High Voltage Modular Training Set
High Voltage Modular Training Set
Test room
To have a special test room to house the HV 9000 is very important .
This has to be build by the customer in accordance with below guidelines and level of accuracy required
as well as to follow local safety regulations.
For a single stage kit a floor space of 4 x 5 m is recommended, with a height of 2,5-3 m.
Since voltages in excess of 1000 V are generated the safety regulations must be carefully followed.
Fencing and grounding of the Test Room are very important.
Safety equipment
The test area should be enclosed by a metal fence of at least 1.8 m in height and the mesh width not to
exceed 50 mm.
All doors leading into the test room must be equipped with door contacts, which close when the door is
closed. All contacts shall be connected in series and fed to the appropriate sockets on the control desk.
Red and green warning lamps must be installed on all doors leading into the test room. These lamps
should be connected to the control desk.
Earthing (grounding)
Earthing is very important and absolutely necessary for a good test room, and a must for impulse voltage
test equipment.
Connection should be done with low inductance connections ( copper bands 100x0.6mm) which connect
all components of the high-voltage circuit that must be earthed to ONE earth terminal.
Earthing connections should be done without loops. A modern way of doing earthing is to cover the lab
floor with aluminium sheet ( 2mm thickness) which are bolted to the floor and connected internally with
copper band and to earth.
Also the screening surfaces if any, as well as the test objects and the measuring instruments should be
connected to the same earthing point, which should be inside the safety fence and have a lower earthing
resistance than the surrounding building. In any case the earthing resistance should not exceed 2 ohm.
Screening
Screening or shielding as it is also called is used to eliminate all external and internal interference
and can be achieved by building a Faradays cage. The easiest way to obtain this is to hung a close meshed metal net on the walls of the laboratory. The unavoidable apertures for power and
communication leads must be blocked for high -frequency currents with low-pass filters. Special
attention must be paid to careful shielding of doors and windows. The floor should be covered with
non-skid sheet metal. The erection of a complete Faradays cage is necessary only when
extremely sensitive partial discharge measurements are intended. Experiments in a high -voltage
practical course can normally be performed without screening.
DC TEST
1
2
3
IMPULSE
1
2
3
Control Desk
Stages
Basic Components
HV 9103
HV 9105
HV 9106
HV 9107
Discharge Rod
HV 9108
Connecting Rod
HV 9109
Connecting Cup
HV 9110
Floor Pedestal
HV 9111
HV 9112
10
14
13
21
HV Rectifier
HV 9113
Measuring Resistor
HV 9114
Earthing Switch
HV 9119
Spacer Tube
HV 9120
Load Capacitor
HV 9121
Charging Resistor
HV 9122
HV 9123
12
10
HV 9124
Insulating Rod
HV 9125
Sphere Gap
HV 9126
HV 9138
Top Electrode
HV 9139
Electrode 200
HV 9140
Electrode 300
HV 9141
Measuring Capacitor/100
HV 9142
Measuring Capacitor/200
HV 9143
Measuring Capacitor/300
HV 9131
Triggering Device
HV 9132
HV 9150
AC Peak Voltmeter
HV 9151
DC Voltmeter
HV 9152
11
17
1
1
1
1
1
1
1
Test Apparatus
1
HV 9118
HV 9127
Load Resistor
HV 9133
HV 9134
HV 9135
Corona Cage
HV 9137
HV 9144
HV 9146
Capacitor Coupling
HV 9153
TEST SET-UPS
AC Voltage Typical Configurations
Single-stage
Rated no load voltage
kV (rms)
Output (kVA) :
Continuous
1 Hr. on
100
5
10
Three-stage
200
5
10
300
5
10
50
100
25
50
16
32
Impedance voltage
(%) Approx.
Impedance voltage
(%) Approx.
Impedance voltage
(%) Approx.
11
Frequency (Hz)
50/60
Frequency (Hz)
Two-stage
50/60
Frequency (Hz)
50/60
TEST SET-UPS
DC Voltage Typical Configurations
Single-stage
Rated no load voltage kV
Rated current (mA)
Two-stage
140
13
Three-stage
280
10
400
7.5
TEST SET-UPS
Impulse Voltage Typical Configurations
Single-stage
Rated DC charging voltage
Two-stage
140 kV
245 J
490 J
980 J
Three-stage
280 kV Rated DC charging voltage
420 kV
92 %
Technical data
Supply voltage:
230 V 50/60 Hz
Regulating transformer:
Regulating Voltage:
220V/0-230 V
Output:
5kVA Continuous
10 kVA (short time duty 2 min.)
Size:
1220x105x800mm (h,w,d)
Weight:
275 kg
This control desk is used to control and operate high voltage AC/DC/Impulse test equipment.
The desk contains operating and signal elements for the control circuit of the test equipment for warning
and safety
The control desk is made to house the measuring instruments ( Peak, Impulse and DC Voltmeter) and
also the Trigger Device.
Its fabricated of steel and stands on four wheels.
2x220V/100kV/220V
2x11,4A/50mA/15.2A
Continuous.
Output:
Impedance voltage:
4% approx.
Frequency:
50Hz or 60 Hz
Dimensions
Height:
770 mm
Diameter:
550 mm
Weight:
215 kg
Application.
Test transformer with coupling winding for cascade connection to produce AC, DC and Impulse voltages.
The transformer consist of three windings with insulating shell and top and bottom corona free aluminium
shielding electrodes.
The insulation cylinder is made of epoxy resin with glass fibre reinforcement and coated with anti tracking
varnish.
The exciter winding is divided into two groups viz. 100 %voltage when connected in parallel and 50%
output voltage at rated current when connected in parallel and 50% output voltage at rated current when
connected in series.
The coil is vacuum impregnated and insulated with high quality grade transformer oil.
The third winding known as Coupler Winding is provided for further cascade connection.
Application
HV 9150
Supply voltage
Measuring Range
Weight
Dimensions
:
:
:
:
HV 9152
Supply voltage
:
Measuring Range :
Weight
:
220 V 50 Hz
100-1000kV
3.4 kg
Dimensions
:
:
:
:
HV 9131
:
:
Diameter
Weight
220 V 50 Hz
3.7 kg
142 x 173 x 245
( W X H X D)
Measuring Ranges
(Optional)
HV 9132
Trigger Device
Dimensions
HV 9130
Supply voltage
Measuring Range
Weight
Dimensions
Supply voltage
Weight
HV 9151
220 V 50 Hz
100-1000 / 2 kV
2.8 kg
450 kV
300 kV
150 kV
75 kV
37.5 kV
100 mm
0.7 kg
Application
Impulse capacitor for generation of
impulse voltages. It can also be used
as smoothing capacitor in DC voltage
generation.
HV 9120
Load Capacitor
DC and Impulse voltages :140 kV
Capacitance
:1.2 nF
Weight
:10 kg
HV 9141
Measuring Capacitor
AC voltages
: 100 kV
Capacitance
:100 pF
Weight
:10 kg
of
HV 9142
Measuring Capacitor
AC voltages
: 200 kV
High voltage divider capacitor for
measurement of AC voltages.
Capacitance
: 100 pF
Weight
: (Approx.) 75 kg
HV 9143
Measuring Capacitor
AC voltages
: 300 kV
High voltage divider capacitor for
Capacitance
: 100 pF
measurement of AC voltages.
Weight
: (Approx.) 125 kg
10
HV 9106
Length
HV Connection
: Approx. 0.7 m for 200 kV Flexible metal connection with
Approx. 2.2 m for 300 kV connector for the test transformer
and connecting Cup.For connection of
a multi-stage AC voltage test equipment
with the test transformer.
HV 9114
Impulse Voltage
DC Voltage
: 140 kV
: 140 kV
Service Voltage
: 24 V, 50 / 60 Hz
Weight
HV 9138
6.5 kg
Electrode
Diameter
Weight
300 mm
2 kg
HV 9107
2.5 kg
Floor Pedestal
Aluminum cast
Weight
HV 9108
2.5 m
100
1.5 kg
Connecting cup
Aluminum cast
Weight
HV 9110
Discharge Rod
Length
Discharge Resistance
Weight
HV 9109
1 kg
Connecting Rod
Aluminum tube
Weight
HV 9119
Aluminum tube
Weight
Spacer tube
1 kg
11
12
Measurement of flashover
Impulse setup
13
List of experiments
1. Generation and Measurement of Alternating Voltage
2. Generation and Measurement of Direct Voltages
i.)
ii.)
DC Conductivity
Measurement of Tan delta and Capacitance
14