100% found this document useful (2 votes)
389 views21 pages

User Manul of ECG03C

manual de electrocardiograma

Uploaded by

joan
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
100% found this document useful (2 votes)
389 views21 pages

User Manul of ECG03C

manual de electrocardiograma

Uploaded by

joan
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 21

ECG Machine

Model PRO-ECG03

User Manual

User Manual of ECG Machine

Statement
In order that you can use this product more safely and get full play to its functions and features, please
read the operating instructions carefully before using this unit and keep it properly.
Our company owns all rights to this unpublished work and intends to maintain this work as confidential. We
may also seek to maintain this work as an unpublished copyright. This publication is to be used solely for
the purpose of reference, operation, maintenance, or repair of our equipment. No part of this can be
disseminated for other purposes.
In the event of inadvertent or deliberate publication, our company intends to enforce its rights to this work
under copyright laws as a published work. Those having access to this work may not copy, use, or disclose
the information in this work unless expressly authorized by us to do so.
All information contained in this publication is believed to be correct. Our company shall not be liable for
consequential damages in connection with the furnishing, performance, or use of this material. This
publication may refer to information and protected by copyrights or patents and does not convey any
license under the patent rights of our company, nor the rights of others. Our company does not assume
any liability for arising out of any infringements of patents or other rights of the third parties.
The content of this manual is subject to change without prior notice.

User Manual of ECG Machine

Contents
Chapter 1 Overview........................................................................................................................................... 3
Chapter 2 Security Notice ................................................................................................................................ 3
Chapter 3 Maintenance Regulation ................................................................................................................ 4
Chapter 4 Instrument Characteristics ............................................................................................................. 4
Chapter 5 Panel Sketch Map ........................................................................................................................... 6
Chapter 6 Attention before Operating ............................................................................................................. 8
Chapter 7 Preparation Work before Instrumentation ................................................................................... 8
Chapter 8 Precaution during Operation ......................................................................................................... 8
Chapter 9 Electrode Placement ...................................................................................................................... 9
Chapter 10 Grounding and Power Connection of Instrument ................................................................... 10
Chapter 11 Precaution for Battery Operation ...............................................................................................11
Chapter 12 Control Panel and Key Instruction............................................................................................ 12
Chapter 13 Troubleshooting........................................................................................................................... 17
Chapter 14 Main Technical Specification ..................................................................................................... 19

User Manual of ECG Machine

Chapter 1 Overview
This ECG machine is such a kind of electrocardiograph, which samples 12 leads ECG signals
simultaneously and prints out the ECG waveforms with thermal printing system. Its functions are as follows:
recording and displaying ECG waveforms in auto/manual mode; auto-measuring and auto-diagnosing the
ECG wave parameters; save patient data into SD memory card automatically, prompting the state of lead
off; switching the interface languages (Chinese/English).
Our companys three channel ECG machine includes 192*64 LCD screen, 3.8 inches LCD screen and 5.7
inches LCD screen. This user manual takes 5.7 inches LCD screen for example. They have the same use
method and function

Chapter 2 Security Notice


2.1 The power supply should be grounded properly before operation.
2.2 If the ground cable is not integrated, the device must run with built-in power supply.
2.3 Please pull out power supply plug before changing the fuse.
2.4 This device must be operated and preserved by professional personnel.
2.5 The operator must read this user manual carefully before operation, and operate the device according
to operation regulation strictly.
2.6 The design of this device has fully considered security, but operators should never neglect to device
state and patients situation.
2.7 Please turn off the instrument and pull out power supply plug before cleaning and disinfection.
2.8 Please dont operate this device in environment which contains flammable anesthesia gas.
2.9 If this device is used with cardiac defibrillator or other electric stimulate devices at the same time,
please choose Ag/AgCl chloride chest electrode and ECG lead with prevent-fibrillation function. To prevent
the metal electrode burn patients' skin, the disposable chest electrode should be used if the defibrillation
time is over 55 seconds. It is better that do not use this device with other electric stimulate devices at the
same time. If it must be used at the same time, there must be professional technician guide on the scene.
2.10 When other devices are connected with this ECG instrument, they must be type I devices which
accord with IEC60601-1. Because the total amount of leakage current may hurt patients, the monitoring of
leakage current is carried out and taken charge by connect devices.
ECG machine is designed to carry on ECG trace interpretation immediately after the measurement. It is
this interpretation that does not give report on all possible heart problems and may sometimes not comply
with the doctors diagnosis. Therefore, the final conclusion concerning each patient is up to the doctor
basing on patient symptom, the unit ECGs interpretation and other examination.

User Manual of ECG Machine

Chapter 3 Maintenance Regulation


3.1 Under the condition of normal use according to the user manual and operation notice, if this instrument
has any problem, please contact with our customer service department. Our company has the sales
record and customer archives for each instrument. The customer has one year's warranty service from the
beginning of shipping date according to the below time and condition.
3.2 Our company may adopt the ways of instruction, mailing to company by courier, visiting customers'
company, etc to carry out the maintenance promise.
3.3 Even in the period of free maintenance, we charge for reparation in the following archives:
3.3.1 Faults or damnification caused by misuse because not operate according to user manual and
operation notice.
3.3.2 Faults or damnification caused by dropping accidently when users move after purchasing.
3.3.3 Faults or damnification caused by preparation, reconstruction, decomposition, etc outside of our
company.
3.3.4 Faults or damnification caused by natural disasters such as fire, flood, earthquake, etc.
3.3.5 Faults or damnification caused by unapt thermal recording paper.
3.4 The free maintenance period for spare parts and fray parts is three months. Power cable, recording
paper, operation manual and packing material are excluded.
3.5 Our company is not responsible for the faults of other connecting instruments cause by the faults of
this device directly or indirectly.
3.6 The free maintenance service will be canceled if we find the protection label has been destroyed.
3.7 For charge maintenance beyond the warranty period, our company advises to continue to use
"Maintenance contract regulation". Please consult our customer service department for specific situation.

Chapter 4 Instrument Characteristics


4.1 5.7 inches LCD screen, real time display 12-lead ECG waveforms, and function key control at the
same time, more convenient and shortcut.
4.2 It is high resolution thermo sensitive printer and thermal-array (8 dots/mm), you should not adjust
anything Frequency Response is up to 150Hz.
4.3 The device can record real time clear and exact six channel ECG waveform and remark continually.
The remark includes: lead sign, sensitivity, paper speed, filter state, etc.
4.4 Full digital filter for resisting the baseline driftAC and interference.
4.5 Under automatic mode, just press the button once, it starts record procedure, which can enhance your
work efficiency.
4

User Manual of ECG Machine


4.6 The power supply includes both AC/DC. This device includes built-in lithium rechargeable battery and
automatic protection circuit
4.7 This instrument can record 150 pieces of ECG waveform and print 90 minutes continually under the
best DC state.
4.8 Digital signal processor for effective inhibition of baseline drift, interference, and the like.
4.9 The instrument has function with regular auto-measurement of ECG waveform parameter,
auto-analyze and auto-diagnostic, it will help to reduce doctors burden and improve working efficiency.
4.10 Safety Class: Class I, Type CF applied part. There is defibrillation circuit.
4.11 According to defense degree of deleterious fluid: IPX0
4.12 According to the safe degree used under the condition with flammable anaesthesia gas mixed with air
(or oxygen, nitrous oxide), this device belongs to the device which cant be used under the condition with
flammable anaesthesia gas mixed with air( or oxygen, nitrous oxide) .
4.13 According to the working mode class, this device belongs to continuous working device.
4.14 Explanation of some symbols in this device:
AC

AC work mode

OFF

Power supply is disconnected

ON

Power supply is connected


Equipotential point
Places need to be noticed, please refer to user manual
Device type is CF, which has defibrillation protection function
USB connector

PATIENT

Lead connector

User Manual of ECG Machine

Chapter 5 Panel Sketch Map


Display Screen

5.1 The sketch map and components name

Indicator Light
Control Panel
Print Paper
Lead Cable
Paper Cabinet Cover
Button

Figure 5-1 Front view


RS232 Connector
Lead Connector
USB Connector
Figure 5-2 Side view
Earth Terminal
Power Switch
Power Plug
Figure 5-3 Rear view
Fuse Label

Fuse Holder

Battery Label

Product Label

Figure 5-4 Bottom view


6

Battery Box

User Manual of ECG Machine

5.2 Key Definition

Figure 5-5 Control Panel


Function key: Switch recording mode
Function key: Waveform Reset; upwards
Function key: 1mV Calibration; Downwards

Function key: Patient Information Input; History Data Review

Function key: Lead Switch; Leftwards; Rightwards


Filter key: AC Filter; EMG filter; Baseline Filter
Function key: Menu
Function key: Print

Function key: On/Off


5.3 Indicator Definition
The indicator turns red when there is AC power supply. At the same time, the battery is being
recharged.
7

User Manual of ECG Machine

The Indicator turns green when instrument is power on.

Chapter 6 Attention before Operating


6.1 You are required to read this operation manual carefully before operating so as to ensure taking safe
and effective operation of the instrument.
6.2 Installation and maintenance of the instrument should be carried out as the following
6.2.1 There should be no high voltage cable, X radial instrument, ultrasound instrument and
electrotherapeutics instrument, etc around the ECG instrument.
6.2.2 Do not use or reserve the instrument in the place where the air pressure is too high, temperature and
humidity are over the common standard, the ventilation is not good, dust is too much, there is gas
containing salt and alkali and chemical medicine.
6.3 The instrument should be put on flat place. Take and put it lightly when move it. Avoid too strong
vibration and shock.
6.4 AC frequency and voltage value should accord with requirement ,and has enough current capacity.
6.5 Please put the device at the place where is easy to be grounded. Do not connect the patients and the
patients connecting cables with other conductors including ground or beds which can be conducted well
with ground.

Chapter 7 Preparation Work before Instrumentation


7.1 Check that the instrument properly grounded and that cable connections safe or not.
7.2 Make sure all electrodes directly connected with patient are properly and firm.
7.3 Check the output voltage when choose the DC UPS.
7.4 Smear the gel separately, avoiding the short circuit caused by the chest electrode touch one another.
7.5 AC power cable can not be enlaced with ECG cable.

Chapter 8 Precaution during Operation


8.1 Pay attention to the patient and instrument condition constantly.
8.2 Patient and instrument can only be connected ECG cables.
8.3 Keep close observation of the patient and instrument, to make sure they are not moved during
operation.
8

User Manual of ECG Machine


8.4 Turn off the instrument after using.
8.5 Turn off the power, and remove the ECG cables slightly without force.
8.6 Properly keep the instrument and spare parts for operation next time.
8.7 Paper Loading
8.7.1 Dimension of the high-speed thermal Recording paper used in this instrument is: 80mmW)20m(L)
8.7.2 Open the cover of paper cabinet, take out the paper axis and install recording paper according to the
figure 8-1 into the proper position inside.
8.7.3 Close the cover of paper cabinet. It's recommended to leave 2 cm of recording paper outside.

Figure 8-1 Inside View of Print Cabinet

Chapter 9 Electrode Placement


Advice: Set the chest electrode first, then the limb electrode.
9.1 Chest Electrode (See Figure 9-1)

Figure 9-1
Attach the chest electrodes to the locations as following:
V1: Fourth inter-costal space at right border of sternum.
V2: Fourth inter-costal space at left border of sternum.
V3: Midway between V2 and V4.
V4: Fifth inter-costal space at left mid-clavicular line.
V5: Left anterior axillary line at the horizontal lever of V4.
V6: Left mid-axillary line at the horizontal lever of V4.
Clean the skin where chest electrodes are to be attached with alcohol, then apply ECG cream to here
9

User Manual of ECG Machine


around 25mm in diameter and to the edge of chest electrodes, and press and attach the electrodes to the
positions from V1-V6.
Note:Keep in mind that the electrodes coming into contact with each other or creams overlap
from one position to another is not allowed.
9.2 Limb Electrode
Electrodes should be placed on the soft skin of hands and feet. Clean all the limb electrodes and the
positions around to which limb electrodes are to be attached with alcohol before applying ECG cream to
them, then firmly attach the electrodes to the positions. (See Figure 9-2)

Figure 9-2
Caution: Screw tightly the knob of ECG cable's plug after it inserted to the instrument.
9.3 Check-List for Electrode connection and ECG cable
Electrode Location Electrode Code

Socket Number

Right Alarm

RA/R

Left Alarm

LA/L

l0

Left Leg

LL/F

11

Right Leg

RL/N

14

Chest 1

Vl/Cl

12

Chest 2

V2/C2

Chest 3

V3/C3

Chest 4

V4/C4

Chest 5

V5/C5

Chest 6

V6/C6

Chapter 10 Grounding and Power Connection of Instrument


Make sure the status of the instrument is power off, and then make the instrument be properly
grounded through a 3-prong outlet. When the outlet, a grounding cable may be utilized to connect the
grounding terminal of the instrument. Do not use other pipeline. Properly grounding could guarantee
the safety and prevent from the interference of AC power and electromagnetic wave.

10

User Manual of ECG Machine

Chapter 11 Precaution for Battery Operation


11.1 This instrument is designed with the built-in sealed maintenance-free rechargeable lithium battery,
and has automatic charge and discharge monitoring system. The instrument recharges the battery
automatically when connect to AC power supply. The LCD screen will show the current power state at the
down right corner when the instrument turns on.
11.2 The device can continuously print 90 minutes and work 4 hours without printing after the battery fully
charged. When it working, the LCD screen displays the signal of the battery status in 4 degree. When
the power of battery is too low to operate, the instrument will turn off automatically to avoid damage
to the battery.
11.3 The battery should be recharged in time after exhausted using. For long storage, the battery is to be
recharged every 3 months. The battery life can be extended by doing so.
11.4 When the battery can not be recharged or works no more than 10 minutes after fully charged, please
change the battery.
Attention

z Do not directly connect both + and - poplars of battery with wire, otherwise it might cause fire hazard.
z Possible explosion hazard if it kept nearby the ablaze area.
z You should not open or disassemble the battery.

11

User Manual of ECG Machine

Chapter 12 Control Panel and Key Instruction


12.1 Main Interface
Patient Information

Record Lead Group

Rhythm Lead

HUM Filter

Current Time
EMG Filter

AC Freq.

Drift Filter
Heart Rate

Paper Speed

Power status
Wave Gain
Lead Switch Mode
Figure 12-1 Main Interface
z

The Current Patients ECG Information: The top left corner shows the current patients information.
ECG machine will measure, record and save this patients ECG curve information when recording.

Record Lead GroupDisplay the current recording lead group.

Rhythm Lead: Any lead from I to V6.

Current Time: Display the current time.

Lead Switch Mode: Switch recording mode (Automatic / manual) by pressing F1.

Wave Gain: 5 mm/mV, 10 mm/mV and 20 mm/mV can be chosen in the base setting menu.

Paper Speed: 12.5mm/s, 25mm/s and 50mm/s can be chosen in the base setting menu.

AC Freq.: 50Hz and 60Hz can be chosen in the base setting menu according to local AC power.

AC Filter: AC Filter (on/off) by pressing F7.

EMG Filter: EMG Filter (on/off) by pressing F7.

Drift Filter: Drift Filter (on/off) by set up in the base setting menu.

Heart Rate: Display patients real time heart rate.

Power status: Display the batterys charge status and capacity.

12.2 ECG Measurement


You can enter into the patient information input interface by pressing F4 before measuring. The patient
information input interface is as below Figure 12-2.

12

User Manual of ECG Machine


Figure 12-1Patient Information Input Window
z

Press F2: Leftwards

Press F3: Rightwards

Press F5: Input figure or letter descending

Press F6: Input figure or letter ascending

Press F4/F8: Enter into measuring main interface

The patient information includes the following contents.


z

ID: The system will automatically number, and you can also number manually by pressing F5/F6.

Name: Input patients name by pressing F5/F6.

Gender: Choose male or female by pressing F5/F6.

Age: Input patients age (Year) by pressing F5/F6.

Height: Input patients height (Cm) by pressing F5/F6.

Weight: Input patients weight (Kg) by pressing F5/F6.

BP: Input patients BP (mmHg) by pressing F5/F6.

System will show the patients information you input before after printing ECG wave. If you did not input
patients information before measuring, system will require you to input patients information after printing
ECG wave just as the same as the upper Figure 12-2. Then you can press F9 again to print the ECG
analysis results. If you select SAVE PATIENT DATA as ON, measure data will be saved into SD
memory card autumnally,

12.3 Main Menu


You can enter into the main menu by pressing F8. Main menu is as below Figure 12-3.

Figure 12-3 Main Menu


z

Base settings: You can choose base settings by pressing F2/F3 and enter into base setting menu by
pressing F5/F6. Base Setting Menu is as below Figure 12-4.

Advanced settings: You can choose advanced settings by pressing F2/F3 and enter into advanced
setting menu by pressing F5/F6. Advanced Setting Menu is as below Figure 12-5.

Date/Time Settings: You can choose date/time settings by pressing F2/F3 and enter into date/time
setting menu by pressing F5/F6. Date/time Setting Menu is as below Figure 12-6.

Version Information: Show the version information of ECG machine.

To restore the default value: You can select this menu by pressing F2/F3 and restore the factory
settings by pressing F5/F6.

Patient Database: You can choose patient database by pressing F2/F3 and enter into patient
13

User Manual of ECG Machine


database window by pressing F5/F6. Patient database window is as below Figure 12-7.

12.4 Base Setting Menu


You can choose Base Settings by pressing F2/F3 and enter base setting menu by pressing F5/F6. Base
Setting Menu is as below Figure 12-4.

Figure 12-4 Base Setting Menu


z

Display Wave Gain: 5 mm/mV, 10 mm/mV and 20 mm/mV can be chosen by pressing F5/F6.

Display Wave Speed: 12.5mm/s, 25mm/s and 50mm/s can be chosen by pressing F5/F6.

Record Wave Gain: 5 mm/mV, 10 mm/mV and 20 mm/mV can be chosen by pressing F5/F6.

Record Wave Speed: 25mm/s and 50mm/s can be chosen by pressing F5/F6.

Rhythm Lead: Any lead from I to V6 by pressing F5/F6.

Display ECG Wave Form: Six modes (3 channel, 3 channel + rhythm lead, 3 channel + channel +
V5 channel, six channel of single row, six channel of two rows, twelve channel) can be chosen by
pressing F5/F6.

Record ECG Wave Form: Five modes (3 channel, 3 channel + rhythm lead, 3 channel + channel
+ V5 channel, six channel, six channel + rhythm lead,) can be chosen by pressing F5/F6.

Auto Record Time: The total time (8s/12s/24s/36s) of recording wave in the auto mode by pressing
F5/F6.

Record Mode: Normal mode and Europe mode can be chosen by pressing F5/F6.

Drift Filter: Drift filter is on or off by pressing F5/F6.

Drift Filter Freq.: You can select drift filter freq.(0.05Hz/0.5Hz/1Hz) by pressing F5/F6.

HUM Filter Freq.: 50Hz and 60Hz can be chosen by pressing F5/F6.

EMG Filter Freq.: You can choose EMG filter freq.(25Hz/35Hz/45Hz) by pressing F5/F6.

Anti Baseline Drift: Anti Baseline Drift is on or off by pressing F5/F6.

Report Mode: You can choose report mode (no report/measure report /diagnose report) by pressing
F5/F6.

Save Patient Data: You can choose whether save patients ECG data or not by pressing F5/F6. ON
shows patients data will be saved into SD memory card autumnally after measuring. OFF shows
patient data will not be saved into SD memory card after be measured.

12.5 Advanced Setting Menu


You can choose Advanced Settings by pressing F2/F3 and enter into advanced setting menu by pressing
F5/F6. Advanced Setting Menu is as below Figure 12-5.

14

User Manual of ECG Machine

Figure 12-5 Advanced Setting Menu


z

Print Thick: You can choose whether print thick or not by pressing F5/F6.

Heartbeat Sound: You can choose whether turn on heartbeat sound or not by pressing F5/F6.

Keyboard Sound: You can choose whether turn on keyboard sound or not by pressing F5/F6.

LCD Backlight: LCD Light is always turning on in the process of using by chosen ALWAYS .You also
can turn off LCD backlight by chosen screen protection time. When screen is protected, you can
press any key except

(ON/OFF) to restore LCD backlight.

Demo: Show simulation wave when demo is on by pressing F5/F6.

Language: You can choose English or Chinese by pressing F5/F6.

Signal Out: Whether to open the ECG signal output or not using RS232 port.(No.9 pin).

Signal Out Cal (%): Set the value of signal out calibration (-50% ~ 50%) by pressing F5/F6.

Signal Out Lead: Choose signal out lead from I to V6 by pressing F5/F6.

Signal In: Whether to open the ECG signal input or not using RS232 port (No.1 pin).

Signal In Cal (%)Set the value of signal in calibration (-50% ~ 50%) by pressing F5/F6.

ECG Cal (%): Set the value of ECG calibration (-50% ~ 50%) by pressing F5/F6.

To ECG Station: Whether to connect with ECG station or not using RS232 port.

12.6 Date/Time Setting Menu


You can choose Date/time Settings by pressing F2/F3 and enter into date/time setting menu by pressing
F5/F6. Date/time Setting Menu is as below Figure 12-6.

Figure 12-6 Date/Time Setting Menu


z

Date: Set the current date by pressing F5/F6.

Time: Set the current time by pressing F5/F6.

Date Format: Choose the date format as YY-MM-DD or DD-MM-YY by pressing F5/F6.
15

User Manual of ECG Machine


12.7 Patient Database
You can choose Patient Database by pressing F2/F3 and enter into patient database window by
pressing F5/F6. Patient database window is as below Figure 12-7.

Figure 12-7 Patient Database


This window shows patients history ECG data those were measured and saved before. You can review,
call and print them by the following operation.
z

Select Patient: You can choose patient by pressing F2/F3.

Page Down/ Page Up: You can page down or page up patient history ECG data list by pressing
F5/F6.

Review: You can enter into review window by pressing F4. Patients history ECG data will be showed.

Print: You can print patients history ECG data by pressing F9.

12.8 Patient ECG Data Review


Patients history ECG data will be reviewed by pressing F4. Review window is as below Figure 12-8.

Figure 12-8 Patient ECG Data Review


z

Page Down/ Page Up: You can see ECG wave from left to right or right to left by pressing F2/F3.

Print: You can print patients history ECG data by pressing F9.

16

User Manual of ECG Machine

Chapter 13 Troubleshooting
13.1 Turn off automatically
z

Please check whether the power of battery is used up. Over discharge control circuit of the battery
acts.

Please check whether the alternating current voltage is too high. Over voltage control circuit acts.

Please check whether the alternating current disturbs is too high, whether the fix knob of lead plug is
too tight. Shut automatically is for protecting circuit when overload.

13.2 AC interference

Figure 13-1
z

Is the ECG device ground cable proper?

Are the electrodes and leads connected properly?

Is the electrode and skin covered with enough Gel?

Is the metal bed grounding proper?

Does the patient touch the wall or metal sickbed?

Does other person touch the patient?

Whether there is powerful electric device working beside ECG device? For example: X radial device
or B-Ultrasound devices.

13.3 EMG interference

Figure 13-2
z

Whether the patient room is comfortable.

Is the patient nervous?

Is the sickbed too narrow?

17

User Manual of ECG Machine


13.4 Baseline drift

Figure 13-3

Verify the electrode attachment and lead wire performance.

Check the connection between patient cable and electrodes.

Check the cleaning of electrode and patient skin. Is the electrode and skin covered with enough Gel?

Keep the patient from motion or hyperventilation.

Is the connection between lead and electrode proper?

Please use filter if still having above-mentioned interference.

13.5 Troubleshooting List


Phenomenon

Reason

Resolve method

1.Whether the ground cable proper.

1. Please check the lead, ground cable and power

2.The connection of leads is not stable.

Disturbance too

big, the waveform 3. Whether there


alternating current.
is in disorder

is

disturbance

from

supply.
2. Please dispose the patient in proper state.

4.Patient is nervous
1. Disturbance from alternating current is too
fierce.
Baseline is rough

Wave form is not


regular, with too
great

wave

beeline

or

1.Change a

comfortable environment for patient

2.Patient is nervous and the disturbance of 2.If the sickbed is metal, please change it
EMG too strong

3. The power line and lead is not parallel or too close.

1. The conductivity of electrode is not well.

1. Use alcohol of high quality.

2.Power of battery is used up

2.Clean the electrode and patients skin where touch

3.Contact between electrode and skin is not the electrode


proper.

3.Charge the battery

4. The plug between lead and main unit is


not tight.
5. The contact between lead and electrode is
not proper.

Baseline drift

1.Power of battery is used up

1.Charge the battery

2.Patient is moving

2.Keep patient hold still

1.The printer head is dirty

1. Clean the printer head with alcohol when the power

Waveform is not 2.The paper is not right

is off use the printer head after the alcohol is volatile.

clear.

2. Use the appointed thermal print paper.


18

User Manual of ECG Machine

Chapter 14 Main Technical Specification


The key technical indexes are as below.
CMRR
Input circuit

100dB, with AC Filter


Floating, Protection against
defibrillator effect

Calibrating voltage

1mV 3%

Voltage tolerance

500mV

Input circuit current

0.1 A

Time constant

> 3.2s

Resolution

12bit/1000Hz

Frequent response

0.05Hz~150Hz

Operation mode

Manual/Auto

Noise level

15 Vp-p

Filter

AC. EMG Filter

Threshold

20 V

Drift filter

Anti-Drift System

Paper speed

2550mm/s ( 3%)

Input Impedance

> 50M

Sensitivity

51020mm/mV

Patient current leakage

< 10 A

Recorder

High resolution thermal printer

Specification of fuse

T1.6A 250V 520

Battery

Recording paper

Thermal recording paper

Package

19

Built-in rechargeable Li-ion battery


(11.1V, 3800mAh)
360mm250mm68mm

User Manual of ECG Machine

20

You might also like