [Copy] of [Please specify the drive of FTP-Home].
The main unit software will be copied in the following FTP folder.
When completed, the [Success.] message box appears.
[FTP Root]:¥Program Files¥FujiFilm¥DRYPIX¥DRYPIXLite¥CD-Image
3-5. Click [OK] on [Success.] message box.
3-6. Click [EXIT] to close [DPX Lite Copy Tool] window.
3-7. Unload the application software CD-R form the CD-ROM drive.
3-8. Connect network cable of the PC for servicing to the network to which the equipment is connected.
3-9. Turn ON the power of the equipment.
3-10. Connect the communication of the PC for servicing to the equipment.
URL : http://[Printer IP Address]:20051/USER/Login.htm
User ID : dryprinter
Password : fujifilm
* URL (default IP address)
http://172.16.1.30:20051/USER/Login.htm
ECN No. 2015-E-0326a (3/3)
3-11. [Upgrading] [Execute]
The application software is transferred (version upgrade) in the equipment. When transferring
completed, the [Command completed.] screen appears.
3-12. End the PC-Utility by clicking the [Logout] button on the maintenance tree.
3-13. Close the logout window.
3-14. Reboot the equipment so that the software is upgraded.
3-15. Check that the equipment whose software version has been upgraded starts and operates normally.
* How to check the software version
[System information and setup] [Setting1] [Check Version]
3-16. Shutdown the PC for servicing.
3-17. Disconnect the network cable of the PC for servicing from the network to which the equipment is
connected.