Monitoring equipment in a microbiology laboratory is important to ensure proper conditions for samples. Incubators and refrigerators must be monitored daily to ensure temperatures are maintained. Other equipment like autoclaves, microscopes, and biological safety cabinets also require regular checks and maintenance to ensure proper functioning and safety. Logs should be kept to record equipment use and conditions. Any equipment malfunctions could impact sample results so corrections must be made immediately.

Uploaded by

ahmed
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOC, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
84 views3 pages

3 CD Rom Microbiology Equipment Optional

Monitoring equipment in a microbiology laboratory is important to ensure proper conditions for samples. Incubators and refrigerators must be monitored daily to ensure temperatures are maintained. Other equipment like autoclaves, microscopes, and biological safety cabinets also require regular checks and maintenance to ensure proper functioning and safety. Logs should be kept to record equipment use and conditions. Any equipment malfunctions could impact sample results so corrections must be made immediately.

Uploaded by

ahmed
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOC, PDF, TXT or read online on Scribd
You are on page 1/ 3

Content sheet 3-10: Monitoring Equipment Used in the

Microbiology Laboratory

Rationale A short discussion of equipment specifically used in the microbiology laboratory


is covered briefly in this section because working with living organisms requires
special handling.
Many organisms are sensitive to specific temperatures and/or atmospheric
conditions, so incubators, freezers and refrigerators must be monitored to insure
that the desired conditions are maintained. Other equipment used in the
laboratory such as centrifuges, microscopes, autoclaves, pH meters, balances, and
pipettes must be checked for proper functioning.

Incubators The temperatures of incubators and refrigerators should be recorded as soon


and as they are opened in the morning and before ambient temperature can affect
refrigerators them. To facilitate that process tape a log sheet on the door of the instrument and
complete daily. Some thermometers are built into the door so the temperature can
be seen without exposing the inside to room temperature, but their temperatures
should still be recorded. Incubators that are used to enhance atmospheric
conditions such as carbon dioxide (CO2) and anaerobic conditions should also be
checked and results recorded.
If candle jars are used to generate CO2, a growth control plate with a CO2
sensitive organism should be incubated along with plates from patients’ samples
as it is difficult to monitor the exact amount of CO2 in the jar.
If any problems are noted, corrections should be made immediately. If cultures
have been incubated in incubators in which temperature or atmospheric
conditions were not maintained for a significant length of time, primary samples
should be re-plated. If an adequate sample is not still available a new one should
be obtained from the patient. As a rule, all samples should be kept until the report
is finalized.

Autoclaves Staff operating the autoclave must receive training that covers detailed
instructions for use and safety precautions. Autoclaves are used to sterilize
material used in the microbiology laboratory to prevent contamination of samples
being testing and to decontaminate waste material before it is discarded.
When operating an autoclave it is important to be sure the temperature is high
enough to kill any organisms. Chemically treated tape strips with indicators that
change color at a prescribed temperature are often used. A color change shows
that the autoclave achieved the desired temperature. Cards and discs are also
available and demonstrate similar sensitivity to temperature.

When using these indicators, they should be placed inside the batch – not just on
the outside surface.

Biological indicators use glass vials containing specific concentrations of


Equipment● Module 3 ● Optional Content Sheet 1
temperature resistant spores of Bacillus stereothermophilus. A color change in
the vials shows a pH change indicating that the spores were killed in the
autoclave.

Biological indicators are superior to chemical indicators because they not only show
that the temperature has been met, but they also verify that desired time and pressure
were achieved during the sterilization process.

Remember that if the autoclave was overloaded or if the volume was too great,
the material may not be sterile even if the indicators work. Procedure manuals
should describe the maximum load that can be used in each autoclave.

A log book should be kept for each autoclave or other means of sterilization.
Maintain the following information:

 date that the material was sterilized;


 start and end time of sterilization;
 maximum temperature achieved;
 number of trays or baskets included in the run;
 results of indicator;
 signature of user.

Microscopes The microscope is one of the most valuable tools available to laboratory staff, but
is often poorly maintained. Dirty, contaminated, or damaged lenses make reading
and interpretation of material difficult and unreliable. Always keep non-
immersion lenses free of oil.

Establish Kohler illumination so the light source will illuminate the sample for
maximum viewing. Kohler illumination will eliminate the need to move the
condenser up and down when changing lenses, a common misuse of the
microscope.
Always focus on objects using the 10x lens first and then rotate to other lenses as
needed. This single process will prevent scratching of higher powered lenses.
If the microscope is used to measure the size of a substance such as a parasite
egg, it must be equipped with an ocular micrometer. The micrometer should be
calibrated on a regular basis.
All staff should be trained on proper care and use of the microscope using
accepted references.

Refer to the CD-ROM for: Microscope maintenance, Presentation 3, Optional


Slides 1-26.

Equipment● Module 3 ● Optional Content Sheet 2


Biological Biological safety cabinets (BSC) are enclosed work stations intended to protect
safety cabinet both the worker and the biological sample from contamination.

For information, refer to content sheet 3-9.


Refer to the CD-ROM for:

 Biohazards: Selection, Installation and Use of Biological Safety Cabinets (54-


page pdf-format document );
 Module 4 Requirements for Purchasing a Biological Safety Cabinet Class II
example.