The document provides technical specifications for the Monnal T75 ventilator from Air Liquide Medical Systems, including physical dimensions, standards compliance, operating conditions, monitoring capabilities, and parameter settings. It contains detailed information on the ventilator's pneumatic, electrical, and monitoring specifications.

Uploaded by

ekawahono
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
100% found this document useful (1 vote)
928 views9 pages

Monnal T75 Especificaciones

The document provides technical specifications for the Monnal T75 ventilator from Air Liquide Medical Systems, including physical dimensions, standards compliance, operating conditions, monitoring capabilities, and parameter settings. It contains detailed information on the ventilator's pneumatic, electrical, and monitoring specifications.

Uploaded by

ekawahono
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 9

DATA SHEET

www.airliquidemedicalsystems.com
www.airliquide.com

[1] Monnal T75 / Air Liquide Medical Systems

TECHNICAL SPECIFICATIONS

APPLICATION AREA
___________________________________________
Application:

Intensive care respirator designed for


resuscitation, intensive care and the
postoperative room

Patient
categories:

Adult, Child and Infant

Weight:

3 to 250 kg

STANDARDS AND DIRECTIVES


___________________________________________
CE Mark:

CE 0459. The device complies with the


European requirements in the Directive
93/42/EEC concerning medical devices.

Medical
classification:

Class IIb equipment, according to th


Directive 93/42/EEC.

Standards:

CEI 60601-1
CEI 60601-2-12

Electromagnetic
compatibility
(EMC):

Complying with the CEI 60601-1-2 (2001)

First launch:

France: end of 2008


Other countries: please contact us

Available
Languages

French, English, Italian, German, Russia,


Spanish, Lithuania, Polish, Czech,
Portuguese, Chinese, Japan
Other language : please contact us

PHYSICAL SPECIFICATIONS
___________________________________________
Size:

H 35 x W 30 x D 40 cm

Ventilator with
trolley:

H 135 x W 55 x D 60 cm

Weight:

16 kg - Ventilator
31.8 kg - Ventilator and trolley

Noise level:

48 dB (A) at 1m

OPERATING CONDITIONS
___________________________________________
Operating
temperature:

+10 to +40C (+50 to +104F)

Relative
humidity:

0 to 90 % (at 40C without condensation)

Atmospheric
pressure:

700 (3075 m) to 1060 mbar


Altitude effect compensation

STORAGE CONDITIONS
___________________________________________
Storage
temperature:

-20 to +70C (-4 to +158F)

Storage relative
humidity:

0 to 90 % (at 40C without condensation)

Storage
atmospheric
pressure:

500 (5550 meters) to 1060 mbar

[2] Monnal T75 / Air Liquide Medical Systems

TECHNICAL SPECIFICATIONS

POWER SUPPLY
___________________________________________

PATIENT SYSTEM GAS CONNECTORS


___________________________________________

Power supply:

100-240 VAC 5 %

22 mm male

Frequency:

50 60 Hz

Inspiratory
breathing tube
connector:
Expiratory
breathing tube
connector:

22 mm male

Electrical power
consumed:

250 VA

Electrical class :

Type :

SCREEN
___________________________________________

Connection
Standard
Available:

EU, AFS, JP, AR, BR


Please contact us for more information
concerning other standards

Type:

Flat color touch screen


TFT-LCD Module

Technology:

Resistive analogical technology

20 to 30 V DC

Size:

10.4 inch

Resolution :

640 x 480 pixels

Continuous
external supply:

If no mains are available, switch to external battery


External battery
Type:

Rechargeable, NiMH, 24V, 2x4500 mAh

Autonomy:

2.5 to 3 hours in standard ventilation

Charge time:

12 hours

Interval between
two recharges:

It is advised to perform one complete


discharge cycle per year

If neither the mains nor the external source is


available, switch to internal battery.

Internal battery
Type:

Rechargeable, NiMH, 24V, 2x4500 mAh

Autonomy:

2.5 to 3 hours in standard ventilation

Charge time:

12 hours

Interval between
two recharges:

It is advised to perform one complete


discharge cycle per year

PNEUMATIC SUPPLY
___________________________________________
O2 pneumatic supply
High pressure:

2.8 6 bar / 280 600 kPa / 40 86 psi


Operating possible starting from 1.5 bar

Connection
standards
available:

NF, NIST, DISS


For other standards, please contact us

Low pressure:

0 1.5 bar / 0 150 kPa / 0 21 psi

Ambiant Air
Inlet:

Equipped with HEPA filter Monnal CleanIn


(High Efficiency Particulate Air Filter)

[3] Monnal T75 / Air Liquide Medical Systems

TECHNICAL SPECIFICATIONS

PNEUMATIC SPECIFICATIONS
___________________________________________

MODES INVASIVE VENTILATION


__________________________________________
_

Method of
triggering:

Flow and pressure

PCV

Pressure controlled ventilation or Assist


pressure controlled ventilation

Max. operating
pressure:

90 cmH2O

VCV

Assist volume controlled ventilation

PRVC
Bias flow:

3 L/min (flow by)

Pressure regulated volume controlled


ventilation

Peak inspiratory
flow:

>175 L/min

SIMV

Synchronized intermittent mandatory


ventilation

PSIMV

Pressure support intermittent mandatory


ventilation

PSV

Spontaneous ventilation with pressure


support and PEEP

CPAP

Continuous Positive Airway Pressure

INSPIRATORY CHANNEL
___________________________________________

Pressure drop:

Max. 4 cm H2O at a flow rate of 1L/s


(without any filter)

Gas delivery
system:

Microprocessor controlled valves and


turbine

Duo-Levels

Alternation between two levels of CPAP

Inspiratory flow
range :

0 to 3.3 L/s

PS-Pro

Spontaneous ventilation with inspiratory


pressure support, PEEP and variable
frequencies

Safety
ventilation

Apnea ventilation (VT, RR and T apnea


can be adjusted, constant flow rate,
Ti/Ttot 33%)

EXPIRATORY CHANNEL
___________________________________________

Pressure drop:

Max. 3 cm H2O at a flow rate of 1L/s

Expiratory flow
range:

0 to 3.3 L/s

MODES NON INVASIVE VENTILATION


_________________________________________

PSV-VNI

Non-invasive spontaneous ventilation


with pressure support and PEEP

CPAP

Continuous Positive Pressure

Duo-Levels

Two-level pressure ventilation

[4] Monnal T75 / Air Liquide Medical Systems

TECHNICAL SPECIFICATIONS

PARAMETER SETTINGS
___________________________________________
PARAMETERS:
______________________

SETTINGS RANGE:
_______________________

Inspiratory tidal volume


(VT):

20 to 2000 mL

PARAMETER SETTINGS (NEXT)


___________________________________________
PARAMETERS:
______________________

SETTINGS RANGE:
________________________

Inspiratory plateau (Tplat):

0 to 60 % of TI
Adult: 0 to 60 % of TI
Child: 0 to 40 % of TI

Adult: 100 to 2000 mL

Infant: 0 to 40 % of TI
Child: 50 to 500 mL
Infant: 20 to 75 mL

Inspiratory ratio (TI/Ttot):

3 to 80 %
Adult: 3 to 80 %

Respiratory rate (RR):

4 to 120 Bpm
Child: 5 to 60 %
Adult: 4 to 80 Bpm

Infant: 5 to 60 %

Child: 5 to 120 Bpm


Infant: 10 to 120 Bpm

Inspiratory time (I/E):

1/.5 (=2) to 1/19


Adult: 1/1.5 to 1/19

SIMV respiratory rate:

1 to 120 Bpm

Child: 1/1.5 to 1/19


Infant: 1/0.7 to 1/19

Adult: 1 to 80 Bpm
Child: 1 to 120 Bpm

Inspiratory time (TI):

0.2 to 10 s

Child: 1 to 120 Bpm


Adult: 0.2 to 10 s

Minimum frequency:

Child: 0.2 to 8 s

1 to 100 Bpm

Infant: 0.2 to 3 s
Adult: 1 to 80 Bpm

Child: 1 to 100 Bpm

Maximum inspiratory time in 0.3 to 3.5 s


spontaneous modes (TI
max.):

Infant: 1 to 100 Bpm


Adult: 0.3 to 5 s

Apnea time:

Adjustable VCV mode

Child: 0.3 to 3.3 s


Infant: 0.2 to 2.5 s

Adult: 15 to 60 s
Child: 10 to 60 s
Infant: 2 to 30 s

[5] Monnal T75 / Air Liquide Medical Systems

TECHNICAL SPECIFICATIONS

PARAMETER SETTINGS (NEXT)


___________________________________________
PARAMETERS:
______________________

SETTINGS RANGE:
________________________

Positive end of expiration


pressure (PEEP):

0 to 50 cmH2O

PARAMETER SETTINGS (NEXT)


___________________________________________
PARAMETERS:
______________________

SETTINGS RANGE:
________________________

Inspiratory plateau:

0 to 15 s

Expiratory plateau:

0 to 15 s

Sigh frequency:

1 sigh every 9 to 200 cycles


1 to 2 per VT

Pressure support (PS) :

2 to 40 cmH2O

Tidal volume of sigh:

Insufflation pressure (PI) :

2 to 99 cmH20

Insufflation pressure of sigh: 1 to 2 per PI


Infant: 0 to 90%

Inspiratory slope:

20 to 200 cmH2O/s

Flow rate shape:

Constant, decelerated

O2 Concentration (FiO2):

21 to 100 %

Inspiratory flow trigger:

1 to 10 L/min

Inspiratory trigger in
pressure:

Expiratory trigger:

Always activated and


automatically set in function of
the flow trigger setting

0 (OFF) to 90 % of inspiratory
peak flow

PARAMETER SETTINGS: SPECIAL FEATURES


___________________________________________
NON INVASIVE
VENTILATION
______________________
Positive end of expiration
pressure (PEEP) :

0 to 15 cmH2O

Pressure Support (PS) :

2 to 25 cmH2O

PRCV MODE
______________________
Tidal Volume Target

OFF to maximum of each


patient categories

Insufflation pressure (PI)

AUTO, 2 to 99 cmH2O

Maximum Insufflation
pressure (PImax)

2 to 99 cmH2O

PS-Pro MODE
______________________
Support frequency
(RRsupport)

Adult: 4 to 99 Bpm
Child: 5 to 120 Bpm
Infant: 5 to 120 Bpm

Pressure Support (PS)

AUTO, 2 to 40 cmH2O

Maximum Insufflation
pressure (PImax)

2 to 99 cmH2O

Tidal Volume Target

OFF to maximum of each


patient categories

O2 : INTELLIGENT
SUCTION
______________________
Pre-oxygenation time:

Adjustable from 30 to 900 s

Post-oxygenation time:

Adjustable from 30 to 300 s

FiO2 Target:

21% to 100%

[6] Monnal T75 / Air Liquide Medical Systems

TECHNICAL SPECIFICATIONS

MONITORING
___________________________________________

LOOPS AND WAVEFORMS PRESENTATION


___________________________________________

MEASURED PARAMETERS:

Real-time curves: Pressure curve

______________________

MEASURE
RANGE:
____________

TREND
VALUE*:
__________

Flow curve
Loop curves:

Pressure / Volume

Inspired minute volume


(MVe):

0 to 99 L/min

Yes

Expired tidal volume (Vte):

0 to 5000 mL

Yes

Insufflated tidal volume


(VTi):

0 to 5000 mL

Yes

Respiratory rate (RR):

0 to 120 Bpm

Yes

Peak airway pressure


(Ppeak):

0 to 120 cmH2O

Yes

Positive end of expiration


pressure (PEEP) :

0 to 99 cmH2O

Yes

Plateau pressure (Pplat) :

0 to 99 cmH2O

Yes

FiO2

15 to 100 %

Yes

0 to 99 cmH2O

Yes

Ti/Ttot

0 to 99%

Yes

I/E

1/20 to 1/0,3

Yes

Mean pressure (Pmean)

0 to 99 cmH2O

Yes

RR/Vte

0 to 5000 pm/L

Yes

MVe Spont

0 to 99 L/min

Yes

RR Spont

0 to 120 bpm

Yes

Vpeak insp**

0 to 200 L/min

Yes

Vpeak exp

0 to 200 L/min

Yes

Vleak (in NIV)

0 to 150 L/min

Yes

%spont

-- to 100

Yes

COMMUNICATION / INTERFACE :
___________________________________________

%leak

-- to 100

Yes

Serial port:

RS 232 port (X 2)

P0.1**

Under dev.

Yes

OTP

Open Taema Protocol

Rstat**

Under dev.

Yes

Video output

Screen report to external monitor

Cstat**

Under dev.

Yes

BOW MEDICAL
interface

Communication with DIANE anesthesia


sheet

Auto-Peep**

Under dev.

Yes

DATACAPTOR

Capsuletech

CLINISOFT
Interface

Communication with the GE data


collection software

Volume / Flow
Flow / Pressure

Mean pressure

* Stored trend values for up to 80 hours


** Only in Invasive Ventilation

CONFIGURATION
___________________________________________
Selection of
ventilation mode

Enabled, Disabled (out of VCV)

Alarms

Save alarms current settings

Settings

Save current ventilation settings

Volume setting
alarm

25 to 100 of alarm custom

Minimum
frequency

Enabled, Disabled (only in PS and NIV


PS)

Setting in VCV
mode

Ti/Ttot, I/E, Ti, Flow

Setting in PCV
mode

Ti/Ttot, I/E, Ti

Startup Mode

Possibility of configuring first default


ventilation mode

EVENT HISTORY
___________________________________________
Alarms history :

Alarms triggered
Technical messages with error codes

[7] Monnal T75 / Air Liquide Medical Systems

TECHNICAL SPECIFICATIONS

ALARM SETTINGS
___________________________________________

ALARM SYSTEMS
___________________________________________

RR low

Visual alarms
classified by
priority:

3 levels of priority with specific colours.


1 level of information

Sound alarms
classified by
priority :

3 levels of priority with specific melodies

Technical alarms:

Categorized by severity

Supply alarms:

Default O2 supply

RR low

Adult: 1 to 70 Bpm
Child: 1 to 110 Bpm
Infant: 1 to 110 Bpm
Adult: 11 to 80 Bpm
Child: 11 to 120 Bpm
Infant: 11 to 120 Bpm

Ppeak high

10 to 100 cmH2O

Pplat high

OFF, 1 to 50 cmH2O

Vti low

Adult: OFF to 2900 mL


Child: OFF to 1900 mL
Infant: OFF to 500 mL

Switch over to internal battery


Switch over to external battery
Internal battery exhausted

Vti high

Vte low

Vte high

VMe high

Adult: 100 to 3000 mL


Child: 100 to 2000 mL
Infant: 100 to 800 mL
Adult: OFF to 2000 mL
Child: OFF to 2000 mL
Infant: OFF to 500 mL
Adult: 10 to 3000 mL
Child: 10 to 3000 mL
Infant: 10 to 800 mL

Internal battery power low


Internal battery absent
Alarm indicating total loss of electrical
power
Automatic
thresholds:

Direct access key

Silence:

2 minutes (fixed)

Apnea Ventilation:

Adjustable VCV mode

O2 Boost/ Suction: Automatic ventilation pause and alarm


inhibition during suction
Non-adjustable
patient alarms:

PEEP> PEEP set-point + 5 cmH2O

Adult: 1.5 to 40 L/min


Child: 1.5 to 40 L/min
Infant: 1 to 40 L/min

VMe low

0.5 to 39 L/min

FiO2 low

18 to 95 %

FiO2 high

23 to 100 %

Disconnection

Active patient. Increase inspiratory flow


rate
CLEANING AND STERILIZATION
___________________________________________
Autoclavable:

Reusable expiratory valve


Patient circuit

ALARM SETTINGS : SPECIAL FEATURES OF NIV


___________________________________________
VMe low

OFF to 39 L/min

VMe high

Adult: 1.5 to 60 L/min


Child: 1.1 to 60 L/min
Infant: 1 to 40 L/min

Reusable spirometry sensor


Single Use:

Single use expiratory valve


Patient circuit
Single use spirometry sensor

Turbine inlet:

Monnal Clean In filter (quarterly


inspection)

MAINTENANCE
___________________________________________
Periodic
maintenance:

Annual, performance check

Credit Photo Le Square

Air Liquide Medical Systems


Parc de Haute Technologie
6, rue Georges Besse CE 80
92 182 ANTONY Cedex
France
Tl. : +33 (0)1 40 96 66 00
Fax : +33 (0)1 40 96 67 00

Air Liquide Medical Systems reserves the right to modify the design and the specifications contained herein without prior notice

Contacts