1001a User Manual of The Software
1001a User Manual of The Software
1. When you use this software for the first time, you should install the driver, once you install it
successfully you don’t need to install it again next time. The procedures are on the below:
driver-setup interface.
a. Connect the Optical Power Meter and The PC with USB cable, and then turn on the
After choose the correct COM port, click the OPEN COM button
Click CONNECT, and then will appear prompt ,this means OPM is
successfully connected to the PC.
3. To process data through this software.
e. After connect the OPM and the PC, click button, you can transmit
3. , Enter relevant information in the blank space, and they will be saved together with data.
4. Time Setting
This OPM has built- in clock to record test time during data storage. The built-in time is set as
standard Beijing time, if you need revise the time, you can enter the corresponding time and
click “TIME SET”. Time setting is finished After noted SETTING SUCESSFUL.
All of our OPMs are calibrated according to Shanghai Institute of Measurement and Testing
Technology. So our OPM have the user-recalibration function. You can recalibrate the OPM
1.When recalibrate the OPM, connect the Light source with known power to the Optical
Power Meter, and fill in this out-power, then click “POWER CALIBRATE”.
out-power of the Optical Light Source and also you should confirm all the connectors and
the test patch cords are clean, otherwise it may cause errors. You also can click “RESET”
button return to the factory setting.
When you want to disconnect the port, only need click “EXIT”, and then unplug the USB cable.