On-Board Imager

PortalVision Advanced
Imaging

DICOM Conformance Statement

P/N KC1505D3CS

May 2010

General Information

Abstract

This document provides information about the DICOM Conformance of the On


Board Imager version 1.5/PortalVision Advanced Imaging product.
OBI and PV Advanced Imaging

Manufacturer and
European
Representative

Manufacturer:

European Representative:

Varian Medical Systems, Inc. Ltd.


3100 Hansen Way, Bldg. 4A
Palo Alto, CA 94304-1030, U.S.A.

Varian Medical Systems UK Ltd.


Gatwick Road, Crawley
West Sussex RH10 9RG
United Kingdom

Notice

Information in this document is subject to change without notice and does not
represent a commitment on the part of Varian. Varian is not liable for errors
contained in this document or for incidental or consequential damages in
connection with the furnishing or use of this material.
This document contains proprietary information protected by copyright. No part
of this document may be reproduced, translated, or transmitted without the
express written permission of Varian Medical Systems, Inc.

FDA 21 CFR 820


Quality System
Regulations (CGMPs)

Varian Medical Systems, Oncology Systems products are designed and


manufactured in accordance with the requirements specified within this federal
regulation.

ISO 13485

Varian Medical Systems, Oncology Systems products are designed and


manufactured in accordance with the requirements specified ISO 13485 quality
systems standards.

CE

Varian Medical Systems, Oncology Systems products meet the requirements of


Council Directive MDD 93/42/EEC.

HIPAA

Varians products and services are specifically designed to include features that
help our customers comply with the Health Insurance Portability and
Accountability Act of 1996 (HIPAA). The software application uses a secure
login process, requiring a user name and password, that supports role-based
access. Users are assigned to groups, each with certain access rights, which
may include the ability to edit and add data or may limit access to data. When a
user adds or modifies data within the database, a record is maintained of the
data that was changed, the users ID and the date and time the changes were
made. This establishes an audit trail that can be examined by authorized system
administrators.

Trademarks

ARIA, Eclipse, Varian and VMS are registered trademarks


ARIA, Eclipse, On-Board Imager are trademarks of Varian Medical
Systems, Inc.
Microsoft and Windows are registered trademarks of Microsoft Corporation.
All other trademarks or registered trademark are the property of their respective
owners.

2 of 118

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

General Information
Contacting Support

Support services are available without charge during the initial warranty period.
If you seek information not included in this publication, call Varian Medical
Systems support at the following locations:
United States and Canada telephone support + 1 888 827 4265
United States and Canada Direct telephone support + 1 650-213-1000
European telephone Support + 41 41 749 8844
Fax (US) + 1 702 938 4754
Fax (Service Europe) + 41 41 740 3340
All other countries please call your local service office
To contact the support location nearest you for Service, Parts or Support, see
the list at the Varian Medical Systems website:
Worldwide Listing
http://www.varian.com/us/oncology/services_and_support/contacts.html
Communicating Via the World Wide Web
If you have access to the Internet, you will find Varian Medical System support
at the following location:
Oncology Systems http://my.varian.com
If you have a Varian account, enter your username and password. Otherwise,
first click create new account to get a username and password.
From MyVarian home page, click Contact Us from the Support list along the left
side of the window.
If possible, please send all e-mail inquires through the my.varian.com web site
at http://my.varian.com/contactus; otherwise, use the following e-mails
addresses for support:
Sending E-Mail

North America (North America


and Canada)

[email protected]

Central & South America

[email protected]

Europe (Europe, Middle East,


Africa)

[email protected]

Australia (Australian, New


Zealand, Australasia)

[email protected]

China / Asia (China, Asia)

[email protected]

Japan

[email protected]

Brachy Therapy Systems

[email protected]

Copyright 2010 Varian Medical Systems Inc., Oncology Systems


All rights reserved.

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

3 of 118

General Information

Document History
1.5.05

May 11, 2010

Release version for OBI and PortalVision 1.5

1.4.06

Mar 5, 2008

Release version for OBI 1.4

4 of 118

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

Contents

Contents
Contents ............................................................................................................................. 5
List of Figures .................................................................................................................... 7
List of Tables ...................................................................................................................... 7
1. Introduction .................................................................................................................. 9
1.1
1.2
1.3
1.4
1.5

Audience ........................................................................................................................................... 9
Overview ........................................................................................................................................... 9
Remarks ......................................................................................................................................... 11
References ..................................................................................................................................... 11
Abbreviations .................................................................................................................................. 12

2. Networking ................................................................................................................. 13
2.1

2.2
2.3
2.4

Implementation Model .................................................................................................................... 13


2.1.1 Application Data Flow ........................................................................................................ 13
2.1.2 Functional Definition of AEs.............................................................................................. 14
2.1.3 Sequencing of Real World Activities.................................................................................. 15
AE Specifications............................................................................................................................ 16
2.2.1 OBI/PV AI Entity Specification ........................................................................................... 16
Network Interfaces.......................................................................................................................... 32
Configuration .................................................................................................................................. 33
2.4.1 AE Title/Presentation Address Mapping ............................................................................ 33
2.4.2 Parameters ........................................................................................................................ 34

3. Media Interchange ..................................................................................................... 35


4. Support of Character Sets ........................................................................................ 36
5. Security ...................................................................................................................... 37
5.1
5.2
5.3

Security Profiles .............................................................................................................................. 37


Association Level Security ............................................................................................................. 37
5.2.1 OBI/PV AI Client ................................................................................................................ 37
Application Level Security .............................................................................................................. 37
5.3.1 OBI/PV AI Application ........................................................................................................ 37

Appendix A
A.1

A.2
A.3

A.4

Specialization .......................................................................................... 38

IOD Contents .................................................................................................................................. 38


A.1.1 Created SOP Instances ..................................................................................................... 38
A.1.2 Usage of Attributes from received IODs ........................................................................... 38
Data Dictionary of Private Attributes .............................................................................................. 38
Coded Terminology and Templates ............................................................................................... 40
A.3.1 Context Groups.................................................................................................................. 40
A.3.2 Template Specifications .................................................................................................... 40
A.3.3 Private Code Definitions .................................................................................................... 40
Grayscale Image Consistency ........................................................................................................ 40

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

5 of 118

Contents
A.5
A.6

Standard Extended/Specialized/Private SOP Classes .................................................................. 40


Private Transfer Syntaxes .............................................................................................................. 40

Appendix B
B.1

B.2

Supported IODs .............................................................................................................................. 41


B.1.1 Computed Tomography Image A.3.3 ............................................................................. 41
B.1.2 Spatial Registration A.39.1.3 .......................................................................................... 41
B.1.3 RT Structure Set A.19.3 ................................................................................................. 42
B.1.4 RT Image A.17.3 ............................................................................................................ 42
Modules and Attributes................................................................................................................... 45
B.2.1 Patient C.7.1.1 ................................................................................................................ 45
B.2.2 General Study - C.7.2.1 ..................................................................................................... 47
B.2.3 General Series C.7.3.1 ................................................................................................... 54
B.2.4 Frame of Reference C.7.4.1 ........................................................................................... 69
B.2.5 General Equipment C.7.5.1 ............................................................................................ 70
B.2.6 General Image C.7.6.1 ................................................................................................... 71
B.2.7 Image Plane C.7.6.2 ....................................................................................................... 82
B.2.8 Image Pixel C.7.6.3 ........................................................................................................ 82
B.2.9 CT Image C.8.2.1 ........................................................................................................... 86
B.2.10 RT Image C.8.8.2 ........................................................................................................... 91
B.2.11 Structure Set C.8.8.5 .................................................................................................... 104
B.2.12 Modality LUT C.11.1 ..................................................................................................... 110
B.2.13 Spatial Registration C.20.2........................................................................................... 111

Appendix C
C.1

6 of 118

IOD Details .............................................................................................. 41

Extended Interface ............................................................................... 114

RT Plan / RT Ion Plan .................................................................................................................. 114

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

List of Figures

List of Figures
Figure 2-1 Implementation Model..................................................................................................................... 13
Figure 2-2 Sequence Diagram for CT Set loading ........................................................................................... 15

List of Tables
Table 1-1: Network Services used by OBI ......................................................................................................... 9
Table 1-2: Network Services used by PV AI .................................................................................................... 10
Table 1-3: Media File SOP Classes OBI .......................................................................................................... 10
Table 1-4: Media File SOP Classes PV AI ....................................................................................................... 10
Table 2-1 Supported SCU/SCP SOP Classes for OBI/PV AI Entity ................................................................ 16
Table 2-2 DICOM Application Context for OBI/PV AI Client Application Entity ............................................... 17
Table 2-3 Used implementation class UIDs ..................................................................................................... 17
Table 2-4 Object Categories used by OBI/PV AI Client Storage SCU Application Entity ................................ 18
Table 2-5 Presentation contexts used by OBI/PV AI Client Storage SCU Application Entity .......................... 18
Table 2-6 OBI/PV AI Client C-STORE Response Status Handling Behavior .................................................. 19
Table 2-7 OBI/PV AI Client Storage SCU Communication Failure Behavior ................................................... 19
Table 2-8 Supported Query/Retrieve Levels for Query/Retrieve SCU ............................................................. 21
Table 2-9 Proposed Presentation Contexts for OBI/PV AI Client Q/R SCU Application Entity ....................... 21
Table 2-10 OBI/PV AI Client C-FIND Response Status Handling Behavior .................................................... 21
Table 2-11 OBI/PV AI Client Q/R C-FIND SCU Communication Failure Behavior .......................................... 22
Table 2-12 Study Root Request Identifier for OBI/PV AI Client Q/R C-FIND SCU .......................................... 22
Table 2-13 OBI/PV AI Client C-MOVE Response Status Handling Behavior .................................................. 23
Table 2-14 OBI/PV AI Client Q/R C-MOVE SCU Communication Failure Behavior ....................................... 23
Table 2-15 Study Root Request Identifier for OBI/PV AI Client Q/R C-MOVE SCU ....................................... 23
Table 2-16 Supported Operations .................................................................................................................... 24
Table 2-17 Beam Setup kV extensions ............................................................................................................ 28
Table 2-18 Acceptable Presentation Contexts for OBI Client Storage SCP Application Entity and Receive
Storage Request ....................................................................................................................................... 31
Table 2-19 Response Status of OBI/PV AI Client Storage SCP and Receive Storage Request .................... 32
Table 2-20 AE Title Configuration Table .......................................................................................................... 33
Table 2-21 Configuration Parameters Table .................................................................................................... 34
Table A-1 Private Creator Identification Strings ............................................................................................... 40
Table B-2 Computed Tomography Image IOD Modules Support .................................................................... 41
Table B-3 Spatial Registration Image IOD Modules Support ........................................................................... 42
OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

7 of 118

List of Tables
Table B-4 RT Structure Set IOD Modules Support .......................................................................................... 42
Table B-5 RT Image IOD Modules Support ..................................................................................................... 43
Table C-6 RT Plan Extended Interface .......................................................................................................... 117

8 of 118

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

Specialization

1.

Introduction

1.1

Audience
This document is intended for the following groups of persons:

Marketing and Sales

System Integrators of medical equipment

Other vendors interfacing using DICOM

It is assumed, that the reader is familiar with the DICOM standard.


The document is structured along the template definition as specified in Part 2 of the DICOM
standard.

1.2

Overview
OBI Advanced Imaging and PortalVision Advanced Imaging are applications that are used to
setup patients for treatment purposes. To support this function the following DICOM services for
receiving and sending diagnostic imaging modalities are supported.
The following table lists network services supported by OBI
SOP Classes

User of Service (SCU)

Provider of Service (SCP)

CT Image Storage

Yes

Yes

Spatial Registration Storage

Yes

No

RT Structure Set Storage

Yes

Yes

RT Image Storage

Yes

Yes

RT Plan Storage

No

Yes

Study Root Query/Retrieve Information


Model FIND

Yes

No

Study Root Query/Retrieve Information


Model MOVE

Yes

No

Transfer

Query/Retrieve

Table 1-1: Network Services used by OBI

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

9 of 118

Specialization
The following table lists network services supported by PortalVision Advanced Imaging (PV AI)
SOP Classes

User of Service (SCU)

Provider of Service (SCP)

CT Image Storage

No

Yes

Spatial Registration Storage

Yes

No

RT Structure Set Storage

Yes

Yes

RT Image Storage

Yes

Yes

RT Plan Storage

No

Yes

Study Root Query/Retrieve Information


Model FIND

Yes

No

Study Root Query/Retrieve Information


Model MOVE

Yes

No

Transfer

Query/Retrieve

Table 1-2: Network Services used by PV AI


The following table lists Media File SOP Classes supported for file based export and import.
These SOP classes are only supported when the OBI application is operating in File mode.
SOP Classes

DICOM Media File Export

DICOM Media File Import

CT Image Storage

Yes

Yes

Spatial Registration Storage

Yes

No

RT Structure Set Storage

Yes

Yes

RT Image Storage

Yes

Yes

RT Plan Storage

No

Yes

Table 1-3: Media File SOP Classes OBI


The following table lists Media File SOP Classes supported for file based export and import.
These SOP classes are only supported when the PV AI application is operating in File mode.
SOP Classes

DICOM Media File Export

DICOM Media File Import

CT Image Storage

No

Yes

Spatial Registration Storage

Yes

No

RT Structure Set Storage

Yes

Yes

RT Image Storage

Yes

Yes

RT Plan Storage

No

Yes

Table 1-4: Media File SOP Classes PV AI

10 of 118

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

Specialization

1.3

Remarks
The scope of this Conformance Statement is to facilitate communication of the OBI/PV AI
application and other vendors medical equipment. The Conformance Statement should be read
and understood in conjunction with the DICOM standard [1].
The DICOM standard in the current version evolved in 1993 with DICOM 3.0. The definition of
the DICOM standard for radiotherapy data started in 1994 and has now reached a mature state.
Nowadays DICOM is the primary choice for exchanging data with an open standard protocol for
the majority of vendors and institutions. Varian Medical Systems is committed to this notion of
standard-based cross-vendor interoperability as well as making use of the DICOM protocol
among its own products.
DICOM, by itself, does not guarantee interoperability. However, the Conformance Statement
facilitates a first-level validation for interoperability between different applications supporting the
same DICOM functionality.
This Conformance Statement is not intended to replace validation with other DICOM equipment
to ensure proper exchange of information intended.
Because the DICOM standard is subject to ongoing changes, enhancements and improvements,
Varian Medical Systems reserves the right to advance their products by making use of upcoming
DICOM features.

1.4

References
[1]

Digital Imaging and Communications in Medicine (DICOM), Parts 1-18 (2008)


National Electrical Manufacturers Association (NEMA)
Rosslyn, VA
United States of America

[2]

Treatment Release 8.8.


DICOM Conformance Statement (P/N 10001678-03)
Varian Medical Systems Inc
Palo Alto, CA, USA

[3]

Varian System Server DICOM Conformance Statement (P/N VA8203D3CS)


Varian Medical Systems International AG
Baden, Switzerland

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

11 of 118

Specialization

1.5

Abbreviations
This section provides the definitions of terms, acronyms and abbreviations that are used
throughout the document
AE

Application Entity

DICOM

Digital Imaging and Communications in Medicine

NEMA

National Electrical Manufacturers Association

SCU

Service Class User

SCP

Service Class Provider

SOP

Service Object Pair

Treatment
System

12 of 118

Management

DICOM entity from which OBI/PV AI retrieves structure set data

OBI

Varians On-Board Imager application to provide Image Guided RT


features.

PV AI

PortalVision Advanced Imaging application to provide Image Guided RT


features for MV only machines.

UID

Unique Identifier used to identify an object by a worldwide unique


identifier (DICOM term)

VR

Value Representation, a data encoding method in DICOM

PDU

Protocol Data Unit (DICOM term)

IOD

Information Object Definition (DICOM term)

DIMSE

DICOM Message Service element

TCP/IP

Transmission Control Protocol / Internet Protocol, a widely used


computer networking protocol

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

Specialization

2.

Networking

2.1

Implementation Model

2.1.1

Application Data Flow

Local User
requests Query

FIND-SCU
OBI Application Entity

Remote
Application Entity
Receives Query
Command

Local User
requests Retrieval

MOVE-SCU
OBI Application Entity

Remote
Application Entity
Receives Retrieve
Command

Object Imported

STORAGE-SCP
OBI Application Entity

Requested
Instances sent by
remote Application
Entity

Local User
requests Storage

STORAGE-SCU
OBI Application Entity

Remote
Application Entity
Receives Storage
Command

DICOM
Standard
Interface

Figure 2-1 Implementation Model


Conceptually, the network services may be modeled as the following separate AEs, though in
fact all the AEs share a single (configurable) AE Title:

STORAGE-SCP, receives incoming RT images, CT images, RT Plans and Structure


Sets.

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

13 of 118

Specialization

2.1.2

STORAGE-SCU, sends outbound RT images, CT images, Structure Sets and Spatial


Registrations.

FIND-SCU, which queries remote AEs for lists of studies, series and instances.

MOVE-SCU, which retrieves selected studies, series or instances.

ECHO-SCU which sends outgoing echo request to Remote Application Entity

Functional Definition of AEs


Note that conceptually the network services may be modeled as the following separate AEs,
though in fact all the AEs share a single (configurable) AE Title.

2.1.2.1

OBI/PV AI Client Storage SCU Application Entity


The Storage SCU Application Entity is invoked when

2.1.2.2

A simple 2D, 2D/2D, marker or 3D/3D matching (OBI only) has been performed and
Spatial Registrations or Markers must be sent to the Remote AE.

A 3D/3D matching (OBI only) has been performed and the acquired CT set must be sent
to the Remote AE

Marker Detection has been performed and the markers must be sent to the Remote AE
as Structure set linked to the current reference CT image. Each marker is defined as
own point set with a single point in it.

Marker Match has been performed, the markers are stored as curves in the image object
and the image object must be sent to the Remote AE.

When the user saves an acquired image through the UI, the image is then sent to the
Remote AE.

The treatment fraction is closed and all unsaved acquired images and other IODs are
sent to the Remote AE.

OBI/PV AI Client FIND SCU, MOVE SCU Application Entity


The Query/Retrieve Application Entity is invoked when

2.1.2.3

The patient is loaded in the 4DITC application, and reference images need to be loaded
in OBI/PV AI

Structure Set and CT images must be loaded in order to perform marker


detection/matching or 3D/3D matching (OBI only).

OBI/PV AI Client Storage SCP Application Entity


The Storage SCP Application Entity is invoked in order to receive objects requested from a
remote Query/Retrieve SCP, such as:

14 of 118

Reference Images/Historical Images

Structure Sets

RT Plan

Reference CT image set

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

Specialization

2.1.3

Sequencing of Real World Activities


When a patient is loaded into the OBI/PV AI application, it loads each reference image/historical
image by requesting them from the remote AE.
CT images are loaded into the OBI/PV AI application typically after the user switches to marker
matching or 3D/3D matching state (OBI only). OBI/PV AI invokes a move command for each CT
Image related to the structure set represented by the CT Image Query/Retrieve SCU. For this
purpose, the OBI application provides the CT Image Storage SCP.
When a Marker Matching is performed, Storage SCU entity is invoked and the Marker Data (as
Structure Set) are stored to the remote AE.
When either a Simple 2D, 2D/2D, 3D/3D (OBI only) or marker matching is performed, the
Storage SCU entity is invoked and stores a Spatial Registration to the remote AE
When loading from or saving to the Remote Application Entity, the ECHO-SCU entity is typically
invoked to ensure that the remote AE is alive and responsive.
The loading of a reference CT set is described by the following sequence diagram, which
illustrates, that the plan and the structure set is retrieved as well to follow the appropriate
references:

User

Management System

OBI/PV AI

Request CT Load

Structure Set request


if the structure set IOD is null,
the Structure Set Instance UID
has changed,the plan from 4DC
is out of date, reload the plan.

Structure Set returned

Plan request
Plan returned

CT Slice request
CT Slice returned

CT Slice request
CT Slice returned

DICOM
Standard
Interface
Figure 2-2 Sequence Diagram for CT Set loading

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

15 of 118

Specialization

2.2

AE Specifications

2.2.1

OBI/PV AI Entity Specification

2.2.1.1

SOP Classes
The OBI/PV AI Entity provides standard conformance to the following DICOM SOP classes.
Abstract Syntax

Name

UID

RT Structure
Set Storage

1.2.840.10008.5.1.4.1.1.481.3

Transfer Syntax
Name List

Role

Extended
Negotiation

UID List

DICOM Implicit VR
Little Endian

1.2.840.10008.1.2

SCU

None

DICOM Implicit VR
Little Endian

1.2.840.10008.1.2

SCP

None

DICOM Explicit VR
Little Endian

1.2.840.10008.1.2.1

SCP

None

Spatial
Registration
Storage

1.2.840.10008.5.1.4.1.1.66.1

DICOM Implicit VR
Little Endian

1.2.840.10008.1.2

SCU

None

CT Image
Storage (OBI
only)

1.2.840.10008.5.1.4.1.1.2

DICOM Implicit VR
Little Endian

1.2.840.10008.1.2

SCU

None

DICOM Implicit VR
Little Endian

1.2.840.10008.1.2

SCP

None

DICOM Explicit VR
Little Endian

1.2.840.10008.1.2.1

SCP

None

DICOM Implicit VR
Little Endian

1.2.840.10008.1.2

SCU

None

DICOM Implicit VR
Little Endian

1.2.840.10008.1.2

SCP

None

DICOM Explicit VR
Little Endian

1.2.840.10008.1.2.1

SCP

None

DICOM Implicit VR
Little Endian

1.2.840.10008.1.2

SCP

None

DICOM Explicit VR
Little Endian

1.2.840.10008.1.2.1

SCP

None

RT Image
Storage

1.2.840.10008.5.1.4.1.1.1

RT Plan
Storage

1.2.840.10008.5.1.4.1.1.5

Study Root
Query/Retriev
e information
model- FIND

1.2.840.10008.5.1.4.1.2.2.1

DICOM Implicit VR
Little Endian

1.2.840.10008.1.2

SCU

None

Study Root
Query/Retriev
e information
model- MOVE

1.2.840.10008.5.1.4.1.2.2.2

DICOM Implicit VR
Little Endian

1.2.840.10008.1.2

SCU

None

Table 2-1 Supported SCU/SCP SOP Classes for OBI/PV AI Entity

16 of 118

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

Specialization

2.2.1.2

Association Policies

2.2.1.2.1

General
The DICOM standard application context name for DICOM 3.0 is always proposed.
Application Context Name

1.2.840.10008.3.1.1.1

Table 2-2 DICOM Application Context for OBI/PV AI Client Application Entity

2.2.1.2.2

Number of Associations
The OBI/PV AI Client Application Entity can initiate (by default) up to three associations at a
time. This value is configurable.

2.2.1.2.3

Asynchronous Nature
OBI/PV AI Client Application Entity does not support asynchronous communication.

2.2.1.2.4

Implementation Identifying Information


The following Implementation Class UIDs are used.Remote AE.
UID Value

UID Name

1.2.246.352.70.2.1.9

VMS Console System 6.5 SCU


(VMS Console Interface 1)

1.2.246.352.70.2.1.60.1

VMS OBI Interface 1

1.2.246.352.70.2.1.60.2

VMS OBI interface 2 Duet

Table 2-3 Used implementation class UIDs

2.2.1.3

Association Initiation Policy


The OBI/PV AI Client Application Entity does initiate Associations.

2.2.1.3.1

Activity Store (STORAGE-SCU)

2.2.1.3.1.1 Description and Sequencing of Activities


When the user performs a Simple 2D, 2D/2D, marker or 3D/3D matching (OBI only), the match
transformation will be stored as a Spatial Registration IOD. The OBI/PV AI administration tool
can be configured to disable the storage of spatial registrations.
When the user performs a marker detection on the reference CT image, the positions of the
detected markers will be stored in the Structure Set and linked to the current reference CT
image. The OBI/PV AI administration tool can be configured to disable the storage of the marker
data.
When the user performs a marker match, the positions of the matched markers will be stored in
the acquired RT Image object as curves. The OBI/PV AI administration tool can be configured to
disable the storage of the marker data The Curve Dimensions (5002,0005) will be 2, the Number
of Points (5002,0010) will be 1, Type of Data (5000,0020) will be POLY, Axis Units (5000,0030)
will be PIXL\PIXL, Axis Labels (5000,0040) will be Marker\Marker and Data Value
Representation (5000,0103) will be 3.
OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

17 of 118

Specialization
When the user acquires a verification CT set, each CT Image will be stored. Following isocenters
are stored as Points in the Structure Set which is linked to the acquired verification CT set. The
values of the RT ROI Interpreted Type (3006,00A4) will be as follows:

Initial Laser Isocenter (position before acquisition): INITLASERISO

Acquisition Isocenter (position during acquisition): ACQ_ISOCENTER

Initial Match Isocenter (position after acquisition): INITMATCHISO


The OBI administration tool can be configured to disable the storage of CT images and
isocenters.
The Storage SCU will be invoked and attempts to initiate a new Association. If multiple objects
shall be transferred then multiple C-STORE requests will be issued over this Association.

Object Category

SOP Class Name

SOP Class UID

Structure Set

RT Structure Set Storage

1.2.840.10008.5.1.4.1.1.481.3

Spatial Registration

Spatial Registration Storage

1.2.840.10008.5.1.4.1.1.66.1

RT Image

RT Image Storage

1.2.840.10008.5.1.4.1.1.1

CT Image

CT Image Storage (OBI only)

1.2.840.10008.5.1.4.1.1.2

Table 2-4 Object Categories used by OBI/PV AI Client Storage SCU Application Entity
2.2.1.3.1.2 Proposed Presentation Contexts
OBI/PV AI Client Storage SCU is capable of proposing the Presentation Contexts shown in the
following table.
Presentation Context Table
Abstract Syntax
Name

UID

Transfer Syntax
Name

UID

Role

Ext.
Neg.

Spatial
Registration
Storage

1.2.840.10008.5.1.4.1.1.66.1

DICOM Implicit
VR Little Endian

1.2.840.10008.1.2

SCU

None

RT Structure
Set Storage

1.2.840.10008.5.1.4.1.1.481.3

DICOM Implicit
VR Little Endian

1.2.840.10008.1.2

SCU

None

RT Image
Storage

1.2.840.10008.5.1.4.1.1.1

DICOM Implicit
VR Little Endian

1.2.840.10008.1.2

SCU

None

CT Image
Storage (OBI
only)

1.2.840.10008.5.1.4.1.1.2

DICOM Implicit
VR Little Endian

1.2.840.10008.1.2

SCU

None

Table 2-5 Presentation contexts used by OBI/PV AI Client Storage SCU Application Entity

18 of 118

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

Specialization
2.2.1.3.1.3 SOP Specific Conformance for all Storage SOP Classes
The behavior of OBI/PV AI Client Storage SCU upon storage is summarized in table below
Service
Status

Further Meaning

Error
Code

Behavior

Refused

Out of Resources

A7xx

Failure

Data Set does not


match SOP Class

A9xx

Cannot understand

Cxxx

Coercion of Data
Elements

B000

Data Set does not


match SOP Class

B007

Elements Discarded

B006

Success

Success

0000

The SCP has successfully stored the SOP


Instance. No user feedback is received.

Any other
status
code.

Depending on the type of the Error Code it is


either handled like a Failure or Warning.
However, Error Codes other than listed above
in this table should not occur, as they are not
defined for C-STORE (see [1]), PS 3.4, Table
B.2-1).

Warning

The user is informed that the C-STORE


request has failed. Status Comment
(0000,0902) is logged and displayed along
with additional error information.
The SCP has successfully stored the SOP
Instance. Because a warning status was
received Status Comment (0000,0902) is
logged .

Table 2-6 OBI/PV AI Client C-STORE Response Status Handling Behavior


After sending all SOP Instances, either successfully or not, the Association is released using ARELEASE and the results are presented to the user.
Exception

Behavior

Timeout

The Association is released using A-RELEASE and the


whole task of sending SOP Instances is aborted. The
user is informed about the timeout and available
information about it is logged.

Association aborted by the SCP or


network layers

The whole task of sending SOP Instances is aborted.

Table 2-7 OBI/PV AI Client Storage SCU Communication Failure Behavior


2.2.1.3.1.4 SOP Specific Conformance for RT Structure Set Storage
Detected marker positions will be saved to a new Structure Set. This structure set references the
planning CT frame of reference and contains the detected marker positions as they are projected
on the planning CT. The modified structure set gets a new UID and the private referenced
structure set sequence (3263,1001) is used to denote the original structure set (with the old
UID). This sequence maybe used by the Treatment Management System to combine the new
structure set and the original one to add the markers as detected by OBI/PV AI.

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

19 of 118

Specialization
The ROI Contour Sequence (3006, 0040) shall have the following definitions:

Contour Geometric type

(3006, 0042) = POINT

Number of Contour Points


(3006, 0046) = 1
The ROI observation module shall have the following definitions:
RT ROI Interpreted Type
(3006,00A4) = MARKER.
Marker position may be located between slice positions.
Before exporting modified volumetric structures, they get resampled onto the planes of the 3D
patient model. The volumetric ROI Contours of an exported RT Structure Set will thus always
reference an image slice.
All images used to construct the 3D patient model are referenced in Contour Image Sequence
(3006,0016) of RT Structure Set Module, even if they do not have any contours defined on them.
2.2.1.3.1.5 SOP Specific Conformance for Spatial Registration Storage
The spatial registration of the 2D, 2D/2D and 3D/3D Match will be used in the following way:

The Registration Sequence (0070,0308) will include all images (the reference images
and the acquired verification RT images) and the Frame of Reference UID of the frame
of reference of the reference RT images.

The frame of reference module of the spatial registration will belong to the frame of
reference of the acquired verification RT images.

The transformation matrix type will be RIGID.

The spatial registration of the 3D/3D Match will be used in the following way:

The Registration Sequence (0070,0308) will include all image slices (the reference CT
slices and the acquired verification CT slices) and the Frame of Reference UID of the
frame of reference of the reference CT slices.

The frame of reference module of the spatial registration will belong to the frame of
reference of the acquired verification CT image slices.

The transformation matrix type will be RIGID.

The spatial registration of the Marker Match will be used in the following way:

2.2.1.3.2

The Registration Sequence (0070,0308) will include all image slices (the reference CT
slices and the acquired verification RT images) and the Frame of Reference UID of the
frame of reference of the reference CT slices.

The frame of reference module of the spatial registration will belong to the frame of
reference of the acquired verification RT images.

The transformation matrix type will be RIGID.

Activity Query/Retrieve (FIND-SCU, MOVE-SCU)

2.2.1.3.2.1 Description and Sequencing of Activities


This function is typically used by the application to load the reference images (e.g., DRRs) and
the historical images (Tx images from previous treatment sessions). This function is also used to
load the reference CT and its structure set after the user switches to MarkerMatch (both PVAI
and OBI) or 3D/3D Match state (OBI only). At this time an Association is requested.
20 of 118

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

Specialization
This application supports Query/Retrieve in the SCU role. The table below shows the supported
values for the tag Query/Retrieve Level (0008,0052):
Query/Retrieve Level

Value in (0008,0052)

Composite Object Instance Information

IMAGE

Table 2-8 Supported Query/Retrieve Levels for Query/Retrieve SCU


2.2.1.3.2.2 Proposed Presentation Contexts
Presentation Context Table
Abstract Syntax

Transfer Syntax
Name

Role

Ext.
Neg.

Name

UID

UID

Study Root
Query/Retrieve
information model
FIND

1.2.840.10008.5.1.4.1.2.
2.1

DICOM
Implicit VR
Little Endian

1.2.840.10008.1.2

SCU

None

Study Root
Query/Retrieve
information model
MOVE

1.2.840.10008.5.1.4.1.2.
2.2

DICOM
Implicit VR
Little Endian

1.2.840.10008.1.2

SCU

None

Table 2-9 Proposed Presentation Contexts for OBI/PV AI Client Q/R SCU Application
Entity
2.2.1.3.2.3 SOP Specific Conformance for C-FIND SOP Classes
The behavior of OBI/PV AI Client Query/Retrieve SCU when encountering status codes in a CFIND response is summarized in the table below.
Service
Status

Further Meaning

Error
Code

Behavior

Refused

Out of Resources

A7xx

The object is not found, the user is advised.

Failure

Data Set does not


match SOP Class

A9xx

Cannot understand

Cxxx

Cancel

Matching terminated
due to Cancel
Request

FE00

Cancel is handled like Failure, i.e. the object


is not found and the user is advised.

Success

Matching is
complete No final
Identifier is supplied

0000

The SCP has completed the match, the user


is advised.

Any other
status
code.

Handled like failure.

Table 2-10 OBI/PV AI Client C-FIND Response Status Handling Behavior


OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

21 of 118

Specialization
No C-CANCEL-FIND requests are ever issued.
Relational-queries are not supported.
Specific Character Set is not supported. It is not included in a query and will be ignored when
present in the response.
Exception

Behavior

Timeout

The user is informed that the operation (C-FIND or CMOVE) has timed out.

Association aborted by the SCP or


network layers

When the Association is aborted during a C-MOVE


operation the user is informed.

Table 2-11 OBI/PV AI Client Q/R C-FIND SCU Communication Failure Behavior
The table below lists the Attributes supported by the C-FIND SCU.
Name

Tag

VR

Type

Query/Retrieve Level

(0008,0052)

CS

SOP Instance UID

(0008,0018)

UI

Study Instance UID

(0020,000D)

UI

Series Instance UID

(0020,000E)

UI

Table 2-12 Study Root Request Identifier for OBI/PV AI Client Q/R C-FIND SCU
The Types of Matching column in the above table should be read as follows:
S

Single Value Matching

List of UID Matching

Universal Matching

Wild Card Matching

Range Matching

One or more matching types of the above list may be supported per Attribute.
2.2.1.3.2.4 SOP Specific Conformance for C-MOVE SOP Classes
The behavior of OBI/PV AI Client Query/Retrieve SCU when encountering status codes in a CMOVE response is summarized in the following table.
Service
Status
Refused

22 of 118

Further Meaning

Error
Code

Out of Resources
Unable to calculate
number of matches

A701

Out of Resources
Unable to perform
sub-operations

A702

Move Destination
unknown

A801

OBI and PortalVision Advanced Imaging

Behavior
None of the requested SOP Instances could
be retrieved. The user will be informed about
the failure.

DICOM Conformance Statement, KC1505D3CS

Specialization

Service
Status
Failure

Further Meaning

Error
Code

Behavior

Identifier does not


match SOP Class

A900

Unable to Process

Cxxx

Cancel

Sub-operations
terminated due to
Cancel Indication

FE00

Cancel is handled like Failure.

Warning

Sub-operations
complete One or
more Failures

B000

Some or all SOP Instances have not been


transmitted successfully. The user is
informed.

Success

Sub-operations
complete No
Failures

0000

All SOP Instances have successfully been


transmitted.

Pending

Sub-operations are
continuing

FF00

Transferring requested SOP Instances is


continuing. This message is ignored.

Any other
status
code.

Any other status code is handled like Failure.

Table 2-13 OBI/PV AI Client C-MOVE Response Status Handling Behavior


After having received all requested SOP Instances or when the user aborts the operation the
Association is released using A-RELEASE. All events occurring during querying and retrieving
SOP Instances are logged. If any log entries are marked with internal category Error they will be
shown to the user automatically.
No C-CANCEL-MOVE requests are ever issued.
Exception

Behavior

Timeout

The user is informed that the operation (C-FIND or CMOVE) has timed out.

Association aborted by the SCP or


network layers

When the Association is aborted during a C-MOVE


operation the user is informed.

Table 2-14 OBI/PV AI Client Q/R C-MOVE SCU Communication Failure Behavior
The table below lists the Attributes that will be sent by the C-MOVE SCU.
Name

Tag

VR

Type

Query/Retrieve Level

(0008,0052)

CS

SOP Instance UID

(0008,0018)

UI

Study Instance UID

(0020,000D)

UI

Series Instance UID

(0020,000E)

UI

Table 2-15 Study Root Request Identifier for OBI/PV AI Client Q/R C-MOVE SCU

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

23 of 118

Specialization
2.2.1.3.2.5 Supported Operations
This application performs only the operations listed in the table below. This table also shows
which of the optional key values (see [1]) are used for a request.
Operation

Target IOD

Key value used for Request

C-FIND

RT Structure Set

SOP Instance UID

C-MOVE

RT Structure Set

SOP Instance UID

C-MOVE

RT Plan

SOP Instance UID

C-FIND

RT Image

SOP Instance UID

C-MOVE

RT Image

SOP Instance UID

C-MOVE

CT Image

SOP Instance UID

Table 2-16 Supported Operations

2.2.1.4

Association Acceptance Policy


The OBI/PV AI Client Application Entity does accept storage requests for those objects
requested by the MOVE SCU

2.2.1.4.1

Activity Receive Storage Request

2.2.1.4.1.1 Description and Sequencing of Activities


Received SOP Instances are received and cached on the file system, as well as being passed to
the application.
2.2.1.4.1.2 General Preconditions
The following precondition for matching shall be fulfilled:
In a valid OBI/PV AI plan every treatment beam shall contain the same couch translation values
(vertical, lateral, longitudinal). It is allowed to have different couch rotation values in the plan
Tolerances for couch values being interpreted as equal are: absolute 2 mm for longitudinal,
lateral and vertical axes in space.
The plan should contain valid Imaging Device-Specific Acquisition Parameters (300A,00CC) to
allow automatic initiation of imaging procedures. However, if no Verification Image Sequence
items are present or they do not contain device-specific acquisition parameters as specified
above, it is still possible to initiate imaging procedures manually.
OBI/PV AI differentiates between treatment and setup fields. Treatment fields and setup fields
(kV and MV) can be used for OBI/PV AI use cases.
OBI/PV AI supports the Treatment Delivery Type (300A,00CE) SETUP. This attribute should
be the preferred method.

24 of 118

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

Specialization
For backwards compatibility, the parameter FieldType in the Extended Interface (3253,1000) can
be used to indicate the field type for each beam in the plan as well (please note, that this is a
deprecated approach):
<ExtendedVAPlanInterface>
<Beams>
<Beam>
<ReferencedBeamNumber>1</ReferencedBeamNumber>
<BeamExtension>

<FieldType>SETUP</FieldType>

</BeamExtension>
</Beam>
<Beam>
<ReferencedBeamNumber>2</ReferencedBeamNumber>
<BeamExtension>

<FieldType>TREATMENT</FieldType>

</BeamExtension>
</Beam>
</Beams>
</ExtendedVAPlanInterface>

Note that the aforementioned interface is only a portion of the entire interface, which is available
in Section C.1.
If a beam is a setup field and planned verification image sequence contains a single item
representing a kV image (see definition below), this beam is referred to as a kV beam in the
following.
Tag

Description

Beam Sequence
(300A,00B0)

Introduces sequence
of treatment beams
for current RT Plan.
One or more items
may be included in
this sequence.

Introduces sequence
of planned verification
images to be acquired
during current beam.
One or more items
may be included in
this sequence. See
C.8.8.14.2.

>Planned
Verification
Image
Sequence
(300A,00CA)

OBI and PortalVision Advanced Imaging

Convention / Interpretation

Used to plan RT images (kV or MV) or


(CB)CT images to be acquired with this
beam.

DICOM Conformance Statement, KC1505D3CS

25 of 118

Specialization

Tag

Description

Convention / Interpretation

>>Imaging
Device-Specific
Acquisition
Parameters
(300A,00CC)

User-specified
device-specific
parameters which
describe how the
imager will acquire
the image.

This tag is multivalued, values are


separated with a backslash character (\).
If this tag is missing completely, the planned
verification image is interpreted as film.
The following values are currently used for
portal (MV) imaging:
PortImageHighQuality
PortImageHighQuality\BeamOff
PortImageLowDose
PortImageLowDose\BeamOff
PortImageIntegrated
PortImageContinuous
The following values are currently used for
kV imaging:
KV\<ImageType>
Defined terms for ImageType:
Image
CBCT
Examples:
KV\Image
KV\CBCT

>Beam Limiting
Device
Sequence
(300A,00B6)
>Number of
Wedges
(300A,00D0)
>Number of
Compensators
(300A,00E0)

Introduces sequence
of beam limiting
device (jaw or leaf).

For kV beams, this sequence will not be


interpreted.

Number of wedges
associated with
current Beam.
Number of
compensators
associated with
current Beam.
Number of boli
associated with
current Beam.
Number of shielding
blocks associated
with Beam.
Introduces sequence
of machine
configurations
describing treatment
beam. Two or more
items may be
included in this
sequence. See
C.8.8.14.5 and
C.8.8.14.6.

For kV beams, this number shall be 0 (no


accessories)

For kV beams, this number shall be 0 (no


accessories)

For kV beams, this number shall be 0 (no


accessories)

For kV beams, this number shall be 0 (no


accessories)

For kV beams, this sequences shall always


include exactly 2 items.

>Number of Boli
(300A,00ED)
>Number of
Blocks
(300A,00ED)
>Control Point
Sequence
(300A,0111)

26 of 118

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

Specialization

Tag

Description

Convention / Interpretation

>>Beam Limiting
Device Position
Sequence
(300A,011A)

Introduces sequence
of beam limiting
device (collimator)
jaw or leaf (element)
positions. Required
for first item of Control
Point Sequence, or if
Beam Limiting Device
changes during
Beam. One or more
items may be
included in this
sequence.

1C

For kV beams, this sequence will not be


interpreted.

Beam Limiting Device


angle, i.e. orientation
of IEC BEAM
LIMITING DEVICE
coordinate system
with respect to IEC
GANTRY coordinate
system (degrees).
Required for first item
of Control Point
Sequence, or if Beam
Limiting Device Angle
changes during
Beam.

1C

For kV beams, this number shall be 0.0

Direction of Beam
Limiting Device
Rotation when
viewing beam limiting
device (collimator)
from radiation source,
for segment following
Control Point.
Required for first item
of Control Point
Sequence, or if Beam
Limiting Device
Rotation Direction
changes during
Beam. See
C.8.8.14.8.
Enumerated Values:
CW = clockwise
CC = counterclockwise
NONE = no rotation

1C

For kV beams, this value shall be NONE.

>>Beam Limiting
Device Angle
(300A,0120)

>>Beam Limiting
Device Rotation
Direction
(300A,0121)

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

27 of 118

Specialization

Tag

Description

Convention / Interpretation

>>Gantry Angle
(300A,011E)

Gantry angle of
radiation source, i.e.
orientation of IEC
GANTRY coordinate
system with respect
to IEC FIXED
REFERENCE
coordinate system
(degrees). Required
for first item of Control
Point Sequence, or if
Gantry Angle
changes during
Beam.

1C

For kV beams, this value specifies the


orientation of the kV radiation source with
respect to IEC FIXED coordinate system.
For MV beams, this value specifies the
orientation of the MV radiation source (main
gantry) with respect to IEC FIXED
coordinate system.

>>Gantry
Rotation
Direction
(300A,011F)

Direction of Gantry
Rotation when
viewing gantry from
isocenter, for
segment following
Control Point.
Required for first item
of Control Point
Sequence, or if
Gantry Rotation
Direction changes
during Beam. See
C.8.8.14.8.
Enumerated Values:
CW = clockwise
CC = counterclockwise
NONE = no rotation

1C

For kV beams, this value shall be NONE.

Table 2-17 Beam Setup kV extensions


When saving back on board images to the Treatment Management System, all images acquired
for the same plan and session will share the same frame of reference UID, even if the couch
position has been adjusted in between (under the assumption that the patient has not moved on
the couch significantly).
2.2.1.4.1.3 Preconditions for RT Images
RT Images, which are used as reference images (for 2D, 2D/2D and Marker Match), must be
scaled and contain all geometrical parameters needed to determine the position/orientation of
the Image in IEC. They must contain a frame of reference UID and the isocenter position in their
frame of reference.
The only geometrical parameter, which maybe missing is RT Image Position (3002,0012). If
missing, the image is assumed as centered around the z-axis of the Image Receptor system.

28 of 118

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

Specialization
2.2.1.4.1.4 Preconditions for 2D Match
Besides the General Preconditions, the following condition shall be fulfilled:
In a valid OBI/PV AI plan every treatment beam and every reference image shall contain the
same Isocenter Position (300A, 012C). Tolerance for the same Isocenter Position is that
isocenter positions of any pair of beams/reference images within the plan shall deviate less than
1.0 mm. If the plan contains multiple isocenters, the 2D match can be performed, but the couch
shifts cannot be applied.
The plan shall contain at least one valid setup beam configured as a kV or MV image beam or
one valid treatment beam configured as a MV image beam.
2.2.1.4.1.5 Preconditions for 2D/2D Match
Besides the General Preconditions, the following condition shall be fulfilled:
In a valid OBI/PV AI plan every treatment beam and every reference image shall contain no or
the same Isocenter Position (300A, 012C). Tolerance for the same Isocenter Position is that
isocenter positions of any pair of beams/reference images within the plan shall deviate less than
1.0 mm.
The plan shall contain two valid setup beams configured as a kV or MV image beam.
A delta of 90.0 degrees between the gantry/source angles (300A,011E) of the two beams is
recommended for an optimal match.
2.2.1.4.1.6 Preconditions for Marker Match
Besides the General Preconditions, the following condition shall be fulfilled:
In a valid OBI/PV AI plan every treatment beam and every reference image shall contain the
same Isocenter Position (300A, 012C). Tolerance for the same Isocenter Position is that
isocenter positions of any pair of beams/reference images within the plan shall deviate less than
1.0 mm
The plan shall contain two valid setup beams configured as a kV or MV image beam.
A delta of 90.0 degrees between the gantry/source angles (300A,011E) of the two beams is
recommended for an optimal match.
To load the slices of the reference CT image OBI/PV AI loads the plans structure set, which
corresponds to the Referenced RT Structure Set (300C,0060), from the Treatment Management
System , and subsequently loads all CT slices referenced in the Contour Image Sequence
(3006,0016).
Conditions for CT Slices:

All CT slices shall have the same Frame of Reference.

The spacing between CT slices positions in z-direction (DICOM Patient coordinate


system) may vary, although it is recommended, that they are equal throughout all slices
referenced by the Structure Set.

The maximal difference in x-direction and in y-direction (DICOM Patient coordinate


system) between all CT slices shall be less or equal 0.1 mm.

Conditions for Structure Set (except Structures of Contour Geometric type (3006, 0042)
POINT):

The contour points within a contour shall be positioned on the same contour plane within
-6
a maximal Epsilon of 10 mm in z direction (DICOM Patient coordinate system).

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

29 of 118

Specialization

Structures with contour(s) having contour points that are not positioned on the same
contour plane (see point above, i.e. non-transversal structures) are ignored (they are not
imported and a corresponding warning message is issued) and the import will continue
with the other structures.

Contour Geometric Type (3006, 0042) has to be either CLOSED_PLANAR or


OPEN_PLANAR for all contours of the structure. Structures with mixed contour types are
not supported.

Contour Geometric type (3006, 0042) POINT dont need to fulfill the conditions
mentioned above.
Conditions for Marker Point Structures:

The markers shall be in the structure set referenced by the plan.

Marker position may be located between slice positions.

The ROI Contour Sequence (3006, 0040) shall have the following definitions:
o Contour Geometric type
(3006, 0042) = POINT
o

Number of Contour Points (3006, 0046) = 1

The ROI observation module shall have the following definitions:


o RT ROI Interpreted Type (3006,00A4) = MARKER.
o

Private tag: Marker Subtype (3271,1000) = MARKER

2.2.1.4.1.7 Preconditions for 3D/3D Match (OBI only)


Besides the General Preconditions, the following condition shall be fulfilled:
In a valid OBI plan every treatment beam and every reference image shall contain the same
Isocenter Position (300A, 012C). Tolerance for the same Isocenter Position is that isocenter
positions of any pair of beams/reference images within the plan shall deviate less than 1.0 mm.
The plan shall contain a valid setup beam. The tag Beam Type (300A,00C4) shall have the
value STATIC configured as a CBCT beam. The tag Imaging Device Specific Acquisition
Parameters (300A,00CC) shall have the value KV\CBCT.
To load the slices of the reference CT image OBI loads the plans structure set, which
corresponds to the Referenced RT Structure Set (300C,0060), from the Treatment Management
System , and subsequently loads all CT slices referenced in the Contour Image Sequence
(3006,0016).
Conditions for CT Slices:

All CT slices shall have the same Frame of Reference.

The spacing between CT slices positions in z-direction (DICOM Patient coordinate


system) may vary, although it is recommended, that they are equal throughout all slices
referenced by the Structure Set.

The maximal difference in x-direction and in y-direction (DICOM Patient coordinate


system) between all CT slices shall be less or equal 0.1 mm.

Conditions for Structure Set (except Structures of Contour Geometric type (3006, 0042)
POINT):

30 of 118

The contour points within a contour shall be positioned on the same contour plane within
-6
a maximal Epsilon of 10 in z direction (DICOM Patient coordinate system).

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

Specialization

Structures with contour(s) having contour points that are not positioned on the same
contour plane (see point above, i.e. non-transversal structures) are ignored (they are not
imported and a corresponding warning message is issued) and the import will continue
with the other structures.

Contour Geometric Type (3006, 0042) has to be either CLOSED_PLANAR or


OPEN_PLANAR for all contours of the structure. Structures with mixed contour types are
not supported.

Contour Geometric type (3006, 0042) POINT dont need to fulfill the conditions
mentioned above.

2.2.1.4.1.8 Accepted Presentation Contexts


OBI Client Storage SCP Application Entity accepts Presentation Contexts shown in the following
table.
Presentation Context Table
Abstract Syntax
Name
All Storage
SOP Classes
in Table 4-1

Transfer Syntax

UID
All Storage SOP
Classes in Table
4-1

Name

Role

Ext.
Neg.

SCP

None

UID

DICOM Implicit VR
Little Endian

1.2.840.10008.1.2

DICOM Explicit VR
Little Endian

1.2.840.10008.1.2.1

Table 2-18 Acceptable Presentation Contexts for OBI Client Storage SCP Application
Entity and Receive Storage Request
2.2.1.4.1.9 SOP Specific Conformance for all Storage SOP Classes
2.2.1.4.1.9.1 Presentation Context Acceptance Criterion
OBI/PV AI Client Storage SCP will always accept any Presentation Context for the supported
SOP Classes with the supported Transfer Syntaxes. More than one proposed Presentation
Context will be accepted for the same Abstract Syntax if the Transfer Syntax is supported,
whether or not it is the same as another Presentation Context.
2.2.1.4.1.9.2 Transfer Syntax Selection Policies
If offered a choice of Transfer Syntaxes in a Presentation Context, OBI/PV AI Client Storage
SCP will select the first Transfer Syntax that is listed in the Presentation Context.
OBI/PV AI Client Storage SCP will accept duplicate Presentation Contexts, that is, if it is offered
multiple Presentation Contexts, each of which offers acceptable Transfer Syntaxes, it will accept
all Presentation Contexts, applying the same rule for selecting a Transfer Syntax for each as
described above.

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

31 of 118

Specialization
2.2.1.4.1.9.3 Response Status
OBI/PV AI Client Storage SCP will behave as described in the Table below when generating the
C-STORE response command message.

Service
Status
Refused
Failure

Warning

Success

Further Meaning

Error
Code

Out of Resources

A700

Out of Resources

A7xx

Invalid object
instance

0117

Data Set does not


match SOP Class

A9xx

Invalid C-STORE
request

C000

Coercion of Data
Elements

B000

Data Set does not


match SOP Class

B007

Elements Discarded

B006
0000

Behavior
Failed to receive incoming DICOM Stream.
Problem is logged on the SCP side.

Image successfully stored in temporary file.

Instance successfully stored in temporary file.

Table 2-19 Response Status of OBI/PV AI Client Storage SCP and Receive Storage
Request

2.3

Network Interfaces
See Varian System Server DICOM Conformance Statement [3]

32 of 118

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

Specialization

2.4

Configuration

2.4.1

AE Title/Presentation Address Mapping

2.4.1.1

Local AE Titles
The local OBI/PV AI Client Application Entity uses the AE Title and TCP/IP port number
configured via the Daemon configuration option in the OBI/PV AI Admin tool. Note that
conceptually the network services have been modeled as separate AEs, though in fact all the
AEs share a single (configurable) AE Title.
Application Entity

Default AE Title

Default TCP/IP Port

OBI/PV AI Client Application Entity

StreamService
(configurable)

58051 (configurable)

OBI/PV AI Client Storage SCU

See above

See above

OBI/PV AI Client Query/Retrieve SCU

See above

See above

Table 2-20 AE Title Configuration Table

2.4.1.2

Remote AE Title/Presentation Address Mapping

2.4.1.2.1

OBI/PV AI Client Storage SCU


For the Storage service class SCU, TCP/IP address, called AE title and port number of the
destination are configurable as well as the calling AE title used by OBI/PV AI.

2.4.1.2.2

OBI/PV AI Client Query/Retrieve SCU


For the OBI/PV AI Client Query/Retrieve SCU, TCP/IP address, called AE title and port number
of the provider are configurable. The calling AE title of the local application, which is also the
Move Destination AE title, is configurable too. The local port is the port of the Storage SCP for
receiving the data. The Storage SCP will accept only connection requests from the configured
remote AE Title and IP Address.

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

33 of 118

Specialization

2.4.2

Parameters
The following table shows DICOM relevant configuration parameters. While some of them may
be configured directly via a configuration dialog or a configuration application, other values can
only be accessed via the Registry or can not be configured at all.
Parameter

Configurable
(Yes/No)

Default Value

General Parameters
Max PDU Receive Size

No

Unlimited

Max PDU Send Size

No

32768 Bytes (32kB)

Time-out waiting for a acceptance or rejection


response to an Association request or Association
Release request (Application Level Timeout)

Yes

300s

Spatial Registration formatted along IHE-RO


specification 2008

Yes

No

OBI/PV AI Client Storage SCU Specific Parameters


OBI/PV AI Client Query/Retrieve SCU Specific Parameters
Time-out awaiting a Response to a DIMSE Request
(Low-Level Timeout)

Yes

300s

Maximum number of simultaneously initiated


Associations

No

Table 2-21 Configuration Parameters Table

34 of 118

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

Specialization

3.

Media Interchange
OBI/PV AI Client Application Entity allows importing and exporting DICOM Media Files. Various
import and export filters are available in the OBI/PV AI Client application to read and create
DICOM Media Files. This functionality is only available when the OBI/PV AI Application is
operating in file mode.
The files fully conform to the Part 10 format. However, the Basic Directory IOD as defined in
Media Interchange Application Profiles is not present.

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

35 of 118

Specialization

4.

Support of Character Sets


See Varian System Server DICOM Conformance Statement [3]

36 of 118

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

Specialization

5.

Security

5.1

Security Profiles
No Security Profiles are supported.

5.2

Association Level Security

5.2.1

OBI/PV AI Client
Storage SCU does not support Association Level Security.
The Storage SCP instantiated to receive Instances requested by Query/Retrieve SCU checks
the following additional values when determining whether to accept Association Open Requests:

Called AE Title

Calling AE Title

IP address of Association Request originator

5.3

Application Level Security

5.3.1

OBI/PV AI Application
In order to load patient information into the OBI/PV AI application, the user opens the patient in
the 4DITC application, which requires the user identification in the form of a user name and
password.

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

37 of 118

Specialization

Appendix A Specialization
A.1

IOD Contents

A.1.1

Created SOP Instances


IODs created by the OBI/PV AI Application Entity are listed in IOD Details.

A.1.2

Usage of Attributes from received IODs


The OBI/PV AI Application Entity require all Type 1 Attributes to be present.

A.2

Data Dictionary of Private Attributes


The following table contains a list of private Attributes that are used throughout all Application
Entities. The OBI/PV AI application does not use any private attributes that are not outlined in
this document or [3].

38 of 118

Tag

Name

VR

VM

(300B,ee01)

Distal Target Distance Tolerance

FL

(300B,ee02)

Maximum Collimated Field Diameter (Field Size)

FL

(300B,ee04)

Range (Planned Distal Target Distance)

FL

(300B,ee08)

Beam Line Data Table Version

SH

(300B,ee0E)

Nominal SOBP Width

FL

(300B,ee11)

Nominal SOBP Width Tolerance

FL

(300B,ee17)

Peek Range

FL

(3241,ee00)

Internal Image Id

LO

(3241,ee02)

Internal Pixel Offset

IS

(3241,ee03)

Not Calibrated

US

(3241,ee04)

Image Orientation (Patient)

DS

(3241,ee05)

Image Position (Patient)

DS

(3241,ee06)

Pixel Spacing

DS

(3243,ee09)

Beam Secondary Name

SH

(3243,ee28)

Related Referenced Beam Number

IS

(3247,ee00)

Session Termination Status

SH

(3249,ee00)

Maximum Treatment Time

DS

(3249,ee10)

Referenced Primary Dose Reference UID

UI

(3251,ee00)

Treatment Machine Note

LO

(3253,ee00)

Extended Interface Data

OB

(3253,ee01)

Extended Interface Length

IS

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

Specialization

Tag

Name

VR

VM

(3253,ee02)

Extended Interface Format

LO

(3257,ee01)

Plan Type
Enumerated Values:
ALL_PLANS
PREDECESSOR
NOT_ PREDECESSOR

CS

(3257,ee10)

Reference Image Type


Enumerated Values:
ALL
PRIMARY

CS

(3259,ee00)

Additional Dose Value Sequence

SQ

0..n

(3259,ee02)

Actual Session Dose

DS

(3259,ee04)

Daily Dose

DS

(3259,ee06)

Life Time Total Dose

DS

(3261,ee2A)

Treatment Session Beam Data

OB

(3263,ee00)

Overlap Type
Enumerated Values:
OVERLAPPING
NONOVERLAPPING

CS

(3263,ee01)

Referenced Structure Set Relationship Sequence

SQ

(3263,ee02)

Structure Set Relationship


Defined Terms:
PREDECESSOR
ADDITION

CS

(3265,ee00)

Total Number of Fractions

SL

(3265,ee01)

Last treated Fraction

SL

(3267,ee00)

Referenced Patient Volume Id

SH

(3271,ee00)

Marker Subtype
Enumerated Values:
MARKER
POINT_MATCH
REFERENCE_LINE

CS

(3273,ee00)

RT Image Isocenter Position

DS

(3273,ee01)

RT Image Patient Position

CS

(3275,ee00)

Registration Sub Type

LO

(3277,ee00)

RT Motion Synchronization Sequence

SQ

(3285,ee00)

Primary Fluence Mode Sequence

SQ

(3285,ee01)

Fluence Mode

CS

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

39 of 118

Specialization

Tag

Name

VR

VM

(3285,ee02)

Fluence Mode ID

SH

(3287,ee00)

Plan Integrity Sequence

SQ

(3287,ee01)

Plan Integrity Hash

LO

(3287,ee02)

Plan Integrity Hash Version

SH

Table A-1 Private Creator Identification Strings

A.3

Coded Terminology and Templates

A.3.1

Context Groups
Information will be published in a future version of this document.

A.3.2

Template Specifications
No standard templates are extended and no private templates are used.

A.3.3

Private Code Definitions


There are no private code definitions.

A.4

Grayscale Image Consistency


Not supported.

A.5

Standard Extended/Specialized/Private SOP


Classes
Not Used.

A.6

Private Transfer Syntaxes


No private Transfer Syntaxes are used.

40 of 118

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

IOD Details

Appendix B IOD Details


B.1

Supported IODs
For better reference with DICOM Standard ([1], section titles of the following sub-sections are
listed with section number of DICOM Standard Part 3 where the corresponding IOD Module table
can be found.

B.1.1

Computed Tomography Image A.3.3


IE

Reference

Usage

Patient

C.7.1.1

Clinical Trial Subject

C.7.1.3

General Study

C.7.2.1

Patient Study

C.7.2.2

Not supported

Clinical Trial Study

C.7.2.3

Not supported

General Series

C.7.3.1

Clinical Trial Series

C.7.3.2

Frame of Reference

Frame of Reference

C.7.4.1

Equipment

General Equipment

C.7.5.1

Image

General Image

C.7.6.1

Image Plane

C.7.6.2

Image Pixel

C.7.6.3

Contrast/Bolus

C.7.6.4

CT Image

C.8.2.1

Overlay Plane

C.9.2

VOI LUT

C.11.2

SOP Common

C.12.1

Patient
Study

Series

Module

Presence

Not supported

Not supported

Not supported
Not supported

Table B-2 Computed Tomography Image IOD Modules Support

B.1.2

Spatial Registration A.39.1.3


IE
Patient
Study

Module

Reference

Usage

Patient

C.7.1.1

Clinical Trial Subject

C.7.1.3

General Study

C.7.2.1

Patient Study

C.7.2.2

Not supported

Clinical Trial Study

C.7.2.3

Not supported

OBI and PortalVision Advanced Imaging

Presence

Not supported

DICOM Conformance Statement, KC1505D3CS

41 of 118

IOD Details

IE

Reference

Usage

General Series

C.7.3.1

Clinical Trial Series

C.7.3.2

Spatial Registration
Series

C.20.1

Frame of Reference

Frame of Reference

C.7.4.1

Equipment

General Equipment

C.7.5.1

Spatial Registration

Spatial Registration

C.20.2

Common Instance
Reference

C.12.2

SOP Common

C.12.1

Series

Module

Presence

Not supported

Table B-3 Spatial Registration Image IOD Modules Support

B.1.3

RT Structure Set A.19.3


IE

Reference

Usage

Patient

C.7.1.1

Clinical Trial Subject

C.7.1.3

General Study

C.7.2.1

Patient Study

C.7.2.2

Not supported

Clinical Trial Study

C.7.2.3

Not supported

RT Series

C.8.8.1

Clinical Trial Series

C.7.3.2

Equipment

General Equipment

C.7.5.1

Structure Set

Structure Set

C.8.8.5

ROI Contour

C.8.8.6

RT ROI Observations

C.8.8.8

Approval

C.8.8.16

C.12.1

Reference

Usage

Patient

C.7.1.1

Clinical Trial Subject

C.7.1.3

General Study

C.7.2.1

Patient
Study

Series

Module

SOP Common

Presence

Not supported

Not supported

Not supported

Table B-4 RT Structure Set IOD Modules Support

B.1.4

RT Image A.17.3
IE
Patient
Study

42 of 118

Module

OBI and PortalVision Advanced Imaging

Presence

Not supported

DICOM Conformance Statement, KC1505D3CS

IOD Details

IE

Reference

Usage

Patient Study

C.7.2.2

Not supported

Clinical Trial Study

C.7.2.3

Not supported

RT Series

C.8.8.1

Clinical Trial Series

C.7.3.2

Not supported

Frame of Reference

Frame of Reference

C.7.4.1

Supported and
required for
Import

Equipment

General Equipment

C.7.5.1

Image

General Image

C.7.6.1

Image Pixel

C.7.6.3

Contrast/Bolus

C.7.6.4

Not supported

Cine

C.7.6.5

Not supported

Multi Frame

C.7.6.6

Not supported

RT Image

C.8.8.2

Modality LUT

C.11.1

VOI LUT

C.11.2

Approval

C.8.8.16

Curve (Retired)

C.10.2

SOP Common

C.12.1

N/A

Series

Extended Interface

Module

Extended Interface

Presence

Private

Table B-5 RT Image IOD Modules Support

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

43 of 118

IOD Details

(This page is intentionally left blank.)

44 of 118

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

IOD Details, Patient C.7.1.1

B.2

Modules and Attributes


For better reference with DICOM Standard ([1]section titles of the following sub-sections are decorated with section number of DICOM Standard
Part 3 where the corresponding Module Attribute table can be found.
The first four columns in the following tables contain definitions from the DICOM standard. References in those columns refer to the DICOM
standard and not to this document.
The Handling column describes for each attribute whether it is supported and to which value in the OBI system it maps. A value of *** export not
supported indicates that an attribute is unknown and therefore ignored by the application.
The application creates objects of the type RT Image, CT Image, RT Structure Set and Spatial Registration. These objects are exported through
DICOM. Therefore only the export is described in the following tables.

B.2.1

Patient C.7.1.1
Attribute Name

Tag

Description

Handling

Patients Name

(0010,0010)

Patient's full name.

Export supported for:


All

Patient ID

(0010,0020)

Primary hospital identification number or code for the


patient.

Export supported for:


All

Issuer of Patient ID

(0010,0021)

Identifier of the Assigning Authority that issued the


Patient ID.

*** export not supported

Patient's Birth Date

(0010,0030)

Birth date of the patient.

Export supported for:


All

Patient's Sex

(0010,0040)

Sex of the named patient.


Enumerated Values:
M = male
F = female
O = other

Export supported for:


All
Female: F, Male: M, anything else: O

Referenced Patient
Sequence

(0008,1120)

A sequence that provides reference to a Patient SOP


Class/Instance pair. Only a single Item shall be permitted
in this Sequence.

*** export not supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

45 of 118

IOD Details, Patient C.7.1.1

46 of 118

Attribute Name

Tag

Description

Handling

> Referenced SOP


Class UID

(0008,1150)

1C

Uniquely identifies the referenced SOP Class. Required if


Referenced Patient Sequence (0008,1120) is sent.

*** export not supported

> Referenced SOP


Instance UID

(0008,1155)

1C

Uniquely identifies the referenced SOP Instance.


Required if Referenced Patient Sequence (0008,1120) is
sent.

*** export not supported

Patient's Birth Time

(0010,0032)

Birth time of the Patient.

Export supported for:


All

Other Patient IDs

(0010,1000)

Other identification numbers or codes used to identify the


patient.

*** export not supported

Other Patient
Names

(0010,1001)

Other names used to identify the patient.

*** export not supported

Ethnic Group

(0010,2160)

Ethnic group or race of the patient.

*** export not supported

Patient Comments

(0010,4000)

User-defined additional information about the patient.

*** export not supported

Patient Identify
Removed

(0012,0062)

The true identity of the patient has been removed from


the Attributes and the Pixel Data
Enumerated Values:
YES
NO

*** export not supported

De-identification
Method

(0012,0063)

1C

A description or label of the mechanism or method use to


remove the patient's identity. May be multi-valued if
successive de-identification steps have been performed.
Note:
This may be used to describe the extent or thoroughness
of the de- identification, for example whether or not the
de-identification is for a "Limited Data Set" (as per HIPAA
Privacy Rule).
Required if Patient Identity Removed (0012,0062) is
present and has a value of YES and De-identification
Method Code Sequence (0012,0064) is not present.

*** export not supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

IOD Details, General Study - C.7.2.1

B.2.2

Attribute Name

Tag

Description

Handling

De-identification
Method Code
Sequence

(0012,0064)

1C

A code describing the mechanism or method use to


remove the patient's identity. One or more Items shall be
present. Multiple items are used if successive deidentification steps have been performed
Required if Patient Identity Removed (0012,0062) is
present and has a value of YES and De-identification
Method (0012,0063) is not present.

*** export not supported

> Code Value

(0008,0100)

1C

See Section 8.1. Required if a sequence item is present.

*** export not supported

> Coding Scheme


Designator

(0008,0102)

1C

See Section 8.2. Required if a sequence item is present.

*** export not supported

> Coding Scheme


Version

(0008,0103)

1C

See Section 8.2. Required if a sequence item is present


and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.

*** export not supported

> Code Meaning

(0008,0104)

1C

See Section 8.3. Required if a sequence item is present.

*** export not supported

General Study - C.7.2.1


Attribute Name

Tag

Description

Handling

Study Instance UID

(0020,000D)

Unique identifier for the Study.

Export supported for:


All

Study Date

(0008,0020)

Date the Study started.

Export supported for:


All

Study Time

(0008,0030)

Time the Study started.

Export supported for:


All

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

47 of 118

IOD Details, General Study - C.7.2.1

48 of 118

Attribute Name

Tag

Description

Handling

Referring
Physicians Name

(0008,0090)

Name of the patient's referring physician

*** export not supported

Referring Physician
Identification
Sequence

(0008,0096)

Identification of the patient's referring physician. Only a


single item shall be permitted in this sequence.

*** export not supported

> Person
Identification Code
Sequence

(0040,1101)

A coded entry which identifies a person. The Code


Meaning attribute, though it will be encoded with a VR of
LO, may be encoded according to the rules of the PN VR
(e.g. caret '^' delimiters shall separate name
components), except that a single component (i.e. the
whole name unseparated by caret delimiters) is not
permitted. Name component groups for use with multibyte character sets are permitted, as long as they fit
within the 64 characters (the length of the LO VR). One
or more Items may be permitted in this Sequence.

*** export not supported

>> Code Value

(0008,0100)

1C

See Section 8.1. Required if a sequence item is present.

*** export not supported

>> Coding Scheme


Designator

(0008,0102)

1C

See Section 8.2. Required if a sequence item is present.

*** export not supported

>> Coding Scheme


Version

(0008,0103)

1C

See Section 8.2. Required if a sequence item is present


and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.

*** export not supported

>> Code Meaning

(0008,0104)

1C

See Section 8.3. Required if a sequence item is present.

*** export not supported

> Persons Address

(0040,1102)

Person's mailing address

*** export not supported

> Persons
Telephone Numbers

(0040,1103)

Person's telephone number(s)

*** export not supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

IOD Details, General Study - C.7.2.1

Attribute Name

Tag

Description

Handling

> Institution Name

(0008,0080)

1C

Institution or organization to which the identified individual


is responsible or accountable. Shall not be present if
Institution Code Sequence (0008,0082) is present.

*** export not supported

> Institution Address

(0008,0081)

Mailing address of the institution or organization to which


the identified individual is responsible or accountable.

*** export not supported

> Institution Code


Sequence

(0008,0082)

1C

Institution or organization to which the identified individual


is responsible or accountable. Shall not be present if
Institution Name (0008,0080) is present. Only a single
Item shall be permitted in this Sequence.

*** export not supported

>> Code Value

(0008,0100)

1C

See Section 8.1. Required if a sequence item is present.

*** export not supported

>> Coding Scheme


Designator

(0008,0102)

1C

See Section 8.2. Required if a sequence item is present.

*** export not supported

>> Coding Scheme


Version

(0008,0103)

1C

See Section 8.2. Required if a sequence item is present


and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.

*** export not supported

>> Code Meaning

(0008,0104)

1C

See Section 8.3. Required if a sequence item is present.

*** export not supported

Study ID

(0020,0010)

User or equipment generated Study identifier.

Export supported for:


All

Accession Number

(0008,0050)

A RIS generated number that identifies the


order for the Study.

Export supported for:


All

Study Description

(0008,1030)

Institution-generated description or classification of the


Study (component) performed.

*** export not supported

Physician(s) of
Record

(0008,1048)

Names of the physician(s) who are responsible for overall


patient care at time of Study (see Section C.7.3.1 for
Performing Physician)

*** export not supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

49 of 118

IOD Details, General Study - C.7.2.1

50 of 118

Attribute Name

Tag

Description

Handling

Physician(s) of
Record Identification
Sequence

(0008,1049)

Identification of the physician(s) who are responsible for


overall patient care at time of Study. One or more items
shall be included in this sequence. If more than one Item,
the number and order shall correspond to the value of
Physician(s) of Record (0008,1048), if present.

*** export not supported

> Person
Identification Code
Sequence

(0040,1101)

A coded entry which identifies a person. The Code


Meaning attribute, though it will be encoded with a VR of
LO, may be encoded according to the rules of the PN VR
(e.g. caret '^' delimiters shall separate name
components), except that a single component (i.e. the
whole name unseparated by caret delimiters) is not
permitted. Name component groups for use with multibyte character sets are permitted, as long as they fit
within the 64 characters (the length of the LO VR). One
or more Items may be permitted in this Sequence.

*** export not supported

>> Code Value

(0008,0100)

1C

See Section 8.1. Required if a sequence item is present.

*** export not supported

>> Coding Scheme


Designator

(0008,0102)

1C

See Section 8.2. Required if a sequence item is present.

*** export not supported

>> Coding Scheme


Version

(0008,0103)

1C

See Section 8.2. Required if a sequence item is present


and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.

*** export not supported

>> Code Meaning

(0008,0104)

1C

See Section 8.3. Required if a sequence item is present.

*** export not supported

> Persons Address

(0040,1102)

Person's mailing address

*** export not supported

> Persons
Telephone Numbers

(0040,1103)

Person's telephone number(s)

*** export not supported

> Institution Name

(0008,0080)

1C

Institution or organization to which the identified individual


is responsible or accountable. Shall not be present if
Institution Code Sequence (0008,0082) is present.

*** export not supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

IOD Details, General Study - C.7.2.1

Attribute Name

Tag

Description

Handling

> Institution Address

(0008,0081)

Mailing address of the institution or organization to which


the identified individual is responsible or accountable.

*** export not supported

> Institution Code


Sequence

(0008,0082)

1C

Institution or organization to which the identified individual


is responsible or accountable. Shall not be present if
Institution Name (0008,0080) is present. Only a single
Item shall be permitted in this Sequence.

*** export not supported

>> Code Value

(0008,0100)

1C

See Section 8.1. Required if a sequence item is present.

*** export not supported

>> Coding Scheme


Designator

(0008,0102)

1C

See Section 8.2. Required if a sequence item is present.

*** export not supported

>> Coding Scheme


Version

(0008,0103)

1C

See Section 8.2. Required if a sequence item is present


and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.

*** export not supported

>> Code Meaning

(0008,0104)

1C

See Section 8.3. Required if a sequence item is present.

*** export not supported

Name of
Physician(s)
Reading Study

(0008,1060)

Names of the physician(s) reading the Study.

*** export not supported

Physician(s)
Reading Study
Identification
Sequence

(0008,1062)

Identification of the physician(s) reading the Study. One


or more items shall be included in this sequence. If more
than one Item, the number and order shall correspond to
the value of Name of Physician(s) Reading Study
(0008,1060), if present.

*** export not supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

51 of 118

IOD Details, General Study - C.7.2.1

52 of 118

Attribute Name

Tag

Description

Handling

> Person
Identification Code
Sequence

(0040,1101)

A coded entry which identifies a person. The Code


Meaning attribute, though it will be encoded with a VR of
LO, may be encoded according to the rules of the PN VR
(e.g. caret '^' delimiters shall separate name
components), except that a single component (i.e. the
whole name unseparated by caret delimiters) is not
permitted. Name component groups for use with multibyte character sets are permitted, as long as they fit
within the 64 characters (the length of the LO VR). One
or more Items may be permitted in this Sequence.

*** export not supported

>> Code Value

(0008,0100)

1C

See Section 8.1. Required if a sequence item is present.

*** export not supported

>> Coding Scheme


Designator

(0008,0102)

1C

See Section 8.2. Required if a sequence item is present.

*** export not supported

>> Coding Scheme


Version

(0008,0103)

1C

See Section 8.2. Required if a sequence item is present


and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.

*** export not supported

>> Code Meaning

(0008,0104)

1C

See Section 8.3. Required if a sequence item is present.

*** export not supported

> Persons Address

(0040,1102)

Person's mailing address

*** export not supported

> Persons
Telephone Numbers

(0040,1103)

Person's telephone number(s)

*** export not supported

> Institution Name

(0008,0080)

1C

Institution or organization to which the identified individual


is responsible or accountable. Shall not be present if
Institution Code Sequence (0008,0082) is present.

*** export not supported

> Institution Address

(0008,0081)

Mailing address of the institution or organization to which


the identified individual is responsible or accountable.

*** export not supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

IOD Details, General Study - C.7.2.1

Attribute Name

Tag

Description

Handling

> Institution Code


Sequence

(0008,0082)

1C

Institution or organization to which the identified individual


is responsible or accountable. Shall not be present if
Institution Name (0008,0080) is present. Only a single
Item shall be permitted in this Sequence.

*** export not supported

>> Code Value

(0008,0100)

1C

See Section 8.1. Required if a sequence item is present.

*** export not supported

>> Coding Scheme


Designator

(0008,0102)

1C

See Section 8.2. Required if a sequence item is present.

*** export not supported

>> Coding Scheme


Version

(0008,0103)

1C

See Section 8.2. Required if a sequence item is present


and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.

*** export not supported

>> Code Meaning

(0008,0104)

1C

See Section 8.3. Required if a sequence item is present.

*** export not supported

Referenced Study
Sequence

(0008,1110)

A sequence that provides reference to a Study SOP


Class/Instance pair. The sequence may have zero or
more Items.

*** export not supported

> Referenced SOP


Class UID

(0008,1150)

1C

Uniquely identifies the referenced SOP Class. Required if


Referenced Study Sequence (0008,1110) is sent.

*** export not supported

> Referenced SOP


Instance UID

(0008,1155)

1C

Uniquely identifies the referenced SOP Instance.


Required if Referenced Study Sequence (0008,1110) is
sent.

*** export not supported

Procedure Code
Sequence

(0008,1032)

A Sequence that conveys the type of procedure


performed. One or more Items may be included in this
Sequence.

*** export not supported

> Code Value

(0008,0100)

1C

See Section 8.1. Required if a sequence item is present.

*** export not supported

> Coding Scheme


Designator

(0008,0102)

1C

See Section 8.2. Required if a sequence item is present.

*** export not supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

53 of 118

IOD Details, General Series C.7.3.1

B.2.3

54 of 118

Attribute Name

Tag

Description

Handling

> Coding Scheme


Version

(0008,0103)

1C

See Section 8.2. Required if a sequence item is present


and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.

*** export not supported

> Code Meaning

(0008,0104)

1C

See Section 8.3. Required if a sequence item is present.

*** export not supported

General Series C.7.3.1


Attribute Name

Tag

Description

Handling

Modality

(0008,0060)

Type of equipment that originally acquired the data used


to create the images in this Series. See C.7.3.1.1.1 for
Defined Terms.

Export supported for:


All

Series Instance UID

(0020,000E)

Unique identifier of the Series.

Export supported for:


All

Series Number

(0020,0011)

A number that identifies this Series.

Export supported for:


All
The value is set to 0 for all objects

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

IOD Details, General Series C.7.3.1

Attribute Name

Tag

Description

Handling

Laterality

(0020,0060)

2C

Laterality of (paired) body part examined.


Required if the body part examined is a paired structure
and Image Laterality (0020,0062) or Frame Laterality
(0020,9072) are not sent.
Enumerated Values: R = right
L = left
Note:
Some IODs support Image Laterality (0020,0062) at the
Image level or Frame Laterality(0020,9072) at the Frame
level in the Frame Anatomy functional group macro,
which can provide a more comprehensive mechanism for
specifying the laterality of the body part(s) being
examined.

*** export not supported

Series Date

(0008,0021)

Date the Series started.

Export supported for:


Computed Tomography Image
RT Structure Set
Spatial Registration

Series Time

(0008,0031)

Time the Series started.

Export supported for:


Computed Tomography Image
RT Structure Set
Spatial Registration

Performing
Physicians Name

(0008,1050)

Name of the physician(s) administering the Series.

*** export not supported

Performing
Physician
Identification
Sequence

(0008,1052)

Identification of the physician(s) administering the Series.


One or more items shall be included in this sequence. If
more than one Item, the number and order shall
correspond to the value of Performing Physicians' Name
(0008,1050), if present.

*** export not supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

55 of 118

IOD Details, General Series C.7.3.1

56 of 118

Attribute Name

Tag

Description

Handling

> Person
Identification Code
Sequence

(0040,1101)

A coded entry which identifies a person. The Code


Meaning attribute, though it will be encoded with a VR of
LO, may be encoded according to the rules of the PN VR
(e.g. caret '^' delimiters shall separate name
components), except that a single component (i.e. the
whole name unseparated by caret delimiters) is not
permitted. Name component groups for use with multibyte character sets are permitted, as long as they fit
within the 64 characters (the length of the LO VR). One
or more Items may be permitted in this Sequence.

*** export not supported

>> Code Value

(0008,0100)

1C

See Section 8.1. Required if a sequence item is present.

*** export not supported

>> Coding Scheme


Designator

(0008,0102)

1C

See Section 8.2. Required if a sequence item is present.

*** export not supported

>> Coding Scheme


Version

(0008,0103)

1C

See Section 8.2. Required if a sequence item is present


and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.

*** export not supported

>> Code Meaning

(0008,0104)

1C

See Section 8.3. Required if a sequence item is present.

*** export not supported

> Persons Address

(0040,1102)

Person's mailing address

*** export not supported

> Persons
Telephone Numbers

(0040,1103)

Person's telephone number(s)

*** export not supported

> Institution Name

(0008,0080)

1C

Institution or organization to which the identified individual


is responsible or accountable. Shall not be present if
Institution Code Sequence (0008,0082) is present.

*** export not supported

> Institution Address

(0008,0081)

Mailing address of the institution or organization to which


the identified individual is responsible or accountable.

*** export not supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

IOD Details, General Series C.7.3.1

Attribute Name

Tag

Description

Handling

> Institution Code


Sequence

(0008,0082)

1C

Institution or organization to which the identified individual


is responsible or accountable. Shall not be present if
Institution Name (0008,0080) is present. Only a single
Item shall be permitted in this Sequence.

*** export not supported

>> Code Value

(0008,0100)

1C

See Section 8.1. Required if a sequence item is present.

*** export not supported

>> Coding Scheme


Designator

(0008,0102)

1C

See Section 8.2. Required if a sequence item is present.

*** export not supported

>> Coding Scheme


Version

(0008,0103)

1C

See Section 8.2. Required if a sequence item is present


and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.

*** export not supported

>> Code Meaning

(0008,0104)

1C

See Section 8.3. Required if a sequence item is present.

*** export not supported

Protocol Name

(0018,1030)

User-defined description of the conditions under which


the Series was performed.
Note:
This attribute conveys series-specific protocol
identification and may or may not be identical to the one
presented in the Performed Protocol Code Sequence
(0040,0260).

*** export not supported

Series Description

(0008,103E)

User provided description of the Series

*** export not supported

Operators Name

(0008,1070)

Name(s) of the operator(s) supporting the Series.

*** export not supported

Operator
Identification
Sequence

(0008,1072)

Identification of the operator(s) supporting the Series.


One or more items shall be included in this sequence. If
more than one Item, the number and order shall
correspond to the value of Operators' Name (0008,1070),
if present.

*** export not supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

57 of 118

IOD Details, General Series C.7.3.1

58 of 118

Attribute Name

Tag

Description

Handling

> Person
Identification Code
Sequence

(0040,1101)

A coded entry which identifies a person. The Code


Meaning attribute, though it will be encoded with a VR of
LO, may be encoded according to the rules of the PN VR
(e.g. caret '^' delimiters shall separate name
components), except that a single component (i.e. the
whole name unseparated by caret delimiters) is not
permitted. Name component groups for use with multibyte character sets are permitted, as long as they fit
within the 64 characters (the length of the LO VR). One
or more Items may be permitted in this Sequence.

*** export not supported

>> Code Value

(0008,0100)

1C

See Section 8.1. Required if a sequence item is present.

*** export not supported

>> Coding Scheme


Designator

(0008,0102)

1C

See Section 8.2. Required if a sequence item is present.

*** export not supported

>> Coding Scheme


Version

(0008,0103)

1C

See Section 8.2. Required if a sequence item is present


and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.

*** export not supported

>> Code Meaning

(0008,0104)

1C

See Section 8.3. Required if a sequence item is present.

*** export not supported

> Persons Address

(0040,1102)

Person's mailing address

*** export not supported

> Persons
Telephone Numbers

(0040,1103)

Person's telephone number(s)

*** export not supported

> Institution Name

(0008,0080)

1C

Institution or organization to which the identified individual


is responsible or accountable. Shall not be present if
Institution Code Sequence (0008,0082) is present.

*** export not supported

> Institution Address

(0008,0081)

Mailing address of the institution or organization to which


the identified individual is responsible or accountable.

*** export not supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

IOD Details, General Series C.7.3.1

Attribute Name

Tag

Description

Handling

> Institution Code


Sequence

(0008,0082)

1C

Institution or organization to which the identified individual


is responsible or accountable. Shall not be present if
Institution Name (0008,0080) is present. Only a single
Item shall be permitted in this Sequence.

*** export not supported

>> Code Value

(0008,0100)

1C

See Section 8.1. Required if a sequence item is present.

*** export not supported

>> Coding Scheme


Designator

(0008,0102)

1C

See Section 8.2. Required if a sequence item is present.

*** export not supported

>> Coding Scheme


Version

(0008,0103)

1C

See Section 8.2. Required if a sequence item is present


and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.

*** export not supported

>> Code Meaning

(0008,0104)

1C

See Section 8.3. Required if a sequence item is present.

*** export not supported

Referenced
Performed
Procedure Step
Sequence

(0008,1111)

Uniquely identifies the Performed Procedure Step SOP


Instance to which the Series is related (e.g. a Modality or
General-Purpose Performed Procedure Step SOP
Instance). The Sequence shall have zero or one Item.

*** export not supported

> Referenced SOP


Class UID

(0008,1150)

1C

Uniquely identifies the referenced SOP Class.


Required if Referenced Performed Procedure Step
Sequence (0008,1111) is sent.

*** export not supported

> Referenced SOP


Instance UID

(0008,1155)

1C

Uniquely identifies the referenced SOP Instance.


Required if Referenced Performed Procedure Step
Sequence (0008,1111) is sent.

*** export not supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

59 of 118

IOD Details, General Series C.7.3.1

60 of 118

Attribute Name

Tag

Description

Handling

Related Series
Sequence

(0008,1250)

Identification of Series significantly related to this Series.


Zero or more Items may be present.
Notes:
1. For example, for a combined CT and PET acquisition,
the CT images and PET images would be in separate
series that could cross- reference each other with
multiple purpose of reference codes meaning same
anatomy, simultaneously acquired and same indication.
2. The related series may have different Frames of
Reference and hence require some sort of registration
before spatial coordinates can be directly compared.
3. This attribute is not intended for conveying localizer
reference information, for which Referenced Image
Sequence (0008,1140) should be used.

*** export not supported

> Study Instance


UID

(0020,000D)

Instance UID of Study to which the related Series


belongs

*** export not supported

> Series Instance


UID

(0020,000E)

Instance UID of Related Series

*** export not supported

> Purpose of
Reference Code
Sequence

(0040,A170)

Describes the purpose for which the reference is made.


Zero or more Items may be present. When absent,
implies that the reason for the reference is unknown.

*** export not supported

>> Code Value

(0008,0100)

1C

See Section 8.1. Required if a sequence item is present.

*** export not supported

>> Coding Scheme


Designator

(0008,0102)

1C

See Section 8.2. Required if a sequence item is present.

*** export not supported

>> Coding Scheme


Version

(0008,0103)

1C

See Section 8.2. Required if a sequence item is present


and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.

*** export not supported

>> Code Meaning

(0008,0104)

1C

See Section 8.3. Required if a sequence item is present.

*** export not supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

IOD Details, General Series C.7.3.1

Attribute Name

Tag

Description

Handling

Body Part Examined

(0018,0015)

Text description of the part of the body examined.


Defined Terms:
SKULL, CSPINE, TSPINE, LSPINE, SSPINE, COCCYX,
CHEST, CLAVICLE, BREAST, ABDOMEN, PELVIS, HIP,
SHOULDER, ELBOW, KNEE, ANKLE, HAND, FOOT,
EXTREMITY, HEAD, HEART, NECK, LEG, ARM, JAW
Note:
Some IODs support the Anatomic Region Sequence
(0008,2218), which can provide a more comprehensive
mechanism for specifying the body part being examined.

*** export not supported

Patient Position

(0018,5100)

2C

Patient position descriptor relative to the equipment.


Required for CT and MR images; shall not be present if
Patient Orientation Code Sequence (0054,0410) is
present; may be present otherwise. See C.7.3.1.1.2 for
Defined Terms and further explanation.

Export supported for:


RT Image
Computed Tomography Image

Smallest Pixel Value


in Series

(0028,0108)

The minimum value of all images in this Series.

*** export not supported

Largest Pixel Value


in Series

(0028,0109)

The maximum value of all images in this Series.

*** export not supported

Request Attributes
Sequence

(0040,0275)

Sequence that contains attributes from the Imaging


Service Request. The sequence may have one or more
Items.

*** export not supported

> Requested
Procedure ID

(0040,1001)

Identifier that identifies the Requested Procedure in the


Imaging Service Request.

*** export not supported

> Reason for the


Requested
Procedure

(0040,1002)

Reason for requesting this procedure.

*** export not supported

> Reason for


Requested
Procedure Code
Sequence

(0040,100A)

Coded Reason for requesting this procedure.

*** export not supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

61 of 118

IOD Details, General Series C.7.3.1

62 of 118

Attribute Name

Tag

Description

Handling

>> Code Value

(0008,0100)

1C

See Section 8.1. Required if a sequence item is present.

*** export not supported

>> Coding Scheme


Designator

(0008,0102)

1C

See Section 8.2. Required if a sequence item is present.

*** export not supported

>> Coding Scheme


Version

(0008,0103)

1C

See Section 8.2. Required if a sequence item is present


and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.

*** export not supported

>> Code Meaning

(0008,0104)

1C

See Section 8.3. Required if a sequence item is present.

*** export not supported

> Scheduled
Procedure Step ID

(0040,0009)

Identifier that identifies the Scheduled Procedure Step.

*** export not supported

> Scheduled
Procedure Step
Description

(0040,0007)

Institution-generated description or classification of the


Scheduled Procedure Step to be performed.

*** export not supported

> Scheduled
Protocol Code
Sequence

(0040,0008)

Sequence describing the Scheduled Protocol following a


specific coding scheme. This sequence contains one or
more Items.

*** export not supported

>> Code Value

(0008,0100)

1C

See Section 8.1. Required if a sequence item is present.

*** export not supported

>> Coding Scheme


Designator

(0008,0102)

1C

See Section 8.2. Required if a sequence item is present.

*** export not supported

>> Coding Scheme


Version

(0008,0103)

1C

See Section 8.2. Required if a sequence item is present


and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.

*** export not supported

>> Code Meaning

(0008,0104)

1C

See Section 8.3. Required if a sequence item is present.

*** export not supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

IOD Details, General Series C.7.3.1

Attribute Name

Tag

Description

Handling

>> Protocol Context


Sequence

(0040,0440)

Sequence that specifies the context for the Scheduled


Protocol Code Sequence Item. One or more items may
be included in this sequence.

*** export not supported

>>> Code Value

(0008,0100)

1C

See Section 8.1. Required if a sequence item is present.

*** export not supported

>>> Coding
Scheme Designator

(0008,0102)

1C

See Section 8.2. Required if a sequence item is present.

*** export not supported

>>> Coding
Scheme Version

(0008,0103)

1C

See Section 8.2. Required if a sequence item is present


and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.

*** export not supported

>>> Code Meaning

(0008,0104)

1C

See Section 8.3. Required if a sequence item is present.

*** export not supported

>>> Content Item


Modifier Sequence

(0040,0441)

Sequence that specifies modifiers for a Protocol Context


Content Item. One or more items may be included in this
sequence. See Section C.4.10.1.

*** export not supported

>>>> Code Value

(0008,0100)

1C

See Section 8.1. Required if a sequence item is present.

*** export not supported

>>>> Coding
Scheme Designator

(0008,0102)

1C

See Section 8.2. Required if a sequence item is present.

*** export not supported

>>>> Coding
Scheme Version

(0008,0103)

1C

See Section 8.2. Required if a sequence item is present


and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.

*** export not supported

>>>> Code Meaning

(0008,0104)

1C

See Section 8.3. Required if a sequence item is present.

*** export not supported

Performed
Procedure Step ID

(0040,0253)

User or equipment generated identifier of that part of a


Procedure that has been carried out within this step.

*** export not supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

63 of 118

IOD Details, General Series C.7.3.1

64 of 118

Attribute Name

Tag

Description

Handling

Performed
Procedure Step
Start Date

(0040,0244)

Date on which the Performed Procedure Step started.

*** export not supported

Performed
Procedure Step
Start Time

(0040,0245)

Time on which the Performed Procedure Step started.

*** export not supported

Performed
Procedure Step
Description

(0040,0254)

Institution-generated description or classification of the


Procedure Step that was performed.

*** export not supported

Performed Protocol
Code Sequence

(0040,0260)

Sequence describing the Protocol performed for this


Procedure Step. One or more Items may be included in
this Sequence.

*** export not supported

> Code Value

(0008,0100)

1C

See Section 8.1. Required if a sequence item is present.

*** export not supported

> Coding Scheme


Designator

(0008,0102)

1C

See Section 8.2. Required if a sequence item is present.

*** export not supported

> Coding Scheme


Version

(0008,0103)

1C

See Section 8.2. Required if a sequence item is present


and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.

*** export not supported

> Code Meaning

(0008,0104)

1C

See Section 8.3. Required if a sequence item is present.

*** export not supported

> Protocol Context


Sequence

(0040,0440)

Sequence that specifies the context for the Performed


Protocol Code Sequence Item. One or more items may
be included in this sequence.

*** export not supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

IOD Details, General Series C.7.3.1

Attribute Name

Tag

Description

Handling

>> Value Type

(0040,A040)

The type of the value encoded in this name-value Item.


Defined Terms:
DATETIME
DATE
TIME
PNAME
UIDREF
TEXT
CODE
NUMERIC.

*** export not supported

>> Concept Name


Code Sequence

(0040,A043)

Coded concept name of this name-value Item.

*** export not supported

>>> Code Value

(0008,0100)

1C

See Section 8.1. Required if a sequence item is present.

*** export not supported

>>> Coding
Scheme Designator

(0008,0102)

1C

See Section 8.2. Required if a sequence item is present.

*** export not supported

>>> Coding
Scheme Version

(0008,0103)

1C

See Section 8.2. Required if a sequence item is present


and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.

*** export not supported

>>> Code Meaning

(0008,0104)

1C

See Section 8.3. Required if a sequence item is present.

*** export not supported

>> DateTime

(0040,A120)

1C

Datetime value for this name-value Item. Required if


Value Type (0040,A040) is DATETIME.

*** export not supported

>> Date

(0040,A121)

1C

Date value for this name-value Item. Required if Value


Type (0040,A040) is DATE.

*** export not supported

>> Time

(0040,A122)

1C

Time value for this name-value Item. Required if Value


Type (0040,A040) is TIME.

*** export not supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

65 of 118

IOD Details, General Series C.7.3.1

66 of 118

Attribute Name

Tag

Description

Handling

>> Person Name

(0040,A123)

1C

Person name value for this name-value Item. Required if


Value Type (0040,A040) is PNAME.

*** export not supported

>> UID

(0040,A124)

1C

UID value for this name-value Item. Required if Value


Type (0040,A040) is UIDREF.

*** export not supported

>> Text Value

(0040,A160)

1C

Text value for this name-value Item. Required if Value


Type (0040,A040) is TEXT.

*** export not supported

>> Concept Code


Sequence

(0040,A168)

1C

Coded concept value of this name-value Item. Required if


Value Type (0040,A040) is CODE.

*** export not supported

>>> Code Value

(0008,0100)

1C

See Section 8.1. Required if a sequence item is present.

*** export not supported

>>> Coding
Scheme Designator

(0008,0102)

1C

See Section 8.2. Required if a sequence item is present.

*** export not supported

>>> Coding
Scheme Version

(0008,0103)

1C

See Section 8.2. Required if a sequence item is present


and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.

*** export not supported

>>> Code Meaning

(0008,0104)

1C

See Section 8.3. Required if a sequence item is present.

*** export not supported

>> Numeric Value

(0040,A30A)

1C

Numeric value for this name-value Item. Required if


Value Type (0040,A040) is NUMERIC.

*** export not supported

>> Measurement
Units Code
Sequence

(0040,08EA)

1C

Units of measurement for a numeric value in this


namevalue Item. Required if Value Type (0040,A040) is
NUMERIC.

*** export not supported

>>> Code Value

(0008,0100)

1C

See Section 8.1. Required if a sequence item is present.

*** export not supported

>>> Coding
Scheme Designator

(0008,0102)

1C

See Section 8.2. Required if a sequence item is present.

*** export not supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

IOD Details, General Series C.7.3.1

Attribute Name

Tag

Description

Handling

>>> Coding
Scheme Version

(0008,0103)

1C

See Section 8.2. Required if a sequence item is present


and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.

*** export not supported

>>> Code Meaning

(0008,0104)

1C

See Section 8.3. Required if a sequence item is present.

*** export not supported

>> Content Item


Modifier Sequence

(0040,0441)

Sequence that specifies modifiers for a Protocol Context


Content Item. One or more items may be included in this
sequence. See Section C.4.10.1.

*** export not supported

>>> Value Type

(0040,A040)

The type of the value encoded in this name-value Item.


Defined Terms:
DATETIME
DATE
TIME
PNAME
UIDREF
TEXT
CODE
NUMERIC.

*** export not supported

>>> Concept Name


Code Sequence

(0040,A043)

Coded concept name of this name-value Item.

*** export not supported

>>>> Code Value

(0008,0100)

1C

See Section 8.1. Required if a sequence item is present.

*** export not supported

>>>> Coding
Scheme Designator

(0008,0102)

1C

See Section 8.2. Required if a sequence item is present.

*** export not supported

>>>> Coding
Scheme Version

(0008,0103)

1C

See Section 8.2. Required if a sequence item is present


and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.

*** export not supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

67 of 118

IOD Details, General Series C.7.3.1

68 of 118

Attribute Name

Tag

Description

Handling

>>>> Code Meaning

(0008,0104)

1C

See Section 8.3. Required if a sequence item is present.

*** export not supported

>>> DateTime

(0040,A120)

1C

Datetime value for this name-value Item. Required if


Value Type (0040,A040) is DATETIME.

*** export not supported

>>> Date

(0040,A121)

1C

Date value for this name-value Item. Required if Value


Type (0040,A040) is DATE.

*** export not supported

>>> Time

(0040,A122)

1C

Time value for this name-value Item. Required if Value


Type (0040,A040) is TIME.

*** export not supported

>>> Person Name

(0040,A123)

1C

Person name value for this name-value Item. Required if


Value Type (0040,A040) is PNAME.

*** export not supported

>>> UID

(0040,A124)

1C

UID value for this name-value Item. Required if Value


Type (0040,A040) is UIDREF.

*** export not supported

>>> Text Value

(0040,A160)

1C

Text value for this name-value Item. Required if Value


Type (0040,A040) is TEXT.

*** export not supported

>>> Concept Code


Sequence

(0040,A168)

1C

Coded concept value of this name-value Item. Required if


Value Type (0040,A040) is CODE.

*** export not supported

>>>> Code Value

(0008,0100)

1C

See Section 8.1. Required if a sequence item is present.

*** export not supported

>>>> Coding
Scheme Designator

(0008,0102)

1C

See Section 8.2. Required if a sequence item is present.

*** export not supported

>>>> Coding
Scheme Version

(0008,0103)

1C

See Section 8.2. Required if a sequence item is present


and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.

*** export not supported

>>>> Code Meaning

(0008,0104)

1C

See Section 8.3. Required if a sequence item is present.

*** export not supported

>>> Numeric Value

(0040,A30A)

1C

Numeric value for this name-value Item. Required if


Value Type (0040,A040) is NUMERIC.

*** export not supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

IOD Details, Frame of Reference C.7.4.1

B.2.4

Attribute Name

Tag

Description

Handling

>>> Measurement
Units Code
Sequence

(0040,08EA)

1C

Units of measurement for a numeric value in this


namevalue Item. Required if Value Type (0040,A040) is
NUMERIC.

*** export not supported

>>>> Code Value

(0008,0100)

1C

See Section 8.1. Required if a sequence item is present.

*** export not supported

>>>> Coding
Scheme Designator

(0008,0102)

1C

See Section 8.2. Required if a sequence item is present.

*** export not supported

>>>> Coding
Scheme Version

(0008,0103)

1C

See Section 8.2. Required if a sequence item is present


and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.

*** export not supported

>>>> Code Meaning

(0008,0104)

1C

See Section 8.3. Required if a sequence item is present.

*** export not supported

Comments on the
Performed
Procedure Step

(0040,0280)

User-defined comments on the Performed Procedure


Step.

*** export not supported

Frame of Reference C.7.4.1


Attribute Name

Tag

Description

Handling

Frame of Reference
UID

(0020,0052)

Uniquely identifies the frame of reference for a Series.


See C.7.4.1.1.1 for further explanation.

Export supported for:


All

Position Reference
Indicator

(0020,1040)

Part of the patient's anatomy used as a reference, such


as the iliac crest, orbital- medial, sternal notch, symphysis
pubis, xiphoid, lower coastal margin, external auditory
meatus. See C.7.4.1.1.2 for further explanation.

*** export not supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

69 of 118

IOD Details, General Equipment C.7.5.1

B.2.5

70 of 118

General Equipment C.7.5.1


Attribute Name

Tag

Description

Handling

Manufacturer

(0008,0070)

Manufacturer of the equipment that produced the


composite instances.

Export supported for:


RT Image
Computed Tomography Image

Institution Name

(0008,0080)

Institution where the equipment that produced the


composite instances is located.

*** export not supported

Institution Address

(0008,0081)

Mailing address of the institution where the equipment


that produced the composite instances is located.

*** export not supported

Station Name

(0008,1010)

User defined name identifying the machine that produced


the composite instances.

Export supported for:


RT Image
Computed Tomography Image

Institutional
Department Name

(0008,1040)

Department in the institution where the equipment that


produced the composite instances is located.

*** export not supported

Manufacturers
Model Name

(0008,1090)

Manufacturer's model name of the equipment that


produced the composite instances.

Export supported for:


RT Image:
OBI: On-Board Imager
PV AI: PortalVision AI
Computed Tomography Image:
OBI Cone-beam CT

Device Serial
Number

(0018,1000)

Manufacturer's serial number of the equipment that


produced the composite instances.

*** export not supported

Software Version(s)

(0018,1020)

Manufacturer's designation of software version of the


equipment that produced the composite instances.

Export supported for:


RT Image
Computed Tomography Image

Spatial Resolution

(0018,1050)

The inherent limiting resolution in mm of the acquisition


equipment for high contrast objects for the data gathering
and reconstruction technique chosen. If variable across
the images of the series, the value at the image center.

*** export not supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

IOD Details, General Image C.7.6.1

B.2.6

Attribute Name

Tag

Description

Handling

Date of Last
Calibration

(0018,1200)

Date when the image acquisition device calibration was


last changed in any way. Multiple entries may be used for
additional calibrations at other times. See C.7.5.1.1.1 for
further explanation.

*** export not supported

Time of Last
Calibration

(0018,1201)

Time when the image acquisition device calibration was


last changed in any way. Multiple entries may be used.
See C.7.5.1.1.1 for further explanation.

*** export not supported

Pixel Padding Value

(0028,0120)

1C

Single pixel value or one limit (inclusive) of a range of


pixel values used in animage to pad to rectangular format
or to signal background that may be suppressed. See
C.7.5.1.1.2 for further explanation.
Required if Pixel Padding Range Limit (0028,0121) is
present. May be present otherwise.
Note: The Value Representation of this Attribute is
determined by the value of Pixel Representation
(0028,0103).

*** export not supported

General Image C.7.6.1


Attribute Name

Tag

Description

Handling

Instance Number

(0020,0013)

A number that identifies this image.


Note:
This Attribute was named Image Number in earlier
versions of this Standard.

Export supported for:


RT Image
Computed Tomography Image
Value is set to 0.

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

71 of 118

IOD Details, General Image C.7.6.1

72 of 118

Attribute Name

Tag

Description

Handling

Patient Orientation

(0020,0020)

2C

Patient direction of the rows and columns of the image.


Required if image does not require Image Orientation
(Patient) (0020,0037) and Image Position (Patient)
(0020,0032). See C.7.6.1.1.1 for further explanation.
Note:
IOD's may have attributes other than Patient Orientation,
Image Orientation, or Image Position (Patient) to describe
orientation in which case this attribute will be zero length.

*** export not supported

Content Date

(0008,0023)

2C

The date the image pixel data creation started. Required


if image is part of a series in which the images are
temporally related.
Note:
This Attribute was formerly known as Image Date.

Export supported for:


RT Image
Computed Tomography Image

Content Time

(0008,0033)

2C

The time the image pixel data creation started. Required


if image is part of a series in which the images are
temporally related.

Export supported for:


RT Image
Computed Tomography Image

Image Type

(0008,0008)

Image identification characteristics. See C.7.6.1.1.2 for


Defined Terms and further explanation.

Export supported for:


RT Image
kV:
ORIGINAL\PRIMARY\PORTAL\OBI
MV: ORIGINAL\PRIMARY\PORTAL
Computed Tomography Image:
ORIGINAL\PRIMARY\AXIAL
ORIGINAL\PRIMARY\LOCALIZER

Acquisition Number

(0020,0012)

A number identifying the single continuous gathering of


data over a period of time that resulted in this image.

Export supported for:


Computed Tomography Image

Acquisition Date

(0008,0022)

The date the acquisition of data that resulted in this


image started

*** export not supported

Acquisition Time

(0008,0032)

The time the acquisition of data that resulted in this


image started

*** export not supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

IOD Details, General Image C.7.6.1

Attribute Name

Tag

Description

Handling

Acquisition Datetime

(0008,002A)

The date and time that the acquisition of data that


resulted in this image started.
Note:
The synchronization of this time with an external clock is
specified in the Synchronization Module in Acquisition
Time Synchronized (0018,1800).

*** export not supported

Referenced Image
Sequence

(0008,1140)

A sequence that references other images significantly


related to this image (e.g. post-localizer CT image or
Mammographic biopsy or partial view images). One or
more Items may be included in this sequence.

*** export not supported

> Referenced SOP


Class UID

(0008,1150)

Uniquely identifies the referenced SOP Class.

*** export not supported

> Referenced SOP


Instance UID

(0008,1155)

Uniquely identifies the referenced SOP Instance.

*** export not supported

> Referenced
Frame Number

(0008,1160)

Identifies the frame numbers within the Referenced SOP


Instance to which the reference applies. The first frame
shall be denoted as frame number 1.
Note: This Attribute may be multi-valued.
Required if the Referenced SOP Instance is a multi-frame
image and the reference does not apply to all frames.

*** export not supported

> Purpose of
Reference Code
Sequence

(0040,A170)

Describes the purpose for which the reference is made.


Only a single Item shall be permitted in this sequence.

*** export not supported

>> Code Value

(0008,0100)

1C

See Section 8.1. Required if a sequence item is present.

*** export not supported

>> Coding Scheme


Designator

(0008,0102)

1C

See Section 8.2. Required if a sequence item is present.

*** export not supported

>> Coding Scheme


Version

(0008,0103)

1C

See Section 8.2. Required if a sequence item is present


and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.

*** export not supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

73 of 118

IOD Details, General Image C.7.6.1

74 of 118

Attribute Name

Tag

Description

Handling

>> Code Meaning

(0008,0104)

1C

See Section 8.3. Required if a sequence item is present.

*** export not supported

Derivation
Description

(0008,2111)

A text description of how this image was derived. See


C.7.6.1.1.3 for further explanation.

*** export not supported

Derivation Code
Sequence

(0008,9215)

A coded description of how this image was derived. See


C.7.6.1.1.3 for further explanation. One or more Items
may be included in this Sequence. More than one Item
indicates that successive derivation steps have been
applied.

*** export not supported

> Code Value

(0008,0100)

1C

See Section 8.1. Required if a sequence item is present.

*** export not supported

> Coding Scheme


Designator

(0008,0102)

1C

See Section 8.2. Required if a sequence item is present.

*** export not supported

> Coding Scheme


Version

(0008,0103)

1C

See Section 8.2. Required if a sequence item is present


and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.

*** export not supported

> Code Meaning

(0008,0104)

1C

See Section 8.3. Required if a sequence item is present.

*** export not supported

Source Image
Sequence

(0008,2112)

A Sequence that identifies the set of Image SOP


Class/Instance pairs of the Images that were used to
derive this Image. Zero or more Items may be included in
this Sequence. See C.7.6.1.1.4 for further explanation.

*** export not supported

> Referenced SOP


Class UID

(0008,1150)

Uniquely identifies the referenced SOP Class.

*** export not supported

> Referenced SOP


Instance UID

(0008,1155)

Uniquely identifies the referenced SOP Instance.

*** export not supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

IOD Details, General Image C.7.6.1

Attribute Name

Tag

Description

Handling

> Referenced
Frame Number

(0008,1160)

Identifies the frame numbers within the Referenced SOP


Instance to which the reference applies. The first frame
shall be denoted as frame number 1.
Note: This Attribute may be multi-valued.
Required if the Referenced SOP Instance is a multi-frame
image and the reference does not apply to all frames.

*** export not supported

> Purpose of
Reference Code
Sequence

(0040,A170)

Describes the purpose for which the reference is made,


that is what role the source image or frame(s) played in
the derivation of this image. Only a single Item shall be
permitted in this sequence.

*** export not supported

>> Code Value

(0008,0100)

1C

See Section 8.1. Required if a sequence item is present.

*** export not supported

>> Coding Scheme


Designator

(0008,0102)

1C

See Section 8.2. Required if a sequence item is present.

*** export not supported

>> Coding Scheme


Version

(0008,0103)

1C

See Section 8.2. Required if a sequence item is present


and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.

*** export not supported

>> Code Meaning

(0008,0104)

1C

See Section 8.3. Required if a sequence item is present.

*** export not supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

75 of 118

IOD Details, General Image C.7.6.1

76 of 118

Attribute Name

Tag

Description

Handling

> Spatial Locations


Preserved

(0028,135A)

Whether or not the spatial locations of all pixels are


preserved during the processing of the source image that
resulted in the current image
Enumerated Values:
YES
NO
Note:
This applies not only to images with a known relationship
to a 3D space, but also to projection images. For
example, a projection radiograph such as a mammogram
that is processed by a point image processing operation
such as contrast enhancement, or a smoothing or edge
enhancing convolution, would have a value of YES for
this attribute. A projection radiograph that had been
magnified or warped geometrically would have a value of
NO for this attribute. This attribute is typically of
importance in relating images with Presentation Intent
Type (0008,0068) values of FOR PROCESSING and
FOR PRESENTATION.

*** export not supported

Referenced
Instance Sequence

(0008,114A)

A sequence which provides reference to a set of nonimage SOP Class/Instance pairs significantly related to
this Image, including waveforms that may or may not be
temporally synchronized with this image . One or more
Items may be included in this sequence.

*** export not supported

> Referenced SOP


Class UID

(0008,1150)

Uniquely identifies the referenced SOP Class.

*** export not supported

> Referenced SOP


Instance UID

(0008,1155)

Uniquely identifies the referenced SOP Instance.

*** export not supported

> Purpose of
Reference Code
Sequence

(0040,A170)

Code describing the purpose of the reference to the


Instance(s). Only a single Item shall be permitted in this
sequence.

*** export not supported

>> Code Value

(0008,0100)

1C

See Section 8.1. Required if a sequence item is present.

*** export not supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

IOD Details, General Image C.7.6.1

Attribute Name

Tag

Description

Handling

>> Coding Scheme


Designator

(0008,0102)

1C

See Section 8.2. Required if a sequence item is present.

*** export not supported

>> Coding Scheme


Version

(0008,0103)

1C

See Section 8.2. Required if a sequence item is present


and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.

*** export not supported

>> Code Meaning

(0008,0104)

1C

See Section 8.3. Required if a sequence item is present.

*** export not supported

Images in
Acquisition

(0020,1002)

Number of images that resulted from this acquisition of


data

*** export not supported

Image Comments

(0020,4000)

User-defined comments about the image

*** export not supported

Quality Control
Image

(0028,0300)

Indicates whether or not this image is a quality control or


phantom image.
Enumerated Values:
YES
NO If this Attribute is absent, then the image may or may
not be a quality control or phantom image.

*** export not supported

Burned In
Annotation

(0028,0301)

Indicates whether or not image contains sufficient burned


in annotation to identify the patient and date the image
was acquired.
Enumerated Values:
YES
NO If this Attribute is absent, then the image may or may
not contain burned in annotation.

*** export not supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

77 of 118

IOD Details, General Image C.7.6.1

78 of 118

Attribute Name

Tag

Description

Handling

Lossy Image
Compression

(0028,2110)

Specifies whether an Image has undergone lossy


compression.
Enumerated Values:
00 = Image has NOT been subjected to lossy
compression.
01 = Image has been subjected to lossy compression.
See C.7.6.1.1.5

*** export not supported

Lossy Image
Compression Ratio

(0028,2112)

Describes the approximate lossy compression ratio(s)


that have been applied to this image. See C.7.6.1.1.5 for
further explanation. May be multivalued if successive
lossy compression steps have been applied.
Notes:
1. For example, a compression ratio of 30:1 would be
described in this Attribute with a single value of 30.
2. For historical reasons, the lossy compression ratio may
also be described in Derivation Description (0008,2111).

*** export not supported

Lossy Image
Compression
Method

(0028,2114)

A label for the lossy compression method(s) that have


been applied to this image. See C.7.6.1.1.5 for further
explanation. May be multivalued if successive lossy
compression steps have been applied; the value order
shall correspond to the values of Lossy Image
Compression Ratio (0028,2112).
Note:
For historical reasons, the lossy compression method
may also be described in Derivation Description
(0008,2111).

*** export not supported

Icon Image
Sequence

(0088,0200)

This icon image is representative of the Image.

*** export not supported

> Samples per Pixel

(0028,0002)

Number of samples (planes) in this image. See


C.7.6.3.1.1 for further explanation.

*** export not supported

> Photometric
Interpretation

(0028,0004)

Specifies the intended interpretation of the pixel data.


See C.7.6.3.1.2 for further explanation.

*** export not supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

IOD Details, General Image C.7.6.1

Attribute Name

Tag

Description

Handling

> Rows

(0028,0010)

Number of rows in the image.

*** export not supported

> Columns

(0028,0011)

Number of columns in the image.

*** export not supported

> Bits Allocated

(0028,0100)

Number of bits allocated for each pixel sample. Each


sample shall have the same number of bits allocated.
See PS 3.5 for further explanation.

*** export not supported

> Bits Stored

(0028,0101)

Number of bits stored for each pixel sample. Each


sample shall have the same number of bits stored. See
PS 3.5 for further explanation.

*** export not supported

> High Bit

(0028,0102)

Most significant bit for pixel sample data. Each sample


shall have the same high bit. See PS 3.5 for further
explanation.

*** export not supported

> Pixel
Representation

(0028,0103)

Data representation of the pixel samples. Each sample


shall have the same pixel representation. Enumerated
Values:
0000H = unsigned integer.
0001H = 2's complement

*** export not supported

> Pixel Data

(7FE0,0010)

1C

A data stream of the pixel samples that comprise the


Image. See C.7.6.3.1.4 for further explanation. Required
if Pixel Data Provider URL (0028,7FE0) is not present.

*** export not supported

> Planar
Configuration

(0028,0006)

1C

Indicates whether the pixel data are sent color-by-plane


or color-by-pixel. Required if Samples per Pixel
(0028,0002) has a value greater than 1. See C.7.6.3.1.3
for further explanation.

*** export not supported

> Pixel Aspect Ratio

(0028,0034)

1C

Ratio of the vertical size and horizontal size of the pixels


in the image specified by a pair of integer values where
the first value is the vertical pixel size, and the second
value is the horizontal pixel size. Required if the aspect
ratio is not 1 and the Image Plane Module or the Pixel
Measures Macro is not applicable to this Image. See
C.7.6.3.1.7.

*** export not supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

79 of 118

IOD Details, General Image C.7.6.1

80 of 118

Attribute Name

Tag

Description

Handling

> Smallest Image


Pixel Value

(0028,0106)

The minimum actual pixel value encountered in this


image.

*** export not supported

> Largest Image


Pixel Value

(0028,0107)

The maximum actual pixel value encountered in this


image.

*** export not supported

> Red Palette Color


Lookup Table
Descriptor US or

(0028,1101)

1C

Specifies the format of the Red Palette Color Lookup


Table Data (0028,1201) Required if Photometric
Interpretation (0028,0004) has a value of PALETTE
COLOR or Pixel Presentation (0008,9205) at the image
level equals COLOR or MIXED. See C.7.6.3.1.5 for
further explanation.

*** export not supported

> Green Palette


Color Lookup Table
Descriptor US or

(0028,1102)

1C

Specifies the format of the Green Palette Color Lookup


Table Data (0028,1202) Required if Photometric
Interpretation (0028,0004) has a value of PALETTE
COLOR or Pixel Presentation (0008,9205) at the image
level equals COLOR or MIXED. See C.7.6.3.1.5 for
further explanation.

*** export not supported

> Blue Palette Color


Lookup Table
Descriptor US or

(0028,1103)

1C

Specifies the format of the Blue Palette Color Lookup


Table Data (0028,1203) Required if Photometric
Interpretation (0028,0004) has a value of PALETTE
COLOR or Pixel Presentation (0008,9205) at the image
level equals COLOR or MIXED. See C.7.6.3.1.5 for
further explanation.

*** export not supported

> Red Palette Color


Lookup Table Data

(0028,1201)

1C

Red Palette Color Lookup Table Data. Required if


Photometric Interpretation (0028,0004) has a value of
PALETTE COLOR or Pixel Presentation (0008,9205) at
the image level equals COLOR or MIXED. See
C.7.6.3.1.6 for further explanation.

*** export not supported

> Green Palette


Color Lookup Table
Data

(0028,1202)

1C

Green Palette Color Lookup Table Data. Required if


Photometric Interpretation (0028,0004) has a value of
PALETTE COLOR or Pixel Presentation (0008,9205) at
the image level equals COLOR or MIXED. See
C.7.6.3.1.6 for further explanation.

*** export not supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

IOD Details, General Image C.7.6.1

Attribute Name

Tag

Description

Handling

> Blue Palette Color


Lookup Table Data

(0028,1203)

1C

Blue Palette Color Lookup Table Data. Required if


Photometric Interpretation (0028,0004) has a value of
PALETTE COLOR or Pixel Presentation (0008,9205) at
the image level equals COLOR or MIXED. See
C.7.6.3.1.6 for further explanation.

*** export not supported

> ICC Profile

(0028,2000)

An ICC Profile encoding the transformation of devicedependent color stored pixel values into PCS-Values.
See Section C.11.15.1.1.1. When present, defines the
color space of color Pixel Data (7FE0,0010) values, and
the output of Palette Color Lookup Table. Data
(0028,1201-1203). Note: The profile applies only to the
Pixel Data (7FE0,0010) attribute at the same level of the
dataset and not to any icons nested within sequences,
which may or may not have their own ICC profile
specified.

*** export not supported

Presentation LUT
Shape

(2050,0020)

When present, specifies an identity transformation for the


Presentation LUT such that the output of all grayscale
transformations, if any, are defined to be in P-Values.
Enumerated Values are:
IDENTITY = output is in P-Values - shall be used if
Photometric Interpretation (0028,0004) is
MONOCHROME2 or any color photometric interpretation.
INVERSE = output after inversion is in P- Values - shall
be used if Photometric Interpretation (0028,0004) is
MONOCHROME1. When this attribute is used with a
color photometric interpretation then the luminance
component is in P-Values.

*** export not supported

Irradiation Event
UID

(0008,3010)

Unique identification of the irradiation event(s) associated


with the acquisition of this image. See C.7.6.1.1.7.

*** export not supported

Image Remarks

(3281,XX00)

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

81 of 118

IOD Details, Image Plane C.7.6.2

B.2.7

B.2.8

82 of 118

Image Plane C.7.6.2


Attribute Name

Tag

Description

Handling

Pixel Spacing

(0028,0030)

Physical distance in the patient between the center of


each pixel, specified by a numeric pair - adjacent row
spacing (delimiter) adjacent column spacing in mm.

Export supported for:


Computed Tomography Image

Image Orientation
(Patient)

(0020,0037)

The direction cosines of the first row and the first column
with respect to the patient. See C.7.6.2.1.1 for further
explanation.

Export supported for:


Computed Tomography Image

Image Position
(Patient)

(0020,0032)

The x, y, and z coordinates of the upper left hand corner


(center of the first voxel transmitted) of the image, in mm.
See C.7.6.2.1.1 for further explanation.

Export supported for:


Computed Tomography Image

Slice Thickness

(0018,0050)

Nominal slice thickness, in mm.

Export supported for:


Computed Tomography Image

Slice Location

(0020,1041)

Relative position of exposure expressed in mm.


C.7.6.2.1.2 for further explanation.

*** export not supported

Image Pixel C.7.6.3


Attribute Name

Tag

Description

Handling

Samples per Pixel

(0028,0002)

Number of samples (planes) in this image. See


C.7.6.3.1.1 for further explanation.

Export supported for:


RT Image
Computed Tomography Image
Value set to 1.

Photometric
Interpretation

(0028,0004)

Specifies the intended interpretation of the pixel data.


See C.7.6.3.1.2 for further explanation.

Export supported for:


RT Image
Computed Tomography Image
Value set to MONOCHROME2

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

IOD Details, Image Pixel C.7.6.3

Attribute Name

Tag

Description

Handling

Rows

(0028,0010)

Number of rows in the image.

Export supported for:


RT Image
Computed Tomography Image

Columns

(0028,0011)

Number of columns in the image.

Export supported for:


RT Image
Computed Tomography Image

Bits Allocated

(0028,0100)

Number of bits allocated for each pixel sample. Each


sample shall have the same number of bits allocated.
See PS 3.5 for further explanation.

Export supported for:


RT Image
Computed Tomography Image

Bits Stored

(0028,0101)

Number of bits stored for each pixel sample. Each


sample shall have the same number of bits stored. See
PS 3.5 for further explanation.

Export supported for:


RT Image
Computed Tomography Image

High Bit

(0028,0102)

Most significant bit for pixel sample data. Each sample


shall have the same high bit. See PS 3.5 for further
explanation.

Export supported for:


RT Image
Computed Tomography Image

Pixel
Representation

(0028,0103)

Data representation of the pixel samples. Each sample


shall have the same pixel representation. Enumerated
Values:
0000H = unsigned integer.
0001H = 2's complement

Export supported for:


RT Image
Computed Tomography Image

Pixel Data

(7FE0,0010)

1C

A data stream of the pixel samples that comprise the


Image. See C.7.6.3.1.4 for further explanation. Required
if Pixel Data Provider URL (0028,7FE0) is not present.

Export supported for:


RT Image
Computed Tomography Image

Planar Configuration

(0028,0006)

1C

Indicates whether the pixel data are sent color-by-plane


or color-by-pixel. Required if Samples per Pixel
(0028,0002) has a value greater than 1. See C.7.6.3.1.3
for further explanation.

*** export not supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

83 of 118

IOD Details, Image Pixel C.7.6.3

84 of 118

Attribute Name

Tag

Description

Handling

Pixel Aspect Ratio

(0028,0034)

1C

Ratio of the vertical size and horizontal size of the pixels


in the image specified by a pair of integer values where
the first value is the vertical pixel size, and the second
value is the horizontal pixel size. Required if the aspect
ratio is not 1 and the Image Plane Module or the Pixel
Measures Macro is not applicable to this Image. See
C.7.6.3.1.7.

*** export not supported

Smallest Image
Pixel Value

(0028,0106)

The minimum actual pixel value encountered in this


image.

*** export not supported

Largest Image Pixel


Value

(0028,0107)

The maximum actual pixel value encountered in this


image.

*** export not supported

Red Palette Color


Lookup Table
Descriptor US or

(0028,1101)

1C

Specifies the format of the Red Palette Color Lookup


Table Data (0028,1201) Required if Photometric
Interpretation (0028,0004) has a value of PALETTE
COLOR or Pixel Presentation (0008,9205) at the image
level equals COLOR or MIXED. See C.7.6.3.1.5 for
further explanation.

*** export not supported

Green Palette Color


Lookup Table
Descriptor US or

(0028,1102)

1C

Specifies the format of the Green Palette Color Lookup


Table Data (0028,1202) Required if Photometric
Interpretation (0028,0004) has a value of PALETTE
COLOR or Pixel Presentation (0008,9205) at the image
level equals COLOR or MIXED. See C.7.6.3.1.5 for
further explanation.

*** export not supported

Blue Palette Color


Lookup Table
Descriptor US or

(0028,1103)

1C

Specifies the format of the Blue Palette Color Lookup


Table Data (0028,1203) Required if Photometric
Interpretation (0028,0004) has a value of PALETTE
COLOR or Pixel Presentation (0008,9205) at the image
level equals COLOR or MIXED. See C.7.6.3.1.5 for
further explanation.

*** export not supported

Red Palette Color


Lookup Table Data

(0028,1201)

1C

Red Palette Color Lookup Table Data. Required if


Photometric Interpretation (0028,0004) has a value of
PALETTE COLOR or Pixel Presentation (0008,9205) at
the image level equals COLOR or MIXED. See
C.7.6.3.1.6 for further explanation.

*** export not supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

IOD Details, Image Pixel C.7.6.3

Attribute Name

Tag

Description

Handling

Green Palette Color


Lookup Table Data

(0028,1202)

1C

Green Palette Color Lookup Table Data. Required if


Photometric Interpretation (0028,0004) has a value of
PALETTE COLOR or Pixel Presentation (0008,9205) at
the image level equals COLOR or MIXED. See
C.7.6.3.1.6 for further explanation.

*** export not supported

Blue Palette Color


Lookup Table Data

(0028,1203)

1C

Blue Palette Color Lookup Table Data. Required if


Photometric Interpretation (0028,0004) has a value of
PALETTE COLOR or Pixel Presentation (0008,9205) at
the image level equals COLOR or MIXED. See
C.7.6.3.1.6 for further explanation.

*** export not supported

ICC Profile

(0028,2000)

An ICC Profile encoding the transformation of devicedependent color stored pixel values into PCS-Values.
See Section C.11.15.1.1.1. When present, defines the
color space of color Pixel Data (7FE0,0010) values, and
the output of Palette Color Lookup Table. Data
(0028,1201-1203). Note: The profile applies only to the
Pixel Data (7FE0,0010) attribute at the same level of the
dataset and not to any icons nested within sequences,
which may or may not have their own ICC profile
specified.

*** export not supported

Pixel Data Provider


URL

(0028,7FE0)

1C

A URL of a provider service that supplies the pixel data of


the Image.
Required if the image is to be transferred in one of the
following presentation contexts identified by Transfer
Syntax UID: 1.2.840.10008.1.2.4.94 (DICOM JPIP
Referenced Transfer Syntax) 1.2.840.10008.1.2.4.95
(DICOM JPIP Referenced Deflate Transfer Syntax)

*** export not supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

85 of 118

IOD Details, CT Image C.8.2.1

B.2.9

86 of 118

CT Image C.8.2.1
Attribute Name

Tag

Description

Handling

Image Type

(0008,0008)

Image identification characteristics. See C.8.2.1.1.1 for


specialization.

Export supported
ORIGINAL\PRIMARY\
AXIAL
ORIGINAL\PRIMARY\
LOCALIZER

Samples per Pixel

(0028,0002)

Number of samples (planes) in this image. See


C.8.2.1.1.2 for specialization.

Export supported
Value set to 1

Photometric
Interpretation

(0028,0004)

Specifies the intended interpretation of the pixel data.


See C.8.2.1.1.3 for specialization.

Export supported
Value set to MONOCHROME2

Bits Allocated

(0028,0100)

Number of bits allocated for each pixel sample. Each


sample shall have the same number of bits allocated.
See C.8.2.1.1.4 for specialization.

Export supported

Bits Stored

(0028,0101)

Number of bits stored for each pixel sample. Each


sample shall have the same number of bits stored. See
C.8.2.1.1.5 for specialization.

Export supported

High Bit

(0028,0102)

Most significant bit for pixel sample data. Each sample


shall have the same high bit. See C.8.2.1.1.6 for
specialization.

Export supported

Rescale Intercept

(0028,1052)

The value b in relationship between stored values (SV)


and Hounsfield (HU). HU = m*SV+b

Export supported

Rescale Slope

(0028,1053)

m in the equation specified in Rescale Intercept


(0028,1052).

Export supported

KVP

(0018,0060)

Peak kilo voltage output of the x-ray generator used

*** export not supported

Acquisition Number

(0020,0012)

A number identifying the single continuous gathering of


data over a period of time which resulted in this image

Export supported

Scan Options

(0018,0022)

Parameters of scanning sequence.

Export supported
Value set to TOMO

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

IOD Details, CT Image C.8.2.1

Attribute Name

Tag

Description

Handling

Data Collection
Diameter

(0018,0090)

The diameter in mm of the region over which data were


collected

Export supported

Reconstruction
Diameter

(0018,1100)

Diameter in mm of the region from within which data were


used in creating the reconstruction of the image. Data
may exist outside this region and portions of the patient
may exist outside this region.

Export supported

Distance Source to
Detector

(0018,1110)

Distance in mm from source to detector center.


Note:
This value is traditionally referred to as Source Image
Receptor Distance (SID).

Export supported

Distance Source to
Patient

(0018,1111)

Distance in mm from source to isocenter (center of field


of view).
Note:
This value is traditionally referred to as Source Object
Distance (SOD).

Export supported
Value set to 0.

Gantry/Detector Tilt

(0018,1120)

Nominal angle of tilt in degrees of the scanning gantry.


Not intended for mathematical computations.

Export supported
Value set to 0.

Table Height

(0018,1130)

The distance in mm of the top of the patient table to the


center of rotation; below the center is positive.

Export supported

Table Top
Longitudinal
Position

(300A,0129)

Export supported

Table Top Lateral


Position

(300A,012A)

Export supported

Patient Support
Angle

(300A,0122)

Export supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

87 of 118

IOD Details, CT Image C.8.2.1

88 of 118

Attribute Name

Tag

Description

Handling

Rotation Direction

(0018,1140)

Direction of rotation of the source when relevant, about


nearest principal axis of equipment.
Enumerated Values:
CW = clockwise
CC = counter clockwise

*** export not supported

Exposure Time

(0018,1150)

Time of x-ray exposure in msec

Export supported

X-ray Tube Current

(0018,1151)

X-ray Tube Current in mA.

Export supported

Exposure

(0018,1152)

The exposure expressed in mAs, for example calculated


from Exposure Time and X-ray Tube Current.

Export supported

Exposure in uAs

(0018,1153)

The exposure expressed in As, for example calculated


from Exposure Time and X-ray Tube Current.

*** export not supported

Filter Type

(0018,1160)

Label for the type of filter inserted into the x-ray beam.

*** export not supported

Generator Power

(0018,1170)

Power in kW to the x-ray generator.

Export supported

Focal Spot(s)

(0018,1190)

Size of the focal spot in mm. For devices with variable


focal spot or multiple focal spots, small dimension
followed by large dimension.

Export supported

Convolution Kernel

(0018,1210)

A label describing the convolution kernel or algorithm


used to reconstruct the data

Export supported

Revolution Time

(0018,9305)

The time in seconds of a complete revolution of the


source around the gantry orbit.

*** export not supported

Single Collimation
Width

(0018,9306)

The width of a single row of acquired data (in mm).


Note:
Adjacent physical detector rows may have been
combined to form a single effective acquisition row.

*** export not supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

IOD Details, CT Image C.8.2.1

Attribute Name

Tag

Description

Handling

Total Collimation
Width

(0018,9307)

The width of the total collimation (in mm) over the area of
active x-ray detection.
Note:
This will be equal the number of effective detector rows
multiplied by single collimation width.

*** export not supported

Table Speed

(0018,9309)

The distance in mm that the table moves in one second


during the gathering of data that resulted in this image.

*** export not supported

Table Feed per


Rotation

(0018,9310)

Motion of the table (in mm) during a complete revolution


of the source around the gantry orbit.

*** export not supported

Spiral Pitch Factor

(0018,9311)

Ratio of the Table Feed per Rotation (0018,9310) to the


Total Collimation Width (0018,9307).

*** export not supported

Exposure
Modulation Type

(0018,9323)

A label describing the type of exposure modulation used


for the purpose of limiting the dose.
Defined Terms:
NONE

*** export not supported

Estimated Dose
Saving

(0018,9324)

A percent value of dose saving due to the use of


Exposure Modulation Type (0018,9323). A negative
percent value of dose savings reflects an increase of
exposure.

*** export not supported

CTDIvol

(0018,9345)

Computed Tomography Dose Index (CTDIvol), im mGy


according to IEC 60601-2-44, Ed.2.1 (Clause
29.1.103.4), The Volume CTDIvol. It describes the
average dose for this image for the selected CT
conditions of operation.

*** export not supported

Anatomic Region
Sequence

(0008,2218)

Sequence that identifies the anatomic region of interest in


this Instance (i.e. external anatomy, surface anatomy, or
general region of the body).
Only a single Item shall be permitted in this sequence.

*** export not supported

> Code Value

(0008,0100)

1C

See Section 8.1. Required if a sequence item is present.

*** export not supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

89 of 118

IOD Details, CT Image C.8.2.1

90 of 118

Attribute Name

Tag

Description

Handling

> Coding Scheme


Designator

(0008,0102)

1C

See Section 8.2. Required if a sequence item is present.

*** export not supported

> Coding Scheme


Version

(0008,0103)

1C

See Section 8.2. Required if a sequence item is present


and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.

*** export not supported

> Code Meaning

(0008,0104)

1C

See Section 8.3. Required if a sequence item is present.

*** export not supported

> Anatomic Region


Modifier Sequence

(0008,2220)

Sequence of Items that modifies the anatomic region of


interest of this Instance One or more Items may be
included in this Sequence.

*** export not supported

>> Code Value

(0008,0100)

1C

See Section 8.1. Required if a sequence item is present.

*** export not supported

>> Coding Scheme


Designator

(0008,0102)

1C

See Section 8.2. Required if a sequence item is present.

*** export not supported

>> Coding Scheme


Version

(0008,0103)

1C

See Section 8.2. Required if a sequence item is present


and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.

*** export not supported

>> Code Meaning

(0008,0104)

1C

See Section 8.3. Required if a sequence item is present.

*** export not supported

Primary Anatomic
Structure Sequence

(0008,2228)

Sequence of Items that identifies the primary anatomic


structure(s) of interest in this Instance.
One or more Items may be included in this Sequence.

*** export not supported

> Code Value

(0008,0100)

1C

See Section 8.1. Required if a sequence item is present.

*** export not supported

> Coding Scheme


Designator

(0008,0102)

1C

See Section 8.2. Required if a sequence item is present.

*** export not supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

IOD Details, RT Image C.8.8.2

B.2.10

Attribute Name

Tag

Description

Handling

> Coding Scheme


Version

(0008,0103)

1C

See Section 8.2. Required if a sequence item is present


and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.

*** export not supported

> Code Meaning

(0008,0104)

1C

See Section 8.3. Required if a sequence item is present.

*** export not supported

> Primary Anatomic


Structure Modifier
Sequence

(0008,2230)

Sequence of Items that modifies the primary anatomic


structure of interest in this Instance.
One or more Items may be included in this Sequence.

*** export not supported

>> Code Value

(0008,0100)

1C

See Section 8.1. Required if a sequence item is present.

*** export not supported

>> Coding Scheme


Designator

(0008,0102)

1C

See Section 8.2. Required if a sequence item is present.

*** export not supported

>> Coding Scheme


Version

(0008,0103)

1C

See Section 8.2. Required if a sequence item is present


and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.

*** export not supported

>> Code Meaning

(0008,0104)

1C

See Section 8.3. Required if a sequence item is present.

*** export not supported

RT Image C.8.8.2
Attribute Name

Tag

Description

Handling

Samples per Pixel

(0028,0002)

Number of samples (planes) in this image. See


C.8.8.2.6.1 for specialization.

Export supported

Photometric
Interpretation

(0028,0004)

Specifies the intended interpretation of the pixel data.


See C.7.6.3.1.2 for further explanation.

Export supported
Value set to MONOCHROME2

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

91 of 118

IOD Details, RT Image C.8.8.2

92 of 118

Attribute Name

Tag

Description

Handling

Bits Allocated

(0028,0100)

Number of bits allocated for each pixel sample. Each


sample shall have the same number of bits allocated.
See PS 3.5 for further explanation.

Export supported

Bits Stored

(0028,0101)

Number of bits stored for each pixel sample. Each


sample shall have the same number of bits stored. See
PS 3.5 for further explanation.

Export supported

High Bit

(0028,0102)

Most significant bit for pixel sample data. Each sample


shall have the same high bit. See PS 3.5 for further
explanation.

Export supported

Pixel
Representation

(0028,0103)

Data representation of the pixel samples. Each sample


shall have the same pixel representation. Enumerated
Values:
0000H = unsigned integer.
0001H = 2's complement

Export supported

Pixel Intensity
Relationship

(0028,1040)

The relationship between the Pixel sample values and


the X-Ray beam intensity.
Enumerated Values:
LIN = Linearly proportional to X-Ray beam intensity
LOG = Logarithmically proportional to X- Ray beam
intensity
See C.8.11.3.1.2 for further explanation.

*** export not supported

Pixel Intensity
Relationship Sign

(0028,1041)

1C

The sign of the relationship between the Pixel sample


values stored in Pixel Data (7FE0,0010) and the X-Ray
beam intensity.
Required if Pixel Intensity Relationship (0028,1040) is
present. Enumerated Values; 1 = Lower pixel values
correspond to less X-Ray beam intensity -1 = Higher pixel
values correspond to less X-Ray beam intensity See
C.8.11.3.1.2 for further explanation.

*** export not supported

RT Image Label

(3002,0002)

User-defined label for RT Image.

Export supported
Used as Image ID.

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

IOD Details, RT Image C.8.8.2

Attribute Name

Tag

Description

Handling

RT Image Name

(3002,0003)

User-defined name for RT Image.

*** export not supported

RT Image
Description

(3002,0004)

User-defined description of RT Image.

Export supported for MV image only

Operators Name

(0008,1070)

Name of operator(s) acquiring or creating RT Image.

*** export not supported

Image Type

(0008,0008)

Image identification characteristics (see Section


C.7.6.1.1.2). RT Images shall use one of the following
Defined Terms for Value 3: DRR = digitally reconstructed
radiograph PORTAL = digital portal image or portal film
image SIMULATOR = conventional simulator image
RADIOGRAPH = radiographic image BLANK = image
pixels set to background value FLUENCE = fluence map

Export supported
kV:
ORIGINAL\PRIMARY\PORTAL\OBI
MV: ORIGINAL\PRIMARY\PORTAL
Value for acquired portal dose:
ORIGINAL\PRIMARY\PORTAL\ACQUI
RED_DOSE

Conversion Type

(0008,0064)

Describes the kind of image conversion.


Defined Terms:
DV = Digitized Video
DI = Digital Interface
DF = Digitized Film
WSD = Workstation

Export supported
Value set to DI

Reported Values
Origin

(3002,000A)

2C

Describes the origin of the parameter values reported in


the image. Required if Value 3 of Image Type
(0008,0008) is SIMULATOR or PORTAL.
Enumerated Values:
OPERATOR = manually entered by operator
PLAN = planned parameter values
ACTUAL = electronically recorded

Export supported
Value set to ACTUAL

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

93 of 118

IOD Details, RT Image C.8.8.2

94 of 118

Attribute Name

Tag

Description

Handling

RT Image Plane

(3002,000C)

Describes whether or not image plane is normal to beam


axis.
Enumerated Values:
NORMAL = image plane normal to beam axis
NON_NORMAL = image plane non-normal to beam axis

Export supported
Value set to NORMAL

X-Ray Image
Receptor
Translation

(3002,000D)

Position in (x,y,z) coordinates of origin of IEC X-RAY


IMAGE RECEPTOR System in the IEC GANTRY
coordinate system (mm). See Note 2.

Export supported

X-Ray Image
Receptor Angle

(3002,000E)

X-Ray Image Receptor Angle i.e. orientation of IEC XRAY IMAGE RECEPTOR coordinate system with respect
to IEC GANTRY coordinate system (degrees). See
C.8.8.2.2.

Export supported

RT Image
Orientation

(3002,0010)

2C

The direction cosines of the first row and the first column
with respect to the IEC XRAY IMAGE RECEPTOR
coordinate system. Required if RT Image Plane
(3002,000C) is NON_NORMAL. May be present
otherwise.

Export supported

Image Plane Pixel


Spacing

(3002,0011)

Physical distance (in mm) between the center of each


image pixel, specified by a numeric pair - adjacent row
spacing (delimiter) adjacent column spacing. See
C.8.8.2.3.

Export supported

RT Image Position

(3002,0012)

The x and y coordinates (in mm) of the upper left hand


corner of the image, in the IEC X-RAY IMAGE
RECEPTOR coordinate system. This is the center of the
first pixel transmitted. See C.8.8.2.7.

Export supported

Radiation Machine
Name

(3002,0020)

User-defined name identifying radiation machine used in


acquiring or computing image (i.e. name of conventional
simulator, electron accelerator, X-ray device, or machine
modeled when calculating DRR).

*** export not supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

IOD Details, RT Image C.8.8.2

Attribute Name

Tag

Description

Handling

Primary Dosimeter
Unit

(300A,00B3)

Measurement unit of machine dosimeter.


Enumerated Values:
MU = Monitor Unit
MINUTE = minute

Export supported
Value for MV Image: MU
Value for kV Image: MINUTE

Radiation Machine
SAD

(3002,0022)

Radiation source to Gantry rotation axis distance of


radiation machine used in acquiring or computing image
(mm).

Export supported

Radiation Machine
SSD

(3002,0024)

Source to patient surface distance (in mm) of radiation


machine used in acquiring or computing image.

*** export not supported

RT Image SID

(3002,0026)

Distance from radiation machine source to image plane


(in mm) along radiation beam axis. See C.8.8.2.3.

Export supported

Source to
Reference Object
Distance

(3002,0028)

Source to reference object distance (in mm), as used for


magnification calculation of RADIOGRAPH and
SIMULATOR images.

*** export not supported

Referenced RT Plan
Sequence

(300C,0002)

Introduces sequence of one Class/Instance pair


describing RT Plan associated with image. Only a single
item shall be permitted in this sequence.

Export supported

> Referenced SOP


Class UID

(0008,1150)

1C

Uniquely identifies the referenced SOP Class. Required if


Referenced RT Plan Sequence (300C,0002) is sent.

Export supported
Supported values: SOP Class UID of
RT Plan and RT Ion Plan.

> Referenced SOP


Instance UID

(0008,1155)

1C

Uniquely identifies the referenced SOP Instance.


Required if Referenced RT Plan Sequence (300C,0002)
is sent.

Export supported
Plan UID of the plan where this image
is a field image.

Referenced Beam
Number

(300C,0006)

Uniquely identifies the corresponding N- segment


treatment beam specified by Beam Number (300A,00C0)
within Beam Sequence in RT Beams Module within the
RT Plan referenced in Referenced RT Plan Sequence
(300C,0002).

Export supported

Referenced Fraction
Group Number

(300C,0022)

Identifier of Fraction Group within RT Plan referenced in


Referenced RT Plan Sequence (300C,0002).

*** export not supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

95 of 118

IOD Details, RT Image C.8.8.2

96 of 118

Attribute Name

Tag

Description

Handling

Fraction Number

(3002,0029)

Fraction Number of fraction during which image was


acquired, within Fraction Group referenced by
Referenced Fraction Group Number (300C,0022) within
RT Plan referenced in Referenced RT Plan Sequence
(300C,0002).

Export supported
Value set to 1.

Start Cumulative
Meterset Weight

(300C,0008)

Cumulative Meterset Weight within Beam referenced by


Referenced Beam Number (300C,0006) at which image
acquisition starts.

Export supported for MV Image only

End Cumulative
Meterset Weight

(300C,0009)

Cumulative Meterset Weight within Beam referenced by


Referenced Beam Number (300C,0006) at which image
acquisition ends.

*** export not supported

Exposure Sequence

(3002,0030)

Introduces sequence of Exposure parameter sets,


corresponding to exposures used in generating the
image. One or more items may be included in this
sequence. See C.8.8.2.4.

Export supported.
Only 1 item supported

> Referenced
Frame Number

(0008,1160)

1C

Identifies corresponding image frame in multi-frame


image. Required if Exposure Sequence (3002,0030) is
sent, there is more than one item in Exposure Sequence
(3002,0030), and image is a multi-frame image.

*** export not supported

> KVP

(0018,0060)

2C

Peak kilo voltage output (kV) of X-ray generator used to


acquire image. Required if Value 3 of Image Type
(0008,0008) is PORTAL, SIMULATOR or RADIOGRAPH
and Exposure Sequence (3002,0030) is sent.

Export supported

> X-ray Tube


Current

(0018,1151)

2C

Imaging device X-ray Tube Current (mA).


Required if Value 3 of Image Type (0008,0008) is
SIMULATOR or RADIOGRAPH and Exposure Sequence
(3002,0030) is sent.

Export supported for kV Image

> Exposure Time

(0018,1150)

2C

Time of X-ray exposure (msec). Required if Value 3 of


Image Type (0008,0008) is SIMULATOR or
RADIOGRAPH and Exposure Sequence (3002,0030) is
sent.

Export supported for kV Image

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

IOD Details, RT Image C.8.8.2

Attribute Name

Tag

Description

Handling

> Meterset
Exposure

(3002,0032)

2C

Treatment machine Meterset duration over which image


has been acquired, specified in Monitor units (MU) or
minutes as defined by Primary Dosimeter Unit
(300A,00B3).
Required if Value 3 of Image Type (0008,0008) is
PORTAL and Exposure Sequence (3002,0030) is sent.

Export supported

> Diaphragm
Position

(3002,0034)

Positions of diaphragm jaw pairs (in mm) in IEC BEAM


LIMITING DEVICE coordinate axis in the IEC order X1,
X2, Y1, Y2.

*** export not supported

> Beam Limiting


Device Sequence

(300A,00B6)

Introduces sequence of beam limiting device (collimator)


jaw or leaf (element) positions for given exposure. One or
more items may be included in this sequence.

Export supported

>> RT Beam
Limiting Device
Type

(300A,00B8)

1C

Type of beam limiting device (collimator).


Required if Beam Limiting Device Sequence (300A,00B6)
is sent.
Enumerated Values:
X = symmetric jaw pair in IEC X direction
Y = symmetric jaw pair in IEC Y direction
ASYMX = asymmetric jaw pair in IEC X direction
ASYMY = asymmetric pair in IEC Y direction
MLCX = multileaf (multi-element) jaw pair in IEC X
direction
MLCY = multileaf (multi-element) jaw pair in IEC Y
direction

Export supported for MV Image only


Supported values: X, Y, ASYMX,
ASYMY.

>> Source to Beam


Limiting Device
Distance

(300A,00BA)

Radiation source to beam limiting device (collimator)


distance (mm).

*** export not supported

>> Number of
Leaf/Jaw Pairs

(300A,00BC)

1C

Number of leaf (element) or jaw pairs (equal to 1 for


standard beam limiting device jaws). Required if Beam
Limiting Device Sequence (300A,00B6) is sent.

Export supported for MV Image only


Value set to 1.

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

97 of 118

IOD Details, RT Image C.8.8.2

98 of 118

Attribute Name

Tag

Description

Handling

>> Leaf Position


Boundaries

(300A,00BE)

2C

Boundaries (in mm) of beam limiting device (collimator)


leaves (elements) in IEC BEAM LIMITING DEVICE
coordinate axis appropriate to RT Beam Limiting Device
Type (300A,00B8), i.e. X-axis for MLCY, Y- axis for
MLCX. Contains N+1 values, where N is the Number of
Leaf/Jaw Pairs (300A,00BC), starting from Leaf
(Element) Pair 1. Required if RT Beam Limiting Device
Type (300A,00B8) is MLCX or MLCY.

*** export not supported

>> Leaf/Jaw
Positions

(300A,011C)

1C

Positions of beam limiting device (collimator) leaf or jaw


(element) pairs (in mm) in IEC BEAM LIMITING DEVICE
coordinate axis appropriate to RT Beam Limiting Device
Type (300A,00B8), e.g. X- axis for MLCX, Y-axis for
MLCY). Contains 2N values, where N is the Number of
Leaf/Jaw Pairs (300A,00BC), in IEC leaf (element)
subscript order 101, 102, ... 1N, 201, 202, ... 2N.
Required if Beam Limiting Device Sequence (300A,00B6)
is sent.

Export supported for MV Image only

> Applicator
Sequence

(300A,0107)

Introduces sequence of Applicators associated with


Beam. Only a single item shall be permitted in this
sequence.

*** export not supported

>> Applicator ID

(300A,0108)

1C

User or machine supplied identifier for Applicator.


Required if Applicator Sequence (300A,0107) is sent.

*** export not supported

>> Applicator Type

(300A,0109)

1C

Type of Applicator. Required if Applicator Sequence


(300A,0107) is sent.
Defined Terms:
ELECTRON_SQUARE = square electron applicator
ELECTRON_RECT = rectangular electron applicator
ELECTRON_CIRC = circular electron applicator
ELECTRON_SHORT = short electron applicator
ELECTRON_OPEN = open (dummy) electron applicator
INTRAOPERATIVE = intraoperative (custom) applicator
STEREOTACTIC = stereotactic applicator

*** export not supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

IOD Details, RT Image C.8.8.2

Attribute Name

Tag

Description

Handling

>> Applicator
Description

(300A,010A)

User-defined description for Applicator.

*** export not supported

> Number of Blocks

(300A,00F0)

1C

Number of shielding blocks associated with Beam.


Required if Exposure Sequence (3002,0030) is sent.

Export supported
Value always 0.

> Block Sequence

(300A,00F4)

2C

Introduces sequence of blocks associated with Beam.


Required if Number of Blocks (300A,00F0) is non-zero.
One or more items may be included in this sequence.

*** export not supported

>> Block Tray ID

(300A,00F5)

User-supplied identifier for block tray.

*** export not supported

>> Source to Block


Tray Distance

(300A,00F6)

2C

Radiation Source to attachment edge of block tray


assembly (mm). Required if Block Sequence
(300A,00F4) is sent.

*** export not supported

>> Block Type

(300A,00F8)

1C

Type of block. Required if Block Sequence (300A,00F4)


is sent.
Enumerated Values:
SHIELDING = blocking material is inside contour
APERTURE = blocking material is outside contour

*** export not supported

>> Block
Divergence

(300A,00FA)

2C

Indicates presence or otherwise of geometrical


divergence. Required if Block Sequence (300A,00F4) is
sent.
Enumerated Values:
PRESENT = block edges are shaped for beam
divergence
ABSENT = block edges are not shaped for beam
divergence

*** export not supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

99 of 118

IOD Details, RT Image C.8.8.2

100 of 118

Attribute Name

Tag

Description

Handling

>> Block Mounting


Position

(300A,00FB)

Indicates on which side of the Block Tray the block is


mounted.
Enumerated Values:
PATIENT_SIDE = the block is mounted on the side of the
Block Tray which is towards the patient.
SOURCE_SIDE = the block is mounted on the side of the
Block Tray which is towards the radiation source.

*** export not supported

>> Block Number

(300A,00FC)

1C

Identification Number of the Block. The value of Block


Number (300A,00FC) shall be unique within the Beam in
which it is created. Required if Block Sequence
(300A,00F4) is sent.

*** export not supported

>> Block Name

(300A,00FE)

User-defined name for block.

*** export not supported

>> Material ID

(300A,00E1)

2C

User-supplied identifier for material used to manufacture


Block. Required if Block Sequence (300A,00F4) is sent.

*** export not supported

>> Block Thickness

(300A,0100)

Physical thickness of block (in mm) parallel to radiation


beam axis.

*** export not supported

>> Block Number of


Points

(300A,0104)

2C

Number of (x,y) pairs defining the block edge. Required if


Block Sequence (300A,00F4) is sent.

*** export not supported

>> Block Data

(300A,0106)

2C

A data stream of (x,y) pairs which comprise the block


edge. The number of pairs shall be equal to Block
Number of Points (300A,0104), and the vertices shall be
interpreted as a closed polygon. Coordinates are
projected onto the machine isocentric plane in the IEC
BEAM LIMITING DEVICE coordinate system (mm).
Required if Block Sequence (300A,00F4) is sent.

*** export not supported

> Primary Fluence


Mode Sequence

(3285,XX00)

Sequence defining whether the primary fluence of the


treatment beam uses a non-standard fluence-shaping.
Only a single Item shall be permitted in this sequence.

*** export not supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

IOD Details, RT Image C.8.8.2

Attribute Name

Tag

Description

Handling

>> Fluence Mode

(3285,XX01)

Describes whether the fluence shaping is the standard


mode for the beam or an alternate.
Enumerated Values:
STANDARD = Uses standard fluence-shaping
NON_STANDARD = Uses a non-standard fluenceshaping mode

*** export not supported

>> Fluence Mode ID

(3285,XX02)

1C

Identifier for the specific fluence-shaping mode. Required


if Fluence Mode (3285,xx01) has value
NON_STANDARD.

*** export not supported

Fluence Map
Sequence

(3002,0040)

1C

A Sequence of data describing the fluence map attributes


for a radiotherapy beam. Only one item may be included
in this sequence.
Required if the third value of Image Type (0008,0008) is
FLUENCE.

*** export not supported

> Fluence Data


Source

(3002,0041)

Source of fluence data.


Enumerated Values:
CALCULATED = Calculated by a workstation
MEASURED = Measured by exposure to a film or
detector.

*** export not supported

> Fluence Data


Scale

(3002,0042)

The meterset corresponding with a fluence map cell


value of 1.0 expressed in units specified by Primary
Dosimeter Units (300A,00B3). This is the meterset value
used for treatment, not the meterset used to expose the
film as defined by Meterset Exposure (3002,0032).

*** export not supported

Gantry Angle

(300A,011E)

Treatment machine gantry angle, i.e. orientation of IEC


GANTRY coordinate system with respect to IEC FIXED
REFERENCE coordinate system (degrees).

Export

Beam Limiting
Device Angle

(300A,0120)

Treatment machine beam limiting device (collimator)


angle, i.e. orientation of IEC BEAM LIMITING DEVICE
coordinate system with respect to IEC GANTRY
coordinate system (degrees).

Export supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

101 of 118

IOD Details, RT Image C.8.8.2

102 of 118

Attribute Name

Tag

Description

Handling

Patient Support
Angle

(300A,0122)

Patient Support angle, i.e. orientation of IEC PATIENT


SUPPORT coordinate system with respect to IEC FIXED
REFERENCE coordinate system (degrees).

Export supported

Table Top Eccentric


Axis Distance

(300A,0124)

Distance (positive) from the IEC PATIENT SUPPORT


vertical axis to the IEC TABLE TOP ECCENTRIC vertical
axis (mm).

*** export not supported

Table Top Eccentric


Angle

(300A,0125)

Table Top (non-isocentric) angle, i.e. orientation of IEC


TABLE TOP ECCENTRIC coordinate system with
respect to IEC PATIENT SUPPORT system (degrees).

*** export not supported

Table Top Vertical


Position

(300A,0128)

Table Top Vertical position in IEC TABLE TOP


coordinate system (mm).

Export supported

Table Top
Longitudinal
Position

(300A,0129)

Table Top Longitudinal position in IEC TABLE TOP


coordinate system (mm).

Export supported

Table Top Lateral


Position

(300A,012A)

Table Top Lateral position in IEC TABLE TOP coordinate


system (mm).

Export supported

Isocenter Position

(300A,012C)

Isocenter coordinates (x,y,z), in mm. Specifies the


location of the machine isocenter in the patient-based
coordinate system associated with the Frame of
Reference. It allows transformation from the equipmentbased IEC coordinate system to the patient-based
coordinate system.

Export supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

IOD Details, RT Image C.8.8.2

Attribute Name

Tag

Description

Handling

Patient Position

(0018,5100)

1C

Patient position descriptor relative to the patient support


device.
Required if Isocenter Position (300A,012C) is present.
May be present otherwise. See Section C.7.3.1.1.2 for
Defined Terms and further explanation.
Note:
The orientation of the patient relative to the patient
support device is denoted in the same manner as in the
RT Patient Setup module. It defines the relation of the
patient-based DICOM coordinate system identified by the
frame of reference module of the RT Image to the IEC
coordinate system and together with the Isocenter
Position (300A,012C) allows the RT Image to be placed
into the patient frame of reference. It also allows a
system using an RT Image to verify that the patient is
setup in a similar position relative to the patient support
device.

Export supported

RT Image Isocenter
Position

(3273,XX00)

The isocenter position in the FOR as defined in the FOR


module. For DRRs, this is typically the FOR of the
diagnostic image and the isocenter coordinate
correspond to those of the associated beam. For
treatment / simulator images, the acquisition systems
create the FOR and define it's origin (typically the
isocenter position of the first image acquired).

Export supported

RT Image Patient
Position

(3273,XX01)

The patient position at the time, when the image was


taken (or created in case of DRRs).

Export supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

103 of 118

IOD Details, Structure Set C.8.8.5

Attribute Name

Tag

Description

Handling

LT Archive Primary
Reference Image
Flag

(3279,XX00)

If present indicates if image is primary reference image of


beam referenced by Referenced RT Plan Sequence
(300C,0002) and Referenced Beam Number
(300C,0006). If absent then it is not a primary reference
image.

*** export not supported

Enumerated Values:
PRIMARY
NOT_PRIMARY

B.2.11

104 of 118

Structure Set C.8.8.5


Attribute Name

Tag

Description

Handling

Structure Set Label

(3006,0002)

User-defined label for Structure Set.

Export supported

Structure Set Name

(3006,0004)

User-defined name for Structure Set.

*** export not supported

Structure Set
Description

(3006,0006)

User-defined description for Structure Set.

*** export not supported

Instance Number

(0020,0013)

A number that identifies this object instance.

*** export not supported

Structure Set Date

(3006,0008)

Date at which Structure Set was last modified.

Export supported

Structure Set Time

(3006,0009)

Time at which Structure Set was last modified.

Export supported

Referenced Frame
of Reference
Sequence

(3006,0010)

Introduces sequence of items describing Frames of


Reference in which the ROIs are defined. One or more
items may be included in this sequence. See C.8.8.5.1.

Export supported

> Frame of
Reference UID

(0020,0052)

1C

Uniquely identifies Frame of Reference within Structure


Set. Required if Referenced Frame of Reference
Sequence (3006,0010) is sent.

Export supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

IOD Details, Structure Set C.8.8.5

Attribute Name

Tag

Description

Handling

> Frame of
Reference
Relationship
Sequence

(3006,00C0)

Introduces sequence of transforms that relate other


Frames of Reference to this Frame of Reference.

*** export not supported

>> Related Frame


of Reference UID

(3006,00C2)

1C

Frame of Reference Coordinate System to be


transformed to the current Frame of Reference. Required
if Frame of Reference Relationship Sequence
(3006,00C0) is sent.

*** export not supported

>> Frame of
Reference
Transformation
Type

(3006,00C4)

1C

Type of Transformation. Required if Frame of Reference


Relationship Sequence (3006,00C0) is sent.
Defined Terms:
HOMOGENEOUS

*** export not supported

>> Frame of
Reference
Transformation
Matrix

(3006,00C6)

1C

Four-by-four transformation Matrix from Related Frame of


Reference to current Frame of Reference. Matrix
elements shall be listed in row-major order. Required if
Frame of Reference Relationship Sequence (3006,00C0)
is sent. See C.8.8.5.2.

*** export not supported

>> Frame of
Reference
Transformation
Comment

(3006,00C8)

Comment regarding the transformation between the


related and current Frames of Reference.

*** export not supported

> RT Referenced
Study Sequence

(3006,0012)

Introduces sequence of Studies containing series to be


referenced. One or more items may be included in this
sequence.

Export supported

>> Referenced SOP


Class UID

(0008,1150)

1C

Uniquely identifies the referenced SOP Class. Required if


RT Referenced Study Sequence (3006,0012) is sent.

Export supported

>> Referenced SOP


Instance UID

(0008,1155)

1C

Uniquely identifies the referenced SOP Instance.


Required if RT Referenced Study Sequence (3006,0012)
is sent.

Export supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

105 of 118

IOD Details, Structure Set C.8.8.5

106 of 118

Attribute Name

Tag

Description

Handling

>> RT Referenced
Series Sequence

(3006,0014)

1C

Introduces sequence of items describing series of images


within the referenced study which are used in defining the
Structure Set. Required if RT Referenced Study
Sequence (3006,0012) is sent. One or more items may
be included in this sequence.

Export supported

>>> Series Instance


UID

(0020,000E)

1C

Unique identifier for the series containing the images.


Required if RT Referenced Series Sequence (3006,0014)
is sent.

Export supported

>>> Contour Image


Sequence

(3006,0016)

1C

Introduces sequence of items describing images in a


given series used in defining the Structure Set (typically
CT or MR images). Required if RT Referenced Series
Sequence (3006,0014) is sent. One or more items may
be included in this sequence.

Export supported

>>>> Referenced
SOP Class UID

(0008,1150)

Uniquely identifies the referenced SOP Class.

Export supported

>>>> Referenced
SOP Instance UID

(0008,1155)

Uniquely identifies the referenced SOP Instance.

Export supported

>>>> Referenced
Frame Number

(0008,1160)

Identifies the frame numbers within the Referenced SOP


Instance to which the reference applies. The first frame
shall be denoted as frame number 1.
Note: This Attribute may be multi-valued.
Required if the Referenced SOP Instance is a multi-frame
image and the reference does not apply to all frames.

*** export not supported

Structure Set ROI


Sequence

(3006,0020)

Introduces sequence of ROIs for current Structure Set.


One or more items may be included in this sequence.

Export supported

> ROI Number

(3006,0022)

1C

Identification number of the ROI. The value of ROI


Number (3006,0022) shall be unique within the Structure
Set in which it is created. Required if Structure Set ROI
Sequence (3006,0020) is sent.

Export supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

IOD Details, Structure Set C.8.8.5

Attribute Name

Tag

Description

Handling

> Referenced
Frame of Reference
UID

(3006,0024)

1C

Uniquely identifies Frame of Reference in which ROI is


defined, specified by Frame of Reference UID
(0020,0052) in Referenced Frame of Reference
Sequence (3006,0010). Required if Structure Set ROI
Sequence (3006,0020) is sent.

Export supported

> ROI Name

(3006,0026)

2C

User-defined name for ROI. Required if Structure Set


ROI Sequence (3006,0020) is sent.

Export supported

> ROI Description

(3006,0028)

User-defined description for ROI.

*** export not supported

> ROI Volume

(3006,002C)

Volume of ROI (cubic centimeters).

*** export not supported

> ROI Generation


Algorithm

(3006,0036)

2C

Type of algorithm used to generate ROI.


Required if Structure Set ROI Sequence (3006,0020) is
sent.
Defined Terms:
AUTOMATIC = calculated ROI
SEMIAUTOMATIC = ROI calculated with user assistance
MANUAL = user-entered ROI

Export supported

> ROI Generation


Description

(3006,0038)

User-defined description of technique used to generate


ROI.

*** export not supported

ROI Contour
Sequence

(3006,0039)

Introduces sequence of Contour Sequences defining


ROIs. One or more items may be included in this
sequence.

Export supported

> Referenced ROI


Number

(3006,0084)

Uniquely identifies the referenced ROI described in the


Structure Set ROI Sequence (3006,0020).

Export supported

> ROI Display Color

(3006,002A)

RGB triplet color representation for ROI, specified using


the range 0-255.

Export supported

> Contour
Sequence

(3006,0040)

Introduces sequence of Contours defining ROI. One or


more items may be included in this sequence.

Export supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

107 of 118

IOD Details, Structure Set C.8.8.5

Attribute Name

Tag

Description

Handling

>> Contour
Geometric Type

(3006,0042)

Geometric type of contour.


Enumerated Values:
POINT = single point

Export supported

>> Number of
Contour Points

(3006,0046)

Number of points (triplets) in Contour Data (3006,0050).

Export supported

>> Contour Number

(3006,0048)

Identification number of the contour. The value of


Contour Number (3006,0048) shall be unique within the
Contour Sequence (3006,0040) in which it is defined. No
semantics or ordering shall be inferred from this attribute.

Export supported

>> Contour Data

(3006,0050)

Sequence of (x,y,z) triplets defining a contour in the


patient based coordinate system (mm).

Export supported

Note: Contour Data may not be properly encoded if Explicit-VR


transfer syntax is used and the VL of this attribute exceeds
65534 bytes.

108 of 118

RT ROI
Observations
Sequence

(3006,0080)

Introduces sequence of observations related to ROIs


defined in the ROI Module. One or more items may be
included in this sequence.

Export supported

> Observation
Number

(3006,0082)

Identification number of the Observation. The value of


Observation Number (3006,0082) shall be unique within
the RT ROI Observations Sequence (3006,0080).

Export supported

> Referenced ROI


Number

(3006,0084)

Uniquely identifies the referenced ROI described in the


Structure Set ROI Sequence (3006,0020).

Export supported

> ROI Observation


Label

(3006,0085)

User-defined label for ROI Observation.


Defined Terms:
InitLaserIso
InitMatchIso
AcqIsocenter

Export supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

IOD Details, Structure Set C.8.8.5

Attribute Name

Tag

Description

Handling

> RT ROI
Interpreted Type

(3006,00A4)

Type of ROI.
Defined Terms:
INITLASERISO
INITMATCHISO
ACQ_ISOCENTER

Export supported

> ROI Interpreter

(3006,00A6)

Name of person performing the interpretation.

Export supported

> ROI Physical


Properties
Sequence

(3006,00B0)

Introduces sequence describing physical properties


associated with current ROI interpretation. One or more
items may be included in this sequence.

Export supported

>> ROI Physical


Property

(3006,00B2)

Physical property specified by ROI Physical Property


Value (3006,00B4).
Defined Terms:
PATSUPPORT_ANGLE

Export supported

>> ROI Physical


Property Value

(3006,00B4

User-assigned value for physical property.

Export supported

Referenced
Structure Set
Relationship
Sequence

(3263,XX01)

Introduces sequence of related SOP Class/Instance pairs


describing related instances of structure sets. One or
more items may be included in this sequence.

Export supported

> Referenced SOP


Class UID

(0008,1150)

Uniquely identifies the referenced SOP Class.

Export supported

> Referenced SOP


Instance UID

(0008,1155)

Uniquely identifies the referenced SOP Instance.

Export supported

> Structure Set


Relationship

(3263,XX02)

Relationship of referenced structure set with respect to


current structure set. Required if Referenced Structure
Set Sequence (3263,1002) is sent.
Defined Terms:
PREDECESSOR = structure set used in derivation of
current structure set
ADDITION = structure set, for which the current structure
set is an addition

Export supported ADDITION is used


for new marker structures defined
during marker detection

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

109 of 118

IOD Details, Modality LUT C.11.1

B.2.12

110 of 118

Modality LUT C.11.1


Attribute Name

Tag

Description

Handling

Modality LUT
Sequence

(0028,3000)

1C

Defines a sequence of Modality LUTs. Only one Item


may be present. Shall not be present if Rescale Intercept
(0028,1052) is present.

*** export not supported

> LUT Descriptor


US or

(0028,3002)

1C

Specifies the format of the LUT Data in this Sequence.


See C.11.1.1 for further explanation. Required if the
Modality LUT Sequence (0028,3000) is sent.

*** export not supported

> LUT Explanation

(0028,3003)

Free form text explanation of the meaning of the LUT.

*** export not supported

> Modality LUT


Type

(0028,3004)

1C

Specifies the output values of this Modality LUT. See


C.11.1.1.2 for further explanation. Required if the
Modality LUT Sequence (0028,3000) is sent.

*** export not supported

> LUT Data US or

(0028,3006)

1C

LUT Data in this Sequence. Required if the Modality LUT


Sequence (0028,3000) is sent.

*** export not supported

Rescale Intercept

(0028,1052)

1C

The value b in relationship between stored values (SV)


and the output units specified in Rescale Type
(0028,1054). Output units = m*SV + b. Required if
Modality LUT Sequence (0028,3000) is not present. Shall
not be present otherwise.

Export supported for:


RT Image

Rescale Slope

(0028,1053)

1C

m in the equation specified by Rescale Intercept


(0028,1052). Required if Rescale Intercept is present.

Export supported for:


RT Image

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

IOD Details, Spatial Registration C.20.2

Attribute Name

Tag

Description

Handling

Rescale Type

(0028,1054)

1C

Specifies the output units of Rescale Slope (0028,1053)


and Rescale Intercept (0028,1052). See C.11.1.1.2 for
further explanation. Required if Rescale Intercept is
present.

Export supported for:


RT Image
Defined Terms supported additionally:
CU = Calibration Units
Gy/MU = Gray per MU
RD = Relative Dose
For portal dose images roundtrip is
supported for CU, Gy/MU and RD. All
other values are mapped to US.
For all other images all values are
supported.

B.2.13

Spatial Registration C.20.2


Attribute Name

Tag

Description

Handling

Content Date

(0008,0023)

The date the content creation started.

Export supported

Content Time

(0008,0033)

The time the content creation started.

Export supported

Instance Number

(0020,0013)

A number that identifies this instance

Export supported Value is set to 0.

Content Label

(0070,0080)

A label that is used to identify this registration.

Export supported

Content Description

(0070,0081)

A description of this registration.

Export supported

Content Creators
Name

(0070,0084)

Name of operator performing the registration (such as a


technologist or physician).

*** export not supported

Registration
Sequence

(0070,0308)

A sequence of one or more registration items. Each item


defines a spatial registration to the referenced images in
that item. All referenced images are in the same spatial
frame of reference or atlas.

Export supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

111 of 118

IOD Details, Spatial Registration C.20.2

112 of 118

Attribute Name

Tag

Description

Handling

> Frame of
Reference UID

(0020,0052)

1C

Identifies a Frame of Reference that may or may not be


an image set (e.g. atlas or physical space). See
C.7.4.1.1.1 for further explanation. Required if
Referenced Image Sequence (0008,1140) is absent. May
be present otherwise.

Export supported

> Referenced Image


Sequence

(0008,1140)

1C

Identifies the set of images registered in this sequence


item. One or more items shall be present. Required if
Frame of Reference UID (0020,0052) is absent. May be
present otherwise.

Export supported

>> Referenced SOP


Class UID

(0008,1150)

Uniquely identifies the referenced SOP Class.

Export supported

>> Referenced SOP


Instance UID

(0008,1155)

Uniquely identifies the referenced SOP Instance.

Export supported

>> Referenced
Frame Number

(0008,1160)

Identifies the frame numbers within the Referenced SOP


Instance to which the reference applies. The first frame
shall be denoted as frame number 1.
Note: This Attribute may be multi-valued.
Required if the Referenced SOP Instance is a multi-frame
image and the reference does not apply to all frames.

*** export not supported

> Matrix
Registration
Sequence

(0070,0309)

A sequence that specifies one spatial registration. Exactly


one item shall be present

Export supported

>> Frame of
Reference
Transformation
Comment

(3006,00C8)

User description or comments about the Transformation


Comment registration.

*** export not supported

>> Registration
Type Code
Sequence

(0070,030D)

Describes the information input into the registration


process. Only one item may be present.

Export supported

>>> Code Value

(0008,0100)

1C

See Section 8.1. Required if a sequence item is present.

Export supported

>>> Coding
Scheme Designator

(0008,0102)

1C

See Section 8.2. Required if a sequence item is present.

Export supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

IOD Details, Spatial Registration C.20.2

Attribute Name

Tag

Description

Handling

>>> Coding
Scheme Version

(0008,0103)

1C

See Section 8.2. Required if a sequence item is present


and the value of Coding Scheme Designator (0008,0102)
is not sufficient to identify the Code Value (0008,0100)
unambiguously.

Export supported

>>> Code Meaning

(0008,0104)

1C

See Section 8.3. Required if a sequence item is present.

Export supported

>> Matrix Sequence

(0070,030A)

One or more items shall be present. Each item specifies


a transformation. The item order is significant and
corresponds to matrix multiplication order. See
C.20.2.1.1.

Export supported
Always one item is exported which
incorporates the combination of all
matrices of the original sequence.

>>> Frame of
Reference
Transformation
Matrix

(3006,00C6)

A 4x4 homogeneous transformation matrix that registers


the referenced images to the local RCS. Matrix elements
shall be listed in row-major order. See C.20.2.1.1.

Export supported

>>> Frame of
Reference
Transformation
Matrix Type

(0070,030C)

Type of Frame of Reference Transformation Matrix


(3006,00C6).
Defined terms:
RIGID
RIGID_SCALE
AFFINE
See C.20.2.1.2

Export supported
Value set to RIGID

>> Registration Sub


Type

(3275,XX00)

Identifies the application or application context in which


this registration object has been created.

Export supported

> Used Fiducials


Sequence

(0070,0314)

The fiducials used to determine the Frame of Reference


Transformation Matrix. One or more Items may be
present.

*** export not supported

>> Fiducial UID

(0070,031A)

The UID that identifies the fiducial used as registration


input.

*** export not supported

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

113 of 118

Extended Interface

Appendix C Extended Interface


The following IODs have private Attributes containing an XML data structure for transporting structured non-DICOM data:
RT Plan
CR/CT/MR/SC/RT Image
The following sections specify Attributes containing the XML data and describe the structure of the XML data.

C.1

114 of 118

RT Plan / RT Ion Plan


Node Name

Min
occurs

Max
occurs

Data
Type

Node Description

ExtendedVAPlanInterface

> Beams

>> Beam

Unbounded

>>> ReferencedBeamNumber

int

Beam identifier

>>> BeamExtension

>>>> FieldType

string

TREATMENT or SETUP

>>>> RelatedTreatBeamNumber

int

Relation from the image beams to the treatment and/or setup


beams.

>>>> FieldOrder

int

Order in which the fields are shown

>>>> GantryRtnExtendedStart

int

When true, gantry angle position is in the extended range, which


means it has an over travel (a Gantry can move most typical
400' and not only 360').

>>>> GantryRtnExtendedStop

int

When true, gantry angle position is in the extended range, which


means it has an over travel (a Gantry can move most typical
400' and not only 360').

>>>> MUSubtraction

int

MUSubtraction defines whether some MUs have to be


subtracted from the related treatment beam to take images in the

Additional items provided from Vision to expand the features of


RT Beams.

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

Extended Interface

Node Name

Min
occurs

Max
occurs

Data
Type

Node Description
before phase and/or after phase.

>>>> FieldSetupPhotos

>>>>> FieldSetupPhoto

>>>>>> SetupPhotoNumber

int

Increasing number of setup photos. The value has to start from 1


and up.

>>>>>> SetupPhotoFormat

string

The format of the pixel information. Supported values are 'JPG'


and 'BMP'

>>>>>> SetupPhotoPicture

hex
binary

Pixel information of the photo in the corresponding format.

>>>>>> PhotoModified

string

Photo has been modified.


PhotoModified may have true or false

>>> CustomAddOns

>>>> CustomAddOn

>>>>> CustomAddOnType

string

Three Add On Type to choose from:


BLOCK
COMPENSATOR
TRAY

>>>>> CustomCode

int

Codes for accessory which may be custom made or factory


made (in that case the internal code in equal the custom code.)

>>>>> ReferencedAddOnNumber

int

Reference to the identification number of the Block or


Compensator.

>>>>> TraySpecification

>>>>>> TrayID

string

User-supplied identifier for tray. Max length is 16 characters.

>>>>>> SourceTrayDistance

int

The distance from source to slot. This information is needed to


distinguish the slot

> ToleranceTables

OBI and PortalVision Advanced Imaging

Two setup photos max

In case the CustomAddOn is a tray, it gets reported back


through this node

DICOM Conformance Statement, KC1505D3CS

115 of 118

Extended Interface

116 of 118

Node Name

Min
occurs

Max
occurs

>> ToleranceTable

Unbounded

>>> ReferencedToleranceTableNumber

>>> ToleranceTableExtension

>>>> GantryRtnSetup

>>>> CollRtnSetup

Data
Type

Node Description
Additional items provided from Vision to expand the features of
Tolerance Tables with AutoSetup capabilities

int

Refers to the tolerance table in RT Tolerance Table module

string

Setup attributes may have following type:


Automatic
Manual
Remote

string

>>>> CollXSetup

string

>>>> CollYSetup

string

>>>> PatientSupportAngleSetup

string

>>>> CouchLngSetup

string

>>>> CouchVrtSetup

string

>>>> CouchLatSetup

string

>>>> TableTopEccentricAngleSetup

string

> DoseReferences

>> DoseReference

Unbounded

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

Extended Interface

Node Name

Min
occurs

Max
occurs

Data
Type

Node Description

>>> ReferencedDoseReferenceNumber

int

Reference to the identification number of the Dose Reference


(300A,0012) in the RT Prescription module which is a
representation of our reference point

>>> DoseReferenceExtension

>>>> DailyDoseLimit

decim
al

Unit is [Gy]

>>>> SessionDoseLimit

decim
al

Unit is [Gy]

>>>> Breakpoints

>>>>> Breakpoint

Unbounded

>>>>>> BreakpointDose

decim
al

Unit is [Gy]

>>>>>> BreakpointWarning

string

Text which describes the breakpoint. Max length is 254


characters

Additional items provided from Vision to expand the features of


reference points.

Breakpoint extensions, since DICOM has just one


DeliveryWarningDose without a description this is needed.

Table C-6 RT Plan Extended Interface

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

117 of 118

Extended Interface

(This page is intentionally left blank.)

118 of 118

OBI and PortalVision Advanced Imaging

DICOM Conformance Statement, KC1505D3CS

You might also like