MV110
ISOIL
I N D U S T R I A
MV110_IT_EN_DS053REV08_IS
INDEX
TECHNICAL DATA 3
OVERALL DIMENSIONS 4
The manufacturer guarantees only English text available on our web site [Link]
2/26 MV110_IT_EN_DS053REV08_IS
TECHNICAL DATA
OVERALL FEATURES
Suitable For All the ISOMAG® sensors up to the DN 1000
Minimum conductivity 5 µS/cm
Altitude -200 m up to 4000 m
STANDARD FEATURES
Compact
Version
Separate
Protection Rate IP 67
OPTIONAL FEATURES
(CHECK HOW TO ORDER, AT LAST PAGE, FOR MORE DETAILS)
Protection Rate IP 68
Conn. Sensor Cable CABLE C014 for separate version
LCD Display Graphic display 128x64 pixels back light, 3 programming keys
Power supply : 24 … 36 VAC/VDC 0...45/66 Hz
Power supply : 12…48 VDC
Power Supply/Consumption Power supply : 100 ... 240 VAC 44/66 Hz + 1 Rechargeable Battery
Power supply : 24 … 36 VAC/VDC 0...44/66 Hz + 1 Rechargeable Battery
Power supply : 12…48 VDC + 1 Rechargeable Battery
Outputs: N°1 digital Output , 1250 Hz, 100mA, 30 Vdc
Pulses/Frequence/Alarms N°2 digital Outputs , 1250 Hz, 100mA, 30 Vdc
n° 1 Analogue output 0/4...20/22 mA (Hart optional)
Analog Output
n° 2 Analogue outputs 0/4...20/22 mA (Hart optional over Out.1)
RS 485
Communication Gateway
Wi-Fi (for programming)
ACCURACY
Flow rate (volume) = ±0,05% c.r
Measurements tolerance Out 4/20 mA = ± 0,08 % c.r
Frequency Out = ± 0,08% c.r
Accuracy (Whole System
ISOIL
See table below
Converter+Sensor)
3/26 MV110_IT_EN_DS053REV08_IS I N D U S T R I A
The manufacturer guarantees only English text available on our web site [Link]
4/26
180 200 Compact version
OVERALL DIMENSIONS
114
114
MV110_IT_EN_DS053REV08_IS
195
164
167
290
The manufacturer guarantees only English text available on our web site [Link]
295
5/26
Separate (wall) version
30
Ø7
MV110_IT_EN_DS053REV08_IS
28
I
142
N
D
U
S
T
R
ISOIL
120
I
A
Horizontal version
144 56
200
Junction box
Junction box
The manufacturer guarantees only English text available on our web site [Link]
Vertical version
87
270
75
Junction box Junction box
6/26 MV110_IT_EN_DS053REV08_IS
DESCRIPTION 3
POS.
PA6 VERSION ALUMINIUM VERSION 4
22 SCREW M6x16
23 GROWER WASHER Ø6 13
24 PG11 CABLE GLAND
CABLE DIAMETER: Ø5-Ø10mm
Main housing
TERMINAL BLOCK 14
SOLID WIRE: 26-16 AWG / 0.129-1.31 mm²
25 STRANDED WIRE: 26-16 AWG / 0.129-1.31 mm² 15
TORQUE: 3.0 [Link] / 0.34 Nm
24
26 DISPLAY FLAT CABLE 16
17
18
19
20
21
22
ISOIL
23
7/26 MV110_IT_EN_DS053REV08_IS I N D U S T R I A
Different visualization
possibilities by simply
pressing of a key.
The manufacturer guarantees only English text available on our web site [Link]
8/26 MV110_IT_EN_DS053REV08_IS
ELECTRICAL CONNECTIONS
1 2 3 4
ELECTRODES
E1
E2
C SH
29 28 27 26
RS485
B
A
0V SH
The manufacturer guarantees only English text available on our web site [Link]
25 24 23 22 21 20
4-20 OUTPUTS
OUT2 (+) 4-20 SH
OUT1 (+) 4-20* +24V
OUT1-2 (-) 4-20 0V
* If the option HART is enable
19 18 17 16 15 14
13 12 11
COILS
C1
C2 SH
POWER SUPPLY
L (+)
N( - )
9/26 MV110_IT_EN_DS053REV08_IS
ISOIL
I N D U S T R I A
DIGITAL INPUT
10k
on/off input
15 (+)
(external power supply)
16 (-)
21 (+)
The manufacturer guarantees only English text available on our web site [Link]
16 (-)
22 (-)
DIGITAL OUTPUTS
18 OUT 1(+)
19 OUT 2(+)
10/26 MV110_IT_EN_DS053REV08_IS
ANALOG OUTPUTS
mA
24 OUT 1(+)
25 OUT 2(+)
11/26 MV110_IT_EN_DS053REV08_IS
ISOIL
I N D U S T R I A
FUNCTIONS
MENU
SENSOR
S. model = 0 1.1 Sensors model: Enter the first two characters of the serial number of the sensor
Lining = UNSPEC. 1.2 Flow sensor lining material type
S. type = FULL BORE 1.3 Type of sensor: fullbore or insertion
u. type = METRIC 1.4 Type of measure units for sensor parameter: metric or imperial
Diam mm 00025.0 1.5 Insert ND of sensor (0-2500)
KA = +00.9637
1.6 Calibration data of sensor visualized on sensor’s label
KA- = -04.4904
KZ = +0000000 1.7 Sensor coefficient KZ (zero point)
KD= +0000000 1.8 Sensor coefficient KD
Ins. position= 0 1.9 Insertion position
KP Dinamic= OFF 1.10 KP dynamic, coefficient for insertion
Ki= 01.8727 1.11 Sensor coefficient Ki
SENSOR Kp= 01.0000 1.12 Sensor coefficient Kp
KC= 1.00000 1.13 Sensor coefficient KC
[Link] = 025.0 1.14 CW volume total. decimal point position
The manufacturer guarantees only English text available on our web site [Link]
UNITS
Diam. mm 2.1 Nominal diameter measure unit
[Link] m 2.2 Cable length on separate version
[Link] METRIC 2.3 Flowrate type measure unit: metric or not metric
Pl1 unit METRIC 2.4 Pulse 1 type measure unit: metric or not metric
Pl2 unit METRIC 2.5 Pulse 2 type measure unit: metric or not metric
T+ unit METRIC 2.6 Total direct totalizer measure unit type: metric or not metric
T+ unit g 2.7 Total direct totalizer measure unit
T+ D.P 3 2.8 Total direct totalizer decimal point position
P+ unit METRIC 2.9 Partial direct totalizer measure unit type: metric or not metric
UNITS P+ unit g 2.10 Partial direct totalizer measure unit
P+ D.P 3 2.11 Partial direct totalizer decimal point position
T- unit METRIC 2.12 Total reverse totalizer measure unit type: metric or not metric
T- unit g 2.13 Total reverse totalizer measure unit
T- D.P 3 2.14 Total reverse totalizer decimal point position
P- unit METRIC 2.15 Partial reverse totalizer measure unit type: metric or not metric
P- unit g 2.16 Partial reverse totalizer measure unit
P- D.P 3 2.17 Partial reverse totalizer decimal point position
[Link] °C 2.18 Temperature measure unit
Mass units ON 2.19 Enable/disable the selection of mass units on full scale set
Sg=kg/dm3 1.0000 2.20 Specific gravity coefficient
12/26 MV110_IT_EN_DS053REV08_IS
SCALES
FS1 g/s 4908.7 3.1 Full scale flow rate 1
FS2 g/s 4908.7 3.2 Full scale flow rate 2
Pls1=g 1000.00 3.3 Pulse value on channel 1
SCALES Tpls1=ms 0050.0 3.4 Duration of the pulse generated on channel 1
Pls2=g 1000.00 3.5 Pulse value on channel 2
Tpls2=g 0050.0 3.6 Duration of the pulse generated on channel 2
Frq1=Hz 1000.00 3.7 Full scale frequency for channel 1 (0.1Hz-1000.0Hz)
Frq2=Hz 1000.00 3.8 Full scale frequency for channel 2 (0.1Hz-1000.0Hz)
The manufacturer guarantees only English text available on our web site [Link]
MEASURES
Damping SMART 4.1 Measure filter
4.2 Low flow zero threshold: 0-25% of full scale value
MEASURES Cut-off= %
DT Min.
00.1
ON 4.3 Automatic calibration verify
T1HC enable ON 4.4 Automatic change of measurement range
T1HC OFF 4.5 High immunity inputs
ALARMS
Max+ = dm3/s OFF 5.1 Maximum value alarm set for direct flow rate
Max- = dm3/s OFF 5.2 Maximum value alarm set for reverse flow rate
5.3 Minimum value alarm set for direct flow rate
ALARMS Min+= dm3/s
Min-= dm3/s
OFF
OFF 5.4 Minimum value alarm set for reverse flow rate
Hysterisis=% 03 5.5 Hysteresis threshold set for the minimum and maximum flow rate alarms
mA [Link]=% 000 5.6 Current output value in case of failure
Hz [Link]=% 000 5.7 Frequency output value in case of alarms
INPUTS
T+ reset OFF 6.1 Total direct (positive) flow totalizer reset enable
P+ reset OFF 6.2 Partial direct (positive) flow totalizer reset enable
T- reset OFF 6.3 Total reverse (negative) flow totalizer reset enable
INPUTS P- reset OFF 6.4 Partial reverse (negative) flow totalizer reset enable
Count lock OFF 6.5 Totalizer counting lock command
[Link] OFF 6.6 Measure zero lock command
Calibration OFF 6.7 Calibration external command
Range change OFF 6.8 Range change external command
13/26 MV110_IT_EN_DS053REV08_IS
ISOIL
I N D U S T R I A
OUTPUTS
Out1 PULSES+ 7.1 Output 1 functions
Out2 PULSES- 7.2 Output 2 functions
OUTPUTS Out mA1 4_22 +/- 7.3 Choice of the function and the range of current output n.1
Out mA2 4_22 +/- 7.4 Choice of the function and the range of current output n.2
A1S 4.9087 7.5 Full Scale value for analog out1
A2S 4.9087 7.6 Full Scale value for analog out2
The manufacturer guarantees only English text available on our web site [Link]
COMMUNICATION
HART pr. 05 8.1 HART packet byte preambles
HART O. C. ON 8.2 HART bus output control
8.3 Device communication address number
COMMUN. Dev. Addr
Speed=bps
001
9600 8.4 MODBUS link speed
Parity= NO 8.5 MODBUS link parity
Delay=ms 00 8.6 MODBUS reply delay
C. timeout 2 8.7 Maximum delay between chars (frames) MODBUS
DISPLAY
Language EN 9.1 Choice of the language
Contrast 5 9.2 Display contrast
[Link]=s 020 9.3 Display/keyboard inactivity time
[Link]=Hz 5 9.4 Display updating frequency: 1-2-5-10 Hz
9.5 Display function number
DISPLAY Disp. Fn.
[Link] ON
1
9.6 Display function selection lock
Part. Tot ON 9.7 Partial totalizer enable
Neg. Tot. ON 9.8 Negative totalizer enable
Net tot. ON 9.9 Net totalizer enable
[Link] ON 9.10 Time and date display enable
Quick start ON 9.11 Quick start menu visualization
14/26 MV110_IT_EN_DS053REV08_IS
DATA LOGGER
[Link] en. ON 10.1 Data logger enabling
Meas. units ON 10.2 Measure units recording enable
Field separat. ; 10.3 Field separator character
Decimal separ. . 10.4 Decimal separator character
Interv. [Link] 10.5 Sampling interval
Log T+ ON 10.6 Enable logging of total direct totalizer
Log P+ ON 10.7 Enable logging of partial direct totalizer
Log T- ON 10.8 Enable logging of total reverse totalizer
Log P- ON 10.9 Enable logging of partial reverse totalizer
DATA Log TN ON 10.10 Enable logging of total net totalizer
FUNCTIONS
T+ reset 11.1 Execute immediate reset of total direct totalizer
P+ reset 11.2 Execute immediate reset of partial direct totalizer
T- reset 11.3 Execute immediate reset of total reverse totalizer
11.4 Execute immediate reset of partial reverse totalizer
FUNCTIONS P- reset
Load Sens. f. def 11.5 Load sensor factory default
Load Conv. f. def 11.6 Load converter factory default
Save Sens. f. def 11.7 Save sensor factory default values
Save Conv. f. def 11.8 Save converter factory default values
Calibration 11.9 Execute immediate internal circuit calibration
DIAGNOSTIC
Self test 12.1 Self test diagnostic function
Test display 12.2 Function tests physical display
Sens. verify 12.3 Sensor verify diagnostic function
Flow sim. = ON 12.4 Flow rate simulation enabling
Display measures 12.5 Display internal measured value
DIAGNOSTIC Disp. Coom. Vars 12.6 Display comm. diagnostic values
Display graphs 12.7 Display measure as graphs
Gen. sens. set 12.8 Generic sensor parameters set
SD card info 12.9 Sd card status informations
Firmware info 12.10 Firmware version/revision
S/N= 999001 12.11 Board serial number
WT= 0002 : 21 : 00 : 2 2 12.12 Total working time
15/26 MV110_IT_EN_DS053REV08_IS
ISOIL
I N D U S T R I A
SYSTEM
Dayl. Saving = ON 13.1 Daylight saving time change
Time zone = h + 01.00 13.2 Localized time zone
2016/04/04-16:07 13.3 System date and time
L1 code = ******** 13.4 Access level 1 code
L2 code = ******** 13.5 Access level 2 code
L3 code = ******** 13.6 Access level 3 code
L4 code = ******** 13.7 Access level 4 code
L5 code = ******** 13.8 Access level 5 code
L6 code = ******** 13.9 Access level 6 code
Restr. Access= ON 13.10 Restricted access level
SYSTEM 010 . 011 . 012 . 013 13.11 Device IP network address
010 . 011 . 012 . 014 13.12 Client IP network address
255 . 255 . 255 . 000 13.13 Network mask
KT 0.96469 13.14 Calibration coefficient KT
KS 1.00000 13.15 Calibration coefficient KF
KR 1.00000 13.16 Calibration coefficient KR
DAC1 (°C) 13.17 DAC1 out 4mA calibration point
DAC1 (°C) 13.18 DAC1 out 20mA calibration point
DAC2 661 13.19 DAC2 out 4mA calibration point
The manufacturer guarantees only English text available on our web site [Link]
16/26 MV110_IT_EN_DS053REV08_IS
ACCURACY
e%
Typical error profile (pure indication only)
SPEED
0,33 1,64 16,4 32,8
ft/sec ft/sec ft/sec ft/sec
where:
e = ACCURACY
V = WORKING SPEED
* = 0,4 ( special )
**= 0,2(m/s) ; 0,66(ft/s) - special
Insertion sensor
See sensor DATA SHEET
17/26 MV110_IT_EN_DS053REV08_IS
ISOIL
I N D U S T R I A
SIZE Q3 Q2 Q1 R SIZE Q3 Q2 Q1 R
(**) : Calibration flowrate 1400 m3/h - as for max test rig flowrate L7
(***) Calibration flowrate 1100 m3/h - as for max test rig flowrate L6
18/26 MV110_IT_EN_DS053REV08_IS
SIZE Q3 Q2 Q1 R SIZE Q3 Q2 Q1 R
(**) : Calibration flowrate 1400 m3/h - as for max test rig flowrate L7
(***) Calibration flowrate 1100 m3/h - as for max test rig flowrate L6
19/26 MV110_IT_EN_DS053REV08_IS
ISOIL
I N D U S T R I A
SIZE Q3 Q2 Q1 R
mm inch m3/h Q3/Q1
25 1 16 0,06 0,04
32 1¼ 25 0,10 0,06
40 1½ 40 0,16 0,10
50 2 63 0,25 0,16
65 2½ 100 0,40 0,25
80 3 160 0,64 0,40
100 4 250 1,0 0,63
The manufacturer guarantees only English text available on our web site [Link]
(**) : Calibration flowrate 1400 m3/h - as for max test rig flowrate L7
(***) Calibration flowrate 1100 m3/h - as for max test rig flowrate L6
20/26 MV110_IT_EN_DS053REV08_IS
0,1°C≤T≤30°C
3 T>30°C
1
MPE
0
/%
-1
-2
The manufacturer guarantees only English text available on our web site [Link]
-3
-4
0,1°C≤T≤30°C
5
T>30°C
4
1
MPE
0
/%
-2
-3
-4
-5
-6
21/26 MV110_IT_EN_DS053REV08_IS
ISOIL
I N D U S T R I A
22/26 MV110_IT_EN_DS053REV08_IS
2,00
1,00
qi 0,1 qp qp
MPE
0,00
/%
q/qp [%]
-1,00
The manufacturer guarantees only English text available on our web site [Link]
-2,00
Ef class1
-3,00
1,0 10,0 100,0
23/26 MV110_IT_EN_DS053REV08_IS
ISOIL
I N D U S T R I A
HOW TO ORDER
CODE
EXAMPLE CODE/DESCRIPTION
Display
A Blind version (without display and programming keys, cable USB type A/USB MINI B is REQUIRED TO PROGRAMMING)
A
B Graphic LCD WSTN - B/W - back light display, point matrix 128 x 64, 8 line/16 characters and 3 programming keys (mandatory for MI001)
Painted aluminum die casting, protection rate IP68 1,5 meters under water, Compact Version, n° 1 IP 68 MIL connector for power supply (CONNECTORS
2 SUPPLIED: MALE + FEMALE)
Painted aluminum die casting, protection rate IP68 1,5 meters under water, Compact Version, Complete of n° 1 of 10 poles IP68 MIL connector (outputs
0 3 connections to be specified) and n° 1 IP 68 MIL connector for power supply (CONNECTORS SUPPLIED: MALE + FEMALE)
Painted aluminum die casting, protection rate IP68 1,5 meters under water, Separate Version, Complete of n° 1 IP 68 MIL connectors for cable from the sensor
4 and n° 1 IP 68 MIL connector for power supply (CONNECTORS SUPPLIED: MALE + FEMALE)
Painted aluminum die casting, protection rate IP68 1,5 meters under water, Separate Version, Complete of n° 1 IP 68 MIL connectors for cable from the sensor,
5 n° 1 of 10 poles IP68 MIL connector (outputs connections to be specified) and n° 1 IP 68 MIL connector for power supply (CONNECTORS SUPPLIED: MALE +
FEMALE)
Version
The manufacturer guarantees only English text available on our web site [Link]
A Compact version with sensor MS.... (liquid maximum temperature 100 °C)
A B Separate version for wall monting, complete with mounting accessories ( CABLE C014 )
Power supply
1 Power supply : 100 ... 240 VAC 44/66 Hz (NOT FOR MI001)
Power supply : 100 ... 240 VAC 44/66 Hz + 1 Rechargeable back-up Battery ( the use of battery supports only the measure UP TO 30 days; all the outputs are
4 set to OFF) (NOT FOR MI001)
Power supply : 24 … 36 VAC/VDC 0...44/66 Hz + 1 Rechargeable back-up Battery ( the use of battery supports only the measure UP TO 30 days; all the outputs
5 are set to OFF) (NOT FOR MI001)
Power supply : 12…48 VDC + 1 Rechargeable back-up Battery ( the use of battery supports only the measure UP TO 30 days; all the outputs are set to OFF)
6 (NOT FOR MI001)
1 Power supply : 100 ... 240 VAC 44/66 Hz + n° 1 SETTINGS FOR Rechargeable back-up Battery ( the Rechargeable Battery is NOT included) (NOT FOR
7 MI001)
Power supply : 24 … 36 VAC/VDC 0...44/66 Hz + n° 1 SETTINGS FOR Rechargeable back-up Battery ( the Rechargeable Battery is NOT included) (NOT FOR
8 MI001)
9 Power supply : 12…48 VDC + n° 1 SETTINGS FOR Rechargeable back-up Battery ( the Rechargeable Battery is NOT included) (NOT FOR MI001)
Power supply : 100 ... 240 VAC 44/66 Hz + 1 Pack of n° 2 SUPERCAP ( the use of it supports only the measure UP TO 3 minutes; all the outputs are set to
a OFF) Suitable for MI001
Power supply : 24 … 36 VAC/VDC 0...44/66 Hz + 1 Pack of n° 2 SUPERCAP ( the use of battery supports only the measure UP TO 3 minutes; all the outputs are
b set to OFF) Suitable for MI0011
Power supply : 12…48 VDC + 1 Pack of n° 2 SUPERCAP ( the use of battery supports only the measure UP TO 3 minutes; all the outputs are set to OFF)
C Suitable for MI001
Analogue output
A Without Analogue output
Digital Input/Output
0 With Digital Input only
24/26 MV110_IT_EN_DS053REV08_IS
Communication Gateway
A Without Gateway
E Others
Protocols
0 Without Protocol
0
1 Modbus (over RS485) requires RS485 port
Accuracy
A Standard accuracy 0,8 %
Data Logger
0 Without Data Logger
The manufacturer guarantees only English text available on our web site [Link]
0 2 MicroSD Memory 4 GB : Data Logger + RTC (Real Time Clock) + BIV (Built In Verificator)
3 MicroSD Memory 4 GB : Data Logger + RTC (Real Time Clock) + Meter Data (Real Time Converter & Sensor Data on SD Memory)
4 MicroSD Memory 4 GB : Data Logger + RTC (Real Time Clock) + BIV + Meter Data
Special Features
A NONE
MID Approval
0 NONE
3 MI-004-CLASS 1
Complete code
MV110-A0A1A0A0A0A0
example for
order
25/26 MV110_IT_EN_DS053REV08_IS
ISOIL
I N D U S T R I A
If you want to find the complete list of our distributors access at the following link:
[Link]
If you want to find the complete list of our distributors access at the following link:
[Link]
The manufacturer guarantees only English text available on our web site [Link]
Due to the constant technical development and improvement of its products, the man-
ufacturer reserves the right to make changes and/or modify the information contained
in this document without notice.