0% found this document useful (0 votes)
71 views1 page

Central Monitor Spec

The document outlines the technical specifications for a central monitoring system, including 10 units of the main unit, LAN cables, UPS, hub, and server. Key features include a 24-inch display, the capability to monitor 16 patients simultaneously, and various vital parameters. Additional requirements specify that the manufacturer must match that of the patient monitors and that installation should be handled by the supplier.

Uploaded by

Classic Bobby
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
71 views1 page

Central Monitor Spec

The document outlines the technical specifications for a central monitoring system, including 10 units of the main unit, LAN cables, UPS, hub, and server. Key features include a 24-inch display, the capability to monitor 16 patients simultaneously, and various vital parameters. Additional requirements specify that the manufacturer must match that of the patient monitors and that installation should be handled by the supplier.

Uploaded by

Classic Bobby
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd

TECHNICAL SPECIFICATION

CS ICU-3, NS ICU-3
Code No. SCU-3, NCU-3, CCU-2 Description : Central monitor Quantity : 10 units
SICU-3
Composition
1. Main unit : 1 unit x 10 units = 10 units
2. LAN cable between central monitor and patient monitor (necessary length)
: 1 set x 10 units = 10 sets
3. UPS : 1 set x 10 units = 10 sets
4. Hub : 1 set x 10 units = 10 sets
5. Server : 1 unit x 10 units = 10 units

Specifications
1. Main unit
(1) Display : 24 inch wide colour LCD
(2) No. of patients monitoring at once : 16 patients
(3) Parameters : ECG, BP, respiration, EEG,
Ventilation CO2 fluctuation line、outer input, pulse,
FLOW, Paw
other connecting devices to bedside monitors
(4) Numerical indication Parameters : heartrate, VPC number, ventilation, pulse, BP
(max.,min., average), NIBP(max.,min., average)
temp., difference temp, and blood temp, ETCO2
tcPO2, tCPCO2, SpO2, ST level, FiO2, CO, CCO
Ventilator, Anesthetic gas, BIS
other connecting devices to bedside monitors
(5) Power requirement : AC230V / 50Hz, single phase

2. LAN cable between central monitor and patient monitor (necessary length) : provided

3. UPS
(1) Power requirement : AC230V / 50Hz, single phase
(2) Capacity : Suitable for this item
(3) Response time : 0.8 second or less after power failure
(4) Backup time : 10 minutes or more

4. Hub
(1) Port : 10BASE-T/100BASE-TX (RJ-45 connecter)
Auto-negotiation, MDI/MDI-X automatic-transfer
(2) Number of Port : 16 or more
(3) Speed : 10Mbps/100Mbps
(4) Performance : Switching delay, Switching fabric
Memory capacity 160KB or more
(5) Power requirement : AC230V / 50Hz, single phase

5. Server
(1) Data saving period : Crimp, ST recall, trend:120 hours or more/per patient
Arrhythmia recall:8 sec. 1500 file or more/per patient
(2) Recorder : Recorder or Network printer

Spare parts
Nil

Consumables
Nil

Notes
Nil
*Manufacturer of this item should be same with Patient Monitor for ICU and HCU.
*Connecting work between central monitor and bedside monitor should be done by Supplier.