5E-MAC6710 Service Manual
5E-MAC6710 Service Manual
Service Manual
KY2.789.120AS
The Service Manual introduces the installation and diagnostics of this instrument.
Maintenance personnel should read and fully understand this manual and follow all the
precautions and warnings in this Manual. CKIC assumes no responsibility for any
accident or damage caused by failure, or use in disregard of instructions in this Manual.
Maintenance personnel shall have professional technical training by CKIC and have the
skill to complete the relevant operations under the guidance of the Manual.
• Ensure the equipment is well grounded. Do use grounding plug and socket. Do NOT
change the original standard power cord.
• The thunder and lightning may cause a risk on you. Do not operate this equipment
during thunderstorm and it must be disconnect from facility power and external
internet wire.
• Regularly check the power cord and power plug and make sure they are in a good
condition and pressed by other components.
1
1.1.2 Safety of Gas Supply
• When the gas required (O2, N2, He, CO2, mixed gas of CO and CO2) is high
pressure gas, contacts with flammable and explosive materials are not allowed in
operations, otherwise there may be fire or violent explosion.
• When using the high-pressure gas cylinder, the gas source must be kept away from
heat and open flames. Operations must be strictly in accordance with the safety
procedures of high-pressure cylinders to ensure there is no gas leakage to avoid
the explosion, personal poisoning, suffocation and other dangerous occurrences.
2
1.2. Caution Symbols
The following symbols may be found on CKIC equipment or their components. These
symbols indicate the use of specific safety guidelines. Operator and service personal
must follow these instructions for personal safely and to prevent damage to the
equipment.
3
1.3. Package
After receiving instrument, please check the packing and items in your package. For any
damage or lack of components, please contact forwarder and CKIC to confirm
responsibility. The packing list for MAC6710 is as follows.
Note: All components listed below are subject to change without prior notice.
Please refer to the packing list received with instrument.
4
Standard Configuration List
LASERJET PRO
2 402003 Laser Printer 1 PCS Optional
P1106
3 110911 5E-MAC6710 5E-MAC6710 1 Set
2)Main Accessories
6
instruments of
CKIC
50 403193 Instruction Manual 5E-MAC6710 PCS
51 Software 5E-MAC
7
2. Introduction
2.1. Application
When it is used to analyze the moisture, ash, volatile matter and fixed carbon in coal, it
conforms to the following standards:
When it is used to analyze the moisture, ash, volatile matter and fixed carbon in coke, it
When it is used to analyze the ignition loss of cements, it conforms to the following
standards:
Equipped with infrared heating equipment and electronic balance to weigh the sample
during heating under special atmosphere, specified temperature and specified time to
calculate the moisture, ash and volatile etc. of the sample.
• Calibration of standard coal: when the reference material is used for analysis,
the analysis results should be within the uncertainty of the reference material.
The instrument consists of one analyzer, one PC and one printer. (See figure 2-1)
9
Analyzer
Monitor
Furnace SCM
RS232
Balance Gearing
PC
Mouse
Printer
Keyboard
Figure 2-1
10
2.5. General Structure
2.5.1. Inner Structure (see figure 2-2)
Crucible
Furnace
Carousel
Positioning Pillar
Rotating Shaft
Ventilation Ring
Balance Arm
Y-type Connector
Gas Cylinder
Rotating Mechanism
Balance Thermocouple
Function:
1 Crucible Load the sample.
2 Carousel Hold the crucibles.
3 Balance Arm Support the crucible during the weighing.
4 Furnace Heat the sample.
5 Balance High precision: 0.0001 g.
Support the carousel and connect the carousel and rotating
6 Ceramic Shaft
mechanism.
7 Positioning Pillar Connect the carousel and ceramic shaft.
8 Thermocouple Collect the internal temperature of high temperature furnace.
Rotating
9 Rotate mechanism for carousel.
Mechanism
10 Gas Cylinder Automatically open and close the furnace lid.
11 Ventilation Ring Connect the gas uniformly to the furnace.
11
2.5.2 Front Panel (Figure 2-3)
Flowmeter
Lid OPEN/CLOSE
Y-Type Connector
Gas Cylinder
Rotary Mechanism
Balance Thermocouple
Figure 2-3
Function:
12
2.5.3 Rear Panel
Smoke Exhausting Fan
Cooling Fan
Heating Switch
Figure 2-4
Function:
Discharge the smoke and dust generated in the
1 Smoke Exhausting Fan
analysis.
Upper Furnace Power Input power supply for the heating parts of upper
2
Input furnace.
3 Cooling Fan Cool down the instrument.
4 Heating Switch Control the power of heating parts.
Upper Furnace Power Output power supply for the heating parts of upper
5
Output furnace.
Connect compress gas , whose gas pressure is
6 Gas Cylinder Inlet 0.2~0.3MPa.The gas pressure can be adjusted to
control the speed of opening and closing furnace lid.
Input power supply for circuit boards and other
7 Circuit Power Supply
components.
8 Heating Power Supply Input power supply for heating furnace.
Connect the O2/Compress Gas to offer gas for the
9 O2/Compressed Gas Inlet
analysis.
13
10 N2 Inlet Connect N2 to offer gas for the analysis.
Connect the RS232 Serial Port to the PC Port with
11 RS232 Serial Port
the serial cable.
12 Pump Power Supply Input power supply for the pump components.
14
3. Installation
Insert the disk into the CD driver, enter “5E-MAC” program, double click the “Setup.exe”,
and then click Next to complete the installation.
15
(Figure 3-1)
Click Next to pop up the dialog box as figure 3-2.
Figure 3-2
16
Click Next to pop up the dialog box as figure 3-3
Figure 3-3
Click Finish. If the operating system has not installed the Data Access, double click the
“setup.exe” in the “Data Component Installation Disk”. Click Next according to the
prompt to finish the installation. Click Restart to restart the computer.
Main Interface
The main interface includes Title Bar, menu Bar, Manufacturer Bar, and Status Bar as
the figure below:
17
Figure 3-4
1. Title Bar——Display the software title. In this dialog box, the title is 5E
Automatic Proximate Analysis System.
2. Menu Bar——Menu bar includes Settings, Start analysis, Function, View,
Data Management, Help. When the contents under the menu is gray, the
menu is not valid temporarily;
3. Tool Bar——Tool bar includes System Settings, Start Analysis, Data Query,
Calculation, Exit.
4. Status Bar——Display the parameters and working status.
18
[Settings(S)] Menu
\\\\\\
Figure 3-5
After entering [System Settings], it will display the dialog box as shown in Figure 3-6.
1. Moisture analysis includes normal method and quick method. The operator
can select the method according to needs.
2. Ash analysis includes normal method and quick method. The operator can
select the method according to needs.
3. The sample weighing method of ash and moisture includes continuous and
separate weighing. The operator can select the method according to needs.
4. The print method includes default and Excel. The operator can select the
method according to needs.
5. Analysis Settings/Print Report : not print, proximate analysis result, etc. The
operator can select the mode according to needs.
6. Analysis Settings/Numbering Method : Auto/manual. The operator can select
the method according to needs.
7. Mad% Default: When the moisture test value is negative, the display and
reference conversion are used.
8. Analysis content: include coal / coke / ash, measured ash combustible,
measured calcined petroleum coke, measured loss of cement, iron ore loss,
and measured biomass fuel.The operator can select moisture analysis , ash
analysis or volatile analysis according to needs.
9. Temperature Calibration: include moisture calibration, ash calibration, furnace
calibration of volatile matter, and oven calibration of volatile matter.
10. The atmosphere setting of "moisture analysis" includes: "compression pump"
19
and "nitrogen bottle". The parameters should be set according to the gas
source configuration.
11. The residue weighing of moisture and ash are used to set the precision and
weighing interval.
12. Instrument Model is used to set the model of instrument.
13. Temperature control and time control is used to set the temperatures, time,
and heating rate.
Figure 3-6
It is used to select the balance type and serial port through which a computer
communicates with the analyzer. The serial port includes Com1/ Com2/ Com3/ Com4.
The balance type includes BS224S (Sartorius)/ WKE204(METTLER)/ ATX224
20
(SHIMADZU)/ CP214(OHAUS).
Figure 3-7
[Transmit Settings]
Click transmit settings to pop up the dialog box.
This menu is used to set the upload serial port, files to be transmitted, etc.
Figure 3-8
21
Database Password Settings (K)
Figure 3-9
Report Settings(R)
Click Settings-Report Settings to pop up the interface as figure 3-10.
Report settings is mainly used to set the items to be printed. Meanwhile , the report
title, instrument model, standards and so on can also be modified.
22
Figure 3-10
Print (P)
Print is used to print the results(It can be selected by mouse).
Click Print to pop out the following dialog box.
23
Figure 3-11
Print Preview(V)
Figure 3-12
24
User Settings
Click User Settings to set the information of users.
Exit (X)
Click the Exit menu to pop up the dialog box. Click [Yes] to exit from the program
immediately.
Start Analysis(E)
Figure 3-13
Click the [Start Analysis(E)] menu to enter the sample adding interface.
Input the crucible location, sample amount, sample name, operator name, etc.
Note: 19 samples only per batch. Click Next to enter the analysis interface. In the
analysis interface, operator can weigh the empty crucible, weigh the sample, analysis, or
stop the analysis according to needs.
25
Figure 3-14
Figure 3-15
[Function(F)] Menu
26
Figure 3-16
The debugging function of the program helps debug the analyzer and test whether
hardware works normally. The hardware debugging window includes Carousel “Move
Up”, “Move Down”, “Rotate”, “Stop”, Nitrogen control”, “Test Heating Control”, “Fan
Control”, etc.
27
Figure 3-17
Online Aging
Click Function-Online Aging.
This function used to test the status of valve, carousel, etc.
28
Figure 3-18
29
图 3-19
Ash Analysis Calibration by Normal Method
Figure 3-20
30
Volatile Analysis Calibration by Quick Method
图 3-21
31
Figure 3-22
Furnace Temp. Calibration(L)
Figure 3-23
32
It is used to burn and clean the crucibles.
View (V)
It is used to display or hide the Status Bar, Tool bar and Statistic Window
Figure 3-24
Data Manager(M)
33
Figure 3-25
The query condition window is selected according to the conditions of the test, the data
backup and data import. The query results are shown as the figure below. The query
result can be printed or previewed directly or generate the report form and export data.
Figure 3-26
Figure 3-27
Help(H)”
34
Figure 3-28
Press F1 key to view the on-line help, the detailed operation instructions, and click on
“About” sub-menu to view the version and copyright information.
Note:
The power supply of the heating part should be 25A.
In the debugging process, when the power supply and heating power supply
disconnect, the circuit breaker should be kept off.
3. Connect the host RS232 serial port to the computer's RS232 serial port.
Temporary fastening device is used to temporarily tighten some parts to prevent bumps
from causing damages to parts.
Please pay attention to the warning signs on the parts before lifting the rotating
mechanism and operate according to the figure below. Remove the fixed seat first and
then carry out the debugging.
35
Figure 3-29
Take out the pump components from the box first. Please pay attention to the signs and
labels on the parts, and remove the fastening screws in the bottom according to the
figure below to return the four rubber pads to the natural status.
36
Figure 3-30
Remove the balance arm and balance bracket (Figure 3-31), then open the left cover of
the analyzer, and put the balance in the analyzer—the three grounding bolts of the
balance should be inserted into the three holes for locking the balance bolts properly,
then install the balance pan bracket, balance pan to the balance in turn, then insert the
balance power wire and signal wire into the power interface and data interface
correspondingly. The position where balance locates can be regulated by adjusting the
height of three lock bolts inside the analyzer, and level of the balance can be adjusted by
the four grounding bolts of the analyzer.
37
Balance Arm
Shielding Cover
Fastening Screw
Shielding Ring
Adjusting Screw
Fastening Screw
Flat Gasket
Nut
Supporting Board
Pillar
Figure 3-31
38
D. The glass vent pipe is long legs downward and is inserted into the furnace body,
and the long legs are inserted into the reserved hole beside the thermocouple until
the glass air tight ring pipe is placed flat on the ceramic plate at the bottom of the
furnace body
E. Connect the glass vent pipe to the blue gas tube with a hose.
Carousel
Shaft
Pressure Block
Index Plate
Pressure Plate
Bearing
Connecting Shaft
Gear Motor
Rotating Motor
Camshaft
Figure 3-32
39
3.10 Balance Arm
3.10.1 Installation
Put the balance base on balance pan and cover it with shield. Insert balance arm into the
hole on base. Do not press balance beam with force. Then tighten fastening screw on
balance base(Refer to figure 3-31).
To view the balance arm to ensure it is in the center of hole (if touch the furnace) firstly,
and if balance arm is not in the center, the balance beam must be adjusted to be in the
center and not to touch the furnace by adjusting the two front grounding bolts.
Then run the analysis program, and open “Hardware Debugging” under “Function”, and
turn on “Carousel Reset” , then put crucible in the Hole 0 of the carousel, then click on
“Carousel Down” to view if the crucibles have been propped up completely (the crucible
can not touch the carousel) and if the balance readings is stable, if not please adjust the
height and central position of the balance beam. The height of the balance arm can be
adjusted by changing the balance height or the depth of balance beam inserted into the
balance pedestal. If the balance arm is not upright to the carousel, it also can be
adjusted by turning the balance pedestal.
40
Figure 3-33
Click the "OK" button to save the parameter to the DSP motherboard.
41
Figure 3-34
Select 5E-MAC6710 and click OK.
Select Function-Hardware Debugging in the menu bar to pop up the dialog box as
the figure below:
42
Figure 3-35
D. Carousel
Set up the debug interface, turn "1 hole" at the bottom, click "start to rotate" button,
the turntable should rise to high position, then turn 1 hole, and stop the movement.
At this point the turntable stays at the high position.
Click Down , the carousel will drop to the low position and stop moving.
Click UP, the carousel will rise to the high position and stop moving.
If the carousel continuously makes ascending and descending movements, the
upper optical slot or the lower optical slot beside the carousel lifting mechanism can
be detected to be effectively shaded by the shutter plate moving up and down with
the carousel.
If the high level optical slot is not shaded, the fixed high or low positions of optical
slots can be adjusted to the shaded position.
43
If the low level optical slot is not shaded, the overall height of the bracket can rise to
the position that can be shaded.
If there is no guarantee that the upper optical slot can be shaded and the lower
optical slot can be shaded when descending, the relative position of the upper and
lower slots on the bracket shall be shortened.When the carousel rises, the shade
stops at the upper optical slot, and when the carousel drops, the shade stops at the
lower optical slot.
Note: If you do not rotate a hole first, but directly push the Down button, the carousel
has no drop action, it may be the reason that the positioning optical slot is
blocked,or carousel is not aligned with crucible hole, or hardware protection
prohibits the carousel descending.
a) Power off the instrument, disconnect the Addin Pair of stepper motor, and
power on the instrument. Rotate the dial by hand, and turn the dial through the
groove of the optical slot. Keep it in the middle position. If not, adjust the counting
slot holder. Similarly, the single needle reset indicator on the dial shall be kept in the
middle position through the groove position of the reset optical slot. If not, adjust the
reset optical slot, and the fixed seat or the reset pointer.
Figure 3-36
b) The counter slot is on and the reset slot is off. The reset optical slot is shaded
by the reset pointer, the counter is full of the grooves. In this case, the counter slot is
effective, the reset slot is valid.
44
Figure 3-37
c) Both the counter slot and reset slot are off.The reset optical slot is shaded by
the reset pointer, while the counter slot is shaded by the dial. The current state is In
this case, the counter slot is invalid, and the reset slot is valid.
Figure 3-38
d) The counter slot is off, and the reset slot is on; The reset slot is not shaded by
reset pointer, while the counter slot is shaded by dial. In this case, both the counter
slot and reset slot are invalid.
45
Figure 3-39
e) Both the counter slot and reset slot are on; the reset slot is not shaded by the
reset pointer, and the counter slot gets through the dial grooves; In this case, the
counter slot is valid, while the reset slot is invalid.
Figure 3-40
Note:The most important thing to emphasize in the above procedure is that when
reset slot is shaded by the reset pointer, the process of counter slot is not allowed
to be destroyed. If this occurs, the relative position between the reset slot and the
counter slot shall be adjusted until the requirements are met.
46
F. Carousel Reset Debugging
Power on the instrument, click rotate reset button, the carousel should rise to the upper
position and start rotation until it reaches the reset position. At this moment , the carousel
stays at the upper position.
Check whether the rod passes through the center of the hole at the bottom of the
furnace. If the rod is deflected, adjust the horizontal adjustment screw of the balance
base or the balance itself, and make sure that the sample rod is vertical through the
center of the hole at the bottom of the furnace.
Click the "Reset" button, after the carousel reset is completed, press "Down"". When the
carousel drops to a low position, it stops motion. Check the distance between the top
crucible and the crucible hole to make sure there is no contact and sufficient safety
redundancy. Turn the 1 crucible in turn and repeat the above check. In case of contact
danger, adjust the balance level adjustment screw to ensure that the top of the crucible
is not contacted with the hole wall of the crucible.
Click "Start the instrument will automatically complete the following actions:
A) automatic reset.
B) automatic weighing 1 crucible one time, and read the weight of the crucible shown
in the blank column on the right.
Observe the weighing process, check the weight of the crucible weight data, if any,
re-check the debugging processes of carousel lift, carousel reset and turn, optical slot
and optical slot reset , weighing rod and balance components until the analog weighing
is normal.
47
I. Fan, solenoid valve debugging
Figure 3-41
Normally, the compressed air supplied with the pump is fed from the oxygen inlet and
provides the function of blast during the moisture test, providing sufficient air combustion
during the ash test. When using the air pump to provide air supply, the manual two-way
48
valve 3 open, manual two-way valve 5 off, the pump out of the compressed air is not
through the instrument panel flow meter directly into the bottom of the furnace bore
circular tube. Click the "Oxygen valve open" button, the pump starts to work. According
to the flow meter on the pump, adjust the ventilation volume of the largest 10 ~ 15L / min,
then the instrument panel flow meter without flow display.
If the customer requests nitrogen and oxygen to do the experiment, the external supply
of gas were received to the nitrogen and oxygen inlet, manual two-way valve 3 off,
manual two-way valve 5 open, nitrogen and oxygen are required through the instrument
panel flow meter , Into the circular vent tube. The pressure gauge on the external
cylinder should be adjusted to 0.2MP. Click the "Oxygen Valve Open" button to adjust
the throttle valve 7, the flowmeter on the panel shows the flow rate of 4L / min, click the
"oxygen valve off" button. Click the "nitrogen valve open" button, adjust the throttle valve
7, the panel flow meter shows the flow rate of 4L / min, click the "nitrogen valve off"
button. Note that oxygen and nitrogen can not be opened at the same time.
Figure 3-42
49
As shown above, the ventilation throttle can control the opening speed of the furnace
cover and adjust the exhaust throttle to control the closing speed of the furnace lid.
Figure 3-43
50
Turn on the furnace and the red indicator of the two solid state relays on the base plate is
bright. Turn off the furnace and the red indicator of 2 solid state relay will be off. In this
case, the heating circuit works normally. If abnormal, check if the connections of solid
state relays, solid state relay, DSP board, control panel and switch power supply that
provides 4-channel output are normal.
Connect the power supply to the main engine with 220V/25A power, turn off the circuit
breaker behind the closed instrument and close the lid.
Turn on the furnace again, the 2 heating indicator on the instrument panel should be
bright. Turn off the furnace and the two lights on the panel are out. If abnormal, check the
heating power relay 1 and the heating power relay 2, the proximity switch of the door, the
metal indicator panel,and the indicator.
51
4. Operation
4.1. Test operation
1) Click 【Start Test】 menu to enter the test data input interface, as shown in Figure
4-1. The interface dialog box includes "data input" and "position input"; "data input "Box
includes " sample number "," sample name "," laboratory "and other input box.
Figure 4-1
2) "Position input"
Click the green circle on the circular dial under "Position input", and a red dot appears in
the center of the small circle to indicate that position is selected. Click the green circle
again and the red dot disappears to indicate that the location is unchecked. Click the
"Select All" button, all the center of the circle will appear a red dot, that select all the
location, this button to switch to "undo all selected", click "Undo Select" button, all red
dot disappears, that canceled All selected location, the button to switch to "all selected".
Green small circle on behalf of the instrument turntable cruise, No. 0 hole to correct the
crucible position, you must put the calibration crucible, can not be placed in the sample,
52
otherwise the measurement results.
3) "Data Input"
Under "Data Input", enter the corresponding "Sample No.", "Sample Name",
"Laboratory". (If you select "Auto Number" in [System Settings], you only need to enter
the corresponding "Sample Name" / "Laboratory"; enter the "Test Name" / "Laboratory"
drop-down box Name "/" laboratory technician "system automatically save. Enter all the
sample information, click 【Next】 button, enter the test work interface, as shown in
Figure 4-36.
4) "Query Modification"
If you enter the wrong sample information, you can "query changes" to find the wrong
location, you can correct. (Must be modified before clicking the [Next] button). Of course,
you can also modify the test interface number, sample name,operators, CO2%, Mad% (if
this experiment does not include moisture analysis).
5) "Sample name" and "laboratory technician" drop-down box maintenance
Time, the "sample name" / "laboratory" drop-down box there are many no longer needed
data, then you can clear the drop-down box information, click the "management
drop-down box" below the [sample name] and [ Laboratory technician], according to the
prompts to complete the operation.
Figure 4-2
6) Analysis Interface is the main interface of this program. In top half of interface the
53
analyzing or weighing crucible number (expressed in red dot) is in the left upper side.
The Temp.-Mass curve is in the bottom half part. In the bottom half of the interface, data
form and real time information, total sample number, error hint (if there is error), current
actual temp. of stove, balance reading and analysis time will be displayed.
Click [Weigh Empty Crucible with Lid for X Analysis] to weigh the empty crucible
with lid. But before weighing, the empty crucible must be put in order from No. 0 to
the last one for analysis with lids. If clicked the Weigh Empty Crucible with X button,
and it has been under weighing, but if you don’t want to start weighing the empty
crucible X with lids again, at that time there are two selections available: One is to
select “No” when the system hints put sample or not after weighing empty crucible X
with lid. The other one is to press “Stop X Analysis” during weighing empty crucible
X with lid, and the system will hint if stop the current analysis or not, select “No” ,
there will be a hint “restart weighing it or not”, then press “Yes”, it will be OK. If the
following items are required to be analyzed, input corresponding value before
clicking the [Weigh Empty Crucible with Lid for X Analysis] button.
If the ash residue is to be analyzed, the CO2 content in CO2% will be entered before
clicking on “Weigh Empty Crucible with Lid”, and if moisture analysis is included
(only ash), the corresponding moisture content must be entered in Mar%. And the
system will prompt you to confirm.
If it is the first time to test the items above, for the convenience of inputting
corresponding value, adjust the title name, sequence. and width.
After having automatically weighed vacant crucibles, the program will prompt you to
put samples in (prompt music will be heard). Click the [OK] button of the pop-up
54
dialog box. When the Carousel resets and stops rotating (the Carousel stops at the
location of #0 crucible), press the red “Rotate” button on the front panel of the
analyzer to rotate the Carousel and put samples into crucibles in turn. After that,
click the [Weigh Moisture and Ash Samples] button. Then, the program
automatically weighs samples
(Note: #0 crucible is used for correction only. No sample can be put into it! ).
If you have pressed down the [Weigh Moisture and Ash Samples] button, the
system has begun to weigh moisture and ash samples, but needs to do it again. In
this case, you also have two choices: After the system has weighed moisture and
ash samples, it will give a prompt of whether to begin analysis and you should select
[No]; at any time before the system has weighed moisture and ash samples, click
the [Stop Moisture and Ash Analysis] button. The system gives a prompt of whether
to terminate the current test and you should select [No]. The system gives another
prompt of whether to weigh moisture and ash samples again and you should select
[YES]. Do not turn on the fan before weighing moisture and ash samples lest the
weighing should be affected.
9) Start X Analysis
If click on “Stop X Analysis” button during analysis, the system will hint if pause the
analysis, if you click on “OK” button, the system w ill hint if exit, if click on “OK’
again, then it will exit from the analysis.
55
11) Prompts and Operation Skills during the Test
If the current test does not determine moisture or CO2, you must manually input
numeric values in the grids of Mad% and CO2%. That is, click the column where you
want to input data and do not input data until the vertical line cursor flashes. You
may move upward or downward by pressing ↑ or ↓ on the keyboard. At this time,
you may press Enter and input data in this column.
b) You may press and hold a heading (for example, date) in a form with the left key of
the mouse. When the red mark “▲” appears, drag the heading leftward or rightward
to the place you want and then loosen the left key. The heading automatically
moves to the place and rearrangement is made possible. If there are 3 dots behind
some data, it indicates that this column is not wide enough and you only need to put
the mouse between the heading of this column and that of next column. When “ ”
appears, press and hold the left key of the mouse and do not loosen it until there is
a proper width. In this way, all content will display.
c) During a test, you may set the number of columns displayed in a form. That is, move
the cursor onto the form and right click to select a corresponding item to be
modified.
1) Modify the heading of a form: Many data items, such as No., sample name and
sample mass, are defined. Some users need to change them, for example, to
change a sample name into a coal sample code; some want to adjust order;
some need three items while others need four. For the sake of users, they may
select whatever field they need to display or modify a field name or data display
font. Operating procedures: adjust the order of fields: set the field to be
displayed. Note: One field is one column in a form. First, right click the
heading of a form (The heading is the first line of the form. It is invariably grey).
The interface will pop up. Click the [Set Column of Form] menu to enter the
56
interface (Figure 4-30).
The program defines altogether 39 fields (some irrelevant fields are not listed).
In Figure 4-31, the fields in the left white square will not displayed in a form
while those in the right will. If you want to add a field in the left square to the
right side, you may double click the field. Or you may select the field (turn into
blue) and then click the [Add->] button. In this way, you may add the field to the
end of the right square. Likewise, you may remove a field in the left square in
the same way. If you want to adjust the order of the fields in the right square,
you may select them (turn into blue) and click the [Up] or [Down] button. Then,
you may click the [OK] button to save your setting. When you enter next time,
you will see them in the set order. Needless to say, when you click the [OK]
button, the program will immediately update its display as adjusted. If you click
[Cancel], your modification will not be validated.
Figure 4-3
57
Figure 4-4
(2)Adjust the width of the field, modify the field name. Similarly, click Figure 4-3 in the
[set table header] menu, you can enter the interface as shown in Figure 4-5. After
entering this interface you can only see the fields that appear in the table. If you want to
modify the field name or width of a field or the position of the header in the column, you
can select it first (click the field with the left mouse button) Field, so that the right side of
the change in the label of the text box will appear in your selected field name, the field
name to modify the desired field name, you can also modify the width, modify the
position in the title, that is, alignment, modify
After all the titles and then click the "OK" button, it will save the changes just, and
immediately update the table shows the next time you enter the changes just as
effective.
Figure 4-5
(3)Modify the display font of the data in the table. Similarly, click Figure 4-37 in the [set
the font] menu, you can enter the interface as shown in Figure 4-6. Into this interface, it
should be noted that some Western fonts do not choose, because some Western fonts
do not support Chinese, will show garbled, and the color settings are invalid. Changes to
58
the font will affect all of the same table in the program, but the field name, field width and
order changes will only affect the current table, and will not affect the other parts of the
table. For example, in the results of the query to modify the font for the fifth, then all other
forms of data fonts are five, but if the results in the query only need to show the number
and Vad%, modified only the results of the query shows the two Field, and the number of
fields in the test form does not change. However, the tables in the general query and the
Figure 4-6
(4) Modify the title of the report for the system. (excluding the template) as shown in
figure 4-7.
Figure 4-7
The "company" dialog box may enter the device's name, and the input will automatically
be added to the header of the printed result. The "department" dialog box can enter the
department that uses the instrument.
59
4. Routine Maintenance
Routine maintenance has a very positive effect on the normal use of the equipment.
4.1. Crucible|
The crucible is made of ceramic material. It is advisable to remove the residue
when the furnace has been cooled to ambient temperature. Put the crucible into
the muffle furnace under the temperature of 850 ℃ for 15 minutes. Take out the
crucible from the furnace, clean it, and put it back to the carousel for next use.
4.2. Thermocouple
Thermocouple is a consumable with a service life of around 12 months. It must be
abnormal power failure occurs, please do not turn the carousel with your hands.
When it gets the power on, check the instrument, and be sure that the carousel is
check the value of gas cylinder. Generally speaking , 0.2MPa is normal. If the
60
5. Professional Maintenance and Troubleshooting
This chapter focuses on the analysis of some common fault phenomena, fault
Troubleshooting :
1) Check whether the CP - end of the stepper motor drive is strung into a 500 ~ 1K
ohm resistor.
2) Check the reset slot and counter slot. Refer to the chapter 3.13.4 Adjustment
of reset slot and counter slot.
Check the instrument according to procedures and meet the demand as mentioned
above, then the dislocation can be solved.
A. The lifting and rotating mechanism does not do the lifting movement:
Troubleshooting:
1) Replace the optical slot
2) Replace the lift motor
3) Fasten the screws
4) Adjust the lifting shaft
5) Replace the control panel
61
B. Lifting and rotating mechanism falls automatically during the test:
1) Check whether the lifting motor is a motor with a damper disc (the rear of the
motor has a metal disk as a damper disc ).
2) Check whether the lifting motor is damaged or not.
Troubleshooting:
The lifting motor must be replaced in all the above phenomena.
C. The lifting and rotating mechanism has abnormal sound in the lifting process:
1) Check whether the deformation of the shaft and plastic friction cause
abnormal sound.
2) Check whether the lifting shaft fastening screws are loose and produce
abnormal sound.
3) Check whether there is excessive noise in the lifting rail or the ball in the slider.
4) Check whether the noise is caused by lifting motor.
Troubleshooting:
1) Adjust the gap between the carousel shaft and the plastic or replace the
carousel shaft.
2) Fasten the screw
3) Add lubricating oil or replace the new guide block after cleaning.
4) Replace lifting motor
1) Check whether the optical slots in the disc position are damaged.
2) Check whether the optical slots in the disc position are loose.
3) Check whether the control panel is damaged or not.
Troubleshooting:
1) Replace the optical slot.
2) Connect all the wire
3) Replace the control panel.
62
5.3. Balance Failure
Troubleshooting:
1) Check whether the balance is properly installed. Refer to Chapter 3.7, "Balance
Installation".
2) Check whether the balance beam is properly installed. Refer to Chapter 3.13.8,
"Adjustment of Balance Beam".
3) Check whether the communication Settings are correct. Refer to Chapter 3.13.1,
"COMM Settings".
Troubleshooting:
1) Check the communication serial port between the PC and the DSP board.
2) Check whether the power supply on the DSP board is normal.
63
Figure 6-1
Check if the thermocouple wiring is correct. The red line is connected to the positive pole
of the thermocouple, and the black wire is connected to the negative pole of the
thermocouple.
When the over-temperature alarms, the temperature does not reach the alarm
temperature yet.
64
Figure 6-2
The TP1 voltage is detected by a multimeter, and the adjustable resistance R30 is
adjusted so that the voltage of the TP1 is stabilized at about 2.64V. This can set high
temperature furnace overheating alarm temperature.
The TP2 voltage is detected by a multimeter, and the adjustable resistance R32 is
adjusted so that the voltage of the TP2 is stabilized at about 0.51V. Thus, the
temperature of the constant temperature furnace and the temperature of the alarm can
be set.
Troubleshooting:
1) check whether the main power fuse is in good condition;
65
2) check the power output of the switch;
3) check the power connector;
4) check whether there is a line of short-term, such as balance, power cord is under the
balance of the stent, broken and so on;
5) check whether the heating circuit breaker is closed;
6) check whether the heating relay SFD-112DM is pulled in. As shown in the figure, the
DC control terminal has power on the 5V voltage, while the AC 220V between the 2 feet
is still 220V, the relay is bad, replaced.
Figure 6-3
Fault:
1) the lifting motor of the carousel has no instruction to move up and down, or do not
move with the instruction;
2) the sample rod motor has no instruction to move up and down, or do not move with
the instruction;
3) the insulation board motor has no instruction to open and close,or do not move with
the instruction;
4) solenoid valve has no instruction to pull off, or do not move with the instruction;
Troubleshooting:
66
2) replace the control panel;
67
Furnace Lid
Gas
Gas Cylinder
Cylinder
68
Inlet
Y-Tee Gas Inlet Throttle Valve Compressed Gas
P
A
R
附录 A 气路图
Gas Outlet
Throttle Valve
Outlet
汽缸气路图
通气气路图
69
Appendix B Wiring Diagram
Wiring Diagram 1
70
71
Wiring Diagram 2
72
73