ENDOSCOPE CLEANING GUIDE
VISERA PEF-V Transnasal Esophagovideoscope
Refer to companion ENDOSCOPE DISINFECTION GUIDE for disinfection information
WARNING: This guide is only a summary of the steps necessary to properly clean your endoscope. Be sure to follow
the detailed steps outlined in the INSTRUCTIONS manual that was included with your endoscope when purchased.
PRECLEANING - IMMEDIATELY AFTER USE
1 2
33
MAJ-1218
Wipe down the insertion tube with a detergentsoaked cloth.
Using the suction valve, aspirate detergent, then air through suction channel.
Using the air/water valve, flush water, then air, through air/water channels.
Disconnect all detachable parts and cleaning equipment. Discard single-use biopsy valve (MAJ-1218).
LEAKAGE TESTING
5
CAUTION: If a leak is detected, the endoscope must be repaired. Follow manufacturer's instructions.
MANUAL CLEANING
6 7 8
Attach leakage tester. Turn on pump. Confirm bending section has expanded. Immerse entire endoscope in clean water. Perform leakage test. Angulate tip during test. Remove from water. Detach tester after tip has deflated.
Immerse in freshly prepared detergent solution. Clean all external surfaces. Wipe endoscope distal tip.
Brush the insertion tube portion of the suction channel. Repeat until all debris is removed.
Brush the universal cord portion of the suction channel. Repeat until all debris is removed.
10
12
13
Suction Cleaning Adapter
Injection Tube Channel Plug
12
Brush the suction cylinder and instrument channel port. Repeat until all debris is removed.
Aspirate detergent solution into suction channel.
Attach injection tube & channel plug. Flush detergent solution into air, water and suction channels.
Soak in detergent solution for the time recommended by the detergent manufacturer.
2003 Olympus America, Inc. / Printed in USA - 08/03/ PWC-100003
c
RINSING
13
Injection Tube Channel Plug
Channel Plug
IMP ORT ANT
14
Injection Tube
15
Wear all appropriate
personal protective equipment.
Meticulous cleaning is
essential for effective disinfection.
Be sure to reprocess all
removable parts (e.g., valves) according to the reprocessing manual.
Rinse the entire instrument in clean water. Flush clean water through all channels.
Remove from water. Flush air through all channels.
Use a soft, lint-free cloth to remove excess moisture from the endoscope and cleaning accessories in preparation for disinfection.
Follow cleaning with appropriate high-level disinfection and alcohol-air drying.