May 2010
General Information
Abstract
This document provides information about the DICOM Conformance of the On
Board Imager version 1.5/PortalVision Advanced Imaging product.
OBI and PV Advanced Imaging
Manufacturer and
European
Representative
Manufacturer:
European Representative:
Varian Medical Systems, Inc. Ltd.
3100 Hansen Way, Bldg. 4A
Palo Alto, CA 94304-1030, U.S.A.
Varian Medical Systems UK Ltd.
Gatwick Road, Crawley
West Sussex RH10 9RG
United Kingdom
Notice
Information in this document is subject to change without notice and does not
represent a commitment on the part of Varian. Varian is not liable for errors
contained in this document or for incidental or consequential damages in
connection with the furnishing or use of this material.
This document contains proprietary information protected by copyright. No part
of this document may be reproduced, translated, or transmitted without the
express written permission of Varian Medical Systems, Inc.
FDA 21 CFR 820
Quality System
Regulations (CGMPs)
Varian Medical Systems, Oncology Systems products are designed and
manufactured in accordance with the requirements specified within this federal
regulation.
ISO 13485
Varian Medical Systems, Oncology Systems products are designed and
manufactured in accordance with the requirements specified ISO 13485 quality
systems standards.
CE
Varian Medical Systems, Oncology Systems products meet the requirements of
Council Directive MDD 93/42/EEC.
HIPAA
Varians products and services are specifically designed to include features that
help our customers comply with the Health Insurance Portability and
Accountability Act of 1996 (HIPAA). The software application uses a secure
login process, requiring a user name and password, that supports role-based
access. Users are assigned to groups, each with certain access rights, which
may include the ability to edit and add data or may limit access to data. When a
user adds or modifies data within the database, a record is maintained of the
data that was changed, the users ID and the date and time the changes were
made. This establishes an audit trail that can be examined by authorized system
administrators.
Trademarks
ARIA, Eclipse, Varian and VMS are registered trademarks
ARIA, Eclipse, On-Board Imager are trademarks of Varian Medical
Systems, Inc.
Microsoft and Windows are registered trademarks of Microsoft Corporation.
All other trademarks or registered trademark are the property of their respective
owners.
2 of 118
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
General Information
Contacting Support
Support services are available without charge during the initial warranty period.
If you seek information not included in this publication, call Varian Medical
Systems support at the following locations:
United States and Canada telephone support + 1 888 827 4265
United States and Canada Direct telephone support + 1 650-213-1000
European telephone Support + 41 41 749 8844
Fax (US) + 1 702 938 4754
Fax (Service Europe) + 41 41 740 3340
All other countries please call your local service office
To contact the support location nearest you for Service, Parts or Support, see
the list at the Varian Medical Systems website:
Worldwide Listing
http://www.varian.com/us/oncology/services_and_support/contacts.html
Communicating Via the World Wide Web
If you have access to the Internet, you will find Varian Medical System support
at the following location:
Oncology Systems http://my.varian.com
If you have a Varian account, enter your username and password. Otherwise,
first click create new account to get a username and password.
From MyVarian home page, click Contact Us from the Support list along the left
side of the window.
If possible, please send all e-mail inquires through the my.varian.com web site
at http://my.varian.com/contactus; otherwise, use the following e-mails
addresses for support:
Sending E-Mail
North America (North America
and Canada)
[email protected]
Central & South America
[email protected]
Europe (Europe, Middle East,
Africa)
[email protected]
Australia (Australian, New
Zealand, Australasia)
[email protected]
China / Asia (China, Asia)
[email protected]
Japan
[email protected]
Brachy Therapy Systems
[email protected]
Copyright 2010 Varian Medical Systems Inc., Oncology Systems
All rights reserved.
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
3 of 118
General Information
Document History
1.5.05
May 11, 2010
Release version for OBI and PortalVision 1.5
1.4.06
Mar 5, 2008
Release version for OBI 1.4
4 of 118
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
Contents
Contents
Contents ............................................................................................................................. 5
List of Figures .................................................................................................................... 7
List of Tables ...................................................................................................................... 7
1. Introduction .................................................................................................................. 9
1.1
1.2
1.3
1.4
1.5
Audience ........................................................................................................................................... 9
Overview ........................................................................................................................................... 9
Remarks ......................................................................................................................................... 11
References ..................................................................................................................................... 11
Abbreviations .................................................................................................................................. 12
2. Networking ................................................................................................................. 13
2.1
2.2
2.3
2.4
Implementation Model .................................................................................................................... 13
2.1.1 Application Data Flow ........................................................................................................ 13
2.1.2 Functional Definition of AEs.............................................................................................. 14
2.1.3 Sequencing of Real World Activities.................................................................................. 15
AE Specifications............................................................................................................................ 16
2.2.1 OBI/PV AI Entity Specification ........................................................................................... 16
Network Interfaces.......................................................................................................................... 32
Configuration .................................................................................................................................. 33
2.4.1 AE Title/Presentation Address Mapping ............................................................................ 33
2.4.2 Parameters ........................................................................................................................ 34
3. Media Interchange ..................................................................................................... 35
4. Support of Character Sets ........................................................................................ 36
5. Security ...................................................................................................................... 37
5.1
5.2
5.3
Security Profiles .............................................................................................................................. 37
Association Level Security ............................................................................................................. 37
5.2.1 OBI/PV AI Client ................................................................................................................ 37
Application Level Security .............................................................................................................. 37
5.3.1 OBI/PV AI Application ........................................................................................................ 37
Appendix A
A.1
A.2
A.3
A.4
Specialization .......................................................................................... 38
IOD Contents .................................................................................................................................. 38
A.1.1 Created SOP Instances ..................................................................................................... 38
A.1.2 Usage of Attributes from received IODs ........................................................................... 38
Data Dictionary of Private Attributes .............................................................................................. 38
Coded Terminology and Templates ............................................................................................... 40
A.3.1 Context Groups.................................................................................................................. 40
A.3.2 Template Specifications .................................................................................................... 40
A.3.3 Private Code Definitions .................................................................................................... 40
Grayscale Image Consistency ........................................................................................................ 40
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
5 of 118
Contents
A.5
A.6
Standard Extended/Specialized/Private SOP Classes .................................................................. 40
Private Transfer Syntaxes .............................................................................................................. 40
Appendix B
B.1
B.2
Supported IODs .............................................................................................................................. 41
B.1.1 Computed Tomography Image A.3.3 ............................................................................. 41
B.1.2 Spatial Registration A.39.1.3 .......................................................................................... 41
B.1.3 RT Structure Set A.19.3 ................................................................................................. 42
B.1.4 RT Image A.17.3 ............................................................................................................ 42
Modules and Attributes................................................................................................................... 45
B.2.1 Patient C.7.1.1 ................................................................................................................ 45
B.2.2 General Study - C.7.2.1 ..................................................................................................... 47
B.2.3 General Series C.7.3.1 ................................................................................................... 54
B.2.4 Frame of Reference C.7.4.1 ........................................................................................... 69
B.2.5 General Equipment C.7.5.1 ............................................................................................ 70
B.2.6 General Image C.7.6.1 ................................................................................................... 71
B.2.7 Image Plane C.7.6.2 ....................................................................................................... 82
B.2.8 Image Pixel C.7.6.3 ........................................................................................................ 82
B.2.9 CT Image C.8.2.1 ........................................................................................................... 86
B.2.10 RT Image C.8.8.2 ........................................................................................................... 91
B.2.11 Structure Set C.8.8.5 .................................................................................................... 104
B.2.12 Modality LUT C.11.1 ..................................................................................................... 110
B.2.13 Spatial Registration C.20.2........................................................................................... 111
Appendix C
C.1
6 of 118
IOD Details .............................................................................................. 41
Extended Interface ............................................................................... 114
RT Plan / RT Ion Plan .................................................................................................................. 114
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
List of Figures
List of Figures
Figure 2-1 Implementation Model..................................................................................................................... 13
Figure 2-2 Sequence Diagram for CT Set loading ........................................................................................... 15
List of Tables
Table 1-1: Network Services used by OBI ......................................................................................................... 9
Table 1-2: Network Services used by PV AI .................................................................................................... 10
Table 1-3: Media File SOP Classes OBI .......................................................................................................... 10
Table 1-4: Media File SOP Classes PV AI ....................................................................................................... 10
Table 2-1 Supported SCU/SCP SOP Classes for OBI/PV AI Entity ................................................................ 16
Table 2-2 DICOM Application Context for OBI/PV AI Client Application Entity ............................................... 17
Table 2-3 Used implementation class UIDs ..................................................................................................... 17
Table 2-4 Object Categories used by OBI/PV AI Client Storage SCU Application Entity ................................ 18
Table 2-5 Presentation contexts used by OBI/PV AI Client Storage SCU Application Entity .......................... 18
Table 2-6 OBI/PV AI Client C-STORE Response Status Handling Behavior .................................................. 19
Table 2-7 OBI/PV AI Client Storage SCU Communication Failure Behavior ................................................... 19
Table 2-8 Supported Query/Retrieve Levels for Query/Retrieve SCU ............................................................. 21
Table 2-9 Proposed Presentation Contexts for OBI/PV AI Client Q/R SCU Application Entity ....................... 21
Table 2-10 OBI/PV AI Client C-FIND Response Status Handling Behavior .................................................... 21
Table 2-11 OBI/PV AI Client Q/R C-FIND SCU Communication Failure Behavior .......................................... 22
Table 2-12 Study Root Request Identifier for OBI/PV AI Client Q/R C-FIND SCU .......................................... 22
Table 2-13 OBI/PV AI Client C-MOVE Response Status Handling Behavior .................................................. 23
Table 2-14 OBI/PV AI Client Q/R C-MOVE SCU Communication Failure Behavior ....................................... 23
Table 2-15 Study Root Request Identifier for OBI/PV AI Client Q/R C-MOVE SCU ....................................... 23
Table 2-16 Supported Operations .................................................................................................................... 24
Table 2-17 Beam Setup kV extensions ............................................................................................................ 28
Table 2-18 Acceptable Presentation Contexts for OBI Client Storage SCP Application Entity and Receive
Storage Request ....................................................................................................................................... 31
Table 2-19 Response Status of OBI/PV AI Client Storage SCP and Receive Storage Request .................... 32
Table 2-20 AE Title Configuration Table .......................................................................................................... 33
Table 2-21 Configuration Parameters Table .................................................................................................... 34
Table A-1 Private Creator Identification Strings ............................................................................................... 40
Table B-2 Computed Tomography Image IOD Modules Support .................................................................... 41
Table B-3 Spatial Registration Image IOD Modules Support ........................................................................... 42
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
7 of 118
List of Tables
Table B-4 RT Structure Set IOD Modules Support .......................................................................................... 42
Table B-5 RT Image IOD Modules Support ..................................................................................................... 43
Table C-6 RT Plan Extended Interface .......................................................................................................... 117
8 of 118
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
Specialization
1.
Introduction
1.1
Audience
This document is intended for the following groups of persons:
Marketing and Sales
System Integrators of medical equipment
Other vendors interfacing using DICOM
It is assumed, that the reader is familiar with the DICOM standard.
The document is structured along the template definition as specified in Part 2 of the DICOM
standard.
1.2
Overview
OBI Advanced Imaging and PortalVision Advanced Imaging are applications that are used to
setup patients for treatment purposes. To support this function the following DICOM services for
receiving and sending diagnostic imaging modalities are supported.
The following table lists network services supported by OBI
SOP Classes
User of Service (SCU)
Provider of Service (SCP)
CT Image Storage
Yes
Yes
Spatial Registration Storage
Yes
No
RT Structure Set Storage
Yes
Yes
RT Image Storage
Yes
Yes
RT Plan Storage
No
Yes
Study Root Query/Retrieve Information
Model FIND
Yes
No
Study Root Query/Retrieve Information
Model MOVE
Yes
No
Transfer
Query/Retrieve
Table 1-1: Network Services used by OBI
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
9 of 118
Specialization
The following table lists network services supported by PortalVision Advanced Imaging (PV AI)
SOP Classes
User of Service (SCU)
Provider of Service (SCP)
CT Image Storage
No
Yes
Spatial Registration Storage
Yes
No
RT Structure Set Storage
Yes
Yes
RT Image Storage
Yes
Yes
RT Plan Storage
No
Yes
Study Root Query/Retrieve Information
Model FIND
Yes
No
Study Root Query/Retrieve Information
Model MOVE
Yes
No
Transfer
Query/Retrieve
Table 1-2: Network Services used by PV AI
The following table lists Media File SOP Classes supported for file based export and import.
These SOP classes are only supported when the OBI application is operating in File mode.
SOP Classes
DICOM Media File Export
DICOM Media File Import
CT Image Storage
Yes
Yes
Spatial Registration Storage
Yes
No
RT Structure Set Storage
Yes
Yes
RT Image Storage
Yes
Yes
RT Plan Storage
No
Yes
Table 1-3: Media File SOP Classes OBI
The following table lists Media File SOP Classes supported for file based export and import.
These SOP classes are only supported when the PV AI application is operating in File mode.
SOP Classes
DICOM Media File Export
DICOM Media File Import
CT Image Storage
No
Yes
Spatial Registration Storage
Yes
No
RT Structure Set Storage
Yes
Yes
RT Image Storage
Yes
Yes
RT Plan Storage
No
Yes
Table 1-4: Media File SOP Classes PV AI
10 of 118
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
Specialization
1.3
Remarks
The scope of this Conformance Statement is to facilitate communication of the OBI/PV AI
application and other vendors medical equipment. The Conformance Statement should be read
and understood in conjunction with the DICOM standard [1].
The DICOM standard in the current version evolved in 1993 with DICOM 3.0. The definition of
the DICOM standard for radiotherapy data started in 1994 and has now reached a mature state.
Nowadays DICOM is the primary choice for exchanging data with an open standard protocol for
the majority of vendors and institutions. Varian Medical Systems is committed to this notion of
standard-based cross-vendor interoperability as well as making use of the DICOM protocol
among its own products.
DICOM, by itself, does not guarantee interoperability. However, the Conformance Statement
facilitates a first-level validation for interoperability between different applications supporting the
same DICOM functionality.
This Conformance Statement is not intended to replace validation with other DICOM equipment
to ensure proper exchange of information intended.
Because the DICOM standard is subject to ongoing changes, enhancements and improvements,
Varian Medical Systems reserves the right to advance their products by making use of upcoming
DICOM features.
1.4
References
[1]
Digital Imaging and Communications in Medicine (DICOM), Parts 1-18 (2008)
National Electrical Manufacturers Association (NEMA)
Rosslyn, VA
United States of America
[2]
Treatment Release 8.8.
DICOM Conformance Statement (P/N 10001678-03)
Varian Medical Systems Inc
Palo Alto, CA, USA
[3]
Varian System Server DICOM Conformance Statement (P/N VA8203D3CS)
Varian Medical Systems International AG
Baden, Switzerland
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
11 of 118
Specialization
1.5
Abbreviations
This section provides the definitions of terms, acronyms and abbreviations that are used
throughout the document
AE
Application Entity
DICOM
Digital Imaging and Communications in Medicine
NEMA
National Electrical Manufacturers Association
SCU
Service Class User
SCP
Service Class Provider
SOP
Service Object Pair
Treatment
System
12 of 118
Management
DICOM entity from which OBI/PV AI retrieves structure set data
OBI
Varians On-Board Imager application to provide Image Guided RT
features.
PV AI
PortalVision Advanced Imaging application to provide Image Guided RT
features for MV only machines.
UID
Unique Identifier used to identify an object by a worldwide unique
identifier (DICOM term)
VR
Value Representation, a data encoding method in DICOM
PDU
Protocol Data Unit (DICOM term)
IOD
Information Object Definition (DICOM term)
DIMSE
DICOM Message Service element
TCP/IP
Transmission Control Protocol / Internet Protocol, a widely used
computer networking protocol
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
Specialization
2.
Networking
2.1
Implementation Model
2.1.1
Application Data Flow
Local User
requests Query
FIND-SCU
OBI Application Entity
Remote
Application Entity
Receives Query
Command
Local User
requests Retrieval
MOVE-SCU
OBI Application Entity
Remote
Application Entity
Receives Retrieve
Command
Object Imported
STORAGE-SCP
OBI Application Entity
Requested
Instances sent by
remote Application
Entity
Local User
requests Storage
STORAGE-SCU
OBI Application Entity
Remote
Application Entity
Receives Storage
Command
DICOM
Standard
Interface
Figure 2-1 Implementation Model
Conceptually, the network services may be modeled as the following separate AEs, though in
fact all the AEs share a single (configurable) AE Title:
STORAGE-SCP, receives incoming RT images, CT images, RT Plans and Structure
Sets.
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
13 of 118
Specialization
2.1.2
STORAGE-SCU, sends outbound RT images, CT images, Structure Sets and Spatial
Registrations.
FIND-SCU, which queries remote AEs for lists of studies, series and instances.
MOVE-SCU, which retrieves selected studies, series or instances.
ECHO-SCU which sends outgoing echo request to Remote Application Entity
Functional Definition of AEs
Note that conceptually the network services may be modeled as the following separate AEs,
though in fact all the AEs share a single (configurable) AE Title.
2.1.2.1
OBI/PV AI Client Storage SCU Application Entity
The Storage SCU Application Entity is invoked when
2.1.2.2
A simple 2D, 2D/2D, marker or 3D/3D matching (OBI only) has been performed and
Spatial Registrations or Markers must be sent to the Remote AE.
A 3D/3D matching (OBI only) has been performed and the acquired CT set must be sent
to the Remote AE
Marker Detection has been performed and the markers must be sent to the Remote AE
as Structure set linked to the current reference CT image. Each marker is defined as
own point set with a single point in it.
Marker Match has been performed, the markers are stored as curves in the image object
and the image object must be sent to the Remote AE.
When the user saves an acquired image through the UI, the image is then sent to the
Remote AE.
The treatment fraction is closed and all unsaved acquired images and other IODs are
sent to the Remote AE.
OBI/PV AI Client FIND SCU, MOVE SCU Application Entity
The Query/Retrieve Application Entity is invoked when
2.1.2.3
The patient is loaded in the 4DITC application, and reference images need to be loaded
in OBI/PV AI
Structure Set and CT images must be loaded in order to perform marker
detection/matching or 3D/3D matching (OBI only).
OBI/PV AI Client Storage SCP Application Entity
The Storage SCP Application Entity is invoked in order to receive objects requested from a
remote Query/Retrieve SCP, such as:
14 of 118
Reference Images/Historical Images
Structure Sets
RT Plan
Reference CT image set
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
Specialization
2.1.3
Sequencing of Real World Activities
When a patient is loaded into the OBI/PV AI application, it loads each reference image/historical
image by requesting them from the remote AE.
CT images are loaded into the OBI/PV AI application typically after the user switches to marker
matching or 3D/3D matching state (OBI only). OBI/PV AI invokes a move command for each CT
Image related to the structure set represented by the CT Image Query/Retrieve SCU. For this
purpose, the OBI application provides the CT Image Storage SCP.
When a Marker Matching is performed, Storage SCU entity is invoked and the Marker Data (as
Structure Set) are stored to the remote AE.
When either a Simple 2D, 2D/2D, 3D/3D (OBI only) or marker matching is performed, the
Storage SCU entity is invoked and stores a Spatial Registration to the remote AE
When loading from or saving to the Remote Application Entity, the ECHO-SCU entity is typically
invoked to ensure that the remote AE is alive and responsive.
The loading of a reference CT set is described by the following sequence diagram, which
illustrates, that the plan and the structure set is retrieved as well to follow the appropriate
references:
User
Management System
OBI/PV AI
Request CT Load
Structure Set request
if the structure set IOD is null,
the Structure Set Instance UID
has changed,the plan from 4DC
is out of date, reload the plan.
Structure Set returned
Plan request
Plan returned
CT Slice request
CT Slice returned
CT Slice request
CT Slice returned
DICOM
Standard
Interface
Figure 2-2 Sequence Diagram for CT Set loading
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
15 of 118
Specialization
2.2
AE Specifications
2.2.1
OBI/PV AI Entity Specification
2.2.1.1
SOP Classes
The OBI/PV AI Entity provides standard conformance to the following DICOM SOP classes.
Abstract Syntax
Name
UID
RT Structure
Set Storage
1.2.840.10008.5.1.4.1.1.481.3
Transfer Syntax
Name List
Role
Extended
Negotiation
UID List
DICOM Implicit VR
Little Endian
1.2.840.10008.1.2
SCU
None
DICOM Implicit VR
Little Endian
1.2.840.10008.1.2
SCP
None
DICOM Explicit VR
Little Endian
1.2.840.10008.1.2.1
SCP
None
Spatial
Registration
Storage
1.2.840.10008.5.1.4.1.1.66.1
DICOM Implicit VR
Little Endian
1.2.840.10008.1.2
SCU
None
CT Image
Storage (OBI
only)
1.2.840.10008.5.1.4.1.1.2
DICOM Implicit VR
Little Endian
1.2.840.10008.1.2
SCU
None
DICOM Implicit VR
Little Endian
1.2.840.10008.1.2
SCP
None
DICOM Explicit VR
Little Endian
1.2.840.10008.1.2.1
SCP
None
DICOM Implicit VR
Little Endian
1.2.840.10008.1.2
SCU
None
DICOM Implicit VR
Little Endian
1.2.840.10008.1.2
SCP
None
DICOM Explicit VR
Little Endian
1.2.840.10008.1.2.1
SCP
None
DICOM Implicit VR
Little Endian
1.2.840.10008.1.2
SCP
None
DICOM Explicit VR
Little Endian
1.2.840.10008.1.2.1
SCP
None
RT Image
Storage
1.2.840.10008.5.1.4.1.1.1
RT Plan
Storage
1.2.840.10008.5.1.4.1.1.5
Study Root
Query/Retriev
e information
model- FIND
1.2.840.10008.5.1.4.1.2.2.1
DICOM Implicit VR
Little Endian
1.2.840.10008.1.2
SCU
None
Study Root
Query/Retriev
e information
model- MOVE
1.2.840.10008.5.1.4.1.2.2.2
DICOM Implicit VR
Little Endian
1.2.840.10008.1.2
SCU
None
Table 2-1 Supported SCU/SCP SOP Classes for OBI/PV AI Entity
16 of 118
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
Specialization
2.2.1.2
Association Policies
2.2.1.2.1
General
The DICOM standard application context name for DICOM 3.0 is always proposed.
Application Context Name
1.2.840.10008.3.1.1.1
Table 2-2 DICOM Application Context for OBI/PV AI Client Application Entity
2.2.1.2.2
Number of Associations
The OBI/PV AI Client Application Entity can initiate (by default) up to three associations at a
time. This value is configurable.
2.2.1.2.3
Asynchronous Nature
OBI/PV AI Client Application Entity does not support asynchronous communication.
2.2.1.2.4
Implementation Identifying Information
The following Implementation Class UIDs are used.Remote AE.
UID Value
UID Name
1.2.246.352.70.2.1.9
VMS Console System 6.5 SCU
(VMS Console Interface 1)
1.2.246.352.70.2.1.60.1
VMS OBI Interface 1
1.2.246.352.70.2.1.60.2
VMS OBI interface 2 Duet
Table 2-3 Used implementation class UIDs
2.2.1.3
Association Initiation Policy
The OBI/PV AI Client Application Entity does initiate Associations.
2.2.1.3.1
Activity Store (STORAGE-SCU)
2.2.1.3.1.1 Description and Sequencing of Activities
When the user performs a Simple 2D, 2D/2D, marker or 3D/3D matching (OBI only), the match
transformation will be stored as a Spatial Registration IOD. The OBI/PV AI administration tool
can be configured to disable the storage of spatial registrations.
When the user performs a marker detection on the reference CT image, the positions of the
detected markers will be stored in the Structure Set and linked to the current reference CT
image. The OBI/PV AI administration tool can be configured to disable the storage of the marker
data.
When the user performs a marker match, the positions of the matched markers will be stored in
the acquired RT Image object as curves. The OBI/PV AI administration tool can be configured to
disable the storage of the marker data The Curve Dimensions (5002,0005) will be 2, the Number
of Points (5002,0010) will be 1, Type of Data (5000,0020) will be POLY, Axis Units (5000,0030)
will be PIXL\PIXL, Axis Labels (5000,0040) will be Marker\Marker and Data Value
Representation (5000,0103) will be 3.
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
17 of 118
Specialization
When the user acquires a verification CT set, each CT Image will be stored. Following isocenters
are stored as Points in the Structure Set which is linked to the acquired verification CT set. The
values of the RT ROI Interpreted Type (3006,00A4) will be as follows:
Initial Laser Isocenter (position before acquisition): INITLASERISO
Acquisition Isocenter (position during acquisition): ACQ_ISOCENTER
Initial Match Isocenter (position after acquisition): INITMATCHISO
The OBI administration tool can be configured to disable the storage of CT images and
isocenters.
The Storage SCU will be invoked and attempts to initiate a new Association. If multiple objects
shall be transferred then multiple C-STORE requests will be issued over this Association.
Object Category
SOP Class Name
SOP Class UID
Structure Set
RT Structure Set Storage
1.2.840.10008.5.1.4.1.1.481.3
Spatial Registration
Spatial Registration Storage
1.2.840.10008.5.1.4.1.1.66.1
RT Image
RT Image Storage
1.2.840.10008.5.1.4.1.1.1
CT Image
CT Image Storage (OBI only)
1.2.840.10008.5.1.4.1.1.2
Table 2-4 Object Categories used by OBI/PV AI Client Storage SCU Application Entity
2.2.1.3.1.2 Proposed Presentation Contexts
OBI/PV AI Client Storage SCU is capable of proposing the Presentation Contexts shown in the
following table.
Presentation Context Table
Abstract Syntax
Name
UID
Transfer Syntax
Name
UID
Role
Ext.
Neg.
Spatial
Registration
Storage
1.2.840.10008.5.1.4.1.1.66.1
DICOM Implicit
VR Little Endian
1.2.840.10008.1.2
SCU
None
RT Structure
Set Storage
1.2.840.10008.5.1.4.1.1.481.3
DICOM Implicit
VR Little Endian
1.2.840.10008.1.2
SCU
None
RT Image
Storage
1.2.840.10008.5.1.4.1.1.1
DICOM Implicit
VR Little Endian
1.2.840.10008.1.2
SCU
None
CT Image
Storage (OBI
only)
1.2.840.10008.5.1.4.1.1.2
DICOM Implicit
VR Little Endian
1.2.840.10008.1.2
SCU
None
Table 2-5 Presentation contexts used by OBI/PV AI Client Storage SCU Application Entity
18 of 118
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
Specialization
2.2.1.3.1.3 SOP Specific Conformance for all Storage SOP Classes
The behavior of OBI/PV AI Client Storage SCU upon storage is summarized in table below
Service
Status
Further Meaning
Error
Code
Behavior
Refused
Out of Resources
A7xx
Failure
Data Set does not
match SOP Class
A9xx
Cannot understand
Cxxx
Coercion of Data
Elements
B000
Data Set does not
match SOP Class
B007
Elements Discarded
B006
Success
Success
0000
The SCP has successfully stored the SOP
Instance. No user feedback is received.
Any other
status
code.
Depending on the type of the Error Code it is
either handled like a Failure or Warning.
However, Error Codes other than listed above
in this table should not occur, as they are not
defined for C-STORE (see [1]), PS 3.4, Table
B.2-1).
Warning
The user is informed that the C-STORE
request has failed. Status Comment
(0000,0902) is logged and displayed along
with additional error information.
The SCP has successfully stored the SOP
Instance. Because a warning status was
received Status Comment (0000,0902) is
logged .
Table 2-6 OBI/PV AI Client C-STORE Response Status Handling Behavior
After sending all SOP Instances, either successfully or not, the Association is released using ARELEASE and the results are presented to the user.
Exception
Behavior
Timeout
The Association is released using A-RELEASE and the
whole task of sending SOP Instances is aborted. The
user is informed about the timeout and available
information about it is logged.
Association aborted by the SCP or
network layers
The whole task of sending SOP Instances is aborted.
Table 2-7 OBI/PV AI Client Storage SCU Communication Failure Behavior
2.2.1.3.1.4 SOP Specific Conformance for RT Structure Set Storage
Detected marker positions will be saved to a new Structure Set. This structure set references the
planning CT frame of reference and contains the detected marker positions as they are projected
on the planning CT. The modified structure set gets a new UID and the private referenced
structure set sequence (3263,1001) is used to denote the original structure set (with the old
UID). This sequence maybe used by the Treatment Management System to combine the new
structure set and the original one to add the markers as detected by OBI/PV AI.
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
19 of 118
Specialization
The ROI Contour Sequence (3006, 0040) shall have the following definitions:
Contour Geometric type
(3006, 0042) = POINT
Number of Contour Points
(3006, 0046) = 1
The ROI observation module shall have the following definitions:
RT ROI Interpreted Type
(3006,00A4) = MARKER.
Marker position may be located between slice positions.
Before exporting modified volumetric structures, they get resampled onto the planes of the 3D
patient model. The volumetric ROI Contours of an exported RT Structure Set will thus always
reference an image slice.
All images used to construct the 3D patient model are referenced in Contour Image Sequence
(3006,0016) of RT Structure Set Module, even if they do not have any contours defined on them.
2.2.1.3.1.5 SOP Specific Conformance for Spatial Registration Storage
The spatial registration of the 2D, 2D/2D and 3D/3D Match will be used in the following way:
The Registration Sequence (0070,0308) will include all images (the reference images
and the acquired verification RT images) and the Frame of Reference UID of the frame
of reference of the reference RT images.
The frame of reference module of the spatial registration will belong to the frame of
reference of the acquired verification RT images.
The transformation matrix type will be RIGID.
The spatial registration of the 3D/3D Match will be used in the following way:
The Registration Sequence (0070,0308) will include all image slices (the reference CT
slices and the acquired verification CT slices) and the Frame of Reference UID of the
frame of reference of the reference CT slices.
The frame of reference module of the spatial registration will belong to the frame of
reference of the acquired verification CT image slices.
The transformation matrix type will be RIGID.
The spatial registration of the Marker Match will be used in the following way:
2.2.1.3.2
The Registration Sequence (0070,0308) will include all image slices (the reference CT
slices and the acquired verification RT images) and the Frame of Reference UID of the
frame of reference of the reference CT slices.
The frame of reference module of the spatial registration will belong to the frame of
reference of the acquired verification RT images.
The transformation matrix type will be RIGID.
Activity Query/Retrieve (FIND-SCU, MOVE-SCU)
2.2.1.3.2.1 Description and Sequencing of Activities
This function is typically used by the application to load the reference images (e.g., DRRs) and
the historical images (Tx images from previous treatment sessions). This function is also used to
load the reference CT and its structure set after the user switches to MarkerMatch (both PVAI
and OBI) or 3D/3D Match state (OBI only). At this time an Association is requested.
20 of 118
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
Specialization
This application supports Query/Retrieve in the SCU role. The table below shows the supported
values for the tag Query/Retrieve Level (0008,0052):
Query/Retrieve Level
Value in (0008,0052)
Composite Object Instance Information
IMAGE
Table 2-8 Supported Query/Retrieve Levels for Query/Retrieve SCU
2.2.1.3.2.2 Proposed Presentation Contexts
Presentation Context Table
Abstract Syntax
Transfer Syntax
Name
Role
Ext.
Neg.
Name
UID
UID
Study Root
Query/Retrieve
information model
FIND
1.2.840.10008.5.1.4.1.2.
2.1
DICOM
Implicit VR
Little Endian
1.2.840.10008.1.2
SCU
None
Study Root
Query/Retrieve
information model
MOVE
1.2.840.10008.5.1.4.1.2.
2.2
DICOM
Implicit VR
Little Endian
1.2.840.10008.1.2
SCU
None
Table 2-9 Proposed Presentation Contexts for OBI/PV AI Client Q/R SCU Application
Entity
2.2.1.3.2.3 SOP Specific Conformance for C-FIND SOP Classes
The behavior of OBI/PV AI Client Query/Retrieve SCU when encountering status codes in a CFIND response is summarized in the table below.
Service
Status
Further Meaning
Error
Code
Behavior
Refused
Out of Resources
A7xx
The object is not found, the user is advised.
Failure
Data Set does not
match SOP Class
A9xx
Cannot understand
Cxxx
Cancel
Matching terminated
due to Cancel
Request
FE00
Cancel is handled like Failure, i.e. the object
is not found and the user is advised.
Success
Matching is
complete No final
Identifier is supplied
0000
The SCP has completed the match, the user
is advised.
Any other
status
code.
Handled like failure.
Table 2-10 OBI/PV AI Client C-FIND Response Status Handling Behavior
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
21 of 118
Specialization
No C-CANCEL-FIND requests are ever issued.
Relational-queries are not supported.
Specific Character Set is not supported. It is not included in a query and will be ignored when
present in the response.
Exception
Behavior
Timeout
The user is informed that the operation (C-FIND or CMOVE) has timed out.
Association aborted by the SCP or
network layers
When the Association is aborted during a C-MOVE
operation the user is informed.
Table 2-11 OBI/PV AI Client Q/R C-FIND SCU Communication Failure Behavior
The table below lists the Attributes supported by the C-FIND SCU.
Name
Tag
VR
Type
Query/Retrieve Level
(0008,0052)
CS
SOP Instance UID
(0008,0018)
UI
Study Instance UID
(0020,000D)
UI
Series Instance UID
(0020,000E)
UI
Table 2-12 Study Root Request Identifier for OBI/PV AI Client Q/R C-FIND SCU
The Types of Matching column in the above table should be read as follows:
S
Single Value Matching
List of UID Matching
Universal Matching
Wild Card Matching
Range Matching
One or more matching types of the above list may be supported per Attribute.
2.2.1.3.2.4 SOP Specific Conformance for C-MOVE SOP Classes
The behavior of OBI/PV AI Client Query/Retrieve SCU when encountering status codes in a CMOVE response is summarized in the following table.
Service
Status
Refused
22 of 118
Further Meaning
Error
Code
Out of Resources
Unable to calculate
number of matches
A701
Out of Resources
Unable to perform
sub-operations
A702
Move Destination
unknown
A801
OBI and PortalVision Advanced Imaging
Behavior
None of the requested SOP Instances could
be retrieved. The user will be informed about
the failure.
DICOM Conformance Statement, KC1505D3CS
Specialization
Service
Status
Failure
Further Meaning
Error
Code
Behavior
Identifier does not
match SOP Class
A900
Unable to Process
Cxxx
Cancel
Sub-operations
terminated due to
Cancel Indication
FE00
Cancel is handled like Failure.
Warning
Sub-operations
complete One or
more Failures
B000
Some or all SOP Instances have not been
transmitted successfully. The user is
informed.
Success
Sub-operations
complete No
Failures
0000
All SOP Instances have successfully been
transmitted.
Pending
Sub-operations are
continuing
FF00
Transferring requested SOP Instances is
continuing. This message is ignored.
Any other
status
code.
Any other status code is handled like Failure.
Table 2-13 OBI/PV AI Client C-MOVE Response Status Handling Behavior
After having received all requested SOP Instances or when the user aborts the operation the
Association is released using A-RELEASE. All events occurring during querying and retrieving
SOP Instances are logged. If any log entries are marked with internal category Error they will be
shown to the user automatically.
No C-CANCEL-MOVE requests are ever issued.
Exception
Behavior
Timeout
The user is informed that the operation (C-FIND or CMOVE) has timed out.
Association aborted by the SCP or
network layers
When the Association is aborted during a C-MOVE
operation the user is informed.
Table 2-14 OBI/PV AI Client Q/R C-MOVE SCU Communication Failure Behavior
The table below lists the Attributes that will be sent by the C-MOVE SCU.
Name
Tag
VR
Type
Query/Retrieve Level
(0008,0052)
CS
SOP Instance UID
(0008,0018)
UI
Study Instance UID
(0020,000D)
UI
Series Instance UID
(0020,000E)
UI
Table 2-15 Study Root Request Identifier for OBI/PV AI Client Q/R C-MOVE SCU
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
23 of 118
Specialization
2.2.1.3.2.5 Supported Operations
This application performs only the operations listed in the table below. This table also shows
which of the optional key values (see [1]) are used for a request.
Operation
Target IOD
Key value used for Request
C-FIND
RT Structure Set
SOP Instance UID
C-MOVE
RT Structure Set
SOP Instance UID
C-MOVE
RT Plan
SOP Instance UID
C-FIND
RT Image
SOP Instance UID
C-MOVE
RT Image
SOP Instance UID
C-MOVE
CT Image
SOP Instance UID
Table 2-16 Supported Operations
2.2.1.4
Association Acceptance Policy
The OBI/PV AI Client Application Entity does accept storage requests for those objects
requested by the MOVE SCU
2.2.1.4.1
Activity Receive Storage Request
2.2.1.4.1.1 Description and Sequencing of Activities
Received SOP Instances are received and cached on the file system, as well as being passed to
the application.
2.2.1.4.1.2 General Preconditions
The following precondition for matching shall be fulfilled:
In a valid OBI/PV AI plan every treatment beam shall contain the same couch translation values
(vertical, lateral, longitudinal). It is allowed to have different couch rotation values in the plan
Tolerances for couch values being interpreted as equal are: absolute 2 mm for longitudinal,
lateral and vertical axes in space.
The plan should contain valid Imaging Device-Specific Acquisition Parameters (300A,00CC) to
allow automatic initiation of imaging procedures. However, if no Verification Image Sequence
items are present or they do not contain device-specific acquisition parameters as specified
above, it is still possible to initiate imaging procedures manually.
OBI/PV AI differentiates between treatment and setup fields. Treatment fields and setup fields
(kV and MV) can be used for OBI/PV AI use cases.
OBI/PV AI supports the Treatment Delivery Type (300A,00CE) SETUP. This attribute should
be the preferred method.
24 of 118
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
Specialization
For backwards compatibility, the parameter FieldType in the Extended Interface (3253,1000) can
be used to indicate the field type for each beam in the plan as well (please note, that this is a
deprecated approach):
<ExtendedVAPlanInterface>
<Beams>
<Beam>
<ReferencedBeamNumber>1</ReferencedBeamNumber>
<BeamExtension>
<FieldType>SETUP</FieldType>
</BeamExtension>
</Beam>
<Beam>
<ReferencedBeamNumber>2</ReferencedBeamNumber>
<BeamExtension>
<FieldType>TREATMENT</FieldType>
</BeamExtension>
</Beam>
</Beams>
</ExtendedVAPlanInterface>
Note that the aforementioned interface is only a portion of the entire interface, which is available
in Section C.1.
If a beam is a setup field and planned verification image sequence contains a single item
representing a kV image (see definition below), this beam is referred to as a kV beam in the
following.
Tag
Description
Beam Sequence
(300A,00B0)
Introduces sequence
of treatment beams
for current RT Plan.
One or more items
may be included in
this sequence.
Introduces sequence
of planned verification
images to be acquired
during current beam.
One or more items
may be included in
this sequence. See
C.8.8.14.2.
>Planned
Verification
Image
Sequence
(300A,00CA)
OBI and PortalVision Advanced Imaging
Convention / Interpretation
Used to plan RT images (kV or MV) or
(CB)CT images to be acquired with this
beam.
DICOM Conformance Statement, KC1505D3CS
25 of 118
Specialization
Tag
Description
Convention / Interpretation
>>Imaging
Device-Specific
Acquisition
Parameters
(300A,00CC)
User-specified
device-specific
parameters which
describe how the
imager will acquire
the image.
This tag is multivalued, values are
separated with a backslash character (\).
If this tag is missing completely, the planned
verification image is interpreted as film.
The following values are currently used for
portal (MV) imaging:
PortImageHighQuality
PortImageHighQuality\BeamOff
PortImageLowDose
PortImageLowDose\BeamOff
PortImageIntegrated
PortImageContinuous
The following values are currently used for
kV imaging:
KV\<ImageType>
Defined terms for ImageType:
Image
CBCT
Examples:
KV\Image
KV\CBCT
>Beam Limiting
Device
Sequence
(300A,00B6)
>Number of
Wedges
(300A,00D0)
>Number of
Compensators
(300A,00E0)
Introduces sequence
of beam limiting
device (jaw or leaf).
For kV beams, this sequence will not be
interpreted.
Number of wedges
associated with
current Beam.
Number of
compensators
associated with
current Beam.
Number of boli
associated with
current Beam.
Number of shielding
blocks associated
with Beam.
Introduces sequence
of machine
configurations
describing treatment
beam. Two or more
items may be
included in this
sequence. See
C.8.8.14.5 and
C.8.8.14.6.
For kV beams, this number shall be 0 (no
accessories)
For kV beams, this number shall be 0 (no
accessories)
For kV beams, this number shall be 0 (no
accessories)
For kV beams, this number shall be 0 (no
accessories)
For kV beams, this sequences shall always
include exactly 2 items.
>Number of Boli
(300A,00ED)
>Number of
Blocks
(300A,00ED)
>Control Point
Sequence
(300A,0111)
26 of 118
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
Specialization
Tag
Description
Convention / Interpretation
>>Beam Limiting
Device Position
Sequence
(300A,011A)
Introduces sequence
of beam limiting
device (collimator)
jaw or leaf (element)
positions. Required
for first item of Control
Point Sequence, or if
Beam Limiting Device
changes during
Beam. One or more
items may be
included in this
sequence.
1C
For kV beams, this sequence will not be
interpreted.
Beam Limiting Device
angle, i.e. orientation
of IEC BEAM
LIMITING DEVICE
coordinate system
with respect to IEC
GANTRY coordinate
system (degrees).
Required for first item
of Control Point
Sequence, or if Beam
Limiting Device Angle
changes during
Beam.
1C
For kV beams, this number shall be 0.0
Direction of Beam
Limiting Device
Rotation when
viewing beam limiting
device (collimator)
from radiation source,
for segment following
Control Point.
Required for first item
of Control Point
Sequence, or if Beam
Limiting Device
Rotation Direction
changes during
Beam. See
C.8.8.14.8.
Enumerated Values:
CW = clockwise
CC = counterclockwise
NONE = no rotation
1C
For kV beams, this value shall be NONE.
>>Beam Limiting
Device Angle
(300A,0120)
>>Beam Limiting
Device Rotation
Direction
(300A,0121)
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
27 of 118
Specialization
Tag
Description
Convention / Interpretation
>>Gantry Angle
(300A,011E)
Gantry angle of
radiation source, i.e.
orientation of IEC
GANTRY coordinate
system with respect
to IEC FIXED
REFERENCE
coordinate system
(degrees). Required
for first item of Control
Point Sequence, or if
Gantry Angle
changes during
Beam.
1C
For kV beams, this value specifies the
orientation of the kV radiation source with
respect to IEC FIXED coordinate system.
For MV beams, this value specifies the
orientation of the MV radiation source (main
gantry) with respect to IEC FIXED
coordinate system.
>>Gantry
Rotation
Direction
(300A,011F)
Direction of Gantry
Rotation when
viewing gantry from
isocenter, for
segment following
Control Point.
Required for first item
of Control Point
Sequence, or if
Gantry Rotation
Direction changes
during Beam. See
C.8.8.14.8.
Enumerated Values:
CW = clockwise
CC = counterclockwise
NONE = no rotation
1C
For kV beams, this value shall be NONE.
Table 2-17 Beam Setup kV extensions
When saving back on board images to the Treatment Management System, all images acquired
for the same plan and session will share the same frame of reference UID, even if the couch
position has been adjusted in between (under the assumption that the patient has not moved on
the couch significantly).
2.2.1.4.1.3 Preconditions for RT Images
RT Images, which are used as reference images (for 2D, 2D/2D and Marker Match), must be
scaled and contain all geometrical parameters needed to determine the position/orientation of
the Image in IEC. They must contain a frame of reference UID and the isocenter position in their
frame of reference.
The only geometrical parameter, which maybe missing is RT Image Position (3002,0012). If
missing, the image is assumed as centered around the z-axis of the Image Receptor system.
28 of 118
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
Specialization
2.2.1.4.1.4 Preconditions for 2D Match
Besides the General Preconditions, the following condition shall be fulfilled:
In a valid OBI/PV AI plan every treatment beam and every reference image shall contain the
same Isocenter Position (300A, 012C). Tolerance for the same Isocenter Position is that
isocenter positions of any pair of beams/reference images within the plan shall deviate less than
1.0 mm. If the plan contains multiple isocenters, the 2D match can be performed, but the couch
shifts cannot be applied.
The plan shall contain at least one valid setup beam configured as a kV or MV image beam or
one valid treatment beam configured as a MV image beam.
2.2.1.4.1.5 Preconditions for 2D/2D Match
Besides the General Preconditions, the following condition shall be fulfilled:
In a valid OBI/PV AI plan every treatment beam and every reference image shall contain no or
the same Isocenter Position (300A, 012C). Tolerance for the same Isocenter Position is that
isocenter positions of any pair of beams/reference images within the plan shall deviate less than
1.0 mm.
The plan shall contain two valid setup beams configured as a kV or MV image beam.
A delta of 90.0 degrees between the gantry/source angles (300A,011E) of the two beams is
recommended for an optimal match.
2.2.1.4.1.6 Preconditions for Marker Match
Besides the General Preconditions, the following condition shall be fulfilled:
In a valid OBI/PV AI plan every treatment beam and every reference image shall contain the
same Isocenter Position (300A, 012C). Tolerance for the same Isocenter Position is that
isocenter positions of any pair of beams/reference images within the plan shall deviate less than
1.0 mm
The plan shall contain two valid setup beams configured as a kV or MV image beam.
A delta of 90.0 degrees between the gantry/source angles (300A,011E) of the two beams is
recommended for an optimal match.
To load the slices of the reference CT image OBI/PV AI loads the plans structure set, which
corresponds to the Referenced RT Structure Set (300C,0060), from the Treatment Management
System , and subsequently loads all CT slices referenced in the Contour Image Sequence
(3006,0016).
Conditions for CT Slices:
All CT slices shall have the same Frame of Reference.
The spacing between CT slices positions in z-direction (DICOM Patient coordinate
system) may vary, although it is recommended, that they are equal throughout all slices
referenced by the Structure Set.
The maximal difference in x-direction and in y-direction (DICOM Patient coordinate
system) between all CT slices shall be less or equal 0.1 mm.
Conditions for Structure Set (except Structures of Contour Geometric type (3006, 0042)
POINT):
The contour points within a contour shall be positioned on the same contour plane within
-6
a maximal Epsilon of 10 mm in z direction (DICOM Patient coordinate system).
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
29 of 118
Specialization
Structures with contour(s) having contour points that are not positioned on the same
contour plane (see point above, i.e. non-transversal structures) are ignored (they are not
imported and a corresponding warning message is issued) and the import will continue
with the other structures.
Contour Geometric Type (3006, 0042) has to be either CLOSED_PLANAR or
OPEN_PLANAR for all contours of the structure. Structures with mixed contour types are
not supported.
Contour Geometric type (3006, 0042) POINT dont need to fulfill the conditions
mentioned above.
Conditions for Marker Point Structures:
The markers shall be in the structure set referenced by the plan.
Marker position may be located between slice positions.
The ROI Contour Sequence (3006, 0040) shall have the following definitions:
o Contour Geometric type
(3006, 0042) = POINT
o
Number of Contour Points (3006, 0046) = 1
The ROI observation module shall have the following definitions:
o RT ROI Interpreted Type (3006,00A4) = MARKER.
o
Private tag: Marker Subtype (3271,1000) = MARKER
2.2.1.4.1.7 Preconditions for 3D/3D Match (OBI only)
Besides the General Preconditions, the following condition shall be fulfilled:
In a valid OBI plan every treatment beam and every reference image shall contain the same
Isocenter Position (300A, 012C). Tolerance for the same Isocenter Position is that isocenter
positions of any pair of beams/reference images within the plan shall deviate less than 1.0 mm.
The plan shall contain a valid setup beam. The tag Beam Type (300A,00C4) shall have the
value STATIC configured as a CBCT beam. The tag Imaging Device Specific Acquisition
Parameters (300A,00CC) shall have the value KV\CBCT.
To load the slices of the reference CT image OBI loads the plans structure set, which
corresponds to the Referenced RT Structure Set (300C,0060), from the Treatment Management
System , and subsequently loads all CT slices referenced in the Contour Image Sequence
(3006,0016).
Conditions for CT Slices:
All CT slices shall have the same Frame of Reference.
The spacing between CT slices positions in z-direction (DICOM Patient coordinate
system) may vary, although it is recommended, that they are equal throughout all slices
referenced by the Structure Set.
The maximal difference in x-direction and in y-direction (DICOM Patient coordinate
system) between all CT slices shall be less or equal 0.1 mm.
Conditions for Structure Set (except Structures of Contour Geometric type (3006, 0042)
POINT):
30 of 118
The contour points within a contour shall be positioned on the same contour plane within
-6
a maximal Epsilon of 10 in z direction (DICOM Patient coordinate system).
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
Specialization
Structures with contour(s) having contour points that are not positioned on the same
contour plane (see point above, i.e. non-transversal structures) are ignored (they are not
imported and a corresponding warning message is issued) and the import will continue
with the other structures.
Contour Geometric Type (3006, 0042) has to be either CLOSED_PLANAR or
OPEN_PLANAR for all contours of the structure. Structures with mixed contour types are
not supported.
Contour Geometric type (3006, 0042) POINT dont need to fulfill the conditions
mentioned above.
2.2.1.4.1.8 Accepted Presentation Contexts
OBI Client Storage SCP Application Entity accepts Presentation Contexts shown in the following
table.
Presentation Context Table
Abstract Syntax
Name
All Storage
SOP Classes
in Table 4-1
Transfer Syntax
UID
All Storage SOP
Classes in Table
4-1
Name
Role
Ext.
Neg.
SCP
None
UID
DICOM Implicit VR
Little Endian
1.2.840.10008.1.2
DICOM Explicit VR
Little Endian
1.2.840.10008.1.2.1
Table 2-18 Acceptable Presentation Contexts for OBI Client Storage SCP Application
Entity and Receive Storage Request
2.2.1.4.1.9 SOP Specific Conformance for all Storage SOP Classes
2.2.1.4.1.9.1 Presentation Context Acceptance Criterion
OBI/PV AI Client Storage SCP will always accept any Presentation Context for the supported
SOP Classes with the supported Transfer Syntaxes. More than one proposed Presentation
Context will be accepted for the same Abstract Syntax if the Transfer Syntax is supported,
whether or not it is the same as another Presentation Context.
2.2.1.4.1.9.2 Transfer Syntax Selection Policies
If offered a choice of Transfer Syntaxes in a Presentation Context, OBI/PV AI Client Storage
SCP will select the first Transfer Syntax that is listed in the Presentation Context.
OBI/PV AI Client Storage SCP will accept duplicate Presentation Contexts, that is, if it is offered
multiple Presentation Contexts, each of which offers acceptable Transfer Syntaxes, it will accept
all Presentation Contexts, applying the same rule for selecting a Transfer Syntax for each as
described above.
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
31 of 118
Specialization
2.2.1.4.1.9.3 Response Status
OBI/PV AI Client Storage SCP will behave as described in the Table below when generating the
C-STORE response command message.
Service
Status
Refused
Failure
Warning
Success
Further Meaning
Error
Code
Out of Resources
A700
Out of Resources
A7xx
Invalid object
instance
0117
Data Set does not
match SOP Class
A9xx
Invalid C-STORE
request
C000
Coercion of Data
Elements
B000
Data Set does not
match SOP Class
B007
Elements Discarded
B006
0000
Behavior
Failed to receive incoming DICOM Stream.
Problem is logged on the SCP side.
Image successfully stored in temporary file.
Instance successfully stored in temporary file.
Table 2-19 Response Status of OBI/PV AI Client Storage SCP and Receive Storage
Request
2.3
Network Interfaces
See Varian System Server DICOM Conformance Statement [3]
32 of 118
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
Specialization
2.4
Configuration
2.4.1
AE Title/Presentation Address Mapping
2.4.1.1
Local AE Titles
The local OBI/PV AI Client Application Entity uses the AE Title and TCP/IP port number
configured via the Daemon configuration option in the OBI/PV AI Admin tool. Note that
conceptually the network services have been modeled as separate AEs, though in fact all the
AEs share a single (configurable) AE Title.
Application Entity
Default AE Title
Default TCP/IP Port
OBI/PV AI Client Application Entity
StreamService
(configurable)
58051 (configurable)
OBI/PV AI Client Storage SCU
See above
See above
OBI/PV AI Client Query/Retrieve SCU
See above
See above
Table 2-20 AE Title Configuration Table
2.4.1.2
Remote AE Title/Presentation Address Mapping
2.4.1.2.1
OBI/PV AI Client Storage SCU
For the Storage service class SCU, TCP/IP address, called AE title and port number of the
destination are configurable as well as the calling AE title used by OBI/PV AI.
2.4.1.2.2
OBI/PV AI Client Query/Retrieve SCU
For the OBI/PV AI Client Query/Retrieve SCU, TCP/IP address, called AE title and port number
of the provider are configurable. The calling AE title of the local application, which is also the
Move Destination AE title, is configurable too. The local port is the port of the Storage SCP for
receiving the data. The Storage SCP will accept only connection requests from the configured
remote AE Title and IP Address.
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
33 of 118
Specialization
2.4.2
Parameters
The following table shows DICOM relevant configuration parameters. While some of them may
be configured directly via a configuration dialog or a configuration application, other values can
only be accessed via the Registry or can not be configured at all.
Parameter
Configurable
(Yes/No)
Default Value
General Parameters
Max PDU Receive Size
No
Unlimited
Max PDU Send Size
No
32768 Bytes (32kB)
Time-out waiting for a acceptance or rejection
response to an Association request or Association
Release request (Application Level Timeout)
Yes
300s
Spatial Registration formatted along IHE-RO
specification 2008
Yes
No
OBI/PV AI Client Storage SCU Specific Parameters
OBI/PV AI Client Query/Retrieve SCU Specific Parameters
Time-out awaiting a Response to a DIMSE Request
(Low-Level Timeout)
Yes
300s
Maximum number of simultaneously initiated
Associations
No
Table 2-21 Configuration Parameters Table
34 of 118
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
Specialization
3.
Media Interchange
OBI/PV AI Client Application Entity allows importing and exporting DICOM Media Files. Various
import and export filters are available in the OBI/PV AI Client application to read and create
DICOM Media Files. This functionality is only available when the OBI/PV AI Application is
operating in file mode.
The files fully conform to the Part 10 format. However, the Basic Directory IOD as defined in
Media Interchange Application Profiles is not present.
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
35 of 118
Specialization
4.
Support of Character Sets
See Varian System Server DICOM Conformance Statement [3]
36 of 118
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
Specialization
5.
Security
5.1
Security Profiles
No Security Profiles are supported.
5.2
Association Level Security
5.2.1
OBI/PV AI Client
Storage SCU does not support Association Level Security.
The Storage SCP instantiated to receive Instances requested by Query/Retrieve SCU checks
the following additional values when determining whether to accept Association Open Requests:
Called AE Title
Calling AE Title
IP address of Association Request originator
5.3
Application Level Security
5.3.1
OBI/PV AI Application
In order to load patient information into the OBI/PV AI application, the user opens the patient in
the 4DITC application, which requires the user identification in the form of a user name and
password.
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
37 of 118
Specialization
Appendix A Specialization
A.1
IOD Contents
A.1.1
Created SOP Instances
IODs created by the OBI/PV AI Application Entity are listed in IOD Details.
A.1.2
Usage of Attributes from received IODs
The OBI/PV AI Application Entity require all Type 1 Attributes to be present.
A.2
Data Dictionary of Private Attributes
The following table contains a list of private Attributes that are used throughout all Application
Entities. The OBI/PV AI application does not use any private attributes that are not outlined in
this document or [3].
38 of 118
Tag
Name
VR
VM
(300B,ee01)
Distal Target Distance Tolerance
FL
(300B,ee02)
Maximum Collimated Field Diameter (Field Size)
FL
(300B,ee04)
Range (Planned Distal Target Distance)
FL
(300B,ee08)
Beam Line Data Table Version
SH
(300B,ee0E)
Nominal SOBP Width
FL
(300B,ee11)
Nominal SOBP Width Tolerance
FL
(300B,ee17)
Peek Range
FL
(3241,ee00)
Internal Image Id
LO
(3241,ee02)
Internal Pixel Offset
IS
(3241,ee03)
Not Calibrated
US
(3241,ee04)
Image Orientation (Patient)
DS
(3241,ee05)
Image Position (Patient)
DS
(3241,ee06)
Pixel Spacing
DS
(3243,ee09)
Beam Secondary Name
SH
(3243,ee28)
Related Referenced Beam Number
IS
(3247,ee00)
Session Termination Status
SH
(3249,ee00)
Maximum Treatment Time
DS
(3249,ee10)
Referenced Primary Dose Reference UID
UI
(3251,ee00)
Treatment Machine Note
LO
(3253,ee00)
Extended Interface Data
OB
(3253,ee01)
Extended Interface Length
IS
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
Specialization
Tag
Name
VR
VM
(3253,ee02)
Extended Interface Format
LO
(3257,ee01)
Plan Type
Enumerated Values:
ALL_PLANS
PREDECESSOR
NOT_ PREDECESSOR
CS
(3257,ee10)
Reference Image Type
Enumerated Values:
ALL
PRIMARY
CS
(3259,ee00)
Additional Dose Value Sequence
SQ
0..n
(3259,ee02)
Actual Session Dose
DS
(3259,ee04)
Daily Dose
DS
(3259,ee06)
Life Time Total Dose
DS
(3261,ee2A)
Treatment Session Beam Data
OB
(3263,ee00)
Overlap Type
Enumerated Values:
OVERLAPPING
NONOVERLAPPING
CS
(3263,ee01)
Referenced Structure Set Relationship Sequence
SQ
(3263,ee02)
Structure Set Relationship
Defined Terms:
PREDECESSOR
ADDITION
CS
(3265,ee00)
Total Number of Fractions
SL
(3265,ee01)
Last treated Fraction
SL
(3267,ee00)
Referenced Patient Volume Id
SH
(3271,ee00)
Marker Subtype
Enumerated Values:
MARKER
POINT_MATCH
REFERENCE_LINE
CS
(3273,ee00)
RT Image Isocenter Position
DS
(3273,ee01)
RT Image Patient Position
CS
(3275,ee00)
Registration Sub Type
LO
(3277,ee00)
RT Motion Synchronization Sequence
SQ
(3285,ee00)
Primary Fluence Mode Sequence
SQ
(3285,ee01)
Fluence Mode
CS
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
39 of 118
Specialization
Tag
Name
VR
VM
(3285,ee02)
Fluence Mode ID
SH
(3287,ee00)
Plan Integrity Sequence
SQ
(3287,ee01)
Plan Integrity Hash
LO
(3287,ee02)
Plan Integrity Hash Version
SH
Table A-1 Private Creator Identification Strings
A.3
Coded Terminology and Templates
A.3.1
Context Groups
Information will be published in a future version of this document.
A.3.2
Template Specifications
No standard templates are extended and no private templates are used.
A.3.3
Private Code Definitions
There are no private code definitions.
A.4
Grayscale Image Consistency
Not supported.
A.5
Standard Extended/Specialized/Private SOP
Classes
Not Used.
A.6
Private Transfer Syntaxes
No private Transfer Syntaxes are used.
40 of 118
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
IOD Details
Appendix B IOD Details
B.1
Supported IODs
For better reference with DICOM Standard ([1], section titles of the following sub-sections are
listed with section number of DICOM Standard Part 3 where the corresponding IOD Module table
can be found.
B.1.1
Computed Tomography Image A.3.3
IE
Reference
Usage
Patient
C.7.1.1
Clinical Trial Subject
C.7.1.3
General Study
C.7.2.1
Patient Study
C.7.2.2
Not supported
Clinical Trial Study
C.7.2.3
Not supported
General Series
C.7.3.1
Clinical Trial Series
C.7.3.2
Frame of Reference
Frame of Reference
C.7.4.1
Equipment
General Equipment
C.7.5.1
Image
General Image
C.7.6.1
Image Plane
C.7.6.2
Image Pixel
C.7.6.3
Contrast/Bolus
C.7.6.4
CT Image
C.8.2.1
Overlay Plane
C.9.2
VOI LUT
C.11.2
SOP Common
C.12.1
Patient
Study
Series
Module
Presence
Not supported
Not supported
Not supported
Not supported
Table B-2 Computed Tomography Image IOD Modules Support
B.1.2
Spatial Registration A.39.1.3
IE
Patient
Study
Module
Reference
Usage
Patient
C.7.1.1
Clinical Trial Subject
C.7.1.3
General Study
C.7.2.1
Patient Study
C.7.2.2
Not supported
Clinical Trial Study
C.7.2.3
Not supported
OBI and PortalVision Advanced Imaging
Presence
Not supported
DICOM Conformance Statement, KC1505D3CS
41 of 118
IOD Details
IE
Reference
Usage
General Series
C.7.3.1
Clinical Trial Series
C.7.3.2
Spatial Registration
Series
C.20.1
Frame of Reference
Frame of Reference
C.7.4.1
Equipment
General Equipment
C.7.5.1
Spatial Registration
Spatial Registration
C.20.2
Common Instance
Reference
C.12.2
SOP Common
C.12.1
Series
Module
Presence
Not supported
Table B-3 Spatial Registration Image IOD Modules Support
B.1.3
RT Structure Set A.19.3
IE
Reference
Usage
Patient
C.7.1.1
Clinical Trial Subject
C.7.1.3
General Study
C.7.2.1
Patient Study
C.7.2.2
Not supported
Clinical Trial Study
C.7.2.3
Not supported
RT Series
C.8.8.1
Clinical Trial Series
C.7.3.2
Equipment
General Equipment
C.7.5.1
Structure Set
Structure Set
C.8.8.5
ROI Contour
C.8.8.6
RT ROI Observations
C.8.8.8
Approval
C.8.8.16
C.12.1
Reference
Usage
Patient
C.7.1.1
Clinical Trial Subject
C.7.1.3
General Study
C.7.2.1
Patient
Study
Series
Module
SOP Common
Presence
Not supported
Not supported
Not supported
Table B-4 RT Structure Set IOD Modules Support
B.1.4
RT Image A.17.3
IE
Patient
Study
42 of 118
Module
OBI and PortalVision Advanced Imaging
Presence
Not supported
DICOM Conformance Statement, KC1505D3CS
IOD Details
IE
Reference
Usage
Patient Study
C.7.2.2
Not supported
Clinical Trial Study
C.7.2.3
Not supported
RT Series
C.8.8.1
Clinical Trial Series
C.7.3.2
Not supported
Frame of Reference
Frame of Reference
C.7.4.1
Supported and
required for
Import
Equipment
General Equipment
C.7.5.1
Image
General Image
C.7.6.1
Image Pixel
C.7.6.3
Contrast/Bolus
C.7.6.4
Not supported
Cine
C.7.6.5
Not supported
Multi Frame
C.7.6.6
Not supported
RT Image
C.8.8.2
Modality LUT
C.11.1
VOI LUT
C.11.2
Approval
C.8.8.16
Curve (Retired)
C.10.2
SOP Common
C.12.1
N/A
Series
Extended Interface
Module
Extended Interface
Presence
Private
Table B-5 RT Image IOD Modules Support
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
43 of 118
IOD Details
(This page is intentionally left blank.)
44 of 118
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
IOD Details, Patient C.7.1.1
B.2
Modules and Attributes
For better reference with DICOM Standard ([1]section titles of the following sub-sections are decorated with section number of DICOM Standard
Part 3 where the corresponding Module Attribute table can be found.
The first four columns in the following tables contain definitions from the DICOM standard. References in those columns refer to the DICOM
standard and not to this document.
The Handling column describes for each attribute whether it is supported and to which value in the OBI system it maps. A value of *** export not
supported indicates that an attribute is unknown and therefore ignored by the application.
The application creates objects of the type RT Image, CT Image, RT Structure Set and Spatial Registration. These objects are exported through
DICOM. Therefore only the export is described in the following tables.
B.2.1
Patient C.7.1.1
Attribute Name
Tag
Description
Handling
Patients Name
(0010,0010)
Patient's full name.
Export supported for:
All
Patient ID
(0010,0020)
Primary hospital identification number or code for the
patient.
Export supported for:
All
Issuer of Patient ID
(0010,0021)
Identifier of the Assigning Authority that issued the
Patient ID.
*** export not supported
Patient's Birth Date
(0010,0030)
Birth date of the patient.
Export supported for:
All
Patient's Sex
(0010,0040)
Sex of the named patient.
Enumerated Values:
M = male
F = female
O = other
Export supported for:
All
Female: F, Male: M, anything else: O
Referenced Patient
Sequence
(0008,1120)
A sequence that provides reference to a Patient SOP
Class/Instance pair. Only a single Item shall be permitted
in this Sequence.
*** export not supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
45 of 118
IOD Details, Patient C.7.1.1
46 of 118
Attribute Name
Tag
Description
Handling
> Referenced SOP
Class UID
(0008,1150)
1C
Uniquely identifies the referenced SOP Class. Required if
Referenced Patient Sequence (0008,1120) is sent.
*** export not supported
> Referenced SOP
Instance UID
(0008,1155)
1C
Uniquely identifies the referenced SOP Instance.
Required if Referenced Patient Sequence (0008,1120) is
sent.
*** export not supported
Patient's Birth Time
(0010,0032)
Birth time of the Patient.
Export supported for:
All
Other Patient IDs
(0010,1000)
Other identification numbers or codes used to identify the
patient.
*** export not supported
Other Patient
Names
(0010,1001)
Other names used to identify the patient.
*** export not supported
Ethnic Group
(0010,2160)
Ethnic group or race of the patient.
*** export not supported
Patient Comments
(0010,4000)
User-defined additional information about the patient.
*** export not supported
Patient Identify
Removed
(0012,0062)
The true identity of the patient has been removed from
the Attributes and the Pixel Data
Enumerated Values:
YES
NO
*** export not supported
De-identification
Method
(0012,0063)
1C
A description or label of the mechanism or method use to
remove the patient's identity. May be multi-valued if
successive de-identification steps have been performed.
Note:
This may be used to describe the extent or thoroughness
of the de- identification, for example whether or not the
de-identification is for a "Limited Data Set" (as per HIPAA
Privacy Rule).
Required if Patient Identity Removed (0012,0062) is
present and has a value of YES and De-identification
Method Code Sequence (0012,0064) is not present.
*** export not supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
IOD Details, General Study - C.7.2.1
B.2.2
Attribute Name
Tag
Description
Handling
De-identification
Method Code
Sequence
(0012,0064)
1C
A code describing the mechanism or method use to
remove the patient's identity. One or more Items shall be
present. Multiple items are used if successive deidentification steps have been performed
Required if Patient Identity Removed (0012,0062) is
present and has a value of YES and De-identification
Method (0012,0063) is not present.
*** export not supported
> Code Value
(0008,0100)
1C
See Section 8.1. Required if a sequence item is present.
*** export not supported
> Coding Scheme
Designator
(0008,0102)
1C
See Section 8.2. Required if a sequence item is present.
*** export not supported
> Coding Scheme
Version
(0008,0103)
1C
See Section 8.2. Required if a sequence item is present
and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.
*** export not supported
> Code Meaning
(0008,0104)
1C
See Section 8.3. Required if a sequence item is present.
*** export not supported
General Study - C.7.2.1
Attribute Name
Tag
Description
Handling
Study Instance UID
(0020,000D)
Unique identifier for the Study.
Export supported for:
All
Study Date
(0008,0020)
Date the Study started.
Export supported for:
All
Study Time
(0008,0030)
Time the Study started.
Export supported for:
All
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
47 of 118
IOD Details, General Study - C.7.2.1
48 of 118
Attribute Name
Tag
Description
Handling
Referring
Physicians Name
(0008,0090)
Name of the patient's referring physician
*** export not supported
Referring Physician
Identification
Sequence
(0008,0096)
Identification of the patient's referring physician. Only a
single item shall be permitted in this sequence.
*** export not supported
> Person
Identification Code
Sequence
(0040,1101)
A coded entry which identifies a person. The Code
Meaning attribute, though it will be encoded with a VR of
LO, may be encoded according to the rules of the PN VR
(e.g. caret '^' delimiters shall separate name
components), except that a single component (i.e. the
whole name unseparated by caret delimiters) is not
permitted. Name component groups for use with multibyte character sets are permitted, as long as they fit
within the 64 characters (the length of the LO VR). One
or more Items may be permitted in this Sequence.
*** export not supported
>> Code Value
(0008,0100)
1C
See Section 8.1. Required if a sequence item is present.
*** export not supported
>> Coding Scheme
Designator
(0008,0102)
1C
See Section 8.2. Required if a sequence item is present.
*** export not supported
>> Coding Scheme
Version
(0008,0103)
1C
See Section 8.2. Required if a sequence item is present
and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.
*** export not supported
>> Code Meaning
(0008,0104)
1C
See Section 8.3. Required if a sequence item is present.
*** export not supported
> Persons Address
(0040,1102)
Person's mailing address
*** export not supported
> Persons
Telephone Numbers
(0040,1103)
Person's telephone number(s)
*** export not supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
IOD Details, General Study - C.7.2.1
Attribute Name
Tag
Description
Handling
> Institution Name
(0008,0080)
1C
Institution or organization to which the identified individual
is responsible or accountable. Shall not be present if
Institution Code Sequence (0008,0082) is present.
*** export not supported
> Institution Address
(0008,0081)
Mailing address of the institution or organization to which
the identified individual is responsible or accountable.
*** export not supported
> Institution Code
Sequence
(0008,0082)
1C
Institution or organization to which the identified individual
is responsible or accountable. Shall not be present if
Institution Name (0008,0080) is present. Only a single
Item shall be permitted in this Sequence.
*** export not supported
>> Code Value
(0008,0100)
1C
See Section 8.1. Required if a sequence item is present.
*** export not supported
>> Coding Scheme
Designator
(0008,0102)
1C
See Section 8.2. Required if a sequence item is present.
*** export not supported
>> Coding Scheme
Version
(0008,0103)
1C
See Section 8.2. Required if a sequence item is present
and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.
*** export not supported
>> Code Meaning
(0008,0104)
1C
See Section 8.3. Required if a sequence item is present.
*** export not supported
Study ID
(0020,0010)
User or equipment generated Study identifier.
Export supported for:
All
Accession Number
(0008,0050)
A RIS generated number that identifies the
order for the Study.
Export supported for:
All
Study Description
(0008,1030)
Institution-generated description or classification of the
Study (component) performed.
*** export not supported
Physician(s) of
Record
(0008,1048)
Names of the physician(s) who are responsible for overall
patient care at time of Study (see Section C.7.3.1 for
Performing Physician)
*** export not supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
49 of 118
IOD Details, General Study - C.7.2.1
50 of 118
Attribute Name
Tag
Description
Handling
Physician(s) of
Record Identification
Sequence
(0008,1049)
Identification of the physician(s) who are responsible for
overall patient care at time of Study. One or more items
shall be included in this sequence. If more than one Item,
the number and order shall correspond to the value of
Physician(s) of Record (0008,1048), if present.
*** export not supported
> Person
Identification Code
Sequence
(0040,1101)
A coded entry which identifies a person. The Code
Meaning attribute, though it will be encoded with a VR of
LO, may be encoded according to the rules of the PN VR
(e.g. caret '^' delimiters shall separate name
components), except that a single component (i.e. the
whole name unseparated by caret delimiters) is not
permitted. Name component groups for use with multibyte character sets are permitted, as long as they fit
within the 64 characters (the length of the LO VR). One
or more Items may be permitted in this Sequence.
*** export not supported
>> Code Value
(0008,0100)
1C
See Section 8.1. Required if a sequence item is present.
*** export not supported
>> Coding Scheme
Designator
(0008,0102)
1C
See Section 8.2. Required if a sequence item is present.
*** export not supported
>> Coding Scheme
Version
(0008,0103)
1C
See Section 8.2. Required if a sequence item is present
and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.
*** export not supported
>> Code Meaning
(0008,0104)
1C
See Section 8.3. Required if a sequence item is present.
*** export not supported
> Persons Address
(0040,1102)
Person's mailing address
*** export not supported
> Persons
Telephone Numbers
(0040,1103)
Person's telephone number(s)
*** export not supported
> Institution Name
(0008,0080)
1C
Institution or organization to which the identified individual
is responsible or accountable. Shall not be present if
Institution Code Sequence (0008,0082) is present.
*** export not supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
IOD Details, General Study - C.7.2.1
Attribute Name
Tag
Description
Handling
> Institution Address
(0008,0081)
Mailing address of the institution or organization to which
the identified individual is responsible or accountable.
*** export not supported
> Institution Code
Sequence
(0008,0082)
1C
Institution or organization to which the identified individual
is responsible or accountable. Shall not be present if
Institution Name (0008,0080) is present. Only a single
Item shall be permitted in this Sequence.
*** export not supported
>> Code Value
(0008,0100)
1C
See Section 8.1. Required if a sequence item is present.
*** export not supported
>> Coding Scheme
Designator
(0008,0102)
1C
See Section 8.2. Required if a sequence item is present.
*** export not supported
>> Coding Scheme
Version
(0008,0103)
1C
See Section 8.2. Required if a sequence item is present
and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.
*** export not supported
>> Code Meaning
(0008,0104)
1C
See Section 8.3. Required if a sequence item is present.
*** export not supported
Name of
Physician(s)
Reading Study
(0008,1060)
Names of the physician(s) reading the Study.
*** export not supported
Physician(s)
Reading Study
Identification
Sequence
(0008,1062)
Identification of the physician(s) reading the Study. One
or more items shall be included in this sequence. If more
than one Item, the number and order shall correspond to
the value of Name of Physician(s) Reading Study
(0008,1060), if present.
*** export not supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
51 of 118
IOD Details, General Study - C.7.2.1
52 of 118
Attribute Name
Tag
Description
Handling
> Person
Identification Code
Sequence
(0040,1101)
A coded entry which identifies a person. The Code
Meaning attribute, though it will be encoded with a VR of
LO, may be encoded according to the rules of the PN VR
(e.g. caret '^' delimiters shall separate name
components), except that a single component (i.e. the
whole name unseparated by caret delimiters) is not
permitted. Name component groups for use with multibyte character sets are permitted, as long as they fit
within the 64 characters (the length of the LO VR). One
or more Items may be permitted in this Sequence.
*** export not supported
>> Code Value
(0008,0100)
1C
See Section 8.1. Required if a sequence item is present.
*** export not supported
>> Coding Scheme
Designator
(0008,0102)
1C
See Section 8.2. Required if a sequence item is present.
*** export not supported
>> Coding Scheme
Version
(0008,0103)
1C
See Section 8.2. Required if a sequence item is present
and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.
*** export not supported
>> Code Meaning
(0008,0104)
1C
See Section 8.3. Required if a sequence item is present.
*** export not supported
> Persons Address
(0040,1102)
Person's mailing address
*** export not supported
> Persons
Telephone Numbers
(0040,1103)
Person's telephone number(s)
*** export not supported
> Institution Name
(0008,0080)
1C
Institution or organization to which the identified individual
is responsible or accountable. Shall not be present if
Institution Code Sequence (0008,0082) is present.
*** export not supported
> Institution Address
(0008,0081)
Mailing address of the institution or organization to which
the identified individual is responsible or accountable.
*** export not supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
IOD Details, General Study - C.7.2.1
Attribute Name
Tag
Description
Handling
> Institution Code
Sequence
(0008,0082)
1C
Institution or organization to which the identified individual
is responsible or accountable. Shall not be present if
Institution Name (0008,0080) is present. Only a single
Item shall be permitted in this Sequence.
*** export not supported
>> Code Value
(0008,0100)
1C
See Section 8.1. Required if a sequence item is present.
*** export not supported
>> Coding Scheme
Designator
(0008,0102)
1C
See Section 8.2. Required if a sequence item is present.
*** export not supported
>> Coding Scheme
Version
(0008,0103)
1C
See Section 8.2. Required if a sequence item is present
and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.
*** export not supported
>> Code Meaning
(0008,0104)
1C
See Section 8.3. Required if a sequence item is present.
*** export not supported
Referenced Study
Sequence
(0008,1110)
A sequence that provides reference to a Study SOP
Class/Instance pair. The sequence may have zero or
more Items.
*** export not supported
> Referenced SOP
Class UID
(0008,1150)
1C
Uniquely identifies the referenced SOP Class. Required if
Referenced Study Sequence (0008,1110) is sent.
*** export not supported
> Referenced SOP
Instance UID
(0008,1155)
1C
Uniquely identifies the referenced SOP Instance.
Required if Referenced Study Sequence (0008,1110) is
sent.
*** export not supported
Procedure Code
Sequence
(0008,1032)
A Sequence that conveys the type of procedure
performed. One or more Items may be included in this
Sequence.
*** export not supported
> Code Value
(0008,0100)
1C
See Section 8.1. Required if a sequence item is present.
*** export not supported
> Coding Scheme
Designator
(0008,0102)
1C
See Section 8.2. Required if a sequence item is present.
*** export not supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
53 of 118
IOD Details, General Series C.7.3.1
B.2.3
54 of 118
Attribute Name
Tag
Description
Handling
> Coding Scheme
Version
(0008,0103)
1C
See Section 8.2. Required if a sequence item is present
and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.
*** export not supported
> Code Meaning
(0008,0104)
1C
See Section 8.3. Required if a sequence item is present.
*** export not supported
General Series C.7.3.1
Attribute Name
Tag
Description
Handling
Modality
(0008,0060)
Type of equipment that originally acquired the data used
to create the images in this Series. See C.7.3.1.1.1 for
Defined Terms.
Export supported for:
All
Series Instance UID
(0020,000E)
Unique identifier of the Series.
Export supported for:
All
Series Number
(0020,0011)
A number that identifies this Series.
Export supported for:
All
The value is set to 0 for all objects
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
IOD Details, General Series C.7.3.1
Attribute Name
Tag
Description
Handling
Laterality
(0020,0060)
2C
Laterality of (paired) body part examined.
Required if the body part examined is a paired structure
and Image Laterality (0020,0062) or Frame Laterality
(0020,9072) are not sent.
Enumerated Values: R = right
L = left
Note:
Some IODs support Image Laterality (0020,0062) at the
Image level or Frame Laterality(0020,9072) at the Frame
level in the Frame Anatomy functional group macro,
which can provide a more comprehensive mechanism for
specifying the laterality of the body part(s) being
examined.
*** export not supported
Series Date
(0008,0021)
Date the Series started.
Export supported for:
Computed Tomography Image
RT Structure Set
Spatial Registration
Series Time
(0008,0031)
Time the Series started.
Export supported for:
Computed Tomography Image
RT Structure Set
Spatial Registration
Performing
Physicians Name
(0008,1050)
Name of the physician(s) administering the Series.
*** export not supported
Performing
Physician
Identification
Sequence
(0008,1052)
Identification of the physician(s) administering the Series.
One or more items shall be included in this sequence. If
more than one Item, the number and order shall
correspond to the value of Performing Physicians' Name
(0008,1050), if present.
*** export not supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
55 of 118
IOD Details, General Series C.7.3.1
56 of 118
Attribute Name
Tag
Description
Handling
> Person
Identification Code
Sequence
(0040,1101)
A coded entry which identifies a person. The Code
Meaning attribute, though it will be encoded with a VR of
LO, may be encoded according to the rules of the PN VR
(e.g. caret '^' delimiters shall separate name
components), except that a single component (i.e. the
whole name unseparated by caret delimiters) is not
permitted. Name component groups for use with multibyte character sets are permitted, as long as they fit
within the 64 characters (the length of the LO VR). One
or more Items may be permitted in this Sequence.
*** export not supported
>> Code Value
(0008,0100)
1C
See Section 8.1. Required if a sequence item is present.
*** export not supported
>> Coding Scheme
Designator
(0008,0102)
1C
See Section 8.2. Required if a sequence item is present.
*** export not supported
>> Coding Scheme
Version
(0008,0103)
1C
See Section 8.2. Required if a sequence item is present
and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.
*** export not supported
>> Code Meaning
(0008,0104)
1C
See Section 8.3. Required if a sequence item is present.
*** export not supported
> Persons Address
(0040,1102)
Person's mailing address
*** export not supported
> Persons
Telephone Numbers
(0040,1103)
Person's telephone number(s)
*** export not supported
> Institution Name
(0008,0080)
1C
Institution or organization to which the identified individual
is responsible or accountable. Shall not be present if
Institution Code Sequence (0008,0082) is present.
*** export not supported
> Institution Address
(0008,0081)
Mailing address of the institution or organization to which
the identified individual is responsible or accountable.
*** export not supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
IOD Details, General Series C.7.3.1
Attribute Name
Tag
Description
Handling
> Institution Code
Sequence
(0008,0082)
1C
Institution or organization to which the identified individual
is responsible or accountable. Shall not be present if
Institution Name (0008,0080) is present. Only a single
Item shall be permitted in this Sequence.
*** export not supported
>> Code Value
(0008,0100)
1C
See Section 8.1. Required if a sequence item is present.
*** export not supported
>> Coding Scheme
Designator
(0008,0102)
1C
See Section 8.2. Required if a sequence item is present.
*** export not supported
>> Coding Scheme
Version
(0008,0103)
1C
See Section 8.2. Required if a sequence item is present
and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.
*** export not supported
>> Code Meaning
(0008,0104)
1C
See Section 8.3. Required if a sequence item is present.
*** export not supported
Protocol Name
(0018,1030)
User-defined description of the conditions under which
the Series was performed.
Note:
This attribute conveys series-specific protocol
identification and may or may not be identical to the one
presented in the Performed Protocol Code Sequence
(0040,0260).
*** export not supported
Series Description
(0008,103E)
User provided description of the Series
*** export not supported
Operators Name
(0008,1070)
Name(s) of the operator(s) supporting the Series.
*** export not supported
Operator
Identification
Sequence
(0008,1072)
Identification of the operator(s) supporting the Series.
One or more items shall be included in this sequence. If
more than one Item, the number and order shall
correspond to the value of Operators' Name (0008,1070),
if present.
*** export not supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
57 of 118
IOD Details, General Series C.7.3.1
58 of 118
Attribute Name
Tag
Description
Handling
> Person
Identification Code
Sequence
(0040,1101)
A coded entry which identifies a person. The Code
Meaning attribute, though it will be encoded with a VR of
LO, may be encoded according to the rules of the PN VR
(e.g. caret '^' delimiters shall separate name
components), except that a single component (i.e. the
whole name unseparated by caret delimiters) is not
permitted. Name component groups for use with multibyte character sets are permitted, as long as they fit
within the 64 characters (the length of the LO VR). One
or more Items may be permitted in this Sequence.
*** export not supported
>> Code Value
(0008,0100)
1C
See Section 8.1. Required if a sequence item is present.
*** export not supported
>> Coding Scheme
Designator
(0008,0102)
1C
See Section 8.2. Required if a sequence item is present.
*** export not supported
>> Coding Scheme
Version
(0008,0103)
1C
See Section 8.2. Required if a sequence item is present
and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.
*** export not supported
>> Code Meaning
(0008,0104)
1C
See Section 8.3. Required if a sequence item is present.
*** export not supported
> Persons Address
(0040,1102)
Person's mailing address
*** export not supported
> Persons
Telephone Numbers
(0040,1103)
Person's telephone number(s)
*** export not supported
> Institution Name
(0008,0080)
1C
Institution or organization to which the identified individual
is responsible or accountable. Shall not be present if
Institution Code Sequence (0008,0082) is present.
*** export not supported
> Institution Address
(0008,0081)
Mailing address of the institution or organization to which
the identified individual is responsible or accountable.
*** export not supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
IOD Details, General Series C.7.3.1
Attribute Name
Tag
Description
Handling
> Institution Code
Sequence
(0008,0082)
1C
Institution or organization to which the identified individual
is responsible or accountable. Shall not be present if
Institution Name (0008,0080) is present. Only a single
Item shall be permitted in this Sequence.
*** export not supported
>> Code Value
(0008,0100)
1C
See Section 8.1. Required if a sequence item is present.
*** export not supported
>> Coding Scheme
Designator
(0008,0102)
1C
See Section 8.2. Required if a sequence item is present.
*** export not supported
>> Coding Scheme
Version
(0008,0103)
1C
See Section 8.2. Required if a sequence item is present
and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.
*** export not supported
>> Code Meaning
(0008,0104)
1C
See Section 8.3. Required if a sequence item is present.
*** export not supported
Referenced
Performed
Procedure Step
Sequence
(0008,1111)
Uniquely identifies the Performed Procedure Step SOP
Instance to which the Series is related (e.g. a Modality or
General-Purpose Performed Procedure Step SOP
Instance). The Sequence shall have zero or one Item.
*** export not supported
> Referenced SOP
Class UID
(0008,1150)
1C
Uniquely identifies the referenced SOP Class.
Required if Referenced Performed Procedure Step
Sequence (0008,1111) is sent.
*** export not supported
> Referenced SOP
Instance UID
(0008,1155)
1C
Uniquely identifies the referenced SOP Instance.
Required if Referenced Performed Procedure Step
Sequence (0008,1111) is sent.
*** export not supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
59 of 118
IOD Details, General Series C.7.3.1
60 of 118
Attribute Name
Tag
Description
Handling
Related Series
Sequence
(0008,1250)
Identification of Series significantly related to this Series.
Zero or more Items may be present.
Notes:
1. For example, for a combined CT and PET acquisition,
the CT images and PET images would be in separate
series that could cross- reference each other with
multiple purpose of reference codes meaning same
anatomy, simultaneously acquired and same indication.
2. The related series may have different Frames of
Reference and hence require some sort of registration
before spatial coordinates can be directly compared.
3. This attribute is not intended for conveying localizer
reference information, for which Referenced Image
Sequence (0008,1140) should be used.
*** export not supported
> Study Instance
UID
(0020,000D)
Instance UID of Study to which the related Series
belongs
*** export not supported
> Series Instance
UID
(0020,000E)
Instance UID of Related Series
*** export not supported
> Purpose of
Reference Code
Sequence
(0040,A170)
Describes the purpose for which the reference is made.
Zero or more Items may be present. When absent,
implies that the reason for the reference is unknown.
*** export not supported
>> Code Value
(0008,0100)
1C
See Section 8.1. Required if a sequence item is present.
*** export not supported
>> Coding Scheme
Designator
(0008,0102)
1C
See Section 8.2. Required if a sequence item is present.
*** export not supported
>> Coding Scheme
Version
(0008,0103)
1C
See Section 8.2. Required if a sequence item is present
and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.
*** export not supported
>> Code Meaning
(0008,0104)
1C
See Section 8.3. Required if a sequence item is present.
*** export not supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
IOD Details, General Series C.7.3.1
Attribute Name
Tag
Description
Handling
Body Part Examined
(0018,0015)
Text description of the part of the body examined.
Defined Terms:
SKULL, CSPINE, TSPINE, LSPINE, SSPINE, COCCYX,
CHEST, CLAVICLE, BREAST, ABDOMEN, PELVIS, HIP,
SHOULDER, ELBOW, KNEE, ANKLE, HAND, FOOT,
EXTREMITY, HEAD, HEART, NECK, LEG, ARM, JAW
Note:
Some IODs support the Anatomic Region Sequence
(0008,2218), which can provide a more comprehensive
mechanism for specifying the body part being examined.
*** export not supported
Patient Position
(0018,5100)
2C
Patient position descriptor relative to the equipment.
Required for CT and MR images; shall not be present if
Patient Orientation Code Sequence (0054,0410) is
present; may be present otherwise. See C.7.3.1.1.2 for
Defined Terms and further explanation.
Export supported for:
RT Image
Computed Tomography Image
Smallest Pixel Value
in Series
(0028,0108)
The minimum value of all images in this Series.
*** export not supported
Largest Pixel Value
in Series
(0028,0109)
The maximum value of all images in this Series.
*** export not supported
Request Attributes
Sequence
(0040,0275)
Sequence that contains attributes from the Imaging
Service Request. The sequence may have one or more
Items.
*** export not supported
> Requested
Procedure ID
(0040,1001)
Identifier that identifies the Requested Procedure in the
Imaging Service Request.
*** export not supported
> Reason for the
Requested
Procedure
(0040,1002)
Reason for requesting this procedure.
*** export not supported
> Reason for
Requested
Procedure Code
Sequence
(0040,100A)
Coded Reason for requesting this procedure.
*** export not supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
61 of 118
IOD Details, General Series C.7.3.1
62 of 118
Attribute Name
Tag
Description
Handling
>> Code Value
(0008,0100)
1C
See Section 8.1. Required if a sequence item is present.
*** export not supported
>> Coding Scheme
Designator
(0008,0102)
1C
See Section 8.2. Required if a sequence item is present.
*** export not supported
>> Coding Scheme
Version
(0008,0103)
1C
See Section 8.2. Required if a sequence item is present
and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.
*** export not supported
>> Code Meaning
(0008,0104)
1C
See Section 8.3. Required if a sequence item is present.
*** export not supported
> Scheduled
Procedure Step ID
(0040,0009)
Identifier that identifies the Scheduled Procedure Step.
*** export not supported
> Scheduled
Procedure Step
Description
(0040,0007)
Institution-generated description or classification of the
Scheduled Procedure Step to be performed.
*** export not supported
> Scheduled
Protocol Code
Sequence
(0040,0008)
Sequence describing the Scheduled Protocol following a
specific coding scheme. This sequence contains one or
more Items.
*** export not supported
>> Code Value
(0008,0100)
1C
See Section 8.1. Required if a sequence item is present.
*** export not supported
>> Coding Scheme
Designator
(0008,0102)
1C
See Section 8.2. Required if a sequence item is present.
*** export not supported
>> Coding Scheme
Version
(0008,0103)
1C
See Section 8.2. Required if a sequence item is present
and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.
*** export not supported
>> Code Meaning
(0008,0104)
1C
See Section 8.3. Required if a sequence item is present.
*** export not supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
IOD Details, General Series C.7.3.1
Attribute Name
Tag
Description
Handling
>> Protocol Context
Sequence
(0040,0440)
Sequence that specifies the context for the Scheduled
Protocol Code Sequence Item. One or more items may
be included in this sequence.
*** export not supported
>>> Code Value
(0008,0100)
1C
See Section 8.1. Required if a sequence item is present.
*** export not supported
>>> Coding
Scheme Designator
(0008,0102)
1C
See Section 8.2. Required if a sequence item is present.
*** export not supported
>>> Coding
Scheme Version
(0008,0103)
1C
See Section 8.2. Required if a sequence item is present
and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.
*** export not supported
>>> Code Meaning
(0008,0104)
1C
See Section 8.3. Required if a sequence item is present.
*** export not supported
>>> Content Item
Modifier Sequence
(0040,0441)
Sequence that specifies modifiers for a Protocol Context
Content Item. One or more items may be included in this
sequence. See Section C.4.10.1.
*** export not supported
>>>> Code Value
(0008,0100)
1C
See Section 8.1. Required if a sequence item is present.
*** export not supported
>>>> Coding
Scheme Designator
(0008,0102)
1C
See Section 8.2. Required if a sequence item is present.
*** export not supported
>>>> Coding
Scheme Version
(0008,0103)
1C
See Section 8.2. Required if a sequence item is present
and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.
*** export not supported
>>>> Code Meaning
(0008,0104)
1C
See Section 8.3. Required if a sequence item is present.
*** export not supported
Performed
Procedure Step ID
(0040,0253)
User or equipment generated identifier of that part of a
Procedure that has been carried out within this step.
*** export not supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
63 of 118
IOD Details, General Series C.7.3.1
64 of 118
Attribute Name
Tag
Description
Handling
Performed
Procedure Step
Start Date
(0040,0244)
Date on which the Performed Procedure Step started.
*** export not supported
Performed
Procedure Step
Start Time
(0040,0245)
Time on which the Performed Procedure Step started.
*** export not supported
Performed
Procedure Step
Description
(0040,0254)
Institution-generated description or classification of the
Procedure Step that was performed.
*** export not supported
Performed Protocol
Code Sequence
(0040,0260)
Sequence describing the Protocol performed for this
Procedure Step. One or more Items may be included in
this Sequence.
*** export not supported
> Code Value
(0008,0100)
1C
See Section 8.1. Required if a sequence item is present.
*** export not supported
> Coding Scheme
Designator
(0008,0102)
1C
See Section 8.2. Required if a sequence item is present.
*** export not supported
> Coding Scheme
Version
(0008,0103)
1C
See Section 8.2. Required if a sequence item is present
and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.
*** export not supported
> Code Meaning
(0008,0104)
1C
See Section 8.3. Required if a sequence item is present.
*** export not supported
> Protocol Context
Sequence
(0040,0440)
Sequence that specifies the context for the Performed
Protocol Code Sequence Item. One or more items may
be included in this sequence.
*** export not supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
IOD Details, General Series C.7.3.1
Attribute Name
Tag
Description
Handling
>> Value Type
(0040,A040)
The type of the value encoded in this name-value Item.
Defined Terms:
DATETIME
DATE
TIME
PNAME
UIDREF
TEXT
CODE
NUMERIC.
*** export not supported
>> Concept Name
Code Sequence
(0040,A043)
Coded concept name of this name-value Item.
*** export not supported
>>> Code Value
(0008,0100)
1C
See Section 8.1. Required if a sequence item is present.
*** export not supported
>>> Coding
Scheme Designator
(0008,0102)
1C
See Section 8.2. Required if a sequence item is present.
*** export not supported
>>> Coding
Scheme Version
(0008,0103)
1C
See Section 8.2. Required if a sequence item is present
and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.
*** export not supported
>>> Code Meaning
(0008,0104)
1C
See Section 8.3. Required if a sequence item is present.
*** export not supported
>> DateTime
(0040,A120)
1C
Datetime value for this name-value Item. Required if
Value Type (0040,A040) is DATETIME.
*** export not supported
>> Date
(0040,A121)
1C
Date value for this name-value Item. Required if Value
Type (0040,A040) is DATE.
*** export not supported
>> Time
(0040,A122)
1C
Time value for this name-value Item. Required if Value
Type (0040,A040) is TIME.
*** export not supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
65 of 118
IOD Details, General Series C.7.3.1
66 of 118
Attribute Name
Tag
Description
Handling
>> Person Name
(0040,A123)
1C
Person name value for this name-value Item. Required if
Value Type (0040,A040) is PNAME.
*** export not supported
>> UID
(0040,A124)
1C
UID value for this name-value Item. Required if Value
Type (0040,A040) is UIDREF.
*** export not supported
>> Text Value
(0040,A160)
1C
Text value for this name-value Item. Required if Value
Type (0040,A040) is TEXT.
*** export not supported
>> Concept Code
Sequence
(0040,A168)
1C
Coded concept value of this name-value Item. Required if
Value Type (0040,A040) is CODE.
*** export not supported
>>> Code Value
(0008,0100)
1C
See Section 8.1. Required if a sequence item is present.
*** export not supported
>>> Coding
Scheme Designator
(0008,0102)
1C
See Section 8.2. Required if a sequence item is present.
*** export not supported
>>> Coding
Scheme Version
(0008,0103)
1C
See Section 8.2. Required if a sequence item is present
and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.
*** export not supported
>>> Code Meaning
(0008,0104)
1C
See Section 8.3. Required if a sequence item is present.
*** export not supported
>> Numeric Value
(0040,A30A)
1C
Numeric value for this name-value Item. Required if
Value Type (0040,A040) is NUMERIC.
*** export not supported
>> Measurement
Units Code
Sequence
(0040,08EA)
1C
Units of measurement for a numeric value in this
namevalue Item. Required if Value Type (0040,A040) is
NUMERIC.
*** export not supported
>>> Code Value
(0008,0100)
1C
See Section 8.1. Required if a sequence item is present.
*** export not supported
>>> Coding
Scheme Designator
(0008,0102)
1C
See Section 8.2. Required if a sequence item is present.
*** export not supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
IOD Details, General Series C.7.3.1
Attribute Name
Tag
Description
Handling
>>> Coding
Scheme Version
(0008,0103)
1C
See Section 8.2. Required if a sequence item is present
and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.
*** export not supported
>>> Code Meaning
(0008,0104)
1C
See Section 8.3. Required if a sequence item is present.
*** export not supported
>> Content Item
Modifier Sequence
(0040,0441)
Sequence that specifies modifiers for a Protocol Context
Content Item. One or more items may be included in this
sequence. See Section C.4.10.1.
*** export not supported
>>> Value Type
(0040,A040)
The type of the value encoded in this name-value Item.
Defined Terms:
DATETIME
DATE
TIME
PNAME
UIDREF
TEXT
CODE
NUMERIC.
*** export not supported
>>> Concept Name
Code Sequence
(0040,A043)
Coded concept name of this name-value Item.
*** export not supported
>>>> Code Value
(0008,0100)
1C
See Section 8.1. Required if a sequence item is present.
*** export not supported
>>>> Coding
Scheme Designator
(0008,0102)
1C
See Section 8.2. Required if a sequence item is present.
*** export not supported
>>>> Coding
Scheme Version
(0008,0103)
1C
See Section 8.2. Required if a sequence item is present
and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.
*** export not supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
67 of 118
IOD Details, General Series C.7.3.1
68 of 118
Attribute Name
Tag
Description
Handling
>>>> Code Meaning
(0008,0104)
1C
See Section 8.3. Required if a sequence item is present.
*** export not supported
>>> DateTime
(0040,A120)
1C
Datetime value for this name-value Item. Required if
Value Type (0040,A040) is DATETIME.
*** export not supported
>>> Date
(0040,A121)
1C
Date value for this name-value Item. Required if Value
Type (0040,A040) is DATE.
*** export not supported
>>> Time
(0040,A122)
1C
Time value for this name-value Item. Required if Value
Type (0040,A040) is TIME.
*** export not supported
>>> Person Name
(0040,A123)
1C
Person name value for this name-value Item. Required if
Value Type (0040,A040) is PNAME.
*** export not supported
>>> UID
(0040,A124)
1C
UID value for this name-value Item. Required if Value
Type (0040,A040) is UIDREF.
*** export not supported
>>> Text Value
(0040,A160)
1C
Text value for this name-value Item. Required if Value
Type (0040,A040) is TEXT.
*** export not supported
>>> Concept Code
Sequence
(0040,A168)
1C
Coded concept value of this name-value Item. Required if
Value Type (0040,A040) is CODE.
*** export not supported
>>>> Code Value
(0008,0100)
1C
See Section 8.1. Required if a sequence item is present.
*** export not supported
>>>> Coding
Scheme Designator
(0008,0102)
1C
See Section 8.2. Required if a sequence item is present.
*** export not supported
>>>> Coding
Scheme Version
(0008,0103)
1C
See Section 8.2. Required if a sequence item is present
and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.
*** export not supported
>>>> Code Meaning
(0008,0104)
1C
See Section 8.3. Required if a sequence item is present.
*** export not supported
>>> Numeric Value
(0040,A30A)
1C
Numeric value for this name-value Item. Required if
Value Type (0040,A040) is NUMERIC.
*** export not supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
IOD Details, Frame of Reference C.7.4.1
B.2.4
Attribute Name
Tag
Description
Handling
>>> Measurement
Units Code
Sequence
(0040,08EA)
1C
Units of measurement for a numeric value in this
namevalue Item. Required if Value Type (0040,A040) is
NUMERIC.
*** export not supported
>>>> Code Value
(0008,0100)
1C
See Section 8.1. Required if a sequence item is present.
*** export not supported
>>>> Coding
Scheme Designator
(0008,0102)
1C
See Section 8.2. Required if a sequence item is present.
*** export not supported
>>>> Coding
Scheme Version
(0008,0103)
1C
See Section 8.2. Required if a sequence item is present
and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.
*** export not supported
>>>> Code Meaning
(0008,0104)
1C
See Section 8.3. Required if a sequence item is present.
*** export not supported
Comments on the
Performed
Procedure Step
(0040,0280)
User-defined comments on the Performed Procedure
Step.
*** export not supported
Frame of Reference C.7.4.1
Attribute Name
Tag
Description
Handling
Frame of Reference
UID
(0020,0052)
Uniquely identifies the frame of reference for a Series.
See C.7.4.1.1.1 for further explanation.
Export supported for:
All
Position Reference
Indicator
(0020,1040)
Part of the patient's anatomy used as a reference, such
as the iliac crest, orbital- medial, sternal notch, symphysis
pubis, xiphoid, lower coastal margin, external auditory
meatus. See C.7.4.1.1.2 for further explanation.
*** export not supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
69 of 118
IOD Details, General Equipment C.7.5.1
B.2.5
70 of 118
General Equipment C.7.5.1
Attribute Name
Tag
Description
Handling
Manufacturer
(0008,0070)
Manufacturer of the equipment that produced the
composite instances.
Export supported for:
RT Image
Computed Tomography Image
Institution Name
(0008,0080)
Institution where the equipment that produced the
composite instances is located.
*** export not supported
Institution Address
(0008,0081)
Mailing address of the institution where the equipment
that produced the composite instances is located.
*** export not supported
Station Name
(0008,1010)
User defined name identifying the machine that produced
the composite instances.
Export supported for:
RT Image
Computed Tomography Image
Institutional
Department Name
(0008,1040)
Department in the institution where the equipment that
produced the composite instances is located.
*** export not supported
Manufacturers
Model Name
(0008,1090)
Manufacturer's model name of the equipment that
produced the composite instances.
Export supported for:
RT Image:
OBI: On-Board Imager
PV AI: PortalVision AI
Computed Tomography Image:
OBI Cone-beam CT
Device Serial
Number
(0018,1000)
Manufacturer's serial number of the equipment that
produced the composite instances.
*** export not supported
Software Version(s)
(0018,1020)
Manufacturer's designation of software version of the
equipment that produced the composite instances.
Export supported for:
RT Image
Computed Tomography Image
Spatial Resolution
(0018,1050)
The inherent limiting resolution in mm of the acquisition
equipment for high contrast objects for the data gathering
and reconstruction technique chosen. If variable across
the images of the series, the value at the image center.
*** export not supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
IOD Details, General Image C.7.6.1
B.2.6
Attribute Name
Tag
Description
Handling
Date of Last
Calibration
(0018,1200)
Date when the image acquisition device calibration was
last changed in any way. Multiple entries may be used for
additional calibrations at other times. See C.7.5.1.1.1 for
further explanation.
*** export not supported
Time of Last
Calibration
(0018,1201)
Time when the image acquisition device calibration was
last changed in any way. Multiple entries may be used.
See C.7.5.1.1.1 for further explanation.
*** export not supported
Pixel Padding Value
(0028,0120)
1C
Single pixel value or one limit (inclusive) of a range of
pixel values used in animage to pad to rectangular format
or to signal background that may be suppressed. See
C.7.5.1.1.2 for further explanation.
Required if Pixel Padding Range Limit (0028,0121) is
present. May be present otherwise.
Note: The Value Representation of this Attribute is
determined by the value of Pixel Representation
(0028,0103).
*** export not supported
General Image C.7.6.1
Attribute Name
Tag
Description
Handling
Instance Number
(0020,0013)
A number that identifies this image.
Note:
This Attribute was named Image Number in earlier
versions of this Standard.
Export supported for:
RT Image
Computed Tomography Image
Value is set to 0.
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
71 of 118
IOD Details, General Image C.7.6.1
72 of 118
Attribute Name
Tag
Description
Handling
Patient Orientation
(0020,0020)
2C
Patient direction of the rows and columns of the image.
Required if image does not require Image Orientation
(Patient) (0020,0037) and Image Position (Patient)
(0020,0032). See C.7.6.1.1.1 for further explanation.
Note:
IOD's may have attributes other than Patient Orientation,
Image Orientation, or Image Position (Patient) to describe
orientation in which case this attribute will be zero length.
*** export not supported
Content Date
(0008,0023)
2C
The date the image pixel data creation started. Required
if image is part of a series in which the images are
temporally related.
Note:
This Attribute was formerly known as Image Date.
Export supported for:
RT Image
Computed Tomography Image
Content Time
(0008,0033)
2C
The time the image pixel data creation started. Required
if image is part of a series in which the images are
temporally related.
Export supported for:
RT Image
Computed Tomography Image
Image Type
(0008,0008)
Image identification characteristics. See C.7.6.1.1.2 for
Defined Terms and further explanation.
Export supported for:
RT Image
kV:
ORIGINAL\PRIMARY\PORTAL\OBI
MV: ORIGINAL\PRIMARY\PORTAL
Computed Tomography Image:
ORIGINAL\PRIMARY\AXIAL
ORIGINAL\PRIMARY\LOCALIZER
Acquisition Number
(0020,0012)
A number identifying the single continuous gathering of
data over a period of time that resulted in this image.
Export supported for:
Computed Tomography Image
Acquisition Date
(0008,0022)
The date the acquisition of data that resulted in this
image started
*** export not supported
Acquisition Time
(0008,0032)
The time the acquisition of data that resulted in this
image started
*** export not supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
IOD Details, General Image C.7.6.1
Attribute Name
Tag
Description
Handling
Acquisition Datetime
(0008,002A)
The date and time that the acquisition of data that
resulted in this image started.
Note:
The synchronization of this time with an external clock is
specified in the Synchronization Module in Acquisition
Time Synchronized (0018,1800).
*** export not supported
Referenced Image
Sequence
(0008,1140)
A sequence that references other images significantly
related to this image (e.g. post-localizer CT image or
Mammographic biopsy or partial view images). One or
more Items may be included in this sequence.
*** export not supported
> Referenced SOP
Class UID
(0008,1150)
Uniquely identifies the referenced SOP Class.
*** export not supported
> Referenced SOP
Instance UID
(0008,1155)
Uniquely identifies the referenced SOP Instance.
*** export not supported
> Referenced
Frame Number
(0008,1160)
Identifies the frame numbers within the Referenced SOP
Instance to which the reference applies. The first frame
shall be denoted as frame number 1.
Note: This Attribute may be multi-valued.
Required if the Referenced SOP Instance is a multi-frame
image and the reference does not apply to all frames.
*** export not supported
> Purpose of
Reference Code
Sequence
(0040,A170)
Describes the purpose for which the reference is made.
Only a single Item shall be permitted in this sequence.
*** export not supported
>> Code Value
(0008,0100)
1C
See Section 8.1. Required if a sequence item is present.
*** export not supported
>> Coding Scheme
Designator
(0008,0102)
1C
See Section 8.2. Required if a sequence item is present.
*** export not supported
>> Coding Scheme
Version
(0008,0103)
1C
See Section 8.2. Required if a sequence item is present
and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.
*** export not supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
73 of 118
IOD Details, General Image C.7.6.1
74 of 118
Attribute Name
Tag
Description
Handling
>> Code Meaning
(0008,0104)
1C
See Section 8.3. Required if a sequence item is present.
*** export not supported
Derivation
Description
(0008,2111)
A text description of how this image was derived. See
C.7.6.1.1.3 for further explanation.
*** export not supported
Derivation Code
Sequence
(0008,9215)
A coded description of how this image was derived. See
C.7.6.1.1.3 for further explanation. One or more Items
may be included in this Sequence. More than one Item
indicates that successive derivation steps have been
applied.
*** export not supported
> Code Value
(0008,0100)
1C
See Section 8.1. Required if a sequence item is present.
*** export not supported
> Coding Scheme
Designator
(0008,0102)
1C
See Section 8.2. Required if a sequence item is present.
*** export not supported
> Coding Scheme
Version
(0008,0103)
1C
See Section 8.2. Required if a sequence item is present
and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.
*** export not supported
> Code Meaning
(0008,0104)
1C
See Section 8.3. Required if a sequence item is present.
*** export not supported
Source Image
Sequence
(0008,2112)
A Sequence that identifies the set of Image SOP
Class/Instance pairs of the Images that were used to
derive this Image. Zero or more Items may be included in
this Sequence. See C.7.6.1.1.4 for further explanation.
*** export not supported
> Referenced SOP
Class UID
(0008,1150)
Uniquely identifies the referenced SOP Class.
*** export not supported
> Referenced SOP
Instance UID
(0008,1155)
Uniquely identifies the referenced SOP Instance.
*** export not supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
IOD Details, General Image C.7.6.1
Attribute Name
Tag
Description
Handling
> Referenced
Frame Number
(0008,1160)
Identifies the frame numbers within the Referenced SOP
Instance to which the reference applies. The first frame
shall be denoted as frame number 1.
Note: This Attribute may be multi-valued.
Required if the Referenced SOP Instance is a multi-frame
image and the reference does not apply to all frames.
*** export not supported
> Purpose of
Reference Code
Sequence
(0040,A170)
Describes the purpose for which the reference is made,
that is what role the source image or frame(s) played in
the derivation of this image. Only a single Item shall be
permitted in this sequence.
*** export not supported
>> Code Value
(0008,0100)
1C
See Section 8.1. Required if a sequence item is present.
*** export not supported
>> Coding Scheme
Designator
(0008,0102)
1C
See Section 8.2. Required if a sequence item is present.
*** export not supported
>> Coding Scheme
Version
(0008,0103)
1C
See Section 8.2. Required if a sequence item is present
and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.
*** export not supported
>> Code Meaning
(0008,0104)
1C
See Section 8.3. Required if a sequence item is present.
*** export not supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
75 of 118
IOD Details, General Image C.7.6.1
76 of 118
Attribute Name
Tag
Description
Handling
> Spatial Locations
Preserved
(0028,135A)
Whether or not the spatial locations of all pixels are
preserved during the processing of the source image that
resulted in the current image
Enumerated Values:
YES
NO
Note:
This applies not only to images with a known relationship
to a 3D space, but also to projection images. For
example, a projection radiograph such as a mammogram
that is processed by a point image processing operation
such as contrast enhancement, or a smoothing or edge
enhancing convolution, would have a value of YES for
this attribute. A projection radiograph that had been
magnified or warped geometrically would have a value of
NO for this attribute. This attribute is typically of
importance in relating images with Presentation Intent
Type (0008,0068) values of FOR PROCESSING and
FOR PRESENTATION.
*** export not supported
Referenced
Instance Sequence
(0008,114A)
A sequence which provides reference to a set of nonimage SOP Class/Instance pairs significantly related to
this Image, including waveforms that may or may not be
temporally synchronized with this image . One or more
Items may be included in this sequence.
*** export not supported
> Referenced SOP
Class UID
(0008,1150)
Uniquely identifies the referenced SOP Class.
*** export not supported
> Referenced SOP
Instance UID
(0008,1155)
Uniquely identifies the referenced SOP Instance.
*** export not supported
> Purpose of
Reference Code
Sequence
(0040,A170)
Code describing the purpose of the reference to the
Instance(s). Only a single Item shall be permitted in this
sequence.
*** export not supported
>> Code Value
(0008,0100)
1C
See Section 8.1. Required if a sequence item is present.
*** export not supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
IOD Details, General Image C.7.6.1
Attribute Name
Tag
Description
Handling
>> Coding Scheme
Designator
(0008,0102)
1C
See Section 8.2. Required if a sequence item is present.
*** export not supported
>> Coding Scheme
Version
(0008,0103)
1C
See Section 8.2. Required if a sequence item is present
and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.
*** export not supported
>> Code Meaning
(0008,0104)
1C
See Section 8.3. Required if a sequence item is present.
*** export not supported
Images in
Acquisition
(0020,1002)
Number of images that resulted from this acquisition of
data
*** export not supported
Image Comments
(0020,4000)
User-defined comments about the image
*** export not supported
Quality Control
Image
(0028,0300)
Indicates whether or not this image is a quality control or
phantom image.
Enumerated Values:
YES
NO If this Attribute is absent, then the image may or may
not be a quality control or phantom image.
*** export not supported
Burned In
Annotation
(0028,0301)
Indicates whether or not image contains sufficient burned
in annotation to identify the patient and date the image
was acquired.
Enumerated Values:
YES
NO If this Attribute is absent, then the image may or may
not contain burned in annotation.
*** export not supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
77 of 118
IOD Details, General Image C.7.6.1
78 of 118
Attribute Name
Tag
Description
Handling
Lossy Image
Compression
(0028,2110)
Specifies whether an Image has undergone lossy
compression.
Enumerated Values:
00 = Image has NOT been subjected to lossy
compression.
01 = Image has been subjected to lossy compression.
See C.7.6.1.1.5
*** export not supported
Lossy Image
Compression Ratio
(0028,2112)
Describes the approximate lossy compression ratio(s)
that have been applied to this image. See C.7.6.1.1.5 for
further explanation. May be multivalued if successive
lossy compression steps have been applied.
Notes:
1. For example, a compression ratio of 30:1 would be
described in this Attribute with a single value of 30.
2. For historical reasons, the lossy compression ratio may
also be described in Derivation Description (0008,2111).
*** export not supported
Lossy Image
Compression
Method
(0028,2114)
A label for the lossy compression method(s) that have
been applied to this image. See C.7.6.1.1.5 for further
explanation. May be multivalued if successive lossy
compression steps have been applied; the value order
shall correspond to the values of Lossy Image
Compression Ratio (0028,2112).
Note:
For historical reasons, the lossy compression method
may also be described in Derivation Description
(0008,2111).
*** export not supported
Icon Image
Sequence
(0088,0200)
This icon image is representative of the Image.
*** export not supported
> Samples per Pixel
(0028,0002)
Number of samples (planes) in this image. See
C.7.6.3.1.1 for further explanation.
*** export not supported
> Photometric
Interpretation
(0028,0004)
Specifies the intended interpretation of the pixel data.
See C.7.6.3.1.2 for further explanation.
*** export not supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
IOD Details, General Image C.7.6.1
Attribute Name
Tag
Description
Handling
> Rows
(0028,0010)
Number of rows in the image.
*** export not supported
> Columns
(0028,0011)
Number of columns in the image.
*** export not supported
> Bits Allocated
(0028,0100)
Number of bits allocated for each pixel sample. Each
sample shall have the same number of bits allocated.
See PS 3.5 for further explanation.
*** export not supported
> Bits Stored
(0028,0101)
Number of bits stored for each pixel sample. Each
sample shall have the same number of bits stored. See
PS 3.5 for further explanation.
*** export not supported
> High Bit
(0028,0102)
Most significant bit for pixel sample data. Each sample
shall have the same high bit. See PS 3.5 for further
explanation.
*** export not supported
> Pixel
Representation
(0028,0103)
Data representation of the pixel samples. Each sample
shall have the same pixel representation. Enumerated
Values:
0000H = unsigned integer.
0001H = 2's complement
*** export not supported
> Pixel Data
(7FE0,0010)
1C
A data stream of the pixel samples that comprise the
Image. See C.7.6.3.1.4 for further explanation. Required
if Pixel Data Provider URL (0028,7FE0) is not present.
*** export not supported
> Planar
Configuration
(0028,0006)
1C
Indicates whether the pixel data are sent color-by-plane
or color-by-pixel. Required if Samples per Pixel
(0028,0002) has a value greater than 1. See C.7.6.3.1.3
for further explanation.
*** export not supported
> Pixel Aspect Ratio
(0028,0034)
1C
Ratio of the vertical size and horizontal size of the pixels
in the image specified by a pair of integer values where
the first value is the vertical pixel size, and the second
value is the horizontal pixel size. Required if the aspect
ratio is not 1 and the Image Plane Module or the Pixel
Measures Macro is not applicable to this Image. See
C.7.6.3.1.7.
*** export not supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
79 of 118
IOD Details, General Image C.7.6.1
80 of 118
Attribute Name
Tag
Description
Handling
> Smallest Image
Pixel Value
(0028,0106)
The minimum actual pixel value encountered in this
image.
*** export not supported
> Largest Image
Pixel Value
(0028,0107)
The maximum actual pixel value encountered in this
image.
*** export not supported
> Red Palette Color
Lookup Table
Descriptor US or
(0028,1101)
1C
Specifies the format of the Red Palette Color Lookup
Table Data (0028,1201) Required if Photometric
Interpretation (0028,0004) has a value of PALETTE
COLOR or Pixel Presentation (0008,9205) at the image
level equals COLOR or MIXED. See C.7.6.3.1.5 for
further explanation.
*** export not supported
> Green Palette
Color Lookup Table
Descriptor US or
(0028,1102)
1C
Specifies the format of the Green Palette Color Lookup
Table Data (0028,1202) Required if Photometric
Interpretation (0028,0004) has a value of PALETTE
COLOR or Pixel Presentation (0008,9205) at the image
level equals COLOR or MIXED. See C.7.6.3.1.5 for
further explanation.
*** export not supported
> Blue Palette Color
Lookup Table
Descriptor US or
(0028,1103)
1C
Specifies the format of the Blue Palette Color Lookup
Table Data (0028,1203) Required if Photometric
Interpretation (0028,0004) has a value of PALETTE
COLOR or Pixel Presentation (0008,9205) at the image
level equals COLOR or MIXED. See C.7.6.3.1.5 for
further explanation.
*** export not supported
> Red Palette Color
Lookup Table Data
(0028,1201)
1C
Red Palette Color Lookup Table Data. Required if
Photometric Interpretation (0028,0004) has a value of
PALETTE COLOR or Pixel Presentation (0008,9205) at
the image level equals COLOR or MIXED. See
C.7.6.3.1.6 for further explanation.
*** export not supported
> Green Palette
Color Lookup Table
Data
(0028,1202)
1C
Green Palette Color Lookup Table Data. Required if
Photometric Interpretation (0028,0004) has a value of
PALETTE COLOR or Pixel Presentation (0008,9205) at
the image level equals COLOR or MIXED. See
C.7.6.3.1.6 for further explanation.
*** export not supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
IOD Details, General Image C.7.6.1
Attribute Name
Tag
Description
Handling
> Blue Palette Color
Lookup Table Data
(0028,1203)
1C
Blue Palette Color Lookup Table Data. Required if
Photometric Interpretation (0028,0004) has a value of
PALETTE COLOR or Pixel Presentation (0008,9205) at
the image level equals COLOR or MIXED. See
C.7.6.3.1.6 for further explanation.
*** export not supported
> ICC Profile
(0028,2000)
An ICC Profile encoding the transformation of devicedependent color stored pixel values into PCS-Values.
See Section C.11.15.1.1.1. When present, defines the
color space of color Pixel Data (7FE0,0010) values, and
the output of Palette Color Lookup Table. Data
(0028,1201-1203). Note: The profile applies only to the
Pixel Data (7FE0,0010) attribute at the same level of the
dataset and not to any icons nested within sequences,
which may or may not have their own ICC profile
specified.
*** export not supported
Presentation LUT
Shape
(2050,0020)
When present, specifies an identity transformation for the
Presentation LUT such that the output of all grayscale
transformations, if any, are defined to be in P-Values.
Enumerated Values are:
IDENTITY = output is in P-Values - shall be used if
Photometric Interpretation (0028,0004) is
MONOCHROME2 or any color photometric interpretation.
INVERSE = output after inversion is in P- Values - shall
be used if Photometric Interpretation (0028,0004) is
MONOCHROME1. When this attribute is used with a
color photometric interpretation then the luminance
component is in P-Values.
*** export not supported
Irradiation Event
UID
(0008,3010)
Unique identification of the irradiation event(s) associated
with the acquisition of this image. See C.7.6.1.1.7.
*** export not supported
Image Remarks
(3281,XX00)
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
81 of 118
IOD Details, Image Plane C.7.6.2
B.2.7
B.2.8
82 of 118
Image Plane C.7.6.2
Attribute Name
Tag
Description
Handling
Pixel Spacing
(0028,0030)
Physical distance in the patient between the center of
each pixel, specified by a numeric pair - adjacent row
spacing (delimiter) adjacent column spacing in mm.
Export supported for:
Computed Tomography Image
Image Orientation
(Patient)
(0020,0037)
The direction cosines of the first row and the first column
with respect to the patient. See C.7.6.2.1.1 for further
explanation.
Export supported for:
Computed Tomography Image
Image Position
(Patient)
(0020,0032)
The x, y, and z coordinates of the upper left hand corner
(center of the first voxel transmitted) of the image, in mm.
See C.7.6.2.1.1 for further explanation.
Export supported for:
Computed Tomography Image
Slice Thickness
(0018,0050)
Nominal slice thickness, in mm.
Export supported for:
Computed Tomography Image
Slice Location
(0020,1041)
Relative position of exposure expressed in mm.
C.7.6.2.1.2 for further explanation.
*** export not supported
Image Pixel C.7.6.3
Attribute Name
Tag
Description
Handling
Samples per Pixel
(0028,0002)
Number of samples (planes) in this image. See
C.7.6.3.1.1 for further explanation.
Export supported for:
RT Image
Computed Tomography Image
Value set to 1.
Photometric
Interpretation
(0028,0004)
Specifies the intended interpretation of the pixel data.
See C.7.6.3.1.2 for further explanation.
Export supported for:
RT Image
Computed Tomography Image
Value set to MONOCHROME2
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
IOD Details, Image Pixel C.7.6.3
Attribute Name
Tag
Description
Handling
Rows
(0028,0010)
Number of rows in the image.
Export supported for:
RT Image
Computed Tomography Image
Columns
(0028,0011)
Number of columns in the image.
Export supported for:
RT Image
Computed Tomography Image
Bits Allocated
(0028,0100)
Number of bits allocated for each pixel sample. Each
sample shall have the same number of bits allocated.
See PS 3.5 for further explanation.
Export supported for:
RT Image
Computed Tomography Image
Bits Stored
(0028,0101)
Number of bits stored for each pixel sample. Each
sample shall have the same number of bits stored. See
PS 3.5 for further explanation.
Export supported for:
RT Image
Computed Tomography Image
High Bit
(0028,0102)
Most significant bit for pixel sample data. Each sample
shall have the same high bit. See PS 3.5 for further
explanation.
Export supported for:
RT Image
Computed Tomography Image
Pixel
Representation
(0028,0103)
Data representation of the pixel samples. Each sample
shall have the same pixel representation. Enumerated
Values:
0000H = unsigned integer.
0001H = 2's complement
Export supported for:
RT Image
Computed Tomography Image
Pixel Data
(7FE0,0010)
1C
A data stream of the pixel samples that comprise the
Image. See C.7.6.3.1.4 for further explanation. Required
if Pixel Data Provider URL (0028,7FE0) is not present.
Export supported for:
RT Image
Computed Tomography Image
Planar Configuration
(0028,0006)
1C
Indicates whether the pixel data are sent color-by-plane
or color-by-pixel. Required if Samples per Pixel
(0028,0002) has a value greater than 1. See C.7.6.3.1.3
for further explanation.
*** export not supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
83 of 118
IOD Details, Image Pixel C.7.6.3
84 of 118
Attribute Name
Tag
Description
Handling
Pixel Aspect Ratio
(0028,0034)
1C
Ratio of the vertical size and horizontal size of the pixels
in the image specified by a pair of integer values where
the first value is the vertical pixel size, and the second
value is the horizontal pixel size. Required if the aspect
ratio is not 1 and the Image Plane Module or the Pixel
Measures Macro is not applicable to this Image. See
C.7.6.3.1.7.
*** export not supported
Smallest Image
Pixel Value
(0028,0106)
The minimum actual pixel value encountered in this
image.
*** export not supported
Largest Image Pixel
Value
(0028,0107)
The maximum actual pixel value encountered in this
image.
*** export not supported
Red Palette Color
Lookup Table
Descriptor US or
(0028,1101)
1C
Specifies the format of the Red Palette Color Lookup
Table Data (0028,1201) Required if Photometric
Interpretation (0028,0004) has a value of PALETTE
COLOR or Pixel Presentation (0008,9205) at the image
level equals COLOR or MIXED. See C.7.6.3.1.5 for
further explanation.
*** export not supported
Green Palette Color
Lookup Table
Descriptor US or
(0028,1102)
1C
Specifies the format of the Green Palette Color Lookup
Table Data (0028,1202) Required if Photometric
Interpretation (0028,0004) has a value of PALETTE
COLOR or Pixel Presentation (0008,9205) at the image
level equals COLOR or MIXED. See C.7.6.3.1.5 for
further explanation.
*** export not supported
Blue Palette Color
Lookup Table
Descriptor US or
(0028,1103)
1C
Specifies the format of the Blue Palette Color Lookup
Table Data (0028,1203) Required if Photometric
Interpretation (0028,0004) has a value of PALETTE
COLOR or Pixel Presentation (0008,9205) at the image
level equals COLOR or MIXED. See C.7.6.3.1.5 for
further explanation.
*** export not supported
Red Palette Color
Lookup Table Data
(0028,1201)
1C
Red Palette Color Lookup Table Data. Required if
Photometric Interpretation (0028,0004) has a value of
PALETTE COLOR or Pixel Presentation (0008,9205) at
the image level equals COLOR or MIXED. See
C.7.6.3.1.6 for further explanation.
*** export not supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
IOD Details, Image Pixel C.7.6.3
Attribute Name
Tag
Description
Handling
Green Palette Color
Lookup Table Data
(0028,1202)
1C
Green Palette Color Lookup Table Data. Required if
Photometric Interpretation (0028,0004) has a value of
PALETTE COLOR or Pixel Presentation (0008,9205) at
the image level equals COLOR or MIXED. See
C.7.6.3.1.6 for further explanation.
*** export not supported
Blue Palette Color
Lookup Table Data
(0028,1203)
1C
Blue Palette Color Lookup Table Data. Required if
Photometric Interpretation (0028,0004) has a value of
PALETTE COLOR or Pixel Presentation (0008,9205) at
the image level equals COLOR or MIXED. See
C.7.6.3.1.6 for further explanation.
*** export not supported
ICC Profile
(0028,2000)
An ICC Profile encoding the transformation of devicedependent color stored pixel values into PCS-Values.
See Section C.11.15.1.1.1. When present, defines the
color space of color Pixel Data (7FE0,0010) values, and
the output of Palette Color Lookup Table. Data
(0028,1201-1203). Note: The profile applies only to the
Pixel Data (7FE0,0010) attribute at the same level of the
dataset and not to any icons nested within sequences,
which may or may not have their own ICC profile
specified.
*** export not supported
Pixel Data Provider
URL
(0028,7FE0)
1C
A URL of a provider service that supplies the pixel data of
the Image.
Required if the image is to be transferred in one of the
following presentation contexts identified by Transfer
Syntax UID: 1.2.840.10008.1.2.4.94 (DICOM JPIP
Referenced Transfer Syntax) 1.2.840.10008.1.2.4.95
(DICOM JPIP Referenced Deflate Transfer Syntax)
*** export not supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
85 of 118
IOD Details, CT Image C.8.2.1
B.2.9
86 of 118
CT Image C.8.2.1
Attribute Name
Tag
Description
Handling
Image Type
(0008,0008)
Image identification characteristics. See C.8.2.1.1.1 for
specialization.
Export supported
ORIGINAL\PRIMARY\
AXIAL
ORIGINAL\PRIMARY\
LOCALIZER
Samples per Pixel
(0028,0002)
Number of samples (planes) in this image. See
C.8.2.1.1.2 for specialization.
Export supported
Value set to 1
Photometric
Interpretation
(0028,0004)
Specifies the intended interpretation of the pixel data.
See C.8.2.1.1.3 for specialization.
Export supported
Value set to MONOCHROME2
Bits Allocated
(0028,0100)
Number of bits allocated for each pixel sample. Each
sample shall have the same number of bits allocated.
See C.8.2.1.1.4 for specialization.
Export supported
Bits Stored
(0028,0101)
Number of bits stored for each pixel sample. Each
sample shall have the same number of bits stored. See
C.8.2.1.1.5 for specialization.
Export supported
High Bit
(0028,0102)
Most significant bit for pixel sample data. Each sample
shall have the same high bit. See C.8.2.1.1.6 for
specialization.
Export supported
Rescale Intercept
(0028,1052)
The value b in relationship between stored values (SV)
and Hounsfield (HU). HU = m*SV+b
Export supported
Rescale Slope
(0028,1053)
m in the equation specified in Rescale Intercept
(0028,1052).
Export supported
KVP
(0018,0060)
Peak kilo voltage output of the x-ray generator used
*** export not supported
Acquisition Number
(0020,0012)
A number identifying the single continuous gathering of
data over a period of time which resulted in this image
Export supported
Scan Options
(0018,0022)
Parameters of scanning sequence.
Export supported
Value set to TOMO
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
IOD Details, CT Image C.8.2.1
Attribute Name
Tag
Description
Handling
Data Collection
Diameter
(0018,0090)
The diameter in mm of the region over which data were
collected
Export supported
Reconstruction
Diameter
(0018,1100)
Diameter in mm of the region from within which data were
used in creating the reconstruction of the image. Data
may exist outside this region and portions of the patient
may exist outside this region.
Export supported
Distance Source to
Detector
(0018,1110)
Distance in mm from source to detector center.
Note:
This value is traditionally referred to as Source Image
Receptor Distance (SID).
Export supported
Distance Source to
Patient
(0018,1111)
Distance in mm from source to isocenter (center of field
of view).
Note:
This value is traditionally referred to as Source Object
Distance (SOD).
Export supported
Value set to 0.
Gantry/Detector Tilt
(0018,1120)
Nominal angle of tilt in degrees of the scanning gantry.
Not intended for mathematical computations.
Export supported
Value set to 0.
Table Height
(0018,1130)
The distance in mm of the top of the patient table to the
center of rotation; below the center is positive.
Export supported
Table Top
Longitudinal
Position
(300A,0129)
Export supported
Table Top Lateral
Position
(300A,012A)
Export supported
Patient Support
Angle
(300A,0122)
Export supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
87 of 118
IOD Details, CT Image C.8.2.1
88 of 118
Attribute Name
Tag
Description
Handling
Rotation Direction
(0018,1140)
Direction of rotation of the source when relevant, about
nearest principal axis of equipment.
Enumerated Values:
CW = clockwise
CC = counter clockwise
*** export not supported
Exposure Time
(0018,1150)
Time of x-ray exposure in msec
Export supported
X-ray Tube Current
(0018,1151)
X-ray Tube Current in mA.
Export supported
Exposure
(0018,1152)
The exposure expressed in mAs, for example calculated
from Exposure Time and X-ray Tube Current.
Export supported
Exposure in uAs
(0018,1153)
The exposure expressed in As, for example calculated
from Exposure Time and X-ray Tube Current.
*** export not supported
Filter Type
(0018,1160)
Label for the type of filter inserted into the x-ray beam.
*** export not supported
Generator Power
(0018,1170)
Power in kW to the x-ray generator.
Export supported
Focal Spot(s)
(0018,1190)
Size of the focal spot in mm. For devices with variable
focal spot or multiple focal spots, small dimension
followed by large dimension.
Export supported
Convolution Kernel
(0018,1210)
A label describing the convolution kernel or algorithm
used to reconstruct the data
Export supported
Revolution Time
(0018,9305)
The time in seconds of a complete revolution of the
source around the gantry orbit.
*** export not supported
Single Collimation
Width
(0018,9306)
The width of a single row of acquired data (in mm).
Note:
Adjacent physical detector rows may have been
combined to form a single effective acquisition row.
*** export not supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
IOD Details, CT Image C.8.2.1
Attribute Name
Tag
Description
Handling
Total Collimation
Width
(0018,9307)
The width of the total collimation (in mm) over the area of
active x-ray detection.
Note:
This will be equal the number of effective detector rows
multiplied by single collimation width.
*** export not supported
Table Speed
(0018,9309)
The distance in mm that the table moves in one second
during the gathering of data that resulted in this image.
*** export not supported
Table Feed per
Rotation
(0018,9310)
Motion of the table (in mm) during a complete revolution
of the source around the gantry orbit.
*** export not supported
Spiral Pitch Factor
(0018,9311)
Ratio of the Table Feed per Rotation (0018,9310) to the
Total Collimation Width (0018,9307).
*** export not supported
Exposure
Modulation Type
(0018,9323)
A label describing the type of exposure modulation used
for the purpose of limiting the dose.
Defined Terms:
NONE
*** export not supported
Estimated Dose
Saving
(0018,9324)
A percent value of dose saving due to the use of
Exposure Modulation Type (0018,9323). A negative
percent value of dose savings reflects an increase of
exposure.
*** export not supported
CTDIvol
(0018,9345)
Computed Tomography Dose Index (CTDIvol), im mGy
according to IEC 60601-2-44, Ed.2.1 (Clause
29.1.103.4), The Volume CTDIvol. It describes the
average dose for this image for the selected CT
conditions of operation.
*** export not supported
Anatomic Region
Sequence
(0008,2218)
Sequence that identifies the anatomic region of interest in
this Instance (i.e. external anatomy, surface anatomy, or
general region of the body).
Only a single Item shall be permitted in this sequence.
*** export not supported
> Code Value
(0008,0100)
1C
See Section 8.1. Required if a sequence item is present.
*** export not supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
89 of 118
IOD Details, CT Image C.8.2.1
90 of 118
Attribute Name
Tag
Description
Handling
> Coding Scheme
Designator
(0008,0102)
1C
See Section 8.2. Required if a sequence item is present.
*** export not supported
> Coding Scheme
Version
(0008,0103)
1C
See Section 8.2. Required if a sequence item is present
and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.
*** export not supported
> Code Meaning
(0008,0104)
1C
See Section 8.3. Required if a sequence item is present.
*** export not supported
> Anatomic Region
Modifier Sequence
(0008,2220)
Sequence of Items that modifies the anatomic region of
interest of this Instance One or more Items may be
included in this Sequence.
*** export not supported
>> Code Value
(0008,0100)
1C
See Section 8.1. Required if a sequence item is present.
*** export not supported
>> Coding Scheme
Designator
(0008,0102)
1C
See Section 8.2. Required if a sequence item is present.
*** export not supported
>> Coding Scheme
Version
(0008,0103)
1C
See Section 8.2. Required if a sequence item is present
and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.
*** export not supported
>> Code Meaning
(0008,0104)
1C
See Section 8.3. Required if a sequence item is present.
*** export not supported
Primary Anatomic
Structure Sequence
(0008,2228)
Sequence of Items that identifies the primary anatomic
structure(s) of interest in this Instance.
One or more Items may be included in this Sequence.
*** export not supported
> Code Value
(0008,0100)
1C
See Section 8.1. Required if a sequence item is present.
*** export not supported
> Coding Scheme
Designator
(0008,0102)
1C
See Section 8.2. Required if a sequence item is present.
*** export not supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
IOD Details, RT Image C.8.8.2
B.2.10
Attribute Name
Tag
Description
Handling
> Coding Scheme
Version
(0008,0103)
1C
See Section 8.2. Required if a sequence item is present
and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.
*** export not supported
> Code Meaning
(0008,0104)
1C
See Section 8.3. Required if a sequence item is present.
*** export not supported
> Primary Anatomic
Structure Modifier
Sequence
(0008,2230)
Sequence of Items that modifies the primary anatomic
structure of interest in this Instance.
One or more Items may be included in this Sequence.
*** export not supported
>> Code Value
(0008,0100)
1C
See Section 8.1. Required if a sequence item is present.
*** export not supported
>> Coding Scheme
Designator
(0008,0102)
1C
See Section 8.2. Required if a sequence item is present.
*** export not supported
>> Coding Scheme
Version
(0008,0103)
1C
See Section 8.2. Required if a sequence item is present
and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.
*** export not supported
>> Code Meaning
(0008,0104)
1C
See Section 8.3. Required if a sequence item is present.
*** export not supported
RT Image C.8.8.2
Attribute Name
Tag
Description
Handling
Samples per Pixel
(0028,0002)
Number of samples (planes) in this image. See
C.8.8.2.6.1 for specialization.
Export supported
Photometric
Interpretation
(0028,0004)
Specifies the intended interpretation of the pixel data.
See C.7.6.3.1.2 for further explanation.
Export supported
Value set to MONOCHROME2
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
91 of 118
IOD Details, RT Image C.8.8.2
92 of 118
Attribute Name
Tag
Description
Handling
Bits Allocated
(0028,0100)
Number of bits allocated for each pixel sample. Each
sample shall have the same number of bits allocated.
See PS 3.5 for further explanation.
Export supported
Bits Stored
(0028,0101)
Number of bits stored for each pixel sample. Each
sample shall have the same number of bits stored. See
PS 3.5 for further explanation.
Export supported
High Bit
(0028,0102)
Most significant bit for pixel sample data. Each sample
shall have the same high bit. See PS 3.5 for further
explanation.
Export supported
Pixel
Representation
(0028,0103)
Data representation of the pixel samples. Each sample
shall have the same pixel representation. Enumerated
Values:
0000H = unsigned integer.
0001H = 2's complement
Export supported
Pixel Intensity
Relationship
(0028,1040)
The relationship between the Pixel sample values and
the X-Ray beam intensity.
Enumerated Values:
LIN = Linearly proportional to X-Ray beam intensity
LOG = Logarithmically proportional to X- Ray beam
intensity
See C.8.11.3.1.2 for further explanation.
*** export not supported
Pixel Intensity
Relationship Sign
(0028,1041)
1C
The sign of the relationship between the Pixel sample
values stored in Pixel Data (7FE0,0010) and the X-Ray
beam intensity.
Required if Pixel Intensity Relationship (0028,1040) is
present. Enumerated Values; 1 = Lower pixel values
correspond to less X-Ray beam intensity -1 = Higher pixel
values correspond to less X-Ray beam intensity See
C.8.11.3.1.2 for further explanation.
*** export not supported
RT Image Label
(3002,0002)
User-defined label for RT Image.
Export supported
Used as Image ID.
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
IOD Details, RT Image C.8.8.2
Attribute Name
Tag
Description
Handling
RT Image Name
(3002,0003)
User-defined name for RT Image.
*** export not supported
RT Image
Description
(3002,0004)
User-defined description of RT Image.
Export supported for MV image only
Operators Name
(0008,1070)
Name of operator(s) acquiring or creating RT Image.
*** export not supported
Image Type
(0008,0008)
Image identification characteristics (see Section
C.7.6.1.1.2). RT Images shall use one of the following
Defined Terms for Value 3: DRR = digitally reconstructed
radiograph PORTAL = digital portal image or portal film
image SIMULATOR = conventional simulator image
RADIOGRAPH = radiographic image BLANK = image
pixels set to background value FLUENCE = fluence map
Export supported
kV:
ORIGINAL\PRIMARY\PORTAL\OBI
MV: ORIGINAL\PRIMARY\PORTAL
Value for acquired portal dose:
ORIGINAL\PRIMARY\PORTAL\ACQUI
RED_DOSE
Conversion Type
(0008,0064)
Describes the kind of image conversion.
Defined Terms:
DV = Digitized Video
DI = Digital Interface
DF = Digitized Film
WSD = Workstation
Export supported
Value set to DI
Reported Values
Origin
(3002,000A)
2C
Describes the origin of the parameter values reported in
the image. Required if Value 3 of Image Type
(0008,0008) is SIMULATOR or PORTAL.
Enumerated Values:
OPERATOR = manually entered by operator
PLAN = planned parameter values
ACTUAL = electronically recorded
Export supported
Value set to ACTUAL
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
93 of 118
IOD Details, RT Image C.8.8.2
94 of 118
Attribute Name
Tag
Description
Handling
RT Image Plane
(3002,000C)
Describes whether or not image plane is normal to beam
axis.
Enumerated Values:
NORMAL = image plane normal to beam axis
NON_NORMAL = image plane non-normal to beam axis
Export supported
Value set to NORMAL
X-Ray Image
Receptor
Translation
(3002,000D)
Position in (x,y,z) coordinates of origin of IEC X-RAY
IMAGE RECEPTOR System in the IEC GANTRY
coordinate system (mm). See Note 2.
Export supported
X-Ray Image
Receptor Angle
(3002,000E)
X-Ray Image Receptor Angle i.e. orientation of IEC XRAY IMAGE RECEPTOR coordinate system with respect
to IEC GANTRY coordinate system (degrees). See
C.8.8.2.2.
Export supported
RT Image
Orientation
(3002,0010)
2C
The direction cosines of the first row and the first column
with respect to the IEC XRAY IMAGE RECEPTOR
coordinate system. Required if RT Image Plane
(3002,000C) is NON_NORMAL. May be present
otherwise.
Export supported
Image Plane Pixel
Spacing
(3002,0011)
Physical distance (in mm) between the center of each
image pixel, specified by a numeric pair - adjacent row
spacing (delimiter) adjacent column spacing. See
C.8.8.2.3.
Export supported
RT Image Position
(3002,0012)
The x and y coordinates (in mm) of the upper left hand
corner of the image, in the IEC X-RAY IMAGE
RECEPTOR coordinate system. This is the center of the
first pixel transmitted. See C.8.8.2.7.
Export supported
Radiation Machine
Name
(3002,0020)
User-defined name identifying radiation machine used in
acquiring or computing image (i.e. name of conventional
simulator, electron accelerator, X-ray device, or machine
modeled when calculating DRR).
*** export not supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
IOD Details, RT Image C.8.8.2
Attribute Name
Tag
Description
Handling
Primary Dosimeter
Unit
(300A,00B3)
Measurement unit of machine dosimeter.
Enumerated Values:
MU = Monitor Unit
MINUTE = minute
Export supported
Value for MV Image: MU
Value for kV Image: MINUTE
Radiation Machine
SAD
(3002,0022)
Radiation source to Gantry rotation axis distance of
radiation machine used in acquiring or computing image
(mm).
Export supported
Radiation Machine
SSD
(3002,0024)
Source to patient surface distance (in mm) of radiation
machine used in acquiring or computing image.
*** export not supported
RT Image SID
(3002,0026)
Distance from radiation machine source to image plane
(in mm) along radiation beam axis. See C.8.8.2.3.
Export supported
Source to
Reference Object
Distance
(3002,0028)
Source to reference object distance (in mm), as used for
magnification calculation of RADIOGRAPH and
SIMULATOR images.
*** export not supported
Referenced RT Plan
Sequence
(300C,0002)
Introduces sequence of one Class/Instance pair
describing RT Plan associated with image. Only a single
item shall be permitted in this sequence.
Export supported
> Referenced SOP
Class UID
(0008,1150)
1C
Uniquely identifies the referenced SOP Class. Required if
Referenced RT Plan Sequence (300C,0002) is sent.
Export supported
Supported values: SOP Class UID of
RT Plan and RT Ion Plan.
> Referenced SOP
Instance UID
(0008,1155)
1C
Uniquely identifies the referenced SOP Instance.
Required if Referenced RT Plan Sequence (300C,0002)
is sent.
Export supported
Plan UID of the plan where this image
is a field image.
Referenced Beam
Number
(300C,0006)
Uniquely identifies the corresponding N- segment
treatment beam specified by Beam Number (300A,00C0)
within Beam Sequence in RT Beams Module within the
RT Plan referenced in Referenced RT Plan Sequence
(300C,0002).
Export supported
Referenced Fraction
Group Number
(300C,0022)
Identifier of Fraction Group within RT Plan referenced in
Referenced RT Plan Sequence (300C,0002).
*** export not supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
95 of 118
IOD Details, RT Image C.8.8.2
96 of 118
Attribute Name
Tag
Description
Handling
Fraction Number
(3002,0029)
Fraction Number of fraction during which image was
acquired, within Fraction Group referenced by
Referenced Fraction Group Number (300C,0022) within
RT Plan referenced in Referenced RT Plan Sequence
(300C,0002).
Export supported
Value set to 1.
Start Cumulative
Meterset Weight
(300C,0008)
Cumulative Meterset Weight within Beam referenced by
Referenced Beam Number (300C,0006) at which image
acquisition starts.
Export supported for MV Image only
End Cumulative
Meterset Weight
(300C,0009)
Cumulative Meterset Weight within Beam referenced by
Referenced Beam Number (300C,0006) at which image
acquisition ends.
*** export not supported
Exposure Sequence
(3002,0030)
Introduces sequence of Exposure parameter sets,
corresponding to exposures used in generating the
image. One or more items may be included in this
sequence. See C.8.8.2.4.
Export supported.
Only 1 item supported
> Referenced
Frame Number
(0008,1160)
1C
Identifies corresponding image frame in multi-frame
image. Required if Exposure Sequence (3002,0030) is
sent, there is more than one item in Exposure Sequence
(3002,0030), and image is a multi-frame image.
*** export not supported
> KVP
(0018,0060)
2C
Peak kilo voltage output (kV) of X-ray generator used to
acquire image. Required if Value 3 of Image Type
(0008,0008) is PORTAL, SIMULATOR or RADIOGRAPH
and Exposure Sequence (3002,0030) is sent.
Export supported
> X-ray Tube
Current
(0018,1151)
2C
Imaging device X-ray Tube Current (mA).
Required if Value 3 of Image Type (0008,0008) is
SIMULATOR or RADIOGRAPH and Exposure Sequence
(3002,0030) is sent.
Export supported for kV Image
> Exposure Time
(0018,1150)
2C
Time of X-ray exposure (msec). Required if Value 3 of
Image Type (0008,0008) is SIMULATOR or
RADIOGRAPH and Exposure Sequence (3002,0030) is
sent.
Export supported for kV Image
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
IOD Details, RT Image C.8.8.2
Attribute Name
Tag
Description
Handling
> Meterset
Exposure
(3002,0032)
2C
Treatment machine Meterset duration over which image
has been acquired, specified in Monitor units (MU) or
minutes as defined by Primary Dosimeter Unit
(300A,00B3).
Required if Value 3 of Image Type (0008,0008) is
PORTAL and Exposure Sequence (3002,0030) is sent.
Export supported
> Diaphragm
Position
(3002,0034)
Positions of diaphragm jaw pairs (in mm) in IEC BEAM
LIMITING DEVICE coordinate axis in the IEC order X1,
X2, Y1, Y2.
*** export not supported
> Beam Limiting
Device Sequence
(300A,00B6)
Introduces sequence of beam limiting device (collimator)
jaw or leaf (element) positions for given exposure. One or
more items may be included in this sequence.
Export supported
>> RT Beam
Limiting Device
Type
(300A,00B8)
1C
Type of beam limiting device (collimator).
Required if Beam Limiting Device Sequence (300A,00B6)
is sent.
Enumerated Values:
X = symmetric jaw pair in IEC X direction
Y = symmetric jaw pair in IEC Y direction
ASYMX = asymmetric jaw pair in IEC X direction
ASYMY = asymmetric pair in IEC Y direction
MLCX = multileaf (multi-element) jaw pair in IEC X
direction
MLCY = multileaf (multi-element) jaw pair in IEC Y
direction
Export supported for MV Image only
Supported values: X, Y, ASYMX,
ASYMY.
>> Source to Beam
Limiting Device
Distance
(300A,00BA)
Radiation source to beam limiting device (collimator)
distance (mm).
*** export not supported
>> Number of
Leaf/Jaw Pairs
(300A,00BC)
1C
Number of leaf (element) or jaw pairs (equal to 1 for
standard beam limiting device jaws). Required if Beam
Limiting Device Sequence (300A,00B6) is sent.
Export supported for MV Image only
Value set to 1.
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
97 of 118
IOD Details, RT Image C.8.8.2
98 of 118
Attribute Name
Tag
Description
Handling
>> Leaf Position
Boundaries
(300A,00BE)
2C
Boundaries (in mm) of beam limiting device (collimator)
leaves (elements) in IEC BEAM LIMITING DEVICE
coordinate axis appropriate to RT Beam Limiting Device
Type (300A,00B8), i.e. X-axis for MLCY, Y- axis for
MLCX. Contains N+1 values, where N is the Number of
Leaf/Jaw Pairs (300A,00BC), starting from Leaf
(Element) Pair 1. Required if RT Beam Limiting Device
Type (300A,00B8) is MLCX or MLCY.
*** export not supported
>> Leaf/Jaw
Positions
(300A,011C)
1C
Positions of beam limiting device (collimator) leaf or jaw
(element) pairs (in mm) in IEC BEAM LIMITING DEVICE
coordinate axis appropriate to RT Beam Limiting Device
Type (300A,00B8), e.g. X- axis for MLCX, Y-axis for
MLCY). Contains 2N values, where N is the Number of
Leaf/Jaw Pairs (300A,00BC), in IEC leaf (element)
subscript order 101, 102, ... 1N, 201, 202, ... 2N.
Required if Beam Limiting Device Sequence (300A,00B6)
is sent.
Export supported for MV Image only
> Applicator
Sequence
(300A,0107)
Introduces sequence of Applicators associated with
Beam. Only a single item shall be permitted in this
sequence.
*** export not supported
>> Applicator ID
(300A,0108)
1C
User or machine supplied identifier for Applicator.
Required if Applicator Sequence (300A,0107) is sent.
*** export not supported
>> Applicator Type
(300A,0109)
1C
Type of Applicator. Required if Applicator Sequence
(300A,0107) is sent.
Defined Terms:
ELECTRON_SQUARE = square electron applicator
ELECTRON_RECT = rectangular electron applicator
ELECTRON_CIRC = circular electron applicator
ELECTRON_SHORT = short electron applicator
ELECTRON_OPEN = open (dummy) electron applicator
INTRAOPERATIVE = intraoperative (custom) applicator
STEREOTACTIC = stereotactic applicator
*** export not supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
IOD Details, RT Image C.8.8.2
Attribute Name
Tag
Description
Handling
>> Applicator
Description
(300A,010A)
User-defined description for Applicator.
*** export not supported
> Number of Blocks
(300A,00F0)
1C
Number of shielding blocks associated with Beam.
Required if Exposure Sequence (3002,0030) is sent.
Export supported
Value always 0.
> Block Sequence
(300A,00F4)
2C
Introduces sequence of blocks associated with Beam.
Required if Number of Blocks (300A,00F0) is non-zero.
One or more items may be included in this sequence.
*** export not supported
>> Block Tray ID
(300A,00F5)
User-supplied identifier for block tray.
*** export not supported
>> Source to Block
Tray Distance
(300A,00F6)
2C
Radiation Source to attachment edge of block tray
assembly (mm). Required if Block Sequence
(300A,00F4) is sent.
*** export not supported
>> Block Type
(300A,00F8)
1C
Type of block. Required if Block Sequence (300A,00F4)
is sent.
Enumerated Values:
SHIELDING = blocking material is inside contour
APERTURE = blocking material is outside contour
*** export not supported
>> Block
Divergence
(300A,00FA)
2C
Indicates presence or otherwise of geometrical
divergence. Required if Block Sequence (300A,00F4) is
sent.
Enumerated Values:
PRESENT = block edges are shaped for beam
divergence
ABSENT = block edges are not shaped for beam
divergence
*** export not supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
99 of 118
IOD Details, RT Image C.8.8.2
100 of 118
Attribute Name
Tag
Description
Handling
>> Block Mounting
Position
(300A,00FB)
Indicates on which side of the Block Tray the block is
mounted.
Enumerated Values:
PATIENT_SIDE = the block is mounted on the side of the
Block Tray which is towards the patient.
SOURCE_SIDE = the block is mounted on the side of the
Block Tray which is towards the radiation source.
*** export not supported
>> Block Number
(300A,00FC)
1C
Identification Number of the Block. The value of Block
Number (300A,00FC) shall be unique within the Beam in
which it is created. Required if Block Sequence
(300A,00F4) is sent.
*** export not supported
>> Block Name
(300A,00FE)
User-defined name for block.
*** export not supported
>> Material ID
(300A,00E1)
2C
User-supplied identifier for material used to manufacture
Block. Required if Block Sequence (300A,00F4) is sent.
*** export not supported
>> Block Thickness
(300A,0100)
Physical thickness of block (in mm) parallel to radiation
beam axis.
*** export not supported
>> Block Number of
Points
(300A,0104)
2C
Number of (x,y) pairs defining the block edge. Required if
Block Sequence (300A,00F4) is sent.
*** export not supported
>> Block Data
(300A,0106)
2C
A data stream of (x,y) pairs which comprise the block
edge. The number of pairs shall be equal to Block
Number of Points (300A,0104), and the vertices shall be
interpreted as a closed polygon. Coordinates are
projected onto the machine isocentric plane in the IEC
BEAM LIMITING DEVICE coordinate system (mm).
Required if Block Sequence (300A,00F4) is sent.
*** export not supported
> Primary Fluence
Mode Sequence
(3285,XX00)
Sequence defining whether the primary fluence of the
treatment beam uses a non-standard fluence-shaping.
Only a single Item shall be permitted in this sequence.
*** export not supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
IOD Details, RT Image C.8.8.2
Attribute Name
Tag
Description
Handling
>> Fluence Mode
(3285,XX01)
Describes whether the fluence shaping is the standard
mode for the beam or an alternate.
Enumerated Values:
STANDARD = Uses standard fluence-shaping
NON_STANDARD = Uses a non-standard fluenceshaping mode
*** export not supported
>> Fluence Mode ID
(3285,XX02)
1C
Identifier for the specific fluence-shaping mode. Required
if Fluence Mode (3285,xx01) has value
NON_STANDARD.
*** export not supported
Fluence Map
Sequence
(3002,0040)
1C
A Sequence of data describing the fluence map attributes
for a radiotherapy beam. Only one item may be included
in this sequence.
Required if the third value of Image Type (0008,0008) is
FLUENCE.
*** export not supported
> Fluence Data
Source
(3002,0041)
Source of fluence data.
Enumerated Values:
CALCULATED = Calculated by a workstation
MEASURED = Measured by exposure to a film or
detector.
*** export not supported
> Fluence Data
Scale
(3002,0042)
The meterset corresponding with a fluence map cell
value of 1.0 expressed in units specified by Primary
Dosimeter Units (300A,00B3). This is the meterset value
used for treatment, not the meterset used to expose the
film as defined by Meterset Exposure (3002,0032).
*** export not supported
Gantry Angle
(300A,011E)
Treatment machine gantry angle, i.e. orientation of IEC
GANTRY coordinate system with respect to IEC FIXED
REFERENCE coordinate system (degrees).
Export
Beam Limiting
Device Angle
(300A,0120)
Treatment machine beam limiting device (collimator)
angle, i.e. orientation of IEC BEAM LIMITING DEVICE
coordinate system with respect to IEC GANTRY
coordinate system (degrees).
Export supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
101 of 118
IOD Details, RT Image C.8.8.2
102 of 118
Attribute Name
Tag
Description
Handling
Patient Support
Angle
(300A,0122)
Patient Support angle, i.e. orientation of IEC PATIENT
SUPPORT coordinate system with respect to IEC FIXED
REFERENCE coordinate system (degrees).
Export supported
Table Top Eccentric
Axis Distance
(300A,0124)
Distance (positive) from the IEC PATIENT SUPPORT
vertical axis to the IEC TABLE TOP ECCENTRIC vertical
axis (mm).
*** export not supported
Table Top Eccentric
Angle
(300A,0125)
Table Top (non-isocentric) angle, i.e. orientation of IEC
TABLE TOP ECCENTRIC coordinate system with
respect to IEC PATIENT SUPPORT system (degrees).
*** export not supported
Table Top Vertical
Position
(300A,0128)
Table Top Vertical position in IEC TABLE TOP
coordinate system (mm).
Export supported
Table Top
Longitudinal
Position
(300A,0129)
Table Top Longitudinal position in IEC TABLE TOP
coordinate system (mm).
Export supported
Table Top Lateral
Position
(300A,012A)
Table Top Lateral position in IEC TABLE TOP coordinate
system (mm).
Export supported
Isocenter Position
(300A,012C)
Isocenter coordinates (x,y,z), in mm. Specifies the
location of the machine isocenter in the patient-based
coordinate system associated with the Frame of
Reference. It allows transformation from the equipmentbased IEC coordinate system to the patient-based
coordinate system.
Export supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
IOD Details, RT Image C.8.8.2
Attribute Name
Tag
Description
Handling
Patient Position
(0018,5100)
1C
Patient position descriptor relative to the patient support
device.
Required if Isocenter Position (300A,012C) is present.
May be present otherwise. See Section C.7.3.1.1.2 for
Defined Terms and further explanation.
Note:
The orientation of the patient relative to the patient
support device is denoted in the same manner as in the
RT Patient Setup module. It defines the relation of the
patient-based DICOM coordinate system identified by the
frame of reference module of the RT Image to the IEC
coordinate system and together with the Isocenter
Position (300A,012C) allows the RT Image to be placed
into the patient frame of reference. It also allows a
system using an RT Image to verify that the patient is
setup in a similar position relative to the patient support
device.
Export supported
RT Image Isocenter
Position
(3273,XX00)
The isocenter position in the FOR as defined in the FOR
module. For DRRs, this is typically the FOR of the
diagnostic image and the isocenter coordinate
correspond to those of the associated beam. For
treatment / simulator images, the acquisition systems
create the FOR and define it's origin (typically the
isocenter position of the first image acquired).
Export supported
RT Image Patient
Position
(3273,XX01)
The patient position at the time, when the image was
taken (or created in case of DRRs).
Export supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
103 of 118
IOD Details, Structure Set C.8.8.5
Attribute Name
Tag
Description
Handling
LT Archive Primary
Reference Image
Flag
(3279,XX00)
If present indicates if image is primary reference image of
beam referenced by Referenced RT Plan Sequence
(300C,0002) and Referenced Beam Number
(300C,0006). If absent then it is not a primary reference
image.
*** export not supported
Enumerated Values:
PRIMARY
NOT_PRIMARY
B.2.11
104 of 118
Structure Set C.8.8.5
Attribute Name
Tag
Description
Handling
Structure Set Label
(3006,0002)
User-defined label for Structure Set.
Export supported
Structure Set Name
(3006,0004)
User-defined name for Structure Set.
*** export not supported
Structure Set
Description
(3006,0006)
User-defined description for Structure Set.
*** export not supported
Instance Number
(0020,0013)
A number that identifies this object instance.
*** export not supported
Structure Set Date
(3006,0008)
Date at which Structure Set was last modified.
Export supported
Structure Set Time
(3006,0009)
Time at which Structure Set was last modified.
Export supported
Referenced Frame
of Reference
Sequence
(3006,0010)
Introduces sequence of items describing Frames of
Reference in which the ROIs are defined. One or more
items may be included in this sequence. See C.8.8.5.1.
Export supported
> Frame of
Reference UID
(0020,0052)
1C
Uniquely identifies Frame of Reference within Structure
Set. Required if Referenced Frame of Reference
Sequence (3006,0010) is sent.
Export supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
IOD Details, Structure Set C.8.8.5
Attribute Name
Tag
Description
Handling
> Frame of
Reference
Relationship
Sequence
(3006,00C0)
Introduces sequence of transforms that relate other
Frames of Reference to this Frame of Reference.
*** export not supported
>> Related Frame
of Reference UID
(3006,00C2)
1C
Frame of Reference Coordinate System to be
transformed to the current Frame of Reference. Required
if Frame of Reference Relationship Sequence
(3006,00C0) is sent.
*** export not supported
>> Frame of
Reference
Transformation
Type
(3006,00C4)
1C
Type of Transformation. Required if Frame of Reference
Relationship Sequence (3006,00C0) is sent.
Defined Terms:
HOMOGENEOUS
*** export not supported
>> Frame of
Reference
Transformation
Matrix
(3006,00C6)
1C
Four-by-four transformation Matrix from Related Frame of
Reference to current Frame of Reference. Matrix
elements shall be listed in row-major order. Required if
Frame of Reference Relationship Sequence (3006,00C0)
is sent. See C.8.8.5.2.
*** export not supported
>> Frame of
Reference
Transformation
Comment
(3006,00C8)
Comment regarding the transformation between the
related and current Frames of Reference.
*** export not supported
> RT Referenced
Study Sequence
(3006,0012)
Introduces sequence of Studies containing series to be
referenced. One or more items may be included in this
sequence.
Export supported
>> Referenced SOP
Class UID
(0008,1150)
1C
Uniquely identifies the referenced SOP Class. Required if
RT Referenced Study Sequence (3006,0012) is sent.
Export supported
>> Referenced SOP
Instance UID
(0008,1155)
1C
Uniquely identifies the referenced SOP Instance.
Required if RT Referenced Study Sequence (3006,0012)
is sent.
Export supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
105 of 118
IOD Details, Structure Set C.8.8.5
106 of 118
Attribute Name
Tag
Description
Handling
>> RT Referenced
Series Sequence
(3006,0014)
1C
Introduces sequence of items describing series of images
within the referenced study which are used in defining the
Structure Set. Required if RT Referenced Study
Sequence (3006,0012) is sent. One or more items may
be included in this sequence.
Export supported
>>> Series Instance
UID
(0020,000E)
1C
Unique identifier for the series containing the images.
Required if RT Referenced Series Sequence (3006,0014)
is sent.
Export supported
>>> Contour Image
Sequence
(3006,0016)
1C
Introduces sequence of items describing images in a
given series used in defining the Structure Set (typically
CT or MR images). Required if RT Referenced Series
Sequence (3006,0014) is sent. One or more items may
be included in this sequence.
Export supported
>>>> Referenced
SOP Class UID
(0008,1150)
Uniquely identifies the referenced SOP Class.
Export supported
>>>> Referenced
SOP Instance UID
(0008,1155)
Uniquely identifies the referenced SOP Instance.
Export supported
>>>> Referenced
Frame Number
(0008,1160)
Identifies the frame numbers within the Referenced SOP
Instance to which the reference applies. The first frame
shall be denoted as frame number 1.
Note: This Attribute may be multi-valued.
Required if the Referenced SOP Instance is a multi-frame
image and the reference does not apply to all frames.
*** export not supported
Structure Set ROI
Sequence
(3006,0020)
Introduces sequence of ROIs for current Structure Set.
One or more items may be included in this sequence.
Export supported
> ROI Number
(3006,0022)
1C
Identification number of the ROI. The value of ROI
Number (3006,0022) shall be unique within the Structure
Set in which it is created. Required if Structure Set ROI
Sequence (3006,0020) is sent.
Export supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
IOD Details, Structure Set C.8.8.5
Attribute Name
Tag
Description
Handling
> Referenced
Frame of Reference
UID
(3006,0024)
1C
Uniquely identifies Frame of Reference in which ROI is
defined, specified by Frame of Reference UID
(0020,0052) in Referenced Frame of Reference
Sequence (3006,0010). Required if Structure Set ROI
Sequence (3006,0020) is sent.
Export supported
> ROI Name
(3006,0026)
2C
User-defined name for ROI. Required if Structure Set
ROI Sequence (3006,0020) is sent.
Export supported
> ROI Description
(3006,0028)
User-defined description for ROI.
*** export not supported
> ROI Volume
(3006,002C)
Volume of ROI (cubic centimeters).
*** export not supported
> ROI Generation
Algorithm
(3006,0036)
2C
Type of algorithm used to generate ROI.
Required if Structure Set ROI Sequence (3006,0020) is
sent.
Defined Terms:
AUTOMATIC = calculated ROI
SEMIAUTOMATIC = ROI calculated with user assistance
MANUAL = user-entered ROI
Export supported
> ROI Generation
Description
(3006,0038)
User-defined description of technique used to generate
ROI.
*** export not supported
ROI Contour
Sequence
(3006,0039)
Introduces sequence of Contour Sequences defining
ROIs. One or more items may be included in this
sequence.
Export supported
> Referenced ROI
Number
(3006,0084)
Uniquely identifies the referenced ROI described in the
Structure Set ROI Sequence (3006,0020).
Export supported
> ROI Display Color
(3006,002A)
RGB triplet color representation for ROI, specified using
the range 0-255.
Export supported
> Contour
Sequence
(3006,0040)
Introduces sequence of Contours defining ROI. One or
more items may be included in this sequence.
Export supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
107 of 118
IOD Details, Structure Set C.8.8.5
Attribute Name
Tag
Description
Handling
>> Contour
Geometric Type
(3006,0042)
Geometric type of contour.
Enumerated Values:
POINT = single point
Export supported
>> Number of
Contour Points
(3006,0046)
Number of points (triplets) in Contour Data (3006,0050).
Export supported
>> Contour Number
(3006,0048)
Identification number of the contour. The value of
Contour Number (3006,0048) shall be unique within the
Contour Sequence (3006,0040) in which it is defined. No
semantics or ordering shall be inferred from this attribute.
Export supported
>> Contour Data
(3006,0050)
Sequence of (x,y,z) triplets defining a contour in the
patient based coordinate system (mm).
Export supported
Note: Contour Data may not be properly encoded if Explicit-VR
transfer syntax is used and the VL of this attribute exceeds
65534 bytes.
108 of 118
RT ROI
Observations
Sequence
(3006,0080)
Introduces sequence of observations related to ROIs
defined in the ROI Module. One or more items may be
included in this sequence.
Export supported
> Observation
Number
(3006,0082)
Identification number of the Observation. The value of
Observation Number (3006,0082) shall be unique within
the RT ROI Observations Sequence (3006,0080).
Export supported
> Referenced ROI
Number
(3006,0084)
Uniquely identifies the referenced ROI described in the
Structure Set ROI Sequence (3006,0020).
Export supported
> ROI Observation
Label
(3006,0085)
User-defined label for ROI Observation.
Defined Terms:
InitLaserIso
InitMatchIso
AcqIsocenter
Export supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
IOD Details, Structure Set C.8.8.5
Attribute Name
Tag
Description
Handling
> RT ROI
Interpreted Type
(3006,00A4)
Type of ROI.
Defined Terms:
INITLASERISO
INITMATCHISO
ACQ_ISOCENTER
Export supported
> ROI Interpreter
(3006,00A6)
Name of person performing the interpretation.
Export supported
> ROI Physical
Properties
Sequence
(3006,00B0)
Introduces sequence describing physical properties
associated with current ROI interpretation. One or more
items may be included in this sequence.
Export supported
>> ROI Physical
Property
(3006,00B2)
Physical property specified by ROI Physical Property
Value (3006,00B4).
Defined Terms:
PATSUPPORT_ANGLE
Export supported
>> ROI Physical
Property Value
(3006,00B4
User-assigned value for physical property.
Export supported
Referenced
Structure Set
Relationship
Sequence
(3263,XX01)
Introduces sequence of related SOP Class/Instance pairs
describing related instances of structure sets. One or
more items may be included in this sequence.
Export supported
> Referenced SOP
Class UID
(0008,1150)
Uniquely identifies the referenced SOP Class.
Export supported
> Referenced SOP
Instance UID
(0008,1155)
Uniquely identifies the referenced SOP Instance.
Export supported
> Structure Set
Relationship
(3263,XX02)
Relationship of referenced structure set with respect to
current structure set. Required if Referenced Structure
Set Sequence (3263,1002) is sent.
Defined Terms:
PREDECESSOR = structure set used in derivation of
current structure set
ADDITION = structure set, for which the current structure
set is an addition
Export supported ADDITION is used
for new marker structures defined
during marker detection
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
109 of 118
IOD Details, Modality LUT C.11.1
B.2.12
110 of 118
Modality LUT C.11.1
Attribute Name
Tag
Description
Handling
Modality LUT
Sequence
(0028,3000)
1C
Defines a sequence of Modality LUTs. Only one Item
may be present. Shall not be present if Rescale Intercept
(0028,1052) is present.
*** export not supported
> LUT Descriptor
US or
(0028,3002)
1C
Specifies the format of the LUT Data in this Sequence.
See C.11.1.1 for further explanation. Required if the
Modality LUT Sequence (0028,3000) is sent.
*** export not supported
> LUT Explanation
(0028,3003)
Free form text explanation of the meaning of the LUT.
*** export not supported
> Modality LUT
Type
(0028,3004)
1C
Specifies the output values of this Modality LUT. See
C.11.1.1.2 for further explanation. Required if the
Modality LUT Sequence (0028,3000) is sent.
*** export not supported
> LUT Data US or
(0028,3006)
1C
LUT Data in this Sequence. Required if the Modality LUT
Sequence (0028,3000) is sent.
*** export not supported
Rescale Intercept
(0028,1052)
1C
The value b in relationship between stored values (SV)
and the output units specified in Rescale Type
(0028,1054). Output units = m*SV + b. Required if
Modality LUT Sequence (0028,3000) is not present. Shall
not be present otherwise.
Export supported for:
RT Image
Rescale Slope
(0028,1053)
1C
m in the equation specified by Rescale Intercept
(0028,1052). Required if Rescale Intercept is present.
Export supported for:
RT Image
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
IOD Details, Spatial Registration C.20.2
Attribute Name
Tag
Description
Handling
Rescale Type
(0028,1054)
1C
Specifies the output units of Rescale Slope (0028,1053)
and Rescale Intercept (0028,1052). See C.11.1.1.2 for
further explanation. Required if Rescale Intercept is
present.
Export supported for:
RT Image
Defined Terms supported additionally:
CU = Calibration Units
Gy/MU = Gray per MU
RD = Relative Dose
For portal dose images roundtrip is
supported for CU, Gy/MU and RD. All
other values are mapped to US.
For all other images all values are
supported.
B.2.13
Spatial Registration C.20.2
Attribute Name
Tag
Description
Handling
Content Date
(0008,0023)
The date the content creation started.
Export supported
Content Time
(0008,0033)
The time the content creation started.
Export supported
Instance Number
(0020,0013)
A number that identifies this instance
Export supported Value is set to 0.
Content Label
(0070,0080)
A label that is used to identify this registration.
Export supported
Content Description
(0070,0081)
A description of this registration.
Export supported
Content Creators
Name
(0070,0084)
Name of operator performing the registration (such as a
technologist or physician).
*** export not supported
Registration
Sequence
(0070,0308)
A sequence of one or more registration items. Each item
defines a spatial registration to the referenced images in
that item. All referenced images are in the same spatial
frame of reference or atlas.
Export supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
111 of 118
IOD Details, Spatial Registration C.20.2
112 of 118
Attribute Name
Tag
Description
Handling
> Frame of
Reference UID
(0020,0052)
1C
Identifies a Frame of Reference that may or may not be
an image set (e.g. atlas or physical space). See
C.7.4.1.1.1 for further explanation. Required if
Referenced Image Sequence (0008,1140) is absent. May
be present otherwise.
Export supported
> Referenced Image
Sequence
(0008,1140)
1C
Identifies the set of images registered in this sequence
item. One or more items shall be present. Required if
Frame of Reference UID (0020,0052) is absent. May be
present otherwise.
Export supported
>> Referenced SOP
Class UID
(0008,1150)
Uniquely identifies the referenced SOP Class.
Export supported
>> Referenced SOP
Instance UID
(0008,1155)
Uniquely identifies the referenced SOP Instance.
Export supported
>> Referenced
Frame Number
(0008,1160)
Identifies the frame numbers within the Referenced SOP
Instance to which the reference applies. The first frame
shall be denoted as frame number 1.
Note: This Attribute may be multi-valued.
Required if the Referenced SOP Instance is a multi-frame
image and the reference does not apply to all frames.
*** export not supported
> Matrix
Registration
Sequence
(0070,0309)
A sequence that specifies one spatial registration. Exactly
one item shall be present
Export supported
>> Frame of
Reference
Transformation
Comment
(3006,00C8)
User description or comments about the Transformation
Comment registration.
*** export not supported
>> Registration
Type Code
Sequence
(0070,030D)
Describes the information input into the registration
process. Only one item may be present.
Export supported
>>> Code Value
(0008,0100)
1C
See Section 8.1. Required if a sequence item is present.
Export supported
>>> Coding
Scheme Designator
(0008,0102)
1C
See Section 8.2. Required if a sequence item is present.
Export supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
IOD Details, Spatial Registration C.20.2
Attribute Name
Tag
Description
Handling
>>> Coding
Scheme Version
(0008,0103)
1C
See Section 8.2. Required if a sequence item is present
and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.
Export supported
>>> Code Meaning
(0008,0104)
1C
See Section 8.3. Required if a sequence item is present.
Export supported
>> Matrix Sequence
(0070,030A)
One or more items shall be present. Each item specifies
a transformation. The item order is significant and
corresponds to matrix multiplication order. See
C.20.2.1.1.
Export supported
Always one item is exported which
incorporates the combination of all
matrices of the original sequence.
>>> Frame of
Reference
Transformation
Matrix
(3006,00C6)
A 4x4 homogeneous transformation matrix that registers
the referenced images to the local RCS. Matrix elements
shall be listed in row-major order. See C.20.2.1.1.
Export supported
>>> Frame of
Reference
Transformation
Matrix Type
(0070,030C)
Type of Frame of Reference Transformation Matrix
(3006,00C6).
Defined terms:
RIGID
RIGID_SCALE
AFFINE
See C.20.2.1.2
Export supported
Value set to RIGID
>> Registration Sub
Type
(3275,XX00)
Identifies the application or application context in which
this registration object has been created.
Export supported
> Used Fiducials
Sequence
(0070,0314)
The fiducials used to determine the Frame of Reference
Transformation Matrix. One or more Items may be
present.
*** export not supported
>> Fiducial UID
(0070,031A)
The UID that identifies the fiducial used as registration
input.
*** export not supported
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
113 of 118
Extended Interface
Appendix C Extended Interface
The following IODs have private Attributes containing an XML data structure for transporting structured non-DICOM data:
RT Plan
CR/CT/MR/SC/RT Image
The following sections specify Attributes containing the XML data and describe the structure of the XML data.
C.1
114 of 118
RT Plan / RT Ion Plan
Node Name
Min
occurs
Max
occurs
Data
Type
Node Description
ExtendedVAPlanInterface
> Beams
>> Beam
Unbounded
>>> ReferencedBeamNumber
int
Beam identifier
>>> BeamExtension
>>>> FieldType
string
TREATMENT or SETUP
>>>> RelatedTreatBeamNumber
int
Relation from the image beams to the treatment and/or setup
beams.
>>>> FieldOrder
int
Order in which the fields are shown
>>>> GantryRtnExtendedStart
int
When true, gantry angle position is in the extended range, which
means it has an over travel (a Gantry can move most typical
400' and not only 360').
>>>> GantryRtnExtendedStop
int
When true, gantry angle position is in the extended range, which
means it has an over travel (a Gantry can move most typical
400' and not only 360').
>>>> MUSubtraction
int
MUSubtraction defines whether some MUs have to be
subtracted from the related treatment beam to take images in the
Additional items provided from Vision to expand the features of
RT Beams.
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
Extended Interface
Node Name
Min
occurs
Max
occurs
Data
Type
Node Description
before phase and/or after phase.
>>>> FieldSetupPhotos
>>>>> FieldSetupPhoto
>>>>>> SetupPhotoNumber
int
Increasing number of setup photos. The value has to start from 1
and up.
>>>>>> SetupPhotoFormat
string
The format of the pixel information. Supported values are 'JPG'
and 'BMP'
>>>>>> SetupPhotoPicture
hex
binary
Pixel information of the photo in the corresponding format.
>>>>>> PhotoModified
string
Photo has been modified.
PhotoModified may have true or false
>>> CustomAddOns
>>>> CustomAddOn
>>>>> CustomAddOnType
string
Three Add On Type to choose from:
BLOCK
COMPENSATOR
TRAY
>>>>> CustomCode
int
Codes for accessory which may be custom made or factory
made (in that case the internal code in equal the custom code.)
>>>>> ReferencedAddOnNumber
int
Reference to the identification number of the Block or
Compensator.
>>>>> TraySpecification
>>>>>> TrayID
string
User-supplied identifier for tray. Max length is 16 characters.
>>>>>> SourceTrayDistance
int
The distance from source to slot. This information is needed to
distinguish the slot
> ToleranceTables
OBI and PortalVision Advanced Imaging
Two setup photos max
In case the CustomAddOn is a tray, it gets reported back
through this node
DICOM Conformance Statement, KC1505D3CS
115 of 118
Extended Interface
116 of 118
Node Name
Min
occurs
Max
occurs
>> ToleranceTable
Unbounded
>>> ReferencedToleranceTableNumber
>>> ToleranceTableExtension
>>>> GantryRtnSetup
>>>> CollRtnSetup
Data
Type
Node Description
Additional items provided from Vision to expand the features of
Tolerance Tables with AutoSetup capabilities
int
Refers to the tolerance table in RT Tolerance Table module
string
Setup attributes may have following type:
Automatic
Manual
Remote
string
>>>> CollXSetup
string
>>>> CollYSetup
string
>>>> PatientSupportAngleSetup
string
>>>> CouchLngSetup
string
>>>> CouchVrtSetup
string
>>>> CouchLatSetup
string
>>>> TableTopEccentricAngleSetup
string
> DoseReferences
>> DoseReference
Unbounded
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
Extended Interface
Node Name
Min
occurs
Max
occurs
Data
Type
Node Description
>>> ReferencedDoseReferenceNumber
int
Reference to the identification number of the Dose Reference
(300A,0012) in the RT Prescription module which is a
representation of our reference point
>>> DoseReferenceExtension
>>>> DailyDoseLimit
decim
al
Unit is [Gy]
>>>> SessionDoseLimit
decim
al
Unit is [Gy]
>>>> Breakpoints
>>>>> Breakpoint
Unbounded
>>>>>> BreakpointDose
decim
al
Unit is [Gy]
>>>>>> BreakpointWarning
string
Text which describes the breakpoint. Max length is 254
characters
Additional items provided from Vision to expand the features of
reference points.
Breakpoint extensions, since DICOM has just one
DeliveryWarningDose without a description this is needed.
Table C-6 RT Plan Extended Interface
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS
117 of 118
Extended Interface
(This page is intentionally left blank.)
118 of 118
OBI and PortalVision Advanced Imaging
DICOM Conformance Statement, KC1505D3CS