The document discusses different types of x-ray generator configurations - single phase, three phase with 6 pulses per second, three phase with 12 pulses per second, and high-frequency. Single phase is the least efficient as it produces a voltage ripple of 120 pulses per second and the voltage drops to zero twice per cycle. Three phase with 6 pulses per second produces a 13-25% ripple and voltage never drops below 75-87% of peak. Three phase with 12 pulses per second produces an even smaller 4-10% ripple. High-frequency generators produce the smallest ripple of 3-4% and are the most efficient. The power of an x-ray generator is calculated as voltage multiplied by amperage
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0 ratings0% found this document useful (0 votes)
38 views
Notebook 8
The document discusses different types of x-ray generator configurations - single phase, three phase with 6 pulses per second, three phase with 12 pulses per second, and high-frequency. Single phase is the least efficient as it produces a voltage ripple of 120 pulses per second and the voltage drops to zero twice per cycle. Three phase with 6 pulses per second produces a 13-25% ripple and voltage never drops below 75-87% of peak. Three phase with 12 pulses per second produces an even smaller 4-10% ripple. High-frequency generators produce the smallest ripple of 3-4% and are the most efficient. The power of an x-ray generator is calculated as voltage multiplied by amperage
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 3
Notebook 8
Single Phase (2p)
- produces voltage ripple of 2p/Hz or 120p/sec - 100% ripple: drops to 0 twice per cycle - least efficient
Three Phase (6p)
produces voltage ripple of 6p/Hz or
360p/sec - 13-25% ripple: voltage in x-ray tube never drops below 75-87% of peak kilovoltage setting - produces 35% more average photon energy than full-wave rectified, single phase - 6 rectifiers
Three Phase (12p)
-
produces voltage ripple 12p/Hz or 720p/sec
4-10% ripple 12 rectifiers
19
High-Frequency -
3-4% ripple Most efficient
Equation is calculate as below:
P= Watt P= voltge (V) X amperage (A) -when generator has low average photon emission (single-phase) use constant: V X A X 0.7 = W
20
References 1. Carlton, R. R., Adler, A. M., & Frank, E. D. (2006). Principles of radiographic imaging: An art and a science. Clifton Park, NY: Thomson Delmar Learning.
Pages From Valerie F. Andolina, Shelly L. Lillé-Mammographic Imaging - A Practical Guide (Point (Lippincott Williams & Wilkins) ) - Lippincott Williams & Wilkins (2010)