English
Signet 2839 to 2842 Conductivity Sensors
3-2850.090
3-2850.090
Rev F 06/03 English
WARNING
SAFETY INSTRUCTIONS
1. Depressurize and vent system prior to installation or removal.
2. Conrm chemical compatibility before use.
3. Do not exceed maximum temperature/pressure specications.
4. Wear safety goggles or faceshield during installation/service.
5. Do not alter product construction.
6. When using chemicals or solvents care should be taken and
appropriate eye, face, hand, body, and/or respiratory protection should be used.
Table of Contents
1. Specications
2. Cell constant
selection
3. In-line Installation
4. Submersible or Tank
installation
5. Wiring
6. Maintenance
7. Ordering information
1. Specications
Dimensions
2839: 0.01 cell
2840: 0.1 cell
2841: 1.0 cell
2842: 10.0 cell
38.9 mm/
1.530 in.
38.9 mm/
1.53 in.
)
38.9 mm/
1.53 in.
144 mm/
5.66 in.
140.7 mm/
5.54 in.
109 mm/
4.29 in.
3-2840-1
3-2840-1D
0CR
905
890
886
0-X
885
0
590
0CR
580
285
Shipping Weight:
2839:
2840, 2841, 2842:
Process connection:
-1 and -2 versions:
-1D and -2D versions:
0.34 kg (0.74 lb)
0.30 kg (0.66 lb)
Temperature measurement:
Temperature response time ():
2839: 5 s
2841: 20 s
3-2841-2
3-2841-2D
3-2842-2
3-2842-2D
75
50
25
C
-20 0 20
60
100
140
-4
140
212
284
32 68
Operating temperature/pressure:
(with thread engagement per ANSI B1.20.1)
/4 in. NPT
ISO 7/1-R3/4
Cable (28XX-1X only):
4.6 m (15 ft.) std., 3 cond w/shld 22 AWG,
max length 30 m (100 ft.)
(For resistivity measurements above 10 M or
below 20C, maximum cable length is 25 ft. (7.6 m)
Performance:
Accuracy:
41.9 mm/
1.65 in.
BAR PSI
7 100
CR Electrodes
with DryLoc (-2X)
Quality Standard:
Manufactured under ISO 9001, ISO 14001
CE
26.8 mm/
1.13 in.
3-2841-1
3-2841-1D
3-2842-1
3-2842-1D
3-2840-2
3-2840-2D
General
CR Electrodes
with Cable (-1X)
35.8 mm/
1.41 in.
15.9 mm/
0.625 in.
21.5 mm/
0.85 in.
15.9 mm/
0.625 in.
15.9 mm/
0.625 in.
35.8 mm/
1.41 in.
15.9 mm/
0.625 in.
41.9 mm/
1.65 in.
73.7 mm/
2.90 in.
3-2839-2
3-2839-2D
Compatibility:
112 mm/
4.41 in.
105.9 mm
4.17 in.
All -2X versions
3-2839-1
3-2839-1D
59.3 mm/
2.33 in.
15.9 mm/
0.625 in.
73.7 mm/
2.90 in.
15.9 mm/
0.625 in.
102.7 mm/
4.05 in.
2% of cell value
PT1000
2840:
2842:
10 s
30 s
-1X versions:
-10C to 100C @ 6.9 bar (14F to 212F @ 100 psi)
-10C to 131C @ 2.76 bar (14F to 268F @ 40 psi)
-2X versions:
-10C to 85C @ 6.9 bar (14F to 185F @ 100 psi)
Storage temperature:
-20C to 131C (-4F to 268F)
Wetted materials:
DryLoc connector (-2 versions only):
CPVC
Threaded tting:
PEEK
Insulator:
PEEK
Insulator O-ring (2841, 2842):
FPM
Electrode contacts:
316L stainless steel
PEEK is a trademark of Victrex plc
2. Cell constant selection
2839 (0.01 cell):
2840 (0.1 cell):
2841 (1.0 cell):
2842 (10.0 cell):
0.010 to 100 S (10 k /to 100 M)
1 to 1000 S (1 M to 1 k)
10 to 10,000 S
100 to 200,000 S
+GF+ SIGNET Conductivity/Resistivity Electrode Ranges
Electrode Models
The nominal process value should be near the center of
the range. Ranges below are for use with +GF+ SIGNET
Conductivity instruments:
2842 10.0 cell
2841 1.0 cell
2840 0.1 cell
2839 0.01 cell
0.010s 1s 10s 100s 200s 1000s 10 000s
(10 k)
(100 M)
200 000s 400,000s
Conductivity Range (S)
3. In-line installation for all -1X electrodes
Inspect threads to ensure integrity. Do not install an electrode with damaged threads.
Apply sealant or PTFE tape to threads.
Wetted materials include 316L stainless steel, PEEK and FPM (FPM O-ring inside 2841, 2842).
Check for chemical compatibility before installing electrode.
The -1X electrodes are supplied with 5 m (15 ft.) of cable. It may be extended to a maximum 30 m (100 ft.)
For resistivity measurements above 10 M or below 20C, maximum cable length is 25 ft. (7.6 m)
If the electrode is mounted
vertically in a tee, do not recess
the openings inside the tee.
Mounting upside down may help
prevent air entrapment.
An oversized tee may also be
helpful for inline installations.
The preferred installation for in-line
applications directs ow straight into the
electrode. This conguration dislodges
entrapped air bubbles, and provides the best
continuous sampling of the uid content.
At least 4 threads (ANSI B1.20.1)
must be engaged to provide
pressure capacity per published
specications.
4. Submersible installation for all -1X electrodes
Wetted materials include 316L stainless steel, PEEK (FPM O-ring inside 2841, 2842).
Check for chemical compatibility before installing electrode.
The -1X electrodes are supplied with 15 ft. of cable. It may be extended to a maximum 30 m (100 ft.)
For resistivity measurements above 10 M or below 20C, maximum cable length is 25 ft. (7.6 m)
1.
2.
Feed cable into watertight conduit.
Apply thread sealant to the electrode before threading
conduit onto electrode. Avoid twisting the cable.
Secure cable with conduit or cable gland.
For additional defense against possible accumulation of
condensation at the back seal area of the electrode, ll the
lower 3-4 inches (75-100 mm) of conduit or extension pipe
with a exible sealant such as silicone.
3.
4.
In aerated vessels install the electrode
in a stillwell to prevent air from being
trapped inside the electrode.
3/4 in. NPT or
ISO 7/1-R3/4
Product name
5. Wiring for all -1X electrodes
Do not route electrode cable in conduit containing AC power wiring. Electrical noise may interfere with electrode signal.
Routing electrode cable in grounded metal conduit will help prevent electrical noise and mechanical damage.
Seal cable entry points to prevent moisture damage.
For resistivity measurements above 10 M or below 20C, maximum cable length is 25 ft. (7.6 m)
8850
Conductivity Transmitter
Integral installation detail for all -1X electrodes
3-8052 Integral Kit and 3-9000.392-X Liquid Tight Connector
kit are required. (See Parts and Accessories on back page)
Cut the cable to approx. 15 cm (6 in.)
Strip outer cable cover back 5 cm (2 in.)
Strip each conductor to expose 1 cm (3/8 in.) of bare wire.
Tin each conductor with solder for best results.
locking ring
conduit base
integral
adapter
electrode
All -2X DryLoc electrodes connect directly to the 2850
Conductivity Sensor with no interconnecting wiring.
Consult the 2850 Instruction manual for details.
ProcessPro
To 8850-1
ProPoint
10
Sensr Gnd
(SHIELD)
Iso. Gnd
(BLACK)
SILVER (SHLD)
Temp. IN
(WHITE)
WHITE
RED
Signal IN
(RED)
BLACK
14
Sensr Gnd
(SHIELD)
13
Iso. Gnd
(BLACK)
12
Temp. IN
(WHITE)
11
Signal IN
(RED)
Shld
To 8850-2 and 8850-3
Temp. IN
Iso. Gnd
Signal IN
To 5800CR
Intelek-Pro
To 9050CR
RLY3
ANL2
RLY4
NO
NO
R+
SERIAL
CH2
CH1
T1
SG
9 10 11 12 13
14
R-
Tx GND Rx
S2
T2
S1
To 8860
BLACK
WHITE
RED
SHLD
CH 2
ISO GND
TEMP 1
Ch 1
SHIELD (Gnd)
RED
WHITE
BLACK
CLEAR (SHLD)
SGNL 1
RED
WHITE
BLACK
WHITE
RED
BLACK
SHLD
CH 1
ISO GND
TEMP 1
CLEAR (SHLD)
Ch 2
SHIELD (Gnd)
SGNL 1
Installation for -2X DryLoc electrodes
Installation instructions for the -2 DryLoc version of these electrodes is located in the associated instrument manual.
product name
6. Maintenance
Conductivity electrodes require little maintenance except for periodic cleaning in installations where contaminants are present.
Keep metal surfaces clean and free of coatings.
6.1 Replacement Insulator, 2842 electrode only
2842 electrodes have a removable insulator so the internal
cavity can be cleaned.
After the insulator snaps into position it cannot be removed
without damage.
Order insulator replacement kit 3-2842.390 before attempting
maintenance.
Insulator removal and replacement
A. Thread the screw into the insulator (8-32 thread)
B. Pull to remove the old insulator.
C. Clean any coating or deposits inside the electrode.
D. Insert the new insulator and press into place.
Electrode tip
(side view)
Electrode tip
(end view)
9 mm
0.34 in.
Electrode
14 mm
0.55 in.
11 mm
0.44 in.
Cotton swab
C
Clean internal and external surfaces
Internal contact
25.4 mm/1.0 in.
41 mm/1.6 in.
Insulator
B
8-32 thread for removal
New Insulator
3-2842.390 Insulator replacement kit
D
electrode tip
(side view)
7. Ordering Information
Mfr. Part No.
3-2839-1
3-2839-1D
3-2840-1
3-2840-1D
3-2841-1
3-2841-1D
3-2842-1
3-2842-1D
Code
159 000 921
159 000 923
159 000 786
159 000 788
159 000 790
159 000 792
159 000 794
159 000796
Description
Cell 0.01, 15 ft cable, NPT
Cell 0.01, 15 ft cable, ISO
Cell 0.1, 15 ft cable, NPT
Cell 0.1, 15 ft cable, ISO
Cell 1.0, 15 ft cable, NPT
Cell 1.0, 15 ft cable, ISO
Cell 10.0, 15 ft cable, NPT
Cell 10.0, 15 ft cable, ISO
3-2839-2
3-2839-2D
3-2840-2
3-2840-2D
3-2841-2
3-2841-2D
3-2842-2
3-2842-2D
159 000 922
159 000 924
159 000 787
159 000 789
159 000 791
159 000 793
159 000 795
159 000 797
Cell 0.01, DryLoc, NPT
Cell 0.01, DryLoc, ISO
Cell 0.1, DryLoc, NPT
Cell 0.1, DryLoc, ISO
Cell 1.0, DryLoc, NPT
Cell 1.0, DryLoc, ISO
Cell 10.0, DryLoc, NPT
Cell 10.0, DryLoc, ISO
Parts and Accessories
3-8052
3-9000.392-1
3-9000.392-2
3-2842.390
159 000 188
159 000 839
159 000 841
159 000 925
3-2850-1
3-2850-2
3-2850-3
3-2850-4
159 000 783
159 000 784
159 000 785
159 000 857
Conductivity Sensor, In-line
Conductivity Sensor, In-line w/EasyCal
Conductivity Sensor, Submersible
Conductivity Sensor, Submersible
/4 in. Integral mounting kit
Liquid-tight connector kit, 1 set, 1/2 in. NPT
Liquid-tight connector kit, 1 set, PG 13.5
2842 replacement insulator, PEEK with FPM O-ring
George Fischer Signet Inc., 3401 Aerojet Avenue, El Monte, CA 91731-2882 U.S.A. Tel. (626) 571-2770 Fax (626) 573-2057
For Worldwide Sales and Service, visit our website: www.gfsignet.com Or call (in the U.S.): (800) 854-4090
3-2840.090 Rev. F 06/03 English
George Fischer Signet, Inc. 2003
Printed in U.S.A. on recycled paper