SWD
[Link] Bhagat
MPT(cardiopulmonary)
A short wave diathermic current has a frequency of
between 107 and 108 Hz and sets up radio waves with a
wave length of between 30 and 3 m.
The use of medical current within this range is classed as
short-wave diathermy, but that commonly used for
medical work has a frequency of 27.12MHz and sets up
radio waves with a wave length of 11m
this current is generated in a machine circuit which is in
turn coupled to a patient circuit which is used to treat the
patient
It provides the deepest form of heat available to the
physiotherapist
2
principles
This type of current can only be produced by discharging a
condenser through an inductance of low ohmic resistance
If a current of very high frequency is required, the
capacitance and inductance should be small and if a
current of low frequency is required the capacitance and
inductance should be large
This is the mechanism of production of high frequency
current
3
Construction
1.
The machine circuit
2.
The patient circuit
The machine circuit (Oscillator circuit)
Consist of a condenser and an inductance
Currents of different frequencies are obtained by
selecting suitable condensers and inductance
In order to produce the high frequency current, the
condenser must be made to discharge repeatedly, and to
achieve this the oscillator is incorporated in a valve circuit
The patient circuit (Resonator circuit)
This is coupled to the machine circuit by inductors, i.e. a
matching high frequency current is produced in the
resonator circuit by electromagnetic induction
For this to happen the oscillator and resonator must be in
resonance with each other, which requires that the
product of inductance and capacitance must be the same
for both curcuits
When short wave diathermy is applied by the condenser
field method, the electrodes and the patients tissues
form a capacitor, the capacitance of which depends on the
size of the electrodes and on the distance and material
between them, and so is different for each application
When the cable electrode is used it forms an inductance
the value of which varies according to its arrangement
Consequently either the capacitance or the inductance of
the patients circuit is varied at each treatment, and a
variable condenser is incorporated in the patient circuit to
compensate for this
7
When the electrodes have been arranged in position, the
capacitance of the variable condenser is adjusted(the
process of tuning) until the product of the inductance and
capacitance of the resonator circuit is equal to that in the
oscillator circuit
When the oscillator and resonator circuits are in tune with
each other, there is maximum power transfer to the
patient circuit
Indications:
1.
An indicator light on the equipment either comes on or
changes color
2.
An ammeter wired into the resonator circuit shows a
maximum reading, which is diminished by turning the
knob controlling the variable capacitor either way
3.
A tube containing a small amount of neon gas placed
within the electric field between the electrodes or the
ends of the cable will glow at maximum intensity when
the circuits are in resonance
Physiological effects of diathermy
current
A high frequency current does not stimulate motor or
sensory nerves
The shorter the duration of the impulse the less is the
effect on the nerves, 0.01ms being the shortest duration
of impulse generally used
A high frequency current has a frequency of more than
approximately 500kHz. This provides one million impulses
per second, so each has a duration of 0.001ms, which is
beyond the range used for nerve stimulation
10
Thus when such a current is passed through the body there
is no discomfort and no muscle contractions are produced
The current is evenly alternating, therefore there is no
danger of chemical burns
Consequently it is possible to pass through the tissues
currents of a much greater intensity than can be used with
low frequency currents
The intensity of the current can be great enough to
produce a direct heating effect on the tissues
11
Methods of application
1.
Capacitor field method
2.
Cable method
12
Techniques of short wave diathermy
Testing of the machine:
Condenser electrodes- are arranged opposite to each
other with a gap in between. The operator places her
hand between the electrodes, switches on and tunes the
machine, then increases the current until a comfortable
warmth is felt
Cable- may be arranged in single loop and tested with a
neon tube which lights up when the two circuits are in
resonance, or the operator may place her hand between
the ends of the cable and tur up the current until warmth
is felt
13
Preparation of the patient
The couch, chair or table that is used for supporting the
patient should not contain metal, as this is liable to
distort the electric field and to be heated by currents
which may be induced in it
14
Clothing should be removed from the area of treatment,
for a number of reasons
It may be slightly damp from perspiration, and its presence
would interfere with the circulation of air, which aids the
evaporation of any sweat which may form during treatment
Tight clothing may interfere with the flow of the blood
through the area, causing overheating
If the clothing is not removed, the necessary inspection of
the skin before and after treatment is not possible, and the
skin-electrode distance and position of electrodes cannot be
judged accurately
Difficult for the patient to appreciate the sensation of
warmth
15
Metal objects in the clothing may easily pass
undetected. Metal and moisture both have a high
dielectric constant, and a localized are of either causes
concentration of an electric field, with consequent overheating
Metal objects, and anything that is damp, should be
removed from the vicinity of the area to be treated
16
Wounds and sinuses must be cleansed and covered with
a dry dressing before commencing treatment
The area to be treated must be dry, if the area is damp,
the moisture on the surface of the skin is heated quickly
and gives rise to a sensation of warmth which limits the
intensity of current that can be applied
The patient must be comfortable and the part to be
treated must be fully supported, as movement may alter
the skin electrode distance
17
The ammeter is no guide to the amount of heating of
the tissues, merely of value for tuning the circuits
The dose is important for the patient to understand the
degree of warmth that he should feel, that undue heat
should be reported, and there is a danger of burns if the
heat becomes excessive
Skin sensations must be tested before the first
treatment
Hearing aid must be removed and left well away from
the machine, induced currents may cause severe damage
to them
18
Types of electrodes
1.
Several types, but each consist of a metal plate
surrounded by some form of insulating material
One type has a cover within which the position of the
metal plate can be adjusted
Commonly circular, but special shapes are made for some
irregular areas like axilla
Arranged in position on supporting arms and it is advisable
to leave a small gap between the cover and the skin to
allow for the circulation of air
19
2.
Another type of electrodes consist of a rigid metal plate
coated with a thin layer of insulating material, either
rubber or plastic
Frequently convex at the edges, which provides a more even
electric field than a flat disc
Because an electric charge concentrates at the edges of a
conductor and sets up a more intense electric field in this
area than elsewhere; with the convex electrodes the edges
are further from the skin than is the centre so the peripheral
part of the field has room to spread before reaching the skin
This electrodes are arranged in position on supporting arms
and separated from the skin by an air gap
Have an adjustable device projecting from the centre to
ensure correct spacing
20
3.
Third type of electrode consist of a malleable metal plate
covered with a thin layer of rubber
Can be moulded to the part, but should not be bent sharply
or the metal plate may crack
Separated from the skin by perforated felt and their position
is maintained by body weight
Undue pressure, which would interfere with the blood flow
should be avoided
Felt is perforated so that it contains a proportion of air,
which is the most satisfactory spacing material, but the
impossibility of entirely air spacing is one of the
disadvantage
21
4.
The cable electrode consists of a thick wire covered with
rubber
Separated from the skin by at least four layers of dry
Turkish toweling, forming a thickness of at least 1 cm
Toweling is necessary to absorb any perspiration which
may be produced by the heat and result in a scald
Turns of the cable should be at least 2.5 cm apart and
may be secured wit spacers made of insulating material
22
Position and size of the electrodes
The position of the electrodes should be chosen with the aim
of directing the electric field through the structures to be
treated
If the structure is of high impedance the electrodes should
be arranged, as far as possible, so that the different tissues
lie in series with each other, i.e. at right angles to the
electric field
To heat a structure of low impedance it is most satisfactory
if the tissues are parallel to the field
23
If the tissues lie in parallel to the field the heating is in
fact mainly confined to the blood vessels and muscles
Such heating is therefore satisfactory for treatment of the
soft structures
The electrodes should be placed parallel to the skin,
otherwise the field concentrates on the area of tissue
lying closet to the electrode
Placing the electrodes parallel to the skin may result in
their not lying parallel to each other, but provided that
the electrode is through the body tissues, this has little
effect on field distibution
24
Distance between the should be greater than the total
spacing
Electrodes should be placed over an even surface of the
body
if the surface is irregular, the field tends to concentrate
on the more prominent parts
Where an irregular surface cannot be avoided, the
concentration can be reduced by using wide spacing
25
1.
Contraplanar
2.
Coplanar
3.
Cross-fire
4.
Monopolar
26
1.
Contraplanar positioning of electrodes-
. Most
satisfactory method for the treatment of deeply
placed structures
. Placed
over opposite aspects of trunk or limb so that the
electric field id directed through the deep tissues
2.
Coplanar positioning of electrodes-
. Placed
side by side on the same aspect of the part,
provided that there is adequate distance between them,
as the pathway through the tissues offers less impedance
to the lines of forces than through the air between the
electrodes
27
The distribution of the resulting electric field
The distance between the electrodes is more than the
total width of spacing, otherwise the electric field will
not pass through the tissues at all
Heating is more superficial than with the contraplanar
method
Method is particularly suitable for the treatment of
superficial structures where some factor contraindicates
the placing of an electrode immediately over the lesion
28
3.
Cross-fire method-
. Half
the treatment is given with the electrode in one
position, then the arrangement is changed so that the
electric field lies at right angles to that obtained during
the first part of the treatment
. Used
to treat the walls of cavities containing air, e.g. the
frontal, maxillary and ethmoid sinuses
. The
lines of force pass through the tissues between the
electrodes but avoid the cavity, as the air within it has a
low dielectric constant
. Thus
those walls of the cavity which face the electrodes
are not treated
29
If the position of the electrodes is then changed so that
the field lies at right angles to previous one, these walls
are heated
Contact lenses should be removed because the heating
effect may cause melting of the lens
May also be used for deeply placed structures, particularly
if they lie in extensive vascular areas, e.g. pelvic organs
The dielectric constant of the vascular tissues is very high
and the cross sectional area of the part is larger than the
electrodes, so the field spreads in the deep tissues, which
consequently receive less heating than the superficial
ones
30
By passing the field through the area in two directions,
thee deep tissues receive twice as long an exposure as the
skin
31
4.
Monopolar technique-
. The
active electrode is placed over the site of the lesion
and the indifferent electrode is applied to some distant
part of the body, or may not be used at all
. A
separate electric field is set up under each electrode,
the lines of force radiating from the electrode
. Thus
the density of the field becomes less as the distance
from the electrode increases, and the heating is
superficial
32
Connecting leads
The leads or the cable must be of the correct length for
the particular electrodes and machine that are used
The leads should lie parallel to each other, at least as far
apart as the terminals of the machine, and not approach
close to any conductor
Currents may induced in any conductor which lies near to
the leads, with consequent loss of energy from the leads
and possible damage to the conductor from over-heating
Must be separated from pts skin by a distance at least as
great as the electrodes spacing, otherwise they induced
currents in the tissues and cause heating in this area
33
Application of current
When the pt, electrodes, and leads are in position, the
current is turned on and the circuits tuned
The current is then turned up slowly, to allow
vasodilatation to occur and for the patient to appreciate
the degree of heating
The operator should remain within call of the patient
throughout the treatment, and turn the current off
immediately if the heating becomes excessive
34
A the end of the treatment the controls are turned to
zero, the current is switched off and the electrodes are
removed
Skin may be faintly pink, but there should be no strong
reaction
35
Dosage
Intensity of the application should be sufficient to cause a
comfortable warmth and the duration of the treatment
should be 20-30 minutes (except for the treatment of
chronic inflammatory lesions, when a duration of at least
30 minutes is desirable)
Daily or on alternate days
For the treatment of acute inflammation, or recent injury,
the application should be less intense than that suggested
above, but may be carried out more frequently
36
The current used may be that which produces a mild
sensation of warmth, or it may be increased until mild
warmth is felt then reduced to the point at which the
sensation is no longer perceptible
The duration is limited to 5-10 minutes; the dose is
increased progressively but cautiously according too the
effects observed
When the inflammation is within a confirmed space, such
as the air sinuses of the face, it is particularly important
that excess treatment should be avoided, as rise in
tension in such an area seriously aggravates the
symptoms
37
Therapeutic uses of SWD
Effects on inflammation
Dilatation of the arterioles and capillaries
Increased blood flow to the area
Available a increased supply of oxygen and nutritive
materials, more antibodies, and WBCs
Vasodilatation increases the exudation of fluid into the
tissues and followed by increased absorption which,
together with the increased flow of blood through the
area, assist in the removal of waste products
Helps to bring resolution of inflammation
38
In acute stage of inflammation, caution should be
exercised in applying the treatment to areas in which
there is already marked vasodilatation and exudation of
fluid, as an increase in these processes may aggravate the
symptoms
In sub acute stages, stronger doses may be applied with
considerable benefit
When the inflammation is chronic, a thermal dose of fairly
long duration must be used to be effective
39
Effects in bacterial infections
Inflammation is the normal response of the tissues to the
presence of bacteria, the principle features being
vasodilatation, exudation of fluid into the tissues, and an
increase in the concentration of WBCs and antibodies in
the area
Heating the tissues augments these changes and so
reinforces the bodys normal mechanism of dealing with
the infecting organisms, therefore SWD may be of value
in the treatment of bacterial infections such as boils,
carbuncles and abscesses
40
Treatment in the early stages may occasionally bring
about resolution of the inflammation without pus
formation occurring, failing this, the development of the
inflammatory response is accelerated
Until there is free drainage, the treatment should be given
cautiously, as in all cases of acute inflammation
When the abscess is draining freely, stronger doses may be
applied, the increased blood supply assisting the healing
processes once the infection has been overcome
41
In some cases SWD appears to aggravate the condition,
but increased discharge for a few days is an indication of
acceleration of the changes occurring in the tissues, and
not a contraindication to treatment
However, should the increased discharge persist it may be
an indication that the bodys defense mechanism is
already taxed to its uttermost, so that it is impossible to
reinforce its action
This is most likely to occur in case of long standing
infection, and under these circumstances no benefit is
derived from the application of SWD
42
Traumatic conditions
Effects of SWD on traumatic lesions ae similar to those
produced in inflammation
The exudation of fluid (followed by increased absorption)
and the increased flow of blood through the area assist in
the removal of waste products, while the improved blood
supply makes available more nutritive materials, so
assisting the healing process
Recent injuries should be treated with the same caution as
acute inflammation, as excessive heating is liable to
increase the exudation of fluid from the damaged vessels
43
Stiff joints and other after effects of injury require
stronger doses, the treatment being a preliminary to the
exercises which is usually the essential part of the
treatment
44
Reducing healing time
To promote the healing of, for instance, a wound, an
increased blood supply to the tissues may be of assistance,
provided that the vascular responses to heat are normal
45
Relief of pain
A mild degree of heating is effective in relieving pain,
presumably as a result of a sedative effect
Pain may be due to the accumulation in the tissues of the
waste products of metabolism and that the increased flow
of blood through the area assist in removing these
substances
When pain is due to inflammatory processes, resolution of
the inflammation is accompanied by relief of pain
46
However, strong heating in these cases may cause an
increase of pain, especially in acute inflammation, if the
increased blood flow and exudation of fluid cause an
increase of tension in the tissues
47
Effects on muscle tissue
The heating of the tissues induces muscle relaxation, so
SWD may be used for the relief of muscle spasm
associated with inflammation and trauma, usually as a
preliminary to movements
Increased efficiency of muscle action should also aid the
satisfactory performance of active exercises
48
Dangers
Burns
In severe cases there is coagulation and destruction of the
tissues, and the burn appears as a white patch surrounded
by a reddened area of inflammation
49