User Manul of ECG03C
User Manul of ECG03C
Model PRO-ECG03
User Manual
Statement
In order that you can use this product more safely and get full play to its functions and features, please
read the operating instructions carefully before using this unit and keep it properly.
Our company owns all rights to this unpublished work and intends to maintain this work as confidential. We
may also seek to maintain this work as an unpublished copyright. This publication is to be used solely for
the purpose of reference, operation, maintenance, or repair of our equipment. No part of this can be
disseminated for other purposes.
In the event of inadvertent or deliberate publication, our company intends to enforce its rights to this work
under copyright laws as a published work. Those having access to this work may not copy, use, or disclose
the information in this work unless expressly authorized by us to do so.
All information contained in this publication is believed to be correct. Our company shall not be liable for
consequential damages in connection with the furnishing, performance, or use of this material. This
publication may refer to information and protected by copyrights or patents and does not convey any
license under the patent rights of our company, nor the rights of others. Our company does not assume
any liability for arising out of any infringements of patents or other rights of the third parties.
The content of this manual is subject to change without prior notice.
Contents
Chapter 1 Overview........................................................................................................................................... 3
Chapter 2 Security Notice ................................................................................................................................ 3
Chapter 3 Maintenance Regulation ................................................................................................................ 4
Chapter 4 Instrument Characteristics ............................................................................................................. 4
Chapter 5 Panel Sketch Map ........................................................................................................................... 6
Chapter 6 Attention before Operating ............................................................................................................. 8
Chapter 7 Preparation Work before Instrumentation ................................................................................... 8
Chapter 8 Precaution during Operation ......................................................................................................... 8
Chapter 9 Electrode Placement ...................................................................................................................... 9
Chapter 10 Grounding and Power Connection of Instrument ................................................................... 10
Chapter 11 Precaution for Battery Operation ...............................................................................................11
Chapter 12 Control Panel and Key Instruction............................................................................................ 12
Chapter 13 Troubleshooting........................................................................................................................... 17
Chapter 14 Main Technical Specification ..................................................................................................... 19
Chapter 1 Overview
This ECG machine is such a kind of electrocardiograph, which samples 12 leads ECG signals
simultaneously and prints out the ECG waveforms with thermal printing system. Its functions are as follows:
recording and displaying ECG waveforms in auto/manual mode; auto-measuring and auto-diagnosing the
ECG wave parameters; save patient data into SD memory card automatically, prompting the state of lead
off; switching the interface languages (Chinese/English).
Our companys three channel ECG machine includes 192*64 LCD screen, 3.8 inches LCD screen and 5.7
inches LCD screen. This user manual takes 5.7 inches LCD screen for example. They have the same use
method and function
AC work mode
OFF
ON
PATIENT
Lead connector
Indicator Light
Control Panel
Print Paper
Lead Cable
Paper Cabinet Cover
Button
Fuse Holder
Battery Label
Product Label
Battery Box
Figure 9-1
Attach the chest electrodes to the locations as following:
V1: Fourth inter-costal space at right border of sternum.
V2: Fourth inter-costal space at left border of sternum.
V3: Midway between V2 and V4.
V4: Fifth inter-costal space at left mid-clavicular line.
V5: Left anterior axillary line at the horizontal lever of V4.
V6: Left mid-axillary line at the horizontal lever of V4.
Clean the skin where chest electrodes are to be attached with alcohol, then apply ECG cream to here
9
Figure 9-2
Caution: Screw tightly the knob of ECG cable's plug after it inserted to the instrument.
9.3 Check-List for Electrode connection and ECG cable
Electrode Location Electrode Code
Socket Number
Right Alarm
RA/R
Left Alarm
LA/L
l0
Left Leg
LL/F
11
Right Leg
RL/N
14
Chest 1
Vl/Cl
12
Chest 2
V2/C2
Chest 3
V3/C3
Chest 4
V4/C4
Chest 5
V5/C5
Chest 6
V6/C6
10
z Do not directly connect both + and - poplars of battery with wire, otherwise it might cause fire hazard.
z Possible explosion hazard if it kept nearby the ablaze area.
z You should not open or disassemble the battery.
11
Rhythm Lead
HUM Filter
Current Time
EMG Filter
AC Freq.
Drift Filter
Heart Rate
Paper Speed
Power status
Wave Gain
Lead Switch Mode
Figure 12-1 Main Interface
z
The Current Patients ECG Information: The top left corner shows the current patients information.
ECG machine will measure, record and save this patients ECG curve information when recording.
Lead Switch Mode: Switch recording mode (Automatic / manual) by pressing F1.
Wave Gain: 5 mm/mV, 10 mm/mV and 20 mm/mV can be chosen in the base setting menu.
Paper Speed: 12.5mm/s, 25mm/s and 50mm/s can be chosen in the base setting menu.
AC Freq.: 50Hz and 60Hz can be chosen in the base setting menu according to local AC power.
Drift Filter: Drift Filter (on/off) by set up in the base setting menu.
12
ID: The system will automatically number, and you can also number manually by pressing F5/F6.
System will show the patients information you input before after printing ECG wave. If you did not input
patients information before measuring, system will require you to input patients information after printing
ECG wave just as the same as the upper Figure 12-2. Then you can press F9 again to print the ECG
analysis results. If you select SAVE PATIENT DATA as ON, measure data will be saved into SD
memory card autumnally,
Base settings: You can choose base settings by pressing F2/F3 and enter into base setting menu by
pressing F5/F6. Base Setting Menu is as below Figure 12-4.
Advanced settings: You can choose advanced settings by pressing F2/F3 and enter into advanced
setting menu by pressing F5/F6. Advanced Setting Menu is as below Figure 12-5.
Date/Time Settings: You can choose date/time settings by pressing F2/F3 and enter into date/time
setting menu by pressing F5/F6. Date/time Setting Menu is as below Figure 12-6.
To restore the default value: You can select this menu by pressing F2/F3 and restore the factory
settings by pressing F5/F6.
Patient Database: You can choose patient database by pressing F2/F3 and enter into patient
13
Display Wave Gain: 5 mm/mV, 10 mm/mV and 20 mm/mV can be chosen by pressing F5/F6.
Display Wave Speed: 12.5mm/s, 25mm/s and 50mm/s can be chosen by pressing F5/F6.
Record Wave Gain: 5 mm/mV, 10 mm/mV and 20 mm/mV can be chosen by pressing F5/F6.
Record Wave Speed: 25mm/s and 50mm/s can be chosen by pressing F5/F6.
Display ECG Wave Form: Six modes (3 channel, 3 channel + rhythm lead, 3 channel + channel +
V5 channel, six channel of single row, six channel of two rows, twelve channel) can be chosen by
pressing F5/F6.
Record ECG Wave Form: Five modes (3 channel, 3 channel + rhythm lead, 3 channel + channel
+ V5 channel, six channel, six channel + rhythm lead,) can be chosen by pressing F5/F6.
Auto Record Time: The total time (8s/12s/24s/36s) of recording wave in the auto mode by pressing
F5/F6.
Record Mode: Normal mode and Europe mode can be chosen by pressing F5/F6.
Drift Filter Freq.: You can select drift filter freq.(0.05Hz/0.5Hz/1Hz) by pressing F5/F6.
HUM Filter Freq.: 50Hz and 60Hz can be chosen by pressing F5/F6.
EMG Filter Freq.: You can choose EMG filter freq.(25Hz/35Hz/45Hz) by pressing F5/F6.
Report Mode: You can choose report mode (no report/measure report /diagnose report) by pressing
F5/F6.
Save Patient Data: You can choose whether save patients ECG data or not by pressing F5/F6. ON
shows patients data will be saved into SD memory card autumnally after measuring. OFF shows
patient data will not be saved into SD memory card after be measured.
14
Print Thick: You can choose whether print thick or not by pressing F5/F6.
Heartbeat Sound: You can choose whether turn on heartbeat sound or not by pressing F5/F6.
Keyboard Sound: You can choose whether turn on keyboard sound or not by pressing F5/F6.
LCD Backlight: LCD Light is always turning on in the process of using by chosen ALWAYS .You also
can turn off LCD backlight by chosen screen protection time. When screen is protected, you can
press any key except
Signal Out: Whether to open the ECG signal output or not using RS232 port.(No.9 pin).
Signal Out Cal (%): Set the value of signal out calibration (-50% ~ 50%) by pressing F5/F6.
Signal Out Lead: Choose signal out lead from I to V6 by pressing F5/F6.
Signal In: Whether to open the ECG signal input or not using RS232 port (No.1 pin).
Signal In Cal (%)Set the value of signal in calibration (-50% ~ 50%) by pressing F5/F6.
ECG Cal (%): Set the value of ECG calibration (-50% ~ 50%) by pressing F5/F6.
To ECG Station: Whether to connect with ECG station or not using RS232 port.
Date Format: Choose the date format as YY-MM-DD or DD-MM-YY by pressing F5/F6.
15
Page Down/ Page Up: You can page down or page up patient history ECG data list by pressing
F5/F6.
Review: You can enter into review window by pressing F4. Patients history ECG data will be showed.
Print: You can print patients history ECG data by pressing F9.
Page Down/ Page Up: You can see ECG wave from left to right or right to left by pressing F2/F3.
Print: You can print patients history ECG data by pressing F9.
16
Chapter 13 Troubleshooting
13.1 Turn off automatically
z
Please check whether the power of battery is used up. Over discharge control circuit of the battery
acts.
Please check whether the alternating current voltage is too high. Over voltage control circuit acts.
Please check whether the alternating current disturbs is too high, whether the fix knob of lead plug is
too tight. Shut automatically is for protecting circuit when overload.
13.2 AC interference
Figure 13-1
z
Whether there is powerful electric device working beside ECG device? For example: X radial device
or B-Ultrasound devices.
Figure 13-2
z
17
Figure 13-3
Check the cleaning of electrode and patient skin. Is the electrode and skin covered with enough Gel?
Reason
Resolve method
Disturbance too
is
disturbance
from
supply.
2. Please dispose the patient in proper state.
4.Patient is nervous
1. Disturbance from alternating current is too
fierce.
Baseline is rough
wave
beeline
or
1.Change a
2.Patient is nervous and the disturbance of 2.If the sickbed is metal, please change it
EMG too strong
Baseline drift
2.Patient is moving
clear.
Calibrating voltage
1mV 3%
Voltage tolerance
500mV
0.1 A
Time constant
> 3.2s
Resolution
12bit/1000Hz
Frequent response
0.05Hz~150Hz
Operation mode
Manual/Auto
Noise level
15 Vp-p
Filter
Threshold
20 V
Drift filter
Anti-Drift System
Paper speed
2550mm/s ( 3%)
Input Impedance
> 50M
Sensitivity
51020mm/mV
< 10 A
Recorder
Specification of fuse
Battery
Recording paper
Package
19