Notebook 11
Notebook 11
KILOVOLTAGE
o Accuracy
o Reproducibility
GRID ALIGNMENT
FOCAL SPOT
Objective:
o To check misalignment of the light field and the x ray field
Equipment:
o Energized radiographic unit
o RMI Collimator Test Tool and the RMI Beam Alignment Test Tool
o Cassette 8 X 10 inch
o The bubble level
The Beam Alignment Test Tool provides a simple test of the x-ray beam alignment.
When used with the Collimator Test Tool, X-ray beam misalignments of 1 and 2%
can be visualized without the need for measuring or calculating.
Objective:
o To check misalignment of the light field and the x ray field
Equipment:
o Energized radiographic unit
o RMI Collimator Test Tool and the RMI Beam Alignment Test Tool
o Cassette 8 X 10 inch
o The bubble level
Evaluate collimator light field and x-ray field congruence plus ensure accurate x-
ray beam alignment
Objective:
o To assess the x-ray beam quality and determine the adequacy of filtration.
Equipment:
o Energized radiographic unit
o Digital dosimeter
o Lead backing sheet
o Aluminium filters
NOTEBOOK 11
For Radiation energies of 140 to 400 kVp, copper is used for specifying the half
value layer of the x-ray beam. 10x10cm copper sheets are used, with thickness
that range from 0.1 to 2.0 mm.
Good film-screen contact across the entire area of the screen is needed for quality
diagnosis. Routine testing of all of the cassettes in the department detects areas
of poor film-screen contact as part of the QA process before blurred areas interfere
with patient care. Such testing is often a regulatory requirement.
The focal spot size of an X-ray tube is a key factor in determining system
resolution. The smaller the focal spot size, the sharper the image at high
geometric magnification. The cylinder provides accurate and reproducible target-
to-image receptor spacing. The visually observed limiting resolution can be simply
converted into an effective focal spot size.
Objective:
Equipment:
o Densitometer
Improves alignment of the radiographic grid and central ray of the x-ray tube and provides
increased image contrast, improved shading in image density, and can reduce
unnecessarily increased patient dosage.
Problems can occur when the x-ray tubes alignment indicators (light/detents) are
misadjusted. Additional problems occur if the x-ray tube is tilted or the grid is installed
upside down. Used to test proper grid alignment with respect to the central ray of the x-
ray tube. Regular quality control testing can help detect misalignments before they
present a patient safety issue.
NOTEBOOK 11
kV Reproducibility
Objective:
o To determine the variation in average kVp over a number of exposures at the
same generator setting.
Equipment:
o Digital kV meter/multifunction meter
o Lead backing sheet
o Energized radiographic unit
Reproducibility
Ability of radiographic unit to produce the same exposure even though in the short
intervals.
mA Linearity
Objective:
o To evaluate the linearity of a generator for commonly used exposure settings
Equipment:
o Energized radiographic unit
o Digital dosimeter
Timer Accuracy
Objective:
o To evaluate the accuracy of an exposure timer.
Equipment:
o Energized radiographic unit
o Digital exposure timer (multi-function meter)