440 SERIES
RoHS 3
COMPLIANT
2015/863/EU
VERSATILE 4-IN-1
FUNCTIONALITY The 440 Series provides advanced 4-in-1 test
capability in a convenient one-box solution. This
INTUITIVE USER INTERFACE new series performs AC Hipot (448 - 500 VA), DC
Hipot, Insulation Resistance and 40A AC Ground
Bond tests while taking up minimal production line
40A GROUND BOND space. The 440 Series is simple and easy-to-use;
CAPABILITY reducing setup time and increasing production
line throughput for your application. With multiple
memories and an optional USB port for remote BUS
20 PROGRAMMABLE
communication so you can quickly perform tests
MEMORIES
on a variety of DUTs from the front panel or with a
PC.
EASILY AUTOMATE FOR
DATA COLLECTION
40A
REMOTE SAFETY
Ground
INTERLOCK AC Hipot DC Hipot Insulation
Resistance Bond
446 • • • •
448 500VA • • •
EASILY SAFEGUARD YOUR
WORKSTATION WITH PPE
ACCESSORIES
4 Electrical Safety Compliance Simplified.® hipot.com
440 SERIES SPECS FOUND ON PAGE 10 & 11
RELEVANT APPLICATIONS SERIES FEATURES
APPLIANCE CONTRACT MANUFACTURING
INDUSTRIAL EQUIPMENT LABORATORY EQUIPMENT
INFORMATION TECHNOLOGY 50Hz 60Hz
Test Setup Frequency
WHAT'S IN THE BOX Memories Selection
33189 Power Cord (10A) 6ft - *446 Model
38071 Power Cord (15A) 6ft - *448 Model
99-10783-01 Fuse 10A Slow Blow 20mm - *446 Model
99-10168-01 Fuse 15A Fast Blow 30mm - *448 Model
38075 Interlock Connector Male Ramp Dwell
40428 Ground Bond Test Lead
40429 Ground Bond Return Lead
04040A-08 High Voltage Test Lead 6ft
99-10797-01 USB A-B 1.8M Cable
All testers come with the accessories you need to run
a test right out of the box. Low Current Safety Agency
Sense Listed
OPTIONS
Description 446 448
Rear Outputs • •
USB Port • • USB
PLC Remote (option-
al)
On the Go Intuitive User
Portability Interface
SCI 448 Model
Toll Free +1-800-504-0055 Phone +1-847-932-3662 5
INDEX
440 SERIES SPECIFICATIONS
INPUT (446 and 448) DIELECTRIC WITHSTAND TEST MODE (Cont.)
Voltage 100 - 120Vac / 200 - 240Vac±10% Auto Range Output
Regulation ± (1% of output + 5V), From no load to full load
Frequency 50/60Hz ± 5%
Dwell Timer Range: 0, 0.2 - 60.0 (0=continuous)
Fuse 446 10A / 250Vac Slow-Blow Resolution: 0.1
Accuracy: ± (0.1% of setting + 0.05 sec)
448 15A / 250Vac Fast-Blow
DIELECTRIC WITHSTAND TEST MODE
Ramp Timer Range: 0.2-180.0
Output Rating 446 5 KV @ 20 mA AC
6 KV @ 5 mA DC Resolution: 0.1
Accuracy: ± (0.1% of setting + 0.05 sec)
448 5 KV @ 99.99 mA AC
6 KV @ 10 mA DC
Voltage Range 0 – 5.00 KV AC
Setting/Display
INSULATION RESISTANCE TEST MODE
0 – 6.00 KV DC
Output Voltage, VDC Range: 100 - 1000
Resolu- 0.01 Resolution: 1
tion Accuracy: ± (2% of setting + 5V)
Accuracy ± (2% of setting + 5V)
Hi-Limit Range: 0, 1 - 1000 (0 = OFF)
Current Display 446 Range: 0 - 20.00 mA AC, 0 - 5.00 mA DC resistance, MΩ Resolution: 1
Resolution: 0.01 mA Accuracy: 100-499V ± (7% of setting + 2
Accuracy: ± (2% of reading + 0.02 mA) counts)
448 Range: 0 - 99.99 mA AC, 0 - 10.00 mA DC Lo-Limit Range: 1 - 1000
Resolution: 0.01 mA resistance, MΩ Resolution: 1
Accuracy: ± (2% of reading + 0.06 mA) Accuracy: 500-1000V ± (3% of setting + 2
counts)
Hi-Limit 446 AC Range: Lo-Limit 0 - 20.00 mA, Hi-Limit 0.10 –
20.00 mA Ramp Time , second Range: 0.1 or 2.0
Lo-Limit Resolution: 0.01 mA Resolution: 0.1
Accuracy: ± (2% of setting + 2 counts) Accuracy: ± (0.1% of setting + 0.05 sec)
446 DC Range: Lo-Limit 0 - 5.00 mA, Hi-Limit 0.02 –
5.00 mA Delay Time, second Range: 0, 0.5 - 999.9 (0=continuous)
Resolution: 0.01 mA Resolution: 0.1
Accuracy: ± (2% of setting + 2 counts) Accuracy: ± (0.1% of setting + 0.05 sec)
448 AC Range: Lo-limit 0 - 99.99 mA, Hi-Limit 0.10 –
99.99 mA For reading specifications, please refer to the user manual.
Resolution: 0.01 mA
Accuracy: ± (2% of reading + 6 counts) Specifications subject to change without notice.
448 DC Range: Lo-Limit 0 - 10.00 mA, Hi-Limit 0.02 –
10.00mA
Resolution: 0.01 mA
Accuracy: ± (2% of reading + 6 counts)
Failure Detector Audible and Visual
DC Output 446 <5 % ( 6KV / 5mA at Resistive Load )
Ripple
448 <5 % ( 6KV / 10mA at Resistive Load )
Discharge Time 446 < 50 ms for no load, < 200 ms for capacitor load
448 < 50 ms for no load, < 100 ms for capacitor load
Max. Capacitive 1µF < 1KV 0.08µF < 4KV
Load in DC Mode 0.75µF < 2KV 0.04µF < 5KV
0.5µF < 3KV 0.015uF < 6KV
AC Wave Form Sine Wave, Crest Factor = 1.3 - 1.5 and output voltage > 300V
AC Output 50Hz/60Hz ± 0.1%, User Selection
Frequency
10 Electrical Safety Compliance Simplified.® hipot.com
INDEX
440 SERIES SPECIFICATIONS
GROUND BOND GENERAL SPECIFICATIONS
Output AC Range: 1.0 – 40.0 Memories Allows storage of up to 20 different test programs and
Current, A Resolution: 0.1 a single step mode
Accuracy: ± (2 % of setting
+ 2 counts)
Remote I/O Input: Test, Reset, Interlock & Recall Mem-
Output AC ory 1 – 6
8V(Fixed)
Voltage, V
Output: Pass, Fail, Test-in-Process
Output
Frequency, Hz 50Hz/60Hz ± 0.1%, User Selectable
Maximum Interface
1.0–10.0A/0–600mΩ,10.1–30.0A/0–200mΩ, USB (Optional)
Loading
30.1–40.0A/0–150mΩ
Security
Lockout capability to avoid unauthorized access to
Offset, mΩ Range: 0-100 test set-up programs
Resolution: 1
Accuracy: ± (2 % of setting + 2 counts) Calibration Software and adjustments made through the front
panel
HI and LO-Limit Range: 0 – 150 (30.1-40.0A)
Resistance, mΩ Resolution: 1 Dimension 446 280mm(W) ×89mm(H) × 400 mm(D)
Accuracy: ± (2 % of setting + 2 counts)
448 430mm(W) ×132mm(H) × 400 mm(D)
Range: 0 – 200 (10.1-30.0A) Weight 446 18lbs (8kg)
Resolution: 1
Accuracy: ± (2 % of setting + 2 counts) 448 53lbs (24kg)
For reading specifications, please refer to the user manual.
Range: 0 – 600 (1.0-10.0A)
Resolution: 1 Specifications subject to change without notice.
Accuracy: ± (2 % of setting + 2 counts)
Fix Ramp Timer, second Range: 0.4
Resolution: 0.1
Accuracy ± 0.05 sec
0-250mOhm = 0.1sec ramp up
251-300mOhm = 0.2sec ramp up
301-450mOhm = 0.3sec ramp up
>450mOhm = 0.4sec
Dwell Timer, second Range: 0, 0.1 - 240.0 (0 = continuous)
Resolution: 0.1
Accuracy ± 0.05 sec
Toll Free +1-800-504-0055 Phone +1-847-932-3662 11