HDL - Mindray BS 480

This document provides instructions and guidelines for using an automatic analyzer to measure HDL cholesterol levels using a direct method. It lists the test name and parameters, including reaction type, wavelengths, increase in absorbance over time, reference ranges, and reagent and sample volumes. Calibration is done using a two-point linear method. The reagent is ready-to-use and should be refrigerated between 2-8 degrees Celsius.

Uploaded by

mrashrafi
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
901 views1 page

HDL - Mindray BS 480

This document provides instructions and guidelines for using an automatic analyzer to measure HDL cholesterol levels using a direct method. It lists the test name and parameters, including reaction type, wavelengths, increase in absorbance over time, reference ranges, and reagent and sample volumes. Calibration is done using a two-point linear method. The reagent is ready-to-use and should be refrigerated between 2-8 degrees Celsius.

Uploaded by

mrashrafi
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 1

BaharAndish Co.

Greiner - Guidelines for automatic analyzers – Mindray-BS 480 Edition 2016-10-21

Application: HDL -C
DIRECT METHOD

Preparation: The reagent is ready to use.


Storage: Refrigerate at 2-8°C
Stability: Please see universal package insert sheet

Test Information Judgment Criteria


Test Name HDL-G Absorbance - -
No. ** Incre. Test -
Full Name HDL Decre. Test -
Std. No. ** Lin. Range 1 200
Reaction Parameters Lin. Limit -
Reac. Type Endpoint Subs. Limit -
Pri. Wave. 605nm Reference
Sec. Wave. 700 nm Serum Low High
Direction Increase Male 35 80
Rgt. Blank 48 49 Female 35 80
Reac. Time 80 82 CALIBRATION
Result Setup Rule Two-point linear
Decimal 0 Replicate 2
Unit mg/dl Factor -
Slope 1
Inter. 0 ** User defined

Reagent Volume
R1 (µL) 180
R2 (µL) 60
Sample Volume
Standard (µL) 2.3 - -
Increased (µL) 4.6 - -
Decreased (µL) 2.0 - -

BaharAndish Co. 304 B - Borj Gardoon – Jomhouri Sq. – Tehran – Iran www.baharandish.com

Greiner Diagnostic GmbH - Unter Gereuth 10 – D-79353 Bahlingen – Germany www.greiner-diagnostic.com