KERN & Sohn GmbH
Ziegelei 1 D-72336 Balingen email: [email protected] Phone: +49-[0]7433- 9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com
Operating Manual Counting balances
KERN CKE
Version 1.3 01/2009 GB
CKE-BA-e-0913
GB
KERN CKE
Version 1.3 01/2009
Operating Manual Counting balances
Table of Contents 1 2 3 MODE MEN...............................................................................................................4 Technical Data ..............................................................................................................5
2.1 3.1 3.2 3.3 3.4 KERN CKE .......................................................................................................................... 5 Proper use .......................................................................................................................... 7 Improper Use...................................................................................................................... 7 Warranty ............................................................................................................................. 7 Monitoring of Test Resources .......................................................................................... 7 Pay attention to the instructions in the Operation Manual ............................................ 8 Personnel training ............................................................................................................. 8 Testing upon acceptance.................................................................................................. 8 Packaging........................................................................................................................... 8 Installation Site, Location of Use ..................................................................................... 9 Unpacking .......................................................................................................................... 9
Placing ............................................................................................................................................ 9
Basic Information (General).........................................................................................7
Basic Safety Precautions .............................................................................................8
4.1 4.2
Transport and storage..................................................................................................8
5.1 5.2
Unpacking, Setup and Commissioning ......................................................................9
6.1 6.2
6.2.1
6.3 6.4 6.5 6.6 6.7 6.8
Mains connection .............................................................................................................. 9 Connection of peripheral devices .................................................................................... 9 Initial Commissioning ....................................................................................................... 9 Adjustment ....................................................................................................................... 10 Adjusting (see chapter 7.4.1).......................................................................................... 10 Verification ....................................................................................................................... 10 Control panel CKE ........................................................................................................... 11 Operation.......................................................................................................................... 12
Operation.....................................................................................................................11
7.1 7.2
7.2.1 Volume assessment - KERN QKE only !!!................................................................................... 14 7.2.1.1 Volume assessment without manual input .......................................................................... 14 7.2.1.2 Volume assessment with manual tare input ........................................................................ 14 7.2.1.3 Volume assessment with manual tare and gross input ....................................................... 14 7.2.2 Display background illumination................................................................................................... 14
7.3
Data output RS 232 C ...................................................................................................... 15
CKE-BA-e-0913
7.4
7.4.1 7.4.2 7.4.3
Interface RS 232C.............................................................................................................16
There are 4 kinds of data output via RS 232C ............................................................................. 16 Explanation of the data transfer.................................................................................................... 16 Numerator ..................................................................................................................................... 16
7.5 7.6
Printer................................................................................................................................17 Underfloor weighing ........................................................................................................17 Cleaning ............................................................................................................................18 Service, maintenance.......................................................................................................18 Disposal ............................................................................................................................18
Service, maintenance, disposal ................................................................................ 18
8.1 8.2 8.3
9 10
Instant help ................................................................................................................. 18 Declaration of conformity....................................................................................... 19
CKE-BA-e-0913
1 MODE MEN
How to invoke the mode menu: Turn on balance, press and hold the tare key and press the ON/OFF key. Let go of the tare key.
Factory settings: 9600bd: Fast: YES 3
Keyboard overview / Function No Yes
CKE-BA-e-0913
2 Technical Data
2.1 KERN CKE CKE 6k0.02
0,02 g 6.100 g 6.100 g 0,02 g 0,06 g 0,02 g 2/4/5/6 kg 5 kg
KERN
Readability (d) Weighing range (max) Taring range (subtractive) Reproducibility Linearity Smallest piece weight Adjustment point Recommended adjustment weight (F1), not added Appropriate for verification Stabilization time (typical) Allowable ambient temperature Housing (B x D x H) mm Vibration filter Weighing plate, stainless steel mm Units Weight kg (net) Data interface
CKE 8k0.05
0,05 g 8.100 g 8.100 g 0,05 g 0,15 g 0,05 g 2/4/5/7/8 kg 5 kg + 2 kg
CKE 16k0.05
0,05 g 16.100 g 16.100 g 0,05 g 0,15 g 0,05 g 5/10/15/16 kg 10 kg + 5 kg
CKE 16k0.1
0,1 g 16.100 g 16.100 g 0,1g 0,3 g 0,1 g 5/10/15/16 kg 10 kg + 5 kg
max. 80% rel. (non-condensing) 3 sec. +10 C ... + 40 C 350 x 390 x 120 ja 340 x 240 see menu 6,5 ja (RS232)
CKE-BA-e-0913
KERN
Readability (d) Weighing range (max) Taring range (subtractive) Reproducibility Linearity Smallest piece weight Adjustment point Recommended adjustment weight (F1), not added Appropriate for verification Stabilization time (typical) Allowable ambient temperature Housing (B x D x H) mm Vibration filter Weighing plate, stainless steel Units Weight kg (net) Data interface
CKE 36k0.1
0,1 g 36.100 g 36.100 g 0,1 g 0,5 g 0,1 g 10/20/30/36 kg 20 kg + 10 kg
CKE 36k0.2
0,2 g 36.100 g 36.100 g 0,2 g 0,6 g 0,2 g 10/20/30/36 kg 20 kg + 10 kg
CKE 65k0.2
0,2 g 65.100 65.100 0,2 g 0,6 g 0,2 g 20/30/50/60 kg 50 kg
CKE 65k0.5
0,5 g 65.100 65.100 0,5 g 1,5 g 0,5 g 20/30/50/60 kg 50 kg
max. 80% rel. (non-condensing) 3 sec. +10 C ... + 40 C 350 x 390 x 120 ja 340 x 240 see menu 6,5 ja (RS232)
CKE-BA-e-0913
3 Basic Information (General)
It is absolutely necessary that you read and understand the operating instructions prior to installation and commissioning and follow the instructions during the process!
3.1
Proper use
The balance you purchased is intended to determine the weighing value of material to be weighed. It is intended to be used as a non-automatic balance, i.e. the material to be weighed is manually and carefully placed in the centre of the weighing plate. As soon as a stable weighing value is reached the weighing value can be read.
3.2
Improper Use
Do not use balance for dynamic weighings. In the event that small quantities are removed or added to the material to be weighed, incorrect weighing results can be displayed due to the stability compensation in the balance. (Example: Slowly draining fluids from a container on the balance.) Do not leave permanent load on the weighing plate. This may damage the measuring system. Impacts and overloading exceeding the stated maximum load (max) of the balance, minus a possibly existing tare load, must be strictly avoided. Balance may be damage by this. Never operate balance in explosive environment. The serial version is not explosion protected. The structure of the balance may not be modified. This may lead to incorrect weighing results, safety-related faults and destruction of the balance. The balance may only be used according to the described conditions. Other areas of use must be released by KERN in writing.
3.3
Warranty
Warranty claims shall be voided in case - Our conditions in the operation manual are ignored - The appliance is used outside the described uses - The appliance is modified or opened - mechanical damage and damage caused by media, liquids - natural wear and tear - The appliance is improperly set up or incorrectly electrically connected - The measuring system is overloaded
3.4
Monitoring of Test Resources
In the framework of quality assurance the measuring-related properties of the balance and, if applicable, the testing weight, must be checked regularly. The responsible user must define a suitable interval as well as type and scope of this test. Information is available on KERNs home page (www.kern-sohn.com) with regard to the monitoring of balance test substances and the test weights required for this. Our accredited DKD calibration laboratory offers fast and inexpensive adjustment for test weights and weighing balances (reset to national normal weight).
CKE-BA-e-0913
4 Basic Safety Precautions
4.1 Pay attention to the instructions in the Operation Manual
Carefully read this operation manual before setup and commissioning, even if you are already familiar with KERN balances.
4.2
Personnel training
The appliance may only be operated and maintained by trained personnel.
5 Transport and storage
5.1 Testing upon acceptance
When receiving the appliance, please check packaging immediately, and the appliance itself when unpacking for possible visible damage. In case of visible damage have the damage verified by the messengers signature. Do not alter goods or packaging and do not remove any parts of the delivery. Report the damage immediately (within 24 hours) in writing to the parcel service.
5.2
Packaging
Keep all parts of the original packaging in case you need to return the appliance. Only use original packaging for returning. Prior to dispatch disconnect all connected cables and remove loose/mobile parts and the weighing plate. Attach possibly existing transport safeguards. Secure all parts such as glass wind screen, weighing plate, power pack etc. against slipping and damage.
CKE-BA-e-0913
6 Unpacking, Setup and Commissioning
6.1 Installation Site, Location of Use
The balances are designed in a way that reliable weighing results are achieved in common conditions of use. You will work accurately and fast, if you select the right location for your balance. Therefore, observe the following for the installation site: Place the balance on a firm, level surface; Avoid extreme heat as well as temperature fluctuation caused by installing next to a radiator or in the direct sunlight; Protect the balance against direct draughts due to open windows and doors; Avoid jarring during weighing; Protect the balance against high humidity, vapours and dust; Do not expose the device to extreme dampness for longer periods of time. Non-permitted condensation (condensation of air humidity on the appliance) may occur if a cold appliance is taken to a considerably warmer environment. In this case, acclimatize the disconnected appliance for ca. 2 hours at room temperature. Avoid static charging of the material to be weighed, weighing container and windshield. If electro-magnetic fields or static charge occur, or if the power supply is unstable major deviations on the display (incorrect weighing results) are possible. In that case, the location must be changed.
6.2
Unpacking
Carefully remove the balance from the packaging, remove plastic cover and setup balance at the intended workstation.
6.2.1 Placing
The balance must be installed in a way that the weighing plate is exactly in horizontal position.
6.3
Mains connection
Power is supplied via the external mains adapter. The stated voltage value must be the same as the local voltage. Only use original KERN mains adapters. Using other makes requires consent by KERN.
6.4
Connection of peripheral devices
Before connecting or disconnecting of additional devices (printer, PC) to the data interface, always disconnect the balance from the power supply. With your balance, only use accessories and peripheral devices by KERN, as they are ideally tuned to your balance.
6.5
Initial Commissioning
A warming up time of 2 hours after switching on stabilizes the measuring values. The accuracy of the balance depends on the local acceleration of gravity. Strictly observe hints in chapter Adjustment.
CKE-BA-e-0913
6.6
Adjustment
As the acceleration value due to gravity is not the same at every location on earth, each balance must be coordinated - in compliance with the underlying physical weighing principle - to the existing acceleration due to gravity at its place of location (only if the balance has not already been adjusted to the location in the factory). This adjustment process must be carried out during the initial start-up, after change in location and variation of surrounding temperature. To receive accurate measuring values it is also recommended to adjust the balance periodically in weighing operation.
6.7
Adjusting (see chapter 7.4.1)
With an adjustment weight, the weighing accuracy can be checked and re-adjusted at any time. Attention: In the verified balances the adjustment is not possible. Procedure when adjusting: Observe stable environmental conditions. A short warming up time of ca. 15 minutes is recommended for stabilization.
6.8
Verification
General introduction: According to EU directive 90/384/EEC balances must be officially verified if they are used as follows (legally controlled area): a) For commercial transactions if the price of goods is determined by weighing. b) For the production of medicines in pharmacies as well as for analyses in the medical and pharmaceutical laboratory. c) For official purposes d) For manufacturing final packages In cases of doubt, please contact your local trade in standard. Verification instructions An EU type approval exists for balances described in their technical data as verifyable. If a balance is used where obligation to verify exists as described above, it must officially verified and re-verified in regular intervals. Re-verification of a balance is carried out according to the respective national regulations. The validity for verification of balances in Germany is e.g. 2 years. The legal regulation of the country where the balance is used must be observed!
10
CKE-BA-e-0913
7 Operation
7.1 Control panel CKE
ON/OFF Taring; Invoke the mode menu by pressing the ON/OFF key. Generate reference using 5 parts; In mode menu: NO function Generate reference using 10 parts Generate reference using 20 parts; In mode menu: YES function Switchover pcs g; In menu: Mode function. Print out weighing result.
Display icon ==OVERLOAD== ============= << . .
Significance Overload: Weighing range exceeded Underload: Weighing range not achieved In add-up and % mode: Part too light Balance in add-up mode show the weight value of the counted quantity
CKE-BA-e-0913
11
7.2
Operation
12
CKE-BA-e-0913
CKE-BA-e-0913
13
7.2.1 Volume assessment - KERN QKE only !!! 7.2.1.1 Volume assessment without manual input
Switch on balance Place an empty vessel and tare Place a vessel containing ideal filling (100%) and save the value by pressing the MAX key It is now possible to qualify additional filled vessels by using the percentage display (%)
7.2.1.2 Volume assessment with manual tare input
Switch on balance Press any arrow key Select decade by pressing the horizontal arrow keys Select numeric value by pressing the vertical arrow keys Save input value by pressing the tare key Place a vessel containing ideal filling (100%) and save the value by pressing the MAX key It is now possible to qualify additional filled vessels by using the percentage display (%)
7.2.1.3 Volume assessment with manual tare and gross input
Switch on balance Press any arrow key Select decade by pressing the horizontal arrow keys Select numeric value by pressing the vertical arrow keys Save input of tare value by pressing the tare key Save input of gross value by pressing the MAX key It is now possible to qualify additional filled vessels by using the percentage display (%)
7.2.2
Display background illumination
Switch on the balance and the zero display, then call-up the balance menu as specified in chapter 1. Select menu item Backlight using the button necessary for the respective model. Acknowledge by pressing the YES button in order to switch on the background illumination permanently. By pressing the NO button, the background illumination is switched off. If the background illumination shall be switched off time-controlled (to save the battery), press the respective button for continuing the menu to select the menu point Backlight auto and confirm by pressing the YES button. The background illumination will be switched off automatically 10 sec after having reached a stable weighing value.
14
CKE-BA-e-0913
7.3 Data output RS 232 C Technical Data
8-bit ASCII Code 1 start bit, 8 data bits, 1 stop bit, no parity bit Baud rate selectable from 2400, 4800, 9600 Baud (factory setting) and 19200 Baud. Sub-D plug 9-channel required For operation with interface faultless operation is only ensured with the correct KERN interface cable (max. 2m)
Pin allocation of the balance output socket (front view)
Sub-D jack 9-channel
Pin 2: Pin 3: Pin 5:
Transmit data Receive data Signal ground
9 Baud rate
The MODE key is used to set the Baud rate used for the transfer of measured values. In the following example the baud rate is set to 4800 baud.
Setting Baud rate KERN CKE
1. Switch on balance 2. Press and hold tare key 3. Tap ON/OFF key and let go of tare key "Cal?" will appear on the display. 4. Keep holding the switch over key pcs g until "Printer" appears on the display and confirm by pressing the "20x" key. 2.400 Bd will appear on the display. 5. Press the switchover key "pcs g" to select the Baud rate and confirm by the 20x key. To return to weighing mode, press the tare key.
Display
PRINTER? 2400 Baud 4800 Baud 4800 Baud X 0.00 oz
CKE-BA-e-0913
15
7.4
Interface RS 232C
Data output via interface RS 232C General Information The previous condition for the data transfer between balance and a peripherical device (e.g. printer, PC ...) is that the appliances are set to the same interface parameters (e.g. baud rate, parity ...).
7.4.1 There are 4 kinds of data output via RS 232C
Data output using the PRINT key
The printing process can be triggered by pressing the PRINT key. The settings AUTOPRINT and AUTOPRINT PC should be disabled for this process.
AUTOPRINT (data output according to weight application)
The setting AUTOPRINT is located on the PRINTER path where you can turn it on or off. When AUTOPRINT is active, the current weighing value will be sent via the RS 232 date interface after unloading and subsequent loading of the balance as soon as the balance is in resting position.
AUTOPRINT PC (continuous data output)
The setting AUTOPRINT PC is located on the PRINTER path and where you can turn it on or off. When AUTOPRINT PC is active, the current weighing values will be sent continuously via the RS 232 data interface.
Data output and remote control commands
Remote control commands transferred as ASCII characters to the balance can be used to trigger the following functions on the balance (always finish with CR, LF!): t Taring w The balance sends a weighing value (also unstable) via the serial interface. s The balance sends a stable weighing value via the serial interface. After receiving either character w or s, the balance will send without a printer pause between the characters.
7.4.2 Explanation of the data transfer
Each data transfer is structured as follows: Without numerator Bit no. 1 2 B B* B: B, 0, ., g: CR: LF: = = = = 3 B 4 B 5 B 6 B 7 B 8 B 9 B 10 0 11
.
12 0
13 B
14 g
15 B
16 B
17 CR
18 LF
Blank or for ? autotare on in zero range. Blank or weighing value giving unit according to loading of the balance Carriage Return Line Feed
With numerator: Bit no. 1 2 3 4 N N N B* N: = Numerator
5 B
6 B
7 B
8 B
9 B
10 B
11 B
12 0
13
.
14 0
15 B
16 G
17 B
18 B
19 20 CR LF
7.4.3 Numerator
The numerator is situated under the menu item "Printer" where it can be activated or deactivated. When data output takes places via the printer key, it will be increased by one digit per key stroke.
16
CKE-BA-e-0913
7.5
Printer
The serial interface RS 232 facilitates the connection of a printer. The printout shows the weight in grams. In counting mode either the piece number or the weight details will be printed out. In percentage mode the percentage proportion or the weight details will be printed out. Printout will take place after pressing the PRINT key. It is possible to number each printout continuously with the help of the numerator. The numerator will be reset to (000) each time the balance is turned off or the CLEAR function is actuated.
7.6
Underfloor weighing
Objects which are unsuitable for placement on the weighing tray due to their size or shape can be weighed with the help of the underfloor weighing facility. Proceed as follows: Switch off balance. Turn over the balance and in doing so take care that the weighing plate is not loaded. Open the closing lid on the bottom of your balance. Mount the hooks for underfloor weighing. Put the balance over an opening Suspend the goods to be weighed from the hook and carry out the weighing.
! Caution!
Pay attention that the hooks used for underfloor weighing are stable enough to provide a secure support for the desired items to be weighed (risk of breakage). Always ensure that there are no persons, animals or objects that might be damaged underneath the load.
! Note!
After completing the underfloor weighing the opening on the bottom of the balance must always be closed (dust protection).
CKE-BA-e-0913
17
8 Service, maintenance, disposal
8.1 Cleaning
Before cleaning, please disconnect the appliance from the operating voltage. Please do not use aggressive cleaning agents (solvents or similar agents), but a cloth dampened with mild soap suds. Ensure that no liquid penetrates into the device and wipe with a dry soft cloth. Loose residue sample/powder can be removed carefully with a brush or manual vacuum cleaner. Spilled weighing goods must be removed immediately.
8.2
Service, maintenance
The appliance may only be opened by trained service technicians who are authorized by KERN. Before opening, disconnect from power supply.
8.3
Disposal
Disposal of packaging and appliance must be carried out by operator according to valid national or regional law of the location where the appliance is used.
9 Instant help
In case of an error in the program process, briefly turn off the balance and disconnect from power supply. The weighing process must then be restarted from the beginning. Help:
Fault
The displayed weight does not glow.
Possible cause
The balance is not switched on. The mains supply connection has been interrupted (mains cable not plugged in/faulty). Power supply interrupted. Draught/air movement Table/floor vibrations Weighing plate has contact with other objects. Electromagnetic fields / static charging (choose different location/switch off interfering device if possible) The display of the balance is not at zero Adjustment is no longer correct. Great fluctuations in temperature. Electromagnetic fields / static charging (choose different location/switch off interfering device if possible)
The displayed changing -
weight
is
permanently
The weighing value is obviously wrong
Should other error messages occur, switch balance off and then on again. If the error message remains inform manufacturer.
18
CKE-BA-e-0913
10 Declaration of conformity
KERN & Sohn GmbH
D-72322 Balingen-Frommern Postfach 4052 E-Mail: [email protected] Tel: 0049-[0]7433- 9933-0 Fax: 0049-[0]7433-9933-149 Internet: www.kern-sohn.de
Konformittserklrung
EC-Konformittserklrung EC- Dclaration de conformit EC-Dichiarazione di conformit EC- Declarao de conformidade EC-Deklaracja zgodnoci EC-Declaration of -Conformity EC-Declaracin de Conformidad EC-Conformiteitverklaring EC- Prohlen o shode EC-
D GB CZ E F I NL P PL RUS
Konformittserklrung Declaration of conformity Prohlen o shode Declaracin de conformidad Dclaration de conformit Dichiarazione di conformit Conformiteitverklaring Declarao de conformidade Deklaracja zgodnoci
Wir erklren hiermit, dass das Produkt, auf das sich diese Erklrung bezieht, mit den nachstehenden Normen bereinstimmt. We hereby declare that the product to which this declaration refers conforms with the following standards. Tmto prohlaujeme, e vrobek, kterho se toto prohlen tk, je v souladu s ne uvedenmi normami. Manifestamos en la presente que el producto al que se refiere esta declaracin est de acuerdo con las normas siguientes Nous dclarons avec cela responsabilit que le produit, auquel se rapporte la prsente dclaration, est conforme aux normes cites ci-aprs. Dichiariamo con ci che il prodotto al quale la presente dichiarazione si riferisce conforme alle norme di seguito citate. Wij verklaren hiermede dat het product, waarop deze verklaring betrekking heeft, met de hierna vermelde normen overeenstemt. Declaramos por meio da presente que o produto no qual se refere esta declarao, corresponde s normas seguintes. Niniejszym owiadczamy, e produkt, ktrego niniejsze owiadczenie dotyczy, jest zgodny z poniszymi normami. , , , .
Electronic Balance: KERN CKE
Mark applied EU Directive 2004/108/EC Standards EN 61000-6-3 :2007 EN 55011:1998+A1:1999+A2 :2002 EN 61000-3-2 :2006-04 EN61000-3-3 :1995-01+A1:2001-06+A2 :2005-11 EN45501 :1992-10+AC :1993-08 OIML R 76-1 :2006 EN60950
2006/95/EC
Date: 26.11.2008
Signature: Gottl. KERN & Sohn GmbH Management
Gottl. KERN & Sohn GmbH, Ziegelei 1, D-72336 Balingen, Tel. +49-[0]7433/9933-0,Fax +49-[0]7433/9933-149
CKE-BA-e-0913
19