Manual Az8922

1. This digital sound level meter measures sound levels from 30 to 130dB with 0.1dB resolution. It can select between fast and slow response times and A and C weightings. 2. The meter allows for recording maximum sound levels and includes a background noise absorber to measure equipment noise accurately even in high background noise. 3. Key features include an LCD display, automatic or manual ranging selection, AC and DC output jacks, and an RS-232 interface to log data on a PC. It meets ANSI and IEC standards for Type 2 sound level meters.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
203 views18 pages

Manual Az8922

1. This digital sound level meter measures sound levels from 30 to 130dB with 0.1dB resolution. It can select between fast and slow response times and A and C weightings. 2. The meter allows for recording maximum sound levels and includes a background noise absorber to measure equipment noise accurately even in high background noise. 3. Key features include an LCD display, automatic or manual ranging selection, AC and DC output jacks, and an RS-232 interface to log data on a PC. It meets ANSI and IEC standards for Type 2 sound level meters.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 18

Digital Sound Level Meter

Model: 8921
8922
INTRODUCTION
Your digital sound level meter provides
automatic or manual ranging in six
meas-urement ranges from 30 to 130dB.
The unit meets ANSI S1.4 and IEC 651
Type 2 standards, and features 0.1dB
resolution.

A background noise absorber permits


you to measure sound levels accurately
even in the presence of high background
noise.
The meter allows you to select between
fast and slow response times and A and
C weighting. A maximum hold function
is provided.

Jacks on the meter provide both AC and


DC analog output, while an RS-232
interface allows you to use an optional
cable to capture sound level data on a
PC.
FUNCTIONS:
1. Selecting A & C weighting.
2. Selecting the response time.
3. Freezing the maximum sound
level reading.
4. Recording the maximum and
minimum measurements.
5. Using the backgound noise
absorber.
6. Selecting Automatic and manual
ranging.
7. Automatic shut off.
8. Easy to replace battery.
FRONT PANEL DESCRIPTION
1 MICROPHONE
2 LCD DISPLAY
3 ON/OFF - Power button
4 REC - Records sound level
readings
5 MAXHLD - Freezes the maximum
sound level digital
reading
6 C/A - A/C frequency
weighting selector
7 BA MODE Background noise
-
absorber
8 F/S - Fast/Slow response
selector
9 DOWN - Adjusts the
measurement range
10 UPPER - Adjusts the
measurement range
11 DC 9V - DC adapter jack
12 CAL - Calibration screw
13 AC OUT - AC analog output jack
14 DC OUT - DC analog output jack
15 RS232 - RS-232 output jack
16 BACKLIT - Back light
(Model 8922 only)
The meter will display all segments
when it is first turned on. Though you
might seen SEL, Leq, DOSE, PEAK....
These are not available for the meter
you purchased, please contact the
store or the place you purchased for
further models with these additional
functions.
FRONT PANEL DIAGRAM

0
16
MEASURING SOUND LEVELS
Sound levels are displayed both digitally
and in a bar graph. The digital display
is updated every 160ms, while the bar
graph is updated every 40ms.

Press the ON/OFF key to turn the


meter on. The unit will first display
the full screen and then count down
to zero.(See Fig. 1)

90dB
50dB
+2 0 +30 +40
+10 +5
90 0

Fig.1

The meter will now begin measuring


the current sound levels.(See Fig.2)

+2 0 +30 +40
+10 +5
50 0

Fig.2

SPL

Point the microphone toward the source


of the sound to be measured.
SELECTING A AND C WEIGHTING
When you turn the meter on, it will be in
A weighting mode.

A weighting enables the meter to


respond in the same manner as the
human ear, which increases and
decreases amplitude over the
frequency spectrum.

Applications for A weighting include


OSHA regulatory testing, environ-
mental measurement, workplace
design, and law enforcement.

C weighting is suitable for flat response


measurements with no increase or
decrease of amplitude over the
frequency spectrum. Applications for
C weighting include the sound level
analysis of engines and machinery.

Press the C/A key to toggle between A


and C weighting. A small A or C will be
displayed on the right side of the screen
to indicate the current mode.

40 +10

Fig.3
SPL
SELECTING THE RESPONSE TIME
You can select fast or slow response time
to suit different applications and standards.
For example, most OSHA-related testing is
done using slow response time and A
weighting. When you turn the meter on, it
will be in FAST response mode. Press the
F/S key to toggle between FAST and
SLOW response. A small SLOW will be
displayed on the right side of the screen to
indicate the current mode.(See Fig.4)

+10
40

Fig.4
SPL

FREEZING THE MAXIMUM SOUND


LEVEL READING
1 Press the ON/OFF key to turn the
meter on.
2 When measuring sound levels, press
the MAXHLD key to freeze the
maximum reading. MAX HOLD will be
displayed. (See Fig.5) The digital
display will remain unchanged until a
higher reading is detected. Note that
the bar graph will continue to record
the current reading.
3 Press the MAXHLD key again to exit
maximum hold mode.
60

FAST
Fig.5
A

RECORDING THE MAXIMUM AND


MINIMUM MEASUREMENTS
1 Press the ON/OFF key to turn the
meter on.
2 Press the REC key. REC will be
displayed on the bottom of the
screen(See Fig.6) . The meter will
begin tracking the maximum and
minimum sound level
measurements.

+10
40

Fig.6
SPL

3 Press the REC key again. MIN will


appear on the bottom of the screen
(See Fig.7) and the mini. sound level
measurement will be displayed. The
unit isn't recording at this time, but the
bar graph will continue to show the
current reading.
+10
40

Fig.7
SPL
4 Press the REC key again. MAX will
appear on the bottom of the screen
and the maximum sound level
measurement will be displayed.
The unit is not recording at this
time,but the bar graph will continue
to show the current reading.

+10
40

Fig.8
SPL

5 Press the REC key again to resume


recording and repeat the process.

6 Press and hold the REC key until


the REC indicator disappears to
exit recording mode.

USING THE BACKGROUND NOISE


ABSORBER

This feature allows you to measure


equipment noise accurately, even in
the presence of high background
noise.

1 Press the ON/OFF key to turn the


meter on.
2 Press the MAXHLD key. MAX HOLD
will be displayed.(See Fig.5)
3 Press the BA MODE key. F will be
displayed to the left of the SPL
(sound pressure level) icon. The
digital display will show the
background noise level.MAX HOLD
will be disappeared now. (See Fig.9)

+10
60

UNDR FAST
Fig.9
A

4 Press the MAXHLD key again and


MAXHOLD will be displayed. (See
Fig.10) The meter is now ready to
measure the actual machine noise.

+10
60

Fig.10 UNDR FAST

MAX HOLD

5 Turn on the machine you want to


measure and note the new sound
level reading. This number represents
the sound level of the device without
the background noise. If there is no
change in the reading, the
background noise is greater than the
noise of the device.

6 Press the MAXHLD key and then the


BA MODE key to exit background
noise absorber mode.
BACKLIT KEY (8922 only)
Offer a light for approximate 5 seconds
to make it easier to see the display in
the dark.

SELECTING AUTOMATIC AND


MANUAL RANGING

The meter features six measurement


ranges in 10dB steps:
30~80dB, 40~90dB, 50~100dB,
60~110dB, 70~120dB, 80~130dB.

When you turn the meter on, it will be in


automatic range mode and a small AUTO
will be displayed on the left side of the
screen (See Fig.11) . In this mode, the
meter will adjust the measurement range
automatically for accuracy. The two digit
number to the left of the bar graph on
the LCD will show the low end of the
current range.
You can also set the range manually. This
is helpful when you know the measurement
range in advance. The meter will be able to
take readings more quickly, because the
unit does not need to first establish the
range before displaying the measurement.

+ 10 + 10

SPL
SPL

Fig.11 Fig.12
TO ADJUST THE RANGE MANUALLY:
1 When measuring sound levels,press
the DOWN and UPPER keys as
needed to adjust the measurement
range. MANU will appear on the
display (See Fig.12) .Note that the
two digit number to the left of the
bar graph will change to reflect the
low of the newly selected range.

If the meter is operating in manual


range and UNDR is displayed (See
Fig.13) , the sound is too low or the
range. If UPER is displayed (See
Fig.14) ,the sound is too loud.
In either case, you must adjust the
measurement range or your readings
will be incaccurate.

+1 0 + 10

UNDR UPER

SPL
SPL

Fig.13 Fig.14
2 Press and hold the DOWN or UPPER
key to switch back to automatic
ranging.(See Fig.11)
AUTOMATIC SHUTOFF
The meter will turn off automatically
after 20 minutes to preserve the
battery.
To override the Auto Shut Off feature:
1 Make sure the unit is turned off.
2 Press the ON/OFF and MAXHLD
buttons simultaneously.
3 When the full display appears,
release the MAXHLD button first.
"n" will appear on the screen in
one second (See Fig.15) , then
enter the power on mode ; i.e.
count down from:

99.9 88.8 77.7 66.6 55.5


44.4 33.3 22.2 11.1 00.0
and then a current sound level
measurement is started.

Fig.15

4 Release the ON/OFF key. The meter


will remain on until the ON/OFF
button is pressed again.
The automatic shutoff feature will
resume the next time the meter is
turned on.
REPLACING THE BATTERY
When the entire display flashes, the
9V battery has fallen to a critically low
voltage level and should be replaced
as soon as possible. Use a screwdriver
to unscrew the back battery compart-
ment cover. Insert a fresh 9V battery
and replace the cover.
SPECIFICATIONS
Applicable IEC 651 Type 2, ANSI
Standards S1.4 Type 2
Measurement 31.5Hz~8KHz
Frequency Range 1.5dB(Under reference
Accuracy conditions)
Measurement
Level A 30dB~130dB
Weighting
Measurement
Level C 35dB~130dB
Weighting
6 ranges in 10dB steps:
Measurement 30~80dB, 40~90dB,
Level Range 50~100dB, 60~110dB,
70~120dB, 80~130dB
Automatic Range 30~130dB
Time Weighting Fast and slow
Segment Range 50dB
3 1/2 digit LCD, 0.1dB
Digital Display resolution updated
every 0.5 seconds

1dB display steps,


Quasi-Analog 50dB display range,
BarIndicator updated every 50ms
6mm diameter Electret
Microphone condenser microphone
AC: 0.707Vrms (at full
Analog Output
scale), DC: 10mvDC/dB
80mm x 256mm x 38mm
Size (3" x 11" x 1.5")
Weight 240g(10.3oz)
Operating 4~50 oC, 10~90%
Temperature Relative Humidity
Storage
Temperauture -20~60 oC
Battery 9V Battery
RS232 OUTPUT:
The meter can link with personal
computer to capture on-line datas ,
display presure records with real-time
output, you can retrieve file, save the
datas for operating data analysis,
records statistic, multi-files display in
the screen,....versatile functions for
your choice.

Connection procedures:
1. Plug the optional accessory RS232
cable onto the DC jack port ( at the
right side of the meter)
2. Instert the D-sub 9P type connector
onto computer's Com.1 or 2 port or....
3. Start to set up RS232 software by
inserting the CD-ROM or Floppy
diskette.
4. When installing the RS232 software,
please follow the operation manual
procedure in the software package.
MATERIAL SUPPLIED
This standard package contains:
1. The meter x 1
2. Battery x 1 (9.0 volt)
3. Operation manual x 1
4. Buffer x 1
5. Hard Carrying case x 1
Optional accessory:
a) RS232 software CD-R.
b) D-sub connector.
CE CERTIFICATION
The meter conforms to the following
standards:
EN 50081-1/1992 : EN 55022
EN 50082-1/1997 :
(EN 61000-4-2/-3/-8,ENV 50204)
The meter complies with the essential
protection requirements of Council
Directive 89/336/EEC on the approxi-
mation of the laws of the Member
States relating to electromagnetic
compatibility.

RETURN AUTHORIZATION
Authorization must be obtained from
the supplier before returning items for
any reason . When requiring a RA (
Return Authorization) , please include
data regarding the defective reason,
the meters are to be returned along
with good packing to prevent any
damage in shipment and insured
against possible damage or loss .
WARRANTY
The meter is warranted to be free from
defects in material and workmanship for
a period of one year from the date of
purchase. This warranty covers normal
operation and does not cover battery ,
misuse , abuse , alteration , tampering ,
neglect , improper maintenance , or
damage resulting from leaking batteries .
Proof of purchase is required for
warranty repairs . Warranty is void if the
meter has been opened .
Accuracy, the Zenith of
Measuring / Testing Instruments !
Hygrometer
Thermometer
Anemometer
Sound Level Meter
Air Flow meter
Infrared Thermometer
K type Thermometer
K.J.T. type Thermometer
K.J.T.R.S.E. type Thermometer
pH Meter
Conductivity Meter
T.D.S. Meter
D.O. Meter
Saccharimeter
Manometer
Tacho Meter
Lux / Light Meter
Moisture Meter
Data logger
Temp./ RH transmitter
Wireless Transmitter ..........

More products available !


http://www.az-instrument.com.tw

You might also like