Ozone Sensor
Basic User Manual
P9.3 CD10.1, OZ10.1, P10.1
October 2023 (EN)
V6
Table of contents
1 Information about these operating instructions ...................................... 4
1.1 Symbols and displays................................................................................ 4
1.2 Associated documents ............................................................................. 5
2 Information on this product.............................................................................. 6
2.1 Product description..................................................................................... 6
2.2 Scope of supply.......................................................................................... 8
2.3 Product overview........................................................................................ 9
2.4 Name plate ............................................................................................. 10
3 Safety ..................................................................................................................... 11
3.1 Use for the intended purpose................................................................. 11
3.2 Use other than for the intended purpose.............................................. 11
3.3 Personal qualifications ........................................................................... 12
3.4 Rebuilding and modifications ................................................................. 12
3.5 Residual risks .......................................................................................... 12
4 Commissioning................................................................................................... 16
4.1 Installation requirements ........................................................................ 16
4.2 Preparation of the sensors .................................................................... 16
4.3 Insertion into the flow chamber ............................................................. 18
4.4 Electrical connection ............................................................................... 19
4.5 Initial calibration........................................................................................ 21
5 Calibration ........................................................................................................... 22
6 Removal................................................................................................................. 24
7 Maintenance ........................................................................................................ 25
7.1 Maintenance overview............................................................................ 25
2 / 43
Table of contents
7.2 Changing the electrolyte and membrane cap ..................................... 25
8 Troubleshooting ................................................................................................ 27
8.1 Fault overview ......................................................................................... 28
8.2 Special checks ....................................................................................... 35
9 Technical data..................................................................................................... 38
10 Deinstallation and storage.............................................................................. 38
11 Disposal................................................................................................................. 38
12 Warranty ............................................................................................................... 39
13 Liability disclaimer ........................................................................................... 39
14 Appendix .............................................................................................................. 40
3 / 43
Information about these operating instructions
1 Information about these operating instructions
1.1 Symbols and displays
1.1.1 Safety and warning instructions
The hazard symbols and signal words listed below are used in these
operating instructions. They help you use the product safely, protect the
operating personnel against injuries and protect the operating company
against damage to property and additional costs.
Signal word Meaning
DANGER means a hazard with a high degree
of risk which if not avoided will lead to death or
DANGER!
serious injury.
WARNING means a hazard with a medium
WARNING! degree of risk which if not avoided may lead to
death or serious injury.
CAUTION means a hazard with a low degree of
CAUTION! risk which if not avoided may lead to minor or
moderate injury.
NOTE NOTE warns against damage to property.
Tab. 1: Signal words
4 / 43
Information about these operating instructions
1.1.2 Labels in the text
Symbol Meaning
This symbol is the general warning symbol and warns you
about risks of injury. Take all the actions that are
indicated by this warning symbol.
This symbol indicates tips and helpful information for
optimum and economic use of the product.
This symbol indicates actions to be performed by the
personnel.
. This symbol indicates the result of an action.
This symbol indicates individual bullet points.
This symbol indicates a precondition before
performing an action.
Tab. 2: Symbols in the text
1.2 Associated documents
Data sheets on the individual types of sensors can be found at supplier.
5 / 43
Information on this product
2 Information on this product
2.1 Product description
The product range consists of sensors with covering membranes.
They are 2-electrode systems for measuring the concentrations in water of
the disinfectants chlorine dioxide, ozone or peracetic acid. The area of
application of these sensors extends to almost all water qualities.
The sensors are not suitable for checking the absence of chlorine
dioxide, ozone or peracetic acid.
A complete measuring and/or control system normally consists of the
following components:
. Sensor
. Electrical leads and connectors
. Flow chambers and connections
. Measuring and control device
. Dosing equipment
. Analytical equipment
These operating instructions relate exclusively to the sensor.
> Comply with the operating instructions for the peripheral
devices.
6 / 43
Information on this product
2.1.1 Chlorine dioxide CD10.1
The sensor measures the concentration of chlorine dioxide in the water
being measured. Such chlorine dioxide arises from the application of
chlorine dioxide (such as the acid/chlorite process, chlorine/chlorite
process). The chlorine dioxide sensor is insensitive to chlorine.
2.1.2 Ozone OZ10.1
The sensor measures the concentration of dissolved ozone in the water
being measured. The sensor is virtually insensitive to chlorine.
2.1.3 Peressigsäure P9.3 and P10.1
The sensor measures the concentration of peracetic acid in the water
being measured resulting from the application of peracetic acid.
7 / 43
Information on this product
2.2 Scope of supply
Keep the all the packaging for the sensor.
In the event of repair or warranty please return the sensor in the
original packaging.
Check that the delivery is complete and undamaged.
If it is damaged:
Please contact your supplier.
Component Quantity Sensor with Sensor with Sensor with
voltage output Modbus signal
4 - 20 mA
signal transmission
current loop output signal
(0…+/-2000 mV) (2-pole Screw terminal (5-pole-M12-
connection) connection)
Sensor with membrane 1
Cap
(depending on the type)
Electrolyte 1 bottle
(depending on the type)
mA hood with O-ring 1
20x1.5 – – –
Special emery paper 1
(depending on the type)
Operating instructions 1
Tab. 3: Scope of supply
8 / 43
Information on this product
2.3 Product overview
Fig. 1: Product overview
1 Electrical connection
2 Sensor body
3 Pressure compensation opening
4 Working electrode
5 O-ring 20 x 1.5
6 Membrane disc
7 Protective cap
8 Membrane cap
9 Electrode finger
10 Reference electrode
9 / 43
Information on this product
2.4 Name plate
A name plate is glued to each sensor, this shows the following
information:
Fig. 2: Example of a name plate
1 Measured variables
2 Sensor designation, sensor name
3 Nominal measuring range of the sensor
4 Permissible temperature range of the measuring water
5 Maximum permissible pressure of the water being measured
6 Signal transmission
7 Power supply
8 Serial number
9 Serial number as barcode
10 / 43
Safety
3 Safety
The sensor is manufactured using the latest technology.
Nevertheless, improper use can give rise to the following risks:
. Effects on health
. Falsification of measured values, which can lead to dangerous dosing of
incorrect quantities of the disinfectant.
Comply with the safety instructions in these operating instructions.
3.1 Use for the intended purpose
The sensor is intended to be used for measuring the concentration of a
specific disinfectant in water.
The sensor may be used only under the following conditions:
. For the disinfectant specified in the respective data sheet
. Under the conditions of use specified on the respective data sheet
. Upright installation in a suitable flow chamber (electrical connection top).
. Restricted to the activities described in these operating instructions.
. Use only when in fault-free condition
. Use of original accessories and spare parts
3.2 Use other than for the intended purpose
The sensor may not be used for measurements to demonstrate the
absence of the disinfectant.
The sensor must not be used for applications beyond the intended
purpose (see section 3.1, p. 11).
11 / 43
Safety
3.3 Personal qualifications
The user must hold the following personal qualifications:
. He must have read and understood the operating instructions.
. He must have received training in the handling of the sensor.
3.4 Rebuilding and modifications
Opening the sensor and making modifications to it which can affect the
safety and functionality of the sensor may be performed only by the
manufacturer.
3.5 Residual risks
3.5.1 Slippage of the sensor
If the sensor is inadequately secured, it may become loose due to the
pressure of the water or due to vibration. This results in the following risks:
. Due to the pressure of the water the sensor may slip out of the flow
chamber.
. The sensor may slip down into the flow chamber due to its own
weight.
Depending on recommendation in the data sheet, use a variant with a
retaining ring (see section 1.2, p. 5).
Make sure that the fixation of the sensor in the flow chamber cannot be
loosened during operation.
Check the sensor regularly for secure attachment.
3.5.2 Water pressure that is high or fluctuating
The membrane may be damaged if the water pressure exceeds the
maximum permissible value, or if the water pressure fluctuates greatly.
Comply with the permissible pressure stated on the data sheet (see section
1.2, p. 5).
Keep the pressure constant.
12 / 43
Safety
3.5.3 Impacts, shocks and improper touching
Impacts or shaking of the sensor, such as by dropping it, can damage it.
Avoid impacts and shocks.
Do not allow the sensor to be dropped.
Do not touch the sensor electrodes at any time.
Touching the reference electrode, or using emery paper on it, can
damage it.
Do not touch the reference electrode.
Should it be necessary for maintenance work, only emery the working
electrode, not the reference electrode.
3.5.4 Defective membrane cap
It the pressure compensation opening is full, this can damage the
membrane beyond repair.
Empty the pressure compensation opening (see section 7.2, p. 25).
3.5.5 Electrical interference
A lack of galvanic isolation can falsify the measured value and even
damage the sensor beyond repair.
Ensure the electrical connection has galvanic isolation.
Electrical interference on the signal lead can damage the electronics.
Ensure the connection is made correctly (see section 4.4, p. 19).
3.5.6 Lack of disinfectant
If for a prolonged period there is no disinfectant in the water, a film of biological
matter can accumulate on the membrane. This makes the measured value
incorrect, and means the membrane cap can no longer be used.
Make sure that the period during which there is no disinfectant
present is not longer than specified on the data sheet (see section
1.2, p. 5).
13 / 43
Safety
3.5.7 Loss of measured values when the sensor is removed
After the sensor has been removed there is no longer a measured value,
which can lead to incorrect dosing of the disinfectant.
Switch off the measurement and control system or switch it over to
manual operation.
3.5.8 Oxidant
Oxidants in the water interfere with measurement and can lead to
measuring errors.
Make sure there are no disruptive oxidants in the water.
Comply with the instructions on the data sheet (see section 1.2, p. 5).
3.5.9 Temperature and fluctuations in temperature
If the ambient temperature or the temperature of the medium lies outside the
permissible range, the sensor and the electrolyte may be damaged.
Make sure that at all stages of the operation the permissible
temperature ranges as per the data sheet are adhered to (see
section 1.2, p. 5).
The measured value may be incorrect if the temperature in the medium
fluctuates abruptly.
Make sure that the temperature in the water changes only slowly.
3.5.10 Impermissible installation position
If the sensor is not installed upright the measured value can be falsified.
Install the sensor upright (electrical connection top).
14 / 43
Safety
3.5.11 Incorrect chemical analytical methods
Incorrect determination of the concentration of the disinfectant will lead to
incorrect calibration of the sensor.
Observe the recommended analysis methods as per the data sheet
(see section 1.2, p. 5).
Perform analytical work in accordance with the specifications in the
manufacturer’s operating instructions for the analytical equipment.
15 / 43
Commissioning
4 Commissioning
4.1 Installation requirements
The following installation requirements must be satisfied:
. Continuous power supply and presence of water being measured
. Flow rate as specified on the data sheet
. Constant through flow rate
. There must be disinfectants present in the water being measured.
. Galvanic separation of the electrical connection (if not present in the
sensor, see data sheet, section 1.2, p. 5)
. Make sure that the measuring water does not evolve gas at the
measurement point.
4.2 Preparation of the sensors
Pull the protective cap [3] off the membrane cap [2].
Unscrew the membrane cap [2] from the sensor body [1].
Fig. 3: Unscrewing the membrane cap
1 Sensor body
2 Membrane cap
3 Protective cap
16 / 43
Commissioning
Put down the membrane cap on a clean surface.
Fill the membrane cap with electrolyte to the brim.
Fig. 4: Filling the membrane cap
Place the sensor body [1] upright on the membrane cap [2].
Rotate the sensor body anticlockwise until the thread is felt to
engage.
Fig. 5: Placing the sensor body on the sensor cap
1 Sensor body
2 Membrane cap
Slowly screw the sensor body into the membrane cap.
> Make sure that the membrane cap is fully screwed into the
sensor body, right up to the stop.
Once the membrane cap has been fully screwed on:
Neither touch nor strike the membrane.
Use mains water to rinse off any electrolyte residues adhering to the
sensor.
The sensor is now prepared for commissioning.
17 / 43
Commissioning
Fig. 6: Prepared sensor
4.3 Insertion into the flow chamber
The sensor must have been prepared for installation (see section 4.2, p.
16).
Insert the sensor into a flow chamber or any other suitable flow
chamber.
In order to insert the sensor correctly into the flow chamber:
Comply with the instructions in the operating instructions for the flow
chamber that is used.
18 / 43
Commissioning
4. 4 Electrical connection
The sensor is inserted into the flow chamber (see section 4.3, p. 18). The
following types of electrical connections to the sensor are available:
4.4.1 Connection with 0…+/-2000-mV signal output
The sensor is provided with a 5-pin M12 screwed plug protected against
polarity reversal.
The connection pins are assigned as follows:
Fig. 7: Connection pin assignment (5-pin)
Analog signal processing Digital signal processing
0…-2000 mV 0…+2000 mV 0…+/-2000 mV
1 Measuring signal Measuring signal Measuring signal
2 +U +U +U
3 -U Power GND Power GND
4 Signal GND Signal GND Signal GND
5 (not assigned) (not assigned) (not assigned)
19 / 43
Commissioning
4.4.2 Connection with 4…20 mA signal output
M12 screwed plug
The sensor is provided with a 5-pin M12 screwed plug protected against
polarity reversal.
The connection pins are assigned as follows:
Fig. 8: Connection pin assignment (5-pin)
1 (not assigned)
2 +U
3 -U
4 (not assigned)
5 (not assigned)
Connection with a 2-pole screwed terminal block
The sensor is provided with a 2-pole screwed terminal block.
Insert the sensor cable through the cable gland of the
hood.
Connect the cores to the terminals in the sensor electronics.
Screw the hood finger-tight into the sensor body until the O-
ring seal is made.
Tighten the cable gland so as to secure the cable.
20 / 43
Commissioning
4.4.3 Connection with Modbus signal transmission
The sensor is provided with a 5-pin M12 screwed plug protected against
polarity reversal. There are no termination resistors within the sensor.
The connection pins are assigned as follows:
Fig. 9: Connection pin assignment (5-pin)
1 (reserved)
2 +9…+30 V
3 Power GND
4 RS485 B
5 RS485 A
4.5 Initial calibration
The sensor must have been connected electrically (see section 4.4, p.
19).
The run-in time has been complied with as per the data sheet (see
section 1.2, p. 5).
Perform calibration (see section 5, p. 22).
After one day, repeat the calibration.
21 / 43
Calibration
5 Calibration
The sensor outputs a signal proportional to the concentration of the
disinfectant in the water being measured. In order to assign the value of the
sensor signal to the concentration of the disinfectant in the
water being measured, the sensor must be calibrated.
The flow rate must be constant.
The temperature of the water being measured must be constant.
Acclimatisation of the temperature of the sensor to that of the water being
measured must be complete (this takes about 20 minutes after a change
in temperature).
The sensor must have completed running in.
No other oxidant may be present in the water being measured.
Take the analytical sample of the water being measured from near to the
sensor.
Using appropriate methods, determine the concentration of the
disinfectant in the measuring water (see the manufacturer’s operating
instructions for the analytical equipment).
In the calibration menu of the measuring and control device, mark up the
sensor signal against the value determined by the analytical
procedure (see the operating instructions for the device).
Repeat the calibration at regular intervals (see section 7.1, p. 25).
Comply with the applicable national regulations for calibration
intervals.
22 / 43
Calibration
Measured variables Recommended analytical methods
Chlorine dioxide DPD-1 Photometer for chlorine
dioxide
DPD-1 + DPD-3
Ozone Photometer for ozone
DPD-4
Peracetic acid Two-stage sulphuric acid titration with
potassium permanganate and sodium
thiosulphate (for the procedure see appendix)
Tab. 4: Recommended analytical methods
Chlorine dioxide can also be determined using a photometer intended for
chlorine. The result must be multiplied by a factor of 1.9.
Ozone can also be determined using a photometer intended for
chlorine. The result must be multiplied by a factor of 0.68.
At higher concentrations of disinfectant the DPD colouration may fail to
appear.
23 / 43
Removal
6 Removal
Removal of the sensor can lead to an incorrect measured value at the
input to the measuring and control device, which can cause the control
circuit to apply uncontrolled dosing.
Before removing the sensor:
Switch off the measurement and control system or switch it over to manual
operation.
Close the inlet of the water being measured.
Close the outlet of the water being measured.
Remove the electrical connection.
To disconnect a sensor with a 2-pole screwed terminal block:
Undo the cable gland.
The cable is now free to move.
Unscrew the hood with the cable gland from the sensor.
Release the cable cores from the terminals.
Undo the screw fastening and carefully pull the sensor out.
24 / 43
Maintenance
7 Maintenance
7.1 Maintenance overview
To ensure optimum performance of the sensor:
Perform the following actions at regular intervals.
Maintenance task Interval
Change the electrolyte 3-6 months
Change the membrane cap b Annually
Perform calibration . Weekly
. After the electrolyte and/or the
membrane cap has been
changed
Tab. 5: Maintenance overview
7.2 Changing the electrolyte and membrane cap
Unscrew the membrane cap.
Empty the electrolyte out of the membrane cap.
Rinse the membrane cap with mains water.
Rinse the electrode finger with mains water.
Shake the sensor body dry several times (see Fig. 10, p. 25).
This empties the pressure compensation opening [1].
Fig. 10: Shaking the sensor body dry
1 Pressure compensation opening
25 / 43
Maintenance
Lay a piece of special emery paper on a clean, smooth surface.
Hold the sensor upright.
Hold the special emery paper in place and move the tip of the
working electrode over it at least twice. Use a fresh area of the emery
paper for each pass.
Fig. 11: Emerying the working electrode
When changing the membrane cap:
> Use a new membrane cap.
Perform the same operations as for commissioning (see section 4.2, p.
16).
Maintenance has now been completed and the sensor can be put
back into use.
26 / 43
Troubleshooting
8 Troubleshooting
Various factors in the environment can affect the sensor. If irregularities occur,
it may be useful to check these factors:
. Flow rate
. Measuring cable
. Measuring and control device
. Calibration
. Dosing equipment
. Concentration of the disinfectant in the dosing container
. Suitability of the sensor for measuring the disinfectant that is being
dosed
. Concentration of the disinfectant in the water being measured
(determined by analytical methods)
. pH value of the water being measured
. Temperature of the water being measured
. Pressure in the flow chamber
. Analytical methods
27 / 43
Troubleshooting
8.1 Fault overview
Fault Cause Corrective action
Sensor cannot Run-in time too short. Wait until the run-in
be calibrated / time has elapsed (see
the measured section 4.5, p. 21).
value deviates Repeat the calibration
from the after a few hours.
analytical
Membrane torn. Change the
measurement
membrane cap (see
section 7.2, p. 25).
Membrane cap Change the
damaged. membrane cap (see
section 7.2, p. 25).
Disruptive substances in Check the water for
the water contents disruptive substances
(observe data sheet).
Provide remedy.
Consult the supplier.
Short circuit / defect in Locate and eliminate
the measuring lead the short circuit /
defect.
Exchange the
measuring lead.
Distance between Screw the membrane
working electrode and cap on fully to the
membrane is too great. stop.
28 / 43
Troubleshooting
Fault Cause Corrective action
The DPD/titration Use fresh
chemicals are out of DPD/titration
date. chemicals.
Repeat the calibration
(see section 5, p. 22).
Unsuitable titration Repeat the titration
method1 using a suitable
method (see
appendix, p. 43).
Deposits on the Change the
membrane membrane cap (see
section 7.2, p. 25).
Air pockets between the Unscrew the
membrane/ membrane cap.
working electrode Repeat
commissioning (see
section 4.2, p. 16).
Gas bubbles on the Temporarily increase
outside face of the the flow rate.
membrane Check the installation
and modify it.
No electrolyte in the Fill the membrane cap
membrane cap with electrolyte.
Prepare the sensor
(see section 4.2,
p. 16).
1 Suitable only for peracetic acid/hydrogen peroxide
29 / 43
Troubleshooting
Fault Cause Corrective action
The concentration of Check the system.
disinfectant exceeds the Remedy the faults.
upper limit of the
Repeat the calibration
measuring range. (see section 5, p. 22).
Lack of galvanic Create galvanic
isolation isolation.
Return the sensor to
the supplier for
checking /
reconditioning.
The sensor is defective. Return the sensor to
the supplier for
checking /
reconditioning.
Unstable Membrane torn. Change the
measured value membrane cap (see
section 7.2, p. 25).
Air pockets between the Unscrew the
membrane/ membrane cap.
working electrode Repeat
commissioning (see
section 4.2, p. 16).
Gas bubbles on the Temporarily increase
outside face of the the flow rate.
membrane Check the installation
and modify it.
Pressure fluctuations in Check the type of
the measuring water installation and modify it.
30 / 43
Troubleshooting
Fault Cause Corrective action
Lack of galvanic Create galvanic
isolation isolation.
Return the sensor to
the supplier for
checking /
reconditioning.
The reference electrode Return the sensor to
is exhausted and/or the supplier for
contaminated.2 checking /
reconditioning.
Overdriving3 Excessive concentration Check the system.
of disinfectant in the Remedy the faults.
measuring water
Calibrate the sensor
(see section 5, p. 22).
Perform maintenance
on the sensor (see
section 7, p. 25).
Run-in time too short. Wait until the run-in
time has elapsed (see
section 4.5, p. 21).
The membrane is Change the
damaged. membrane cap (see
section 7.2, p. 25).
Flow rate too high Check the system.
Reduce the flow rate.
2 The reference electrode has a silvery sheen or is white. The usual
colour on the other hand is brown/grey.
3 See Tab. 7, p. 34
31 / 43
Troubleshooting
Fault Cause Corrective action
Lack of galvanic Create galvanic
isolation isolation.
Return the sensor to
the supplier for
checking /
reconditioning.
The sensor is defective. Return the sensor to
the supplier for
checking /
reconditioning.
Underdriving4 Run-in time too short. Wait until the run-in
time has elapsed (see
section 4.5, p. 21).
The working electrode is Perform maintenance
contaminated. on the sensor (see
section 7, p. 25).
Lack of galvanic Create galvanic
isolation isolation.
Return the sensor to
the supplier for
checking /
reconditioning.
The sensor is defective. Return the sensor to
the supplier for
checking /
reconditioning.
Green LED Defective Provide the correct
flickering or Power supply power supply.
failing to light up5
4 See Tab. 7, p. 34
5 Only for sensors with digital electronics
32 / 43
Troubleshooting
Fault Cause Corrective action
The sensor is defective. Return the sensor to
the supplier for
checking /
reconditioning.
No signal The sensor is connected Connect the sensor
to the measuring and correctly to the
control device with the
measuring and control
wrong polarity.6
device.
The measuring lead is Exchange the
broken. measuring lead.
The sensor is not Provide the correct
receiving any power power supply.
supply.
The sensor is defective. Return the sensor to
the supplier for
checking /
reconditioning.
Tab. 6: Faults overview
6 Only for sensors with 4…20-mA signal output
33 / 43
Troubleshooting
Electronics Signal Underdriving Overdriving
transmission
Analogue 4 … 20 mA <4 mA >20 mA
0 … +2000 mV <0 mV >+2000 mV
0 … -2000 mV >0 mV <-2000 mV
Digital Modbus RTU <0 ppm/ % Measured value >
<0 mA Measurement range
0 … +2000 mV Orange LED . >+2000 mV
lights up7 . Orange LED
flashes
regularly
0 … -2000 mV Orange LED . <-2000 mV
lights up7 . Orange LED
flashes
regularly
Tab. 7: Output signal of the sensor when over driven/under driven
7 The displayed output signal must be multiplied by a factor of -1.
34 / 43
Troubleshooting
8.2 Special checks
8.2.1 Tightness of the membrane cap
Unscrew the membrane cap from the sensor (see section 7, p. 25).
Dry the outside of the membrane cap.
Prepare the membrane cap (see section 4.2, p. 16).
When screwing the membrane cap on, watch out for liquid escaping
through the membrane.
If liquid does escape through the membrane:
Use a new membrane cap.
If the sensor does not respond:
Return the sensor to the supplier for checking.
8.2.2 Electronics
Unscrew the membrane cap.
Rinse the electrode finger with mains water.
Using a clean cloth, carefully dry the electrode finger.
Connect the sensor to the measuring and control device.
Connect a suitable measuring device to the original sensor signal.
Wait five minutes.
Read the original sensor signal at the measuring device.
Mark up the values that were read against the following target values:
. Sensor (mV): approx. +/- 0 mV
. Sensor (mA): approx. 4 mA
. Sensor (Modbus): approx. 0 ppm or 0%
If the sensor signal approximately corresponds to the aforementioned value,
the electronics can be provisionally regarded as OK.
If the measured value deviates significantly from the above value:
Return the sensor to the supplier for checking.
35 / 43
Troubleshooting
8.2.3 Checking the zero point
The electronics must have been tested and found to be OK.
Prepare the sensor (see section 4.2, p. 16).
Connect the sensor to the measuring and control device.
Fill a glass beaker with mains water (without any disinfectant!).
Stir the sensor round in the glass beaker for 30 seconds.
Carefully put the sensor down obliquely in the glass beaker.
Wait 30 minutes.
Read the measured value.
If the measured value is close to the value 0, the zero point can
provisionally be regarded as OK.
If the measured value deviates significantly from zero:
Perform maintenance on the sensor (see section 7, p. 25) and repeat the
zero point test.
A freshly cleaned working electrode has a relatively high zero point. The
sensor takes a few days to settle back to its lowest zero point.
If after maintenance has been performed on the sensor measured value is
not close to zero:
Return the sensor to the supplier for checking.
This completes the zero point checking.
36 / 43
Troubleshooting
8.2.4 Signal
The zero point checking must have been performed successfully.
Add the relevant disinfectant to the mains water in the glass beaker
(see section 8.2.3, p. 36).
Stir the sensor steadily round in the glass beaker for five minutes.
Monitor the measured value throughout this time.
If the measured value increases, the sensor can provisionally be
regarded as OK.
If the measured value does not change:
Perform maintenance on the sensor (see section 7, p. 25) and repeat the
signal test.
This completes the signal test. The sensor can be put back into use. If
after maintenance the sensor shows no response to the disinfectant:
Return the sensor to the supplier for checking.
37 / 43
Technical data
9 Technical data
Information on the technical data can be found at the supplier.
10 Deinstallation and storage
To deinstall a sensor and prepare it for storage, proceed as follows:
Unscrew the membrane cap.
Use mains water to rinse the electrolyte out of the membrane cap.
Shake the sensor body dry several times (see Fig. 10, p. 25).
Rinse the electrode finger with mains water.
Dry the membrane cap and sensor body in a dust-free place.
For protection, loosely screw the dry membrane cap onto the sensor body.
Make sure that the membrane is not lying in contact with the working
electrode.
If the membrane cap has been in use for one day or longer we
recommend that it is not used when the sensor is recommissioned.
> Perform a change of membrane cap (see section 7.2, p. 25).
11 Disposal
Comply with the local regulations on disposal.
38 / 43
Warranty
12 Warranty
We grant a manufacturer’s warranty of two years on the sensor body and the
electronics, subject to correct handling. The warranty does not apply to the
membrane cap (wearing part), electrolyte (expendable material)
and service work to be performed (cleaning the parts in contact with the
electrolyte, renewing the reference electrode and cleaning the electrode tip
with fine grade emery paper). If there is mechanical damage or the serial
number is illegible, the warranty becomes void.
Returning a sensor for checking/factory reconditioning:
Shipments will be accepted only if they are returned carriage paid.
Otherwise it will be returned to the sender.
Each return shipment must be accompanied by a completed declaration of
clearance. missing or not completed declaration of clearance, a cleaning fee
will be charged.
On checked/reconditioned sensors we grant a warranty of one year from the
date of checking/reconditioning. The warranty is on the electrode
body and the electronics, subject to correct handling.
If there is mechanical damage or the serial number is illegible, this
warranty becomes void.
13 Liability disclaimer
The sensor is manufactured with great care and is subjected to a
documented function test. Should any malfunctions occur in the sensor
despite this, no liability claims may be lodged against the manufacturer for
damages resulting from this malfunction.
39 / 43
Appendix
14 Appendix
Recommended titration procedure
For the determination of hydrogen peroxide (H2O2), only the 1st titration
stage need to be performed.
For the determination of peracetic acid (PAA) the 1st and 2nd titration
stages need to be performed.
Procedure:
. The titration must be performed quickly.
. After the first addition of potassium permanganate it may take a few
seconds before the sample shows colour.
. Renewed colouration of the sample after completion of the 2nd
titration stage is disregarded.
. In the 1st titration stage, take care to avoid a large excess of
potassium permanganate (intense violet colouration of the sample).
In the 2nd titration stage the potassium permanganate will be
codetermined as peracetic acid and thus falsifies the results.
40 / 43
Appendix
H2O2 H2O2 H2O2
Peracetic acid Peracetic acid Peracetic acid
Concentration 0…200 >200…2000 >2000…20000 (2%)
range in ppm
Material . 25 ml sample of the water being . 25 ml sample of the water being . 5 ml sample of the water being
measured measured measured
. Sulphuric acid (25%) . Sulphuric acid (25%) . Sulphuric acid (25%)
. Potassium permanganate (0.01 n) . Potassium permanganate (0.1 n) . Potassium permanganate (0.1 n)
. Potassium iodide 0.3-0.5 g . Potassium iodide 0.3-0.5 g . Potassium iodide 0.3-0.5 g
(powder) (powder) (powder)
. Thiosulphate (0.01 n) . Thiosulphate (0.1 n) . Thiosulphate (0.1 n)
. Starch solution (1%) . Starch solution (1%) . Starch solution (1%)
1st titration Add 20 ml sulphuric acid to the sample of the water being measured.
stage Stirring continuously, titrate with potassium permanganate until the sample colour turns faintly violet.
H2O2 Make a note of the consumption (A) of potassium permanganate in ml.
41 / 43
Appendix
H2O2 H2O2 H2O2
Peracetic acid Peracetic acid Peracetic acid
2nd titration After the 1st titration stage, add potassium iodide.
stage Stirring continuously, titrate with thiosulphate until the sample colour turns pale yellow.
PAA Add 2 ml starch solution.
The sample colour turns blue.
Stirring continuously, titrate with thiosulphate until the sample turns colourless.
Make a note of the consumption (B) of thiosulphate in ml.
Calculation of A * 6.8 A * 68 A * 340
the H2O2 = concentration in ppm H2O2 = concentration in ppm H2O2 = concentration in ppm H2O2
Calculation of B * 15.2 B * 152 B * 760
the PAA = concentration in ppm PAA = concentration in ppm PAA = concentration in ppm PAA
42 / 43