Microplate Reader
Easy operation
Convenient and efficient
Product introduction
The BGMR-1000 is an enzyme-linked microplate reader developed by Bio-Gener company. This product uses the
principle of optoelectronics to perform enzyme immunization measurement of samples. The detection wavelength
is 300-800nm. Specific application. The detection wavelength is 400-750nm, which is suitable for various applications
such as enzyme dynamic research. The instrument can be measured quickly and accurately, with the fastest single
wave length of 6 seconds/96 wells, and the double wave length 10 seconds/96 wells. The instrument also comes
standard with a linear shock plate function. The type incubation temperature with the incubator can be as high as
50 . Visualized workflow, simple data analysis and export function can maximize your experience. This product
can be widely used in the fields of research, agriculture, animal husbandry, feed companies, food companies and
other testing.
10.1 -inch touch screen, can operate, store, print without external computers
Panel visualization operation, data displayed completely
8 -channel optical filter measurement system, you can manually replace the filter
You can measure the end point method, dynamics, suppression rate, etc.
Three linear vibration board functions
The reagent is open and there is no specific restrictions.
Store standard curve automatically, users can correct the standard curve
Choose auction of inhalation, CUT-OFF qualitative calculation, single-point fixed standard,
folding line regression, linear regression, index return, pair return, double-pair regression,
log-logit, power regression, four parameter regression, and other calculation methods
Provide a variety of printing methods and reports, and can customize various reports
Layout and import functions, greatly improving the test efficiency.
Model BGMR-1000
Light source 6v10W long life halogen lamp
wavelength 300-800nm
Four filters of 405, 450, 492 and 630nm are equipped as standard,
and up to eight filters can be loaded
Filter
Filter width 6-8nm
Absorbent 0.000-4.000 OD
Resolution 0.001 OD
The correlation coefficient is greater than 0.999 (0,2.0),
and the correlation coefficient is greater than 0.99 (2.0,4.0)
Linearity
Wavelength error ≤±3nm
≤±1.5nm
Wavelength
repetitiveness
Repetitiveness ≤0.3%(0,3.0),≤2%(3.0,4.0)
Stability ≤0.3%(0,3.0),≤2%(3.0,4.0)
Accuracy ≤±0.005 OD(0,2.0),≤±1%(2.0,3.0),≤±1.5%(3.0,4.0)
Sensitivity sensitivity≥0.010 OD
Channel difference <0.01 OD (1.0 OD standard filter)
Measuring speed Single wavelength 6 seconds/96 wells, double wavelength 10 seconds/96 wells
Vibrating plate 3 speeds (high, medium, low)
Temperature range RT+5℃~65℃
Control Method 10.1 -inch touch screen, embedded software
Data export excel,word
Print Report can be printed (optional USB thermal printer)
Communication A type USBx2,B type USB,Ethernet port,serial port
Dimension 300×470×200mm
weight 13KG