DOC0235290643 3ed
DOC0235290643 3ed
90643
01/2024, Edition 3
Table of Contents
1
Table of Contents
2
Section 1 Maintenance
DANGER
Multiple hazards. Only qualified personnel must conduct the tasks described in this
section of the document.
Hazardous voltage. This symbol indicates that hazardous voltages are present where a risk of electrical shock
exists.
Hot surface. This symbol indicates that the marked item can be hot and should not be touched without care.
Corrosive substance. This symbol identifies the presence of a strong corrosive or other hazardous substance and a
risk of chemical harm. Only individuals qualified and trained to work with chemicals should handle chemicals or
perform maintenance on chemical delivery systems associated with the equipment.
Toxic. This symbol indicates a toxic/poisonous substance hazard.
This symbol indicates the presence of devices sensitive to Electro-static Discharge (ESD) and indicates that care
must be taken to prevent damage with the equipment.
Protective earth. This symbol indicates a terminal which is intended for connection to an external conductor for
protection against electric shock in case of a fault (or the terminal of a protective earth (ground) electrode).
Noiseless (clean) earth. This symbol indicates a functional earthing (grounding) terminal (e.g., a specially designed
earthing (grounding) system) to avoid a malfunction of the equipment.
This symbol indicates there is a lifting hazard because the object is heavy.
3
Maintenance
Electrical equipment marked with this symbol may not be disposed of in European domestic or public disposal
systems. Return old or end-of-life equipment to the manufacturer for disposal at no charge to the user.
DANGER
Indicates a potentially or imminently hazardous situation which, if not avoided, will result in death
or serious injury.
WARNING
Indicates an instruction for a potentially hazardous situation that may result in death or a serious
injury.
CAUTION
Indicates that a precaution must be followed for a potentially hazardous situation that may result in
minor or moderate injury.
NOTICE
Indicates a situation which, if not avoided, may cause damage to the instrument. Information that
requires special emphasis.
CAUTION
Ozone inhalation hazard. This instrument produces ozone that is contained within the
equipment, specifically within the internal plumbing. The ozone could be released under
fault conditions.
It is recommended to plumb the exhaust gas port to a fume hood or to the building
exterior in accordance with local, regional and national requirements.
Exposure to even low concentrations of ozone can damage delicate nasal, bronchial and
pulmonary membrane. In sufficient concentration, ozone can cause headaches,
coughing, eye, nose and throat irritation. Immediately, move the victim to uncontaminated
air and seek first aid.
The type and harshness of symptoms are based on the concentration and exposure time
(n). Ozone poisoning includes one or more of the symptoms that follow.
• Irritation or burning of the eyes, nose or throat
• Lassitude
• Frontal headache
• Sensation of sub-sternal pressure
• Constriction or oppression
4
Maintenance
Select MAINTENANCE > DIAGNOSTICS > SIMULATE. Select MFC. Set the flow to 20 L/h. Push to start
the mass flow controller (MFC). The measured flow shows on the display.
Make sure that the oxygen pressure regulator shows 400 mbar at 20 L/h. Refer to Analysis enclosure
on page 31 for the location.
Make sure that the reagent levels are sufficient. Fill or replace reagent containers as necessary. Refer to Fill
or replace the reagents on page 6.
1 Refer to the documentation supplied with the maintenance kit for instructions.
5
Maintenance
Task Initial
Make sure that there are no leaks at the reagent pumps. Refer to Analysis enclosure on page 31 for the
location.
Make sure that there are no leaks at the sample pump.
Make sure that the oxidized sample catch-pot has no leaks.
Make sure that there are no leaks at the valves in the analyzer. Refer to Analysis enclosure on page 31 for
the location.
Make sure that there are no blockages in the sample lines to the analyzer or the sample lines in the analyzer.
Make sure that there are no blockages in the drain lines from the analyzer or the drain lines in the analyzer.
Make sure that there is sufficient sample flow to the oxidized sample catch-pot or the sample tubing for a fresh
sample for each analysis cycle.
Make sure that there is no blockage in the exhaust tubing.
Make sure that there are no blockages in the filter in the fan housing and the vent housing on the side of the
analyzer.
If a sampler is used, make sure that the operation of the sampler is correct. Make sure that there is sufficient
flow to the sample pipe.
CAUTION
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the
personal protective equipment appropriate to the chemicals that are handled. Refer to
the current safety data sheets (MSDS/SDS) for safety protocols.
CAUTION
Chemical exposure hazard. Dispose of chemicals and wastes in accordance with local,
regional and national regulations.
Fill or replace the acid and base reagent containers as necessary when the analyzer is
stopped.
1. Select OPERATION > START,STOP > FINISH & STOP or EMERGENCY STOP.
2. Fill or replace the reagents.
3. Select MAINTENANCE > COMMISSIONING > REAGENTS MONITOR.
4. Set the reagents volumes.
5. Select OPERATION > REAGENTS SETUP> INSTALL NEW REAGENTS to prime
the reagent tubing and do a zero calibration.
DANGER
Electrocution hazard. Isolate all power to the instrument and disconnect all power from
the instrument and relay connections before this maintenance task is started
6
Maintenance
DANGER
Electrocution hazard. Use the same type and current rating to replace fuses.
Replace a blown fuse for correct operation. Refer to Figure 1 for the fuse locations. Refer
to Table 2 for the fuse specifications.
In addition, a diagram of the fuse locations is available on the top door.
7
Maintenance
Key:
A—Amperes
F—Fuse
H—High interrupt
ID—Identification
L— Low interrupt
mA—Milliamperes
PCB—Printed circuit board
T—Time lag (time delay)
V—Volts
8
Maintenance
Task Initial
Do the steps that follow:
1. Select MAINTENANCE > DIAGNOSTICS > SIMULATE > CLEANING VALVE. Select ON to open the
cleaning valve.
2. Make sure that all of the stream, manual and calibration valves are closed.
3. Select SAMPLE PUMP, then select REV to set the pump to operate in reverse. Operate the sample pump
in reverse until the sample lines and the oxidized sample catch-pot are empty.
CAUTION
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the
personal protective equipment appropriate to the chemicals that are handled. Refer to
the current safety data sheets (MSDS/SDS) for safety protocols.
CAUTION
Chemical exposure hazard. Dispose of chemicals and wastes in accordance with local,
regional and national regulations.
1. Put on the personal protective equipment identified in the safety data sheets
(MSDS/SDS).
2. Remove the tubes from the ACID and BASE ports on the side of the analyzer.
3. Plumb the ACID and BASE ports to a deionized water container. If deionized water is
not available, use tap water.
4. Select CALIBRATION > ZERO CALIBRATION > RUN REAGENTS PURGE to start a
purge cycle.
5. Do step 4 a second time.
The analyzer replaces the reagents in the reagents lines with water.
6. When the reagent purge cycle is complete, remove the tubing from the deionized
water container and put them in open air.
7. Do step 4 two times.
The analyzer replaces the water in the reagents lines with air.
9
Maintenance
10
Section 2 Troubleshooting
02_LOW O2 FLOW - The oxygen flow through the sample • Oxygen supply problem. The oxygen pressure
SO out (SO) valve (MV5) was less than should be 400 mbar (±10 mbar) at 20 L/h MFC flow.
50% of the MFC setpoint for more than Select MAINTENANCE > DIAGNOSTICS > O2-
the LOW O2 FLOW TIME setting. CTRL STATUS.
Refer to MAINTENANCE > SYSTEM • Failure of or blockage in the sample out valve
CONFIGURATION > FAULT SETUP > • Failure of or blockage in the exhaust valve (MV1)
LOW O2 FLOW TIME.
• Failure of the MFC. Do a flow test. Refer to Do a flow
test on page 21.
03_HIGH O2 FLOW The oxygen flow through the exhaust • Failure of the MFC
valve (MV1) was more than 50% of the
• Oxygen supply problem. The oxygen pressure
MFC setpoint for more than the HIGH
should be 400 mbar (±10 mbar) at 20 L/h MFC flow.
O2 FLOW TIME setting.
Select MAINTENANCE > DIAGNOSTICS > O2-
Refer to MAINTENANCE > SYSTEM CTRL STATUS.
CONFIGURATION > FAULT SETUP >
HIGH O2 FLOW TIME.
04_NO REACTION No TOC (or TC) CO2 peak or the CO2 • The acid reagent and/or base reagent are the
(can be set as a fault or peak is less than the CO2 LEVEL incorrect concentration.
warning) setting for three consecutive reactions.
• The acid reagent container and/or base reagent
Refer to MAINTENANCE > SYSTEM container are empty.
CONFIGURATION > REACTION • The acid and/or base reagent lines have a blockage
CHECK > CO2 LEVEL. or air bubbles.
• The operation of the acid pump and/or base pump is
incorrect.
• The operation of the mixer reactor is incorrect. Do a
pH test. Refer to Do a pH test on page 23.
11
Troubleshooting
09_OZONE PCB The operation of the ozone board is Replace the ozone board. Contact technical support.
FAULT incorrect.
11_CO2 ANALYZER The operation of the CO2 analyzer is Examine the 24V DC input power to the CO2 analyzer
FAULT incorrect. from the motherboard wires 101 and 102). Refer to
Control enclosure components on page 35 for the
location.
Examine the signal from the CO2 analyzer. Open the CO2
analyzer and clean the lenses.
Remove and then apply power to the analyzer.
For more tests, refer to the information sheet T019.
BioTector CO2 Analyzer Troubleshooting.
12_HIGH CO2 IN O2 There is a high level of CO2 in the input Select MAINTENANCE > DIAGNOSTICS > SIMULATE.
oxygen gas. If the CO2 value on the display is more than 250 to
300 ppm, examine the oxygen purity.
Identify if there is CO2 contamination in the oxygen
supply. Refer to Examine the oxygen supply in the
Operation and Installation Manual.
If the oxygen purity is sufficient, open the CO2 analyzer
and clean the lenses. If the problem continues, replace
the CO2 analyzer filters.
If the oxygen purity is not satisfactory, replace the oxygen
concentrator.
13_SMPL VALVE SEN The sample valve sensors are in a Identify if faults 14_SAMPLE VALVE SEN1, 15_SAMPLE
SEQ wrong sequence. The sample valve VALVE SEN2 or 16_SAMPLE VALVE SEN3 have
sensors should be in the sequence occurred.
Sensor 1, 2 and 3. Examine fuse F6 on the Relay PCB. Select
MAINTENANCE > DIAGNOSTICS > SAMPLE VALVE.
Examine the operation of the sample valve. Examine the
sample valve sensor wiring.
12
Troubleshooting
13
Troubleshooting
23_FLOW WARNING – The oxygen flow through the sample out • Oxygen supply problem. The oxygen pressure
SO valve (MV5) decreased to less than the should be 400 mbar (±10 mbar) at 20 L/h MFC
FLOW WARNING setting during the flow. Select MAINTENANCE > DIAGNOSTICS >
pressure test. O2-CTRL STATUS.
Refer to MAINTENANCE > SYSTEM • Failure of or blockage in the sample out valve
CONFIGURATION > SEQUENCE • Blockage in the tube after MFC
PROGRAM > PRESSURE/FLOW TEST
• Failure of the MFC. Do a flow test. Refer to Do a
> FLOW WARNING.
flow test on page 21.
26_PRESSURE TEST The MFC flow did not decrease to less • The analyzer has a gas and/or liquid leak.
WARN than the PRESSURE TEST WARN
• A valve has a leak.
setting during the pressure test.
• Examine the sample out valve, sample (ARS)
Refer to MAINTENANCE > SYSTEM valve and analyzer fittings for leaks.
CONFIGURATION > SEQUENCE
• Examine the mixer reactor for leaks. Do a
PROGRAM > PRESSURE/FLOW TEST
pressure test. Refer to Do a pressure test
> PRESSURE TEST WARN.
on page 21.
28_NO PRESSURE The pressure test was not done during The analyzer was started with a quick startup. The
TEST the system startup sequence. RIGHT arrow key was pushed when START was
Note: The warning stays active until a selected.
pressure test is passed.
29_PRESSURE TEST The daily pressure test and flow test Set the pressure test and flow test functions to on in
OFF functions are set to off. the MAINTENANCE > SYSTEM CONFIGURATION >
SEQUENCE PROGRAM > PRESSURE/FLOW TEST
menu.
30_TOC SPAN CAL FAIL The result of the TIC or TOC span Make sure that the concentration of the prepared
31_TIC SPAN CAL FAIL calibration is not within the TIC BAND or standard solution is correct. Make sure that the
TOC BAND setting. settings in the CALIBRATION > SPAN
Refer to MAINTENANCE > SYSTEM CALIBRATION menu are correct. Examine the
CONFIGURATION > SEQUENCE operation of the analyzer.
PROGRAM > SPAN PROGRAM > TIC
BAND or TOC BAND.
33_TOC SPAN CHCK The result of the TIC or TOC span check
FAIL is not within the TIC BAND or TOC BAND
34_ TIC SPAN CHCK setting.
FAIL Refer to MAINTENANCE > SYSTEM
CONFIGURATION > SEQUENCE
PROGRAM > SPAN PROGRAM > TIC
BAND or TOC BAND.
14
Troubleshooting
53_TEMPERATURE The analyzer temperature is more than Identify the internal analyzer temperature. Examine
ALARM the TEMPERATURE ALARM setting. the filters in the fan and the vent. Examine the
Refer to MAINTENANCE > SYSTEM operation of the fan.
Note: At temperatures below 25 °C (77 °F), the analyzer
CONFIGURATION > FAULT SETUP >
sets the fan to off.
TEMPERATURE ALARM.
Note: The analyzer fan operates in backup
mode until the warning is acknowledged.
54_COOLER LOW TEMP The cooler temperature is less than 2 °C Look at the flashing LED 3 on the signal board to
for more than 600 seconds. examine the operation of the cooler. The operation of
the temperature sensor is incorrect. Replace the
cooler.
15
Troubleshooting
16
Troubleshooting
17
Troubleshooting
128_REACT PURGE The gas flow is not normal. There is an • Blockage in the mixer reactor, the sample out
WARN instrument air or oxygen supply problem. valve or the sample out tubing and fittings
• Blockage in the tube after MFC
• Failure of the MFC
18
Troubleshooting
2.3 Notifications
Select OPERATION > FAULT ARCHIVE to see the notifications. When "SYSTEM NOTE"
shows in the top-left corner of the Reaction Data screen or the Reagent Status screen, a
notification has occurred. Refer to Table 5.
Table 5 Notifications
Message Description Solution
85_LOW REAGENTS The calculated reagent levels identify Replace the reagents. Refer to Fill or replace the reagents
(can be set as a warning that the reagent containers are at a on page 6.
or note) low level. To increase the number of days before a LOW
REAGENTS notification occurs, select MAINTENANCE >
COMMISSIONING > REAGENTS MONITOR > LOW
REAGENTS AT.
86_POWER UP Power was supplied to the analyzer This notification is automatically acknowledged. No action
or a power reboot was done after the is necessary.
processor watchdog timeout.
87_SERVICE TIME The service counter has been set to This notification is automatically acknowledged. No action
RESET 200 days (default). is necessary.
RESET SERVICE COUNTER was
selected.
116_LOW/NO SAMPLE 1 The sample sensor does not sense Examine the sample liquid level and the sampling system
117_LOW/NO SAMPLE 2 sample or the sample quantity is less for each sample source.
118_LOW/NO SAMPLE 3 than the limit for the sample source Select MAINTENANCE > DIAGNOSTICS > SIMULATE >
(default: 75%). SAMPLE PUMP. Select PUMP FORWARD TEST.
Examine the sample delivery and the sample bypass
tubing.
Identify if there are air bubbles in the sample tubing.
122_SAMPLE FAULT 1 An external device sent a sample Examine the external sample liquid level and sampling
123_SAMPLE FAULT 2 fault input signal to the analyzer. system for the sample channel.
124_SAMPLE FAULT 3 Examine the external sample monitoring device and the
external input signal wiring.
19
Troubleshooting
20
Section 3 Diagnostics
1. Select MAINTENANCE > DIAGNOSTICS > PROCESS TEST > PRESSURE TEST.
2. Select PRESSURE TEST, then push .
A pressure test starts (60 seconds). The information that follows shows.
Item Description
TIME Shows the remaining time for the test.
MFC Shows the mass flow controller (MFC) setting for the test (default: 40 L/hr).
SETPOINT
MFC FLOW Shows the flow from the MFC. If there is no gas leak, the flow will slowly
decrease to near 0 L/hr after 25 seconds.
STATUS Shows the results of the test.
TESTING—Test in progress
PASS—The flow from the MFC at the end of the test is less than 4 L/hr
(default).
WARNING—The flow from the MFC at the end of the test is more than
4 L/hr but less than 6 L/hr (default).
FAIL—The flow from the MFC at the end of the test is more than 6 L/hr
(default).
Note: To change the default limits for the test, select MAINTENANCE >
SYSTEM CONFIGURATION > SEQUENCE PROGRAM >
PRESSURE/FLOW TEST.
3. If the pressure test fails, select PRESSURIZE REACTOR, then push to find the
location of a leak. A longer test starts (999 seconds).
1. Select MAINTENANCE > DIAGNOSTICS > PROCESS TEST > FLOW TEST.
2. Select EXHAUST TEST, then push .
A flow test starts (30 seconds). The information that follows shows.
Item Description
TIME Shows the remaining time for the test.
MFC Shows the mass flow controller (MFC) setting for the test (default: 80 L/hr).
SETPOINT
MFC FLOW Shows the flow from the MFC. If there is no blockage, the flow is
approximately 80 L/hr.
STATUS Shows the results of the test.
TESTING—Test in progress
PASS—The flow from the MFC at the end of the test is more than 72 L/hr
(default).
WARNING—The flow from the MFC at the end of the test is less than
72 L/hr but more than 40 L/hr (default).
FAIL—The flow from the MFC at the end of the test is less than 40 L/hr
(default).
Note: To change the default limits for the test, select MAINTENANCE >
SYSTEM CONFIGURATION > SEQUENCE PROGRAM >
PRESSURE/FLOW TEST.
21
Diagnostics
3. If the exhaust test fails, select EXHAUST FLOW, then push to find the location of
the blockage (e.g., at the exhaust valve). A longer test starts (999 seconds).
4. Select SAMPLE OUT TEST, then push .
A sample out test is started. The test identifies if there is a blockage sample out lines.
5. If the sample out test fails, select SAMPLE OUT FLOW, then push to find the
location of the blockage (e.g., at the sample out valve). A longer test starts
(999 seconds).
1. Install the ozone tester in the analyzer. Refer to information sheet T029. Procedure to
check the ozone level in a BioTector B3500 and B7000 using a universal ozone
tester..
2. Select MAINTENANCE > DIAGNOSTICS > PROCESS TEST > OZONE TEST.
3. Select START TEST.
The analyzer does a pressure test. Then the ozone generator is set to on. An ozone
warning message shows on the display.
4. When the O-ring in the tester breaks, select STOP TEST.
The analyzer removes all of the ozone from the ozone tester (30 seconds). The test
results show on the display.
Item Description
TIME Shows the time for the O-ring to break.
STATUS Shows the results of the test.
TESTING—Test in progress
PASS—The time to break the O-ring was less than 18 seconds (default).
LOW OZONE—The time to break the O-ring was more than 18 seconds but less
than 60 seconds (default).
FAIL—The time to break the O-ring was more than 60 seconds.
Note: To change the default limits for the test, select MAINTENANCE > SYSTEM
CONFIGURATION > FAULT SETUP > OZONE TEST TIME.
1. Select MAINTENANCE > DIAGNOSTICS > PROCESS TEST > SAMPLE PUMP
TEST.
2. Select an option.
Option Description
VALVE Sets the SAMPLE or MANUAL fitting used for the test. For example, to
select the SAMPLE 1 fitting, select STREAM VALVE 1.
22
Diagnostics
Option Description
PUMP Starts the sample pump in the forward direction.
FORWARD Note: First select PUMP REVERSE TEST to empty the sample lines, then
TEST select PUMP FORWARD TEST.
1. Push to stop the timer when the sample is through the sample
(ARS) valve and the sample drips into the drain pipe on the side of the
analyzer.
2. Record the time on the display. The time is the correct forward time for
the selected stream.
SAMPLE Goes to the MAINTENANCE > COMMISSIONING > SAMPLE PUMP menu
PUMP to set the forward and reverse times for each sample stream.
3.5 Do a pH test
CAUTION
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the
personal protective equipment appropriate to the chemicals that are handled. Refer to
the current safety data sheets (MSDS/SDS) for safety protocols.
CAUTION
Chemical exposure hazard. Dispose of chemicals and wastes in accordance with local,
regional and national regulations.
Do a pH test to identify if the pH of the solution in the reactor is correct at the different
steps of a reaction.
Items to collect:
• pH paper
• Glass beaker
• Personal protective equipment (refer to MSDS/SDS)
1. Put on the personal protective equipment identified in the safety data sheet
(MSDS/SDS).
2. Select MAINTENANCE > DIAGNOSTICS > PROCESS TEST > pH TEST.
3. Select RANGE,VALVE.
4. Set the operation range (e.g., 1) and the stream (e.g., STREAM 1) to use for the test.
Refer to the OPERATION > SYSTEM RANGE DATA screen to see the operation
ranges. Select the operation range that agrees with normal measurements for the
sample stream.
5. Select MODE.
6. Select the test mode (e.g., TIC+TOC or TC).
7. Select START TEST.
8. Push again to confirm that the previous reaction completed normally.
23
Diagnostics
1. Select MAINTENANCE > DIAGNOSTICS > PROCESS TEST > SAMPLE VALVE
TEST.
2. Select TEST FIRST FAILURE to start the test.
The analyzer rotates the sample valve from sensor position 1, 2, then 3.
"COMPLETE" shows when the test is completed.
• LOOP COUNT —Shows the number of loops the sample ball valve is rotated for
each adjustment point for each sensor position during the test (default: 2).
• CURRENTLY TESTING —Shows the adjustment points (the time delay
implemented by the software) for each sensor during the test. The adjustment
points are from 0 to 15 with 1 point increments.
• FIRST FAILURE POINT —Shows the adjustment point at which the analyzer
does not detect the position of the valve.
3. Select ADJUST SAMPLE VALVE to set the sample valve stop position to align the
sample ball valve with the sample valve ports. Follow the instructions on the display.
24
Diagnostics
The analyzer shows the position of the valve (e.g., SENS 1) when the adjustment
values are entered.
If a 17_SMPL VALVE NOT SYNC fault occurs, refer to the information sheets T018.
BioTector Sample Valve Not Synchronized Fault Troubleshooting after Valve
Replacement and TT002. BioTector Sample Valve Not Sync Fault Quick
Troubleshooting.
Note: When the sample valve is replaced, refer to the information sheet M046. Sample Valve
Adjustment and Sample Tube Positioning Guidelines.
3.7 Do simulations
Do simulations to identify if a component (e.g., pumps, valves and mass flow controller)
operation is correct.
Note: Each time a component is activated, the analyzer will stop the operation of other devices as
necessary to prevent damage to the analyzer.
When the back key is pushed to exit the menu, the analyzer does a pump
synchronization process.
Option Description
MFC Sets the mass flow controller (MFC) flow (e.g., 40 L/h). Set the flow.
Push to start the mass flow controller (MFC). The measured flow
shows at the top of the display.
Note: If the flow shown is 0.0 L/h, the MFC is off.
ACID PUMP Sets the acid pump to on or off. Sets the number of pulses (½
revolution).
When the pump is in operation, the actual pulse time (external
brackets) and the set pulse time (internal brackets) show.
ACID VALVE Sets the acid valve to on or off.
BASE PUMP Sets the base pump to on or off. Sets the number of pulses (½
revolution).
When the pump is in operation, the actual pulse time (external
brackets) and the set pulse time (internal brackets) show.
PH ADJUST VALVE Sets the pH adjust valve to on or off.
SAMPLE VALVE Sets the sample (ARS) valve to the selected position. Options:
SEN1 (sample pump to bypass), SEN2 (sample pump to reactor) or
SEN3 (acid or base to reactor).
25
Diagnostics
Option Description
SAMPLE PUMP Sets the sample pump to the selected operation mode. Options: FWD
(forward), REV (reverse), P-FWD (pulse control forward) or P-REV
(pulse control reverse).
If P-FWD or P-REV is selected, set the number of pulses (½
revolution of the pump roller).
When the pump is in operation, the actual pulse time (external
brackets) and the set pulse time (internal brackets) show.
INJECTION VALVE Sets the injection valve to on or off.
REACTOR MOTOR Sets the mixer reactor motor to on or off.
SAMPLE OUT Sets the sample out valve to on or off.
VALVE
EXHAUST VALVE Sets the exhaust valve to on or off.
CLEANING VALVE Sets the cleaning valve to on or off.
CALIBRATION Sets the zero or span calibration valve to on or off. Options: ZERO,
VALVE (optional) SPAN or OFF.
STREAM VALVE Sets a sample stream valve to on or off. Select the number of the
stream valve. Only one stream valve can be set to on at one time.
Note: The stream valves can be controlled from the programmable
relays or from the stream expansion (auxiliary) board.
MANUAL VALVE Sets a manual valve to on or off. Select the manual valve. Only one
manual valve can be set to on at one time.
COOLER Sets the cooler to on, off or automatic to identify if the cooler relay
operation is correct.
LEAK DETECTOR The LEAK DETECTOR option cannot be selected. The condition of
the Liquid Leak Detector alarm input shows on the display.
FAN Sets the fan to on, off or automatic to identify if the fan relay operation
is correct. The analyzer temperature shows on the display.
When FAN is set to AUTO, the analyzer sets the fan to off when the
analyzer temperature is less than 25 °C. The fan operates
continuously when the analyzer temperature is more than 25 °C.
TEMP SWITCH Sets the temperature switch to on, off or automatic to identify if the
temperature switch operation is correct.
When TEMP SWITCH is set to AUTO, the analyzer sets the
temperature switch to on when the analyzer temperature is 25 °C
(default) or higher. The temperature switch stays on until the analyzer
temperature is less than 25 °C.
SAMPLER FILL Sets the signal to fill the sampler to on or off. The signal stays on until
set to off.
SAMPLER EMPTY Sets the signal to make the sampler empty to on or off. The signal
stays on for 5 seconds.
SAMPLER ERROR Sets the signal for a sampler error to on or off. The sampler error
signal is normally sent from the sampler when there is an error in the
sampler.
SAMPLE SENSOR The SAMPLE SENSOR option cannot be selected. The condition of
the sample sensor shows on the display.
REACTOR PURGE Starts the reactor purge operation.
26
Diagnostics
Option Description
RUN REAGENTS Starts the reagent prime operation, which fills the reagent tubing with
PURGE reagent.
INPUT/OUTPUT Goes to the MAINTENANCE > DIAGNOSTICS > INPUT/OUTPUT
STATUS STATUS menu. The INPUT/OUTPUT STATUS menu shows the
conditions of the digital inputs, digital outputs, analog inputs and
analog outputs.
27
Diagnostics
DIGITAL Shows the digital signal at the digital outputs (1 = active, 0 = not active). "DO"
OUTPUT followed by two digits identifies the digital outputs. For example, DO21 is Digital
Output 21.
The digital output number is followed by the digital signal at the output and then
the function. "[PROGRAMMABLE]" identifies the configurable digital outputs.
Note: When the analyzer is set to on, all of the digital outputs are set to 0.
Note: DO21 has a digital signal of 1 when the cooler is on and 0 when the
cooler is off. The cooler operates for approximately 3 seconds and then is off
for 7 seconds.
ANALOG Shows the ADC converter digital value, input voltage and function of each
INPUT analog input. The analyzer uses a 12-bit ADC, so the range of the digital value
is 0 to 4095. The input voltage range is 0 to 5.00 V.
ANALOG Shows the DAC converter digital value, output voltage and function of each
OUTPUT analog output. The analyzer uses a 12-bit DAC, so the range of the digital value
is 0 to 4095. The output voltage range is 0 to 10.00 V.
28
Diagnostics
• VALVE1, 2, 3—The oxygen controller valve outputs for Valves 1, 2 and 3 (1 = on,
0 = off). Valve 1 is the air isolation valve. Valve 2 and 3 are reserved.
• ROTARY VALVE —The operation of the rotary valve (FORWARD, REVERSE or
STOP).
Note: Approximately September 2022, the rotary valve was removed from the analyzer.
• ROTARY VALVE SENSOR —The sensor position of the rotary valve (1 = rotary
valve is on the sensor, 0 = the valve is not on the sensor).
Note: Approximately September 2022, the rotary valve sensor was removed from the
analyzer.
• MFC SETPOINT —Sets the setpoint for the mass flow controller. Push enter
(checkmark icon), select the setpoint, then push enter again. The MFC flow
shows at the top of the display. The MFC is off when the flow is 0.0 L/h.
• MFC FLOW —The first value is the MFC flow. The second value is the voltage
reading from the MFC. When the analyzer is stopped or in remote standby, the
MFC setpoint is 1 L/h.
BUS COM ERROR Shows the number of corrupted or not fully received Modbus
COUNT messages that the Modbus received.
Note: When the count is 65,535, the subsequent message
received sets the count to 1.
MANUFACTURE ID Shows the manufacturer ID for the instrument (e.g., 1 for Hach).
DEVICE ID Shows the class or family of the instrument, if entered (default:
1234).
SERIAL NUMBER Shows the serial number of the instrument.
LOCATION TAG Shows the location of the instrument.
FIRMWARE REV Shows the firmware revision installed on the instrument.
REGISTERS MAP REV Shows the Modbus register map version used by the
instrument. Refer to the Modbus register maps in the Advanced
Configuration Manual.
After the menu options, the first 17 bytes of the last received (RX) and transmitted (TX)
Modbus message show.
29
Diagnostics
3. Select MAINTENANCE > DIAGNOSTICS > MODBUS STATUS > BUS COM ERROR
COUNT. Look at the bus transmission error count.
The bus error count should increase each time the analyzer reads an invalid or not
fully received Modbus message.
Note: Valid messages that are not addressed to the instrument do not increase the counter.
4. For the Modbus RTU option, make sure that the wire connected to terminal D+ is
positively biased compared to the wire connected to terminal D– when the bus is in
an idle condition.
5. Make sure that there is a jumper installed on J15 of the motherboard at the end of the
bus to terminate the bus. The motherboard is in the electronic enclosure on the door
behind the stainless steel cover.
6. For the Modbus TCP option, open the web interface. Refer to Configure the Modbus
TCP/IP module in the Installation and Operation Manual. If the web interface does not
open, do the steps that follow:
a. Make sure that the network settings are correct.
b. Make sure that the Ethernet cable connectors are fully installed in the Ethernet
ports.
c. Make sure that the LED for the Modbus TCP/IP (RJ45) connector is green.
30
Section 4 Analysis enclosure
Approximately September 2022, the oxygen concentrator parts changed.
Figure 2 shows the pumps and components in the analysis enclosure after the change.
Figure 3 shows the valves in the analysis enclosure after the change.
Figure 4 shows the pumps and components in the analysis enclosure before the change.
Figure 5 shows the valves in the analysis enclosure before the change.
Figure 2 Analysis enclosure—Pumps and components
31
Analysis enclosure
32
Analysis enclosure
33
Analysis enclosure
34
Section 5 Control enclosure components
Figure 6 Control enclosure components
35
Control enclosure components
36
Section 6 Replacement parts and accessories
WARNING
Personal injury hazard. Use of non-approved parts may cause personal injury, damage
to the instrument or equipment malfunction. The replacement parts in this section are
approved by the manufacturer.
Note: Product and Article numbers may vary for some selling regions. Contact the appropriate
distributor or refer to the company website for contact information.
Consumables
20 L
Acid reagent, 1.8 N Sulfuric Acid containing 80mg/L Manganese Sulfate Monohydrate 25255061
(5.2 gallons)
20 L
Base reagent, 1.2 N Sodium Hydroxide 2985562
(5.2 gallons)
Deionized water 4 L (1 gallon) 27256
TOC standard, 50.0 mg/L 4L 5847200
TOC standard, 100 mg/L 1L LCW843
TOC standard, 200 mg/L 1L LCW845
TOC standard, 250 mg/L 1L LCW848
TOC standard, 500 mg/L 1L LCW846
TOC standard, 500 mg/L 4L 5847300
TOC standard, 1000 mg/L 4L 5846900
TOC standard, 5000 mg/L 4L 5847400
37
Replacement parts and accessories
Replacement parts
Quantity to
Description2 Item no.
stock
6-month maintenance kit, B7000i TOC analyzer 1 19-KIT-131
Air isolation valve, N/C 0 19-B5C-012
Acid pump or base pump, SR25 0 19-ASF-004
ARM main board, Rev 9, includes:
0 19-PCB-053
processor and LCD
CO2 analyzer, Hastelloy, 0–15000 ppm 0 20-CO2-011
Cooler , B4M with glass bead filter 0 19-BAS-018
Instrument air filter pack, B5C 0 10-SMC-001
Filter pack elements for air supply, B5C 13 12-SMC-001
Isolation amplifier 1 10-KNK-001
Mass flow controller (MFC) 0 12-PCP-001
Mixer reactor motor, B4M, 24 VDC, complete with leak detection 1 19-BAS-015
Mixer reactor, B4M, PTFE, complete with 24 VDC motor 0 19-BAS-016
Mixer reactor, B4M, PTFE 0 19-BAS-017
Ozone destructor heater 0 10-HAW-001
Ozone generator module B7000i 0 20-OZN-003
Ozone tube subassembly (with PCBA) 0 20-OZN-002
Ozone tube subassembly (without PCBA) 0 20-OZN-006
Ozone PCBA 0 ZBA81204326
Oxidized sample catch-pot (OSCP), cleaning vessel, glassware, 50 mm 1 10-KBS-018
Oxygen control board, complete 0 20-PCB-136
Power board, 115 VAC analyzer, B7000 1 19-PCB-160
Power board, 230 VAC analyzer, B7000 1 19-PCB-250
PTFE diaphragm for mixer reactor 1 10-KNF-038
PTFE ferrule and PEEK locking ring set, 1 x 3/16-in. 5 10-EMT-136
PTFE ferrule and PEEK locking ring set, 1 x 1/4-in. 5 10-EMT-114
Sample pump, WMM60, B4M 13 19-MAX-104
Safety board holder assembly 0 20-OZN-001
Tubing, PFA, 3/16-in. OD x 1/8-inch ID, 1 m length 5 m length 10-SCA-002
Tubing, PFA, 1/4-in. OD x 4 mm ID, 1 m length 5 m length 10-SCA-003
Tubing, PFA, 1/4-in. OD x 1/8-in. ID (6.35-mm OD x 3.18-mm ID), 1 m length 5 m length 10-SCA-006
Tubing, PFA, 3/16-in. OD x 1/16-inch ID, 1 m length 1 m length 10-SCA-007
Tubing, EMPP 562, 6.4 mm OD x 3.2 mm ID, 1 m length 2 m length 10-REH-002
2 Consumables/wearing parts: EMPP tubing, Y tube fittings, filters for the fan and vent, FPM/FKM tubing in the
ozone generator, catalyst in the ozone destructor, CO2 filter for the base reagent container, 24 V relay on the
ozone board, 24 V plug-in relays in the relay board (81204001), sample out valve, exhaust valve, acid valve,
diaphragm in the mixer reactor and the wetted parts of the sample valve (ARS valve).
3 Normally replaced at 24 month intervals.
38
Replacement parts and accessories
Quantity to
Description2 Item no.
stock
Tubing, EMPP, 5.6 mm OD x 2.4 mm ID, 1 m length 1 m length 10-REH-003
Tubing, EMPP, 6.9 mm OD x 4.1 mm ID, 1 m length 1 m length 10-REH-008
Valve, N/C with plug, Type 6606 Burkert 1 19-EMC-001
Valve, N/O with plug, Type 6606 Burkert 1 19-EMC-002
Valve, C/O with plug, Type 6606 Burkert 1 19-EMC-003
Valve, non-return (check valve), 1 psi 1 10-SMR-001
Valve, pinch, B4M, C/O, complete 0 12-BIO-001
Valve, sample, PEEK ARS, 3.0 mm with integrated fittings 13 10-EMT-089
Valve, sample, PEEK ARS, 2.5 mm with integrated fittings 13 10-EMT-090
2 Consumables/wearing parts: EMPP tubing, Y tube fittings, filters for the fan and vent, FPM/FKM tubing in the
ozone generator, catalyst in the ozone destructor, CO2 filter for the base reagent container, 24 V relay on the
ozone board, 24 V plug-in relays in the relay board (81204001), sample out valve, exhaust valve, acid valve,
diaphragm in the mixer reactor and the wetted parts of the sample valve (ARS valve).
39
Replacement parts and accessories
40
HACH COMPANY World Headquarters HACH LANGE GMBH HACH LANGE Sàrl
P.O. Box 389, Loveland, CO 80539-0389 U.S.A. Willstätterstraße 11 6, route de Compois
Tel. (970) 669-3050 D-40549 Düsseldorf, Germany 1222 Vésenaz
(800) 227-4224 (U.S.A. only) Tel. +49 (0) 2 11 52 88-320 SWITZERLAND
Fax (970) 669-2932 Fax +49 (0) 2 11 52 88-210 Tel. +41 22 594 6400
[email protected] [email protected] Fax +41 22 594 6499
www.hach.com www.de.hach.com
© Hach Company/Hach Lange GmbH, 2020, 2022, 2024. All rights reserved. Printed in Ireland.
*DOC023.52.90643*