TECHNICAL SPECIFICATION
CS ICU-3, NS ICU-3
Code No. SCU-3, NCU-3, CCU-2 Description : Central monitor Quantity : 10 units
SICU-3
Composition
1. Main unit : 1 unit x 10 units = 10 units
2. LAN cable between central monitor and patient monitor (necessary length)
: 1 set x 10 units = 10 sets
3. UPS : 1 set x 10 units = 10 sets
4. Hub : 1 set x 10 units = 10 sets
5. Server : 1 unit x 10 units = 10 units
Specifications
1. Main unit
(1) Display : 24 inch wide colour LCD
(2) No. of patients monitoring at once : 16 patients
(3) Parameters : ECG, BP, respiration, EEG,
Ventilation CO2 fluctuation line、outer input, pulse,
FLOW, Paw
other connecting devices to bedside monitors
(4) Numerical indication Parameters : heartrate, VPC number, ventilation, pulse, BP
(max.,min., average), NIBP(max.,min., average)
temp., difference temp, and blood temp, ETCO2
tcPO2, tCPCO2, SpO2, ST level, FiO2, CO, CCO
Ventilator, Anesthetic gas, BIS
other connecting devices to bedside monitors
(5) Power requirement : AC230V / 50Hz, single phase
2. LAN cable between central monitor and patient monitor (necessary length) : provided
3. UPS
(1) Power requirement : AC230V / 50Hz, single phase
(2) Capacity : Suitable for this item
(3) Response time : 0.8 second or less after power failure
(4) Backup time : 10 minutes or more
4. Hub
(1) Port : 10BASE-T/100BASE-TX (RJ-45 connecter)
Auto-negotiation, MDI/MDI-X automatic-transfer
(2) Number of Port : 16 or more
(3) Speed : 10Mbps/100Mbps
(4) Performance : Switching delay, Switching fabric
Memory capacity 160KB or more
(5) Power requirement : AC230V / 50Hz, single phase
5. Server
(1) Data saving period : Crimp, ST recall, trend:120 hours or more/per patient
Arrhythmia recall:8 sec. 1500 file or more/per patient
(2) Recorder : Recorder or Network printer
Spare parts
Nil
Consumables
Nil
Notes
Nil
*Manufacturer of this item should be same with Patient Monitor for ICU and HCU.
*Connecting work between central monitor and bedside monitor should be done by Supplier.