Raman Spectroscopy of Carbon Tetrachloride
Raman Spectroscopy of Carbon Tetrachloride
Chem 334
Purpose:
A plot of intensity of scattered light versus energy difference is a Raman spectrum, this is
what is to be studied in this particular lab. The purpose of this lab was to understand and use the
Raman spectrometer and how using it to measure the polarization of molecule is done by
Procedure:
The samples of cyclohexane and carbon tetrachloride were prepared and Since the Raman
Spectrometer had already been turned on it did not need the 5-minute warm up before use. Once
in the program COM4 was selected as the spectrometer port. The sample of cyclohexane was
placed inside the sample holder. At first, 1s for integration time was selected 10 acquired spectra
with polarization turned off, and then continuous was clicked, there was noticeable noise.
Integration time was bumped up to 25s and 10 spectra to acquire was set. The best was chosen.
This was done under the same parameters for the carbon tetrachloride. The integration time was
then set to 40s and 6 spectra to acquire was set. These parameters were done for both non-
polarized and polarized and the best data set was selected for calculations.
Data:
The original spectrum of cyclohexane was taken to get the parameter readings needed for the
Raman.
Using the graph below generated by the NuSpec, the wavenumber for each peak could be found.
These wavenumbers correspond to v1, v2, v3, and v4. Using this graph v1 was found to be 456cm-1
with v2, v3, and v4 being 216, 787, and 311 cm-1 respectively.
wave numbers of each peak are about the same, however the intensity and width of the peak
changes between the two graphs. The polarized spectrum also has a much noisier baseline
Calculations:
k
4 π 2 v 21= (Equation 1)
mcl
k delta
4 π 2 v 22=
( m3 )(
cl l2
) (Equation 2)
k delta
These equations can also be used to determine the values of K and . For this case K was
l2
calculated to be 4.806*10-16 N/m and 3.581*10-16 N/m respectfully. Using these values Equations
4 mCl 8 mCl 2 kδ
(
4 π 2 ( ν 23 + ν 24 ) = 1+
3 mC )( ) (
k
2 mCl
+ 1+
3 mC mCl )( )( )
l2
(Equation 3)
8 mCl kδ
(
16 π 2 ( ν 23 ν 24 ) = 1+
3 mC )( )( )( )
k
mCl
2
mCl l
2 (Equation 4)
The values of v1, v2, v3, and v4 for carbon tetrachloride can also be calculated theoretically using
GaussView. Below is a table with the theoretical wavenumbers for v1 through v4 using the
The numbers between the two methods vary slightly for all of the values of v with the MP2
values all being higher than the H-F method. These theoretical values can also be compared with
the values obtained using the experimental results and the literature values show below.
Error Analysis:
The numbers found from using the peaks in Figure 2, were compared the literature values, the
Hartree-Fock Method, and the MP2 methods calculated frequencies, a percent error was
As seen in Table 4, the method that was the most different from the other values were the MP2
values, they gave not only the largest percent error for the experimental results but were also the
Discussion:
Compared to the literature values, our experimental data was very close with v1 only
being 0.13%, v2 being 0.91%, v3 being 0.50% and v4 being 0.955% off. With all the values being
less than 1% off of the literature values, the experimental values are considered nearly equivalent
to the literature values. This not only showed accuracy but also precision. When comparing the
experimental values with the H-F method values v1 through v4 are 4.58%, 1.82%, 3.43%, and
0.32% different. Once again, the experimental values are within the 5% margin of error, showing
once again the experimental results are accurate. When the values are compared to the MP2
method values the percent difference between the results were 11.5%, 4.8%, 10.57% and 5.2%
off. When these values are compared the experimental data is not statistically accurate due to
being above 5% margin of error. This is not unexpected however when looking at the data on a
whole. All of the values for MP2 are higher than the values calculated using the H-F method. In
turn the experimental values of v2, v3, and v4 for less than the values for v2, v3, and v4 for the H-F
method, with the experimental v1 only being slightly above the value calculated by the H-F
method. When comparing the experimental and H-F methods the increase is slight enough that
the experimental results still fall in the experimental error range of 5%. However, when the
values increase again when compared to the MP2 method, they increase too much for the percent
To conclude the experimental values for v1, v2, v3, and v4 were all within the 5% margin
of error when compared to the provided literature values and the H-F method of theoretical
calculation. Even though the experimental values were not within the 5% margin when compared
to the MP2 method for theoretical calculation, the experimental values are shown to be accurate
results.
References:
Company, 2003)
Hartree-Fock Method:
Figure 1. ν1 Figure 5. V2
Figure 2. V3 Figure 7. V4
MP2 Method:
Figure 8. V1 Figure 9. V2