Manual 71-0158RK
Manual 71-0158RK
www.rkiinstruments.com
WARNING
Absorbent cartridges
Filter elements, disks, or sheets
Pump diaphragms and valves
Start Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Start-up Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Performing a Fresh Air Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Turning Off the GX-2009 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Normal Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Display Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Combustible Gas Detection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Aspirator Adapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Data Logging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
GX-2009 Setup Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Calibration Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Using Calibration Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Setting the Date & Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Performing a Fresh Air Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Performing a Span Adjustment in AUTO CAL. . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Performing a Span Adjustment in ONE CAL. . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Performing a Bump Test in BUMP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Discharging the Batteries With the REFRESH Function. . . . . . . . . . . . . . . . . . . 46
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Recharging the GX-2009’s NiMH Batteries. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Replacing the GX-2009’s NiMH Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Checking the Combustible Gas Sensor’s Condition . . . . . . . . . . . . . . . . . . . . . . 62
Replacing the Scrubber Filters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Replacing the Sensor Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Replacing a Sensor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Parts List. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
NOTE: Only the combustible gas detector portion of this instrument has
been assessed for performance.
Top Buzzer
Sound Opening
IR Port
LCD
LED Arrays
Front Buzzer
Sound Opening
RKI GX-2009
Wrist Strap
POW ER
AIR MO DE
H2S CO %LEL O2
Control Buttons
Sensor Retainer
Sensor Cover Retainer
Threaded Inserts
Case
The GX-2009’s sturdy, high-impact plastic case is suitable for use in many
environmental conditions, indoors and out. The unit is dust proof and water
resistant. Rubber overmolded surfaces on much of the case help protect it
from impact damage and aid in holding the instrument in your hand.
A clear plastic window through which the LCD can be viewed is located on
the front of the case. Three brass charging contacts that are used when the
GX-2009 is placed in the charging station are on the back of the case. Four
sensor diffusion ports are located on the bottom of the case. The ports are
labelled as follows from left to right: H2S, CO, HC, and O2. Three frosted
windows, one on the left side, one of the right side, and one on top, cover
alarm LED arrays. There are two buzzer sound openings, one on the front
below the LCD and one on the top behind the top LED array. An IR
(infrared) communication port is located on the top left side which is used
to transfer information when the instrument is used with one of the
calibration or downloading accessories.
There are three threaded inserts on the back of the case. These are used
to mount the alligator clip or the optional belt clip.
The alligator clip can be used to attach the GX-2009 to clothing or a belt.
Teeth in the alligator clip’s jaws prevent the unit from slipping off. The
alligator clip can be rotated to change how the instrument is oriented when
worn. The belt clip is used to easily clip the GX-2009 on a belt.
Sensor Cover
The sensor cover is white and protects the scrubber filters and sensors
from dust and water. It fits on the sensor retainer gasket over all the sensor
diffusion ports and is secured in place by the sensor cover retainer. If the
sensor cover becomes dirty, the response time and accuracy of the GX-
2009 will be affected. Replace the sensor cover if it appears dirty.
Sensors
The GX-2009 uses four sensors to monitor combustible gases, O2, CO,
and H2S simultaneously. The sensors are protected from the environment
by the sensor cover which allows ambient air to diffuse to the sensors. The
sensors use different detection principles, as described below.
LCD
The LCD (liquid crystal display) is visible through the GX-2009’s front
window. Target gas concentrations, the time, battery condition, and alarm
indications are displayed on the LCD. When either of the control buttons
are pressed, the LCD backlight comes on for 30 seconds.
Button Function
NiMH Batteries
Two NiMH (nickel metal hydride) batteries, each with an integral holder,
supply 2.4 volts to power the GX-2009. The batteries will run the unit for up
to 20 hours when no alarms have been activated during that time period.
The batteries are designed to be charged while in the GX-2009 with the
GX-2009 Charging Station. The batteries should not be removed from the
GX-2009 unless they need to be replaced because they will no longer hold
a charge. The battery icon in the lower left corner of the LCD shows the
charge remaining in the batteries. See “Recharging the GX-2009’s NiMH
Batteries” on page 56 for instructions to recharge the batteries and
“Replacing the GX-2009’s NiMH Batteries” on page 60 for instructions to
replace the batteries.
Charging Station
The GX-2009’s NiMH batteries are charged with the GX-2009 Charging
Station. There are two types of charging stations available, the standard
AC powered charging station and the optional vehicle plug DC powered
charging station.
Charging
Contacts
Charge LED
Rear View Top View
AC Adapter
Adapter Jack
AC Adapter
Adapter Jack
Vehicle Plug
DC Adapter
Charging Base Adapter Jack
Charging
Contacts
Charge LED
Rear View Top View
Start-up Procedure
1. Press and briefly hold down the POWER MODE button. The backlight
will turn on and all the display segments will turn on. Release the button
when you hear a beep.
2. The vibrator vibrates and the alarm lights flash momentarily.
3. If Cal. Limit Display is turned on, the screen that appears next
depends on how Cal. Limit Check is set. These parameters and the
Auto Zero Adjustment parameter mentioned in Step 9 below cannot
be set using the GX-2009’s instrument menus, but are set using the
GX-2009 Setup Program. See the GX-2009 Setup Program Operator’s
Manual for information regarding changing various instrument
parameters that are not available for adjustment in the instrument’s
operating modes.
• If the unit is due for calibration and Cal. Limit Check is set to
Confirm to use, then the following screen displays.
CAL
C--LIMIT
The alarm LED’s and buzzer will pulse several times. After this,
press and release the POWER MODE button. You will hear a beep
and startup sequence will continue.
• If the unit is due for calibration and Cal. Limit Check is set to Can’t
use, then the following screen displays.
FAIL
C--LIMIT
The alarm LED’s and buzzer will pulse several times. This screen
will then remain on the display until the unit is turned off. The GX-
2009 cannot be used until a complete calibration has been
performed either by selecting AUTO CAL or ONE CAL in the
Calibration Mode menu. See “Performing a Span Adjustment in
AUTO CAL” on page 32 or “Performing a Span Adjustment in ONE
CAL” on page 37 for calibration instructions.
08 - 7.27
15 d
NEXT CAL
NOTE: The following screen only appears if Bump Test Limit Display is
set to On using the User Setup Program. The standard factory
setting for this function is Off.
4. If Bump Test Limit Display is set to On using the User Setup Program,
the next screen will depend on how Bump Limit Check is set using the
User Setup Program.
• If the unit is due for bump testing and Bump Test Limit Check is set
to Confirm to use, then the following screen displays.
tESt
BUMP NOW
The alarm LED’s and buzzer will pulse several times. After this,
press and release the POWER MODE button. You will hear a beep
and startup sequence will continue.
• If the unit is due for bump testing and Bump Test Limit Check is set
to Can’t use, then the following screen displays.
FAIL
BUMP NOW
The alarm LED’s and buzzer will pulse several times. This screen
will then remain on the display until the unit is turned off. The GX-
2009 cannot be used until a bump test has been performed. See
“Performing a Bump Test in BUMP” on page 41 for bump test
procedures.
• If bump testing is not due or Bump Test Limit Check is set to None,
then the following screen appears for a few seconds indicating when
the next bump test is due.
08 - 7.27
15 d
NEXT BUMP
NOTE: If a successful calibration is performed, the next bump test date is
reset and starts over even though a bump test was not performed.
08 - 7.27
3:51
bAtt.
2.4
AL -- H V
The screen displays the current battery voltage and the alarm pattern.
AL -- H indicates latching alarms and AL -- A indicates self-resetting
alarms. See “Alarm Logic/Resetting Alarms” on page 24 for a
description of latching and self-resetting alarms.
7. The display then indicates the following items for about a second each:
• Full scale values for all channels
• Warning setpoint (low gas alarm) for all channels
• Alarm setpoint (high gas alarm) for all channels
• STEL alarm setpoint for the CO and H2S channels
• TWA alarm setpoint for the CO and H2S channels
F A IL
CO ppm
SENSOR
If you wish to continue, press and release the POWER MODE button to
acknowledge the failure. The gas reading for the failed sensor will be
replaced by “- - -”. Replace the failed sensor as soon as possible.
9. If Auto Zero Adjustment is set to On, then the GX-2009 will perform
an automatic fresh air adjustment.
Auto
ZERO
If the fresh air adjustment is successful, the unit will proceed to Normal
Mode. If one or more of the sensors fails the fresh air adjustment, the
display will indicate which sensor or sensors failed. If a failure occurs,
press and release the POWER MODE button to proceed to Normal
Mode. The gas reading for the failed sensor or sensors will be replaced
by “- - -”. Replace any sensors that fail the automatic fresh air
adjustment as soon as possible.
10. The GX-2009 is now operating in Measuring Mode and monitoring for
gas. The Normal Operation Screen appears.
CH4 %LEL OXY vol%
0
CO ppm
20.9
H2S ppm
0 0.0
5:15
The concentrations of the target gasses are displayed along with the
time in the lower right corner and the battery charge level in the lower
left corner.
Air CAL
HOLD AIR
3. After a few seconds the LCD displays the following screen prompting
you to release the AIR button.
Adj
RELEASE
4. The unit will take a few seconds to adjust the fresh air readings, then
return to the Normal Operation Screen.
Normal Operation
When the GX-2009 comes out of its startup sequence, see “Start Up” on
page 14, it enters Measuring Mode and is in normal operation if no alarm
indications are taking place. It is advisable to perform a fresh air
adjustment when the GX-2009 comes out of its startup sequence as
described in “Start Up”. If Auto Zero Adjustment is set to On, be sure to
follow the guidelines in Step 9 in “Start-up Procedure” on page 14.
In the Normal Operation Screen of Measuring Mode, the battery level, time,
and gas concentrations for each channel are displayed on the LCD. The
time is in the lower right corner and the battery level is in the lower left
corner. Combustible gas is displayed as a percentage of the lower
explosive limit (%LEL), oxygen is displayed as a volume percentage
(vol%), H2S and CO as parts per million (ppm). The combustible channel
gas name is CH4 if it is calibrated to methane (CH4) or HC if it is calibrated
to a another combustible gas. See “Combustible Gas Detection” on
page 21 for a discussion of the combustible channel gas name and
calibration.
CH4 %LEL OXY vol%
0
CO ppm
20.9
H2S ppm
0 0.0
5:15
Display Mode
You can access Display Mode while in Measuring Mode by using the
POWER MODE button. In Display Mode you can view and reset the peak
readings for each target gas, view the TWA and STEL values for CO and
H2S, and view the detection range full scale for each target gas.
STEL is an acronym for short-term exposure limit, and it is the average
reading of the target gas during the last 15 minutes. TWA is an acronym for
time-weighted average, and it is the average reading for the target gas
during the last eight (8) hours. If eight (8) hours has not elapsed since the
unit was turned on, the TWA is still calculated over eight hours, with the
missing time assigned a zero (0) value for the readings. Similarly, if the unit
has not been on for 15 minutes, the missing time is assigned a 0 value and
the STEL is calculated over 15 minutes. The peak readings, STEL values,
0
CO ppm
20.9
H2S ppm
0 0.0
PEAK
The peak readings since the GX-2009 was turned on are displayed for
combustible gas, CO, and H2S. The minimum reading is displayed for
oxygen.
3. If you do not want to clear the peak readings, continue to the next step.
If you want to clear the peak readings, press and hold the AIR button.
After a couple of seconds, the LCD will prompt you to hold the AIR
button with the following screen.
CLEAr
HOLD AIR
Hold the AIR button until the following screen appears, then release it.
CLEAr
RELEASE
The peak readings will be cleared and the unit will return to the PEAK
Screen.
4. Press and release the POWER MODE button again to proceed to the
STEL Screen. The STEL values for CO and H2S are displayed.
5. Press and release the POWER MODE button again to proceed to the
TWA Screen. The TWA values for CO and H2S are displayed.
6. Press and release the POWER MODE button again to proceed to the
Full Scale Screen. The detection range full scale value for each channel
is displayed.
NOTE: If you do not press a button for 20 seconds while in Display Mode,
the GX-2009 will return to Measuring Mode automatically and the
backlight will turn off 30 seconds since the last button was
pressed.
Alarms
This section covers alarm indications. It also describes the two types of
alarm logic, how to reset the GX-2009 after an alarm has occurred, and
how to respond to an alarm condition.
Responding to Alarms
This section describes response to gas, over range, battery, and sensor
failure alarms.
Responding to Gas Alarms
1. Determine which gas alarm has been activated.
2. Follow your established procedure for an increasing gas condition or a
decreasing oxygen condition.
3. If your unit is set for latching alarms, reset the alarm using the POWER
MODE button once the alarm condition has been cleared.
Responding to Over Range Alarms
NOTE: Alarms and the back light feature consume battery power and
reduce the amount of operating time remaining.
Aspirator Adapter
The aspirator adapter accessory is used when it is necessary to draw
sample from an area that cannot be entered or that must be checked for
safety before entry, such as a tank or confined space.
O2
PO W ER
MOD E
O2
10 ft. Hose
F low
GX-2009
%LEL
%LEL
Sample Adapter
Plate
CO
CO
RKI
A IR
H2 S
H2S
Probe
Data Logging
The GX-2009 features the ability to log data to its internal memory and
download it to a computer via the infrared communications port on the top
left side. It logs gas readings during normal operation, alarm data,
calibration data, and bump test data.
To utilize the GX-2009’s downloading capability, you will need the GX-2009
Data Logger Management Program and a computer with an infrared port or
a USB port that runs one of the following operating systems: Windows 7,
Windows 8, or Windows 10. If your computer has an infrared port, then no
additional accessories are needed to download data from the GX-2009. If
your computer does not have an infrared port but does have a USB port, a
USB/IrDA adapter cable can be used to download data from the GX-2009
using the USB port. The GX-2009 Data Logger Management Program is
available at www.rkiinstruments.com/gx2009. The USB/IrDA adapter cable
is also available from RKI.
See the GX-2009 Data Logging Management Program Operator’s Manual
for a complete description of the GX-2009 Data Logging Management
Program and procedures for downloading data to a computer.
NOTE: You can set the GX-2009 to alert you during the startup sequence
when calibration or bump testing is due with the GX-2009 User
Setup Program. See the GX-2009 User Setup Program
Operator’s Manual for information on setting the Cal. Limit
Display and Bump Limit Display parameters.
If the combustible channel passes the above response test and does not
require calibration, the unit should still be calibrated periodically. The
optimum frequency of calibration depends heavily on how the GX-2009 is
used. For example, instruments used daily may need to be calibrated
weekly or monthly, while instruments that are used only a few times a year
may need to be calibrated before each use. Typical calibration frequencies
range from monthly to quarterly. Make sure to perform the combustible
channel response test as described above and make sure to develop a
calibration schedule tailored to your application that takes this test and
required calibration resulting from this test into account.
DATE
3. Use the AIR button to move forward through Calibration Mode. When
you get to the last menu item, the START menu item, continuing will
take you back to the beginning of the menu.
4. When you arrive at the item you wish to enter, press and release the
POWER MODE button to enter that item.
5. When you need to adjust the numerical value of a parameter, increase it
or decrease it, use the AIR button to change the value. When adjusting
a numerical parameter value, it is possible to reverse the direction of
adjustment. To switch from increasing to decreasing a value or
decreasing to increasing a value, do the following:
• with the parameter flashing on the screen, press and hold the AIR
button
• immediately press the POWER MODE button and then release both
buttons
• the direction of adjustment when you press the AIR button is now
reversed
START
7. At the START screen, press and release the POWER MODE button.
The GX-2009 will begin its start-up sequence.
The Calibration Mode menu items are described below in the order in
which they appear while moving through Calibration Mode.
DATE
1. When the DATE Screen is displayed, press and release the POWER
MODE button. A screen appears with the year flashing in the upper left,
the month and day in the upper right, and the time in the lower right.
08 - 8. 7
16:06
DATE
2. Use the AIR button to display the desired year.
3. Press and release the POWER MODE button to save the setting. The
month setting flashes.
4. Repeat steps 2 and 3 to enter the month, day, hour and minute setting.
When you save the minute setting, you return to the DATE Screen.
AIR CAL
Perform a fresh air adjustment in Calibration Mode when you are
performing a calibration before proceeding either to the AUTO CAL or ONE
0
CO ppm
20.9
H2S ppm
0 0.0
PUSH AIR
3. Press and hold the AIR button. The LCD prompts you to continue to
hold the AIR button.
Air CAL
HOLD AIR
4. Release the AIR button when the following screen appears.
AdJ
RELEASE
5. The unit will take a few seconds to adjust the fresh air readings, then
indicate “END” to the right of the battery icon and return to the AIR CAL
Screen.
AUTO CAL
Perform a span adjustment as part of a calibration after performing a fresh
air adjustment. Performing a span adjustment requires the use of a
50
CO ppm
12.0
H2S ppm
50 25.0
AUTO CAL
If the displayed concentrations match the calibration cylinder
concentrations, continue with Step 3.
If any of the displayed concentrations do not match the calibration
cylinder concentrations or if you want to turn off the AUTO CAL function
for a channel or channels, do the following:
• Press and hold the AIR button, then press and momentarily hold the
POWER MODE button.
• Release both buttons as soon as you hear the buzzer sound a beep.
The following screen will display.
CH4 %LEL
- - -
AUTO CAL
• You can scroll through all the channels by pressing and releasing
50
AUTO CAL
• Use the AIR button to adjust the calibration gas value to the desired
value.
• If you want to turn the AUTO CAL function for the selected channel
off so that a span adjustment will not be performed on this channel,
set the calibration gas value to OFF by adjusting the reading down
from the lowest value or up from the highest value.
CH4 %LEL
OFF
AUTO CAL
• Press and release the POWER MODE button to accept the
displayed value. The previous screen will return.
CH4 %LEL
- - -
AUTO CAL
• Repeat the steps above for any other channels that require
calibration gas value updates.
• When you are done updating calibration gas values, use the AIR
button to scroll to the following screen.
ESCAPE
• Press and release the POWER MODE button to return to the screen
that displays the calibration gas values.
CH4 %LEL OXY vol%
50
CO ppm
12.0
H2S ppm
50 25.0
AUTO CAL
Calibration Tubing u b
T
T
in g
Adapter Plate
%LEL
H2S
CO
O2
Flow
4. Confirm that the regulator on/off knob is turned all the way clockwise
(closed) and screw the calibration gas cylinder onto the regulator.
5. Push the adapter plate onto the GX-2009’s sensor face. Make sure the
adapter plate is oriented as shown in Figure 10 below with the sensor
names on the adapter plate matching up with the sensor names on the
instrument.
To Fixed Flow
Regulator
RKI GX-2009
POWER
AIR MODE
H 2S CO %LEL O2
Adapter Plate
%LEL
H2S
CO
O2
Flow
0
CO ppm
20.9
H2S ppm
0 0.0
AUTO CAL
7. Turn the regulator on/off knob counterclockwise to open it. Calibration
gas will begin to flow.
8. Allow the gas to flow for two minutes.
9. Press and release the POWER button.
10. The GX-2009 will attempt to make a span adjustment on all channels.
11. If the span adjustment is successful, the LCD will show the following
screen before returning to the AUTO CAL screen.
PASS
12. If the span adjustment fails, the LCD will show the following screen
indicating which channel or channels failed, the alarm LEDs will flash,
and the buzzer will sound a double pulsing tone.
CH4 %LEL
FAIL
In this example, the CH4 channel failed the span adjustment. Press and
release the POWER MODE button to clear the failure indication and
return to the AUTO CAL screen.
AUTO CAL
See “Troubleshooting” on page 55 to investigate the cause of the failure
and replace the failed sensor or sensors if necessary.
13. Turn the regulator on/off knob clockwise to close it.
14. Remove the calibration adapter plate from the instrument.
15. Remove the regulator from the calibration gas cylinder.
16. Leave the regulator connected to the calibration adapter plate for
convenience.
ONE CAL
Perform a span adjustment as part of a calibration after performing a fresh
air adjustment. Performing a span adjustment requires the use of a
calibration kit. The procedure below describes a span adjustment of one
channel using a calibration kit that includes a 4-gas calibration cylinder, a
0.5 LPM (liters per minute) regulator, a calibration adapter plate, and non-
absorbent sample tubing. The standard 4-gas cylinder consists of 50% LEL
methane, 12% oxygen, 25 ppm H2S, 50 ppm CO, and a balance of
nitrogen. If the H2S channel on your GX-2009 is active, then use a 4-gas
cylinder so that you can calibrate all channels when necessary. If the H2S
channel on your GX-2009 is not active, you may use a 3-gas cylinder. The
standard 3-gas cylinder consists of 50% LEL methane, 12% oxygen, 50
ppm CO, and a balance of nitrogen. In the example below, the CH4
channel’s span is adjusted. See “Parts List” on page 67 for available
calibration kits and calibration kit spare parts.
Calibration cylinders that contain only one target gas may also be used for
single calibration. For channels other than oxygen, RKI Instruments, Inc.
recommends that you use a cylinder that contains a concentration of 10-
50% of the full scale value of the target gas. For example, if you are
calibrating the combustible gas channel, your calibration cylinder should
have a combustible gas concentration between 10% LEL and 50% LEL.
For the oxygen channel, RKI Instruments, Inc. recommends using 100%
nitrogen to perform the adjustment but any concentration between 0%
oxygen (100% nitrogen) and 17% oxygen can be used to make the
adjustment.
- - -
ONE CAL
If the CH4 channel is the one you want to span adjust, skip to Step 4.
If you want to span adjust a different channel, continue with Step 3.
3. Scroll through the channels using the AIR button until the channel you
want to span adjust is displayed. The gas channels are displayed in the
following order along with an escape option: CH4, oxygen, H2S, CO,
ESCAPE.
If you want to abort the span adjustment, scroll through the channels
until you arrive at the screen that indicates “ESCAPE” to the right of the
battery icon, then press and release the POWER MODE button to
return to the ONE CAL screen.
4. Use the sample tubing to connect the calibration adapter plate to the
regulator. Attach the tubing to the adapter plate on the inlet side as
shown below in Figure 11.
To Fixed Flow
Regulator
Calibration Tubing u b
T
T
in g
Adapter Plate
%LEL
H2S
CO
O2
Flow
5. Confirm that the regulator on/off knob is turned all the way clockwise
(closed) and screw the calibration gas cylinder onto the regulator.
To Fixed Flow
Regulator
RKI GX-2009
POWER
AIR MODE
H 2S CO %LEL O2
Adapter Plate
%LEL
H2S
CO
O2
Flow
7. Press and release the POWER MODE button, the current gas reading
for the selected channel is displayed.
CH4 %LEL
ONE CAL
8. Turn the regulator on/off knob counterclockwise to open it. Calibration
gas will begin to flow.
9. Allow the gas to flow for two minutes.
10. Adjust the display gas reading using the AIR button to match the
calibration cylinder’s target gas concentration. See Step 5 on page 30
for instructions to adjust a parameter using the AIR button.
11. Press and release the POWER MODE button.
- - -
ONE CAL
The channel that was just span adjusted will be displayed.
13. If you want to span adjust additional channels, repeat Step 3 through
Step 12 for each channel while gas is still flowing. Make the reading
adjustment right away since gas has been flowing for more than two
minutes. If you need to calibrate more than two channels, use the
AUTO CAL menu item to calibrate the GX-2009.
14. Turn the regulator on/off knob clockwise to close it.
15. When you are finished with span adjustments, scroll to the screen that
indicates “ESCAPE.”
ESCAPE
16. Press and release the POWER MODE button to return to the ONE CAL
Screen.
ONE CAL
17. Remove the calibration adapter plate from the instrument.
18. Remove the regulator from the calibration gas cylinder.
19. Leave the regulator connected to the calibration adapter plate for
convenience.
20. Store the components of the calibration kit in a safe and convenient
place.
NOTE: Bump Test Function must be set to On using the GX-2009 User
Setup Program in order for BUMP to appear in Calibration Mode.
If Bump Test Function is set to Off, BUMP will not appear. See
the GX-2009 User Setup Program Operator’s Manual for
instructions. The factory setting for Bump Test Function is Off.
Entering the BUMP menu item allows you to perform a bump test on all
channels simultaneously to determine if the instrument is responding
properly to gas.
BUMP
Performing a bump test requires the use of a calibration kit. The procedure
below describes a bump test of all four channels using a calibration kit that
includes a 4-gas calibration cylinder, a 0.5 LPM (liters per minute)
regulator, a calibration adapter plate, and non-absorbent sample tubing.
The standard 4-gas cylinder consists of 50% LEL methane, 12% oxygen,
25 ppm H2S, 50 ppm CO, and a balance of nitrogen. If the H2S channel on
your GX-2009 is active, then use a 4-gas cylinder. If the H2S channel on
your GX-2009 is not active, you may use a 3-gas cylinder to perform the
bump test. The standard 3-gas cylinder consists of 50% LEL methane,
12% oxygen, 50 ppm CO, and a balance of nitrogen. See “Parts List” on
page 67 for available calibration kits and calibration kit spare parts.
CO ppm
50 12.0
H2S ppm
50 25.0
BUMP 30
If the values are not correct, you can change them by changing the
AUTO CAL gas values in Calibration Mode or User Setup Mode or by
using the Data Logger Management Program or the User Setup
Program. To exit this screen and return to the main menu, press and
release the AIR button.
3. Use the sample tubing to connect the calibration adapter plate to the
regulator. Attach the tubing to the adapter plate on the inlet side as
shown below in Figure 13.
To Fixed Flow
Regulator
Calibration Tubing u b
T
T
in g
Adapter Plate
%LEL
H2S
CO
O2
Flow
4. Confirm that the regulator on/off knob is turned all the way clockwise
(closed) and screw the calibration gas cylinder onto the regulator.
To Fixed Flow
Regulator
RKI GX-2009
POWER
AIR MODE
H 2S CO %LEL O2
Adapter Plate
%LEL
H2S
CO
O2
Flow
6. Press and release the POWER MODE button. The LCD will display the
current gas readings and the instrument will count down from 30.
CH4 %LEL OXY vol%
0
CO ppm
20.9
H2S ppm
0 0.0
BUMP 30
7. Turn the regulator on/off knob counterclockwise to open it. Calibration
gas will begin to flow. Do not delay opening the regulator.
If you wish to cancel the bump test or a calibration initiated because of a
failed bump test, press and hold the AIR button until you return to the
BUMP screen.
8. When the bump test countdown reaches 0, the unit will determine the
bump test results and proceed as described in either Step 9 or Step 10
depending on the setting of Calibration After Bump Test Failed. See
the User Setup Program Operator’s Manual for a complete description
of this parameter.
P
CO ppm
F
H2S ppm
F F
BUMP
• To view the bump test gas readings press the AIR button.
CH4 %LEL OXY vol%
45
CO ppm
18.9
H2S ppm
10 1.0
BUMP
• To return to the BUMP screen in Calibration Mode, press POWER
MODE. You may also press POWER MODE to return to the BUMP
screen without first viewing the bump test gas readings.
BUMP
See “Troubleshooting” on page 55 to investigate the cause of the
failure and replace the failed sensor or sensors if necessary.
• Turn the regulator on/off knob clockwise to close it.
P
CO ppm
P
H2S ppm
P P
BUMP
• To view the bump test gas readings press the AIR button.
CH4 %LEL OXY vol%
45
CO ppm
11.9
H2S ppm
49 23.0
BUMP
• To return to the BUMP screen in Calibration Mode, press POWER
MODE. You may also press POWER MODE to return to the BUMP
screen without first viewing the bump test gas readings.
BUMP
• Turn the regulator on/off knob clockwise to close it.
If any channel fails the bump test,
• The instrument will beep and a calibration is immediately and
automatically started. Continue to apply the calibration gas.
CH4 %LEL OXY vol%
1
CO ppm
20.9
H2S ppm
4 11.0
CAL 60
• At the end of the calibration, the instrument displays the results from
both the bump test and the calibration.
CH4 %LEL OXY vol%
FP
CO ppm
FP
H2S ppm
PP PP
BUMP/CAL
• To view the bump test gas readings, press the AIR button.
CH4 %LEL OXY vol%
0
CO ppm
20.9
H2S ppm
48 24
BUMP/
52
CO ppm
20.9
H2S ppm
48 0.0
/CAL
• To return to the BUMP screen in Calibration Mode, press POWER
MODE at any time.
BUMP
• Turn the regulator on/off knob clockwise to close it.
11. Remove the calibration adapter plate from the instrument.
12. Remove the regulator from the calibration gas cylinder.
13. Leave the regulator connected to the calibration adapter plate for
convenience.
14. Store the components of the calibration kit in a safe and convenient
place.
REFRESH
If the GX-2009’s batteries are repeatedly only partially discharged before
being charged or only partially charged before being discharged, the
instrument’s runtime may be reduced. The NiMH batteries used in the GX-
2009 are resistant to this effect, but if you notice that the runtime is
decreasing, run the REFRESH function two or three times and see if the
runtime recovers. If the batteries are fully charged, the REFRESH function
will take about three hours to discharge the batteries. If the instrument’s
runtime does not recover, the batteries may need to be replaced.
d IS c h G. d IS c h G.
YES/MODE NO /AIR
2. To run the REFRESH function, press and release the POWER MODE
button. The following screen will be displayed and the fill inside the
battery icon will begin to flash. The GX-2009 is now discharging the
batteries.
d IS c h G.
REFRESH
If you do not want to run the REFRESH function, press and release the
AIR button. The unit will return to the REFRESH screen.
3. If you want to abort the battery discharge at any time, press and release
the AIR button and the unit will return to the REFRESH screen.
4. When the unit is done discharging the batteries, the buzzer will sound a
double pulsing tone, the battery icon will flash, and the following screen
will appear.
TURN OFF
5. Press and hold the POWER MODE button until the unit turns off, then
release it. Recharge the GX-2009’s batteries. See “Recharging the GX-
2009’s NiMH Batteries” on page 56.
NOTE: You can set the GX-2009 to alert you during the startup sequence
when calibration or bump testing is due with the GX-2009 Setup
Program. See the GX-2009 Setup Program Operator’s Manual for
information on setting the Cal. Limit Display and Bump Limit
Display parameter.
The description of the menu items that are common to Calibration Mode
and User Setup Mode will refer you to the appropriate pages in the
Calibration Mode section.
0000
PASSWORD
• Use the AIR button to scroll through the numbers 0-9 until you arrive
at the desired value, then press and release the POWER MODE
button to accept the value.
• Repeat the above procedure for each digit.
• When you have set and accepted the last digit of the password, the
first menu item in User Setup Mode will appear, the DATE Screen.
DATE
• If you entered the incorrect password, an error screen will
appear.You must use the POWER MODE button to turn the unit off
and try again if you wish to enter User Setup Mode or turn on the
unit and proceed to Measuring Mode.
5. If the PASSWORD menu item is set to oFF, the first menu item in User
Setup Mode displays, the DATE Screen.
DATE
6. Use the AIR button to move forward through User Setup Mode. When
you get to the last menu item, the START menu item, continuing will
take you to the beginning of the menu.
7. When you arrive at the item you wish to enter, press and release the
POWER MODE button to enter that item.
8. When you need to adjust the numerical value of a parameter, increase it
or decrease it, use the AIR button to change the value. When adjusting
a numerical parameter value, it is possible to reverse the direction of
START
10. At the START screen, press and release the POWER MODE button.
The GX-2009 will begin its start-up sequence.
The User Setup Mode menu items are described below in the order in
which they appear while moving through User Setup Mode.
ALARM--P
Table 5 below lists the factory set alarm points for each channel.
All of the alarms are increasing alarms, the alarm condition actuates when
the gas level increases above the alarm point, except for the O2 Warning
alarm. The O2 Warning alarm is a decreasing alarm. The O2 Warning
condition actuates when the oxygen level decreases below the alarm point.
For the combustibles channel, it is not possible to adjust either the Warning
or Alarm settings to be higher than 60 %LEL. In addition, the Warning
setting cannot be adjusted higher than the Alarm setting for all channels.
1. At the ALARM--P screen, press and release the POWER MODE button.
The following screen will display showing the combustible channel.
CH4 %LEL
- - -
ALARM--P
2. You can scroll through all the channels by pressing and releasing the
AIR button.
10
WARNING
4. Use the AIR button to adjust the Warning setpoint to the desired value.
5. Press and release the POWER MODE button to save the new Warning
setpoint. A screen with the Alarm setpoint flashing is displayed.
CH4 %LEL
50
ALARM
6. Use the AIR button to adjust the Alarm setpoint to the desired value.
7. Press and release the POWER MODE button to save the new Alarm
setpoint. The display will return to the following screen.
CH4 %LEL
- - -
ALARM--P
NOTE: If you are setting the H2S or CO channel’s alarm points, you will
also see screens for setting the TWA and STEL alarms.
8. Repeat Step 2 through Step 7 for any other channels whose alarm
points you want to change.
9. When you are done, scroll to the following screen.
ESCAPE
10. Press and release the POWER MODE button to return to the
ALARM--P screen.
ALARM--P
PASSWORD
The factory setting for the password feature is oFF.
1. At the PASSWORD screen, press and release the POWER MODE
button. A screen will appear that shows the current password feature
setting.
oFF
PASSWORD
2. Press and release the air button to change the displayed setting. When
the desired setting is displayed, press and release the POWER MODE
button to save the setting.
3. If the password feature was set to oFF, the unit will exit the
PASSWORD menu item and return to the PASSWORD screen.
PASSWORD
With the password feature set to oFF, you will be able to enter User
Setup Mode without entering a password.
4. If the password feature was set to on, the following screen will appear
with the first digit flashing prompting you to enter the password you
want to use.
0000
PASSWORD
5. Use the AIR button to scroll through the numbers 0 - 9.
6. When the desired number is displayed for the first digit, press and
release the POWER MODE button to save the number.
7. The next digit will begin flashing.
PASSWORD
Troubleshooting
The troubleshooting table describes error messages, symptoms, probable
causes, and recommended action for problems you may encounter with
the GX-2009.
• The LCD is • The unit may have been 1. To turn on the unit, press
blank. turned off. and hold the POWER
• The batteries may need MODE button until you
recharging. hear a beep.
2. If the unit does not turn on,
the batteries may need to
be recharged. Recharge
the batteries with the
charging station.
3. If the difficulties continue,
contact RKI Instruments,
Inc. for further instruction.
• The unit shows • The GX-2009 may need 1. Calibrate the unit.
abnormally high to be calibrated. 2. Replace the sensor for the
or low readings • The sensor for the affected channel(s).
but other gas affected channel(s) may 3. If the difficulties continue,
detection need replacement. contact RKI Instruments for
instruments do further instruction.
not.
CAUTION: Use with RKI NiMH battery set 49-1609RK. Charge only with
RKI AC charger p/n 49-2170RK-0X (where X = 1, 2, 3, 4, or 5)
or RKI DC charger 49-2171RK. Use of other rechargeable
batteries or chargers or charging of other rechargeable
batteries in the GX-2009 will void the warranty.
AC Power Supply
To AC
Outlet
Charge LED
DC Adapter
To Vehicle
12 VDC Charge LED
Power Socket
Retaining Tabs
RKI GX-2009
POWER
AIR MODE
H2S CO %LEL O2
5. While the unit is being charged, the red charging LED on the top face of
the charger is on.
6. When the charging LED turns off, the GX-2009’s batteries are fully
charged. It takes approximately 3 hours to recharge a fully discharged
battery set.
7. To verify whether the batteries are fully recharged,
remove the unit from the charging base, then press the
POWER MODE button to start up the unit. If the battery
icon on the LCD is fully colored in, then the unit has been
completely recharged.
NOTE: A battery set will have a typical service life of 500 charging cycles.
3. Grasp the instrument by the front case with your left hand.
4. Grasp the rear case near the top where it is deepest with your right
hand and gently pull the top of the rear case slightly away from the top
of the front case, rotating it about the bottom of the rear case.
6. The batteries will be visible in the front case on the main circuit board.
Once the front case is removed, they are no longer held in place and
can be easily removed.
7. Remove each battery by gently pulling it away from the circuit board.
Each battery has an integral holder that provides electrical connection
to the circuit board.
8. Replace each battery with a new one. Note the polarity markings on the
circuit board and batteries when installing each battery. Tabs on the
bottom of each battery match up with holes in the circuit board.
9. Reinstall the rear case to the front case with the four case screws.
Sensor Cover
Retainer
Sensor Cover
Sensor Retainer
Gasket
RKI GX-2009
POWER
AIR MODE
H2S Removal
H2 S CO %LEL O2 Filter Disk
Charcoal
Filter Disk
Figure 23: Removing the Sensor Retainer and Replacing the Scrubber Filters
Replacing a Sensor
O2
GX-2009
%LEL
%LEL
CO
CO
RKI
AIR
H2S
H2S
Figure 24: Removing the Sensor Retainer to Replace a Sensor
H2S CO HC O2
Line up slots in H2S
and CO sensors
with tabs in case.
H2S
CO
HC
02
Figure 25: Replacing the Sensors and Their Locations in the GX-2009
6. Snap the sensor retainer back onto the case, then secure it to the GX-
2009 with the two screws. The sensor retainer’s tabs that snap onto the
case are keyed so you can only snap it to the case one way.
7. Calibrate the new sensor as described in “Performing a Span
Adjustment in AUTO CAL” on page 32 or “Performing a Span
Adjustment in ONE CAL” on page 37.
81-1050RK Regulator, fixed flow, 0.5 LPM, with gauge and knob, for 17 liter
and 34 liter steel cylinders (cylinders with external threads)
81-1051RK Regulator, fixed flow, 0.5 LPM, with gauge and knob, for 34 liter
aluminum, 58 liter, and 103 liter cylinders (cylinders with internal
threads)
81-1107RK Calibration/sample cap, rubber
81-1160RK Hand aspirated sampler, w/10’ hose, probe
ES-1821 Carbon monoxide sensor
ES-1827i Hydrogen sulfide sensor
NC-6264A Combustibles sensor