U1240C Series Handheld Digital Multimeters
U1240C Series Handheld Digital Multimeters
U1240 Series
Handheld digital multimeters
Introduction
For technicians or industrial test engineers, your ultimate goal is to keep the production line running
seamlessly to prevent any unplanned shutdown. The strong and solid Keysight Technologies, Inc.
U1240B and U1240C Series are just the DMMs you need to work efficiently on the bench or in the field.
Certified to IP 67 and with the durability to withstand drops up to 10 feet (3 meters) to suit harsh working
environments, the U1240C can work longer for you with up to 400 hours of battery life. Better yet, you can
optimize productivity with the unique remote data logging functionality via USB. Perform tests confidently
and efficiently — everything you need to get the job done.
Key Features
• 10,000-count display
• Up to 0.09% basic DCV accuracy
• Longer battery life up to 400 hours 1
• Certified to IP 67 for water and dust protection 1
• Tested to withstand a 3-meter (10-ft) drop 1 – CAT III 1000 V / CAT IV 600 V overvoltage protection
• Special features:
o Harmonic ratio measurement to quickly identify the presence of harmonics in AC signals
o ZLOW, low impedance mode 2 to eliminate false readings caused by stray voltages
o Vsense 2 for non-contact voltage detection
o T1 – T2 differential temperature 2 measurement
o Built-in flashlight 3
2
Dual and differential temperature
In the installation, maintenance, and troubleshooting of heating, ventilation, and air-conditioning (HVAC)
systems, temperature measurements are essential. As an example, to ensure boiler temperature adheres
to safety requirements, you would need to measure both boiler and air temperature simultaneously to
achieve accurate real-time readings. With a faulty air conditioning system, viewing the temperature
difference between the warm return air and cool air supply helps to reveal the cooling behavior of the
evaporator with respect to time. Now with the convenience of the U1242B/C DMM, you need only one
instrument to efficiently get dual and differential measurements.
Figure 1. Dual and differential temperature for efficient testing of HVAC systems
3
Prolonged battery life and rugged build
The last thing you want is for your tools to run out of juice when you need it the most. The U1240C Series
of handheld DMMs lets you carry out tests and measurements over a prolonged duration. With up to 400
hours of battery life, you don’t have to worry about the hassles of battery change and just focus on
maximum productivity.
When operating in harsh environments, you need tools that are strong and solid enough to stand up to
the task. The U1240C Series DMMs are housed in robust overmolded enclosures and certified to IP 67
for dust and water resistance. Better yet, it is also designed to absorb the impact of a 3-meter (10-foot)
drop — especially useful in case of accidental drops during installation and maintenance work.
Figure 3. Easily capture test readings with the Keysight Handheld Meter Logger software
Manual-recording
Graph
Table
Single screen
Report generation
4
Take a Closer Look at the U1240B Series
Front panel
Ambient temperature
MINMAX recording
Capacitance:
Resistance: 0.1 nF to 10 mF
0.1 Ω to 100 MΩ ACμA:
50 µA to 10 mA
Backlight:
Two intensity levels Frequency:
1 Hz to 200 kHz
Diode test switch counter
Manual data logging
(U1248B only)
DCV: 0.1 mV to 1000 V
ACmA:
ACV: 50 mV to 1000 V
5 mA to 440 mA
Harmonic ratio
(U1242B only) ACA:
0.5 A to 10 A
T1-T2 Differential temperature
(U1242B only) Temperature:
-40 to 1000 °C
5
Take a Closer Look at the U1240C Series
Front panel
Vsense* red LED indicator for
non-contact voltage detection
Data logging:
Back panel
Probe holder/storage
6
U1240C Series Comparison
Basic features U1241C U1242C
Display resolution 10, 000 10, 000
Auto/manual ranging Yes Yes
Analog bar graph Yes Yes
Backlight Yes Yes
AC bandwidth 2 kHz 2 kHz
True RMS AC AC (Switchable to averaging sense)
Measurements
Range 100 mV to 1000 V 100 mV to 1000 V
Voltage DC
Accuracy 0.09% + 2 cnts 0.09% + 2 cnts
Range 100 mV to 1000 V 100 mV to 1000 V
Voltage AC
Accuracy 1.0% + 3 cnts 1.0% + 3 cnts
Range 1000 µA to 10 A 1000 µA to 10 A
Current DC
Accuracy 0.1% + 2 cnts 0.1% + 2 cnts
Current AC Range 1000 µA to 10 A 1000 µA to 10 A
Accuracy 1% + 3 cnts 1% + 3 cnts
Resistance Range 1000 Ω to 100 MΩ 100 Ω to 100 MΩ
Accuracy 0.2% + 2 cnts 0.2% + 2 cnts
Frequency Range 100 Hz to 10 MHz 100 Hz to 10 MHz
Accuracy 0.02% + 1 cnts 0.02% + 1 cnts
Range 1000 nF to 10 mF 1000 nF to 10 mF
Capacitance
Accuracy 1% + 5 cnts 1% + 5 cnts
Temperature Range –200 °C to 1372 °C –200 °C to 1372 °C
(K-type thermocouple) Accuracy 1% + 1 °C 1% + 1 °C
Continuity with beeper Yes Yes
Diode test Yes Yes
Data management
Min/Max recording Yes Yes
Display hold Yes Yes
Auto hold Yes Yes
Null Yes Yes
PC-Connectivity Infrared (IR)-USB Infrared (IR)-USB
Special features
Harmonic ratio — Yes
Vsense: non-contact voltage detection — Yes
ZLOW - low impedance mode — Yes
T1 - T2 differential temperature measurement — Yes
Safety and regulatory
Over-voltage safety protection CAT III 1000 V / CAT IV 600 V CAT III 1000 V / CAT IV 600 V
General
–20 °C to 55 °C –20 °C to 55 °C
Operating temperature
0% to 80% R.H. 0% to 80% R.H.
Battery (included) 4 x 1.5 V AAA 4 x 1.5 V AAA
Battery life 400 hours 400 hours
Warranty 3-year 3-year
One year, or Two years
One year, or Two years
Calibration (with 1.5 times of one year
(with 1.5 times of one year specification)
specification)
Dimensions (H x W x D) 198 x 96 x 57 mm 198 x 96 x 57 mm
7
U1240C Series General Specifications
General specifications U1240C
Display 4 ½-digit Liquid Crystal Display (LCD) with maximum reading of
11,000-count
Automatic polarity indication
Power consumption 530 mVA maximum with backlight and flashlight.
Battery type 4x 1.5 V AA alkaline batteries (ANSI/NEDA 24A or IEC LR03), or;
4x 1.5 V AA zinc chloride batteries (ANSI/NEDA 24D or IEC R03), or;
4x 1.5 V AA lithium batteries (ANSI/NEDA 24-LF or IEC FR03)
Battery life 400 hours typical based on new alkaline batteries for DC voltage measurement
Connectivity Infrared (IR) port, connect with IR-USB cable
Fuse 10 x 35 mm, 440 mA / 1000 V, 10 kA minimum fast-acting fuse
10 x 38 mm, 11 A / 1000 V, 20 kA minimum fast-acting fuse
Operating environment Full accuracy from –20 °C to 55 °C; and up to 80% R.H. for temperature up to 30 °C decreasing
linearly to 50% R.H. at 55 °C
Pollution degree II
Altitude up to 3000 meters
Storage compliance –40 °C to 70 °C, 0 to 80% R. H. (with battery removed)
Safety compliance Refer to Declaration of Conformity for the latest revisions of regulatory compliance at
[Link]/go/conformity
Measurement category CAT III 1000 V / CAT IV 600 V
Electromagnetic Compatibility (EMC) Refer to Declaration of Conformity for the latest revisions of regulatory compliance at:
[Link]/go/conformity
8
U1240C Series DC specifications
Accuracy ± (% of reading + counts
Function Range Resolution Test current / burden voltage
of least significant digit)
ZLOW2,4
0.1 V 1% + 4 —
(applicable to 1000 V range)
9
2. Built-in buzzer sounds when reading is below 0.05 V approximately, and single tone for normal
forward-biased diode or semiconductor junction as 0.3 V ≤ reading ≤ 0.8 V.
3. The maximum threshold voltage display is less than +2 V.
D. Notes for resistance/audible Continuity specification
1. Maximum open voltage: < +2.4 V
2. Built-in buzzer sounds as transient when resistance less than 28 ±10 Ω. It may capture the
intermittent for longer than 1 ms.
3. 100 Ω range is for U1242C only
4. The accuracy is specified after Math Null, which is used to subtract the test lead resistance and
the thermal effect (by shorting the test leads tips on a flat surface or a null value circuit). Ensure
good ventilation and no heat element close to the meter.
5. For 10 MΩ and 100 MΩ ranges, the R.H. is specified for < 60% at 30 °C.
6. For 100 MΩ range: temperature coefficient is 0.1 x (specified accuracy) /°C.
7. Resistance range 100 Ω is typical characteristic.
10
U1240C Series AC current specifications
Function Range Resolution Accuracy ± (% of reading + counts of least significant digit)
40 Hz to 1 kHz (True RMS) 1 kHz to 2 kHz (Averaging sense)6
11
U1240C Series capacitance specifications
Range Resolution Accuracy ± (% of reading + counts of least significant digit)
12
U1240C Series sensitivity for current measurement
Input range Minimum sensitivity (RMS sine wave)
Maximum input1 for specified accuracy 0.5 Hz to 20 kHz 20 kHz to 30 kHz
100 μA 175 μA 60 μA
10 mA 1.75 mA 0.6 mA
100 mA 17.5 mA 6.0 mA
600 mA 100 mA 38 mA
10 A N/A 1.15 A (< 10 kHz)
Notes:
1. Refer to ‘AC specification’ for specified accuracy of maximum input.
2. 100 mV and 600 mV range available on Temperature T1 terminal.
13
U1240B Series General Specifications
General specifications U1240B
Display Dual display (secondary display is intended for temperature function display only) consists of 4-digit
Liquid Crystal Display (LCD) with maximum reading of 11,000 counts. Automatic polarity indication.
Power consumption 0.22 VA maximum
Battery type and life Four single standard 1.5 V AAA batteries (Alkaline or Zinc Chloride type); 300 hours typical
Operating environment Full accuracy at –10 to 55 °C; and to 80% RH for temperatures up to 30 °C, decreasing linearly to
50% RH at 55 °C
0 – 2000 meters per IEC 61010-1 3rd edition CAT III, 1000 V/CAT IV,
600 V IEC 61010-1 3rd edition
Storage compliance –20 to 70°C
Measurement category CAT III 1000 V/CAT IV 600 V Overvoltage Protection, Pollution Degree 2
Safety & EMC compliance Refer to Declaration of Conformity for the latest revisions of regulatory compliance at:
[Link]/go/conformity
14
U1240B Series DC specifications
Accuracy ±(% of reading + counts of least
Function Range Resolution Test current/burden voltage
significant digit)
15
U1240B Series AC specifications
Test current/burden Accuracy ±(% of reading + counts of least significant digit)
Function` Range Resolution
voltage 40 to 500 Hz 500 Hz to 1 kHz 1 to 2 kHz
1000.0 mV 0.1 mV — 1% + 5 2% + 5 —
AC voltage 1,5 10.000 V 0.001 V — 1% + 5 1% + 5 2% + 5
True RMS 100.00 V 0.01 V — 1% + 5 1% + 5 2% + 5
1000.0 V 0.1 V — 1% + 5 1% + 5 —
0.1 μA
1000.0 μA < 0.06 V (50 Ω)
AC current 2,5
True RMS 10000 μA 1 μA < 0.55 V (50 Ω)
1% + 5 1% + 5 —
100.00 mA 0.01 mA < 0.18 V (0.5 Ω)
440.0 mA 3 0.1 mA < 0.8 V (0.5 Ω)
10.000 A 4 0.001 A < 0.4 V (0.01 Ω)
Notes:
1. Input impedance: 10 MΩ (nominal) in parallel with < 100 pF, with overload protection of 1000 V RMS.
2. Crest factor ≤ 3 at full-scale and decrease reciprocally for overange as 3× Full Scale/Input; except for the 1000 V range, where this
range has crest factor ≤ 1.5 at full scale and decrease reciprocally for overange as 1.5× Full Scale/Input. For non-sinusoidal waveforms,
add (2% of reading + 2% of full scale) typical.
3. Current can be measured from 50 mA to 440 mA continuously. An additional 0.2% needs to be added to the specified accuracy if the
signal measured is in the range of 440 mA to 1100 mA for 30 seconds maximum. After measuring a current of > 440 mA, leave the
meter to cool down for twice the measuring time used before application of low current measurement.
4. Current can be measured from 0.5 A up to 10 A continuously with a maximum operating temperature of 50 °C. An additional 0.3%
needs to be added to the specified accuracy if the signal measured is in the range of 10 A to 19.999 A for 15 seconds maximum. After
measuring a current of >10 A, leave the meter to cool down for 60 seconds before applying a low current measurement.
5. AC voltage and AC current specifications are AC coupled. True RMS measurement is valid from 5% of range to 100% of range.
16
U1240B Series switch counter definition
Switch condition 1,2 Circuit switch Display 3 Switch threshold
Low level Normally close Lo < 370 ohms
Intermittent 4 Close to open Number of switch count Low to high transition
High level Normally open Hi > 430 ohms
Intermittent 5 Open to close Number of switch count High to low transition
Notes:
1. Detects intermittent Opens or Closes lasting for at least 250 μsec.
2. Test current of 0.5 mA with maximum open circuit voltage of 2.8 V is used.
3. Maximum count reading: 199.99 M (display “OL” when achieving 2 x 108 and thereafter).
4. Count only low to high transition for initial switch condition of Low.
5. Count only high to low transition for initial switch condition of High.
100.00 Hz 0.01 Hz
1000.0 Hz 0.1 Hz
0.03% + 3 1 Hz
10.000 kHz 0.001 kHz
100.00 kHz 0.01 kHz
1000.0 kHz 1 0.1 kHz
Note:
1. Effective frequency measurement of up to 200 kHz; refer to frequency sensitivity table below for details.
17
U1240B Series measuring rate
Function Times/second
ACV 7
DCV (V or mV) 7
Ω 14
Diode 14
Capacitance 4 (< 100 μF)
DCA (μA, mA, A) 7
ACA (μA, mA, A) 7
Temperature 7 (single)
Frequency 1 (> 10 Hz)
18
Optional accessories Images
U1168B standard test lead kit
Includes two test leads (red and black), 4-mm test probes, alligator clips,
fine-tip test probes, SMT grabbers and mini grabbers.
Test leads: CAT III 1000 V, CAT IV 600 V, 15 A
Test probe (19-mm tips): CAT II 1000 V, 15 A
Test probe (4-mm tips): CAT III 1000 V, CAT IV 600 V, 15 A (highly
recommended for CAT IV environment)
Alligator clips: CAT III 1000 V, CAT IV 600 V, 15 A
Fine-tip test p [robes: CAT II 300 V, 3 A
SMT grabber: CAT II 300 V, 3 A
Mini grabber: CAT II 300 V, 3 A
19