0% found this document useful (0 votes)
169 views100 pages

Planmeca ProMax Pan Ceph Calibration Manual

Uploaded by

drakon
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
169 views100 pages

Planmeca ProMax Pan Ceph Calibration Manual

Uploaded by

drakon
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd

PLANMECA

ProMax Pan/Ceph

Calibration Manual
Printed copies of this document are considered uncontrolled. 1
22811.20.Rev001 07.31.2019
The manufacturer, assembler, and importer are responsible for the safety,
reliability and performance of the unit only if:
- Installation, calibration, modification and repairs are carried out by
qualified authorized personnel.
- Electrical installations are carried out according to the appropriate
requirements such as IEC364.
- Equipment is used according to the operating instructions

Planmeca pursues a policy of continual product development. Although every


effort is made to produce up-to-date product documentation this publication
should not be regarded as an infallible guide to current specifications. We
reserve the right to make changes without prior notice.

COPYRIGHT PLANMECA USA 2008-05


Publication Revision 4 JO

-2-
Printed copies of this document are considered uncontrolled. 2
22811.20.Rev001 07.31.2018
Table of Contents

Chapter A: General Information


1. Disclaimer 7
2. Required Tools 9
2.1. Calibration Tools 9
2.2. Hand Tools 10

Chapter B: Removing Covers


1. Upper C-Arm 11
2. Tube head 12
3. Shoulder 12
4. Lower C-Arm 13
5. Patient Positioning 13
6. Fixed sensor Holder 14

Chapter C: Computer Connections


1. Ethernet 17
2. Didapi Configuration 17

Chapter D: Beam Alignment


1. Aligning Beam 20
2. Primary Collimator (61.0) 21
2.1. Pan mode X-Collimator (61.1) 23
2.1.a X-pan collimator position 24
2.1.a.i Too Narrow 25
2.1.a.ii Too Wide 26
2.2. Pan mode Y-Collimator (61.2) 26
2.2.a Too High 26
2.2.b Too Low 27
2.2.c Angled 27
2.2.c.i Angled Left/Right 30

Chapter E: Beam Check


1. Setting up Beam Check 31
2. Computer Setup 33

Chapter F: Calibration
1. Dimax 3 Tool 35
1.1 Turning Radiation ON/OFF 36
1.1.a OFF 36
1.1.b ON 37
1.2 kV/mA setting for Panoramic 37
1.3 Changing Binnings 37

-3-
Printed copies of this document are considered uncontrolled. 3
22811.20.Rev001 07.31.2018
Chapter G: Ball Phantom
1. Setting up for a Ball Phantom 39
2. Setting up the Computer 41
2.1 Calibrating / measuring image 41
2.1.a. Center Ball 43
2.1.a.i Too Thin 44
2.1.a.ii Too Fat 46
2.1.b 10th Ball 48
2.1.b.i Too Left 48
2.1.b.ii Too Right 49
2.1.b.iiiMoving the Table 49
2.1.c Shadow Ball 50
2.1.c.i Too Left/Right 53
3. Position Sensor Calibration (Pots) (64.0) 54
3.1 Aligning Pots 56
3.1.a Shoulder 56
3.1.b Elbow 57
3.1.c C-Arm 58
4. Patient Positioning Lights Alignment 59
4.1 Focal Layer Light Beam 61
4.2 Frankfort Light Beam 62
4.3 Midsagittal Light Beam 62

Chapter H: Cephalometric
1. Getting into Cephalometric Mode 63
2. Removing Cephalometric Covers 64
2.1 Top Covers 64
2.2 Secondary Collimator Covers 66
3. Beam Alignment 67
3.1 Ceph Mode C-arm Rotation 68
3.1.a Ceph Mode Beam Right Position 70
3.1.a.i Too Left/Right 70
3.1.b Ceph Mode C-arm Rotation 72
3.1.b.i Too Left/Right 72
3.2 Ceph Mode Beam Position 74
3.2.a Too Left/Right 75
3.3 Ceph y-collimator calibration 76
3.3.a Too High 77
3.3.b Too Low 78
4. Ceph Head Leveling 78
4.1 Too Left 81
4.2 Too Right 82
4.3 Too Low 83
4.4 Too High 83
5. Secondary Collimator (Beaver Tail) 84
5.1 Too Left 85
5.2 Too Right 87
5.3 Too Low 88
5.4 Too High 89
-4-
Printed copies of this document are considered uncontrolled. 4
22811.20.Rev001 07.31.2018
5.5 Angled 90
5.5.a Left 90
5.5.b Right 91
6. Beam Check 92
6.1 Setting up Beam Check 92
6.2 Computer Setup 94
7. Calibrating 95
7.1 kV/mA fo Cephalometric 95
7.2 Computer Setup 97
7.3 Dimax3 Tool 97
7.3.a Not Together 99
7.3.a.i Raising Up 99

-5-
Printed copies of this document are considered uncontrolled. 5
22811.20.Rev001 07.31.2018
-6-
Printed copies of this document are considered uncontrolled. 6
22811.20.Rev001 07.31.2018
Chapter

A General Information

1 DISCLAIMER
This manual contains the information required to setup and calibrate the
Planmeca ProMax Panoramic with Cephalometric X-ray unit for Digital.

WARNING
Protect yourself from radiation when you are checking the beam
alignment and calibrating. Since radiation safety requirements vary from
state to state, country to country, it is the responsibility of the installer to
ensure that the correct precautions are observed.

1.1 The display values shown in this guide are only examples
and should be interpreted as recommended values unless
otherwise specified.

-7-
Printed copies of this document are considered uncontrolled. 7
22811.20.Rev001 07.31.2018
-8-
Printed copies of this document are considered uncontrolled. 8
22811.20.Rev001 07.31.2018
2 REQUIRED TOOLS
2.1. Calibration Tools

 Ball phantom (part number 10004011) used for checking the position of the
Patient positioning mechanism and the Patient positioning lights. (Figure 1)

Figure 1

 Frankfort plane alignment tool (part number 50977) used with the ball
phantom for checking the position of the Frankfort plane light. (Figure 2)

Figure 2

 Beam alignment tool, fluorescent screen, (part number 50972) used for
checking the position of the x-ray beam. (Figure 3)

Figure 3

-9-
Printed copies of this document are considered uncontrolled. 9
22811.20.Rev001 07.31.2018
 Sensor head alignment tool (part number 10002699) used for checking the
beam alignment. (Figure 4)

Figure 4
 Cephalostat calibration tool, hatchet, (part number 10004246). For
calibrating the digital cephalostat. (Figure 5)

Figure 5
 Left and Right ear posts (part number 651003, 651005) used for aligning the
ball and ring on the cephalometric arm. (Figure 6)

Figure 6

2.2 Hand Tools


 1.5mm allen wrench
 2mm allen wrench
 2.5mm allen wrench
 3mm allen wrench
 4mm allen wrench

- 10 -
Printed copies of this document are considered uncontrolled. 10
22811.20.Rev001 07.31.2018
Chapter

B Removing Covers
1 Upper C-arm
Unscrew the eight 2.5mm allen screws on top. (Figure 7)

Figure 7

Twist the cover away from the column and off. (Figure 8)

Figure 8

- 11 -
Printed copies of this document are considered uncontrolled. 11
22811.20.Rev001 07.31.2018
2 Tube head
Unscrew the top of the front tubehead cover with a 3mm allen (Figure 9a-b) and
unscrew the bottom with a 2.5mm allen (Figure 9c) and remove the front cover.

Top What is unscrewed

Figure 9a Figure 9b
Bottom

Figure 9c

3 Shoulder
Pull up on the shoulder cover. (Figure 10)

Figure 10
- 12 -
Printed copies of this document are considered uncontrolled. 12
22811.20.Rev001 07.31.2018
4 Lower C-Arm
Un-screw the four screws with a 2.5 allen. (Figure 11)

Figure 11

5 Patient Positioning
Un-screw the five screws with a 2.5 allen. (Figure 12)

Figure 12

- 13 -
Printed copies of this document are considered uncontrolled. 13
22811.20.Rev001 07.31.2018
6 Fixed Sensor Holder
Pull back on the inner cover of the sensor holder. (Figure 13)

Figure 13

Unscrew the two screws on the top of the rear cover of the holder with a 2.5mm allen.
(Figure 14)

Figure 14

- 14 -
Printed copies of this document are considered uncontrolled. 14
22811.20.Rev001 07.31.2018
Unscrew the two screws on the inside of the rear cover with a 2.5mm allen. (Figure
15)

Figure 15

Take off the rear cover. (Figure 16)

Figure 16

Now turn the unit on and go to Primary Collimator.

- 15 -
Printed copies of this document are considered uncontrolled. 15
22811.20.Rev001 07.31.2018
- 16 -
Printed copies of this document are considered uncontrolled. 16
22811.20.Rev001 07.31.2018
Chapter

C Computer Connections
1 Ethernet
Connect a crossover cable to the Ethernet port on the back of the ProMax. (Figure 17)

Figure 17

Connect the other end of the crossover cable into the Ethernet port on the back of the
computer.
NOTE: If using a PCI card then follow the above instructions but instead use a PCI Interface Cable.

2 Didapi Configuration
Click Start , All Programs, Planmeca and then Didapi Configuration (Figure 18)

Figure 18

Click on Ethernet Interface.

The default ip address is: 192.168.0.130

Click on “ProMax Open device config file” and type in:

Localhost=(ip address):5000

Save the file, click “Refresh” and close Didapi Configuration

- 17 -
Printed copies of this document are considered uncontrolled. 17
22811.20.Rev001 07.31.2018
- 18 -
Printed copies of this document are considered uncontrolled. 18
22811.20.Rev001 07.31.2018
Chapter

D Beam Alignment
Unscrew the two locking screws from the attachment pins on the sensor holder with a
3mm allen and remove the screws and washers.(Figure 19)

Figure 19

Loosen the connector plate screw on the bottom of the holder with a 2.5mm allen.
(Figure 20a-b)

Figure 20a Figure 20b

- 19 -
Printed copies of this document are considered uncontrolled. 19
22811.20.Rev001 07.31.2018
1 Aligning the beam
Place a beam alignment tool into the sensor holder. (Figure 21)

Figure 21

Attach a fluorescent screen to the beam alignment tool. (Figure 22)

Figure 22

- 20 -
Printed copies of this document are considered uncontrolled. 20
22811.20.Rev001 07.31.2018
2 Primary Collimator
Press the i button in the lower left hand corner. (Figure 23)

Figure 23

Press Technical Calibration, 60.0. (Figure 24)

Figure 24

- 21 -
Printed copies of this document are considered uncontrolled. 21
22811.20.Rev001 07.31.2018
Type in 1701 as the password. (Figure 25)

Figure 25

Press Primary Collimator, 61.0. (Figure 26)

Figure 26

- 22 -
Printed copies of this document are considered uncontrolled. 22
22811.20.Rev001 07.31.2018
2.1 Pan mode X-Collimator (61.1) (Figure 27)

Figure 27

Now press the exposure button to see where the beam is hitting the screen.

The beam image should appear within the borders of the rectangle markings
on the screen. (Figure 28)

U P
Figure 28

NOTE: The beam will need to be half of the distance of the box.

- 23 -
Printed copies of this document are considered uncontrolled. 23
22811.20.Rev001 07.31.2018
2.1.a X pan collimator position
If the beam is too left, press the right button (Figure 29)

Figure 29

If the beam is too right, press the left button (Figure 30)

Figure 30

- 24 -
Printed copies of this document are considered uncontrolled. 24
22811.20.Rev001 07.31.2018
Press Test and take another exposure. (Figure 31)

Figure 31

2.1.a.i Too Narrow

Press the X-pan collimator width to the right (Figure 32)

Figure 32

Press Test and take another exposure.

- 25 -
Printed copies of this document are considered uncontrolled. 25
22811.20.Rev001 07.31.2018
2.1.a.ii Too Wide
Press the X-pan collimator width to the left (Figure 33)

Figure 33
Press Test and take another exposure.
If aligned then press Done and move onto the Y
Collimator.

2.2 Pan mode Y Collimator (61.2)


Now check the top and bottom of the beam.
2.2.a Too High
Press the Y-pan collimator top limit button down. (Figure 34)

Figure 34

- 26 -
Printed copies of this document are considered uncontrolled. 26
22811.20.Rev001 07.31.2018
Press Test and take another exposure.
2.2.b Too Low
Press the Y-pan collimator bottom limit button up. (Figure 35)

Figure 35
Press Test and take another exposure.
If aligned then press Done and move onto Ball Phantom.
2.2.c Angled
Press the i button in the lower left hand corner. (Figure 36)

Figure 36
- 27 -
Printed copies of this document are considered uncontrolled. 27
22811.20.Rev001 07.31.2018
Press Technical Calibration, 60.0. (Figure 37)

Figure 37

Press Primary Collimator, 61.0. (Figure 38)

Figure 38

- 28 -
Printed copies of this document are considered uncontrolled. 28
22811.20.Rev001 07.31.2018
Press down once and select Collimator assembly screws visible,
61.9. (Figure 39)

Figure 39

NOTE: 61.9 is available in software versions 1.19.7 or higher

Press Drive and the collimator will move to expose the screws. (Figure
40)

Figure 40

- 29 -
Printed copies of this document are considered uncontrolled. 29
22811.20.Rev001 07.31.2018
The collimator will move to expose one screw then push the lower blade up
and expose the second screw. (Figure 41)

Figure 41
NOTE: This allows the two screws that adjust the angle to be visible.

2.2.c.i Angled Left/Right


Loosen the two hidden screws with a 2.5 mm allen at # 1.
(Figure 42)

Figure 42
Twist collimator to the right if angled Right to Left
Twist collimator to the left if angled Left to Right
Tighten down the screws and take another exposure.
Go back into Pan X-Collimator and Y-Collimator and press Test in
both allowing the collimator to go to the home position.
Take another exposure. Press Done until you get back to the Technical
Calibration Screen.

- 30 -
Printed copies of this document are considered uncontrolled. 30
22811.20.Rev001 07.31.2018
Chapter

E Beam Check
1 Setting up Beam Check
Press the i button. (Figure 43)

Figure 43

Press Special Functions (40.0) (Figure 44)

Figure 44

- 31 -
Printed copies of this document are considered uncontrolled. 31
22811.20.Rev001 07.31.2018
Press Beam Check (47.0) (Figure 45)

Figure 45

Press Panoramic (47.1) (Figure 46)

Figure 46

- 32 -
Printed copies of this document are considered uncontrolled. 32
22811.20.Rev001 07.31.2018
2 Computer Setup
NOTE: Make sure the sensor is in the sensor holder.

Click Start then All Programs then Planmeca then Beam Check. (Figure 47)

Figure 47

Click on ProMax Pan. (Figure 48)

Figure 48

Click on Take the Image

Press Ready on the touch screen (Figure 49) and then take an exposure.

Figure 49
- 33 -
Printed copies of this document are considered uncontrolled. 33
22811.20.Rev001 07.31.2018
Image should be centered on the top, bottom, left, and right. (Figure 50)

Figure 50

NOTE: The smaller image to the left is the entire sensor image.

Refer to pgs. 20-30 if not aligned properly.

- 34 -
Printed copies of this document are considered uncontrolled. 34
22811.20.Rev001 07.31.2018
Chapter

F Calibration
Press Done and Exit

Close Beam Check and open Dimax3 Tool

1 Dimax3 Tool
Click on Start, All Programs, Planmeca then Dimax3 Tool. (Figure 51)

Figure 51

Click on Calibrate and select Pan. (Figure 52)

Figure 52

Follow the instructions on the computer screen.

- 35 -
Printed copies of this document are considered uncontrolled. 35
22811.20.Rev001 07.31.2018
1.1 Turning Radiation On/Off
1.1.a OFF
Press the smile in the middle right of the touch screen. (Figure 53)

Figure 53

Press all the arrows at the bottom and side, turning each from white to
red. (Figure 54)

Figure 54

- 36 -
Printed copies of this document are considered uncontrolled. 36
22811.20.Rev001 07.31.2018
1.1.b ON
Press the smile again. (See Figure 53) Then Press, all the arrows at the
bottom and side, turning each from red to white.

1.2 kV/mA setting for Panoramic

Binnings kV mA
4x4 54 1
3x3 60 1
2x2 60 2

1.3 Changing Binnings


Click Settings and select Binnings then select the appropriate binning.
(Figure 55)

Figure 55

Press Ready on the touch screen. Take an Exposure.

NOTE: Make sure that the correct kV/mA are being used for each binning.

The Image will come out as showing a grey box with a black stripe in the
center. (Figure 56)

Figure 56

NOTE: There is no need to close and re-open Dimax3 Tool before selecting other binnings.
Follow 1.5.c for checking the other binnings.

The calibration files are automatically saved. Close Dimax3 Tool.

- 37 -
Printed copies of this document are considered uncontrolled. 37
22811.20.Rev001 07.31.2018
- 38 -
Printed copies of this document are considered uncontrolled. 38
22811.20.Rev001 07.31.2018
Chapter

G Ball Phantom
1 Setting up for ball phantom
Place the ball phantom in the patient positioning mechanism. (Figure 57)

Figure 57

NOTE: Layer must be at 0mm before doing a ball phantom. If layer is not then adjust the thumb wheel till
layer is at 0mm. (Figure 58)

Figure 58

- 39 -
Printed copies of this document are considered uncontrolled. 39
22811.20.Rev001 07.31.2018
Make sure the sensor is in the sensor holder. (Figure 59)

Figure 59

Setting the kV/mA

Press the kV/mA at the top left of the touch screen. (Figure 60)

Figure 60

- 40 -
Printed copies of this document are considered uncontrolled. 40
22811.20.Rev001 07.31.2018
Select the 60kV / 4mA then click OK. (Figure 61)

Figure 61

2 Setting up the computer


Click on Start, All Programs, Planmeca, Dimaxis 4.x.x. (Figure 62)

Figure 62

Click OK then click OK again. Select New from the bottom of the Select Image
screen.

Press Ready on the touch screen and take an exposure.

Click OK on the computer screen to evaluate the image for alignment.

2.1 Calibrating / measuring image


Click the measurement toolbar at the top. (Figure 63)

Figure 63

- 41 -
Printed copies of this document are considered uncontrolled. 41
22811.20.Rev001 07.31.2018
Click on the first icon, CAL, and calibrate the Center Ball from top to bottom.
(Figure 64)

Figure 64

Enter the number 7 when asked to input a distance. (Figure 65)

Figure 65

There will be an image of 23 balls on the films; one Center Ball and 11 balls
to the left and right. (Figure 66)

Figure 66

NOTE: Make sure Center Ball is round before moving onto the 10th ball distances.

- 42 -
Printed copies of this document are considered uncontrolled. 42
22811.20.Rev001 07.31.2018
2.1.a Center Ball

NOTE: Center ball will be above the midsagittal bar of the ball phantom. (Figure 67)

Figure 67

Click the second icon now and measure the top to bottom and left to
right. (Figure 68)

Figure 68

The ball should be 7mm top to bottom and left to right. (Figure 69)

Figure 69

NOTE: If ball phantom is off, do a Position Sensor Calibration (Pots) on pgs. 49-54.

- 43 -
Printed copies of this document are considered uncontrolled. 43
22811.20.Rev001 07.31.2018
2.1.a.i Too thin

If the ball is too thin (Figure 70) then adjust worm gear in Pots.

Figure 70

Press i then select Technical Calibration (60.0) then select Position


Sensor Calibration. (64.0). (Figure 71)

Figure 71

Press Position Sensor (64.1) (Figure 72)

Figure 72

Press Check and write down what the shoulder position number is.

- 44 -
Printed copies of this document are considered uncontrolled. 44
22811.20.Rev001 07.31.2018
Put the locking pin into the c-arm/elbow. (Figure 73a-b)

Figure 73a Figure 73b

Press Lock (Figure 74)

Figure 74

Press Release then twist the worm gear away from the column. (Figure 75)

Figure 75
- 45 -
Printed copies of this document are considered uncontrolled. 45
22811.20.Rev001 07.31.2018
NOTE: The new shoulder number should be higher than older number.

Press Lock and take the pin out and then press Done. Take another exposure
and check ball size again. If not repeat until correct size.
2.1.a.ii Too Fat
If the ball is too fat (Figure 76) then adjust in Pots.

Figure 76

Press i then select Technical Calibration (60.0) then select Position Sensor
Calibration. (64.0). (Figure 77)

Figure 77

Press Position Sensor (64.1) (Figure 78)

Figure 78
- 46 -
Printed copies of this document are considered uncontrolled. 46
22811.20.Rev001 07.31.2018
Press Check and write down what the shoulder position number is.
Put the locking pin into the c-arm/elbow. (Figure 79a-b)

Figure 79a Figure 79b


Press Lock (Figure 80).

Figure 80

Press Release then twist the worm gear towards the column. (Figure 81)

Figure 81
- 47 -
Printed copies of this document are considered uncontrolled. 47
22811.20.Rev001 07.31.2018
NOTE: The new shoulder number should be lower than old number.

Press Lock and take the pin out and then press Done. Take another exposure
and check ball size again. If not repeat until correct size.

2.1.b 10th Ball


Measure from the center of the Center Ball to the center on the 10th Ball on
the right then measure from the center of the Center Ball to the center of 10th
Ball on the left. (Figure 82)

Figure 82

NOTE: From the Center Ball, the distance left or right to the 10th Ball will always need to be the
same.

2.1.b.i Too Left

If the distance from the Center Ball to the left is more than the
right, then move the table away from the column. (Figure 83).

Figure 83

- 48 -
Printed copies of this document are considered uncontrolled. 48
22811.20.Rev001 07.31.2018
2.1.b.ii Too Right

If the distance from the Center Ball to the right is more than the
left, then move the table toward the column. (Figure 84)

Figure 84

2.1.b.iii Moving the Table

NOTE: Call Planmeca Technical Support before proceeding.

Loosen the four screws on the table with a 4mm allen. (Figure
85)

Figure 85

NOTE: Screws might be a Torx T25 instead of a 4mm allen

Physically push or pull the table to the left or right.

- 49 -
Printed copies of this document are considered uncontrolled. 49
22811.20.Rev001 07.31.2018
NOTE: Best way to adjust the table is to face the column with the table in the center of the body
and gently push or pull depending on what is needed. (Figure 86a-b)

Figure 86a Figure 86b

2.1.c Shadow Ball

NOTE: The oval above the Center Ball is the shadow ball. (Figure 87)

Figure 87

If the shadow ball is too left or too right, then go into Pan mode X-Collimator.

- 50 -
Printed copies of this document are considered uncontrolled. 50
22811.20.Rev001 07.31.2018
Press the i button. (Figure 88)

Figure 88

Press Technical Calibration (Figure 89)

Figure 89

- 51 -
Printed copies of this document are considered uncontrolled. 51
22811.20.Rev001 07.31.2018
Press Primary Collimator (Figure 90)

Figure 90

Press Pan mode X-collimator calibration. (Figure 91)

Figure 91

- 52 -
Printed copies of this document are considered uncontrolled. 52
22811.20.Rev001 07.31.2018
2.1.c.i Too Left/ Too Right
If the shadow ball is too left press the X-Pan collimator pos left
button. (Figure 92)

Figure 92

Press Test then Done then take another exposure.


If the shadow ball is too right then press the X-Pan collimator
post right button. (Figure 93)

Figure 93
Press Test then Done then take another exposure.
- 53 -
Printed copies of this document are considered uncontrolled. 53
22811.20.Rev001 07.31.2018
3 Position Sensor Calibration (Pots)
NOTE: If the ball phantom image is good, then skip this section.

Press i then select Technical Calibration (60.0) then select Position Sensor
Calibration. (64.0). (Figure 94)

Figure 94

Press Position Sensor (64.1) (Figure 95)

Figure 95

- 54 -
Printed copies of this document are considered uncontrolled. 54
22811.20.Rev001 07.31.2018
Press Check, Lock and then Release

Put the locking pins into the elbow/c-arm and in the shoulder (Figure 96a-c)

Figure 96a Figure 96b

Figure 96c

Press Lock (Figure 97)

Figure 97

If the triangles are all green, then remove the pins and press Done
- 55 -
Printed copies of this document are considered uncontrolled. 55
22811.20.Rev001 07.31.2018
If any of the pots are red or not centered, align the according pot before removing the
pins. (Figure 98)

Figure 98

3.1 Aligning Pots


3.1.a Shoulder
Unscrew the set screw with a 1.5mm allen to get the sensor coupling
off of the spindle. (Figure 99)

Figure 99

- 56 -
Printed copies of this document are considered uncontrolled. 56
22811.20.Rev001 07.31.2018
Watch the numbers on the touch screen and rotate spindle until in the
Green and center. (Figure 100)

Figure 100

Put sensor coupling back onto spindle and re-tighten set screw. Re-
check still in the green and center.

When aligned take out the pins then press Done.

3.1.b Elbow
Unscrew the set screw with a 2mm allen to get the sensor holder off of
the cable guide. (Figure 101)

Figure 101

- 57 -
Printed copies of this document are considered uncontrolled. 57
22811.20.Rev001 07.31.2018
Watch the numbers on the touch screen and rotate spindle until in the
Green and center. (Figure 102)

Figure 102

When aligned take out the pins then press Done.

1.6.d.i.*** C-Arm
Unscrew the set screw with a 2mm allen to get the sensor coupling off
of the spindle. (Figure 103)

Figure 103

- 58 -
Printed copies of this document are considered uncontrolled. 58
22811.20.Rev001 07.31.2018
Watch the numbers on the touch screen and rotate spindle until in the
Green and center. (Figure 104)

Figure 104

When aligned take out the pins then press Done.

4 Patient Positioning Lights Alignment


After aligning the Ball Phantom, turn the patient positioning lights on by pressing the
temple support button. (Figure 105)

Figure 105

- 59 -
Printed copies of this document are considered uncontrolled. 59
22811.20.Rev001 07.31.2018
Place the Ball Phantom into the patient positioning mechanism with the Frankfort
plane alignment tool on top of the Ball Phantom. (Figure 106)

Figure 106

NOTE: Make sure the layer light is on the line of the ball phantom then zero out the layer light.

Press i then press Technical Calibration (60.0) (Figure 107)

Figure 107

- 60 -
Printed copies of this document are considered uncontrolled. 60
22811.20.Rev001 07.31.2018
Press Layer laser calibration (66.0) (Figure 108)

Figure 108

Press Done and now layer is zeroed out.

4.1 Focal Layer Light Beam


If the focal layer light is not straight on the Ball Phantom, then finger adjust
the laser until it lines up. (Figure 109)

Figure 109

- 61 -
Printed copies of this document are considered uncontrolled. 61
22811.20.Rev001 07.31.2018
4.2 Frankfort Light Beam
If the Frankfort light is not straight on the Frankfort plane tool, then adjust the
laser. (See Figure 109)

4.3 Midsagittal Light Beam

If the Midsagittal light is not lined up with the center line on the Ball
phantom, then finger adjust the laser until it lines up. (See Figure 109)

- 62 -
Printed copies of this document are considered uncontrolled. 62
22811.20.Rev001 07.31.2018
Chapter d

H Cephalometric
1 Getting into Cephalometric Mode
Press Pan (Figure 110)

Figure 110

Press Ceph (Figure 111)

Figure 111

- 63 -
Printed copies of this document are considered uncontrolled. 63
22811.20.Rev001 07.31.2018
Press Go Ceph on the bottom in the middle. (Figure 112)

Figure 112

2 Removing Cephalometric Covers

2.1 Top covers


.
NOTE: All eight hidden screws are 3mm allens

Screws 1-8

- 64 -
Printed copies of this document are considered uncontrolled. 64
22811.20.Rev001 07.31.2018
Rotate the head support to 45º position towards the front of the column and
unscrew #1. (Figure 113)

Figure 113

Rotate the head support to the 0º position and unscrew #2. (Figure 114)

Figure 114

Rotate the head support to the 45º position away from the front of the column
and unscrew #3. (Figure 115)

Figure 115

- 65 -
Printed copies of this document are considered uncontrolled. 65
22811.20.Rev001 07.31.2018
Rotate the head support to the 135º position and unscrew #4 and #5. (Figure
116)

Figure 116

Rotate the head support to the 150º position and unscrew #6 and #7. (Figure
117)

Figure 117

Unscrew the inner attachment screw and take off the covers. (Figure 118)

Figure 118

- 66 -
Printed copies of this document are considered uncontrolled. 66
22811.20.Rev001 07.31.2018
2.2 Secondary Collimator Covers
Unscrew the 2mm screw on the bottom of the collimator cover. (Figure 119.2)

Figure 119

Unscrew the two 4mm screws holding the cover box to the top of the
collimator. (Figure 119.1)

3 Beam Alignment
Attach a fluorescent screen to the middle of the beam alignment tool. (Figure 120)

Figure 120

- 67 -
Printed copies of this document are considered uncontrolled. 67
22811.20.Rev001 07.31.2018
3.1 Ceph mode C-arm rotation (61.5)
Press i (Figure 121)

Figure 121

Press Technical calibration (60.0) (Figure 122)

Figure 122

- 68 -
Printed copies of this document are considered uncontrolled. 68
22811.20.Rev001 07.31.2018
Press Primary collimator (61.0) (Figure 123)

Figure 123

Press Ceph mode c-arm rotation (61.5) (Figure 124)

Figure 124

- 69 -
Printed copies of this document are considered uncontrolled. 69
22811.20.Rev001 07.31.2018
3.1.a Ceph Mode Beam Right Position
Press Test on the bottom and take an exposure. Beam should be hitting
from the center to the right on the screen. (Figure 125)

Figure 125

If not aligned, adjust Ceph mode C-arm rotation.


3.1.a.i Too Left/Right
If the beam is too left, move the Ceph mode beam right
position to the right. (Figure 126)

Figure 126

- 70 -
Printed copies of this document are considered uncontrolled. 70
22811.20.Rev001 07.31.2018
Press Test on the top and take another exposure. (Figure 127)

Figure 128

If the beam is too right, move the Ceph mode beam right
position to the left. (Figure 129)

Figure 129

Press Test on the bottom and take another exposure.

If aligned then press Done move onto Ceph mode C-arm


rotation.

- 71 -
Printed copies of this document are considered uncontrolled. 71
22811.20.Rev001 07.31.2018
3.1.b Ceph Mode C-arm Rotation
Press Test on the bottom and take an exposure. Beam should be hitting
from the center to the right on the screen. (Figure 130)

Figure 131

If not aligned, adjust Ceph mode C-arm rotation.


3.1.b.i Too Left/Right
If the beam is too left, move the Ceph mode C-arm rotation
to the right. (Figure 132)

Figure 132

- 72 -
Printed copies of this document are considered uncontrolled. 72
22811.20.Rev001 07.31.2018
Press Test on the bottom and take another exposure. (Figure
133)

Figure 133

If the beam is too right, move the Ceph mode C-arm rotation
to the left. (Figure 134)

Figure 134

Press Test on the bottom and take another exposure.

If aligned then press Done move onto Ceph mode beam


position.

- 73 -
Printed copies of this document are considered uncontrolled. 73
22811.20.Rev001 07.31.2018
3.2 Ceph mode beam position (61.6)
Press Ceph mode beam position. (Figure 135)

Figure 135

Press Test and take an exposure. Beam should cover the center of the
fluorescent screen with a bit of overlap on either side of the center lines.
(Figure 136)

Figure 136

Adjust if the beam is off.

- 74 -
Printed copies of this document are considered uncontrolled. 74
22811.20.Rev001 07.31.2018
3.2.a Too Left/Right
If the beam is too left, move the x-right limit to the right. (Figure 137)

Figure 137

Press Test and take another exposure.

If the beam is too right, move the x-right limit to the left. (Figure 138)

Figure 138

Press Test and take another exposure. If the beam is aligned then press
Done and move onto the Ceph y-collimator.

- 75 -
Printed copies of this document are considered uncontrolled. 75
22811.20.Rev001 07.31.2018
3.3 Ceph y-collimator calibration (61.4)
Attach the two fluorescent screens to the beam alignment tool. (Figure 139a-
b)

Figure 139a Figure 139b

Press Ceph y-collimator calibration (61.4) (Figure 140)

Figure 140

Press Test and take an exposure.

- 76 -
Printed copies of this document are considered uncontrolled. 76
22811.20.Rev001 07.31.2018
Image should come out as covering top to bottom on the fluorescent screens.
(Figure 141)

Figure 141

3.3.a Too High


Press the y-top limit down and y-bottom limit down. (Figure 142)

Figure 142

Press Test and take another exposure. If the beam is aligned then press
Done and Move onto the Secondary collimator (beaver tail).

- 77 -
Printed copies of this document are considered uncontrolled. 77
22811.20.Rev001 07.31.2018
3.3.b Too Low
Press the y-top limit up and y-bottom limit up. (Figure 143)

Figure 143

Press Test and take another exposure. If the beam is aligned then press
Done. Move onto the Secondary collimator (beaver tail).

4 Ceph Head Leveling


Put the left and right ear posts into their respective holders. (Figure 144)

Figure 144

- 78 -
Printed copies of this document are considered uncontrolled. 78
22811.20.Rev001 07.31.2018
Click on Start then Run. (Figure 145)

Figure 145

Type in pmsample then click OK. (Figure 146)

Figure 146

Uncheck Use calibration. (Figure 167)

Figure 167

- 79 -
Printed copies of this document are considered uncontrolled. 79
22811.20.Rev001 07.31.2018
Click on Cephalometric. (Figure 168)

Figure 168
Set the kV / mA to 66 kV and 10 mA
Press Ready on the GUI. (Figure 169)

Figure 169

- 80 -
Printed copies of this document are considered uncontrolled. 80
22811.20.Rev001 07.31.2018
Ball should be within ring. (Figure 170)

Figure 170

4.1 Too Left


If the ball is too left then adjust arm at points in Figure 171.

Figure 171

Loosen #4 a quarter of a turn then loosen up #1 and #2.

Turn #3 toward the column and retighten #1 and #2. Also retighten #4 a
quarter of a turn.

Take another exposure. If aligned, re-check Beam check.

- 81 -
Printed copies of this document are considered uncontrolled. 81
22811.20.Rev001 07.31.2018
4.2 Too Right
If the ball is too right then adjust arm at points in Figure 172.

Figure 173

Loosen #4 a quarter of a turn then loosen up #1 and #2.

Turn #3 away from the column and retighten #1 and #2. Also retighten #4 a
quarter of a turn.

Take another exposure. If aligned, re-check Beam check.

- 82 -
Printed copies of this document are considered uncontrolled. 82
22811.20.Rev001 07.31.2018
4.3 Too Low
If the ball is too low then adjust arm at points in Figure 174.

Figure 174

Loosen #1 a quarter of a turn then loosen up #3 and tighten down #2.


Take another exposure. If aligned then tighten #1 a quarter turn. If aligned, re-
check Beam check.

4.4 Too High


If the ball is too high then adjust arm at points in Figure 175.

Figure 175

Loosen #1 a quarter of a turn then loosen up #2 and tighten down #3.

Take another exposure. If aligned then tighten #1 a quarter turn. If aligned, re-
check Beam check

- 83 -
Printed copies of this document are considered uncontrolled. 83
22811.20.Rev001 07.31.2018
5 Secondary Collimator (Beaver Tail)
Unscrew the three 3mm allen screws. (Figure 176a-b)

Figure 176a Figure 176b


Attach the collimator and screw in the three 3mm allens. (Figure 177a-b)

Figure 177a Figure 177b


Press i then Technical calibration then Primary Collimator and select “Ceph mode
beam position” (61.6)
Press Test and take an exposure. (Figure 178)

Figure 178

- 84 -
Printed copies of this document are considered uncontrolled. 84
22811.20.Rev001 07.31.2018
The beam should be in the boxes of the fluorescent screens. (Figure 179)

Figure 179

5.1 Too Left


Loosen the two 3mm screws closest to the column. (Figure 180)

Figure 180

- 85 -
Printed copies of this document are considered uncontrolled. 85
22811.20.Rev001 07.31.2018
Loosen the nut on the left and right locking screws. (Figure 181)

Figure 181

Loosen the right 3mm locking screw and tighten the 3mm left locking screw.
(Figure 182)

Figure 182

Take another exposure. If aligned tighten the right nut and tighten the two
screws on the bottom.

Move onto Beam Check.

- 86 -
Printed copies of this document are considered uncontrolled. 86
22811.20.Rev001 07.31.2018
5.2 Too Right
Loosen the two 3mm screws closest to the column. (Figure 183)

Figure 183

Loosen the nut on the left and right locking screws. (Figure 184)

Figure 184

- 87 -
Printed copies of this document are considered uncontrolled. 87
22811.20.Rev001 07.31.2018
Loosen the left 3mm locking screw and tighten the 3mm right locking screw.
(Figure 185)

Figure 185

Take another exposure. If aligned tighten the left nut and tighten the two
screws on the bottom.
Move onto Beam Check.

5.3 Too Low


Loosen the two 3mm allen screws on the inside of the collimator. (Figure 186)

Figure 186

- 88 -
Printed copies of this document are considered uncontrolled. 88
22811.20.Rev001 07.31.2018
Tighten the 3mm locking screw on the outside, closest to the column. (Figure
187)

Figure 187

Take another exposure. If aligned then retighten the inner screws and move
onto Beam Check.

5.4 Too High


Loosen the two 3mm allen screws on the inside of the collimator. (Figure 188)

Figure 188

- 89 -
Printed copies of this document are considered uncontrolled. 89
22811.20.Rev001 07.31.2018
Loosen the 3mm locking screw on the outside, closet to the column. (Figure
189)

Figure 189

Take another exposure. If aligned then re-tighten the inner screws and move
onto Beam Check.

5.5 Angled
5.5.a Left
If angled left, then loosen the nuts and loosen the right locking screw
and tighten the left locking screw. (Figure 190)

Figure 190

Take another exposure. If aligned then re-tighten the nuts and move
onto Beam Check.

- 90 -
Printed copies of this document are considered uncontrolled. 90
22811.20.Rev001 07.31.2018
5.5.b Right
If angled right, then loosen the nuts and loosen the left locking screw
and tighten the right locking screw. (Figure 191)

Figure 192

Take another exposure. If aligned then re-tighten the nuts and move
onto Beam Check.

- 91 -
Printed copies of this document are considered uncontrolled. 91
22811.20.Rev001 07.31.2018
6 Beam Check

6.1 Setting up beam check

Press the i button. (Figure 193)

Figure 193
Press Special Functions (40.0) (Figure 194)

Figure 194

- 92 -
Printed copies of this document are considered uncontrolled. 92
22811.20.Rev001 07.31.2018
Press Beam Check (47.0) (Figure 195)

Figure 195

Press Cephalometric (47.1) (Figure 196)

Figure 196

- 93 -
Printed copies of this document are considered uncontrolled. 93
22811.20.Rev001 07.31.2018
6.2 Computer Setup
NOTE: Make sure the sensor is in the sensor holder.

Click Start then All Programs then Planmeca then Beam Check. (Figure 197)

Figure 197

Click on ProMax Ceph. (Figure 198)

Figure 198

Press Take the Image.

- 94 -
Printed copies of this document are considered uncontrolled. 94
22811.20.Rev001 07.31.2018
Take an exposure. The image should be the same on the top, bottom, left, and
right. (Figure 199)

Figure 199

7 Calibrating
Press Exit on the touch screen, if just finished doing a Beam Check.

7.1 kV / mA for Cephalometric

Binnings kV mA
4x4 60 4
3x3 60 7

- 95 -
Printed copies of this document are considered uncontrolled. 95
22811.20.Rev001 07.31.2018
Place the Cephalostat calibrating tool (hatchet) into the patient’s right ear post
holder with the point going towards the back or column. (Figure 200)

Figure 200

NOTE: Change binnings by clicking Settings and select Binnings then the appropriate binning in
Dimax3tool. (Figure 201)

Figure 201

- 96 -
Printed copies of this document are considered uncontrolled. 96
22811.20.Rev001 07.31.2018
7.2 Computer Setup
Click on Start, All Programs, Planmeca then Dimax3 Tool. (Figure 202)

Figure 202

7.3 Dimax3 Tool


Click on Calibrate then select Ceph (Figure 203)

Figure 203

Click Ready on the touch screen and take an exposure

- 97 -
Printed copies of this document are considered uncontrolled. 97
22811.20.Rev001 07.31.2018
NOTE: image will come out as being a grey hatchet with black around it. (Figure 204)

Figure 204

Press the “L” key on the keyboard of the computer to give calibration lines.
This gives the point where the images are combined in the computer.

Click the magnify glass (+) to zoom in on the lines to check combination.

NOTE: There is a + / - 2 clicks left or right on the keyboard to get aligned left to right.

- 98 -
Printed copies of this document are considered uncontrolled. 98
22811.20.Rev001 07.31.2018
7.3.a Not together

If the image comes out separated check to make sure that the hatchet is
placed in the holder correctly. Re-do Beam Check.

7.3.a.i Raising up
If hatchet is place correctly, then raise the beaver tail
up.

Loosen the two screws on the inside of the beaver tail


and tighten the vertical screw on the outside of the
beaver tail. (Figure 205a-b)

Figure 205a Figure 205b

- 99 -
Printed copies of this document are considered uncontrolled. 99
22811.20.Rev001 07.31.2018
www.planmecausa.com
- 100 -
Printed copies of this document are considered uncontrolled. 100
22811.20.Rev001 07.31.2018

You might also like