Document No.
DC-2008-013-05P
PrexViewer
Users Manual
Rev. 2.2
PreXion, Inc.
Index
1.
PREFACE
1-1
1-2
1-3
DISCLAIMER
ABOUT PREXVIEWER
RENDERING FEATURES
4
4
4
1-4
1-5
MINIMUM SYSTEM REQUIREMENT
UNSUPPORTED FUNCTIONS
7
7
1.3.1
1.3.2
1.3.3
1.3.4
1.3.5
1.3.6
1.3.7
2.
4
5
5
5
6
6
6
2-1
2-2
2-3
LAUNCH VIEWER
LOADING DATA
DISPLAY ADJUSTMENT
8
10
12
2-4
2-5
3DVR IMAGE ADJUSTMENT
MPR IMAGE ADJUSTMENT
15
16
2.5.1
2.5.2
3.
Window Level
Panning
Zoom In/Out
Rotation
Slice
Window level adjustment
RaySum Display
CUTTING
3-1
3-2
USING AN OUTLINE
USING THE SLAB FUNCTION
3.2.1
4.
Example Use of the Slab Function
CPR FUNCTION
4-1
4-2
CPR MODE SCREEN
CREATING CROSS-SECTION IMAGES
4.2.1
5.
Swiching Cross-Section Images to RaySum Images
MEASUREMENT FUNCTION
5-1
5-2
6.
MEASURING DISTANCE
MEASURING THE ANGLE
PANORAMIC IMAGE DISPLAY
6-1
6-2
6-3
6-4
6-5
6-6
6-7
6-8
3DVR
MPR (Multi Planar Reconstruction)
RaySum (Ray Summation)
MIPMaximum Intensity Projection
MinIP (Minimum Intensity Projection)
Panoramic CPR
Panorama Cross Section
STARTING PREXVIEWER
2.3.1
2.3.2
2.3.3
2.3.4
2.3.5
7.
LOAD PANORAMIC IMAGE
GAMMA SETTING AND WINDOW LEVEL
PRESET
COLOR TEMPLATE
COLOR MAP
EDGE FILTER
ZOOM SHORTCUTS
FLIP
2D IMAGE DISPLAY
12
13
13
14
14
16
16
17
17
19
20
22
22
23
25
26
27
28
29
32
34
36
38
40
43
44
46
48
7-1
7-2
CHANGING THE LAYOUT
ADJUSTMENT OF THE MULTI-SETTING DISPLAY
49
51
8.
SCENE
52
9.
CONTACT INFORMATION
53
PrexViewer Users Manual
1. Preface
1-1
Disclaimer
PrexViewer is a review software application. Please refer to PreXion3D Viewer Software [of the
PreXion3D Scanner] for clinical diagnostics. By running the PrexViewer software you acknowledge
having read the disclaimer.
1-2
About PrexViewer
PrexViewer Data CD(Data CD) is generated by PrexViewer CDMaker(CD Maker).
The PrexViewer can run without installation on Windows computers and laptops.
1-3
Rendering Features
1.3.1 3DVR
This is a 3D volume rendering image. It is abbreviated as VR.
Volume rendering is a visualization method that uses volume data consisting of cubes called voxels. This
is used to visualize the 3D image to the interior, and image is dislplayed by adding colors and shadows
that have been calculate using the light penetration and reflection.
PrexViewer Users Manual
1.3.2 MPR (Multi Planar Reconstruction)
MPR stands for the Multi Planer Reconstruction method. This method stacks the slices to create solid
volume data which can be used to display freely generated cross-section images.
Cross sections from different angles can be displayed based on multiple three-dimensionally imaged
image data such as horizontal cross sections (Axial) and vertical cross sections (Sagittal).
This is used to make a detailed check of the affected area and other applications.
<Axial>
<Coronal>
<Sagittal>
1.3.3 RaySum (Ray Summation)
This is the ray summation method. The CT value is added to the projected rays to create an image that
mimics a simple X-ray photograph.
1.3.4 MIPMaximum Intensity Projection
Maximum Intensity Projection provides a display that emphasizes the areas with the highest CT values.
This is displayed in a 3D monochrome image. This is used to easily identify the overall shape, implant
areas, etc.
CAUTION
An MIP image differs from a simple X-ray projection image.
PrexViewer Users Manual
1.3.5 MinIP (Minimum Intensity Projection)
Minimum Intensity Projection provides a display that emphasizes the areas with the lowest CT values.
This is used to check nerve holes and similar areas.
1.3.6 Panoramic CPR
This is a panoramic image. Discretionary cross sections of 3D volume data are joined together to form a
single image. This makes it possible to see the entire tooth alignment in one view.
CAUTION
This differs from images taken using a panorama X-ray machine.
1.3.7 Panorama Cross Section
This displays the discretionary cross sections of a panorama CPR image.
PrexViewer Users Manual
1-4
Minimum System Requirement
Operating System: WindowsXP Service Pack2, Windows7
CPU: Intel Pentium4 or later (SSE2 feature must be required)
Memory: 512MB
Optical drive: CDROM (x52 or faster recommended)
Display: 1024x768 pixel
1-5
Unsupported Functions
This product does not support following functions.
Printing
2D/3D protocol (movie)
DICOM file import/export
Save Scene
Simple Report/Interactive Report
Multi Data
Mask
Implant Planning
Fusion
PrexViewer Users Manual
2. Starting PrexViewer
These methods are used to adjust the image display. The display settings for angle, color, image
information, and other items can be set.
2-1
Launch Viewer
1
Operation
Insert the Data CD into the
optical drive.
The PrexViewer will start
automatically.
If viewer does not start,
Double click My Computer
icon on Desktop, and then
right click the optical drive.
Operation
Select AutoPlay from menu.
PrexViewer Users Manual
Operation
PrexViewer will be launched
Operation
To exit from PrexViewer,
click Exit button.
PrexViewer Users Manual
2-2
Loading Data
1
Operation
Double click the examination
to be read.
Note
The data can be read from the series list. Double click on the series to be read or right click on the series and then
click on Open 3D View.
Operation
The Review screen is
displayed.
Of the series included in the
selected examination, the
series with the most images is
read.
10
PrexViewer Users Manual
Operation
To return to Patient screen,
click Patient List button.
PrexViewer Users Manual
11
2-3
Display Adjustment
The tool buttons at the top left are used to zoom, rotate, and change the display position of the images.
Note
Adjustment can also be performed from the menu that is displayed by a right click.
2.3.1 Window Level
After clicking on the button, drag on the screen display view image to change the window level.
The window level can also be changed without using a button by dragging on the screw display view
image with the right and left buttons at the same time.
When not using a button
Simultaneous
right and left drag
Left drag
12
PrexViewer Users Manual
2.3.2 Panning
After clicking on the button, drag right or left, up or down on the screen display view image to change
the display position.
The display position can also be changed without using a button by using the mouse right button to drag
right or left, up or down on the screen display view image.
When not using a button
Right drag
Left drag
2.3.3 Zoom In/Out
After clicking on the button, drag on the screen display view image to zoom in or zoom out the display.
The zoom can also be done without using a button by dragging the wheel button up or down on the
screen display view image.
When not using a button
Wheel drag (vertical)
Left drag
PrexViewer Users Manual
13
2.3.4 Rotation
This is used with the 3DVR image display.
After clicking on the button, drag on the screen display view 3DVR image to rotate the image display.
This cannot be used for MPR images.
Select this button when changing the screen.
Left drag
2.3.5 Slice
This is used with MPR image displays.
After clicking on the button, drag on the screen display view MPR image to move the slice. Click on the
[] button to specify the slice movement speed.
This cannot be used for 3DVR images.
Left drag
14
PrexViewer Users Manual
2-4
3DVR Image Adjustment
The 3D Settings tab is used to adjust the 3DVR image display.
<3DVR Example>
PrexViewer Users Manual
15
2-5
MPR Image Adjustment
The MPR image display can be adjusted using the MPR operation panel.
2.5.1 Window level adjustment
Window level adjustment is set using the Window Level tab of the MPR operation panel.
This adjustment can be set by directly inputting the window width value in WW and the window level
value in WL.
In addition, the slide bar can be dragged to set the window level and both knobs of the slide bar can be
dragged to set the window width.
Slide bar
Knobs
2.5.2 RaySum Display
The three basic windows slice thickness can be set using the RaySum tab of the MRP operation panel.
Check the box for RaySum and then select Thickness from the list.
To move by the selected RaySum mode thickness, remove the check from the box for Snap to Slab.
CAUTION
16
When a check is place in the box for Snap to Slab, the normal MPR image slices are moved one at a time.
PrexViewer Users Manual
3. Cutting
Freely cutting any area allows you to remove unnecessary information and display only the area of
interest.
3-1
Using an Outline
An outline and anchor points are displayed when an MPR is displayed.
The outline on the MPR image can be dragged to cut the 3DVR image.
Outline
An outer frame broken line is
displayed on each image.
<Flow when cutting away the front portion>
Move the outline to the cut position
on the MPR (axial) image.
3DVR image before cut
MPR (axial) image
PrexViewer Users Manual
3DVR image after cut
17
Operation
Drag the outline to the area to
be cut on the MPR image.
Operation
The area in front of the drag
is cut off and the cut 3DVR
image is displayed.
18
PrexViewer Users Manual
3-2
Using the Slab Function
This function can be used to cut at a diagonal to the oblique coordinates or create a cut image that leaves
just a specified thickness (slab).
An oblique can be created by dragging the cross line and anchor point.
Aligning the cut with the tooth alignment makes it possible to create an MIP image of just the required
area and a MinIP image that shows an overall image of a neural tube.
Cuts at the thickness specified for the cut line
Cut line
Cuts at the thickness
specified for the cut line
Cut line
Cross line
Anchor points
Note
An oblique is a discretionary cross-section image. A discretionary diagonal image is called a double oblique.
<Flow when cutting using the slab function>
Move the green line on the
MPR (axial) to the place to
be cut.
Use the yellow lines on
both sides of the green
line
to
adjust
the
thickness.
PrexViewer Users Manual
A 3DVR image cut to the
specified thickness is
displayed.
19
3.2.1 Example Use of the Slab Function
1
Operation
Move the center of the cross
lines to the area to be shown
by the Slab function.
The cross lines are moved by
dragging them at their center.
Operation
Click on the Slab tab of the
3D operation panel and place
a check in the box for Slab.
20
PrexViewer Users Manual
Operation
Move the cross lines and
anchor points to adjust the
display area.
Drag the displayed yellow
line to change the thickness of
the displayed area.
Operation
The 3DVR image cut the set
slab thickness is displayed.
Note
Drag the anchor points to change the cut direction.
Right click on the 3DVR image to switch between MPR, MIP and MinIP images.
PrexViewer Users Manual
21
4. CPR Function
Curved Planar Reconstruction (CPR) is one type of MPR method. With MPR straight line cross sections
are displayed, but with CPR free curve cross-section views (cross-section images) are displayed.
This function can be used to create discretionary cross-section images.
4-1
CPR Mode Screen
Click on the [CPR] button from the tool buttons at the top right of the Review screen to switch to the
CPR mode.
A free curve can be drawn in the CPR mode and displayed and analyzed as a continuous cross-section
view.
CPR button
22
PrexViewer Users Manual
4-2
Creating Cross-Section Images
A continuous horizontal cross-section view is called a cross-section image.
When a curve is drawn in the CPR mode screen, a cross-section view that intersects with the drawn
curve (CPR line) is created.
The intersecting lines are called cross-sections and continuous horizontal cross-section views
corresponding to the number of displayed cross sections are created.
The CPR line cross-section view is displayed as a panoramic image.
<Flow when creating a cross-section image>
3DVR Image
Draw the cut line (CPR line)
Put cross sections in the CPR line)
CPR line cross-section view (panorama
display)
Cross-section views
Operation
Click on while holding down
the Ctrl key at the MPR view
and draw a discretionary CPR
line.
A green line will be drawn.
CPR line
Note
To clear the drawn CPR line, click on the [Reset] button from the CPR tool buttons.
The lines shown on both sides of the drawn CPR lines show the CPR line slice.
PrexViewer Users Manual
23
Operation
Drag the anchor points to
correct the drawn CPR line.
To delete the anchor points,
click on while holding down
the Shift key.
Anchor Points
Operation
The cross-section images for
the drawn CPR line are
displayed in the cross-section
view.
The drawn CPR line
panoramic image is displayed
in the panorama view.
24
PrexViewer Users Manual
4.2.1 Swiching Cross-Section Images to RaySum Images
Cross-Section images with CPR mode can be swiched to RaySum Images.
Operation
Display the cross-section
image and click
button
on the top of the screen.
Operation
1. Check the box of
[RaySum].
2. Select the thickness from
the pull down menu.
3. Click [OK] button, the
cross-section image
swiched to RaySum
image is displayed.
(1)
(2)
(3)
PrexViewer Users Manual
25
5. Measurement Function
The discretionary distance on an image, the surface area of an area of interest, signal values, and other
values can be measured in the Review screen.
The Measurement menu is used for measurements. The Measurement menu is displayed by clicking on
the [Measurement] button of the tool buttons at the top left of the Review screen.
Button
Name
Description
Select and move measurement result values, lines, etc.
Select
Delete measurement results, lines, etc.
Delete
Measure straight lines. Drag to draw a straight line. Displays the length.
Distance
Multiple
Distance
Relative
Distance
Angle
26
Line
Measure curve distance. Click on to draw a curve. Double click on the end point
to display the measurement results.
Measure the relative distance of multiple lines. Drag to draw a straight line. The
first line is used as the reference and the distance relative value for the second and
subsequent lines is displayed.
Measure the angle of two straight lines. Drag to draw a straight line. The first line
is used as the reference and the angle with the second is displayed.
PrexViewer Users Manual
5-1
Measuring Distance
This is used to measure the distance of the drawn line.
(1)
Operation
1.
2.
Click on the [Distance]
button.
Drag on the area of the
image to be measured to
draw a straight line.
(2)
Operation
Display the measurement
results.
PrexViewer Users Manual
27
5-2
Measuring the Angle
This is used to measure the angle between the drawn lines.
(1)
Operation
1.
2.
3.
Click on the [Angle]
button.
Drag on the discretionary
position on the image to
draw a straight line.
From the starting point of
the first line drag to draw
a straight line.
(2)
(3)
Operation
The measurement results are
displayed.
28
PrexViewer Users Manual
6. Panoramic Image Display
The Panoramic image can be loaded from the Patient List screen.The image is displayed in various
ways.
Adjustment of Gamma setting and Window Level
The Gamma setting, Window Level (WL) and the Window Width (WW) can be changed.
Refer to 6-2 Gamma Setting and Window Level.
Preset
Only by clicking a button, one of five presets of Gamma setting, Window Level and Window
Width can be applied.
Refer to 6-3 Preset.
PrexViewer Users Manual
29
Color Templates
Color templates such as sepia-toned, inverted color can be applied.
Refer to 6-4 Color Template.
Color Map
The area which has the pixel value of your interest can be tinted with colors.
Refer to 6-5 Color Map.
30
Edge Filter
The image filter can be applied to the image. Two filters of edge enhancement.
PrexViewer Users Manual
Refer to 6-6 Edge Filter.
Flip
The image can be flipped vertically and/or horizontally.
Refer to 6-8 Flip.
PrexViewer Users Manual
31
6-1
Load Panoramic Image
Load a panoramic image into Panoramic Image View from Patient List. When the scan is done, an
panoramic image is generated in Series 100.
Operation
Click the Study data in the
upper list of Patient List
screen to display.the Series
List at the bottom of the
screen.
Double click the series that
has the number 100. It must
have only one image in it.
Operation
The panoramic image is
loaded immediately in the
Panoramic Image View.
32
PrexViewer Users Manual
Operation
Click on [Enable/Disable
Orientation], the orientation
display is displayed.
PrexViewer Users Manual
33
6-2
Gamma Setting and Window Level
The Gamma setting, the Window Level and the Window Width can be adjusted.
Operation
Click on the icon for
Panorama Histogram
floating panel on upper
right of the screen.
Operation
The Panorama Histogram
floating panel is displayed.
You can change Gamma
setting, Window Level and
Window Width here.
Note that Window Level
and Window Width can be
changed by mouse
operation described in 2-3
Display Adjustment.
34
PrexViewer Users Manual
Operation
An example of Gamma
setting. Gamma is set to
0.58.
Alternative Operation
The Window Level and the
Window Width can be
adjusted in the panel on the
right side of the screen, too.
PrexViewer Users Manual
35
6-3
Preset
You can apply the preset Gamma setting, window level and the window width to the image.
Operation
Click one of the buttons in
[Preset] section in the
panel. (e.g. Click [Tooth2])
Operation
The preset Gamma setting,
window level and the
window width is applied to
the image.
36
PrexViewer Users Manual
Operation
To change the preset value,
right click [Preset] button
after adjusting the Gamma,
window level and window
width. Input the new button
name on the Set Panorama
Preset window and click
[OK] button.
[Default] button reverts the
values to the factory setting.
PrexViewer Users Manual
37
6-4
Color Template
Different types of the color template can be applied.
Operation
Click one of the buttons in
[Color Template] section in
the panel.
If you choose [Rainbow],
the image is switched to the
color gradation from blue to
red.
Operation
If you choose [Sepia], the
image will be switched to
brownish-red.
38
PrexViewer Users Manual
Operation
If you choose [Invert], the
color of the image will be
inverted.
PrexViewer Users Manual
39
6-5
Color Map
You can color the area which has the specified pixel value range.
Operation
Click on [Display pixel
value] button.
You can start from the 2nd
operation below without
this step. In that case, the
pixel value cannot be
checked in the 2nd operation
because the pixel value is
not displayed.
Operation
Move the cursor and double
click on the pixel which has
the pixel value you are
interested in. This pixel value
will be the center value.
CAUTION
40
In the panoramic image, the CT value is not displayed but the pixel value.
PrexViewer Users Manual
Operation
The area that has pixel
value close to the center
value (the center value
range) is colored.
The original image can be
observed even if colored,
only when you chose the
Gray color template.
Operation
Double click again to add
another area of pixel value
range. You can have up to
three areas.
PrexViewer Users Manual
41
Operation
You can change the color of
the area.
Right click the row to show
the pull down menu.
Select [Color Edit] to
change colors.
Select [Up] or [Down] to
change the lap order.
Select [Remove] to remove
the colored area.
Operation
Use the range slider bar to
adjust the range of pixel
value to color.
Operation
Click [Remove All] button to
remove all the colored areas.
42
PrexViewer Users Manual
6-6
Edge Filter
You can apply the image filter that enhances edges.
Operation
Click one of the filter buttons
in the [Filter] section of the
panel to apply the image
filter.
Note
[No Filter]:
The original image
[Sharpen 1]: Edge enhance. When you load an image, this filter is applied.
[Sharpen 2]: Stronger edge enhancement than Sharpen 1.
Operation
This is an example of
[Sharpen 2]
PrexViewer Users Manual
43
6-7
Zoom shortcuts
There are two zooming shortcut buttons in the [Zoom] section of the panel.
Operation
Click [Fit] button to make
the image fit the current
viewing area. Left is an
example result.
Click [100%] button to
make
the
image
pixel-by-pixel
magnification.
Pixel-by-pixel means that 1
pixel of image corresponds
to 1 pixel of LCD display.
Note
The display magnification ratio will be changed when you perform another zooming operation.
The display magnification ratio is changed when you show or hide the operation panel on the right side of the
screen.
Operation
Left is an example result of
[100%] zoom.
44
PrexViewer Users Manual
Alternative Operation
Right click on the screen,
the menu is displayed.
The same operation as
using [Zoom] button can be
done from the menu.
PrexViewer Users Manual
45
6-8
Flip
The image can be flipped horizontally and/or vertically.
Please note that the image is displayed the other way round in right and left. Enable the orientation
display and indicate LR code to avoid misidentifying.
Operation
Click [VFlip] button to flip the
image vertically.
The button is highlighted to
indicate the image is flipped.
Operation
Click on [HFlip] button to flip
the image horizontally.
Operation
This is an example of
horizontally flipped image.
Make sure to enable the
orientation display.
Refer to 6-1 Load Panoramic
Image for
display.
46
PrexViewer Users Manual
the
orientation
Operation
Click [Reset] button to revert
to the original status.
PrexViewer Users Manual
47
7. 2D Image Display
Multiple 2D images can be displayed at the same time just like for film diagnosis. Multiple displays of
one slice or simultaneous display of multiple slices are possible.
The types of slice multiple displays are explained below.
Multi-setting Display
This shows multiple displays of one slice. Each display is shown at a different angle, so several
angles can be checked at one time. This is shown by clicking on the [Multi-setting] button on the left
side.
Multi-setting
Multi-setting display view
2D operation panel
Multi-slice display
This displays multiple slices at one time. This allows the differences of each cross section to be
checked. This is shown by clicking on the [Multi-slice] button on the right side.
Multi-slice
Multi-slice display view
48
2D operation panel
PrexViewer Users Manual
7-1
Changing the Layout
The multi-setting display layout can be changed from 11 to 22, and the multi-slice layout can be
changed from 11 up to 55.
This section explains the method for changing the multi-slice display layout.
Operation
Click on the [Multi-slice]
button.
Note
For the Multi-setting display, click on the left side [Multi-setting] button.
Operation
Click on the [Change Layout]
button and drag the number to
be displayed to change to the
number of images you want
to display.
PrexViewer Users Manual
49
CAUTION
The [Change Layout] button when the multi-setting display is used differs from that for the multi-slice display.
The number of displays that can be specified is 11 or 22.
Operation
The specified layout is
displayed.
50
PrexViewer Users Manual
7-2
Adjustment of the Multi-setting Display
The top left tool buttons and 2D operation panel can be used to adjust the 2D image display.
Note
The top left tool buttons are used in the same manner as when working with 3D images.
Window Level Adjustment
This is used to adjust the displayed slice image window width and window level.
This can be set for each displayed slice. The setting method is the same as for 3D display.
The slices are set in the order of top left, top right, bottom left, bottom right from the top of the 2D
operation panel.
(1)
(2)
(1)
(2)
(3)
(4)
(3)
(4)
PrexViewer Users Manual
51
8. Scene
PrexViewer can load Scenes which saved in PreXion3D Viewer.
(1) Operation
4.
Select the examination
containing the scene to
be opened.
The saved scenes are
displayed as thumbnail
images.
5. Select the thumbnail
(2)
image of the scene to be
opened and click on the
[Open] button or double
click on the thumbnail
image.
Operation
The display changes to the
Review screen and the scene
is displayed.
52
PrexViewer Users Manual
9. Contact Information
PreXion, Inc.
Address:
Tel:
Fax:
E-mail:
411 Borel Avenue, Suite 550
San Mateo, CA 94402, USA
+1-650-212-0300
+1-650-212-0310
[email protected]Customer Service
Tel:
+1-650-212-0314
E-mail:
[email protected]
PrexViewer Users Manual
53