U1240C Series Handheld Digital Multimeters

The U1240 Series handheld digital multimeters from Keysight Technologies are designed for technicians and industrial test engineers, offering durability with IP67 certification and a battery life of up to 400 hours. Key features include a 10,000-count display, harmonic ratio measurement, low impedance mode, and non-contact voltage detection, making them suitable for harsh environments and efficient testing. The devices also support data logging and connectivity options for streamlined productivity in various applications.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
30 views20 pages

U1240C Series Handheld Digital Multimeters

The U1240 Series handheld digital multimeters from Keysight Technologies are designed for technicians and industrial test engineers, offering durability with IP67 certification and a battery life of up to 400 hours. Key features include a 10,000-count display, harmonic ratio measurement, low impedance mode, and non-contact voltage detection, making them suitable for harsh environments and efficient testing. The devices also support data logging and connectivity options for streamlined productivity in various applications.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd

DATA SHEET

U1240 Series
Handheld digital multimeters

Introduction
For technicians or industrial test engineers, your ultimate goal is to keep the production line running
seamlessly to prevent any unplanned shutdown. The strong and solid Keysight Technologies, Inc.
U1240B and U1240C Series are just the DMMs you need to work efficiently on the bench or in the field.
Certified to IP 67 and with the durability to withstand drops up to 10 feet (3 meters) to suit harsh working
environments, the U1240C can work longer for you with up to 400 hours of battery life. Better yet, you can
optimize productivity with the unique remote data logging functionality via USB. Perform tests confidently
and efficiently — everything you need to get the job done.
Key Features
• 10,000-count display
• Up to 0.09% basic DCV accuracy
• Longer battery life up to 400 hours 1
• Certified to IP 67 for water and dust protection 1
• Tested to withstand a 3-meter (10-ft) drop 1 – CAT III 1000 V / CAT IV 600 V overvoltage protection
• Special features:
o Harmonic ratio measurement to quickly identify the presence of harmonics in AC signals
o ZLOW, low impedance mode 2 to eliminate false readings caused by stray voltages
o Vsense 2 for non-contact voltage detection
o T1 – T2 differential temperature 2 measurement
o Built-in flashlight 3

Test Confidently and Efficiently


Harmonics ratio
In real life, harmonics are present in AC power due to the fact that many electronic devices are powered
by switched-mode current pulses which travel back into the power source. If these unwanted harmonics
become too large, the undesirable side effects include overheating that shortens the lifespan of motors,
generators, and transformers, premature tripping of circuit breakers, and blown fuses. One of the fastest
ways to detect and gauge the percentage of distortion due to harmonics is by measuring the harmonic
ratio of the incoming AC voltage. You can perform this very easily on the U1242B/C handheld DMM, with
just a button press. The percentage helps you to decide if further analysis of the power source is required
with an oscilloscope or a spectrum analyzer.

Low impedance mode (Zlow)


Stray voltages are usually encountered in non-energized electrical wiring adjacent to powered wires due
to capacitive or inductive coupling between these wires. The low impedance mode helps to reduce false
readings by dissipating these stray voltages, thus improving safety and measurement efficiency during
voltage measurement.

Vsense for non-contact voltage detection


Vsense is a unique method of non-contact voltage detection that safeguards users from exposure to hot
or live wires while making measurements in dangerous working environments. Upon detection of high AC
voltage, it produces a unique combination of beeping audio and blinking LED lights to alert users.

1 U1240C Series only


2 U1242C Only
3 U1242B & U1242C only

2
Dual and differential temperature
In the installation, maintenance, and troubleshooting of heating, ventilation, and air-conditioning (HVAC)
systems, temperature measurements are essential. As an example, to ensure boiler temperature adheres
to safety requirements, you would need to measure both boiler and air temperature simultaneously to
achieve accurate real-time readings. With a faulty air conditioning system, viewing the temperature
difference between the warm return air and cool air supply helps to reveal the cooling behavior of the
evaporator with respect to time. Now with the convenience of the U1242B/C DMM, you need only one
instrument to efficiently get dual and differential measurements.

Figure 1. Dual and differential temperature for efficient testing of HVAC systems

Built-in flashlight with ergonomics in mind


Designed for handheld users working in poorly lit environments, the U1240C Series fits comfortably in
your palm, allowing you to single-handedly illuminate the test area with its easily activated built-in LED
flashlight. With ergonomics in mind, you can swiftly make more measurements without straining your
hands even over a longer period of time.

Figure 2. Ergonomically shaped with a built-in flashlight

3
Prolonged battery life and rugged build
The last thing you want is for your tools to run out of juice when you need it the most. The U1240C Series
of handheld DMMs lets you carry out tests and measurements over a prolonged duration. With up to 400
hours of battery life, you don’t have to worry about the hassles of battery change and just focus on
maximum productivity.

When operating in harsh environments, you need tools that are strong and solid enough to stand up to
the task. The U1240C Series DMMs are housed in robust overmolded enclosures and certified to IP 67
for dust and water resistance. Better yet, it is also designed to absorb the impact of a 3-meter (10-foot)
drop — especially useful in case of accidental drops during installation and maintenance work.

Perform data logging on-DMM, wired or wirelessly


Keep the production lines running smoothly with the U1240 Series DMMs’ data logging capabilities. Users
have two options to record individual readings. The first option is to simply record data into the main DMM
unit by utilizing the built-in internal memory of up to 2,000 readings storage. The second option is to
transmit data to a PC with the IR-USB cable utilizing the Keysight Handheld Meter Logger software. With
this capability, you can ensure every reading gets recorded accurately at intervals you specify. Better yet,
you can eliminate the conventional data entering process and generate error-free automated test reports
in various forms such as a graph, table, and statistical information or limit test results.

Figure 3. Easily capture test readings with the Keysight Handheld Meter Logger software

Manual-recording

Graph

Table

Single screen

Report generation

Figure 4. Get automatic report generated in various formats via PDF

4
Take a Closer Look at the U1240B Series
Front panel

Ambient temperature

MINMAX recording
Capacitance:
Resistance: 0.1 nF to 10 mF
0.1 Ω to 100 MΩ ACμA:
50 µA to 10 mA
Backlight:
Two intensity levels Frequency:
1 Hz to 200 kHz
Diode test switch counter
Manual data logging
(U1248B only)
DCV: 0.1 mV to 1000 V
ACmA:
ACV: 50 mV to 1000 V
5 mA to 440 mA
Harmonic ratio
(U1242B only) ACA:
0.5 A to 10 A
T1-T2 Differential temperature
(U1242B only) Temperature:
-40 to 1000 °C

5
Take a Closer Look at the U1240C Series
Front panel
Vsense* red LED indicator for
non-contact voltage detection

10,000-count dual display

Data logging:

• Manual data logging up to


100 readings, or,
Press-and-hold to switch on the
• Auto/event data logging up to
2,000 readings built-in flashlight

Harmonics ratio* to detect the T1-T2 differential temperature*

presence of harmonics content measurement

Broad range of measurement functions:


ZLow (low input impedance) *
Voltage, current, frequency, capacitance,
to remove ghost or induced
resistance, continuity, diode test and temperature
voltage from your measurement
CAT IV 600 V CAT III 1000 V
overvoltage protection
*U1242C only

Back panel

Built-in flashlight IR-USB connectivity

Probe holder/storage

Easy access to battery and


fuse compartment

Ergonomically designed for better grip

6
U1240C Series Comparison
Basic features U1241C U1242C
Display resolution 10, 000 10, 000
Auto/manual ranging Yes Yes
Analog bar graph Yes Yes
Backlight Yes Yes
AC bandwidth 2 kHz 2 kHz
True RMS AC AC (Switchable to averaging sense)
Measurements
Range 100 mV to 1000 V 100 mV to 1000 V
Voltage DC
Accuracy 0.09% + 2 cnts 0.09% + 2 cnts
Range 100 mV to 1000 V 100 mV to 1000 V
Voltage AC
Accuracy 1.0% + 3 cnts 1.0% + 3 cnts
Range 1000 µA to 10 A 1000 µA to 10 A
Current DC
Accuracy 0.1% + 2 cnts 0.1% + 2 cnts
Current AC Range 1000 µA to 10 A 1000 µA to 10 A
Accuracy 1% + 3 cnts 1% + 3 cnts
Resistance Range 1000 Ω to 100 MΩ 100 Ω to 100 MΩ
Accuracy 0.2% + 2 cnts 0.2% + 2 cnts
Frequency Range 100 Hz to 10 MHz 100 Hz to 10 MHz
Accuracy 0.02% + 1 cnts 0.02% + 1 cnts
Range 1000 nF to 10 mF 1000 nF to 10 mF
Capacitance
Accuracy 1% + 5 cnts 1% + 5 cnts
Temperature Range –200 °C to 1372 °C –200 °C to 1372 °C
(K-type thermocouple) Accuracy 1% + 1 °C 1% + 1 °C
Continuity with beeper Yes Yes
Diode test Yes Yes
Data management
Min/Max recording Yes Yes
Display hold Yes Yes
Auto hold Yes Yes
Null Yes Yes
PC-Connectivity Infrared (IR)-USB Infrared (IR)-USB
Special features
Harmonic ratio — Yes
Vsense: non-contact voltage detection — Yes
ZLOW - low impedance mode — Yes
T1 - T2 differential temperature measurement — Yes
Safety and regulatory
Over-voltage safety protection CAT III 1000 V / CAT IV 600 V CAT III 1000 V / CAT IV 600 V
General
–20 °C to 55 °C –20 °C to 55 °C
Operating temperature
0% to 80% R.H. 0% to 80% R.H.
Battery (included) 4 x 1.5 V AAA 4 x 1.5 V AAA
Battery life 400 hours 400 hours
Warranty 3-year 3-year
One year, or Two years
One year, or Two years
Calibration (with 1.5 times of one year
(with 1.5 times of one year specification)
specification)
Dimensions (H x W x D) 198 x 96 x 57 mm 198 x 96 x 57 mm

7
U1240C Series General Specifications
General specifications U1240C
Display 4 ½-digit Liquid Crystal Display (LCD) with maximum reading of
11,000-count
Automatic polarity indication
Power consumption 530 mVA maximum with backlight and flashlight.
Battery type 4x 1.5 V AA alkaline batteries (ANSI/NEDA 24A or IEC LR03), or;
4x 1.5 V AA zinc chloride batteries (ANSI/NEDA 24D or IEC R03), or;
4x 1.5 V AA lithium batteries (ANSI/NEDA 24-LF or IEC FR03)
Battery life 400 hours typical based on new alkaline batteries for DC voltage measurement
Connectivity Infrared (IR) port, connect with IR-USB cable
Fuse 10 x 35 mm, 440 mA / 1000 V, 10 kA minimum fast-acting fuse
10 x 38 mm, 11 A / 1000 V, 20 kA minimum fast-acting fuse
Operating environment Full accuracy from –20 °C to 55 °C; and up to 80% R.H. for temperature up to 30 °C decreasing
linearly to 50% R.H. at 55 °C
Pollution degree II
Altitude up to 3000 meters
Storage compliance –40 °C to 70 °C, 0 to 80% R. H. (with battery removed)
Safety compliance Refer to Declaration of Conformity for the latest revisions of regulatory compliance at
[Link]/go/conformity
Measurement category CAT III 1000 V / CAT IV 600 V
Electromagnetic Compatibility (EMC) Refer to Declaration of Conformity for the latest revisions of regulatory compliance at:
[Link]/go/conformity

Commercial limits compliance with EN61326-1


Influence of radiated immunity: In RF electromagnetic fields of 3 V/M
DC voltage measurement typical accuracy:
100 mV & 600 mV range; ±0.1% of range
DC current measurement typical accuracy:
1000 uA & 10 mA range; ±0.3% of range
10 A range; ±0.7% of range

Note: If used in close proximity to an RF transmitter or when subjected to continuously present


electromagnetic phenomena, some recoverable degradation of performance may occur.
Ingress protection rating IP-67, protected against dust and the effect of immersion between 15 cm and 1 m
Temperature coefficient 0.05 x (specified accuracy)/°C (from –20 °C to 18 °C or 28 °C to 55 °C)
Common Mode Rejection Ratio (CMRR) > 120 dB at DC, 50/60 Hz ±0.1% (1 kΩ unbalanced)
Normal Mode Rejection Ratio (NMRR) > 60 dB at 50/60 Hz ±0.1%
Dimensions (H x W x D) 198 x 96 x 57 mm
Weight 545 grams (with batteries)
Warranty Three years for main unit. Three months for standard shipped accessories
Calibration cycle One year, or; Two years (with 1.5 times of one year specification)
Specification assumptions
1. Accuracy is given as ± (% of reading + counts of least significant digit) at 23 °C ±5 °C, with relative humidity less
than 80% R.H.
2. Accuracy is specified for 1-year after calibration, at operating temperature of room temperature. Multiply 1.5 times
of the accuracy for 2-year after calibration, example ± (1.5 x% of reading + 1.5 x counts of least significant digit).
3. AC V/mV and AC µA/mA/A specifications are AC coupled.
4. True RMS measurement is valid from 5% of range to 100% of range.
5. For non-sinusoidal waveforms, add (0.5% of reading +0.3% of full scale) typically
6. Crest factor ≤ 3 at full-scale and decrease reciprocally for overange as 3× Full Scale/Input; except for the 100 mV,
600 mV and 1000 V ranges, where these ranges have crest factor ≤ 1.5 at full scale and decrease reciprocally for
overange as 1.5× Full Scale/Input. – Specification is based on 5 times/second of data refresh rate
7. After Z L O W voltage measurement, wait for at least 20 minutes to cool down thermal impact before making next
measurement.
8. CMRR and NMRR are based on 5 times/second of data refresh rate

8
U1240C Series DC specifications
Accuracy ± (% of reading + counts
Function Range Resolution Test current / burden voltage
of least significant digit)

100 mV1,3 0.01 mV 0.09% + 2 —


600 mV 1,3 0.1 mV 0.09% + 2 —
1000 mV 4 0.1 mV 0.09% + 2 —
10 V 4 0.001 V 0.09% + 2 —
Voltage
100 V4 0.01 V 0.09% + 2 —
1000 V4 0.1 V 0.09% + 2 —

ZLOW2,4
0.1 V 1% + 4 —
(applicable to 1000 V range)

1000 µA3 0.1 µA 0.1% + 2 0.032 V (30 Ω)


10 mA3 0.001 mA 0.1% + 2 0.32 V (30 Ω)
Current5 100 mA1,3 0.01 mA 0.2% + 2 0.2 V (0.5 Ω)
600 mA 1,3 0.1 mA 0.2% + 2 0.88 V (0.5 Ω)
10 A 2,4 0.001 A 0.3% + 5 0.5 V (0.01 Ω)
Diode test — 0.001 V 0.5% + 10 < 1.6 mA
Accuracy ± (% of reading + counts
Function Range Resolution Continuity threshold
of least significant digit)

100 Ω3,4,7 0.01 Ω 0.2% + 5 28 ±10 Ω


1000 Ω4 0.1 Ω 0.2% + 2 28 ±10 Ω
10 kΩ 0.001 kΩ 0.2% + 2 0.151 ±0.05 kΩ
Resistance / 100 kΩ 0.01 kΩ 0.2% + 2 1.38 ±0.5 kΩ
audible
continuity 1000 kΩ 0.1 kΩ 0.2% + 2 13.8 ±4.3 kΩ
10 MΩ5 0.001 MΩ 0.8% + 2 0.12 ±0.04 MΩ

1.5% + 3 (< 50 MΩ)


100 MΩ5,6 0.01 MΩ 0.12 ±0.04 MΩ
3.0% + 3 (> 50 MΩ)
Specification assumptions
A. Notes for voltage specification.
1. 100 mV and 600 mV range available on Temperature T1 terminal. The accuracy is specified for
10 MΩ (nominal) input impedance. The accuracy is specified after NULL function is used to zero
out thermal effect (by shorting test leads on a flat surface or a null value circuit).
2. Only available in U1242C only. 1.8 kΩ nominal input impedance for Z L O W mode.
3. Overload protection for 100 mV and 600 mV ranges: 1000 V r m s for circuits < 0.3 A short circuit
current.
4. Overload protection: 1000 V r m s .

B. Notes for current specification


1. Current can be measured up to 440 mA continuously. Maximum of 20 hours for measuring
current more than 440 mA and up to 600 mA. 100 mA and 600 mA ranges have thermal effect of
0.35 µA/mA to be offset after current applied to these ranges. Cool down the meter for at least 6
seconds if 100 mA was applied, and at least 3 minutes if 600 mA was applied; or alternatively
use the NULL function to zero-out thermal effect with open test lead before measuring the signal.
2. Current can be measured up to 10 A continuously. Maximum of 30 seconds for measuring current
more than 10 A to 20 A, add 0.3% to specified accuracy. The multimeter needs to be cool down
after measuring current that is more than 10 A. Cool down the meter for twice the duration of the
measured time and use NULL function to zero-out thermal effect before proceeding with lower
current measurement.
3. 1000 µA to 600 mA ranges (connection with mA terminal) overload protection by 10 x 35 mm,
440 mA/1000 V, and 10 kA minimum fast-acting fuse.
4. 10 A ranges (connection with A terminal) overload protection by 10 x 38 mm, 11 A / 1000 V, 20
kA minimum fast-acting fuse.
5. Ensure good ventilation and no heat element close to the meter.

C. Notes for diode test specification


1. Overload protection: 1000 V r m s for circuits < 0.3 A short circuit current.

9
2. Built-in buzzer sounds when reading is below 0.05 V approximately, and single tone for normal
forward-biased diode or semiconductor junction as 0.3 V ≤ reading ≤ 0.8 V.
3. The maximum threshold voltage display is less than +2 V.
D. Notes for resistance/audible Continuity specification
1. Maximum open voltage: < +2.4 V
2. Built-in buzzer sounds as transient when resistance less than 28 ±10 Ω. It may capture the
intermittent for longer than 1 ms.
3. 100 Ω range is for U1242C only
4. The accuracy is specified after Math Null, which is used to subtract the test lead resistance and
the thermal effect (by shorting the test leads tips on a flat surface or a null value circuit). Ensure
good ventilation and no heat element close to the meter.
5. For 10 MΩ and 100 MΩ ranges, the R.H. is specified for < 60% at 30 °C.
6. For 100 MΩ range: temperature coefficient is 0.1 x (specified accuracy) /°C.
7. Resistance range 100 Ω is typical characteristic.

U1240C Series AC voltage specifications


Function Range Resolution Accuracy ± (% of reading + counts of least significant digit)
40 Hz to 1 kHz (True RMS) 1 kHz to 2 kHz (Averaging sense)6

100 mV2 0.01 mV 1.0% + 3 1.5% + 3


600 mV2 0.1 mV 1.0% + 3 1.5% + 3
1000 mV 0.1 mV 1.0% + 3 1.5% + 3

AC voltage1,4 10 V 0.001 V 1.0% + 3 1.5% + 3


True RMS 100 V 0.01 V 1.0% + 3 1.5% + 3
1000 V 0.1 V 1.0% + 3 1.5% + 3
ZLOW3
(applicable to 0.1 V 2.0% + 4 N/A
1000 V range)
100 mV2 0.01 mV 1.0% + 5 1.5% + 5
600 mV 2 0.1 mV 1.0% + 5 1.5% + 5

AC voltage 1,4 1000 mV 0.1 mV 1.0% + 5 1.5% + 5


Averaging sense 10 V 0.001 V 1.0% + 5 1.5% + 5
100 V 0.01 V 1.0% + 5 1.5% + 5
1000 V 0.1 V 1.0% + 5 1.5% + 5
Notes:
1. Overload protection: 1000 Vrms.
2. 100 mV and 600 mV range available on Temperature T1 terminal. The accuracy is specified for 10 MΩ (nominal) input impedance. The
accuracy is specified after NULL function is used to zero out thermal effect (by shorting test leads on a flat surface or a null value
circuit).
3. Only available in U1242C only. 1.8 kΩ nominal input impedance for ZLOW mode.
4. The input signal is lower than the product of 20,000,000 V x Hz.

10
U1240C Series AC current specifications
Function Range Resolution Accuracy ± (% of reading + counts of least significant digit)
40 Hz to 1 kHz (True RMS) 1 kHz to 2 kHz (Averaging sense)6

1000 µA3 0.1 µA 1.0% + 3 1.2% + 5


10 mA 3 0.001 mA 1.0% + 3 1.2% + 5
AC current5
100 mA3 0.01 mA 1.0% + 3 1.2% + 5
1000 mA1,3 0.1 mA 1.0% + 3 1.2% + 5
10 A2,4 0.001 A 1.2% + 54 1.2% + 5
Notes:
1. Current can be measured up to 440 mA continuously. Maximum of 20 hours for measuring current more than 440 mA and up to 600
mA. 100 mA and 600 mA ranges have thermal effect of 0.35 µA/mA to be offset after current applied to these ranges. Cool down the
meter for at least 6 seconds if 100 mA was applied, and at least 3 minutes if 600 mA was applied; or alternatively use the NULL function
to zero-out thermal effect with open test lead before measuring the signal.
2. Current can be measured up to 10 A continuously. Maximum of 30 seconds for measuring current more than 10 A to 20 A, add 0.3% to
specified accuracy. The multimeter needs to be cool down after measuring current that is more than 10 A. Cool down the meter for
twice the duration of the measured time and use NULL function to zero-out thermal effect before proceeding with lower current
measurement.
3. 1000 µA to 1000 mA ranges (connection with mA terminal) overload protection by 10 x 35 mm, 440 mA / 1000 V, 10 kA minimum fast-
acting fuse.
4. 10 A ranges (connection with A terminal) overload protection by 10 x 38 mm, 11 A / 1000 V, 20 kA minimum fast-acting fuse.
5. Ensure good ventilation and no heat element close to the meter.
6. The averaging sense is calibrated for sine wave only. Add additional 0.05 counts /°C to accuracy from -20 °C to 18°C or –28 °C to 55
°C.

U1240C Series temperature specifications


Thermocouple type Range Resolution Accuracy ± (% of reading + as specified below)

–200°C to 1372°C 0.1°C 1% + 1°C


K
–328°F to 2502°F 0.1°F 1% + 1.8°F
–210°C to 1200°C 0.1°C 1% + 1°C
J5
–346°F to 2192°F 0.1°F 1% + 1.8°F
Notes:
1. The specification above is specified after 60 minutes of warm-up time.
2. The accuracy does not include the tolerance of the thermocouple probe.
3. Do not allow the temperature sensor to contact a surface hat is energized above 30 Vrms or 60 V DC. Such voltages pose a shock
hazard.
4. The temperature calculation is specified according to the safety standards of EN/IEC-60548-1 and NIST 175.
5. Only for U1242C.

11
U1240C Series capacitance specifications
Range Resolution Accuracy ± (% of reading + counts of least significant digit)

1000 nF 0.1 nF 1.0% + 5


10 µF 0.001 µF 1.0% + 5
100 µF 0.01 µF 1.0% + 5
1000 µF 0.01 µF 1.2% + 5
10 mF 0.001 mF 1.2% + 5
Notes:
1. Overload protection: 1000 Vrms for short circuits with < 0.3 A current.
2. The accuracy for all ranges is specified based on a film capacitor or better, and after the Null function is used to subtract the test lead
resistance and thermal effect (by opening the test leads).
3. The maximum display is 12000 counts selectable.

U1240C Series frequency specifications


Minimum input
Range Resolution Accuracy ± (% of reading + counts of least significant digit)
frequency

100.00 Hz 0.01 Hz 0.02% + 11


1000.0 Hz 0.1 Hz 0.02% + 1
10.000 kHz 0.001 kHz 0.02% + 1
0.5 Hz
100.00 kHz 0.01 kHz 0.02% + 1
1000.0 kHz 0.001 kHz 0.02% + 1
10.000 MHz 0.001 MHz 0.02% + 1, < 1 MHz
Note:
1. The frequency measurement is susceptible to error when measuring low-voltage, low-frequency signals. Shielding inputs from external
noise pickup is critical for minimizing measurement errors. Turning on LPF (low pass filter) may help to filter out the noise and achieve a
stable reading.

U1240C Series sensitivity for voltage measurement


Input range Minimum sensitivity (RMS sine wave)
Maximum input1 for
0.5 Hz to 20 kHz 20 kHz to 50 kHz 50 kHz to 100 kHz
specified accuracy
100 mV2 15 mV 7.2 mV 15 mV
600 mV2 15 mV 7.2 mV 15 mV
1000 mV 125 mV 60 mV 125 mV
10 V 1.25 V 0.6 V 1.25 V
100 V 12.5 V 6V 12.5 V
1000 V 60 V 60 V 60 V

12
U1240C Series sensitivity for current measurement
Input range Minimum sensitivity (RMS sine wave)
Maximum input1 for specified accuracy 0.5 Hz to 20 kHz 20 kHz to 30 kHz
100 μA 175 μA 60 μA
10 mA 1.75 mA 0.6 mA
100 mA 17.5 mA 6.0 mA
600 mA 100 mA 38 mA
10 A N/A 1.15 A (< 10 kHz)
Notes:
1. Refer to ‘AC specification’ for specified accuracy of maximum input.
2. 100 mV and 600 mV range available on Temperature T1 terminal.

U1240C Series harmonic ratio (U1242C only)


Range Frequency Voltage
0.0% to 99.9% 40 Hz to 1 kHz 100 mVAC to 1000 VAC

U1240C Series multimeter data refresh rate


Function Slow (times/second) Fast (times/second)
ACV (V or mV) 5 40
DCV (V or mV) 5 40
Ω 5 40
Diode 5 40
Auto diode 1
Capacitance 0.8 (< 1000 μF) —
DC μA, mA or A 5 40
AC μA, mA or A 5 40
Temperature 5 40
Frequency 1 (> 10 Hz) —

13
U1240B Series General Specifications
General specifications U1240B
Display Dual display (secondary display is intended for temperature function display only) consists of 4-digit
Liquid Crystal Display (LCD) with maximum reading of 11,000 counts. Automatic polarity indication.
Power consumption 0.22 VA maximum
Battery type and life Four single standard 1.5 V AAA batteries (Alkaline or Zinc Chloride type); 300 hours typical
Operating environment Full accuracy at –10 to 55 °C; and to 80% RH for temperatures up to 30 °C, decreasing linearly to
50% RH at 55 °C
0 – 2000 meters per IEC 61010-1 3rd edition CAT III, 1000 V/CAT IV,
600 V IEC 61010-1 3rd edition
Storage compliance –20 to 70°C
Measurement category CAT III 1000 V/CAT IV 600 V Overvoltage Protection, Pollution Degree 2
Safety & EMC compliance Refer to Declaration of Conformity for the latest revisions of regulatory compliance at:
[Link]/go/conformity

Commercial limits compliance with EN61326-1


Influence of radiated immunity: In RF electromagnetic fields of 3 V/M
DC voltage measurement typical accuracy:
1000 mV to 1000 V range; ±0.2% of range

DC current measurement typical accuracy:


1000 uA & 10000uA range; ±0.7% of range
100 mA range; ±0.22% of range
440 mA range; ±0.5% of range
10 A range; ±0.11% of range

Note: If used in close proximity to an RF transmitter or when subjected to continuously present


electromagnetic phenomena, some recoverable degradation of performance may occur.
Common mode rejection ratio (CMRR) > 90 dB at DC, 50/60 Hz ±0.1% (1 kΩ unbalanced)
Normal mode rejection ratio (NMRR) > 60 dB at 50/60 Hz ±0.1%
Crest factor Crest factor ≤ 3 at full-scale and decrease reciprocally for overange as 3× Full Scale/Input; except
for the 1000 V range, where this range has crest factor ≤ 1.5 at full scale and decrease reciprocally
for overange as 1.5× Full Scale/Input.
Temperature coefficient 0.1 × (specified accuracy)/°C (from –10 to 18 °C or 28 to 55 °C)
Shock and vibration Tested to IEC/EN 60068-2
Dimensions (H x W x D) 193.8 x 92.2 x 58.0 mm
Weight 450 g with batteries
400 g without batteries
Warranty Three years for main unit.
Three months for standard shipped accessories

14
U1240B Series DC specifications
Accuracy ±(% of reading + counts of least
Function Range Resolution Test current/burden voltage
significant digit)

1000.0 mV 0.1 mV — 0.09% + 5


Voltage 1 10.000 V 0.001 V —
0.09% + 2
100.00 V 0.01 V —
1000.0 V 0.1 V — 0.15% + 5
1000.0 μA 0.1 μA < 0.06 V (50 Ω) 0.1% + 3
10000 μA 1 μA < 0.55 V (50 Ω) 0.1% + 3
Current
100.00 mA 0.01 mA < 0.18 V (0.5 Ω) 0.2% + 3
440.0 mA 2 0.1 mA < 0.8 V (0.5 Ω) 0.5% + 3
10.000 A 3 0.001 A < 0.4 V (0.01 Ω) 0.6% + 5
1000.0 Ω 5 0.1 Ω 0.5 mA
10.000 kΩ 5 0.001 kΩ 50 μA 0.3% + 3
Resistance 4 100.00 kΩ 0.01 kΩ 4.91 μA
1000.0 kΩ 0.1 kΩ 447 nA
10.000 MΩ 0.001 MΩ 112 nA 0.8% + 3
100.00 MΩ 6 0.01 MΩ 112 nA 1.5% + 3
Diode test 7 1V 0.001 V approximately 0.5 mA 0.3% + 2
Notes:
1. Input impedance: 10 MΩ (nominal).
2. Current can be measured up to 440 mA continuously. An additional 0.2% needs to be added to the specified accuracy if the signal
measured is in the range of 440 mA to 1100 mA for 30 seconds maximum. After measuring a current of > 440 mA, leave the meter to
cool down for twice the measuring time used before applying a low current measurement.
3. Current can be measured up to 10 A continuously with a maximum operating temperature of 50 °C. An additional 0.3% needs to be
added to the specified accuracy if the signal measured is in the range of 10 A to 19.999 A for 15 seconds maximum. After measuring a
current of > 10 A, leave the meter to cool down for 60 seconds before applying a low current measurement.
4. The maximum open voltage is < 2.8 V. For instant continuity, the built-in buzzer sounds when resistance is < 10.0 Ω.
5. The accuracy of 1 kΩ and 10 kΩ is specified after Null function, which is used to substrate the test lead resistance and the thermal
effect.
6. For the range of 100 MΩ, the R.H. is specified for < 60%. The temperature coefficient will be 0.15 times of specified accuracy as > 50
MΩ.
7. Overload protection: 1000 V RMS for circuits < 0.3 A short circuit current. The built-in buzzer sounds when reading is approximately
below
50 mV and audible single tone for normal forward biased diode or semiconductor junction as 0.3 V ≤ Reading ≤ 0.8 V.

15
U1240B Series AC specifications
Test current/burden Accuracy ±(% of reading + counts of least significant digit)
Function` Range Resolution
voltage 40 to 500 Hz 500 Hz to 1 kHz 1 to 2 kHz
1000.0 mV 0.1 mV — 1% + 5 2% + 5 —
AC voltage 1,5 10.000 V 0.001 V — 1% + 5 1% + 5 2% + 5
True RMS 100.00 V 0.01 V — 1% + 5 1% + 5 2% + 5
1000.0 V 0.1 V — 1% + 5 1% + 5 —
0.1 μA
1000.0 μA < 0.06 V (50 Ω)
AC current 2,5
True RMS 10000 μA 1 μA < 0.55 V (50 Ω)
1% + 5 1% + 5 —
100.00 mA 0.01 mA < 0.18 V (0.5 Ω)
440.0 mA 3 0.1 mA < 0.8 V (0.5 Ω)
10.000 A 4 0.001 A < 0.4 V (0.01 Ω)
Notes:
1. Input impedance: 10 MΩ (nominal) in parallel with < 100 pF, with overload protection of 1000 V RMS.
2. Crest factor ≤ 3 at full-scale and decrease reciprocally for overange as 3× Full Scale/Input; except for the 1000 V range, where this
range has crest factor ≤ 1.5 at full scale and decrease reciprocally for overange as 1.5× Full Scale/Input. For non-sinusoidal waveforms,
add (2% of reading + 2% of full scale) typical.
3. Current can be measured from 50 mA to 440 mA continuously. An additional 0.2% needs to be added to the specified accuracy if the
signal measured is in the range of 440 mA to 1100 mA for 30 seconds maximum. After measuring a current of > 440 mA, leave the
meter to cool down for twice the measuring time used before application of low current measurement.
4. Current can be measured from 0.5 A up to 10 A continuously with a maximum operating temperature of 50 °C. An additional 0.3%
needs to be added to the specified accuracy if the signal measured is in the range of 10 A to 19.999 A for 15 seconds maximum. After
measuring a current of >10 A, leave the meter to cool down for 60 seconds before applying a low current measurement.
5. AC voltage and AC current specifications are AC coupled. True RMS measurement is valid from 5% of range to 100% of range.

U1240B Series temperature specifications


Thermocouple type Range Resolution Accuracy ±(% of reading + offset error)
K (for U1241B and U1242B) –40 to 1000 ºC/–48 to 1832 ºF 0.1 °C /0.1 °F 1% + 1 ºC/1% + 1.8 °F
J (for U1242B only) –40 to 1000 ºC/–48 to 1832 ºF 0.1 ºC/0.1 ºF 1% + 1 ºC/1% + 1.8 °F

U1240B Series capacitance specifications


Range Resolution Accuracy ±(% of reading + counts of least significant digit)
1000.0 nF 0.1 nF
10.000 μF 0.001 μF 1.2% + 4
100.00 μF 0.01 μF
1000.0 μF 0.1 μF
2% + 4
10.000 mF 0.001 mF

U1240B Series harmonic ratio specifications


Range Frequency Voltage
0.0% to 99.9% 40 to 500 Hz 100 mVAC to 1000 VAC

16
U1240B Series switch counter definition
Switch condition 1,2 Circuit switch Display 3 Switch threshold
Low level Normally close Lo < 370 ohms
Intermittent 4 Close to open Number of switch count Low to high transition
High level Normally open Hi > 430 ohms
Intermittent 5 Open to close Number of switch count High to low transition
Notes:
1. Detects intermittent Opens or Closes lasting for at least 250 μsec.
2. Test current of 0.5 mA with maximum open circuit voltage of 2.8 V is used.
3. Maximum count reading: 199.99 M (display “OL” when achieving 2 x 108 and thereafter).
4. Count only low to high transition for initial switch condition of Low.
5. Count only high to low transition for initial switch condition of High.

U1240B Series frequency specifications


Range Resolution Accuracy Minimum input frequency

100.00 Hz 0.01 Hz
1000.0 Hz 0.1 Hz
0.03% + 3 1 Hz
10.000 kHz 0.001 kHz
100.00 kHz 0.01 kHz
1000.0 kHz 1 0.1 kHz
Note:
1. Effective frequency measurement of up to 200 kHz; refer to frequency sensitivity table below for details.

U1240B Series frequency sensitivity during voltage


measurement
Minimum sensitivity (RMS sine wave)
Input range (Maximum input for specified accuracy = 10 x Range or 1000 V)
20 Hz to 50 kHz 50 to 200 kHz
1000.0 mV 0.3 V 0.6 V
10.000 V 0.5 V 1.8 V
100.00 V 5V 10 V (< 100 kHz)
1000.0 V 50 V 100 V (< 100 kHz)

U1240B Series frequency sensitivity during current


measurement
Minimum sensitivity (RMS sine wave)
Input range
20 Hz to 20 kHz
1000.0 μA 100 μA
10000 μA 500 μA
100.00 mA 10 mA
440.00 mA 50 mA
10.000 A 1A

17
U1240B Series measuring rate
Function Times/second
ACV 7
DCV (V or mV) 7
Ω 14
Diode 14
Capacitance 4 (< 100 μF)
DCA (μA, mA, A) 7
ACA (μA, mA, A) 7
Temperature 7 (single)
Frequency 1 (> 10 Hz)

Optional accessories Images


U1179A IR connectivity bracket
Connects the U1240B Series handheld DMM to the IR-to-USB cable

U1171A magnetic hanging kit


For fastening the DMM to a steel surface so both hands are free

U1174A carrying case, soft


For handheld digital multimeters

U1172A transit case (aluminium-clad)


The robust casing to transport your DMM and accessories
Aluminium-clad, black panel construction
Dimension: 18 inches (H) x 13 inches (W) x 6 inches (D)
Weight: 4 kg

U1595A rugged carrying case


High quality, water and dust proof carrying case designed to store up to
two handheld and accessories

U1161A extended test lead kit


Includes two test leads (red and black), two test probes, medium sized
alligator clips and 4-mm banana plugs.
Test leads: CAT III 1000 V, CAT IV 600 V, 15 A
Test probe (4-mm tips): CAT III 1000 V, CAT IV 600 V, 15 A
Medium-sized alligator clips: CAT III 1000 V, CAT IV 600 V, 15 A
4-mm banana plugs: CAT II 600 V, 10 A

18
Optional accessories Images
U1168B standard test lead kit
Includes two test leads (red and black), 4-mm test probes, alligator clips,
fine-tip test probes, SMT grabbers and mini grabbers.
Test leads: CAT III 1000 V, CAT IV 600 V, 15 A
Test probe (19-mm tips): CAT II 1000 V, 15 A
Test probe (4-mm tips): CAT III 1000 V, CAT IV 600 V, 15 A (highly
recommended for CAT IV environment)
Alligator clips: CAT III 1000 V, CAT IV 600 V, 15 A
Fine-tip test p [robes: CAT II 300 V, 3 A
SMT grabber: CAT II 300 V, 3 A
Mini grabber: CAT II 300 V, 3 A

U1162A alligator clips


One pair of insulated alligator clips (red and black). Recommended for
use with Keysight standard test leads.
Rated CAT III 1000 V, CAT IV 600 V, 15 A

U1163A SMT grabbers


One pair of SMT grabbers (red and black). Recommended for use with
Keysight standard test leads.
Rated CAT II 300 V, 3 A

U1164A fine-tip test probes


One pair of fine-tip test probes (red and black). Recommended for use
with Keysight standard test leads.
Rated CAT II 300 V, 3 A

U1169A test probe leads


Includes two test leads (red and black), and a pair each of 19-mm and
4-mm test probes.
Test leads: CAT III 1000 V, CAT IV 600 V, 15 A
Test probes (19-mm tip): CAT II 1000 V, 15 A
Test probes (4-mm tip): CAT III 1000 V, CAT IV 600 V, 15 A
(highly recommended for CAT IV environment)

U1180A temperature sensors and probes


Includes thermocouple adapter, thermocouple bead J-type and
thermocouple bead K-type.
T/C adapter J/K type
T/C bead J-type: –20 to 200 °C
T/C bead K-type: –20 to 200 °C

U1181A immersion temperature probe


Type K T/C for use in oil and other liquids
Measurement range: –50 to 700 °C
Includes adapter U1184A for connection to DMM

19
Optional accessories Images

U1182A industrial surface temperature probe


Type K T/C for use on still surfaces
Measurement range: –50 to 400 °C
Includes adapter U1184A for connection to DMM

U1183A air temperature probe


Type K T/C for use in air and non-caustic gas
Measurement range: –50 to 800 °C
Includes adapter U1184A for connection to DMM

U1184A temperature probe adapter


Mini-connector-to-banana-plug adapter for use with DMM

U1185A thermocouple (J-type) and temperature probe adapter


T/C adapter J/K type
T/C bead J-type: –20 to 200 °C

U1186A thermocouple (K-type) and temperature probe adapter


T/C adapter J/K type
T/C bead J-type: –20 to 200 °C

Keysight enables innovators to push the boundaries of engineering by quickly solving


design, emulation, and test challenges to create the best product experiences. Start your
innovation journey at [Link].