Intermediate Modifying & Terminating Devices
Intermediate Modifying & Terminating Devices
•In most cases, the mechanical quantity which was detected will
be transduced into an ‘electrical form’. The output of the first
stage has to be modified (signal conditioning) before it is fed to
the third or terminating stage such as indicators, recorders or
control elements.
•Signal conditioning equipment used may be of mechanical,
electrical or electronic. Mechanical types (using elements such as
linkages, gearing, cams, etc.) have many limitations such as friction,
inertia, non linearity, backlash, elastic deformation, etc.
•Hence electrical & electronic systems are used which are free
from these drawbacks. Also they give large voltage & power
amplifications required to drive the recording devices.
ei
kRi
Resistance
type transducer
Rt
= 0.5
Rm
2
io
imax
4
10
k
Ballast Circuit
A ballast circuit is only a simple variation of the
current sensitive circuit. In this case a voltage
sensitive device is connected across the
transducer as shown in fig. It is also called as
‘voltage sensitive circuit’.
A ballast resistor Rb is the resistance of the
measuring circuit excluding the transducer. In the
absence of a ballast resistor, the voltage indicator
will always record the full source voltage ei & hence
some value of resistance Rb is always necessary
for proper functioning of the circuit.
In order to analyze a ballast circuit, we assume that
the voltage indicator has an infinite resistance such
that it does not draw any current.
Schematic of Ballast Circuit
R
b
Resistance
ei type transducer
Voltage indicator
to sense output kRi
voltage eo
By Ohm' s law, the output current is ,
ei
io . If e o is the voltage across kR t , which is indicated by
R b kR t
the voltage indicator, then the output voltage indicated is,
ei kR t
e o i o (kR t ) . This can be written as,
R b kR t
eo kR t k Rt Rb e
. For a ballast circuit, o is a meausure of the output and
ei R b kR t kR t ei
1
R
b
kR t
is a measure of the input.
Rb
1.0
eo 0.5
ei
k
Electronic Amplifiers
Electronic amplifiers are used to provide voltage gain,
current gain, and impedance transformation.
In most transducers, electrical voltage is the output
but the voltage level available from the transducer
is very low, hence a voltage amplifier is required to
increase the level for subsequent processing.
Some times the input signal may be used to drive a
recorder or some control apparatus. In such cases
power must be increased by using current or power
amplifiers.
Also high output impedance leads to noise. Hence it
is desirable to include an amplifier which converts
high impedance input into a low impedance output.
Vacuum tube Amplifiers
Plate
Grid Cathode
Signal
input Amplifier
load
Heater supply
(A)
_ + _ +
Bias voltage Plate supply
(C) (B)
TELEMETRY
Temperature
SCO
TRANSMITTING
+ SYSTEM
_
Transmitter
SCO
Accelration
etc
Sub carrier
Descriminator
SCD
TELEMETRY
Antenna RECEIVING
SCD SYSTEM
Receiver
Recorder SCD
etc
Terminating devices
Usefulness of any measuring system depends on its ability
to present the measured quantity in a form which can
be understood fully by the human operator or any
controlling device. The primary function of a
terminating device is to accept the analogous driving
signal and to provide output for the immediate reading
or for recording.
For direct human interpretation, a terminating device
provides information as;
(1) A relative displacement: For ex, a pointer moving over
a scale, light beam & scale, liquid column & scale, etc.
(2) A digital form: Examples: Odometer in an automobile
speedometer, a rotating drum mechanical counter.
(3) ‘Yes’ or ‘No’ limiting type: Ex, red oil pressure lights in
automobiles, Pilot lamps on equipments.
Most of the dynamic mechanical measurements require
electrical terminating devices due to poor response
characteristics of mechanical, pneumatic or optical
systems.
Meter indicators
Pointer & scale meters are useful for static &
steady state dynamic indications, but not
suitable for transient measurements. This is
due to relatively high inertia of the meter
movement. Meter indicators may be classified
as: (i) Simple D’Arsonval type meter
(ii) Ohm meters & Volt-Ohm milli ammeters
(iii) Vacuum tube voltmeters.
Among these, D’Arsonval type meter is widely
employed as the final indicating device.
D’Arsonval type meter is the common type used
for measuring either current or voltage. It
consists of a coil assembly mounted on a
pivoted shaft whose rotation is constrained by
two spiral springs, one at each end of the shaft
as shown in fig. The coil assembly is mounted in
a magnetic field.
The electric current to be measured is passed
through the coil and the two interacting magnetic
fields result in a torque applied to the pivoted
assembly. Then the resulting displacement of
the pointer on scale is calibrated in terms of
electric current. This principle forms the basis for
most of the electric meters, stylus & light beam
Oscillograph.
D’Arsonval type meter
D’Arsonval type meter
Scale
Pointer
N S
Mechanical Counter
Mechanical
Counter
CATHODE RAY OSCILLOSCOPE (CRO)
CRO is the most versatile readout device for mechanical
measurements. It is used for measurement and analysis of
waveforms and other phenomenon in electrical & electronic
circuits. CRO is a voltage sensitive instrument with an electron
beam striking the fluorescent screen. The extremely low inertia
beam of electrons enables it to be used for following the
rapidly varying voltages.
The heart of the CRO is the Cathode ray tube (CRT), whose
important parts are;
(1) Electron gun assembly: The electron gun assembly produces
a sharply focused beam of electrons which in turn are
accelerated to high velocity. This beam of electrons strikes the
fluorescent screen with sufficient energy to cause a luminous
spot on the screen.
(2) Electron gun: An electron gun emits electrons and makes
them into a beam. It consists of a heater, cathode, grid,
focusing and accelerating anodes. Electrons are emitted from
an indirectly heated cathode. These pass through a small hole
in the control grid. The grid controls the electrons emitted from
the cathode and hence the intensity of the beam. The
CRT (continued…..)
(3) Deflection plates: These are two pairs of electrostatic
plates. A voltage applied to a pair of vertical plates moves
the electron beam vertically up or down. And if the voltage is
applied to the pair of horizontal plates, the electron beam
moves horizontally from one end to other end of the screen.
The CRT is evacuated so that the emitted electrons can
move freely from one end of the tube to the other.
Usually in CRO’s, the horizontal voltage is internally
developed where as the vertical voltage is the voltage under
investigation (input). This voltage moves the luminous spot
up & down in accordance with the instantaneous value of
voltage. In other words, it traces the ‘waveform’ of the input
voltage w.r.t. time.
CRO’s can also be used to visualize various quantities such
as current, strain, acceleration, pressure if they can be
converted into voltages.
Important parts of a Cathode ray tube
Flourescent
Electron gun screen
Vertical
deflection
plates
Cathode
Accelerating
anodes
Heater
Mirror
Reflected light
beam
coil assembly
Lens
N
Input
Photographic
Incident paper
light beam
Light source
Permanent
magnet
Coil assembly
Permanent magnet
N S
Stylus
Paper
Paper movement
Trace
Reference
source
Chopper Amplifier
Pen
Balance drive
Y-Input Attenuator circuit motor
Pen X-diection
A
R
Y-direction M
Graph
paper
Balance Arm
X-Input Attenuator drive
circuit
motor
Chopper Amplifier
Reference
source