DATA SHEET
www.airliquidemedicalsystems.com
www.airliquide.com
[1] Monnal T75 / Air Liquide Medical Systems
TECHNICAL SPECIFICATIONS
APPLICATION AREA
___________________________________________
Application:
Intensive care respirator designed for
resuscitation, intensive care and the
postoperative room
Patient
categories:
Adult, Child and Infant
Weight:
3 to 250 kg
STANDARDS AND DIRECTIVES
___________________________________________
CE Mark:
CE 0459. The device complies with the
European requirements in the Directive
93/42/EEC concerning medical devices.
Medical
classification:
Class IIb equipment, according to th
Directive 93/42/EEC.
Standards:
CEI 60601-1
CEI 60601-2-12
Electromagnetic
compatibility
(EMC):
Complying with the CEI 60601-1-2 (2001)
First launch:
France: end of 2008
Other countries: please contact us
Available
Languages
French, English, Italian, German, Russia,
Spanish, Lithuania, Polish, Czech,
Portuguese, Chinese, Japan
Other language : please contact us
PHYSICAL SPECIFICATIONS
___________________________________________
Size:
H 35 x W 30 x D 40 cm
Ventilator with
trolley:
H 135 x W 55 x D 60 cm
Weight:
16 kg - Ventilator
31.8 kg - Ventilator and trolley
Noise level:
48 dB (A) at 1m
OPERATING CONDITIONS
___________________________________________
Operating
temperature:
+10 to +40C (+50 to +104F)
Relative
humidity:
0 to 90 % (at 40C without condensation)
Atmospheric
pressure:
700 (3075 m) to 1060 mbar
Altitude effect compensation
STORAGE CONDITIONS
___________________________________________
Storage
temperature:
-20 to +70C (-4 to +158F)
Storage relative
humidity:
0 to 90 % (at 40C without condensation)
Storage
atmospheric
pressure:
500 (5550 meters) to 1060 mbar
[2] Monnal T75 / Air Liquide Medical Systems
TECHNICAL SPECIFICATIONS
POWER SUPPLY
___________________________________________
PATIENT SYSTEM GAS CONNECTORS
___________________________________________
Power supply:
100-240 VAC 5 %
22 mm male
Frequency:
50 60 Hz
Inspiratory
breathing tube
connector:
Expiratory
breathing tube
connector:
22 mm male
Electrical power
consumed:
250 VA
Electrical class :
Type :
SCREEN
___________________________________________
Connection
Standard
Available:
EU, AFS, JP, AR, BR
Please contact us for more information
concerning other standards
Type:
Flat color touch screen
TFT-LCD Module
Technology:
Resistive analogical technology
20 to 30 V DC
Size:
10.4 inch
Resolution :
640 x 480 pixels
Continuous
external supply:
If no mains are available, switch to external battery
External battery
Type:
Rechargeable, NiMH, 24V, 2x4500 mAh
Autonomy:
2.5 to 3 hours in standard ventilation
Charge time:
12 hours
Interval between
two recharges:
It is advised to perform one complete
discharge cycle per year
If neither the mains nor the external source is
available, switch to internal battery.
Internal battery
Type:
Rechargeable, NiMH, 24V, 2x4500 mAh
Autonomy:
2.5 to 3 hours in standard ventilation
Charge time:
12 hours
Interval between
two recharges:
It is advised to perform one complete
discharge cycle per year
PNEUMATIC SUPPLY
___________________________________________
O2 pneumatic supply
High pressure:
2.8 6 bar / 280 600 kPa / 40 86 psi
Operating possible starting from 1.5 bar
Connection
standards
available:
NF, NIST, DISS
For other standards, please contact us
Low pressure:
0 1.5 bar / 0 150 kPa / 0 21 psi
Ambiant Air
Inlet:
Equipped with HEPA filter Monnal CleanIn
(High Efficiency Particulate Air Filter)
[3] Monnal T75 / Air Liquide Medical Systems
TECHNICAL SPECIFICATIONS
PNEUMATIC SPECIFICATIONS
___________________________________________
MODES INVASIVE VENTILATION
__________________________________________
_
Method of
triggering:
Flow and pressure
PCV
Pressure controlled ventilation or Assist
pressure controlled ventilation
Max. operating
pressure:
90 cmH2O
VCV
Assist volume controlled ventilation
PRVC
Bias flow:
3 L/min (flow by)
Pressure regulated volume controlled
ventilation
Peak inspiratory
flow:
>175 L/min
SIMV
Synchronized intermittent mandatory
ventilation
PSIMV
Pressure support intermittent mandatory
ventilation
PSV
Spontaneous ventilation with pressure
support and PEEP
CPAP
Continuous Positive Airway Pressure
INSPIRATORY CHANNEL
___________________________________________
Pressure drop:
Max. 4 cm H2O at a flow rate of 1L/s
(without any filter)
Gas delivery
system:
Microprocessor controlled valves and
turbine
Duo-Levels
Alternation between two levels of CPAP
Inspiratory flow
range :
0 to 3.3 L/s
PS-Pro
Spontaneous ventilation with inspiratory
pressure support, PEEP and variable
frequencies
Safety
ventilation
Apnea ventilation (VT, RR and T apnea
can be adjusted, constant flow rate,
Ti/Ttot 33%)
EXPIRATORY CHANNEL
___________________________________________
Pressure drop:
Max. 3 cm H2O at a flow rate of 1L/s
Expiratory flow
range:
0 to 3.3 L/s
MODES NON INVASIVE VENTILATION
_________________________________________
PSV-VNI
Non-invasive spontaneous ventilation
with pressure support and PEEP
CPAP
Continuous Positive Pressure
Duo-Levels
Two-level pressure ventilation
[4] Monnal T75 / Air Liquide Medical Systems
TECHNICAL SPECIFICATIONS
PARAMETER SETTINGS
___________________________________________
PARAMETERS:
______________________
SETTINGS RANGE:
_______________________
Inspiratory tidal volume
(VT):
20 to 2000 mL
PARAMETER SETTINGS (NEXT)
___________________________________________
PARAMETERS:
______________________
SETTINGS RANGE:
________________________
Inspiratory plateau (Tplat):
0 to 60 % of TI
Adult: 0 to 60 % of TI
Child: 0 to 40 % of TI
Adult: 100 to 2000 mL
Infant: 0 to 40 % of TI
Child: 50 to 500 mL
Infant: 20 to 75 mL
Inspiratory ratio (TI/Ttot):
3 to 80 %
Adult: 3 to 80 %
Respiratory rate (RR):
4 to 120 Bpm
Child: 5 to 60 %
Adult: 4 to 80 Bpm
Infant: 5 to 60 %
Child: 5 to 120 Bpm
Infant: 10 to 120 Bpm
Inspiratory time (I/E):
1/.5 (=2) to 1/19
Adult: 1/1.5 to 1/19
SIMV respiratory rate:
1 to 120 Bpm
Child: 1/1.5 to 1/19
Infant: 1/0.7 to 1/19
Adult: 1 to 80 Bpm
Child: 1 to 120 Bpm
Inspiratory time (TI):
0.2 to 10 s
Child: 1 to 120 Bpm
Adult: 0.2 to 10 s
Minimum frequency:
Child: 0.2 to 8 s
1 to 100 Bpm
Infant: 0.2 to 3 s
Adult: 1 to 80 Bpm
Child: 1 to 100 Bpm
Maximum inspiratory time in 0.3 to 3.5 s
spontaneous modes (TI
max.):
Infant: 1 to 100 Bpm
Adult: 0.3 to 5 s
Apnea time:
Adjustable VCV mode
Child: 0.3 to 3.3 s
Infant: 0.2 to 2.5 s
Adult: 15 to 60 s
Child: 10 to 60 s
Infant: 2 to 30 s
[5] Monnal T75 / Air Liquide Medical Systems
TECHNICAL SPECIFICATIONS
PARAMETER SETTINGS (NEXT)
___________________________________________
PARAMETERS:
______________________
SETTINGS RANGE:
________________________
Positive end of expiration
pressure (PEEP):
0 to 50 cmH2O
PARAMETER SETTINGS (NEXT)
___________________________________________
PARAMETERS:
______________________
SETTINGS RANGE:
________________________
Inspiratory plateau:
0 to 15 s
Expiratory plateau:
0 to 15 s
Sigh frequency:
1 sigh every 9 to 200 cycles
1 to 2 per VT
Pressure support (PS) :
2 to 40 cmH2O
Tidal volume of sigh:
Insufflation pressure (PI) :
2 to 99 cmH20
Insufflation pressure of sigh: 1 to 2 per PI
Infant: 0 to 90%
Inspiratory slope:
20 to 200 cmH2O/s
Flow rate shape:
Constant, decelerated
O2 Concentration (FiO2):
21 to 100 %
Inspiratory flow trigger:
1 to 10 L/min
Inspiratory trigger in
pressure:
Expiratory trigger:
Always activated and
automatically set in function of
the flow trigger setting
0 (OFF) to 90 % of inspiratory
peak flow
PARAMETER SETTINGS: SPECIAL FEATURES
___________________________________________
NON INVASIVE
VENTILATION
______________________
Positive end of expiration
pressure (PEEP) :
0 to 15 cmH2O
Pressure Support (PS) :
2 to 25 cmH2O
PRCV MODE
______________________
Tidal Volume Target
OFF to maximum of each
patient categories
Insufflation pressure (PI)
AUTO, 2 to 99 cmH2O
Maximum Insufflation
pressure (PImax)
2 to 99 cmH2O
PS-Pro MODE
______________________
Support frequency
(RRsupport)
Adult: 4 to 99 Bpm
Child: 5 to 120 Bpm
Infant: 5 to 120 Bpm
Pressure Support (PS)
AUTO, 2 to 40 cmH2O
Maximum Insufflation
pressure (PImax)
2 to 99 cmH2O
Tidal Volume Target
OFF to maximum of each
patient categories
O2 : INTELLIGENT
SUCTION
______________________
Pre-oxygenation time:
Adjustable from 30 to 900 s
Post-oxygenation time:
Adjustable from 30 to 300 s
FiO2 Target:
21% to 100%
[6] Monnal T75 / Air Liquide Medical Systems
TECHNICAL SPECIFICATIONS
MONITORING
___________________________________________
LOOPS AND WAVEFORMS PRESENTATION
___________________________________________
MEASURED PARAMETERS:
Real-time curves: Pressure curve
______________________
MEASURE
RANGE:
____________
TREND
VALUE*:
__________
Flow curve
Loop curves:
Pressure / Volume
Inspired minute volume
(MVe):
0 to 99 L/min
Yes
Expired tidal volume (Vte):
0 to 5000 mL
Yes
Insufflated tidal volume
(VTi):
0 to 5000 mL
Yes
Respiratory rate (RR):
0 to 120 Bpm
Yes
Peak airway pressure
(Ppeak):
0 to 120 cmH2O
Yes
Positive end of expiration
pressure (PEEP) :
0 to 99 cmH2O
Yes
Plateau pressure (Pplat) :
0 to 99 cmH2O
Yes
FiO2
15 to 100 %
Yes
0 to 99 cmH2O
Yes
Ti/Ttot
0 to 99%
Yes
I/E
1/20 to 1/0,3
Yes
Mean pressure (Pmean)
0 to 99 cmH2O
Yes
RR/Vte
0 to 5000 pm/L
Yes
MVe Spont
0 to 99 L/min
Yes
RR Spont
0 to 120 bpm
Yes
Vpeak insp**
0 to 200 L/min
Yes
Vpeak exp
0 to 200 L/min
Yes
Vleak (in NIV)
0 to 150 L/min
Yes
%spont
-- to 100
Yes
COMMUNICATION / INTERFACE :
___________________________________________
%leak
-- to 100
Yes
Serial port:
RS 232 port (X 2)
P0.1**
Under dev.
Yes
OTP
Open Taema Protocol
Rstat**
Under dev.
Yes
Video output
Screen report to external monitor
Cstat**
Under dev.
Yes
BOW MEDICAL
interface
Communication with DIANE anesthesia
sheet
Auto-Peep**
Under dev.
Yes
DATACAPTOR
Capsuletech
CLINISOFT
Interface
Communication with the GE data
collection software
Volume / Flow
Flow / Pressure
Mean pressure
* Stored trend values for up to 80 hours
** Only in Invasive Ventilation
CONFIGURATION
___________________________________________
Selection of
ventilation mode
Enabled, Disabled (out of VCV)
Alarms
Save alarms current settings
Settings
Save current ventilation settings
Volume setting
alarm
25 to 100 of alarm custom
Minimum
frequency
Enabled, Disabled (only in PS and NIV
PS)
Setting in VCV
mode
Ti/Ttot, I/E, Ti, Flow
Setting in PCV
mode
Ti/Ttot, I/E, Ti
Startup Mode
Possibility of configuring first default
ventilation mode
EVENT HISTORY
___________________________________________
Alarms history :
Alarms triggered
Technical messages with error codes
[7] Monnal T75 / Air Liquide Medical Systems
TECHNICAL SPECIFICATIONS
ALARM SETTINGS
___________________________________________
ALARM SYSTEMS
___________________________________________
RR low
Visual alarms
classified by
priority:
3 levels of priority with specific colours.
1 level of information
Sound alarms
classified by
priority :
3 levels of priority with specific melodies
Technical alarms:
Categorized by severity
Supply alarms:
Default O2 supply
RR low
Adult: 1 to 70 Bpm
Child: 1 to 110 Bpm
Infant: 1 to 110 Bpm
Adult: 11 to 80 Bpm
Child: 11 to 120 Bpm
Infant: 11 to 120 Bpm
Ppeak high
10 to 100 cmH2O
Pplat high
OFF, 1 to 50 cmH2O
Vti low
Adult: OFF to 2900 mL
Child: OFF to 1900 mL
Infant: OFF to 500 mL
Switch over to internal battery
Switch over to external battery
Internal battery exhausted
Vti high
Vte low
Vte high
VMe high
Adult: 100 to 3000 mL
Child: 100 to 2000 mL
Infant: 100 to 800 mL
Adult: OFF to 2000 mL
Child: OFF to 2000 mL
Infant: OFF to 500 mL
Adult: 10 to 3000 mL
Child: 10 to 3000 mL
Infant: 10 to 800 mL
Internal battery power low
Internal battery absent
Alarm indicating total loss of electrical
power
Automatic
thresholds:
Direct access key
Silence:
2 minutes (fixed)
Apnea Ventilation:
Adjustable VCV mode
O2 Boost/ Suction: Automatic ventilation pause and alarm
inhibition during suction
Non-adjustable
patient alarms:
PEEP> PEEP set-point + 5 cmH2O
Adult: 1.5 to 40 L/min
Child: 1.5 to 40 L/min
Infant: 1 to 40 L/min
VMe low
0.5 to 39 L/min
FiO2 low
18 to 95 %
FiO2 high
23 to 100 %
Disconnection
Active patient. Increase inspiratory flow
rate
CLEANING AND STERILIZATION
___________________________________________
Autoclavable:
Reusable expiratory valve
Patient circuit
ALARM SETTINGS : SPECIAL FEATURES OF NIV
___________________________________________
VMe low
OFF to 39 L/min
VMe high
Adult: 1.5 to 60 L/min
Child: 1.1 to 60 L/min
Infant: 1 to 40 L/min
Reusable spirometry sensor
Single Use:
Single use expiratory valve
Patient circuit
Single use spirometry sensor
Turbine inlet:
Monnal Clean In filter (quarterly
inspection)
MAINTENANCE
___________________________________________
Periodic
maintenance:
Annual, performance check
Credit Photo Le Square
Air Liquide Medical Systems
Parc de Haute Technologie
6, rue Georges Besse CE 80
92 182 ANTONY Cedex
France
Tl. : +33 (0)1 40 96 66 00
Fax : +33 (0)1 40 96 67 00
Air Liquide Medical Systems reserves the right to modify the design and the specifications contained herein without prior notice
Contacts